• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Medical & Radiation Emitting Device Recalls

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | Registration & Listing | Adverse Events | Recalls | PMA | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Recall
Animas 2020 Insulin Infusion Pump
see related information
Date Posted August 12, 2011
Recall Number Z-3017-2011
Product Animas 2020 Insulin Infusion Pump. Animas Corp. An insulin infusion.
Code Information Animas 2020 Insulin Pump Serial numbers 00-89133-12, 00-89207-12, 14-88703-12, 14-88710-12, 15-88704-12, 15-89190-12, 16-89096-12, 16-89191-12, 17-88720-12, 17-89097-12, 17-89185-12, 17-89611-12, 18-88721-12, 18-89612-12, 19-88715, 18-89099-12, 19-89187-12, 19-89194-12, 19-89620-12, 20-88716-12, 20-89188-12, 20-89621-12, 21-88717-12, 21-88731-12, 21-89196-12, 21-89615-12, 22-88732-12, 22-89616-12, 23-88733-12, 24-88734-12, 25-88728-12, 25-88735-12, 27-88737-12, 28-88738-12, 33-89824-12, 35-89819-12, 36-89827-12, 88-89100-12, 91-89110-12, 92-89111-12, 93-89112-12, 94-89106-12, 94-89120-12, 95-89107-12, 95-89114-12, 95-89121-12, 96-89115-12, 21-88386-12, 21-88393-12, 22-88387-12, 96-89122-12, 96-89210-12, 97-89116-12, 97-89123-12, 97-89130-12, 01-89215-12, 02-89216-12, 02-89223-12, 02-89230-12, 03-89217-12, 03-89224-12, 03-89400-12, 04-89218-12, 04-89225-12, 04-89232-12, 04-89401-12, 05-89219-12, 05-89226-12, 06-89227-12, 06-89234-12, 06-89241-12, 06-89410-12, 07-89235-12, 07-89242-12, 07-89404-12, 07-89411-12, 08-89236-12, 08-89412-12, 09-89237-12, 09-89406-12, 09-89413-12, 10-89238-12, 10-89414-12, 11-89239-12, 11-89408-12, 11-89415-12, 12-89409-12, 13-89417-12, 14-89418-12, 14-89520-12, 15-89521-12, 17-89509-12, 17-89516-12, 18-89517-12, 18-98524-12, 18-89531-12, 19-89525-12, 29-88672-12, 32-88668-12, 32-88682-12, 20-89533-12, 20-89540-12, 21-88643-12, 21-89541-12, 22-89528-12, 22-89535-12, 22-89542-12, 23-88638-12, 23-88740-12, 24-88741-12, 24-89544-12, 25-88647-12, 26-88655-12, 26-89553-12, 27-88751-12, 27-89385-12, 27-89554-12, 27-89561-12, 28-88745-12, 28-89555-12, 29-88760-12, 29-89556-12, 29-89563-12, 30-88747-12, 30-88754-12, 30-88761-12, 30-89395-12, 31-88755-12, 31-88762-12, 31-89558-12, 32-88756-12, 32-89397-12, 33-88757-12, 33-88771-12, 34-88758-12, 34-88772-12, 35-88759-12, 35-88766-12, 36-88774-12, 37-88768-12, 39-88777-12, 39-88939-12, 41-88948-12, 42-88787-12, 51-88979-12, 03-88414-12, 10-89245-12, 33-88690-12, 40-88697-12, 98-89131-12, 14-89256-12, 15-89257-12, 16-89258-12, 17-89273-12, 18-89267-12, 19-89444-12, 20-89269-12, 21-89460-12, 22-88901-12, 22-89461-12, 24-89463-12, 26-88905-12, 27-89466-12, 33-88852-12, 34-88853-12, 35-88854-12, 36-88855-12, 37-88856-12, 39-88872-12, 41-88874-12, 43-88869-12, 47-88887-12, 99-88403-12, 00-88404-12, 00-88411-12, 01-88412-12, 02-88406-12, 02-88413-12, 03-88407-12, 04-88408-12, 04-88415-12, 12-88370-12, 13-88371-12, 14-88372-12, 15-88373-12, 16-88367-12, 16-88381-12, 17-88375-12, 18-88376-12, 18-88383-12, 18-88390-12, 19-88377-12, 19-88384-12, 19-88391-12, 20-88378-12, 20-88385-12, 20-88392-12, 44-88789-12, 00-89140-12, 01-89141-12, 64-83146-12, 00-89221-12, 05-89240-12, 17-89266-12, 22-88394-12, 23-88388-12, 23-88395-12, 24-88389-12, 24-88396-12, 25-88397-12, 26-88398-12, 27-88399-12, 96-88400-12, 