Device | AUDALLION BEAMFORMER |
Generic Name | Implant, cochlear |
Applicant | Cochlear Americas 10350 Park Meadows Drive Lone Tree, CO 80124 |
PMA Number | P890027 |
Supplement Number | S046 |
Date Received | 09/21/1998 |
Decision Date | 09/25/1998 |
Product Code |
MCM |
Advisory Committee |
Ear Nose & Throat |
Supplement Type | Real-Time Process |
Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
Expedited Review Granted? | No |
Combination Product | No |
Approval Order Statement Approval of four minor design changes to the Audallion Beamformer Noise Reduction System: 1) switch on/off circuit improvements, 2) replacement of the LC filter with an RC filter, 3) battery spring design change, and 4) a minor software change pertaining to the volume default. With reference to item 1), the resistor value change would improve the switch on/off of the Audiallion. Regarding item 2), the RC filter will improve the rejection of the overall RF interference. For item 3), changing the battery spring design will ensure the proper battery contact position, reducing the likelihood of battery discharge. Item 4) pertains to a software change that will retain the user's volume setting whenever s/he cycles through the available processor settings on the Audallion, enabling the device to be more user friendly. |
|
|