Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceANEURX AAADVANTAGE STENT GRAFT WITH XCELERANT DELIVERY SYSTEM
Generic NameSYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
ApplicantMedtronic Vascular
3576 UNOCAL PLACE
SANTA ROSA, CA 95403-1774
PMA NumberP990020
Supplement NumberS020
Date Received06/24/2005
Decision Date03/17/2006
Withdrawal Date 12/20/2013
Product Code MIH 
Advisory Committee Cardiovascular
Supplement TypeNormal 180 Day Track
Supplement Reason Change Design/Components/Specifications/Material
Expedited Review Granted? No
Combination ProductNo
Approval Order Statement  
APPROVAL FOR ADDITIONAL ILIAC SIZES (INCLUDING 18MM AND 20MM DIAMETERS; FLARED ILIAC AND ILIAC CUFFS IN 16 MM X 20MM, 18MM X 22MM, AND 18MM X 24MM); A LONGER AORTIC BODY OF THE BIFURCATED COMPONENT BY 1 CM; PRE-SHAPED (CONTOURED) STENT RINGS; AND PLATINUM-IRIDIUM RADIOPAQUE MARKETS. THE DEVICE, AS MODIFIED, WILL BE MARKETED UNDER THE TRADE NAME ANEURX AAADVANTAGE STENT GRAFT WITH XCELERANT DELIVERY SYSTEM AND IS INDICATED FOR THE ENDOVASCULAR TREATMENT OF INFRARENAL ABDOMINAL AORTIC OR AORTO-ILIAC ANEURYSMS HAVING: 1) ADEQUATE ILIAC/FEMORAL ACCESS; 2) INFRARENAL, NON-ANEURYSMAL, NECK LENGTH OF GREATER THAN 1 CM AT THE PROXIMAL AND DISTAL ENDS OF THE ANEURYSM AND AN INNER VESSEL DIAMETER APPROXIMATELY 10-20% SMALLER THAN THE LABELED DEVICE DIAMETER; 3) MORPHOLOGY SUITABLE FOR ENDOVASCULAR REPAIR; 4) ONE OF THE FOLLOWING: ANEURYSM DIAMETER > 5 CM; ANEURYSM DIAMETER OF 4-5 CM WHICH HAS ALSO INCREASED IN SIZE BY 0.5 CM IN THE LAST 6 MONTHS; OR ANEURYSM WHICH IS TWICE THE DIAMETER OF THE NORMAL INFRARENAL AORTA.
-
-