• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Eclipse Release 6.5 through 10 (build 7.3.10 through 10.0.34)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
 Class 2 Recall
Eclipse Release 6.5 through 10 (build 7.3.10 through 10.0.34)
see related information
Date Posted July 11, 2011
Recall Status1 Terminated on August 15, 2014
Recall Number Z-2780-2011
Recall Event ID 58969
Premarket Notification
510(K) Number
K093527 
Product Classification System,Planning,Radiation Therapy Treatment - Product Code MUJ
Product Eclipse Treatment Planning System; Release 6.5 through 10 (build 7.3.10 through 10.0.34) Varian Medical Systems, Inc. Used to plan radiotherapy treatments for patients with Malignant or benign diseases. Plans beam irradiation with photon, electron and proton beams, as well as for internal radiation treatments.
Code Information Model number 48; Serial numbers: H480001 H480062 H480125 H480189 H480253 H480319 H480388 H480453 H480526 H480606 H480668 H480733 H480808 H480002 H480063 H480126 H480190 H480254 H480320 H480389 H480454 H480527 H480607 H480669 H480734 H480809 H480003 H480064 H480127 H480191 H480255 H480321 H480390 H480455 H480528 H480608 H480670 H480735 H480810 H480004 H480065 H480128 H480192 H480256 H480323 H480391 H480456 H480529 H480609 H480672 H480738 H480811 H480005 H480066 H480129 H480193 H480257 H480324 H480392 H480458 H480530 H480610 H480673 H480739 H480812 H480006 H480067 H480130 H480194 H480258 H480325 H480393 H480459 H480531 H480611 H480674 H480740 H480813 H480007 H480068 H480131 H480195 H480259 H480326 H480394 H480460 H480532 H480612 H480675 H480745 H480814 H480008 H480069 H480132 H480196 H480260 H480327 H480395 H480461 H480533 H480613 H480676 H480746 H480815 H480009 H480070 H480133 H480197 H480261 H480328 H480396 H480462 H480534 H480614 H480677 H480747 H480816 H480010 H480072 H480137 H480198 H480262 H480329 H480397 H480463 H480535 H480615 H480678 H480748 H480817 H480011 H480073 H480138 H480199 H480263 H480330 H480398 H480464 H480536 H480616 H480679 H480749 H480819 H480012 H480074 H480139 H480200 H480264 H480331 H480399 H480465 H480537 H480617 H480680 H480750 H480821 H480013 H480075 H480140 H480201 H480265 H480332 H480400 H480466 H480538 H480618 H480682 H480751 H480822 H480014 H480076 H480141 H480202 H480266 H480333 H480403 H480467 H480539 H480619 H480683 H480753 H480823 H480015 H480077 H480142 H480203 H480267 H480334 H480404 H480468 H480541 H480620 H480684 H480755 H480824 H480016 H480078 H480143 H480204 H480268 H480335 H480405 H480469 H480542 H480621 H480685 H480756 H480825 H480017 H480079 H480144 H480205 H480269 H480338 H480406 H480470 H480543 H480622 H480686 H480757 H480826 H480018 H480080 H480145 H480206 H480270 H480339 H480408 H480471 H480544 H480623 H480687 H480758 H480827 H480019 H480081 H480147 H480207 H480271 H480340 H480409 H480473 H480545 H480624 H480688 H480759 H480828 H480020 H480082 H480148 H480208 H480272 H480341 H480410 H480474 H480546 H480625 H480689 H480760 H480829 H480021 H480083 H480149 H480209 H480273 H480342 H480411 H480475 H480547 H480626 H480691 H480761 H480830 H480022 H480084 H480150 H480210 H480274 H480343 H480412 H480476 H480548 H480627 H480692 H480762 H480831 H480023 H480085 H480151 H480212 H480275 H480344 H480413 H480477 H480549 H480628 H480693 H480763 H480832 H480024 H480086 H480152 H480213 H480276 H480345 H480414 H480478 H480566 H480629 H480694 H480764 H480833 H480025 H480087 H480153 H480215 H480277 H480346 H480415 H480481 H480567 H480630 H480695 H480765 H480834 H480026 H480088 H480154 H480216 H480278 H480347 H480416 H480482 H480568 H480631 H480696 H480766 H480836 H480027 H480089 H480155 H480217 H480279 H480348 H480417 H480483 H480569 H480632 H480697 H480767 H480837 H480028 H480090 H480156 H480218 H480280 H480349 H480418 H480484 H480570 H480633 H480698 H480769 H480839 H480030 H480091 H480157 H480219 H480281 H480350 H480419 H480486 H480571 H480634 H480699 H480770 H480840 H480031 H480092 H480158 H480220 H480282 H480351 H480420 H480487 H480572 H480635 H480700 H480772 H480841 H480032 H480093 H480159 H480221 H480283 H480352 H480421 H480488 H480573 H480636 H480701 H480774 H480842 H480033 H480094 H480160 H480222 H480284 H480353 H480422 H480489 H480574 H480637 H480702 H480775 H480843 H480034 H480095 H480161 H480223 H480285 H480354 H480423 H480490 H480575 H480638 H480704 H480776 H480844 H480035 H480096 H480162 H480224 H480286 H480355 H480424 H480491 H480576 H480640 H480705 H480779 H480845 H480036 H480098 H480163 H480225 H480287 H480356 H480425 H480492 H480577 H480641 H480707 H480780 H480846 H480037 H480099 H480164 H480226 H480288 H480357 H480427 H480493 H480578 H480642 H480708 H480781 H480847 H480038 H480100 H480165 H480227 H480289 H480358 H480428 H480494 H480579 H480643 H480709 H480782 H480849 H480039 H480101 H480167 H480228 H480293 H480359 H480429 H480495 H480580 H480644 H480710 H480783 H480850 H480040 H480102 H480168 H480229 H480294 H480360 H480430 H480496 H480581 H480645 H480711 H480785 H480852 H480042 H480103 H480169 H480230 H480295 H480361 H480432 H480497 H480582 H480646 H480712 H480786 H480853 H480043 H480104 H480170 H480233 H480296 H480362 H480433 H480498 H480583 H480647 H480713 H480787 H480854 H480044 H480105 H480171 H480234 H480297 H480365 H480434 H480502 H480585 H480648 H480714 H480788 H480855 H480045 H480106 H480172 H480235 H480298 H480366 H480435 H480503 H480587 H480649 H480715 H480789 H480856 H480046 H480107 H480173 H480236 H480300 H480369 H480436 H480504 H480588 H480651 H480716 H480790 H480857 H480047 H480108 H480174 H480237 H480301 H480370 H480437 H480505 H480589 H480652 H480717 H480791 H480858 H480048 H480109 H480175 H480238 H480302 H480371 H480438 H480507 H480590 H480653 H480718 H480792 H480859 H480049 H480110 H480176 H480239 H480303 H480372 H480439 H480508 H480591 H480654 H480719 H480793 H480860 H480050 H480111 H480177 H480240 H480305 H480373 H480440 H480510 H480592 H480655 H480720 H480794 H480861 H480051 H480112 H480178 H480241 H480306 H480374 H480441 H480511 H480593 H480656 H480722 H480795 H480862 H480052 H480113 H480179 H480242 H480307 H480375 H480443 H480512 H480594 H480657 H480723 H480796 H480863 H480053 H480115 H480180 H480244 H480308 H480376 H480444 H480513 H480595 H480658 H480724 H480797 H480865 H480054 H480116 H480181 H480245 H480311 H480377 H480445 H480514 H480596 H480659 H480725 H480798 H480866 H480055 H480117 H480182 H480246 H480312 H480378 H480446 H480518 H480599 H480660 H480726 H480799 H480867 H480056 H480118 H480183 H480247 H480313 H480379 H480447 H480519 H480600 H480661 H480727 H480802 H480868 H480057 H480119 H480184 H480248 H480314 H480380 H480448 H480521 H480601 H480662 H480728 H480803 H480869 H480058 H480121 H480185 H480249 H480315 H480381 H480449 H480522 H480602 H480663 H480729 H480804 H480870 H480059 H480122 H480186 H480250 H480316 H480382 H480450 H480523 H480603 H480664 H480730 H480805 H480871 H480060 H480123 H480187 H480251 H480317 H480383 H480451 H480524 H480604 H480665 H480731 H480806 H480872 H480061 H480124 H480188 H480252 H480318 H480384 H480452 H480525 H480605 H480667 H480732 H480807 H480873 H480874 H480949 H481018 H481084 H481149 H481229 H481299 H481375 H481449 H481520 H481609 H481677 H481746 H480875 H480950 H481019 H481085 H481150 H481230 H481300 H481376 H481450 H481521 H481610 H481678 H481748 H480876 H480952 H481020 H481086 H481151 H481231 H481301 H481377 H481451 H481523 H481611 H481679 H481749 H480877 H480953 H481021 H481087 H481152 H481232 H481304 H481378 H481452 H481525 H481612 H481680 H481750 H480878 H480954 H481022 H481088 H481153 H481233 H481305 H481379 H481453 H481526 H481613 H481681 H481751 H480879 H480955 H481023 H481089 H481154 H481234 H481306 H481380 H481454 H481527 H481614 H481682 H481752 H480881 H480956 H481024 H481090 H481155 H481235 H481307 H481381 H481455 H481528 H481615 H481683 H481753 H480882 H480957 H481025 H481092 H481156 H481236 H481308 H481382 H481456 H481529 H481616 H481684 H481754 H480883 H480958 H481026 H481093 H481157 H481237 H481309 H481383 H481457 H481530 H481617 H481685 H481755 H480884 H480959 H481027 H481094 H481158 H481239 H481310 H481384 H481458 H481531 H481618 H481686 H481756 H480886 H480960 H481028 H481095 H481159 H481240 H481311 H481389 H481459 H481532 H481619 H481687 H481757 H480887 H480962 H481029 H481096 H481160 H481241 H481312 H481391 H481460 H481533 H481620 H481688 H481758 H480889 H480963 H481030 H481097 H481161 H481248 H481313 H481392 H481461 H481534 H481621 H481689 H481759 H480890 H480964 H481031 H481099 H481163 H481249 H481314 H481393 H481462 H481535 H481622 H481690 H481760 H480892 H480965 H481032 H481100 H481164 H481250 H481315 H481394 H481463 H481536 H481625 H481694 H481761 H480893 H480966 H481033 H481101 H481165 H481251 H481316 H481395 H481464 H481539 H481626 H481695 H481762 H480894 H480967 H481034 H481102 H481166 H481252 H481317 H481397 H481465 H481540 H481627 H481696 H481763 H480896 H480968 H481035 H481103 H481168 H481253 H481318 H481398 H481466 H481542 H481628 H481697 H481764 H480897 H480969 H481036 H481104 H481169 H481254 H481319 H481399 H481467 H481543 H481629 H481698 H481765 H480898 H480970 H481037 H481105 H481171 H481255 H481320 H481401 H481468 H481544 H481630 H481699 H481766 H480899 H480971 H481038 H481106 H481172 H481256 H481321 H481402 H481471 H481545 H481631 H481700 H481767 H480901 H480972 H481039 H481107 H481176 H481257 H481322 H481403 H481472 H481547 H481632 H481702 H481768 H480902 H480973 H481040 H481108 H481177 H481259 H481323 H481404 H481473 H481553 H481633 H481703 H481769 H480903 H480974 H481041 H481109 H481180 H481260 H481326 H481405 H481474 H481554 H481634 H481704 H481770 H480906 H480975 H481043 H481110 H481183 H481261 H481327 H481408 H481476 H481555 H481635 H481705 H481771 H480907 H480976 H481044 H481111 H481184 H481262 H481328 H481409 H481477 H481556 H481636 H481706 H481774 H480908 H480977 H481045 H481112 H481188 H481263 H481329 H481410 H481478 H481557 H481638 H481707 H481777 H480910 H480978 H481046 H481113 H481193 H481264 H481330 H481411 H481479 H481558 H481639 H481708 H481778 H480911 H480980 H481047 H481114 H481194 H481265 H481331 H481412 H481480 H481560 H481640 H481709 H481779 H480912 H480981 H481048 H481115 H481195 H481266 H481332 H481413 H481482 H481561 H481641 H481710 H481780 H480913 H480982 H481049 H481116 H481196 H481267 H481333 H481414 H481483 H481562 H481642 H481711 H481781 H480915 H480983 H481050 H481117 H481197 H481268 H481334 H481415 H481484 H481563 H481644 H481712 H481783 H480916 H480984 H481051 H481118 H481198 H481269 H481335 H481416 H481485 H481564 H481645 H481713 H481784 H480917 H480985 H481052 H481119 H481199 H481270 H481338 H481417 H481486 H481565 H481646 H481715 H481787 H480918 H480986 H481054 H481120 H481200 H481271 H481340 H481418 H481487 H481566 H481647 H481716 H481788 H480919 H480987 H481055 H481122 H481201 H481272 H481341 H481419 H481488 H481567 H481648 H481717 H481789 H480920 H480988 H481056 H481123 H481202 H481273 H481342 H481420 H481489 H481569 H481649 H481718 H481791 H480921 H480989 H481057 H481124 H481203 H481274 