| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | MEDTRONIC SYNCHROMED PUMP HYBRID |
| Classification Name | pump, infusion, implanted, programmable |
| Generic Name | implantable infusion pumps |
| Applicant | MEDTRONIC VASCULAR |
| PMA Number | P860004 |
| Supplement Number | S041 |
| Date Received | 11/09/1998 |
| Decision Date | 12/04/1998 |
| Product Code | |
| Advisory Committee |
General Hospital |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - component |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for modifications to the device: 1) replacement of the quartz crystal oscillator with a surface mount crystal oscillator with the same electrical operating specifications, 2) replacement of the ceramic wiring board with a fiberglass printed wiring board, 3)mounting the antenna/reedswitch assembly directly on the wiring board 4) replacement of the ceramic chip carrier packaging of the microprocessor and integrated circuit with plastic packages, 5) attachment of the hybrid printed wiring board to the pump center plate with two sided, acrylic adhesive tape rather than a bracket and plastic cup assembly. |
|
|