| Trade Name | CARPENTIER EDWARDS PERIMOUNT PERICARDIAL BIOPROSTHESIS |
| Classification Name | replacement heart-valve |
| Generic Name | bioprosthetic heart valve |
| Regulation Number | 870.3925 |
| Applicant | EDWARDS LIFESCIENCES, LLC. |
| PMA Number | P860057 |
| Supplement Number | S012 |
| Date Received | 10/04/1999 |
| Decision Date | 10/22/1999 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | 30-day notice |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement The 30-day notice proposed a change in the quality control tests used to declare sterility of finished bioprosthetic heart valves. The change will allow for parametric release following terminal liquid steriliztion in accordance with documentation and release criteria for liquid chemical sterilization process. |