Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
|
Device | Restore, Itrel, Synergy and Intellis Spinal Cord Stimulation Systems and Pisces, Specify, and Vectris Spinal Cord Stimul |
Generic Name | Stimulator, spinal-cord, totally implanted for pain relief |
Applicant | MEDTRONIC NEUROMODULATION 7000 CENTRAL AVENUE NE MINNEAPOLIS, MN 55432 |
PMA Number | P840001 |
Supplement Number | S468 |
Date Received | 08/21/2020 |
Decision Date | 11/03/2020 |
Product Code |
LGW |
Advisory Committee |
Neurology |
Supplement Type | Real-Time Process |
Supplement Reason | Change Design/Components/Specifications/Material |
Expedited Review Granted? | No |
Combination Product | No |
Approval Order Statement Approval to add the use of tin-lead plating for the terminations of both ends of the T8 tantalum capacitors and for the Model 97810 InterStim Micro INS only, change the solder paste stencil tooling for the 1411 capacitor size to increase the amount of solder paste to match the mount used on other capacitor case sizes. |
|
|