H481343 H481423 H481490 H481570 H481650 H481719 H481792 H480923 H480992 H481058 H481125 H481204 H481275 H481344 H481424 H481491 H481581 H481651 H481720 H481793 H480924 H480993 H481059 H481126 H481205 H481276 H481345 H481426 H481492 H481582 H481652 H481722 H481794 H480925 H480994 H481060 H481127 H481206 H481277 H481346 H481427 H481493 H481583 H481653 H481723 H481795 H480926 H480996 H481061 H481128 H481207 H481278 H481347 H481428 H481494 H481585 H481654 H481724 H481796 H480927 H480997 H481062 H481129 H481208 H481279 H481348 H481429 H481495 H481586 H481655 H481725 H481797 H480928 H480998 H481063 H481130 H481209 H481280 H481349 H481430 H481496 H481587 H481656 H481727 H481798 H480929 H480999 H481065 H481131 H481210 H481281 H481351 H481431 H481497 H481588 H481657 H481728 H481799 H480930 H481000 H481066 H481132 H481211 H481282 H481352 H481432 H481498 H481589 H481658 H481729 H481800 H480931 H481002 H481067 H481133 H481212 H481283 H481353 H481433 H481499 H481590 H481659 H481730 H481801 H480932 H481003 H481069 H481134 H481213 H481284 H481354 H481434 H481502 H481591 H481660 H481731 H481802 H480933 H481004 H481070 H481135 H481214 H481285 H481355 H481435 H481503 H481592 H481661 H481732 H481803 H480934 H481005 H481071 H481136 H481215 H481286 H481356 H481436 H481504 H481593 H481662 H481733 H481804 H480935 H481006 H481072 H481137 H481216 H481287 H481357 H481437 H481505 H481594 H481663 H481734 H481805 H480937 H481007 H481073 H481138 H481217 H481288 H481358 H481438 H481506 H481595 H481664 H481735 H481806 H480938 H481008 H481074 H481139 H481218 H481289 H481359 H481439 H481507 H481596 H481665 H481736 H481807 H480939 H481009 H481075 H481140 H481219 H481290 H481360 H481440 H481508 H481597 H481666 H481737 H481808 H480940 H481010 H481076 H481141 H481220 H481291 H481361 H481441 H481509 H481601 H481667 H481738 H481810 H480941 H481011 H481077 H481142 H481221 H481292 H481362 H481442 H481510 H481602 H481670 H481739 H481812 H480942 H481012 H481078 H481143 H481222 H481293 H481363 H481443 H481513 H481603 H481671 H481740 H481813 H480943 H481013 H481079 H481144 H481224 H481294 H481364 H481444 H481514 H481604 H481672 H481741 H481814 H480944 H481014 H481080 H481145 H481225 H481295 H481366 H481445 H481515 H481605 H481673 H481742 H481815 H480945 H481015 H481081 H481146 H481226 H481296 H481367 H481446 H481517 H481606 H481674 H481743 H481816 H480947 H481016 H481082 H481147 H481227 H481297 H481373 H481447 H481518 H481607 H481675 H481744 H481817 H480948 H481017 H481083 H481148 H481228 H481298 H481374 H481448 H481519 H481608 H481676 H481745 H481818 H481819 H481890 H481962 H482024 H482086 H482157 H482226 H482297 H482361 H482445 H482524 H482599 H482687 H481820 H481891 H481963 H482025 H482087 H482158 H482228 H482300 H482362 H482446 H482527 H482601 H482688 H481821 H481892 H481964 H482026 H482088 H482159 H482229 H482301 H482367 H482447 H482528 H482603 H482689 H481822 H481893 H481965 H482027 H482089 H482160 H482230 H482302 H482369 H482448 H482529 H482604 H482690 H481823 H481894 H481966 H482028 H482090 H482161 H482231 H482303 H482374 H482449 H482530 H482605 H482691 H481824 H481896 H481967 H482029 H482091 H482162 H482232 H482304 H482380 H482450 H482531 H482608 H482692 H481825 H481897 H481968 H482030 H482092 H482163 H482233 H482305 H482381 H482451 H482534 H482609 H482693 H481826 H481898 H481969 H482031 H482093 H482164 H482234 H482306 H482382 H482452 H482538 H482610 H482694 H481828 H481899 H481970 H482032 H482094 H482165 H482235 H482307 H482383 H482456 H482539 H482611 H482697 H481829 H481900 H481971 H482033 H482095 H482166 H482236 H482308 H482384 H482457 H482540 H482612 H482698 H481830 H481901 H481972 H482034 H482096 H482167 H482237 H482309 H482385 H482460 H482541 H482613 H482699 H481831 H481902 H481973 H482035 H482097 H482168 H482238 H482310 H482386 H482461 H482542 H482619 H482700 H481832 H481903 H481974 H482036 H482098 H482169 H482239 H482311 H482387 H482462 H482543 H482620 H482701 H481833 H481904 H481975 H482037 H482099 H482170 H482243 H482312 H482388 H482463 H482544 H482621 H482702 H481834 H481905 H481976 H482038 H482100 H482171 H482244 H482313 H482389 H482464 H482545 H482624 H482703 H481835 H481906 H481977 H482039 H482101 H482172 H482245 H482314 H482390 H482465 H482546 H482625 H482704 H481836 H481907 H481978 H482040 H482102 H482173 H482246 H482315 H482391 H482466 H482547 H482626 H482705 H481837 H481908 H481979 H482041 H482103 H482174 H482247 H482316 H482392 H482467 H482548 H482627 H482706 H481838 H481909 H481980 H482042 H482104 H482178 H482248 H482317 H482395 H482470 H482549 H482628 H482709 H481839 H481910 H481981 H482043 H482105 H482179 H482249 H482318 H482396 H482473 H482550 H482629 H482710 H481840 H481911 H481982 H482044 H482106 H482181 H482250 H482319 H482397 H482474 H482551 H482630 H482711 H481841 H481912 H481983 H482045 H482107 H482182 H482251 H482320 H482398 H482475 H482552 H482631 H482712 H481842 H481913 H481984 H482046 H482108 H482183 H482252 H482321 H482399 H482476 H482555 H482633 H482713 H481843 H481914 H481985 H482047 H482109 H482184 H482253 H482322 H482401 H482478 H482556 H482634 H482714 H481846 H481915 H481986 H482048 H482110 H482185 H482254 H482323 H482402 H482479 H482557 H482635 H482715 H481847 H481916 H481987 H482049 H482111 H482186 H482256 H482324 H482403 H482480 H482558 H482636 H482716 H481848 H481918 H481988 H482050 H482112 H482187 H482258 H482325 H482406 H482481 H482559 H482637 H482717 H481849 H481919 H481989 H482051 H482113 H482188 H482259 H482326 H482407 H482484 H482560 H482638 H482718 H481850 H481920 H481990 H482052 H482114 H482189 H482260 H482327 H482408 H482485 H482561 H482641 H482719 H481851 H481921 H481991 H482053 H482115 H482190 H482261 H482328 H482409 H482486 H482562 H482645 H482720 H481852 H481922 H481992 H482054 H482120 H482191 H482262 H482329 H482410 H482487 H482563 H482646 H482722 H481853 H481923 H481993 H482055 H482121 H482192 H482263 H482330 H482411 H482488 H482564 H482647 H482723 H481854 H481924 H481994 H482056 H482122 H482193 H482264 H482331 H482412 H482489 H482566 H482648 H482724 H481855 H481925 H481995 H482057 H482124 H482194 H482265 H482332 H482413 H482490 H482567 H482649 H482725 H481856 H481926 H481996 H482058 H482125 H482195 H482266 H482333 H482414 H482491 H482568 H482650 H482726 H481857 H481927 H481997 H482059 H482126 H482196 H482267 H482334 H482415 H482492 H482569 H482651 H482727 H481858 H481928 H481998 H482060 H482127 H482198 H482268 H482335 H482416 H482494 H482570 H482652 H482728 H481859 H481929 H481999 H482061 H482128 H482199 H482269 H482336 H482417 H482497 H482571 H482653 H482730 H481861 H481930 H482000 H482062 H482129 H482200 H482270 H482337 H482418 H482499 H482572 H482654 H482731 H481862 H481931 H482001 H482063 H482130 H482201 H482271 H482338 H482419 H482500 H482573 H482656 H482732 H481863 H481932 H482002 H482064 H482131 H482202 H482272 H482339 H482420 H482501 H482576 H482657 H482733 H481864 H481933 H482003 H482065 H482132 H482203 H482273 H482340 H482422 H482502 H482577 H482658 H482734 H481865 H481934 H482004 H482066 H482133 H482204 H482274 H482341 H482423 H482503 H482578 H482662 H482735 H481866 H481935 H482005 H482067 H482134 H482205 H482275 H482342 H482426 H482504 H482579 H482663 H482737 H481867 H481936 H482006 H482068 H482135 H482206 H482276 H482343 H482427 H482505 H482580 H482664 H482738 H481868 H481938 H482007 H482069 H482136 H482209 H482277 H482344 H482428 H482506 H482581 H482665 H482739 H481869 H481939 H482008 H482070 H482137 H482210 H482278 H482345 H482429 H482508 H482582 H482666 H482740 H481870 H481940 H482009 H482071 H482138 H482211 H482279 H482346 H482430 H482509 H482583 H482667 H482741 H481871 H481945 H482010 H482072 H482139 H482212 H482281 H482347 H482431 H482510 H482584 H482668 H482742 H481872 H481946 H482011 H482073 H482140 H482213 H482282 H482348 H482432 H482511 H482586 H482669 H482743 H481873 H481950 H482012 H482074 H482141 H482214 H482283 H482349 H482433 H482512 H482587 H482671 H482744 H481874 H481951 H482013 H482075 H482142 H482215 H482284 H482350 H482434 H482513 H482588 H482676 H482745 H481875 H481952 H482014 H482076 H482143 H482216 H482285 H482351 H482435 H482514 H482589 H482677 H482746 H481876 H481953 H482015 H482077 H482144 H482217 H482286 H482352 H482436 H482515 H482590 H482678 H482747 H481877 H481954 H482016 H482078 H482145 H482218 H482287 H482353 H482437 H482516 H482591 H482679 H482748 H481878 H481955 H482017 H482079 H482146 H482219 H482288 H482354 H482438 H482517 H482592 H482680 H482749 H481879 H481956 H482018 H482080 H482147 H482220 H482289 H482355 H482439 H482518 H482593 H482681 H482750 H481881 H481957 H482019 H482081 H482150 H482221 H482290 H482356 H482440 H482519 H482594 H482682 H482751 H481882 H481958 H482020 H482082 H482151 H482222 H482292 H482357 H482441 H482520 H482595 H482683 H482752 H481884 H481959 H482021 H482083 H482152 H482223 H482293 H482358 H482442 H482521 H482596 H482684 H482753 H481885 H481960 H482022 H482084 H482153 H482224 H482294 H482359 H482443 H482522 H482597 H482685 H482754 H481886 H481961 H482023 H482085 H482156 H482225 H482295 H482360 H482444 H482523 H482598 H482686 H482755 H482756 H482830 H482903 H483005 H483097 H483183 H483272 H483342 H483419 H483504 H483584 H483660 H483739 H482757 H482831 H482904 H483006 H483098 H483184 H483273 H483343 H483420 H483505 H483586 H483661 H483740 H482760 H482832 H482906 H483007 H483099 H483185 H483274 H483344 H483421 H483506 H483587 H483662 H483741 H482761 H482833 H482907 H483008 H483100 H483186 H483275 H483345 H483422 H483507 H483588 H483663 H483742 H482762 H482834 H482908 H483009 H483102 H483191 H483276 H483350 H483424 H483509 H483589 H483664 H483743 H482763 H482835 H482909 H483010 H483103 H483192 H483277 H483351 H483425 H483510 H483590 H483665 H483744 H482764 H482836 H482910 H483011 H483104 H483194 H483278 H483352 H483426 H483513 H483591 H483666 H483745 H482765 H482837 H482911 H483012 H483106 H483195 H483279 H483353 H483427 H483514 H483592 H483667 H483746 H482766 H482838 H482912 H483013 H483107 H483196 H483280 H483354 H483428 H483515 H483593 H483668 H483747 H482767 H482839 H482913 H483014 H483109 H483197 H483281 H483355 H483429 H483516 H483594 H483670 H483748 H482768 H482840 H482914 H483015 H483110 H483198 H483282 H483356 H483430 H483517 H483595 H483671 H483749 H482770 H482841 H482915 H483016 H483115 H483199 H483283 H483357 H483431 H483518 H483599 H483672 H483750 H482771 H482842 H482916 H483017 H483117 H483200 H483284 H483359 H483432 H483519 H483600 H483674 H483751 H482772 H482843 H482917 H483018 H483118 H483203 H483285 H483360 H483433 H483521 H483602 H483675 H483752 H482773 H482845 H482918 H483019 H483119 H483204 H483286 H483361 H483434 H483522 H483606 H483676 H483753 H482774 H482846 H482919 H483020 H483120 H483206 H483287 H483362 H483435 H483525 H483607 H483677 H483754 H482775 H482847 H482920 H483021 H483121 H483207 H483288 H483363 H483436 H483526 H483608 H483679 H483755 H482776 H482848 H482921 H483022 H483123 H483208 H483289 H483364 H483437 H483527 H483609 H483680 H483756 H482777 H482849 H482922 H483023 H483124 H483209 H483290 