|
Class 2 Device Recall Ultraview SL Command Module |
|
Date Initiated by Firm |
November 03, 2011 |
Date Posted |
January 11, 2012 |
Recall Status1 |
Terminated 3 on September 19, 2012 |
Recall Number |
Z-0690-2012 |
Recall Event ID |
60593 |
510(K)Number |
K103142
|
Product Classification |
Detector and alarm, arrhythmia - Product Code DSI
|
Product |
Ultraview SL Command Module is the Spacelabs Multi-parameter Module. It is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use.
The Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use. Physiological parameters that may be monitored include ECG with arrhythmia detection, respiration, invasive and noninvasive blood pressure, temperature, oxygen saturation (Sp02) and cardiac output. Acquired data may then be communicated to an information network for display, recording, editing and analysis.
|
Code Information |
Serial numbers: 146-049340, 146-049341, 146-049342, 146-049343, 146-049344, 146-049345, 146-049346, 1476-006611, 1496-000171, 1496-001678, 1496-002176, 1496-002407, 1496-002424, 1496-002427, 1496-002430, 1496-002432, 1496-002433, 1496-002438, 1496-002441, 1496-002443, 1496-002444, 1496-002445, 1496-002447, 1496-002448, 1496-002454, 1496-002455, 1496-002459, 1496-002480, 1496-002489, 1496-002492, 1496-002500, 1496-002507, 1496-002522, 1496-002523, 1496-002525, 1496-002533, 1496-002546, 1496-002548, 1496-002552, 1496-002553, 1496-002570, 1496-002572, 1496-002577, 1496-002579, 1496-002581, 1496-002582, 1496-002583, 1496-002585, 1496-002719, 1496-002720, 1496-002721, 1496-002722, 1496-002723, 1496-002724, 1496-002725, 1496-002726, 1496-002729, 1496-002731, 1496-002737, 1496-003065, 1496-003069, 1496-003070, 1496-003073, 1496-003158, 1496-003160, 1496-003161, 1496-003162, 1496-003163, 1496-003164, 1496-003168, 1496-003169, 1496-003170, 1496-003172, 1496-003173, 1496-003177, 1496-003178, 1496-003179, 1496-003181, 1496-003184, 1496-003187, 1496-003190, 1496-003191, 1496-003192, 1496-003210, 1496-003502, 1496-003503, 1496-003509, 1496-003512, 1496-003579, 1496-003602, 1496-003607, 1496-003739, 1496-003746, 1496-003794, 1496-003795, 1496-003796, 1496-003797, 1496-003798, 1496-003812, 1496-003813, 1496-003815, 1496-003816, 1496-003817, 1496-003818, 1496-003819, 1496-003821, 1496-003822, 1496-003823, 1496-003824, 1496-003825, 1496-004080, 1496-004095, 1496-004101, 1496-004221, 1496-004223, 1496-004302, 1496-004303, 1496-004381, 1496-004382, 1496-004384, 1496-004385, 1496-004386, 1496-004387, 1496-004388, 1496-004390, 1496-004392, 1496-004394, 1496-004622, 1496-004665, 1496-004666, 1496-004667, 1496-004669, 1496-004673, 1496-004674, 1496-004675, 1496-004676, 1496-004678, 1496-004679, 1496-004762, 1496-004885, 1496-004886, 1496-004888, 1496-004890, 1496-004892, 1496-005204, 1496-005342, 1496-005343, 1496-005368, 1496-005385, 1496-006036, 1496-006038, 1496-006048, 1496-006051, 1496-006054, 1496-006057, 1496-006058, 1496-006059, 1496-006062, 1496-006063, 1496-006065, 1496-006067, 1496-006069, 1496-006071, 1496-006075, 1496-006077, 1496-006079, 1496-006082, 1496-006083, 1496-006085, 1496-006086, 1496-006087, 1496-006088, 1496-006089, 1496-006090, 1496-006137, 1496-006145, 1496-006148, 1496-006264, 1496-006425, 1496-006440, 1496-006443, 1496-006445, 1496-006446, 1496-006457, 1496-006458, 1496-006601, 1496-006613, 1496-006969, 1496-006975, 1496-007096, 1496-007139, 1496-007376, 1496-007632, 1496-007633, 1496-007747, 1496-007751, 1496-007752, 1496-008241, 1496-008319, 1496-008370, 1496-008381, 1496-008503, 1496-008583, 1496-008586, 1496-008873, 1496-009676, 1496-009707, 1496-009711, 1496-009729, 1496-009733, 1496-009735, 1496-009737, 1496-009738, 1496-009860, 1496-009873, 1496-009887, 1496-010922, 1496-011010, 1496-011211, 1496-011212, 1496-011317, 1496-011552, 1496-011554, 1496-011555, 1496-011716, 1496-011874, 1496-011894, 1496-011902, 1496-011906, 1496-011908, 1496-011918, 1496-011923, 1496-011934, 1496-011959, 1496-011960, 1496-011961, 1496-011966, 1496-012343, 1496-012372, 1496-012691, 1496-012693, 1496-012707, 1496-012713, 1496-012731, 1496-013213, 1496-014700, 1496-014911, 1496-015421, 1496-015432, 1496-016499, 1496-016781, 1496-016804, 1496-016824, 1496-017795, 1496-018755, 1496-019690, 1496-020426, 1496-021303, 1496-021309, 1496-022147, 1496-022413, 1496-022651, 1496-022652, 1496-023129, 1496-023139, 1496-023718, 1496-025394, 1496-025895, 1496-027098, 1496-027104, 1496-027446, 1496-027449, 1496-029028, 1496-029038, 1496-029042, 1496-029054, 1496-029449, 1496-029981, 1496-031722, 1496-032023, 1496-032024, 1496-032025, 1496-032026, 1496-032027, 1496-032028, 1496-032029, 1496-032030, 1496-032031, 1496-032032, 1496-032033, 1496-032035, 1496-032036, 1496-032037, 1496-032038, 1496-032039, 1496-032040, 1496-032041, 1496-032433, 1496-032691, 1496-032693, 1496-032696, 1496-032700, 1496-032703, 1496-032704, 1496-032705, 1496-032706, 1496-032707, 1496-032709, 1496-032710, 1496-032712, 