| Trade Name | SELENIA FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM |
| Classification Name | full field digital,system,x-ray,mammographic |
| Regulation Number | 892.1715 |
| Applicant | LORAD, A HOLOGIC CO. |
| PMA Number | P010025 |
| Supplement Number | S002 |
| Date Received | 03/28/2003 |
| Decision Date | 09/24/2003 |
| Product Code | |
| Advisory Committee |
Radiology |
| Supplement Type | real-time process |
| Supplement Reason | labeling change - minor |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for the following changes in the selenia qc manual: 1) change the minimum scores for the acr phantom test to 5 fibers, 4 specs, and 4 masses form 4 fibers, 4 specs, and 3 masses. 2) change the signal-to-noise test result minimum from 40 to 35. The 40 value was based on the use of the br-12 phantom, whereas the standard phantom is the act with a minimum value of 35. 3) add a specification of 28 k vp and 65 mas for performance of qc testing. |
|
|