H483365 H483438 H483531 H483610 H483681 H483757 H482778 H482850 H482923 H483024 H483125 H483210 H483291 H483366 H483440 H483532 H483611 H483682 H483758 H482779 H482852 H482924 H483025 H483126 H483211 H483292 H483367 H483441 H483533 H483612 H483683 H483759 H482781 H482853 H482925 H483026 H483127 H483215 H483294 H483369 H483442 H483534 H483615 H483684 H483760 H482782 H482854 H482926 H483027 H483128 H483223 H483295 H483370 H483444 H483535 H483616 H483686 H483761 H482783 H482855 H482927 H483028 H483129 H483224 H483296 H483371 H483445 H483536 H483617 H483687 H483762 H482784 H482856 H482928 H483029 H483130 H483225 H483297 H483372 H483446 H483537 H483619 H483688 H483763 H482785 H482857 H482929 H483040 H483131 H483226 H483298 H483373 H483447 H483538 H483620 H483689 H483764 H482786 H482858 H482931 H483041 H483132 H483227 H483299 H483374 H483449 H483539 H483621 H483690 H483766 H482787 H482859 H482934 H483042 H483133 H483228 H483300 H483375 H483452 H483540 H483622 H483691 H483767 H482788 H482860 H482935 H483043 H483134 H483229 H483301 H483376 H483453 H483541 H483623 H483694 H483768 H482789 H482861 H482936 H483044 H483135 H483233 H483302 H483378 H483454 H483542 H483624 H483695 H483769 H482791 H482862 H482937 H483047 H483136 H483234 H483303 H483379 H483459 H483543 H483625 H483696 H483770 H482792 H482863 H482938 H483048 H483137 H483235 H483306 H483380 H483460 H483544 H483626 H483697 H483771 H482793 H482865 H482939 H483051 H483138 H483236 H483309 H483381 H483461 H483545 H483627 H483698 H483772 H482794 H482866 H482940 H483052 H483139 H483237 H483311 H483382 H483462 H483546 H483628 H483701 H483773 H482795 H482867 H482941 H483053 H483140 H483238 H483312 H483383 H483463 H483547 H483629 H483702 H483774 H482796 H482868 H482942 H483054 H483141 H483240 H483313 H483384 H483465 H483548 H483632 H483705 H483775 H482797 H482869 H482943 H483055 H483142 H483241 H483315 H483385 H483466 H483549 H483633 H483706 H483776 H482799 H482870 H482944 H483057 H483143 H483242 H483317 H483386 H483469 H483550 H483634 H483712 H483777 H482800 H482871 H482945 H483058 H483144 H483243 H483318 H483387 H483470 H483551 H483635 H483713 H483778 H482801 H482874 H482963 H483059 H483145 H483244 H483319 H483388 H483471 H483552 H483636 H483715 H483779 H482802 H482875 H482964 H483060 H483148 H483245 H483320 H483389 H483472 H483554 H483637 H483716 H483780 H482803 H482876 H482965 H483061 H483151 H483246 H483321 H483390 H483473 H483555 H483639 H483717 H483796 H482804 H482877 H482966 H483062 H483152 H483247 H483322 H483391 H483474 H483556 H483640 H483718 H483797 H482805 H482878 H482982 H483063 H483153 H483248 H483323 H483392 H483475 H483557 H483641 H483719 H483798 H482806 H482879 H482983 H483064 H483157 H483249 H483324 H483393 H483477 H483560 H483642 H483720 H483799 H482807 H482880 H482984 H483070 H483158 H483250 H483325 H483394 H483478 H483566 H483643 H483721 H483800 H482812 H482881 H482985 H483071 H483159 H483253 H483326 H483396 H483479 H483567 H483644 H483722 H483801 H482813 H482882 H482986 H483072 H483161 H483254 H483327 H483397 H483485 H483568 H483645 H483723 H483802 H482814 H482883 H482987 H483073 H483162 H483255 H483328 H483401 H483486 H483570 H483646 H483724 H483803 H482815 H482888 H482988 H483075 H483163 H483256 H483329 H483403 H483487 H483571 H483647 H483725 H483804 H482816 H482889 H482989 H483076 H483164 H483257 H483330 H483404 H483488 H483572 H483648 H483726 H483807 H482817 H482890 H482990 H483077 H483165 H483258 H483331 H483405 H483489 H483573 H483649 H483727 H483808 H482818 H482891 H482991 H483078 H483167 H483259 H483332 H483406 H483490 H483574 H483650 H483728 H483809 H482819 H482892 H482992 H483079 H483168 H483260 H483333 H483407 H483491 H483575 H483651 H483730 H483810 H482820 H482893 H482997 H483080 H483173 H483261 H483334 H483408 H483492 H483576 H483652 H483731 H483811 H482821 H482894 H482998 H483081 H483174 H483262 H483335 H483409 H483494 H483577 H483653 H483732 H483812 H482822 H482897 H482999 H483084 H483176 H483265 H483336 H483410 H483495 H483578 H483654 H483733 H483813 H482823 H482898 H483000 H483087 H483178 H483266 H483337 H483411 H483496 H483579 H483655 H483734 H483814 H482824 H482899 H483001 H483088 H483179 H483267 H483338 H483412 H483500 H483580 H483656 H483735 H483816 H482825 H482900 H483002 H483089 H483180 H483268 H483339 H483416 H483501 H483581 H483657 H483736 H483817 H482827 H482901 H483003 H483091 H483181 H483269 H483340 H483417 H483502 H483582 H483658 H483737 H483818 H482829 H482902 H483004 H483092 H483182 H483270 H483341 H483418 H483503 H483583 H483659 H483738 H483820 H483821 H483927 H484004 H484090 H484166 H484235 H484309 H484381 H484452 H484533 H484601 H484681 H484749 H483823 H483928 H484005 H484091 H484167 H484236 H484310 H484382 H484454 H484534 H484602 H484682 H484750 H483824 H483929 H484008 H484092 H484168 H484237 H484311 H484383 H484455 H484535 H484603 H484683 H484751 H483825 H483930 H484009 H484093 H484169 H484238 H484312 H484384 H484456 H484536 H484604 H484684 H484752 H483826 H483932 H484010 H484094 H484170 H484239 H484313 H484385 H484457 H484537 H484605 H484685 H484753 H483827 H483933 H484011 H484095 H484171 H484240 H484314 H484386 H484458 H484538 H484606 H484686 H484754 H483828 H483934 H484012 H484096 H484172 H484241 H484315 H484387 H484459 H484539 H484607 H484687 H484755 H483830 H483935 H484013 H484097 H484173 H484242 H484316 H484388 H484460 H484540 H484608 H484688 H484756 H483831 H483936 H484014 H484098 H484174 H484243 H484317 H484389 H484464 H484541 H484610 H484689 H484757 H483833 H483937 H484015 H484099 H484175 H484244 H484318 H484390 H484465 H484542 H484611 H484691 H484761 H483836 H483938 H484016 H484100 H484177 H484245 H484319 H484391 H484467 H484543 H484612 H484692 H484762 H483837 H483939 H484017 H484103 H484178 H484246 H484320 H484392 H484468 H484544 H484613 H484693 H484763 H483838 H483940 H484018 H484104 H484179 H484247 H484321 H484393 H484469 H484545 H484616 H484694 H484764 H483839 H483941 H484019 H484105 H484180 H484248 H484322 H484394 H484472 H484546 H484617 H484695 H484765 H483840 H483942 H484020 H484107 H484181 H484250 H484323 H484395 H484473 H484547 H484619 H484696 H484766 H483841 H483943 H484023 H484108 H484182 H484251 H484324 H484397 H484474 H484548 H484620 H484697 H484767 H483842 H483944 H484024 H484109 H484185 H484252 H484325 H484398 H484475 H484549 H484621 H484698 H484768 H483843 H483945 H484025 H484111 H484186 H484253 H484326 H484399 H484476 H484550 H484622 H484699 H484769 H483844 H483946 H484026 H484112 H484188 H484254 H484327 H484400 H484477 H484551 H484624 H484701 H484770 H483845 H483947 H484027 H484113 H484189 H484256 H484328 H484401 H484478 H484552 H484625 H484703 H484771 H483847 H483948 H484031 H484114 H484190 H484257 H484329 H484402 H484479 H484553 H484626 H484704 H484772 H483848 H483949 H484032 H484115 H484191 H484258 H484330 H484407 H484481 H484554 H484627 H484705 H484773 H483849 H483950 H484033 H484116 H484192 H484260 H484331 H484408 H484482 H484555 H484628 H484706 H484774 H483850 H483951 H484034 H484117 H484193 H484261 H484332 H484409 H484483 H484556 H484629 H484707 H484775 H483851 H483952 H484035 H484118 H484194 H484263 H484334 H484410 H484485 H484557 H484630 H484708 H484777 H483852 H483953 H484036 H484119 H484195 H484264 H484335 H484411 H484486 H484558 H484631 H484709 H484778 H483853 H483957 H484039 H484120 H484198 H484265 H484336 H484412 H484487 H484559 H484632 H484710 H484779 H483854 H483960 H484040 H484121 H484199 H484266 H484337 H484413 H484495 H484560 H484634 H484711 H484780 H483855 H483964 H484041 H484122 H484200 H484267 H484339 H484414 H484496 H484561 H484635 H484712 H484781 H483856 H483966 H484042 H484123 H484201 H484268 H484340 H484415 H484497 H484562 H484636 H484715 H484782 H483858 H483967 H484043 H484124 H484202 H484269 H484341 H484416 H484498 H484563 H484637 H484716 H484784 H483859 H483968 H484044 H484125 H484203 H484270 H484342 H484418 H484499 H484564 H484638 H484717 H484785 H483860 H483969 H484045 H484126 H484204 H484271 H484343 H484419 H484500 H484565 H484639 H484718 H484786 H483861 H483970 H484046 H484128 H484205 H484272 H484345 H484421 H484501 H484566 H484640 H484719 H484787 H483862 H483971 H484047 H484129 H484206 H484273 H484346 H484422 H484502 H484567 H484641 H484720 H484788 H483866 H483973 H484048 H484130 H484207 H484274 H484348 H484423 H484503 H484568 H484650 H484721 H484789 H483867 H483975 H484049 H484131 H484208 H484275 H484349 H484424 H484504 H484569 H484651 H484722 H484790 H483868 H483976 H484050 H484132 H484209 H484276 H484350 H484425 H484506 H484570 H484652 H484723 H484791 H483869 H483977 H484051 H484133 H484210 H484279 H484351 H484426 H484507 H484571 H484653 H484724 H484792 H483871 H483978 H484052 H484135 H484211 H484280 H484353 H484427 H484510 H484573 H484654 H484725 H484793 H483872 H483979 H484053 H484139 H484212 H484281 H484354 H484429 H484511 H484574 H484656 H484726 H484794 H483873 H483980 H484054 H484140 H484213 H484282 H484355 H484430 H484512 H484575 H484657 H484727 H484795 H483874 H483981 H484055 H484143 H484214 H484283 H484356 H484431 H484513 H484578 H484658 H484728 H484796 H483875 H483982 H484056 H484144 H484215 H484285 H484357 H484433 H484514 H484579 H484659 H484729 H484797 H483876 H483983 H484057 H484145 H484216 H484286 H484358 H484434 H484515 H484580 H484661 H484730 H484798 H483877 H483984 H484058 H484147 H484217 H484287 H484359 H484435 H484516 H484581 H484662 H484731 H484799 H483879 H483986 H484059 H484148 H484218 H484288 H484360 H484436 H484517 H484582 H484663 H484732 H484800 H483880 H483987 H484060 H484149 H484219 H484289 H484361 H484437 H484518 H484583 H484665 H484733 H484801 H483881 H483988 H484061 H484151 H484220 H484290 H484362 H484438 H484519 H484584 H484666 H484734 H484802 H483914 H483991 H484062 H484153 H484221 H484291 H484363 H484439 H484520 H484585 H484667 H484735 H484803 H483915 H483992 H484063 H484154 H484222 H484292 H484367 H484440 H484521 H484586 H484668 H484736 H484804 H483916 H483993 H484072 H484155 H484224 H484293 H484368 H484441 H484522 H484588 H484669 H484737 H484805 H483917 H483994 H484073 H484156 H484225 H484295 H484369 H484442 H484523 H484589 H484670 H484738 H484806 H483918 H483995 H484076 H484157 H484226 H484296 H484370 H484443 H484524 H484590 H484671 H484739 H484807 H483919 H483996 H484077 H484158 H484227 H484297 H484371 H484444 H484525 H484591 H484672 H484740 H484808 H483920 H483997 H484079 H484159 H484228 H484298 H484374 H484445 H484526 H484592 H484673 H484741 H484809 H483921 H483998 H484080 H484160 H484229 H484299 H484375 H484446 H484527 H484593 H484674 H484742 H484810 H483922 H483999 H484081 H484161 H484230 H484300 H484376 H484447 H484528 H484594 H484675 H484744 H484811 H483923 H484000 H484082 H484162 H484231 H484301 H484377 H484448 H484529 H484597 H484677 H484745 H484812 H483924 H484001 H484083 H484163 H484232 H484306 H484378 H484449 H484530 H484598 H484678 H484746 H484813 H483925 H484002 H484084 H484164 H484233 H484307 H484379 H484450 H484531 H484599 H484679 H484747 H484814 H483926 H484003 H484085 H484165 H484234 H484308 H484380 