| Trade Name | THE CLOSER 6 FR. SUTURE MEDIATED CLOSURE (SMC) SYSTEM AND THE CLOSER S 6 FR. SUTURE MEDIATED CLOSURE (SMC) SYSTEM |
| Classification Name | device, hemostasis, vascular |
| Generic Name | vascular hemostasis device |
| Applicant | ABBOTT VASCULAR INC. |
| PMA Number | P960043 |
| Supplement Number | S030 |
| Date Received | 12/26/2000 |
| Decision Date | 02/02/2001 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | real-time process |
| Supplement Reason | design change - minor |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for design changes to the multi-lumen structure of the sheath (from 5 lumens to 3 lumens), as well as combining the exit ramp and pebax beading into a single injection molded pebax component to simplify the manufacturing process. These modifications apply to both the closer 6 fr. Smc system and the closer s 6 fr. Smc system. |