| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | IMAGE CHECKER M1000 SYSTEM |
| Classification Name | analyzer,medical image |
| Generic Name | image analysis system |
| Applicant | R2 TECHNOLOGY, INC. |
| PMA Number | P970058 |
| Supplement Number | S016 |
| Date Received | 05/08/2003 |
| Decision Date | 09/10/2003 |
| Product Code | |
| Advisory Committee |
Radiology |
| Supplement Type | normal 180 day track |
| Supplement Reason | change design/components/specifications - software |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for a new software release (v5. 0) that: 1) provides the user with a choice of two operating points (one with higher sensitivity and increased false marks and the other with lower false marks with decreased sensitivity, chosen at the time of device installation) for use with film/screen images only; 2) when packaged with v3. 1a software in a single image checker housing, allows the processing of both analog and ge ffdm images; 3) allows cad processing of ge ffdm dicom images formatted for presentation; and 4) reduces the number of oversized malignant calcification clusters missed by earlier algorithms. |
|
|