H484451 H484532 H484600 H484680 H484748 H484815 H484816 H484888 H484970 H485040 H485107 H485176 H485249 H485318 H485381 H485462 H485555 H485637 H485716 H484817 H484889 H484971 H485041 H485108 H485177 H485250 H485319 H485382 H485463 H485556 H485638 H485717 H484818 H484890 H484972 H485042 H485109 H485178 H485251 H485320 H485383 H485465 H485557 H485639 H485719 H484819 H484891 H484973 H485043 H485110 H485179 H485252 H485321 H485384 H485466 H485558 H485640 H485720 H484820 H484892 H484974 H485044 H485111 H485180 H485253 H485322 H485385 H485467 H485559 H485641 H485721 H484822 H484893 H484975 H485045 H485113 H485181 H485254 H485323 H485386 H485468 H485560 H485642 H485722 H484823 H484894 H484976 H485046 H485114 H485182 H485255 H485324 H485387 H485469 H485561 H485643 H485723 H484824 H484895 H484977 H485047 H485115 H485183 H485256 H485325 H485388 H485470 H485562 H485644 H485724 H484825 H484896 H484978 H485048 H485116 H485184 H485257 H485326 H485389 H485471 H485563 H485645 H485725 H484826 H484897 H484979 H485049 H485117 H485185 H485258 H485327 H485390 H485473 H485564 H485646 H485726 H484827 H484898 H484980 H485050 H485118 H485186 H485259 H485328 H485391 H485474 H485565 H485647 H485727 H484828 H484899 H484981 H485051 H485119 H485187 H485260 H485329 H485392 H485475 H485569 H485648 H485728 H484829 H484900 H484987 H485052 H485120 H485188 H485261 H485330 H485393 H485476 H485570 H485653 H485729 H484830 H484901 H484988 H485053 H485121 H485189 H485262 H485331 H485396 H485477 H485571 H485654 H485730 H484831 H484903 H484989 H485054 H485122 H485190 H485264 H485332 H485397 H485478 H485572 H485656 H485731 H484832 H484904 H484990 H485055 H485125 H485191 H485265 H485333 H485398 H485481 H485573 H485657 H485732 H484833 H484905 H484991 H485056 H485126 H485193 H485266 H485334 H485399 H485482 H485574 H485658 H485733 H484834 H484906 H484992 H485057 H485127 H485194 H485267 H485335 H485400 H485483 H485576 H485660 H485734 H484835 H484908 H484993 H485058 H485128 H485195 H485268 H485336 H485401 H485484 H485577 H485661 H485736 H484836 H484909 H484994 H485059 H485129 H485196 H485269 H485337 H485402 H485485 H485578 H485663 H485737 H484837 H484910 H484995 H485060 H485130 H485197 H485270 H485338 H485403 H485486 H485579 H485667 H485738 H484838 H484911 H484996 H485061 H485131 H485198 H485271 H485339 H485404 H485487 H485580 H485668 H485739 H484839 H484912 H484997 H485062 H485132 H485199 H485272 H485340 H485405 H485488 H485581 H485669 H485740 H484840 H484913 H484998 H485063 H485133 H485200 H485273 H485341 H485412 H485489 H485582 H485670 H485741 H484841 H484915 H484999 H485064 H485134 H485201 H485274 H485342 H485413 H485490 H485583 H485671 H485742 H484842 H484916 H485000 H485065 H485135 H485202 H485275 H485343 H485418 H485491 H485585 H485672 H485743 H484844 H484917 H485001 H485066 H485137 H485203 H485276 H485344 H485419 H485492 H485586 H485675 H485744 H484845 H484918 H485002 H485067 H485138 H485204 H485277 H485345 H485420 H485495 H485587 H485676 H485745 H484846 H484919 H485003 H485068 H485139 H485205 H485278 H485346 H485421 H485516 H485588 H485677 H485746 H484847 H484920 H485004 H485069 H485140 H485206 H485279 H485347 H485428 H485517 H485589 H485678 H485747 H484849 H484923 H485005 H485070 H485141 H485207 H485283 H485348 H485429 H485518 H485590 H485679 H485748 H484850 H484925 H485006 H485071 H485142 H485209 H485284 H485349 H485430 H485519 H485592 H485681 H485749 H484851 H484926 H485007 H485072 H485143 H485210 H485285 H485350 H485431 H485520 H485593 H485682 H485750 H484852 H484932 H485008 H485073 H485144 H485211 H485286 H485352 H485432 H485521 H485594 H485683 H485751 H484853 H484933 H485009 H485074 H485145 H485212 H485288 H485353 H485433 H485524 H485595 H485684 H485752 H484854 H484934 H485010 H485075 H485146 H485213 H485289 H485354 H485434 H485525 H485596 H485685 H485753 H484856 H484935 H485011 H485076 H485147 H485214 H485291 H485355 H485435 H485526 H485597 H485686 H485754 H484859 H484936 H485012 H485077 H485148 H485215 H485292 H485356 H485436 H485527 H485599 H485687 H485755 H484861 H484937 H485013 H485078 H485151 H485216 H485293 H485357 H485437 H485529 H485601 H485689 H485756 H484862 H484938 H485014 H485079 H485152 H485217 H485294 H485358 H485438 H485530 H485602 H485690 H485757 H484863 H484939 H485015 H485081 H485153 H485218 H485295 H485359 H485439 H485531 H485603 H485691 H485758 H484864 H484940 H485016 H485082 H485154 H485222 H485296 H485360 H485440 H485532 H485604 H485692 H485759 H484865 H484941 H485017 H485083 H485155 H485223 H485297 H485361 H485442 H485533 H485607 H485693 H485760 H484866 H484942 H485018 H485084 H485156 H485224 H485298 H485362 H485443 H485534 H485608 H485694 H485761 H484868 H484943 H485019 H485085 H485157 H485225 H485299 H485363 H485444 H485535 H485609 H485695 H485762 H484869 H484945 H485020 H485087 H485158 H485226 H485301 H485364 H485445 H485536 H485612 H485696 H485763 H484870 H484946 H485021 H485088 H485159 H485227 H485302 H485365 H485446 H485537 H485613 H485697 H485764 H484871 H484947 H485022 H485090 H485160 H485228 H485303 H485366 H485447 H485538 H485614 H485698 H485765 H484872 H484948 H485023 H485091 H485161 H485229 H485304 H485367 H485448 H485540 H485615 H485699 H485766 H484873 H484950 H485024 H485092 H485162 H485230 H485305 H485368 H485449 H485541 H485617 H485700 H485767 H484874 H484951 H485025 H485093 H485163 H485231 H485306 H485369 H485450 H485542 H485618 H485701 H485768 H484876 H484952 H485026 H485094 H485164 H485232 H485307 H485370 H485451 H485543 H485620 H485702 H485769 H484877 H484953 H485027 H485095 H485165 H485233 H485308 H485371 H485452 H485545 H485621 H485703 H485770 H484878 H484954 H485028 H485096 H485166 H485234 H485309 H485372 H485453 H485546 H485622 H485704 H485771 H484879 H484955 H485029 H485099 H485168 H485235 H485310 H485373 H485454 H485547 H485628 H485705 H485772 H484880 H484956 H485030 H485100 H485169 H485240 H485311 H485374 H485455 H485548 H485629 H485706 H485773 H484881 H484957 H485033 H485101 H485170 H485241 H485312 H485375 H485456 H485549 H485630 H485707 H485774 H484882 H484958 H485035 H485102 H485171 H485242 H485313 H485376 H485457 H485550 H485632 H485708 H485775 H484883 H484959 H485036 H485103 H485172 H485243 H485314 H485377 H485458 H485551 H485633 H485711 H485776 H484884 H484966 H485037 H485104 H485173 H485244 H485315 H485378 H485459 H485552 H485634 H485713 H485777 H484886 H484967 H485038 H485105 H485174 H485247 H485316 H485379 H485460 H485553 H485635 H485714 H485778 H484887 H484968 H485039 H485106 H485175 H485248 H485317 H485380 H485461 H485554 H485636 H485715 H485779 H485780 H485851 H485927 H485996 H486074 H486145 H486212 H486286 H486356 H486419 H486483 H486558 H486632 H485781 H485852 H485928 H485997 H486075 H486146 H486213 H486287 H486357 H486420 H486484 H486559 H486633 H485782 H485853 H485929 H485998 H486076 H486147 H486214 H486288 H486358 H486421 H486485 H486560 H486634 H485783 H485854 H485931 H485999 H486077 H486148 H486215 H486289 H486359 H486422 H486486 H486561 H486635 H485784 H485855 H485932 H486000 H486078 H486149 H486216 H486290 H486360 H486423 H486488 H486562 H486636 H485785 H485856 H485933 H486001 H486079 H486150 H486217 H486291 H486361 H486424 H486489 H486563 H486637 H485786 H485857 H485934 H486003 H486080 H486151 H486218 H486292 H486362 H486425 H486491 H486564 H486638 H485787 H485858 H485935 H486004 H486082 H486152 H486219 H486293 H486363 H486426 H486492 H486565 H486639 H485788 H485859 H485936 H486005 H486086 H486153 H486220 H486294 H486364 H486427 H486493 H486566 H486640 H485790 H485860 H485937 H486007 H486087 H486154 H486224 H486295 H486365 H486428 H486494 H486567 H486641 H485791 H485862 H485938 H486009 H486088 H486155 H486225 H486296 H486366 H486429 H486495 H486568 H486642 H485792 H485863 H485939 H486016 H486089 H486156 H486226 H486297 H486367 H486430 H486496 H486570 H486643 H485793 H485867 H485940 H486017 H486091 H486157 H486227 H486299 H486368 H486431 H486497 H486571 H486644 H485795 H485868 H485941 H486018 H486093 H486158 H486228 H486300 H486369 H486432 H486498 H486572 H486645 H485796 H485869 H485942 H486019 H486094 H486159 H486231 H486301 H486370 H486433 H486499 H486579 H486646 H485797 H485870 H485943 H486020 H486095 H486160 H486232 H486302 H486371 H486434 H486500 H486580 H486647 H485798 H485871 H485944 H486021 H486096 H486161 H486233 H486303 H486372 H486435 H486501 H486581 H486648 H485799 H485872 H485945 H486023 H486097 H486162 H486234 H486304 H486373 H486436 H486502 H486582 H486649 H485800 H485873 H485946 H486024 H486098 H486163 H486235 H486306 H486374 H486437 H486503 H486583 H486651 H485801 H485874 H485948 H486026 H486099 H486164 H486236 H486307 H486375 H486438 H486504 H486584 H486652 H485802 H485875 H485949 H486029 H486100 H486165 H486237 H486308 H486376 H486439 H486505 H486585 H486653 H485803 H485876 H485950 H486030 H486101 H486166 H486238 H486309 H486377 H486441 H486506 H486586 H486654 H485804 H485879 H485951 H486031 H486102 H486167 H486239 H486310 H486378 H486442 H486507 H486587 H486655 H485805 H485886 H485953 H486032 H486103 H486168 H486240 H486311 H486379 H486443 H486508 H486588 H486656 H485806 H485887 H485954 H486033 H486104 H486169 H486241 H486312 H486380 H486444 H486509 H486589 H486657 H485807 H485888 H485955 H486034 H486105 H486170 H486242 H486313 H486381 H486445 H486510 H486590 H486659 H485808 H485889 H485956 H486035 H486107 H486171 H486243 H486314 H486382 H486446 H486511 H486591 H486660 H485809 H485890 H485957 H486036 H486108 H486172 H486244 H486315 H486383 H486447 H486517 H486592 H486661 H485810 H485891 H485958 H486037 H486109 H486173 H486245 H486316 H486384 H486448 H486518 H486593 H486662 H485811 H485892 H485960 H486038 H486110 H486174 H486250 H486317 H486385 H486449 H486520 H486594 H486663 H485812 H485893 H485961 H486040 H486111 H486175 H486253 H486318 H486386 H486451 H486522 H486595 H486664 H485814 H485894 H485962 H486041 H486113 H486176 H486254 H486319 H486387 H486452 H486523 H486596 H486665 H485815 H485895 H485963 H486042 H486114 H486177 H486255 H486320 H486388 H486453 H486524 H486597 H486666 H485816 H485896 H485964 H486043 H486115 H486178 H486256 H486322 H486389 H486454 H486525 H486598 H486670 H485817 H485897 H485965 H486044 H486116 H486179 H486257 H486323 H486390 H486455 H486526 H486599 H486671 H485818 H485898 H485966 H486045 H486117 H486180 H486258 H486324 H486391 H486456 H486527 H486600 H486672 H485820 H485899 H485967 H486046 H486118 H486181 H486259 H486325 H486392 H486457 H486528 H486603 H486673 H485821 H485900 H485968 H486047 H486119 H486182 H486260 H486326 H486393 H486458 H486529 H486604 H486674 H485823 H485901 H485969 H486048 H486120 H486183 H486261 H486327 H486394 H486459 H486530 H486606 H486675 H485824 H485902 H485970 H486049 H486121 H486184 H486262 H486328 H486395 H486460 H486531 H486607 H486676 H485825 H485903 H485971 H486050 H486123 H486185 H486263 H486330 H486396 H486461 H486532 H486608 H486677 H485826 H485904 H485972 H486051 H486124 H486186 H486264 H486331 H486397 H486462 H486535 H486609 H486678 H485827 H485905 H485973 H486052 H486125 H486187 H486265 H486333 