| Trade Name | ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM |
| Classification Name | stent, iliac |
| Applicant | ABBOTT VASCULAR INC. |
| PMA Number | P110028 |
| Date Received | 07/25/2011 |
| Decision Date | 02/22/2012 |
| Product Code | |
| Docket Number | 12M-0181 |
| Notice Date | 02/27/2012 |
| Advisory Committee |
Cardiovascular |
| Clinical Trials |
NCT00844532
|
| Expedited Review Granted? | No |
| Combination Product |
No
|
| Information About: |
Labeling, Approval Order, Summary of Safety and Effectiveness |
Approval Order Statement Approval for the absolute pro vascular self-expanding stent system. The device is indicated for improving luminal diameter in patients with de novo or restenotic atherosclerotic lesions in the native common iliac artery and native external iliac artery with reference vessel diameters between 4. 3 mm and 9. 1 mm and lesions lengths up to 90 mm. |
| Approval Order |
Approval Order
|
| Supplements: |
S001 S002 S003 S004 S005 S006 S007 |