1496-032717, 1496-032718, 1496-032719, 1496-032720, 1496-032722, 1496-032723, 1496-032724, 1496-032725, 1496-032726, 1496-032727, 1496-032729, 1496-032730, 1496-032731, 1496-032732, 1496-032733, 1496-032734, 1496-032735, 1496-032736, 1496-032737, 1496-033281, 1496-033283, 1496-033284, 1496-033285, 1496-033294, 1496-033426, 1496-033769, 1496-033840, 1496-033840, 1496-034033, 1496-034557, 1496-034579, 1496-034584, 1496-035141, 1496-035296, 1496-035361, 1496-035426, 1496-035427, 1496-035906, 1496-036413, 1496-036786, 1496-037191, 1496-037461, 1496-037483, 1496-037806, 1496-038387, 1496-046474, 1496-046781, 1496-046782, 1496-046783, 1496-046784, 1496-046785, 1496-046921, 1496-046922, 1496-046923, 1496-046924, 1496-046925, 1496-046926, 1496-047006, 1496-047079, 1496-047080, 1496-047081, 1496-047082, 1496-047083, 1496-047086, 1496-047087, 1496-047088, 1496-047089, 1496-047090, 1496-047091, 1496-047121, 1496-047122, 1496-047145, 1496-047168, 1496-047169, 1496-047170, 1496-047171, 1496-047172, 1496-047173, 1496-047174, 1496-047175, 1496-047230, 1496-047231, 1496-047232, 1496-047233, 1496-047234, 1496-047235, 1496-047236, 1496-047237, 1496-047238, 1496-047239, 1496-047240, 1496-047241, 1496-047242, 1496-047243, 1496-047244, 1496-047245, 1496-047246, 1496-047247, 1496-047248, 1496-047249, 1496-047250, 1496-047251, 1496-047252, 1496-047253, 1496-047254, 1496-047255, 1496-047259, 1496-047260, 1496-047261, 1496-047262, 1496-047263, 1496-047264, 1496-047265, 1496-047266, 1496-047267, 1496-047268, 1496-047269, 1496-047270, 1496-047271, 1496-047272, 1496-047273, 1496-047274, 1496-047275, 1496-047276, 1496-047277, 1496-047278, 1496-047279, 1496-047280, 1496-047281, 1496-047282, 1496-047283, 1496-047284, 1496-047285, 1496-047286, 1496-047287, 1496-047288, 1496-047289, 1496-047290, 1496-047300, 1496-047301, 1496-047302, 1496-047303, 1496-047304, 1496-047305, 1496-047306, 1496-047307, 1496-047327, 1496-047356, 1496-047359, 1496-047399, 1496-047400, 1496-047401, 1496-047402, 1496-047403, 1496-047404, 1496-047405, 1496-047406, 1496-047407, 1496-047408, 1496-047410, 1496-047410, 1496-047413, 1496-047414, 1496-047415, 1496-047416, 1496-047417, 1496-047418, 1496-047419, 1496-047420, 1496-047421, 1496-047422, 1496-047423, 1496-047424, 1496-047425, 1496-047426, 1496-047427, 1496-047428, 1496-047429, 1496-047430, 1496-047431, 1496-047432, 1496-047433, 1496-047434, 1496-047435, 1496-047436, 1496-047437, 1496-047438, 1496-047439, 1496-047440, 1496-047441, 1496-047442, 1496-047443, 1496-047444, 1496-047445, 1496-047446, 1496-047447, 1496-047448, 1496-047449, 1496-047450, 1496-047451, 1496-047452, 1496-047453, 1496-047454, 1496-047455, 1496-047456, 1496-047457, 1496-047458, 1496-047459, 1496-047460, 1496-047461, 1496-047462, 1496-047463, 1496-047464, 1496-047465, 1496-047466, 1496-047467, 1496-047468, 1496-047469, 1496-047470, 1496-047471, 1496-047472, 1496-047473, 1496-047474, 1496-047475, 1496-047476, 1496-047477, 1496-047478, 1496-047479, 1496-047480, 1496-047481, 1496-047482, 1496-047483, 1496-047484, 1496-047485, 1496-047486, 1496-047487, 1496-047488, 1496-047489, 1496-047490, 1496-047491, 1496-047492, 1496-047493, 1496-047494, 1496-047495, 1496-047496, 1496-047497, 1496-047498, 1496-047499, 1496-047500, 1496-047501, 1496-047502, 1496-047503, 1496-047504, 1496-047505, 1496-047506, 1496-047507, 1496-047508, 1496-047509, 1496-047510, 1496-047511, 1496-047512, 1496-047513, 1496-047514, 1496-047515, 1496-047516, 1496-047517, 1496-047518, 1496-047519, 1496-047520, 1496-047521, 1496-047522, 1496-047523, 1496-047524, 1496-047525, 1496-047526, 1496-047527, 1496-047528, 1496-047529, 1496-047530, 1496-047531, 1496-047532, 1496-047533, 1496-047534, 1496-047535, 1496-047536, 1496-047537, 1496-047538, 1496-047539, 1496-047540, 1496-047541, 1496-047542, 1496-047543, 1496-047544, 1496-047545, 1496-047546, 1496-047547, 1496-047548, 1496-047549, 1496-047550, 1496-047551, 1496-047552, 1496-047553, 1496-047554, 1496-047555, 1496-047556, 1496-047557, 1496-047558, 1496-047559, 1496-047560, 1496-047561, 1496-047562, 1496-047563, 1496-047564, 1496-047565, 1496-047566, 1496-047567, 1496-047568, 1496-047569, 1496-047570, 1496-047571, 1496-047572, 1496-047573, 1496-047574, 1496-047575, 1496-047576, 1496-047577, 1496-047578, 1496-047579, 1496-047580, 1496-047581, 1496-047582, 1496-047583, 1496-047584, 1496-047585, 1496-047586, 1496-047587, 1496-047588, 1496-047589, 1496-047590, 1496-047591, 1496-047592, 1496-047593, 1496-047594, 1496-047595, 1496-047596, 1496-047597, 1496-047598, 1496-047599, 1496-047600, 1496-047601, 1496-047602, 1496-047603, 1496-047604, 1496-047605, 1496-047606, 1496-047607, 1496-047608, 1496-047609, 1496-047610, 1496-047611, 1496-047612, 1496-047613, 1496-047614, 1496-047615, 1496-047616, 1496-047617, 1496-047618, 1496-047619, 1496-047620, 1496-047621, 1496-047622, 1496-047623, 1496-047624, 1496-047625, 1496-047626, 1496-047627, 1496-047628, 1496-047629, 1496-047630, 