| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | PROSORBA COLUMN |
| Classification Name | separator for therapeutic purposes, membrane automated blood cell/plasma |
| Generic Name | therapeutic plasma exchange |
| Applicant | FRESENIUS MEDICAL CARE |
| PMA Number | P850020 |
| Supplement Number | S024 |
| Date Received | 09/18/2006 |
| Decision Date | 10/19/2006 |
| Product Code | |
| Advisory Committee |
Gastroenterology/Urology |
| Supplement Type | Special Supplement |
| Supplement Reason | labeling change - instructions |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for labeling changes to add 1) caution and precaution statements relating to the risk of thrombosis, 2) a clarification regarding treatment volumes for idiopathic thrombocytopenic purpura (itp) patients, 3) a clarification that itp, as well as rheumatoid arthritis, patients may not feel the effects of prosorba treatments immediately after initiation of therapy, and 4) a clarification that treatments should begin within one hour of priming. |
|
|