Device | Infinity Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) |
Generic Name | Stimulator, electrical, implanted, for parkinsonian tremor |
Applicant | Abbott Medical 6901 Prestion Road Plano, TX 75024 |
PMA Number | P140009 |
Supplement Number | S071 |
Date Received | 06/29/2021 |
Decision Date | 09/27/2021 |
Product Codes |
MHY NHL PJS |
Advisory Committee |
Neurology |
Supplement Type | Real-Time Process |
Supplement Reason | Change Design/Components/Specifications/Material |
Expedited Review Granted? | No |
Combination Product | No |
Approval Order Statement Approval for 1) Adding an existing Abbott qualified supplier for diode components for 3 types of diodes. 2) Replacing tantalum decoupling capacitors with ceramic capacitors sourced from an existing Abbott qualified supplier; and 3) Reducing the number of tantalum output capacitors on the EPPIC charge pump. |