1496-047631, 1496-047632, 1496-047633, 1496-047634, 1496-047635, 1496-047636, 1496-047637, 1496-047638, 1496-047639, 1496-047640, 1496-047641, 1496-047642, 1496-047643, 1496-047644, 1496-047645, 1496-047646, 1496-047647, 1496-047648, 1496-047649, 1496-047650, 1496-047651, 1496-047652, 1496-047653, 1496-047654, 1496-047655, 1496-047656, 1496-047657, 1496-047658, 1496-047659, 1496-047660, 1496-047661, 1496-047662, 1496-047663, 1496-047664, 1496-047665, 1496-047666, 1496-047667, 1496-047668, 1496-047669, 1496-047670, 1496-047671, 1496-047672, 1496-047673, 1496-047674, 1496-047675, 1496-047676, 1496-047677, 1496-047678, 1496-047679, 1496-047680, 1496-047681, 1496-047682, 1496-047683, 1496-047684, 1496-047685, 1496-047686, 1496-047687, 1496-047688, 1496-047689, 1496-047690, 1496-047691, 1496-047692, 1496-047693, 1496-047694, 1496-047695, 1496-047696, 1496-047697, 1496-047698, 1496-047699, 1496-047700, 1496-047701, 1496-047702, 1496-047703, 1496-047704, 1496-047705, 1496-047706, 1496-047707, 1496-047708, 1496-047709, 1496-047710, 1496-047711, 1496-047712, 1496-047713, 1496-047714, 1496-047715, 1496-047716, 1496-047717, 1496-047718, 1496-047719, 1496-047720, 1496-047721, 1496-047722, 1496-047723, 1496-047724, 1496-047725, 1496-047726, 1496-047727, 1496-047728, 1496-047729, 1496-047730, 1496-047731, 1496-047732, 1496-047733, 1496-047734, 1496-047735, 1496-047736, 1496-047737, 1496-047738, 1496-047739, 1496-047740, 1496-047741, 1496-047742, 1496-047743, 1496-047744, 1496-047745, 1496-047746, 1496-047747, 1496-047748, 1496-047749, 1496-047750, 1496-047751, 1496-047752, 1496-047753, 1496-047754, 1496-047755, 1496-047756, 1496-047757, 1496-047758, 1496-047759, 1496-047760, 1496-047761, 1496-047762, 1496-047763, 1496-047764, 1496-047765, 1496-047766, 1496-047767, 1496-047768, 1496-047769, 1496-047770, 1496-047771, 1496-047772, 1496-047773, 1496-047774, 1496-047775, 1496-047776, 1496-047777, 1496-047778, 1496-047779, 1496-047780, 1496-047781, 1496-047782, 1496-047783, 1496-047784, 1496-047785, 1496-047786, 1496-047787, 1496-047788, 1496-047789, 1496-047790, 1496-047791, 1496-047792, 1496-047793, 1496-047794, 1496-047795, 1496-047796, 1496-047797, 1496-047798, 1496-047799, 1496-047800, 1496-047801, 1496-047802, 1496-047803, 1496-047804, 1496-047805, 1496-047806, 1496-047807, 1496-047808, 1496-047809, 1496-047810, 1496-047811, 1496-047812, 1496-047813, 1496-047814, 1496-047815, 1496-047816, 1496-047817, 1496-047818, 1496-047819, 1496-047820, 1496-047821, 1496-047822, 1496-047823, 1496-047824, 1496-047825, 1496-047826, 1496-047827, 1496-047828, 1496-047829, 1496-047830, 1496-047831, 1496-047832, 1496-047833, 1496-047834, 1496-047835, 1496-047836, 1496-047837, 1496-047838, 1496-047839, 1496-047840, 1496-047841, 1496-047842, 1496-047843, 1496-047844, 1496-047845, 1496-047846, 1496-047847, 1496-047848, 1496-047849, 1496-047850, 1496-047851, 1496-047852, 1496-047853, 1496-047854, 1496-047855, 1496-047856, 1496-047857, 1496-047858, 1496-047859, 1496-047860, 1496-047861, 1496-047862, 1496-047863, 1496-047864, 1496-047865, 1496-047866, 1496-047867, 1496-047868, 1496-047869, 1496-047870, 1496-047871, 1496-047872, 1496-047873, 1496-047874, 1496-047875, 1496-047876, 1496-047877, 1496-047878, 1496-047879, 1496-047880, 1496-047881, 1496-047882, 1496-047883, 1496-047884, 1496-047885, 1496-047886, 1496-047887, 1496-047888, 1496-047889, 1496-047890, 1496-047891, 1496-047892, 1496-047893, 1496-047894, 1496-047895, 1496-047896, 1496-047897, 1496-047898, 1496-047899, 1496-047900, 1496-047901, 1496-047902, 1496-047903, 1496-047904, 1496-047905, 1496-047906, 1496-047907, 1496-047908, 1496-047909, 1496-047910, 1496-047911, 1496-047912, 1496-047913, 1496-047914, 1496-047915, 1496-047916, 1496-047917, 1496-047918, 1496-047919, 1496-047920, 1496-047921, 1496-047922, 1496-047923, 1496-047924, 1496-047925, 1496-047926, 1496-047927, 1496-047928, 1496-047929, 1496-047930, 1496-047931, 1496-047932, 1496-047933, 1496-047934, 1496-047935, 1496-047936, 1496-047937, 1496-047938, 1496-047939, 1496-047940, 1496-047941, 1496-047942, 1496-047943, 1496-047944, 1496-047945, 1496-047946, 1496-047947, 1496-047948, 1496-047949, 1496-047950, 1496-047951, 1496-047952, 1496-047953, 1496-047954, 1496-047955, 1496-047956, 1496-047957, 1496-047958, 1496-047959, 1496-047960, 1496-047961, 1496-047962, 1496-047963, 1496-047964, 1496-047965, 1496-047966, 1496-047967, 1496-047968, 1496-047969, 1496-047970, 1496-047971, 1496-047972, 1496-047973, 1496-047974, 1496-047975, 1496-047976, 1496-047977, 1496-047978, 1496-047979, 1496-047980, 1496-047981, 1496-047982, 1496-047983, 1496-047984, 1496-047985, 1496-047986, 1496-047987, 1496-047988, 1496-047989, 1496-047990, 1496-047991, 1496-047992, 1496-047993, 1496-047994, 1496-047995, 1496-047996, 1496-047997, 1496-047998, 1496-047999, 1496-048000, 1496-048001, 1496-048002, 1496-048003, 1496-048004, 1496-048005, 1496-048006, 1496-048007, 1496-048008, 