• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Premarket Approval (PMA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceAFFINITY/INTEGRITY/VICTORY FAMILY OF PACEMAKERS
Generic NamePulse generator, permanent, implantable
ApplicantAbbott Medical
15900 Valley View Court
Sylmar, CA 91335
PMA NumberP880086
Supplement NumberS142
Date Received04/06/2007
Decision Date05/07/2007
Product Code NVZ 
Advisory Committee Cardiovascular
Supplement TypeReal-Time Process
Supplement Reason Change Design/Components/Specifications/Material
Expedited Review Granted? No
Combination ProductNo
RecallsCDRH Recalls
Approval Order Statement  
APPROVAL FOR MODIFICATIONS TO THE EPIC AND ATLAS FAMILIES OF ICDS AND CRT-DS, INCLUDING: ADDITION OF AN OVER-CURRENT PROTECTION CIRCUIT; ADDED CAPABILITY FOR RAM OPERATION WITH ROM BACKUP WITH DEFIBRILLATION ONLY (DFO) IN THE HARDWARE BACKUP MODE; ADDITION OF DAILY HIGH-VOLTAGE LEAD IMPEDANCE MEASUREMENTS; ADDITION OF LV RING TO RV COIL PACING PULSE CONFIGURATION; ADDITION OF VENTRICULAR INTRINSIC PREFERENCE FEATURE; AND MODIFICATIONS TO PROGRAMMER SOFTWARE AND DISPLAY ENHANCEMENTS. THE DEVICES AS MODIFIED WILL BE MARKETED UNDER THE TRADE NAMES PROMOTE CRT-D DEVICE MODEL NUMBERS 3107-36 AND 3107-30, AND CURRENT DR AND VR ICD DEVICE MODEL NUMBERS 2107-36, 2107-30 AND 1107-36, 1107-30, AND WILL BE SUPPORTED BY THE MODEL 3650 MERLIN PCS WITH MODEL 3330 VERSION 6.0 SOFTWARE.
-
-