| Trade Name | MEDTRONIC SIGMA SERIES PACING SYSTEM AND MEDTRONIC 350 SERIES PACING SYSTEM |
| Generic Name | implantable pulse generator programming software |
| Applicant | MEDTRONIC VASCULAR |
| PMA Number | P980035 |
| Supplement Number | S008 |
| Date Received | 09/14/2000 |
| Decision Date | 10/10/2000 |
| Advisory Committee |
Cardiovascular |
| Supplement Type | real-time process |
| Supplement Reason | process change: manufacturing |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for modifications to the sigma and 350 pacing systems, specifically: modify the bottom of the connector by adding a 0. 020 recess on the underside area between the connector and the can; change interconnect ribbon material; modify connector icr area by adding walls between interconnect ribbons; modify icr weld configuration; change material for set screw blocks; modify 3. 2mm connector bores; increase polyurethane thick ness on outside of connector in ultrasonic weld zones; preformed icrs; minimum adhesive fill line added around the perimeter of the fill area. |