| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | PROSTATRON |
| Classification Name | system, hyperthermia, rf/microwave (benign prostatic hyperplasia),thermotherapy |
| Generic Name | transurethral microwave therotherapy (tumt) |
| Applicant | UROLOGIX, INC. |
| PMA Number | P950014 |
| Supplement Number | S024 |
| Date Received | 01/19/2005 |
| Decision Date | 02/23/2005 |
| Product Code | |
| Advisory Committee |
Gastroenterology/Urology |
| Supplement Type | real-time process |
| Supplement Reason | labeling change - minor |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval to change location balloon material from latex to silicone, to change the method of attaching the location balloon due to the new material, and to remove the current latex sensitivity caution and other references to latex in the labeling. |
|
|