1496-048009, 1496-048010, 1496-048011, 1496-048012, 1496-048013, 1496-048014, 1496-048015, 1496-048016, 1496-048017, 1496-048018, 1496-048019, 1496-048020, 1496-048021, 1496-048022, 1496-048023, 1496-048024, 1496-048025, 1496-048026, 1496-048027, 1496-048028, 1496-048029, 1496-048030, 1496-048031, 1496-048032, 1496-048033, 1496-048034, 1496-048035, 1496-048036, 1496-048037, 1496-048038, 1496-048039, 1496-048040, 1496-048041, 1496-048042, 1496-048043, 1496-048044, 1496-048045, 1496-048046, 1496-048047, 1496-048048, 1496-048049, 1496-048050, 1496-048051, 1496-048052, 1496-048053, 1496-048054, 1496-048055, 1496-048056, 1496-048057, 1496-048058, 1496-048059, 1496-048060, 1496-048061, 1496-048062, 1496-048063, 1496-048064, 1496-048065, 1496-048066, 1496-048067, 1496-048068, 1496-048069, 1496-048070, 1496-048071, 1496-048072, 1496-048073, 1496-048074, 1496-048075, 1496-048076, 1496-048077, 1496-048078, 1496-048079, 1496-048080, 1496-048081, 1496-048082, 1496-048083, 1496-048084, 1496-048085, 1496-048086, 1496-048087, 1496-048088, 1496-048089, 1496-048090, 1496-048091, 1496-048092, 1496-048093, 1496-048094, 1496-048095, 1496-048096, 1496-048097, 1496-048098, 1496-048099, 1496-048100, 1496-048101, 1496-048102, 1496-048103, 1496-048104, 1496-048105, 1496-048106, 1496-048107, 1496-048108, 1496-048109, 1496-048110, 1496-048111, 1496-048112, 1496-048113, 1496-048114, 1496-048115, 1496-048116, 1496-048117, 1496-048118, 1496-048119, 1496-048120, 1496-048121, 1496-048122, 1496-048123, 1496-048124, 1496-048125, 1496-048126, 1496-048127, 1496-048128, 1496-048129, 1496-048130, 1496-048131, 1496-048132, 1496-048133, 1496-048134, 1496-048135, 1496-048136, 1496-048137, 1496-048138, 1496-048139, 1496-048140, 1496-048141, 1496-048142, 1496-048143, 1496-048144, 1496-048145, 1496-048146, 1496-048147, 1496-048148, 1496-048149, 1496-048150, 1496-048151, 1496-048152, 1496-048153, 1496-048154, 1496-048155, 1496-048156, 1496-048157, 1496-048158, 1496-048159, 1496-048160, 1496-048161, 1496-048162, 1496-048163, 1496-048164, 1496-048165, 1496-048166, 1496-048167, 1496-048168, 1496-048169, 1496-048170, 1496-048171, 1496-048172, 1496-048173, 1496-048174, 1496-048175, 1496-048176, 1496-048177, 1496-048178, 1496-048179, 1496-048180, 1496-048181, 1496-048182, 1496-048183, 1496-048184, 1496-048185, 1496-048186, 1496-048187, 1496-048188, 1496-048189, 1496-048190, 1496-048191, 1496-048192, 1496-048193, 1496-048194, 1496-048195, 1496-048196, 1496-048197, 1496-048198, 1496-048199, 1496-048200, 1496-048201, 1496-048202, 1496-048203, 1496-048204, 1496-048205, 1496-048206, 1496-048207, 1496-048208, 1496-048209, 1496-048210, 1496-048211, 1496-048212, 1496-048213, 1496-048214, 1496-048215, 1496-048216, 1496-048217, 1496-048218, 1496-048219, 1496-048220, 1496-048221, 1496-048222, 1496-048223, 1496-048224, 1496-048225, 1496-048226, 1496-048227, 1496-048228, 1496-048229, 1496-048230, 1496-048231, 1496-048232, 1496-048233, 1496-048234, 1496-048235, 1496-048236, 1496-048237, 1496-048238, 1496-048239, 1496-048240, 1496-048241, 1496-048242, 1496-048243, 1496-048244, 1496-048245, 1496-048246, 1496-048247, 1496-048248, 1496-048249, 1496-048250, 1496-048251, 1496-048252, 1496-048253, 1496-048254, 1496-048255, 1496-048256, 1496-048257, 1496-048258, 1496-048259, 1496-048260, 1496-048261, 1496-048262, 1496-048263, 1496-048264, 1496-048265, 1496-048266, 1496-048267, 1496-048268, 1496-048269, 1496-048270, 1496-048271, 1496-048272, 1496-048273, 1496-048274, 1496-048275, 1496-048276, 1496-048277, 1496-048278, 1496-048279, 1496-048280, 1496-048281, 1496-048282, 1496-048283, 1496-048284, 1496-048285, 1496-048286, 1496-048287, 1496-048288, 1496-048289, 1496-048290, 1496-048291, 1496-048292, 1496-048293, 1496-048294, 1496-048295, 1496-048296, 1496-048297, 1496-048298, 1496-048299, 1496-048300, 1496-048301, 1496-048302, 1496-048303, 1496-048304, 1496-048305, 1496-048306, 1496-048307, 1496-048308, 1496-048309, 1496-048310, 1496-048311, 1496-048312, 1496-048313, 1496-048314, 1496-048315, 1496-048316, 1496-048317, 1496-048318, 1496-048319, 1496-048320, 1496-048321, 1496-048322, 1496-048323, 1496-048324, 1496-048325, 1496-048326, 1496-048327, 1496-048328, 1496-048329, 1496-048330, 1496-048331, 1496-048332, 1496-048333, 1496-048334, 1496-048335, 1496-048336, 1496-048337, 1496-048338, 1496-048339, 1496-048340, 1496-048341, 1496-048342, 1496-048343, 1496-048344, 1496-048345, 1496-048346, 1496-048347, 1496-048348, 1496-048349, 1496-048350, 1496-048351, 1496-048352, 1496-048353, 1496-048354, 1496-048355, 1496-048356, 1496-048357, 1496-048358, 1496-048359, 1496-048360, 1496-048361, 1496-048362, 1496-048363, 1496-048364, 1496-048365, 1496-048366, 1496-048367, 1496-048368, 1496-048369, 1496-048370, 1496-048371, 1496-048372, 1496-048373, 1496-048374, 1496-048375, 1496-048376, 1496-048377, 1496-048378, 1496-048379, 1496-048380, 1496-048381, 1496-048382, 1496-048383, 1496-048384, 1496-048385, 1496-048386, 1496-048387, 