Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
|
Device | HIRESOLUTION BIONIC EAR SYSTEM |
Generic Name | Implant, cochlear |
Applicant | ADVANCED BIONICS 28515 WESTINGHOUSE PLACE VALENCIA, CA 91355 |
PMA Number | P960058 |
Supplement Number | S037 |
Date Received | 05/10/2004 |
Decision Date | 07/23/2004 |
Product Code |
MCM |
Advisory Committee |
Ear Nose & Throat |
Supplement Type | Special (Immediate Track) |
Expedited Review Granted? | No |
Combination Product | No |
Recalls | CDRH Recalls |
Approval Order Statement APPROVAL FOR EXPANSION OF THE IN-PROCESS INSPECTION PROCEDURE FOR THE HIRES 90K IMPLANT. NAMELY, THE PMA SUPPLEMENT PROPOSED TO ADD AN ADDITIONAL VISUAL INSPECTION STEP TO ENSURE WELD INTEGRITY BY USING A MINIMUM OF 80X MAGNIFICATION TO VERIFY THE FOLLOWING: 1) COMPLETE CLOSURE OF THE WELD HOLD, 2) ACCEPTABLE WELD QUALITY, AND 3) ABSENCE OF FOREIGN MATERIAL AT THE WELD SITE. |
|
|