| Trade Name | MOSAIC PORCINE BIOPROSTHESIS |
| Generic Name | replacement heart valve |
| Applicant | THE HEART VALVE DIV. MEDTRONIC CARDIOVASCULAR SURG |
| PMA Number | P990064 |
| Supplement Number | S001 |
| Date Received | 07/31/2000 |
| Decision Date | 08/29/2000 |
| Advisory Committee |
Cardiovascular |
| Supplement Type | 30-day notice |
| Supplement Reason | process change: manufacturing |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Change in the method in which a raw material (aoa compound) is synthesized. This will result in the raw material being received in a more purified form. In addition, the quality control test performed on incoming inspection will change from the gravimetric test method to the ninhydrin test method. |