Date Initiated by Firm |
June 02, 2011 |
Date Posted |
September 27, 2011 |
Recall Status1 |
Terminated 3 on March 15, 2013 |
Recall Number |
Z-3281-2011 |
Recall Event ID |
59383 |
510(K)Number |
K033326
|
Product Classification |
System, X-ray, Tomography, computed - Product Code JAK
|
Product |
Brilliance 64 Computed Tomography System, Model #78231, Serial Number: 9592.
The Brilliance 64 is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at difference angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories. |
Code Information |
Model #78231, 510(k) #K03326, Serial Number: 9592. |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Road Cleveland OH 44143-2131
|
For Additional Information Contact |
440-483-7000
|
Manufacturer Reason for Recall |
Component defect. Philips Medical decided to conduct a recall field /correction (rework) of the mounting key locking threaded insert of the Detector Module system (DMS) bolt on one (1) Brilliance 64 CT X-ray unit. The firm initiated this action following Philip's receipt of the customer's complaint reporting that the retaining inserts were pulling out of the rotor during the removal of the UDMS
|
FDA Determined Cause 2 |
Other |
Action |
Philips sent a Field Change Order (FCO) 72800519 on June 13, 2011 to the affected customer. A Field Service Engineer visited the customer site and installed a hardware upgrade to the affected Brilliance 64 device system (Serial Number: 9592). The correction action was completed on August 9, 2011.
For questions regarding this recall call 1-440-483-7000. |
Quantity in Commerce |
1 unit |
Distribution |
Nationwide Distribution including PA |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JAK and Original Applicant = PHILIPS MEDICAL SYSTEMS, INC.
|