H486398 H486463 H486536 H486610 H486679 H485828 H485907 H485974 H486053 H486126 H486188 H486266 H486334 H486399 H486464 H486537 H486611 H486680 H485829 H485908 H485975 H486054 H486127 H486189 H486267 H486336 H486400 H486465 H486538 H486613 H486681 H485830 H485909 H485976 H486055 H486128 H486190 H486269 H486337 H486401 H486466 H486539 H486614 H486683 H485831 H485910 H485977 H486056 H486129 H486191 H486270 H486338 H486402 H486467 H486540 H486615 H486685 H485832 H485911 H485978 H486057 H486130 H486196 H486271 H486339 H486403 H486468 H486541 H486616 H486686 H485833 H485912 H485979 H486058 H486131 H486197 H486272 H486340 H486404 H486469 H486542 H486617 H486687 H485835 H485913 H485980 H486059 H486132 H486198 H486273 H486341 H486405 H486470 H486543 H486618 H486688 H485836 H485914 H485982 H486060 H486133 H486199 H486274 H486342 H486406 H486471 H486544 H486619 H486689 H485837 H485915 H485983 H486061 H486134 H486200 H486275 H486343 H486407 H486472 H486545 H486620 H486690 H485838 H485916 H485984 H486062 H486135 H486201 H486276 H486344 H486408 H486473 H486546 H486621 H486691 H485839 H485918 H485985 H486063 H486136 H486203 H486277 H486345 H486409 H486474 H486547 H486623 H486694 H485840 H485919 H485987 H486064 H486137 H486204 H486278 H486346 H486411 H486475 H486548 H486624 H486695 H485841 H485920 H485988 H486065 H486138 H486205 H486279 H486347 H486412 H486476 H486549 H486625 H486696 H485842 H485921 H485989 H486068 H486139 H486206 H486280 H486348 H486413 H486477 H486550 H486626 H486697 H485843 H485922 H485990 H486069 H486140 H486207 H486281 H486349 H486414 H486478 H486553 H486627 H486698 H485846 H485923 H485991 H486070 H486141 H486208 H486282 H486350 H486415 H486479 H486554 H486628 H486699 H485848 H485924 H485992 H486071 H486142 H486209 H486283 H486352 H486416 H486480 H486555 H486629 H486700 H485849 H485925 H485994 H486072 H486143 H486210 H486284 H486353 H486417 H486481 H486556 H486630 H486701 H485850 H485926 H485995 H486073 H486144 H486211 H486285 H486354 H486418 H486482 H486557 H486631 H486702 H486703 H486788 H486857 H486924 H486989 H487053 H487119 H487185 H487262 H487331 H487438 H487510 H487576 H486704 H486789 H486858 H486926 H486990 H487054 H487120 H487186 H487263 H487332 H487440 H487511 H487577 H486705 H486790 H486859 H486927 H486991 H487055 H487121 H487187 H487264 H487333 H487441 H487512 H487579 H486706 H486791 H486860 H486928 H486992 H487056 H487125 H487188 H487265 H487340 H487442 H487513 H487580 H486707 H486792 H486861 H486929 H486993 H487057 H487126 H487189 H487266 H487343 H487443 H487514 H487581 H486708 H486793 H486862 H486930 H486994 H487058 H487127 H487190 H487267 H487344 H487444 H487515 H487582 H486709 H486794 H486863 H486931 H486995 H487059 H487128 H487191 H487268 H487345 H487445 H487516 H487583 H486710 H486795 H486864 H486932 H486996 H487060 H487129 H487192 H487269 H487346 H487446 H487517 H487584 H486711 H486796 H486865 H486933 H486997 H487061 H487130 H487193 H487270 H487347 H487447 H487518 H487587 H486712 H486797 H486866 H486934 H486998 H487062 H487131 H487195 H487271 H487350 H487448 H487519 H487588 H486713 H486798 H486867 H486935 H486999 H487063 H487132 H487196 H487272 H487351 H487449 H487520 H487590 H486714 H486799 H486868 H486936 H487000 H487064 H487133 H487197 H487273 H487352 H487450 H487521 H487591 H486715 H486800 H486869 H486937 H487001 H487065 H487134 H487198 H487274 H487378 H487452 H487522 H487592 H486716 H486801 H486870 H486938 H487002 H487066 H487135 H487199 H487275 H487379 H487453 H487523 H487593 H486717 H486802 H486871 H486939 H487003 H487067 H487136 H487200 H487276 H487380 H487454 H487524 H487594 H486718 H486803 H486872 H486940 H487004 H487068 H487137 H487201 H487277 H487381 H487455 H487525 H487595 H486719 H486804 H486873 H486941 H487005 H487069 H487138 H487202 H487278 H487383 H487457 H487526 H487596 H486720 H486805 H486874 H486942 H487006 H487070 H487139 H487203 H487279 H487386 H487458 H487527 H487597 H486721 H486806 H486878 H486943 H487007 H487071 H487140 H487204 H487282 H487387 H487459 H487528 H487598 H486722 H486807 H486879 H486944 H487008 H487072 H487141 H487205 H487283 H487392 H487460 H487529 H487599 H486723 H486808 H486880 H486945 H487009 H487073 H487142 H487206 H487284 H487393 H487461 H487530 H487600 H486724 H486809 H486881 H486946 H487010 H487075 H487143 H487210 H487285 H487394 H487466 H487531 H487601 H486727 H486810 H486882 H486947 H487011 H487076 H487144 H487211 H487286 H487395 H487467 H487532 H487602 H486728 H486811 H486883 H486948 H487012 H487077 H487145 H487212 H487287 H487396 H487468 H487533 H487605 H486729 H486812 H486884 H486949 H487013 H487078 H487146 H487213 H487288 H487397 H487469 H487534 H487606 H486730 H486813 H486885 H486950 H487014 H487079 H487147 H487214 H487289 H487398 H487470 H487535 H487608 H486731 H486814 H486886 H486951 H487015 H487080 H487148 H487215 H487290 H487399 H487473 H487536 H487609 H486732 H486815 H486887 H486954 H487016 H487081 H487149 H487216 H487291 H487400 H487474 H487537 H487610 H486733 H486816 H486888 H486955 H487017 H487082 H487150 H487217 H487292 H487401 H487475 H487538 H487611 H486734 H486817 H486889 H486956 H487019 H487083 H487151 H487218 H487293 H487402 H487476 H487539 H487612 H486735 H486818 H486890 H486957 H487020 H487084 H487152 H487219 H487294 H487403 H487477 H487540 H487613 H486736 H486819 H486891 H486958 H487021 H487085 H487153 H487220 H487295 H487404 H487478 H487541 H487614 H486737 H486820 H486892 H486959 H487022 H487086 H487154 H487223 H487296 H487405 H487479 H487543 H487615 H486738 H486821 H486893 H486960 H487023 H487088 H487155 H487225 H487297 H487406 H487480 H487544 H487616 H486739 H486823 H486895 H486961 H487024 H487089 H487156 H487226 H487298 H487407 H487481 H487545 H487617 H486740 H486824 H486896 H486962 H487025 H487090 H487157 H487227 H487299 H487408 H487482 H487546 H487619 H486754 H486825 H486897 H486963 H487026 H487091 H487158 H487228 H487300 H487409 H487483 H487547 H487620 H486755 H486826 H486898 H486964 H487028 H487092 H487159 H487229 H487301 H487410 H487484 H487548 H487621 H486756 H486827 H486899 H486965 H487029 H487093 H487160 H487231 H487302 H487411 H487485 H487549 H487622 H486757 H486828 H486900 H486966 H487030 H487094 H487161 H487232 H487303 H487412 H487486 H487550 H487623 H486758 H486830 H486901 H486967 H487031 H487097 H487162 H487233 H487304 H487413 H487487 H487551 H487624 H486759 H486831 H486902 H486968 H487032 H487098 H487164 H487235 H487305 H487414 H487488 H487554 H487625 H486760 H486832 H486903 H486969 H487033 H487099 H487165 H487236 H487306 H487415 H487489 H487555 H487626 H486766 H486833 H486904 H486970 H487034 H487100 H487166 H487237 H487307 H487416 H487490 H487556 H487627 H486767 H486834 H486906 H486971 H487035 H487101 H487167 H487238 H487308 H487417 H487491 H487557 H487628 H486769 H486835 H486907 H486972 H487036 H487102 H487168 H487239 H487309 H487418 H487493 H487559 H487629 H486771 H486836 H486908 H486973 H487037 H487103 H487169 H487240 H487310 H487419 H487494 H487560 H487630 H486772 H486837 H486909 H486974 H487038 H487104 H487170 H487242 H487311 H487420 H487495 H487561 H487631 H486773 H486838 H486910 H486975 H487039 H487105 H487171 H487247 H487312 H487423 H487496 H487562 H487632 H486774 H486839 H486911 H486976 H487040 H487106 H487172 H487248 H487313 H487425 H487497 H487563 H487633 H486775 H486840 H486912 H486977 H487041 H487107 H487173 H487249 H487314 H487426 H487498 H487564 H487634 H486776 H486841 H486913 H486978 H487042 H487108 H487174 H487250 H487315 H487427 H487499 H487565 H487635 H486777 H486847 H486914 H486979 H487043 H487109 H487175 H487252 H487320 H487428 H487500 H487566 H487636 H486778 H486848 H486915 H486980 H487044 H487110 H487176 H487253 H487321 H487429 H487501 H487567 H487639 H486779 H486849 H486916 H486981 H487045 H487111 H487177 H487254 H487322 H487430 H487502 H487568 H487640 H486780 H486850 H486917 H486982 H487046 H487112 H487178 H487255 H487323 H487431 H487503 H487569 H487641 H486781 H486851 H486918 H486983 H487047 H487113 H487179 H487256 H487324 H487432 H487504 H487570 H487642 H486782 H486852 H486919 H486984 H487048 H487114 H487180 H487257 H487325 H487433 H487505 H487571 H487643 H486784 H486853 H486920 H486985 H487049 H487115 H487181 H487258 H487326 H487434 H487506 H487572 H487644 H486785 H486854 H486921 H486986 H487050 H487116 H487182 H487259 H487327 H487435 H487507 H487573 H487645 H486786 H486855 H486922 H486987 H487051 H487117 H487183 H487260 H487328 H487436 H487508 H487574 H487646 H486787 H486856 H486923 H486988 H487052 H487118 H487184 H487261 H487330 H487437 H487509 H487575 H487647 H487648 H487721 H487789 H487871 H487940 H488019 H488090 H488163 H488234 H488307 H488375 H488442 H488518 H487649 H487722 H487790 H487873 H487941 H488020 H488091 H488164 H488235 H488308 H488376 H488443 H488519 H487650 H487723 H487791 H487874 H487942 H488021 H488092 H488165 H488237 H488309 H488377 H488444 H488520 H487651 H487724 H487792 H487875 H487943 H488022 H488093 H488166 H488238 H488310 H488378 H488445 H488521 H487652 H487725 H487793 H487876 H487944 H488023 H488094 H488169 H488239 H488311 H488379 H488446 H488525 H487653 H487726 H487794 H487877 H487945 H488024 H488096 H488170 H488240 H488312 H488380 H488447 H488526 H487654 H487728 H487795 H487878 H487946 H488025 H488097 H488171 H488241 H488313 H488381 H488448 H488527 H487655 H487729 H487796 H487879 H487948 H488026 H488098 H488172 H488242 H488314 H488382 H488449 H488528 H487656 H487730 H487797 H487880 H487949 H488027 H488099 H488173 H488243 H488315 H488383 H488450 H488529 H487657 H487731 H487798 H487882 H487950 H488028 H488100 H488174 H488244 H488316 H488384 H488452 H488530 H487659 H487732 H487811 H487883 H487951 H488029 H488101 H488175 H488245 H488317 H488385 H488453 H488532 H487660 H487733 H487812 H487884 H487952 H488030 H488102 H488176 H488247 H488318 H488387 H488454 H488533 H487661 H487734 H487815 H487885 H487953 H488031 H488103 H488177 H488248 H488319 H488388 H488455 H488534 H487662 H487735 H487817 H487886 H487954 H488032 H488104 H488178 H488249 H488320 H488389 H488456 H488535 H487663 H487736 H487818 H487887 H487955 H488033 H488106 H488179 H488250 H488321 H488390 H488457 H488536 H487664 H487737 H487819 H487888 H487956 H488034 H488107 H488182 H488251 H488322 H488391 H488459 H488537 H487665 H487739 H487820 H487889 H487957 H488035 H488108 H488183 H488254 H488323 H488392 H488460 H488538 H487666 H487740 H487821 H487890 H487958 H488036 H488109 H488184 H488255 H488324 H488393 H488461 H488539 H487667 H487741 H487822 H487891 H487959 H488037 H488110 H488185 H488256 H488325 H488394 H488462 H488541 H487668 H487742 H487823 H487892 H487960 H488038 H488111 H488186 H488257 H488326 H488395 H488463 H488542 H487669 H487743 H487824 H487893 H487961 H488039 H488112 H488188 H488258 H488327 H488396 H488464 H488543 H487670 H487744 H487825 H487894 H487962 H488040 H488113 H488189 H488261 H488328 H488397 H488465 H488544 H487671 H487745 H487826 H487895 H487963 H488041 H488114 H488190 H488262 H488329 H488398 H488466 H488545 H487672 H487746 H487828 H487897 H487964 H488042 H488115 H488191 