| Trade Name | VITROS IMMUNDIAGNOSTICS PRODUCTS ANTI-HBS REAGENT PACK & CALIBRATOR |
| Classification Name | test, hepatitis b (b core, be antigen, be antibody, b core igm) |
| Generic Name | antibody to hepatitis b surface antigen (anti-hbs) assay |
| Applicant | ORTHO-CLINICAL DIAGNOSTICS, INC. |
| PMA Number | P000014 |
| Supplement Number | S019 |
| Date Received | 09/04/2009 |
| Decision Date | 01/11/2010 |
| Product Code | |
| Advisory Committee |
Microbiology |
| Supplement Type | normal 180 day track |
| Supplement Reason | change design/components/specifications - component |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for revision of the intended use to include use of the vitros 5600 integrated system and vitros 3600 immunodiagnostic system with the vitros immunodiagnostic products anti-hbs assay. The device, as modified, will be marketed under the trade names vitros immunodiagnostic products anti-hbs quantitative reagent pack, vitros immunodiagnostic products anti-hbs calibrator, and is indicated for: vitros immunodiagnostic products anti-hbs quantitative reagent pack: for the quantitative in vitro determination of total antibody to hepatitis b surface antigen (anti- hbs) in human serum using the vitros eci/eciq immunodiagnostic systems, the vitros 3600 immunodiagnostic system and the vitros 5600 integrated system. Assay results may be used as an aid in the determination of susceptibility to hepatitis b virus (hbv) infection for individuals prior to or following hbv vaccination, or where vaccination status is unknown. Assay results may be used with other hbv serological markers for the laboratory' diagnosis of hbv disease associated with hbv infection. A reactive assay result will allow a differential diagnosis in individuals displaying signs and symptoms of hepatitis, in whom etiology is unknown. Vitros immunodiagnostic products anti-hbs calibrator: for use in the calibration of the vitros eci/eciq immunodiagnostic systems, the vitros 3600 immunodiagnostic system and the vitros 5600 integrated system for the quantitative in vitro determination of total antibody to hepatitis b surface antigen (anti-hbs) in human serum using the vitros immunodiagnostic products anti-hbs quantitative reagent pack. |
|
|