| Trade Name | THE ANEURX STENT GRAFT WITH XPEDIENT DELIVERY SYSTEM |
| Classification Name | system, endovascular graft, aortic aneurysm treatment |
| Generic Name | transluminal graft prosthesis for the treatment of abdominal aortic aneurysms |
| Applicant | MEDTRONIC AVE, INC. |
| PMA Number | P990020 |
| Supplement Number | S009 |
| Date Received | 06/03/2002 |
| Decision Date | 11/21/2002 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | normal 180 day track |
| Supplement Reason | change design/components/specifications - other |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for the xpedient delivery system. The device, as modified, will be marketed under the trade name aneurx stent graft with expedient delivery system and is indicated for the endovascular treatment of infrarenal abdominal aortic or aorto-iliac aneurysms having: 1) infrarenal, non-aneurysmal, neck length of greater than 1 cm at the proximal and distal ends of the aneurysm and an inner vessel diameter approximately 10-20% smaller than the labeled device diameter: 2) morphology suitable for endovascular repair; 3) one of the following: a) aneurysm diameter > 5 cm; b) aneurysm diameter of 4-5 cm which has also increased in size by 0. 5 cm in the last 6 months; or c) aneurysm which is twice the diameter of the normal infrarenal aorta. |
|
|