• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Premarket Approval (PMA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceEASYTRAK 3 IS-1 LEADS
Generic NameImplantable pulse generator, pacemaker (non-CRT)
ApplicantBoston Scientific Corp.
4100 HAMLINE AVE NORTH
ST. PAUL, MN 55112
PMA NumberP010012
Supplement NumberS274
Date Received12/02/2011
Decision Date03/05/2012
Product Codes LWP OJX 
Advisory Committee Cardiovascular
Supplement TypeReal-Time Process
Supplement Reason Change Design/Components/Specifications/Material
Expedited Review Granted? No
Combination ProductYes
RecallsCDRH Recalls
Approval Order Statement  
APPROVAL FOR THE FOLLOWING CHANGES: 1) ADDITION OF OPTINOVA AS SUPPLIER OF EXTRUDED POLYURETHANE TUBING; 2) MODIFICATION TO DIMENSIONAL TOLERANCES FOR EXTRUDED POLYURETHANE TUBING; 3) MODIFICATION TO POLYURETHANE TUBING SPECIFICATION FOR FOREIGN MATERIAL, BUBBLES, AND GEL INCLUSIONS; AND 4) REMOVAL OF REDUNDANT METHYLENE DIANILINE (MDA) RESIDUAL LEVEL REQUIREMENTS FOR EXTRUDED TUBING,
-
-