1496-048388, 1496-048389, 1496-048390, 1496-048391, 1496-048392, 1496-048393, 1496-048394, 1496-048395, 1496-048396, 1496-048397, 1496-048398, 1496-048399, 1496-048400, 1496-048401, 1496-048402, 1496-048403, 1496-048404, 1496-048405, 1496-048406, 1496-048407, 1496-048408, 1496-048409, 1496-048410, 1496-048411, 1496-048412, 1496-048413, 1496-048414, 1496-048415, 1496-048416, 1496-048417, 1496-048418, 1496-048419, 1496-048420, 1496-048421, 1496-048422, 1496-048423, 1496-048424, 1496-048425, 1496-048426, 1496-048427, 1496-048428, 1496-048429, 1496-048430, 1496-048431, 1496-048432, 1496-048433, 1496-048434, 1496-048435, 1496-048436, 1496-048437, 1496-048438, 1496-048439, 1496-048440, 1496-048441, 1496-048442, 1496-048443, 1496-048444, 1496-048445, 1496-048457, 1496-048458, 1496-048459, 1496-048475, 1496-048476, 1496-048477, 1496-048478, 1496-048479, 1496-048480, 1496-048481, 1496-048482, 1496-048483, 1496-048484, 1496-048485, 1496-048486, 1496-048487, 1496-048488, 1496-048489, 1496-048495, 1496-048496, 1496-048497, 1496-048498, 1496-048499, 1496-048501, 1496-048502, 1496-048503, 1496-048504, 1496-048505, 1496-048506, 1496-048507, 1496-048508, 1496-048509, 1496-048510, 1496-048511, 1496-048512, 1496-048513, 1496-048514, 1496-048515, 1496-048516, 1496-048517, 1496-048518, 1496-048519, 1496-048520, 1496-048521, 1496-048522, 1496-048523, 1496-048524, 1496-048525, 1496-048526, 1496-048527, 1496-048528, 1496-048529, 1496-048530, 1496-048531, 1496-048532, 1496-048533, 1496-048534, 1496-048535, 1496-048536, 1496-048537, 1496-048538, 1496-048539, 1496-048540, 1496-048541, 1496-048542, 1496-048543, 1496-048544, 1496-048545, 1496-048546, 1496-048547, 1496-048548, 1496-048549, 1496-048550, 1496-048551, 1496-048552, 1496-048553, 1496-048554, 1496-048555, 1496-048556, 1496-048557, 1496-048558, 1496-048559, 1496-048560, 1496-048561, 1496-048562, 1496-048563, 1496-048564, 1496-048565, 1496-048566, 1496-048567, 1496-048568, 1496-048569, 1496-048570, 1496-048571, 1496-048572, 1496-048573, 1496-048574, 1496-048575, 1496-048576, 1496-048577, 1496-048578, 1496-048579, 1496-048580, 1496-048581, 1496-048582, 1496-048583, 1496-048584, 1496-048585, 1496-048586, 1496-048587, 1496-048588, 1496-048589, 1496-048590, 1496-048591, 1496-048592, 1496-048593, 1496-048594, 1496-048595, 1496-048596, 1496-048597, 1496-048598, 1496-048599, 1496-048600, 1496-048601, 1496-048602, 1496-048603, 1496-048605, 1496-048606, 1496-048607, 1496-048608, 1496-048609, 1496-048610, 1496-048611, 1496-048612, 1496-048613, 1496-048614, 1496-048615, 1496-048616, 1496-048617, 1496-048618, 1496-048619, 1496-048620, 1496-048621, 1496-048622, 1496-048623, 1496-048624, 1496-048625, 1496-048626, 1496-048627, 1496-048628, 1496-048629, 1496-048630, 1496-048631, 1496-048632, 1496-048633, 1496-048634, 1496-048635, 1496-048637, 1496-048638, 1496-048639, 1496-048640, 1496-048641, 1496-048642, 1496-048643, 1496-048644, 1496-048645, 1496-048646, 1496-048647, 1496-048648, 1496-048649, 1496-048650, 1496-048651, 1496-048652, 1496-048653, 1496-048654, 1496-048655, 1496-048656, 1496-048657, 1496-048658, 1496-048659, 1496-048660, 1496-048661, 1496-048662, 1496-048663, 1496-048664, 1496-048665, 1496-048666, 1496-048667, 1496-048668, 1496-048669, 1496-048670, 1496-048671, 1496-048672, 1496-048673, 1496-048674, 1496-048675, 1496-048676, 1496-048677, 1496-048678, 1496-048679, 1496-048680, 1496-048681, 1496-048682, 1496-048683, 1496-048684, 1496-048685, 1496-048686, 1496-048687, 1496-048688, 1496-048689, 1496-048690, 1496-048691, 1496-048692, 1496-048693, 1496-048694, 1496-048695, 1496-048696, 1496-048697, 1496-048698, 1496-048699, 1496-048700, 1496-048701, 1496-048702, 1496-048703, 1496-048704, 1496-048705, 1496-048706, 1496-048707, 1496-048708, 1496-048709, 1496-048710, 1496-048711, 1496-048712, 1496-048713, 1496-048714, 1496-048715, 1496-048716, 1496-048717, 1496-048718, 1496-048719, 1496-048720, 1496-048721, 1496-048722, 1496-048723, 1496-048724, 1496-048725, 1496-048726, 1496-048727, 1496-048728, 1496-048729, 1496-048730, 1496-048731, 1496-048732, 1496-048733, 1496-048734, 1496-048735, 1496-048736, 1496-048737, 1496-048738, 1496-048739, 1496-048740, 1496-048741, 1496-048742, 1496-048743, 1496-048744, 1496-048745, 1496-048746, 1496-048747, 1496-048748, 1496-048749, 1496-048750, 1496-048751, 1496-048752, 1496-048753, 1496-048754, 1496-048755, 1496-048756, 1496-048757, 1496-048758, 1496-048759, 1496-048760, 1496-048761, 1496-048762, 1496-048763, 1496-048764, 1496-048765, 1496-048766, 1496-048767, 1496-048768, 1496-048769, 1496-048770, 1496-048771, 1496-048772, 1496-048773, 1496-048774, 1496-048775, 1496-048776, 1496-048777, 1496-048778, 1496-048779, 1496-048780, 1496-048781, 1496-048782, 1496-048783, 1496-048784, 1496-048785, 1496-048786, 1496-048787, 1496-048788, 1496-048789, 1496-048790, 1496-048791, 1496-048792, 1496-048793, 1496-048794, 1496-048795, 1496-048796, 1496-048797, 1496-048798, 1496-048799, 