99-88410-12, 38-88695-12, 41-88698-12, 09-88603-12, 10-88604-12, 12-88606-12, 12-88613-12, 13-88614-12, 15-88616-12, 22-88637-12, 20-88642-12, 22-88644-12, 23-88645-12, 24-88646-12, 12-89180-12, 08-88602-12, 11-88605-12, 11-88612-12, 19-88627-12, 21-88629-12, 33-88588-12, 51-88993-12, 54-88989-12, 54-88996-12, 86-89003-12, 89-89013-12, 90-89007-12, 92-89016-12, 97-89028-12, 98-89029-12, 19-88641-12, 27-88656-12, 27-88663-12, 27-88670-12, 29-88658-12, 35-88678-12, 35-88685-12, 36-88679-12, 36-88686-12, 36-88950-12, 37-88951-12, 38-88952-12, 39-88953-12, 40-88947-12, 40-88954-12, 41-88955-12, 42-88949-12, 42-88963-12, 27-88920-12, 30-88923-12, 32-88932-12, 33-88933-12, 34-88927-12, 35-88928-12, 35-88935-12, 18-88802-12, 21-88805-12, 23-88807-12, 33-88838-12, 02-89135-12, 03-89136-12, 04-89144-12, 06-89160-12, 08-89148-12, 11-89077-12, 11-89158-12, 12-89159-12, 12-89173-12, 13-89167-12, 18-89179-12, 23-89293-12, 24-89294-12, 25-89295-12, 00-89126-12, 02-89128-12, 03-89129-12, 13-88702-12, 16-88705-12, 17-88706-12, 99-89125-12, 25-88742-12, 26-88743-12, 27-88744-12, 28-88752-12, 29-88739-12, 29-88746-12, 31-88748-12, 32-88749-12, 32-88763-12, 32-88770-12, 34-88765-12, 35-88773-12, 35-88780-12, 38-88769-12, 38-88776-12, 40-88778-12, 15-88380-12, 16-88374-12, 17-88368-12, 17-88382-12, 18-88369-12, 21-88379-12, 32-88851-12, 42-88868-12, 07-89330-12, 11-89334-12, 19-88803-12, 19-88810-12, 21-88812-12, 22-88820-12, 24-88808-12, 25-88809-12, 25-88816-12, 25-88830-12, 26-88831-12, 27-88818-12, 27-88832-12, 29-88827-12, 29-88834-12, 30-88828-12, 31-88829-12, 31-88836-12, 32-88837-12, 01-89303-12, 02-89142-12, 03-89150-12, 04-89151-12, 05-89138-12, 05-89145-12, 06-89139-12, 07-89161-12, 08-89155-12, 09-89149-12, 09-89156-12, 10-89083-12, 10-89090-12, 10-89157-12, 10-89171-12, 11-89165-12, 13-89174-12, 13-89181-12, 14-89094-12, 14-89168-12, 14-89175-12, 15-89176-12, 20-89276-12, 20-89290-12, 21-89277-12, 21-89291-12, 22-89278-12, 22-89292-12, 25-89288-12, 26-89289-12, 26-89296-12, 27-89297-12, 29-89299-12, 30-88930-12, 12-88701-12, 05-89233-12, 26-88750-12, 37-88775-12, 10-89164-12, 36-88693-12, 39-88696-12, 32-88675-12, 33-88676-12, and 30-88835-12.
Recalling Firm/
Manufacturer
Animas Corporation
200 Lawrence Dr
West Chester, Pennsylvania 19380-3428
Consumer Instructions Contact the recalling firm for information
For Additional Information Contact Jon Mulberg
877-937-7867 Ext. 1257
Reason for
Recall
Incorrect component (1nF capacitor) was used to produce 191 printed circuit boards used in six pump lots of which 380 pumps were distributed. This issue may result in the pump displaying a "Replace Battery" alarm not related to the condition of the battery.
Action Animas Corporation sent an "Urgent: Insulin Pump Recall" letter dated July, 29 20011 to all affected customers. The letter included affected product and problem. They requested anyone who received the pump to send it back to Animas and a replacement pump will be sent at no charge. For additional information contact the Customer Support Department at (877) 937-7867.
Quantity in Commerce 380
Distribution Worldwide Distribution: Nationwide (USA) distribution including the states of CA, CO, MA, and MD; and the countries of Australia, Finland, France, Israel, Italy, Sweden and the U.K.
 
-
-