| Trade Name | ROTABLATOR ROTATIONAL ANGIOPLASTY SYSTEM |
| Classification Name | catheter, coronary, atherectomy |
| Generic Name | rotational angioplasty system guide wires |
| Applicant | BOSTON SCIENTIFIC CORP. |
| PMA Number | P900056 |
| Supplement Number | S035 |
| Date Received | 01/21/2000 |
| Decision Date | 02/11/2000 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | 30-day notice |
| Supplement Reason | process change: manufacturing |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement The 30-day notice requested two changes;the addition of an inspection of the advancer short drive shaft following the wet test and an additional glue application during assembly of the large hypo tube and liner components. |