Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device.
 
DeviceTALENT AND BRIO PACEMAKERS
Classification Nameimplantable pacemaker pulse-generator
Generic Nameimplantable pacemaker pulse-generator
Regulation Number870.3610
Applicant
SORIN GROUP- CRM
4 avenue reaumur
clamart cedex 92140
PMA NumberP950029
Supplement NumberS015
Date Received10/31/2001
Decision Date11/26/2001
Product Code
DXY[ Registered Establishments with DXY ]
Advisory Committee Cardiovascular
Supplement Typereal-time process
Supplement Reason change design/components/specifications/material
Expedited Review Granted? No
Combination Product No
Approval Order Statement 
Approval to change the silicone header currently in use with the single chamber talent and brio devices to tecothane 75d polyurethane header (rigid header).
-
-