| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | MEDTRONIC MINIMED CONTINUOUS SUBCUTANEOUS GLUCOSE MONITORING SYSTEM |
| Classification Name | sensor, glucose, invasive |
| Applicant | MEDTRONIC MINIMED |
| PMA Number | P980022 |
| Supplement Number | S128 |
| Date Received | 11/14/2012 |
| Decision Date | 02/11/2013 |
| Product Code | |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - component |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for a change to the packaging of a low dropout (ldo) regulator integrated circuit to halogen free and a change of the solder plating to pure tin solder plating for the minilink real-time transmitter (mmt-7703), cgms ipro recorder (mmt- 7709), and ipro2 recorder (mmt-7741) components of the paradigm real-time, paradigm real-time revel, guardian real-time, cgms ipro, and ipro2 professional glucose monitoring systems. |
|
|