Device | HAVAB-M EIA |
Generic Name | Hepatitis a test (antibody and igm antibody) |
Regulation Number | 866.3310 |
Applicant | Abbott Laboratories D-49C, BUILDING AP6C 100 ABBOTT PARK ROAD ABBOTT PARK NORTH, IL 60064-3500 |
PMA Number | P790019 |
Supplement Number | S016 |
Date Received | 10/07/2002 |
Decision Date | 11/06/2002 |
Reclassified Date
|
02/09/2006 |
Product Code |
LOL |
Advisory Committee |
Microbiology |
Supplement Type | 30-Day Notice |
Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
Expedited Review Granted? | No |
Combination Product | No |
Approval Order Statement CHANGES TO 1) REPLACE LYOPHILIZED GOAT ANTI-HUMAN IGM PURCHASED FROM LITTON BIONETICS WITH NON-LYOPHILIZED GOAT ANTI-HUMAN IGM PURCHASED FROM JACKSON IMMUNO RESEARCH. 2) REPLACE RHEOPHORESIS TESTING FOR GOAT ANTI-HUMAN IGM (ANTIBODY) IDENTITY/SPECIFICITY WITH IMMUNOELECTRO-PHORESIS (IEP) TESTING. 3) REPLACE HAVAB-M RIA (RADIOIMMUNOASSAY) ANTIBODY BEAD COAT EFFICACY TESTING WITH SMALL-SCALE ANTIBODY DILUTION FACTOR DETERMINATION AND CONFIRMATION TESTING USING HAVAB-M EIA (ENZYME IMMUNOASSAY) TO ESTABLISH THE OPTIMAL DILUTION FACTOR OF THE GOAT ANTI-HUMAN IGM ANTIBODY TO BE USED FOR FULL-SCALE PRODUCTION BEAD COATING. |