H488263 H488330 H488399 H488467 H488546 H487675 H487747 H487829 H487898 H487965 H488043 H488116 H488192 H488264 H488331 H488401 H488468 H488547 H487676 H487748 H487830 H487899 H487966 H488044 H488117 H488193 H488265 H488332 H488403 H488469 H488548 H487677 H487749 H487831 H487900 H487967 H488045 H488118 H488194 H488266 H488333 H488404 H488470 H488549 H487678 H487750 H487832 H487902 H487968 H488047 H488119 H488196 H488267 H488334 H488405 H488471 H488550 H487679 H487751 H487834 H487903 H487969 H488048 H488120 H488197 H488268 H488335 H488406 H488472 H488551 H487680 H487752 H487835 H487904 H487970 H488049 H488124 H488198 H488269 H488337 H488407 H488473 H488552 H487681 H487753 H487836 H487905 H487971 H488050 H488125 H488199 H488270 H488338 H488408 H488474 H488553 H487682 H487754 H487837 H487906 H487972 H488051 H488126 H488200 H488275 H488339 H488409 H488475 H488558 H487683 H487755 H487838 H487907 H487973 H488052 H488127 H488201 H488276 H488340 H488411 H488476 H488559 H487684 H487757 H487839 H487908 H487974 H488054 H488128 H488202 H488277 H488341 H488412 H488477 H488560 H487685 H487758 H487840 H487909 H487975 H488055 H488129 H488203 H488278 H488342 H488413 H488478 H488561 H487686 H487759 H487841 H487910 H487976 H488056 H488130 H488204 H488279 H488343 H488414 H488481 H488564 H487687 H487760 H487842 H487911 H487977 H488057 H488131 H488205 H488280 H488344 H488415 H488483 H488565 H487688 H487761 H487843 H487912 H487979 H488058 H488132 H488206 H488281 H488345 H488416 H488484 H488566 H487689 H487762 H487844 H487913 H487980 H488059 H488133 H488207 H488282 H488346 H488417 H488485 H488567 H487690 H487763 H487845 H487914 H487981 H488060 H488134 H488208 H488283 H488349 H488418 H488486 H488568 H487691 H487766 H487846 H487915 H487982 H488061 H488135 H488209 H488284 H488353 H488419 H488487 H488569 H487692 H487767 H487847 H487916 H487983 H488062 H488136 H488210 H488285 H488354 H488420 H488488 H488570 H487693 H487768 H487848 H487917 H487984 H488063 H488137 H488211 H488286 H488355 H488421 H488489 H488571 H487694 H487769 H487849 H487919 H487990 H488070 H488138 H488212 H488287 H488356 H488422 H488490 H488573 H487695 H487770 H487850 H487920 H487991 H488071 H488139 H488213 H488288 H488357 H488423 H488491 H488574 H487696 H487771 H487851 H487921 H487992 H488072 H488140 H488214 H488289 H488358 H488424 H488492 H488575 H487697 H487772 H487852 H487922 H487993 H488073 H488141 H488215 H488290 H488359 H488425 H488493 H488576 H487698 H487773 H487853 H487923 H487994 H488074 H488142 H488216 H488291 H488360 H488426 H488494 H488577 H487699 H487774 H487854 H487924 H487995 H488075 H488144 H488217 H488292 H488361 H488427 H488495 H488578 H487700 H487775 H487855 H487925 H487996 H488076 H488150 H488220 H488293 H488362 H488429 H488496 H488579 H487703 H487776 H487858 H487926 H487997 H488077 H488151 H488221 H488295 H488363 H488430 H488497 H488580 H487704 H487777 H487859 H487927 H487998 H488078 H488152 H488222 H488296 H488364 H488431 H488498 H488581 H487705 H487778 H487860 H487928 H487999 H488079 H488153 H488223 H488297 H488365 H488432 H488499 H488582 H487706 H487779 H487861 H487929 H488001 H488080 H488154 H488224 H488298 H488366 H488433 H488505 H488583 H487707 H487780 H487862 H487930 H488002 H488081 H488155 H488225 H488299 H488367 H488434 H488507 H488584 H487708 H487781 H487863 H487933 H488003 H488082 H488156 H488227 H488300 H488368 H488435 H488508 H488585 H487709 H487782 H487865 H487934 H488004 H488083 H488157 H488228 H488301 H488369 H488436 H488512 H488586 H487710 H487783 H487866 H487935 H488005 H488084 H488158 H488229 H488302 H488370 H488437 H488513 H488587 H487711 H487784 H487867 H487936 H488006 H488085 H488159 H488230 H488303 H488371 H488438 H488514 H488588 H487714 H487785 H487868 H487937 H488007 H488086 H488160 H488231 H488304 H488372 H488439 H488515 H488589 H487715 H487786 H487869 H487938 H488008 H488087 H488161 H488232 H488305 H488373 H488440 H488516 H488590 H487720 H487788 H487870 H487939 H488018 H488089 H488162 H488233 H488306 H488374 H488441 H488517 H488591 H488592 H488663 H488730 H488804 H488875 H488938 H489008 H489076 H489150 H489218 H489295 H489359 H489424 H488594 H488664 H488731 H488805 H488876 H488940 H489009 H489077 H489151 H489219 H489296 H489360 H489425 H488595 H488665 H488732 H488806 H488877 H488941 H489010 H489078 H489152 H489220 H489297 H489361 H489426 H488596 H488666 H488734 H488807 H488878 H488942 H489011 H489079 H489153 H489221 H489298 H489362 H489427 H488597 H488667 H488735 H488808 H488879 H488943 H489012 H489080 H489154 H489222 H489299 H489363 H489428 H488598 H488668 H488736 H488809 H488880 H488944 H489013 H489081 H489155 H489230 H489300 H489364 H489429 H488599 H488669 H488738 H488810 H488881 H488945 H489015 H489082 H489156 H489231 H489301 H489365 H489430 H488600 H488672 H488739 H488811 H488882 H488946 H489016 H489083 H489157 H489232 H489302 H489366 H489431 H488601 H488673 H488740 H488812 H488883 H488949 H489017 H489084 H489158 H489233 H489303 H489367 H489432 H488602 H488674 H488741 H488813 H488884 H488951 H489018 H489085 H489159 H489234 H489304 H489368 H489433 H488603 H488675 H488742 H488814 H488885 H488952 H489019 H489086 H489160 H489235 H489305 H489369 H489434 H488612 H488676 H488743 H488815 H488886 H488953 H489020 H489087 H489161 H489236 H489306 H489370 H489435 H488613 H488677 H488744 H488816 H488887 H488954 H489021 H489088 H489162 H489237 H489307 H489371 H489436 H488614 H488678 H488745 H488818 H488888 H488955 H489022 H489090 H489163 H489238 H489308 H489373 H489437 H488615 H488679 H488746 H488820 H488889 H488956 H489023 H489091 H489164 H489239 H489309 H489374 H489438 H488616 H488680 H488747 H488821 H488890 H488957 H489024 H489092 H489166 H489240 H489310 H489375 H489439 H488617 H488681 H488748 H488822 H488891 H488958 H489025 H489093 H489167 H489241 H489311 H489376 H489440 H488618 H488682 H488749 H488823 H488892 H488959 H489026 H489094 H489168 H489243 H489312 H489377 H489441 H488619 H488683 H488750 H488824 H488893 H488960 H489027 H489095 H489169 H489244 H489313 H489378 H489442 H488620 H488684 H488754 H488825 H488894 H488961 H489028 H489096 H489170 H489245 H489314 H489379 H489443 H488621 H488685 H488755 H488826 H488895 H488962 H489029 H489097 H489171 H489246 H489315 H489380 H489444 H488622 H488686 H488756 H488827 H488896 H488963 H489030 H489098 H489172 H489247 H489316 H489381 H489445 H488623 H488687 H488757 H488830 H488897 H488964 H489031 H489099 H489173 H489248 H489317 H489382 H489446 H488624 H488688 H488758 H488831 H488898 H488965 H489032 H489100 H489174 H489249 H489318 H489383 H489447 H488625 H488689 H488759 H488832 H488899 H488966 H489033 H489101 H489175 H489250 H489319 H489384 H489452 H488626 H488690 H488760 H488833 H488900 H488967 H489034 H489102 H489176 H489252 H489320 H489385 H489453 H488627 H488691 H488761 H488834 H488901 H488969 H489037 H489103 H489177 H489253 H489321 H489386 H489454 H488628 H488692 H488763 H488835 H488902 H488970 H489038 H489104 H489178 H489254 H489322 H489387 H489455 H488629 H488693 H488764 H488836 H488903 H488971 H489039 H489105 H489180 H489255 H489323 H489389 H489456 H488630 H488694 H488766 H488837 H488904 H488972 H489040 H489106 H489181 H489256 H489324 H489390 H489457 H488631 H488695 H488767 H488838 H488906 H488973 H489041 H489109 H489182 H489257 H489325 H489391 H489458 H488632 H488696 H488768 H488839 H488907 H488974 H489042 H489110 H489184 H489260 H489326 H489392 H489459 H488633 H488698 H488769 H488840 H488908 H488975 H489043 H489111 H489185 H489261 H489327 H489393 H489460 H488634 H488699 H488770 H488842 H488909 H488976 H489044 H489112 H489187 H489262 H489328 H489394 H489461 H488635 H488700 H488771 H488843 H488910 H488977 H489046 H489113 H489188 H489264 H489329 H489395 H489462 H488636 H488701 H488772 H488844 H488911 H488978 H489047 H489114 H489189 H489265 H489330 H489396 H489463 H488637 H488702 H488773 H488846 H488912 H488979 H489049 H489115 H489190 H489269 H489331 H489397 H489464 H488638 H488703 H488775 H488847 H488913 H488980 H489050 H489116 H489191 H489270 H489332 H489398 H489465 H488639 H488704 H488776 H488848 H488914 H488981 H489051 H489119 H489192 H489271 H489333 H489399 H489466 H488640 H488705 H488777 H488849 H488915 H488982 H489052 H489121 H489193 H489272 H489334 H489401 H489467 H488641 H488706 H488779 H488850 H488916 H488983 H489053 H489124 H489194 H489273 H489335 H489402 H489468 H488642 H488707 H488780 H488851 H488917 H488984 H489054 H489125 H489195 H489274 H489336 H489403 H489469 H488643 H488708 H488782 H488852 H488918 H488985 H489055 H489126 H489196 H489275 H489337 H489404 H489470 H488644 H488709 H488783 H488853 H488919 H488986 H489056 H489127 H489197 H489276 H489338 H489405 H489471 H488645 H488710 H488784 H488854 H488920 H488987 H489057 H489128 H489199 H489277 H489340 H489406 H489472 H488646 H488711 H488785 H488855 H488921 H488988 H489058 H489129 H489200 H489278 H489341 H489407 H489473 H488647 H488712 H488786 H488857 H488922 H488991 H489059 H489130 H489201 H489279 H489342 H489408 H489474 H488648 H488713 H488787 H488858 H488923 H488992 H489060 H489131 H489202 H489280 H489343 H489409 H489475 H488649 H488714 H488788 H488859 H488924 H488993 H489062 H489132 H489203 H489281 H489344 H489410 H489476 H488650 H488715 H488789 H488860 H488925 H488994 H489063 H489133 H489204 H489282 H489346 H489411 H489478 H488651 H488716 H488790 H488861 H488926 H488995 H489064 H489134 H489205 H489283 H489347 H489412 H489479 H488652 H488717 H488791 H488862 H488927 H488996 H489065 H489136 H489206 H489284 H489348 H489413 H489480 H488653 H488718 H488792 H488863 H488928 H488997 H489066 H489138 H489207 H489285 H489349 H489414 H489481 H488654 H488719 H488793 H488864 H488929 H488998 H489067 H489139 H489208 H489286 H489350 H489415 H489482 H488655 H488720 H488794 H488865 H488930 H488999 H489068 H489140 H489209 H489287 H489351 H489416 H489483 H488656 H488721 H488795 H488866 H488931 H489000 H489069 H489141 H489210 H489288 H489352 H489417 H489484 H488657 H488722 H488796 H488867 H488932 H489001 H489070 H489142 H489211 H489289 H489353 H489418 H489485 H488658 H488724 H488798 H488868 H488933 H489003 H489071 H489143 H489212 H489290 H489354 H489419 H489486 H488659 H488725 H488799 H488869 H488934 H489004 H489072 H489144 H489213 H489291 H489355 H489420 H489487 H488660 H488726 H488801 H488872 H488935 H489005 H489073 H489145 H489215 H489292 H489356 H489421 H489488 H488661 H488727 H488802 H488873 H488936 H489006 H489074 H489147 H489216 H489293 H489357 H489422 H489489 H488662 H488729 H488803 H488874 H488937 H489007 H489075 H489149 H489217 H489294 H489358 H489423 H489490 H489491 H489563 H489632 H489697 H489765 H489840 H489913 H489990 H48A060 H48A140 H48A224 H48A298 H489492 H489564 H489633 H489698 H489766 H489841 H489917 H489991 H48A061 H48A141 H48A225 H48A299 H489493 H489565 H489634 H489699 H489767 H489842 H489919 H489992 H48A062 H48A142 H48A226 H48A300 H489494 H489566 H489635 H489700 H489768 H489843 H489920 H489993 H48A063 H48A143 H48A229 H48A301 H489495 H489567 H489636 H489701 H489769 H489844 H489923 H489994 H48A064 H48A144 H48A230 H48A303 H489496 