Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
|
Device | Evera MRI DF-1 ICD, Evera MRI ICD, Evera S DR ICD, Evera S VR ICD, Evera XT DR ICD, Evera XT VR ICD, Mirro MRI DR ICD, M |
Generic Name | Implantable cardioverter defibrillator (non-CRT) |
Applicant | MEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENT 8200 CORAL SEA STREET N.E. MS MVS11 MOUNDS VIEW, MN 55112 |
PMA Number | P980016 |
Supplement Number | S692 |
Date Received | 10/17/2018 |
Decision Date | 01/09/2019 |
Product Code |
LWS |
Advisory Committee |
Cardiovascular |
Supplement Type | Real-Time Process |
Supplement Reason | Change Design/Components/Specifications/Material |
Expedited Review Granted? | No |
Combination Product | Yes |
Approval Order Statement Approval for a minor design change, corresponding manufacturing documentation updates, and the addition of an in-process 100% visual inspection for the advanced valve metal capacitor component. |
|
|