DeviceCADENCE FAMILY OF ICDS
Generic NamePulse generator, permanent, implantable
ApplicantAbbott Medical
15900 Valley View Court
Sylmar, CA 91335
PMA NumberP910023
Supplement NumberS162
Date Received10/24/2007
Decision Date12/12/2007
Product Code NVZ 
Advisory Committee Cardiovascular
Supplement TypeReal-Time Process
Supplement Reason Change Design/Components/Specifications/Material
Expedited Review Granted? No
Combination ProductNo
Approval Order Statement  
APPROVAL FOR TWO CORRECTIVE FIXES TO ADDRESS A HARDWARE ISSUED THAT CAN LEAVE THE CRT-D AND ICD SYSTEMS LISTED ABOVE IN A POST-SENSE REFRACTORY STATE. FOR DEVICES THAT ARE ALREADY IMPLANTED AND/OR ALREADY MANUFACTURED, CHANGES WILL BE MADE TO THE PROGRAMMER SOFTWARE. DURING A PATIENT FOLLOW-UP VISIT, THE PROGRAMMER WILL INTERROGATE THE DEVICE AND DOWNLOAD FIRMWARE RAM CODE TO CORRECT THE ANOMALOUS CONDITION IF IT OCCURS. FOR NEW DEVICES, A MODIFICATION HAS BEEN MADE TO THE SIO2 CHIP HARDWARE TO CHECK THE SENSE REFRACTORY STATE EVERY TWO SECONDS AND END IT IF APPROPRIATE.
-
-