1496-048800, 1496-048801, 1496-048802, 1496-048803, 1496-048804, 1496-048805, 1496-048806, 1496-048807, 1496-048808, 1496-048809, 1496-048810, 1496-048811, 1496-048812, 1496-048813, 1496-048814, 1496-048815, 1496-048816, 1496-048817, 1496-048818, 1496-048819, 1496-048820, 1496-048821, 1496-048822, 1496-048823, 1496-048824, 1496-048825, 1496-048826, 1496-048827, 1496-048828, 1496-048829, 1496-048830, 1496-048831, 1496-048832, 1496-048833, 1496-048834, 1496-048835, 1496-048836, 1496-048837, 1496-048838, 1496-048839, 1496-048840, 1496-048841, 1496-048842, 1496-048843, 1496-048844, 1496-048845, 1496-048846, 1496-048847, 1496-048848, 1496-048849, 1496-048850, 1496-048851, 1496-048852, 1496-048853, 1496-048854, 1496-048855, 1496-048856, 1496-048857, 1496-048858, 1496-048859, 1496-048860, 1496-048861, 1496-048862, 1496-048863, 1496-048864, 1496-048865, 1496-048866, 1496-048867, 1496-048868, 1496-048869, 1496-048870, 1496-048872, 1496-048873, 1496-048874, 1496-048875, 1496-048876, 1496-048877, 1496-048878, 1496-048879, 1496-048880, 1496-048881, 1496-048882, 1496-048883, 1496-048884, 1496-048885, 1496-048886, 1496-048887, 1496-048888, 1496-048889, 1496-048890, 1496-048891, 1496-048892, 1496-048893, 1496-048894, 1496-048895, 1496-048896, 1496-048897, 1496-048898, 1496-048899, 1496-048900, 1496-048901, 1496-048902, 1496-048903, 1496-048904, 1496-048905, 1496-048906, 1496-048907, 1496-048908, 1496-048909, 1496-048910, 1496-048911, 1496-048912, 1496-048913, 1496-048914, 1496-048915, 1496-048916, 1496-048917, 1496-048918, 1496-048924, 1496-048925, 1496-048926, 1496-048927, 1496-048928, 1496-048929, 1496-048930, 1496-048931, 1496-048932, 1496-048933, 1496-048934, 1496-048935, 1496-048936, 1496-048937, 1496-048938, 1496-048939, 1496-048940, 1496-048941, 1496-048942, 1496-048943, 1496-048944, 1496-048945, 1496-048946, 1496-048947, 1496-048948, 1496-048951, 1496-048952, 1496-048953, 1496-048954, 1496-048955, 1496-048956, 1496-048957, 1496-048958, 1496-048959, 1496-048960, 1496-048961, 1496-048962, 1496-048963, 1496-048964, 1496-048965, 1496-048966, 1496-048967, 1496-048968, 1496-048969, 1496-048970, 1496-048971, 1496-048972, 1496-048973, 1496-048974, 1496-048975, 1496-048976, 1496-048977, 1496-048978, 1496-048979, 1496-048980, 1496-048981, 1496-048982, 1496-048983, 1496-048984, 1496-048985, 1496-048986, 1496-048987, 1496-048988, 1496-048989, 1496-048990, 1496-048991, 1496-048992, 1496-048993, 1496-048994, 1496-048995, 1496-048996, 1496-048997, 1496-048998, 1496-048999, 1496-049000, 1496-049001, 1496-049002, 1496-049003, 1496-049004, 1496-049005, 1496-049006, 1496-049007, 1496-049008, 1496-049009, 1496-049010, 1496-049011, 1496-049012, 1496-049013, 1496-049014, 1496-049015, 1496-049016, 1496-049017, 1496-049018, 1496-049019, 1496-049020, 1496-049021, 1496-049022, 1496-049023, 1496-049024, 1496-049025, 1496-049026, 1496-049027, 1496-049028, 1496-049029, 1496-049030, 1496-049031, 1496-049032, 1496-049033, 1496-049034, 1496-049035, 1496-049036, 1496-049037, 1496-049038, 1496-049039, 1496-049040, 1496-049041, 1496-049042, 1496-049043, 1496-049044, 1496-049045, 1496-049046, 1496-049047, 1496-049048, 1496-049049, 1496-049050, 1496-049051, 1496-049052, 1496-049053, 1496-049054, 1496-049055, 1496-049056, 1496-049057, 1496-049058, 1496-049059, 1496-049060, 1496-049061, 1496-049062, 1496-049063, 1496-049064, 1496-049065, 1496-049066, 1496-049067, 1496-049068, 1496-049069, 1496-049070, 1496-049071, 1496-049072, 1496-049073, 1496-049074, 1496-049075, 1496-049076, 1496-049077, 1496-049078, 1496-049079, 1496-049080, 1496-049081, 1496-049082, 1496-049083, 1496-049084, 1496-049085, 1496-049086, 1496-049087, 1496-049088, 1496-049089, 1496-049090, 1496-049091, 1496-049092, 1496-049093, 1496-049094, 1496-049095, 1496-049096, 1496-049097, 1496-049098, 1496-049099, 1496-049100, 1496-049101, 1496-049102, 1496-049103, 1496-049104, 1496-049105, 1496-049106, 1496-049107, 1496-049108, 1496-049109, 1496-049110, 1496-049111, 1496-049112, 1496-049113, 1496-049114, 1496-049115, 1496-049116, 1496-049117, 1496-049118, 1496-049119, 1496-049120, 1496-049121, 1496-049122, 1496-049123, 1496-049124, 1496-049125, 1496-049126, 1496-049127, 1496-049128, 1496-049129, 1496-049130, 1496-049131, 1496-049132, 1496-049133, 1496-049134, 1496-049135, 1496-049136, 1496-049137, 1496-049138, 1496-049139, 1496-049140, 1496-049141, 1496-049142, 1496-049143, 1496-049144, 1496-049145, 1496-049146, 1496-049147, 1496-049148, 1496-049149, 1496-049150, 1496-049151, 1496-049152, 1496-049153, 1496-049154, 1496-049155, 1496-049156, 1496-049157, 1496-049158, 1496-049159, 1496-049160, 1496-049161, 1496-049162, 1496-049163, 1496-049164, 1496-049166, 1496-049167, 1496-049168, 1496-049169, 1496-049170, 1496-049171, 1496-049172, 1496-049173, 1496-049174, 1496-049175, 1496-049176, 1496-049177, 1496-049178, 1496-049179, 1496-049180, 1496-049181, 1496-049182, 1496-049183, 1496-049184, 1496-049185, 1496-049186, 1496-049187, 