| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | FREESTYLE NAVIGATOR CONTIUOUS GLUCOSE MONITORING SYSTEM |
| Classification Name | sensor, glucose, invasive |
| Generic Name | continuous glucose monitor |
| Applicant | ABBOTT LABORATORIES |
| PMA Number | P050020 |
| Supplement Number | S008 |
| Date Received | 11/28/2008 |
| Decision Date | 10/29/2009 |
| Product Code | |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | normal 180 day track |
| Supplement Reason | change design/components/specifications - software |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for a modification to the freestyle navigator continuous glucose monitoring system receiver software. The primary purpose of the new software is to allow users to obtain continuous glucose data as quickly as possible after insertion of the sensor. The new algorithm changes the calibration time to 1, 2, 10, 24 and 72 hours, with glucose data being available at one hour. Other changes are the increase to allowable range for glucose calibration, increase to the allowable glucose rate change and the addition of a calibration lag correction. Additional changes were made to the receiver software/user interface; the backlight will now always be on when the receiver is displaying information, the screen timeout will be 20 seconds and there will be no abandon alarm feature. |
|
|