H489568 H489637 H489702 H489770 H489845 H489924 H489995 H48A065 H48A145 H48A231 H48A304 H489497 H489569 H489638 H489703 H489771 H489846 H489925 H489996 H48A066 H48A146 H48A232 H48A305 H489498 H489570 H489639 H489704 H489772 H489847 H489927 H489997 H48A068 H48A147 H48A233 H48A306 H489499 H489571 H489640 H489706 H489773 H489850 H489928 H489998 H48A069 H48A148 H48A234 H48A307 H489506 H489573 H489641 H489707 H489774 H489854 H489929 H489999 H48A070 H48A149 H48A235 H48A308 H489507 H489574 H489642 H489708 H489775 H489856 H489930 H48A000 H48A072 H48A150 H48A236 H48A309 H489508 H489575 H489643 H489709 H489776 H489857 H489931 H48A001 H48A073 H48A154 H48A237 H48A310 H489509 H489576 H489644 H489710 H489777 H489858 H489932 H48A002 H48A074 H48A155 H48A238 H48A312 H489510 H489577 H489645 H489711 H489778 H489859 H489933 H48A006 H48A075 H48A156 H48A239 H48A315 H489511 H489578 H489646 H489712 H489780 H489860 H489934 H48A008 H48A076 H48A157 H48A241 H48A316 H489512 H489579 H489647 H489713 H489781 H489861 H489935 H48A009 H48A077 H48A158 H48A243 H48A317 H489513 H489580 H489648 H489714 H489782 H489862 H489936 H48A010 H48A078 H48A159 H48A244 H48A318 H489514 H489581 H489649 H489715 H489783 H489863 H489937 H48A011 H48A079 H48A160 H48A245 H48A319 H489515 H489582 H489650 H489716 H489784 H489864 H489938 H48A012 H48A080 H48A161 H48A246 H48A320 H489516 H489583 H489651 H489717 H489785 H489865 H489939 H48A013 H48A081 H48A162 H48A254 H48A321 H489517 H489584 H489652 H489718 H489788 H489866 H489940 H48A014 H48A082 H48A163 H48A255 H48A322 H489518 H489585 H489653 H489719 H489789 H489867 H489943 H48A015 H48A083 H48A164 H48A256 H48A323 H489519 H489586 H489654 H489720 H489790 H489868 H489944 H48A016 H48A084 H48A165 H48A257 H48A325 H489520 H489587 H489655 H489721 H489791 H489869 H489945 H48A017 H48A085 H48A166 H48A258 H48A326 H489521 H489588 H489656 H489724 H489792 H489870 H489946 H48A018 H48A086 H48A170 H48A259 H48A328 H489522 H489589 H489657 H489725 H489793 H489871 H489947 H48A019 H48A087 H48A171 H48A260 H48A329 H489523 H489591 H489658 H489726 H489794 H489872 H489948 H48A020 H48A088 H48A172 H48A261 H48A330 H489524 H489592 H489659 H489727 H489795 H489873 H489950 H48A024 H48A089 H48A173 H48A262 H48A331 H489525 H489593 H489660 H489728 H489796 H489874 H489951 H48A025 H48A090 H48A174 H48A263 H48A332 H489526 H489594 H489661 H489729 H489797 H489875 H489952 H48A026 H48A091 H48A175 H48A264 H48A333 H489527 H489595 H489662 H489731 H489798 H489876 H489953 H48A027 H48A092 H48A176 H48A265 H48A334 H489528 H489596 H489663 H489732 H489799 H489877 H489954 H48A028 H48A093 H48A177 H48A266 H48A337 H489529 H489597 H489664 H489733 H489802 H489878 H489955 H48A029 H48A094 H48A178 H48A267 H48A339 H489530 H489598 H489665 H489734 H489803 H489879 H489956 H48A030 H48A096 H48A179 H48A268 H48A340 H489531 H489599 H489666 H489735 H489804 H489880 H489957 H48A032 H48A099 H48A184 H48A269 H48A341 H489532 H489600 H489667 H489736 H489805 H489881 H489958 H48A033 H48A100 H48A185 H48A270 H48A344 H489533 H489601 H489668 H489737 H489806 H489883 H489959 H48A034 H48A101 H48A186 H48A271 H48A345 H489534 H489602 H489669 H489738 H489807 H489884 H489962 H48A035 H48A102 H48A187 H48A272 H48A346 H489535 H489603 H489670 H489740 H489808 H489885 H489963 H48A036 H48A106 H48A188 H48A273 H48A348 H489536 H489604 H489671 H489741 H489809 H489886 H489964 H48A037 H48A107 H48A194 H48A274 H48A349 H489537 H489605 H489672 H489742 H489810 H489887 H489965 H48A038 H48A108 H48A195 H48A275 H48A351 H489538 H489606 H489673 H489743 H489811 H489888 H489966 H48A039 H48A109 H48A199 H48A276 H48A352 H489539 H489607 H489674 H489744 H489812 H489889 H489970 H48A040 H48A110 H48A201 H48A277 H48A354 H489540 H489612 H489675 H489745 H489813 H489890 H489971 H48A041 H48A111 H48A202 H48A278 H48A355 H489541 H489613 H489676 H489746 H489814 H489891 H489972 H48A042 H48A112 H48A203 H48A279 H48A356 H489542 H489614 H489677 H489747 H489815 H489892 H489973 H48A043 H48A113 H48A204 H48A280 H48A357 H489543 H489615 H489678 H489748 H489818 H489894 H489974 H48A044 H48A114 H48A205 H48A281 H48A358 H489544 H489616 H489679 H489749 H489819 H489895 H489975 H48A045 H48A115 H48A207 H48A282 H48A359 H489545 H489617 H489680 H489750 H489821 H489896 H489976 H48A046 H48A116 H48A209 H48A283 H48A360 H489546 H489619 H489681 H489751 H489822 H489897 H489977 H48A047 H48A117 H48A210 H48A284 H48A362 H489548 H489620 H489682 H489752 H489823 H489898 H489978 H48A048 H48A118 H48A211 H48A285 H48A363 H489549 H489621 H489683 H489753 H489825 H489900 H489979 H48A049 H48A119 H48A212 H48A286 H48A364 H489550 H489622 H489684 H489754 H489826 H489901 H489980 H48A050 H48A120 H48A213 H48A287 H48A365 H489552 H489623 H489685 H489755 H489829 H489903 H489981 H48A051 H48A124 H48A214 H48A288 H48A366 H489553 H489624 H489686 H489756 H489830 H489904 H489982 H48A052 H48A125 H48A215 H48A289 H48A367 H489554 H489625 H489687 H489757 H489831 H489905 H489983 H48A053 H48A126 H48A216 H48A290 H48A368 H489555 H489626 H489691 H489758 H489834 H489906 H489984 H48A054 H48A127 H48A218 H48A291 H48A369 H489558 H489627 H489692 H489760 H489835 H489908 H489985 H48A055 H48A128 H48A219 H48A292 H48A370 H489559 H489628 H489693 H489761 H489836 H489909 H489986 H48A056 H48A136 H48A220 H48A293 H48A371 H489560 H489629 H489694 H489762 H489837 H489910 H489987 H48A057 H48A137 H48A221 H48A294 H48A374 H489561 H489630 H489695 H489763 H489838 H489911 H489988 H48A058 H48A138 H48A222 H48A295 H48A375 H489562 H489631 H489696 H489764 H489839 H489912 H489989 H48A059 H48A139 H48A223 H48A296 H48A376 H48A377 H48A465 H48A538 H48A628 H48A703 H48A779 H48A846 H48A924 H48B003 H48B076 H48B174 H48B248 H48A378 H48A469 H48A539 H48A629 H48A704 H48A781 H48A847 H48A926 H48B004 H48B077 H48B175 H48B249 H48A379 H48A470 H48A541 H48A630 H48A705 H48A782 H48A849 H48A927 H48B005 H48B078 H48B176 H48B250 H48A380 H48A471 H48A550 H48A631 H48A706 H48A783 H48A852 H48A928 H48B006 H48B079 H48B177 H48B251 H48A381 H48A472 H48A551 H48A632 H48A707 H48A784 H48A853 H48A929 H48B007 H48B080 H48B178 H48B252 H48A382 H48A473 H48A552 H48A633 H48A708 H48A785 H48A854 H48A930 H48B008 H48B081 H48B179 H48B253 H48A383 H48A474 H48A553 H48A634 H48A709 H48A786 H48A856 H48A931 H48B010 H48B082 H48B180 H48B254 H48A384 H48A475 H48A554 H48A635 H48A710 H48A787 H48A857 H48A932 H48B011 H48B083 H48B181 H48B255 H48A385 H48A476 H48A555 H48A636 H48A711 H48A788 H48A858 H48A934 H48B012 H48B084 H48B182 H48B256 H48A386 H48A480 H48A556 H48A637 H48A712 H48A789 H48A859 H48A936 H48B013 H48B085 H48B183 H48B257 H48A387 H48A481 H48A557 H48A638 H48A713 H48A790 H48A863 H48A937 H48B014 H48B086 H48B184 H48B258 H48A388 H48A482 H48A560 H48A639 H48A714 H48A792 H48A864 H48A938 H48B015 H48B087 H48B185 H48B259 H48A390 H48A484 H48A561 H48A640 H48A715 H48A793 H48A865 H48A939 H48B017 H48B088 H48B186 H48B260 H48A391 H48A485 H48A562 H48A641 H48A716 H48A794 H48A866 H48A940 H48B018 H48B089 H48B187 H48B261 H48A392 H48A486 H48A563 H48A642 H48A717 H48A795 H48A867 H48A941 H48B019 H48B090 H48B188 H48B262 H48A393 H48A487 H48A564 H48A643 H48A718 H48A796 H48A868 H48A944 H48B020 H48B091 H48B189 H48B263 H48A395 H48A488 H48A565 H48A644 H48A719 H48A797 H48A869 H48A945 H48B022 H48B092 H48B190 H48B264 H48A398 H48A489 H48A566 H48A645 H48A720 H48A798 H48A870 H48A946 H48B023 H48B093 H48B191 H48B267 H48A405 H48A491 H48A567 H48A648 H48A721 H48A799 H48A871 H48A947 H48B024 H48B094 H48B192 H48B268 H48A406 H48A492 H48A568 H48A649 H48A722 H48A800 H48A872 H48A949 H48B026 H48B097 H48B193 H48B269 H48A407 H48A493 H48A569 H48A650 H48A724 H48A801 H48A873 H48A950 H48B027 H48B098 H48B196 H48B270 H48A408 H48A494 H48A570 H48A651 H48A725 H48A802 H48A874 H48A951 H48B028 H48B099 H48B198 H48B273 H48A409 H48A495 H48A571 H48A652 H48A726 H48A803 H48A880 H48A956 H48B029 H48B100 H48B199 H48B274 H48A410 H48A497 H48A572 H48A653 H48A727 H48A804 H48A881 H48A957 H48B030 H48B101 H48B200 H48B275 H48A411 H48A498 H48A573 H48A654 H48A728 H48A805 H48A882 H48A959 H48B031 H48B102 H48B201 H48B276 H48A412 H48A499 H48A574 H48A655 H48A729 H48A806 H48A883 H48A960 H48B032 H48B103 H48B202 H48B277 H48A413 H48A500 H48A575 H48A656 H48A730 H48A807 H48A884 H48A961 H48B033 H48B104 H48B203 H48B278 H48A414 H48A501 H48A576 H48A657 H48A731 H48A808 H48A885 H48A962 H48B034 H48B105 H48B204 H48B279 H48A415 H48A502 H48A577 H48A658 H48A732 H48A809 H48A886 H48A964 H48B035 H48B106 H48B205 H48B280 H48A416 H48A503 H48A578 H48A659 H48A733 H48A810 H48A887 H48A965 H48B036 H48B107 H48B206 H48B281 H48A417 H48A504 H48A587 H48A660 H48A734 H48A811 H48A888 H48A966 H48B037 H48B109 H48B207 H48B286 H48A422 H48A505 H48A588 H48A661 H48A735 H48A812 H48A889 H48A967 H48B038 H48B111 H48B210 H48B288 H48A424 H48A506 H48A589 H48A662 H48A736 H48A815 H48A890 H48A968 H48B039 H48B112 H48B213 H48B289 H48A425 H48A507 H48A590 H48A663 H48A737 H48A816 H48A893 H48A969 H48B040 H48B122 H48B214 H48B290 H48A426 H48A509 H48A591 H48A664 H48A747 H48A817 H48A894 H48A970 H48B041 H48B124 H48B215 H48B293 H48A436 H48A510 H48A592 H48A666 H48A748 H48A818 H48A895 H48A971 H48B042 H48B125 H48B216 H48B294 H48A437 H48A511 H48A593 H48A667 H48A749 H48A819 H48A896 H48A972 H48B043 H48B126 H48B217 H48B295 H48A438 H48A512 H48A594 H48A668 H48A750 H48A820 H48A897 H48A973 H48B044 H48B127 H48B218 H48B297 H48A439 H48A513 H48A595 H48A669 H48A751 H48A821 H48A898 H48A974 H48B046 H48B128 H48B219 H48B298 H48A440 H48A514 H48A596 H48A671 H48A752 H48A822 H48A899 H48A975 H48B048 H48B129 H48B220 H48B299 H48A441 H48A515 H48A597 H48A672 H48A753 H48A823 H48A900 H48A976 H48B049 H48B130 H48B221 H48B300 H48A442 H48A516 H48A598 H48A673 H48A754 H48A824 H48A901 H48A977 H48B050 H48B131 H48B222 H48B301 H48A443 H48A517 H48A599 H48A674 H48A755 H48A825 H48A902 H48A978 H48B053 H48B138 H48B223 H48B303 H48A444 H48A518 H48A600 H48A675 H48A756 H48A826 H48A903 H48A981 H48B054 H48B139 H48B224 H48B304 H48A445 H48A519 H48A601 H48A676 H48A759 H48A827 H48A905 H48A982 H48B055 H48B140 H48B226 H48B305 H48A446 H48A520 H48A604 H48A677 H48A760 H48A828 H48A907 H48A983 H48B056 H48B141 H48B227 H48B306 H48A447 H48A521 H48A605 H48A678 H48A762 H48A829 H48A908 H48A984 H48B057 H48B142 H48B229 H48B307 H48A448 H48A522 H48A607 H48A679 H48A763 H48A830 H48A909 H48A986 H48B059 H48B143 H48B231 H48B308 H48A449 H48A523 H48A610 H48A682 H48A764 H48A831 H48A910 H48A987 H48B060 H48B144 H48B232 H48B312 H48A450 H48A524 H48A612 H48A683 H48A765 H48A832 H48A911 H48A988 H48B061 H48B145 H48B233 H48B314 H48A451 H48A525 H48A614 H48A684 H48A766 H48A833 H48A912 H48A989 H48B062 H48B151 H48B234 H48B320 H48A452 H48A526 H48A616 H48A685 H48A767 H48A834 H48A913 H48A990 H48B063 H48B152 H48B235 H48B321 H48A454 H48A527 H48A617 H48A686 H48A769 H48A835 H48A914 H48A993 H48B064 H48B153 H48B236 H48B323 H48A455 H48A528 H48A618 H48A692 H48A770 H48A836 H48A915 H48A994 H48B065 H48B154 H48B237 H48B325 H48A456 H48A529 H48A619 H48A693 H48A771 H48A837 H48A916 H48A995 H48B066 H48B155 H48B238 H48B326 H48A458 H48A530 H48A620 H48A694 H48A772 H48A838 H48A917 H48A996 H48B067 H48B156 H48B241 H48B327 H48A459 H48A531 H48A621 H48A695 H48A773 H48A839 H48A918 H48A997 H48B068 H48B157 H48B242 