1496-049188, 1496-049189, 1496-049190, 1496-049191, 1496-049192, 1496-049193, 1496-049194, 1496-049195, 1496-049196, 1496-049197, 1496-049198, 1496-049199, 1496-049200, 1496-049201, 1496-049202, 1496-049203, 1496-049204, 1496-049205, 1496-049206, 1496-049207, 1496-049208, 1496-049209, 1496-049210, 1496-049211, 1496-049212, 1496-049213, 1496-049214, 1496-049215, 1496-049216, 1496-049217, 1496-049218, 1496-049219, 1496-049220, 1496-049221, 1496-049222, 1496-049223, 1496-049224, 1496-049225, 1496-049226, 1496-049227, 1496-049228, 1496-049229, 1496-049230, 1496-049231, 1496-049232, 1496-049233, 1496-049234, 1496-049235, 1496-049236, 1496-049237, 1496-049238, 1496-049239, 1496-049240, 1496-049241, 1496-049242, 1496-049243, 1496-049244, 1496-049245, 1496-049246, 1496-049247, 1496-049248, 1496-049249, 1496-049250, 1496-049251, 1496-049252, 1496-049253, 1496-049254, 1496-049255, 1496-049256, 1496-049257, 1496-049258, 1496-049259, 1496-049260, 1496-049261, 1496-049262, 1496-049263, 1496-049264, 1496-049265, 1496-049266, 1496-049267, 1496-049268, 1496-049269, 1496-049270, 1496-049271, 1496-049272, 1496-049273, 1496-049274, 1496-049275, 1496-049276, 1496-049277, 1496-049278, 1496-049279, 1496-049280, 1496-049281, 1496-049282, 1496-049283, 1496-049284, 1496-049285, 1496-049286, 1496-049287, 1496-049288, 1496-049289, 1496-049290, 1496-049291, 1496-049292, 1496-049293, 1496-049294, 1496-049295, 1496-049296, 1496-049297, 1496-049298, 1496-049299, 1496-049300, 1496-049301, 1496-049302, 1496-049303, 1496-049304, 1496-049305, 1496-049306, 1496-049307, 1496-049308, 1496-049309, 1496-049310, 1496-049311, 1496-049312, 1496-049313, 1496-049314, 1496-049315, 1496-049316, 1496-049317, 1496-049318, 1496-049319, 1496-049320, 1496-049321, 1496-049322, 1496-049323, 1496-049324, 1496-049325, 1496-049326, 1496-049327, 1496-049328, 1496-049329, 1496-049330, 1496-049331, 1496-049332, 1496-049333, 1496-049334, 1496-049335, 1496-049336, 1496-049337, 1496-049338, 1496-049339, 1496-049347, 1496-049348, 1496-049349, 1496-049350, 1496-049351, 1496-049352, 1496-049353, 1496-049354, 1496-049355, 1496-049356, 1496-049357, 1496-049358, 1496-049359, 1496-049360, 1496-049361, 1496-049362, 1496-049363, 1496-049364, 1496-049365, 1496-049366, 1496-049367, 1496-049368, 1496-049369, 1496-049370, 1496-049371, 1496-049372, 1496-049373, 1496-049374, 1496-049375, 1496-049376, 1496-049377, 1496-049378, 1496-049379, 1496-049380, 1496-049381, 1496-049382, 1496-049383, 1496-049384, 1496-049385, 1496-049386, 1496-049387, 1496-049388, 1496-049389, 1496-049390, 1496-049391, 1496-049392, 1496-049393, 1496-049394, 1496-049395, 1496-049396, 1496-049397, 1496-049398, 1496-049399, 1496-049400, 1496-049401, 1496-049402, 1496-049403, 1496-049404, 1496-049405, 1496-049406, 1496-049407, 1496-049408, 1496-049409, 1496-049410, 1496-049411, 1496-049412, 1496-049413, 1496-049414, 1496-049415, 1496-049416, 1496-049417, 1496-049418, 1496-049419, 1496-049420, 1496-049422, 1496-049423, 1496-049424, 1496-049425, 1496-049426, 1496-049427, 1496-049428, 1496-049429, 1496-049430, 1496-049431, 1496-049432, 1496-049433, 1496-049434, 1496-049435, 1496-049436, 1496-049437, 1496-049438, 1496-049439, 1496-049440, 1496-049441, 1496-049442, 1496-049443, 1496-049444, 1496-049445, 1496-049446, 1496-049447, 1496-049448, 1496-049449, 1496-049450, 1496-049451, 1496-049452.1496-049453, 1496-049454, 1496-049455, 1496-049456, 1496-049457, 1496-049458, 1496-049459, 1496-049460, 1496-049461, 1496-049462, 1496-049463, 1496-049464, 1496-049465, 1496-049466, 1496-049467, 1496-049468, 1496-049469, 1496-049470, 1496-049471, 1496-049472, 1496-049473, 1496-049474, 1496-049475, 1496-049476, 1496-049477, 1496-049478, 1496-049479, 1496-049480, 1496-049481, 1496-049482, 1496-049483, 1496-049484, 1496-049485, 1496-049486, 1496-049487, 1496-049488, 1496-049489, 1496-049490, 1496-049491, 1496-049492, 1496-049493, 1496-049494, 1496-049495, 1496-049496, 1496-049497, 1496-049498, 1496-049499, 1496-049500, 1496-049501, 1496-049502, 1496-049503, 1496-049504, 1496-049505, 1496-049506, 1496-049507, 1496-049508, 1496-049509, 1496-049510, 1496-049511, 1496-049512, 1496-049513, 1496-049514, 1496-049515, 1496-049516, 1496-049517, 1496-049518, 1496-049519, 1496-049520, 1496-049521, 1496-049522, 1496-049523, 1496-049524, 1496-049525, 1496-049526, 1496-049527, 1496-049528, 1496-049529, 1496-049530, 1496-049531, 1496-049532, 1496-049533, 1496-049534, 1496-049535, 1496-049536, 1496-049537, 1496-049538, 1496-049539, 1496-049540, 1496-049541, 1496-049542, 1496-049543, 1496-049544, 1496-049545, 1496-049546, 1496-049547, 1496-049548, 1496-049549, 1496-049556, 1496-049557, 1496-049558, 1496-049559, 1496-049560, 1496-049561, 1496-049562, 1496-049563, 1496-049564, 1496-049565, 1496-049566, 1496-049567, 1496-049568, 1496-049569, 1496-049570, 1496-049571, 1496-049572, 1496-049573, 1496-049574, 1496-049575, 1496-049576, 1496-049577, 1496-049578, 1496-049579, 