• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Premarket Approval (PMA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device.
 
Trade NameTHERMA CHOICE UTERINE BALLOON THERAPY SYSTEM
Classification Namedevice, thermal ablation, endometrial
Generic Namethermal ballo0n endometrial ablation
ApplicantGYNECARE, INC.
PMA NumberP970021
Supplement NumberS002
Date Received02/03/1999
Decision Date11/02/1999
Product Code
MNB[ Registered Establishments with MNB ]
Advisory Committee Obstetrics/Gynecology
Supplement Typenormal 180 day track
Supplement Reason change design/components/specifications - component
Expedited Review Granted? No
Combination Product No
Approval Order Statement 
Approval for the following modifications: 1) change in balloon material from latex to silicone; 2) addition of fluid circulation mechanism (i. E. , impeller) within catheter; 3) change in software language from assembly to a combination of c and assembly; and 4) other minor modifications to the device system. In addition, changes also include a change in the manufacturer of the catheter controller to rela, inc. , located at 410 s. Sunset, bldg. , d. , longmont, co 80501. The device is intended for the treatment of menorrhagia (excessive uterine bleeding) due to benign causes in premenopausal women for whom child bearing is complete.
-
-