| Trade Name | NUCLEUS 22 CHANNEL COCHLEAR IMPLANT |
| Classification Name | implant, cochlear |
| Generic Name | cochlear implant system for children |
| Applicant | COCHLEAR CORP. |
| PMA Number | P890027 |
| Supplement Number | S025 |
| Date Received | 10/04/1994 |
| Decision Date | 02/13/1996 |
| Product Code | |
| Advisory Committee |
Ear Nose & Throat |
| Supplement Type | normal 180 day track |
| Supplement Reason | change design/components/specifications - other |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for component and process changes to the nucleus 22 channel cochlear implant device. The component change consists of replacing the lower value npo dielectric trim capacitor with the same bx dielectric material as used for the higher value trim capacitor. The process change consists of preheating the capacitors and printed circuit board prior to soldering the capacitors to the pcb. |