• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Premarket Approval (PMA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceACCESS HYBRITECH P2PSA REAGENTS ON THE ACCESS IMMUNOASSAY SYSTEM
Generic NameP2psa
ApplicantBECKMAN COULTER, INC.
1000 LAKE HAZELTINE DR.
CHASKA, MN 55318
PMA NumberP090026
Supplement NumberS001
Date Received10/25/2012
Decision Date11/19/2012
Product Code OYA 
Advisory Committee Immunology
Supplement TypeSpecial (Immediate Track)
Supplement Reason Labeling Change - Indications/instructions/shelf life/tradename
Expedited Review Granted? No
Combination ProductNo
Approval Order Statement  
APPROVAL FOR THE ADDITION OF INFORMATION TO THE MANUFACTURING PROCEDURES OF THE INSTRUMENT (SPECIFICALLY, THE INSTRUCTIONS FOR APPLICATION OF THE ADHESIVE TO THE MIXER PULLEY ASSEMBLY AND THE IMPLEMENTATION OF A VERIFICATION TEST TO ENSURE THE MIXER PULLEY ASSEMBLIES PASS THE TORQUE SPECIFICATION.)
-
-