H48B328 H48A460 H48A532 H48A622 H48A696 H48A774 H48A840 H48A919 H48A998 H48B069 H48B158 H48B243 H48B329 H48A461 H48A533 H48A624 H48A697 H48A775 H48A841 H48A920 H48A999 H48B070 H48B167 H48B244 H48B331 H48A462 H48A534 H48A625 H48A698 H48A776 H48A842 H48A921 H48B000 H48B071 H48B168 H48B245 H48B332 H48A463 H48A535 H48A626 H48A699 H48A777 H48A843 H48A922 H48B001 H48B074 H48B169 H48B246 H48B333 H48A464 H48A537 H48A627 H48A700 H48A778 H48A844 H48A923 H48B002 H48B075 H48B170 H48B247 H48B334 H48B335 H48B427 H48B526 H48B609 H48B703 H48B788 H48B865 H48B955 H48C071 H48C181 H48B336 H48B430 H48B527 H48B610 H48B704 H48B789 H48B866 H48B956 H48C072 H48C182 H48B337 H48B431 H48B528 H48B613 H48B705 H48B790 H48B867 H48B957 H48C082 H48C183 H48B338 H48B432 H48B531 H48B614 H48B707 H48B791 H48B868 H48B958 H48C084 H48C190 H48B339 H48B436 H48B533 H48B615 H48B708 H48B792 H48B869 H48B959 H48C085 H48C196 H48B340 H48B437 H48B534 H48B616 H48B709 H48B793 H48B870 H48B960 H48C086 H48C197 H48B341 H48B438 H48B537 H48B617 H48B712 H48B796 H48B873 H48B962 H48C087 H48C198 H48B342 H48B439 H48B538 H48B618 H48B713 H48B797 H48B874 H48B963 H48C088 H48C199 H48B343 H48B440 H48B539 H48B619 H48B714 H48B798 H48B875 H48B964 H48C089 H48C200 H48B344 H48B441 H48B542 H48B622 H48B715 H48B799 H48B886 H48B965 H48C095 H48C201 H48B345 H48B442 H48B543 H48B623 H48B718 H48B800 H48B887 H48B966 H48C096 H48C202 H48B346 H48B443 H48B544 H48B624 H48B719 H48B801 H48B891 H48B967 H48C097 H48C203 H48B347 H48B444 H48B545 H48B625 H48B720 H48B802 H48B892 H48B968 H48C098 H48C204 H48B348 H48B445 H48B546 H48B626 H48B721 H48B803 H48B893 H48B969 H48C099 H48C205 H48B353 H48B446 H48B547 H48B627 H48B722 H48B804 H48B894 H48B971 H48C100 H48C206 H48B354 H48B447 H48B549 H48B628 H48B723 H48B805 H48B895 H48B973 H48C101 H48C207 H48B360 H48B448 H48B550 H48B629 H48B724 H48B806 H48B896 H48B975 H48C102 H48C208 H48B361 H48B449 H48B551 H48B630 H48B725 H48B807 H48B897 H48B976 H48C103 H48C210 H48B362 H48B450 H48B552 H48B631 H48B726 H48B808 H48B898 H48B977 H48C104 H48C213 H48B363 H48B451 H48B553 H48B632 H48B727 H48B809 H48B899 H48B978 H48C105 H48C214 H48B364 H48B455 H48B554 H48B633 H48B728 H48B810 H48B900 H48B995 H48C114 H48C215 H48B365 H48B456 H48B555 H48B634 H48B729 H48B812 H48B901 H48B996 H48C115 H48C218 H48B366 H48B457 H48B556 H48B635 H48B730 H48B813 H48B906 H48C010 H48C116 H48C219 H48B370 H48B458 H48B557 H48B636 H48B731 H48B814 H48B907 H48C011 H48C118 H48C220 H48B371 H48B459 H48B559 H48B637 H48B732 H48B816 H48B908 H48C012 H48C120 H48C226 H48B372 H48B460 H48B560 H48B638 H48B733 H48B817 H48B909 H48C014 H48C121 H48C227 H48B373 H48B461 H48B561 H48B639 H48B734 H48B818 H48B910 H48C015 H48C122 H48C232 H48B374 H48B471 H48B570 H48B640 H48B738 H48B819 H48B912 H48C016 H48C123 H48C235 H48B375 H48B472 H48B571 H48B641 H48B739 H48B820 H48B913 H48C017 H48C124 H48C236 H48B376 H48B473 H48B572 H48B642 H48B740 H48B822 H48B914 H48C018 H48C125 H48C237 H48B377 H48B477 H48B573 H48B643 H48B741 H48B828 H48B915 H48C019 H48C126 H48C238 H48B378 H48B480 H48B574 H48B647 H48B744 H48B829 H48B916 H48C022 H48C127 H48B379 H48B481 H48B575 H48B648 H48B745 H48B830 H48B917 H48C024 H48C128 H48B380 H48B482 H48B576 H48B649 H48B746 H48B831 H48B918 H48C029 H48C129 H48B381 H48B483 H48B578 H48B650 H48B747 H48B832 H48B919 H48C030 H48C130 H48B382 H48B484 H48B579 H48B651 H48B748 H48B833 H48B926 H48C031 H48C131 H48B383 H48B485 H48B580 H48B652 H48B749 H48B834 H48B927 H48C032 H48C133 H48B384 H48B486 H48B581 H48B653 H48B750 H48B835 H48B928 H48C033 H48C141 H48B386 H48B487 H48B582 H48B654 H48B751 H48B836 H48B929 H48C034 H48C142 H48B387 H48B492 H48B583 H48B655 H48B752 H48B837 H48B930 H48C035 H48C143 H48B388 H48B494 H48B584 H48B657 H48B753 H48B838 H48B931 H48C036 H48C144 H48B389 H48B495 H48B585 H48B658 H48B755 H48B839 H48B932 H48C037 H48C145 H48B390 H48B496 H48B586 H48B659 H48B756 H48B840 H48B933 H48C038 H48C146 H48B391 H48B497 H48B587 H48B660 H48B757 H48B841 H48B934 H48C039 H48C147 H48B392 H48B498 H48B588 H48B661 H48B758 H48B842 H48B935 H48C042 H48C152 H48B394 H48B499 H48B589 H48B666 H48B759 H48B843 H48B936 H48C044 H48C153 H48B395 H48B500 H48B590 H48B667 H48B762 H48B844 H48B937 H48C045 H48C154 H48B399 H48B502 H48B591 H48B676 H48B763 H48B845 H48B938 H48C046 H48C155 H48B402 H48B504 H48B592 H48B687 H48B764 H48B846 H48B939 H48C047 H48C156 H48B407 H48B505 H48B593 H48B688 H48B765 H48B847 H48B941 H48C048 H48C163 H48B408 H48B506 H48B594 H48B689 H48B774 H48B848 H48B942 H48C049 H48C164 H48B409 H48B507 H48B595 H48B690 H48B775 H48B849 H48B943 H48C050 H48C165 H48B410 H48B511 H48B596 H48B691 H48B776 H48B850 H48B944 H48C054 H48C166 H48B411 H48B512 H48B598 H48B692 H48B777 H48B853 H48B945 H48C062 H48C167 H48B412 H48B513 H48B599 H48B693 H48B778 H48B854 H48B946 H48C063 H48C168 H48B416 H48B516 H48B600 H48B694 H48B780 H48B855 H48B947 H48C064 H48C169 H48B417 H48B518 H48B602 H48B695 H48B781 H48B856 H48B948 H48C065 H48C170 H48B418 H48B521 H48B603 H48B696 H48B782 H48B859 H48B949 H48C066 H48C173 H48B423 H48B522 H48B604 H48B697 H48B783 H48B860 H48B950 H48C067 H48C174 H48B424 H48B523 H48B605 H48B698 H48B784 H48B861 H48B951 H48C068 H48C177 H48B425 H48B524 H48B607 H48B699 H48B785 H48B862 H48B952 H48C069 H48C178 H48B426 H48B525 H48B608 H48B702 H48B787 H48B864 H48B954 H48C070 H48C179 ***12-12-12, AMENDED INFO ***: H48F464 H48A954 H48B404 H48F465 H48A955 H48A861 H48F466 H48C385 H48A862 H48F467 H48C386 H48A860 H48F468 H48F697 H48E833 H48F469 H48F698 H48F001 H48F470 H48F699 H48F003 H489400 H48FS16 H48A477 H48CS46 H48FS17 H48B113 H480218 H487787 H48B114 H48C216 H48B769 H48B115 H48C230 H48B770 H489556 H48C262 H488771 H489557 H48C263 H48B772 H484875 H48B940 H48B773 H485149 H480966 H48E837 H485150 H48E955 H48E838 H485523 H48E956 H48E121 H48B025 H489828 H48C719 H48B265 H48B073 H489002 H48B266 H48E868 H48C273 H484403 H48F004 H48C274 H484404 H48F184 H48C275 H483835 H48F185 H48C276 H484405 H48F478 H48F306 H484406 H48F479 H48CS98 H48E340 H48F657 H48B403 H48CS25 H48F658 H489882 H48CS26 H48F659 H48C847 H48CS27 H48F660 H48C848 H48A615 H48F661 H48F717 H48F096 H48F662 H48C968 H48F100 H48F663 H48A608 H48F150 H48F664 H48A609 H488778 H48F665 H480541 H48F420 H48F666 H480635 H487243 H48F667 H48B021 H484277 H48F668 H489146 H488883 H48F669 H488400 H488884 H48F670 H488401 H48F286 H48F671 H483116 H48E772 H48F672 H48C624 H48E773 H48F673 H48C625 H48E774 H480393 H48C626 H48E775 H488871 H48A848 H48E793 H488872 H480441 H487335 H48F177 H48A738 H48FS98 H48C721 H48C061 H48A739 H480922 H48C722 H48E679 H48A740 H48C090 H48C328 H48E907 H48A741 H48FS67 H48C329 H487856 H48A742 H489014 H48A419 H48FOOO H48A743 H488815 H48A420 H48F700 H48A744 H488058 H48A421 H48F701 H48A745 H489949 H488165 H48F702 H48A746 H489942 H488166 H48F703 H480255 H48E861 H488161 H48F704 H48F623 H48E959 H488162 H48F705 H48F624 H48A031 H488163 H48F706 H48F625 H489824 H488164 H48F707 H48F626 H486332 H488421 H48F708 H489848 H48A723 H488422 H487244 H489849 H488650 H48A373 H480399 H48C334 H488651 H48A372 H480400 H488350 H488652 H48F679 H48A496 H488351 H488653 H48F680 H48F638 H488352 H488654 H48F681 H48F639 H488310 H488655 H48F682 H48E199 H48C374 H488656 H480187 H48E200 H488974 H488657 H480192 H48E201 H489117 H488658 H48E210 H48E202 H489118 H488659 H48C355 H48E203 H489120 H488660 H485778 H48F106 H488315 H488661 H485779 H48F107 H488316 H488662 H485780 H48C682 H488317 H488663 H485781 H48C683 H48C868 H488664 H485782 H489833 H48C630 H488665 H485783 H48A958 H48E854 H488666 H485784 H480390 H487027 H488667 H485785 H480223 H48F655 H48A758 H485786 H48E586 H480359 H48A191 H485787 H487341 H480360 H48A189 H48A399 H487342 H48C369 H48A190 H48A400 H48E445 H48A508 H48A192 H48A401 H482242 H489258 H48A193 H48A402 H487280 H489259 H488762 H48A403 H487281 H48A247 H48E938 H48A404 H488171 H48A248 H48F341 H48A467 H488878 H48A249 H48F397 H48A468 H48B287 H48A250 H48CS69 H48C175 H480386 H48A251 H48C224 H48F601 H480387 H48A252 H480416 H48F602 H48F386 H480997 H48F320 H480998 H48F322 H48E064 H48F323 H48EOS6 H48F324 H48E230 H48A681 H48A603 H48C654 H48A602 H487552 H480848 H487553 H480849 H485414 H480984 H485415 H488347 H485416 H488348 H485417 H48A394 H48C623 H480546 H481827 H48C454 H48CS45 H480225 H48F267 H480440 H48FS40 H48ES46 H484960 H48B434 H484961 H48B435 H484962 H48B433 H484963 H48A130 H484964 H48A129 H488148 H48B123 H48C132 H486223 H48C426 H48B146 H48E391 H48F645 H48E392 H48F646 H48C376 H48C092 H48E898 H48C093 H48E899 H48C094 H48B009 H48C091 H48F640 H48F204 H48F641 H48F213 H48F642 H48F249 H48F643 H48E613 H48FS85 H48E756 H488856 H48E758 H48B322 H489926 H48C310 H480584 H48C187 H488253 H480119 H48A985 H485789 H48E800 H48F394 H488488 H48C079 H48B489 H48C080 H488490 H48C477 H485496 H485497 H485498 H485499 H485500 H485501 H485502 H485503 H485504 H485505 H485506 H485507 H485508 H485509 H485510 H485511 H485512 H485513 H485514 H485515 H48A687 H48A688 H48A689 H48A757 H480801 H48F709 H48C321 H48C322 H487265 H48C961 H48C962 H48C963 H48C964 H488318 H488319 H489832 H488271 H488272 H48A980 H48A979 H480053 H48C283 H48E676 H48C707 H48E255 H480521 H487336 H480663 H487421 H48F344 H481548 H48C081 H488491 H480122 H48F718 H487337 H480664 H48E585 H48F396 H481549 H48C923 H48F370 H480123 H48C209 H487338 H48FS06 H48F423 H48EOSS H480864 H48F091 H48F371 H480124 H48E816 H487339 H48FS07 H48C979 H480166 H48E672 H48F092 H48F372 H480125 H48C655 H48C605 H48FS08 H48C980 H48A167 H48C534 H48ES80 H48F373 H480126 H488355 H48C606 H48FS09 H48C391 H48A168 H48C370 H48F429 H48E572 H48COSS H480675 H488420 H48FS10 H48C392 H48A169 H48C371 H48F431 H48F381 H48COS6 H480676 H48F395 H48FS11 H48C331 H489820 H48C372 H483558 H48F382 H48B045 H480697 H488464 H48FS12 H48C332 H48C481 H48C269 H483559 H48F383 H48A948 H480702 H488465 H48FS13 H48C333 H48C621 H48FS92 H48B393 H48F384 H48A952 H480707 H488466 H48FS14 H48FS97 H48C622 H480388 H488716 H48F435 H48A953 H488405 H488467 H48FS15 H48A855 H48C720 H480389 H48B717 H48A623
Recalling Firm/
Manufacturer
Varian Medical Systems, Inc. Oncology Systems
911 Hansen Way
Palo Alto, California 94304-1028
For Additional Information Contact Michael Pignataro
650-424-6471
Manufacturer Reason
for Recall
In affected Eclipse versions, user editing of the MLC Leaf positions may not automatically invalidate calculated dose distributions.
FDA Determined
Cause 2
DESIGN: Software Design
Action Varian Medical Systems, Inc sent a "Urgent Medical Device Correction Urgent Field Safety Notice" letter dated May 13, 2011 to all affected customers. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter included the description of the product, problem and recommendations and includes the release of a service pack in the future. For additional information please contact the Varian Oncology Help Desk at 888-VARIAN5 (888- 827-4265).
Quantity in Commerce 10,097***12-12-12 AMENDED TO: 10664 units***
Distribution Worldwide distribution: Nationwide (USA) and the country of Canada.
Total Product Life Cycle TPLC Device Report

 
1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = MUJ and Original Applicant = VARIAN MEDICAL SYSTEMS, INC.
-
-