| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | HIRESOLUTION BIONIC EAR SYSTEM |
| Classification Name | implant, cochlear |
| Generic Name | cochlear implant (pediatric) |
| Applicant | ADVANCED BIONICS CORP. |
| PMA Number | P960058 |
| Supplement Number | S038 |
| Date Received | 07/28/2004 |
| Decision Date | 08/24/2004 |
| Product Code | |
| Advisory Committee |
Ear Nose & Throat |
| Supplement Type | normal 180 day track |
| Supplement Reason | change design/components/specifications - component |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for an interim design change for the auria powerpak accessory to the hiresolution bionic ear system. Specifically, you propose the addition of an insulating washer to the auria powerpak assembly while you develop a final design solution for the recently discovered safety issue associated with this device. We note that you agree to design, test, manufacture, and distribute the revised auria powerpak model with the final design solution within six months of the date of this letter. We further note that you agree to submit a supplement to fda for the proposed final design once the design and test activities have been complete and will not distribute the revised auria powerpak without fda approval. |
|
|