1496-049580, 1496-049581, 1496-049582, 1496-049583, 1496-049584, 1496-049585, 1496-049586, 1496-049587, 1496-049588, 1496-049589, 1496-049590, 1496-049591, 1496-049592, 1496-049593, 1496-049594, 1496-049595, 1496-049596, 1496-049597, 1496-049598, 1496-049599, 1496-049600, 1496-049601, 1496-049602, 1496-049603, 1496-049604, 1496-049605, 1496-049606, 1496-049607, 1496-049608, 1496-049609, 1496-049610, 1496-049611, 1496-049612, 1496-049613, 1496-049614, 1496-049615, 1496-049616, 1496-049617, 1496-049618, 1496-049619, 1496-049620, 1496-049621, 1496-049622, 1496-049623, 1496-049624, 1496-049625, 1496-049626, 1496-049627, 1496-049628, 1496-049629, 1496-049630, 1496-049631, 1496-049632, 1496-049633, 1496-049634, 1496-049635, 1496-049636, 1496-049637, 1496-049638, 1496-049639, 1496-049640, 1496-049641, 1496-049642, 1496-049643, 1496-049644, 1496-049645, 1496-049646, 1496-049647, 1496-049648, 1496-049649, 1496-049650, 1496-049651, 1496-049652, 1496-049653, 1496-049654, 1496-049655, 1496-049656, 1496-049657, 1496-049658, 1496-049659, 1496-049660, 1496-049661, 1496-049662, 1496-049663, 1496-049664, 1496-049665, 1496-049666, 1496-049667, 1496-049668, 1496-049669, 1496-049670, 1496-049671, 1496-049672, 1496-049673, 1496-049674, 1496-049675, 1496-049676, 1496-049677, 1496-049678, 1496-049679, 1496-049680, 1496-049681, 1496-049682, 1496-049683, 1496-049684, 1496-049685, 1496-049686, 1496-049687, 1496-049688, 1496-049689, 1496-049690, 1496-049691, 1496-049692, 1496-049693, 1496-049694, 1496-049695, 1496-049696, 1496-049697, 1496-049698, 1496-049699, 1496-049700, 1496-049701, 1496-049702, 1496-049703, 1496-049704, 1496-049705, 1496-049706, 1496-049707, 1496-049708, 1496-049709, 1496-049710, 1496-049711, 1496-049712, 1496-049713, 1496-049714, 1496-049715, 1496-049716, 1496-049717, 1496-049718, 1496-049719, 1496-049720, 1496-049721, 1496-049722, 1496-049723, 1496-049724, 1496-049725, 1496-049726, 1496-049727, 1496-049728, 1496-049729, 1496-049730, 1496-049731, 1496-049732, 1496-049762, 1496-049763, 1496-049764, 1496-049765, 1496-049766, 1496-049767, 1496-049768, 1496-049769, 1496-049770, 1496-049771, 1496-049772, 1496-049773, 1496-049774, 1496-049775, 1496-049776, 1496-049777, 1496-049778, 1496-049779, 1496-049780, 1496-049781, 1496-049782, 1496-049783, 1496-049784, 1496-049785, 1496-049786, 1496-049787, 1496-049788, 1496-049789, 1496-049790, 1496-049791, 1496-049792, 1496-049793, 1496-049794, 1496-049795, 1496-049796, 1496-049797, 1496-049798, 1496-049799, 1496-049800, 1496-049801, 1496-049802, 1496-049803, 1496-049804, 1496-049805, 1496-049806, 1496-049807, 1496-049808, 1496-049809, 1496-049810, 1496-049811, 1496-049812, 1496-049813, 1496-049814, 1496-049815, 1496-049816, 1496-049817, 1496-049818, 1496-049819, 1496-049820, 1496-049821, 1496-049822, 1496-049823, 1496-049824, 1496-049825, 1496-049826, 1496-049827, 1496-049828, 1496-049829, 1496-049830, 1496-049831, 1496-049832, 1496-049833, 1496-049834, 1496-049835, 1496-049836, 1496-049837, 1496-049838, 1496-049839, 1496-049840, 1496-049841, 1496-049842, 1496-049843, 1496-049844, 1496-049845, 1496-049846, 1496-049847, 1496-049848, 1496-049849, 1496-049850, 1496-049851, 1496-049852, 1496-049853, 1496-049854, 1496-049855, 1496-049856, 1496-049857, 1496-049858, 1496-049859, 1496-049860, 1496-049861, 1496-049862, 1496-049863, 1496-049864, 1496-049865, 1496-049866, 1496-049867, 1496-049868, 1496-049869, 1496-049870, 1496-049871, 1496-049872, 1496-049873, 1496-049874, 1496-049875, 1496-049876, 1496-049877, 1496-049878, 1496-049879, 1496-049880, 1496-049881, 1496-049882, 1496-049883, 1496-049884, 1496-049885, 1496-049886, 1496-049887, 1496-049888, 1496-049889, 1496-049890, 1496-049891, 1496-049892, 1496-049893, 1496-049894, 1496-049895, 1496-049896, 1496-049897, 1496-049898, 1496-049899, 1496-049900, 1496-049901, 1496-049902, 1496-049903, 1496-049904, 1496-049905, 1496-049906, 1496-049907, 1496-049908, 1496-049909, 1496-049910, 1496-049911, 1496-049912, 1496-049913, 1496-049914, 1496-049915, 1496-049916, 1496-049917, 1496-049918, 1496-049919, 1496-049920, 1496-049921, 1496-049922, 1496-049923, 1496-049924, 1496-049925, 1496-049926, 1496-049927, 1496-049928, 1496-049929, 1496-049930, 1496-049931, 1496-049932, 1496-049933, 1496-049934, 1496-049935, 1496-049936, 1496-049937, 1496-049938, 1496-049939, 1496-049940, 1496-049941, 1496-049942, 1496-049943, 1496-049944, 1496-049945, 1496-049946, 1496-049947, 1496-049948, 1496-049949, 1496-049950, 1496-049951, 1496-049952, 1496-049953, 1496-049954, 1496-049955, 1496-049956, 1496-049957, 1496-049958, 1496-049959, 1496-049960, 1496-049961, 1496-049962, 1496-049963, 1496-049964, 1496-049965, 1496-049966, 1496-049967, 1496-049968, 1496-049969, 1496-049970, 1496-049971, 1496-049972, 1496-049973, 1496-049974, 1496-049975, 1496-049976, 1496-049977, 1496-049978, 1496-049979, 1496-049980, 1496-049981, 1496-049982, 1496-049983, 1496-049984, 1496-049985, 1496-049986, 1496-049987, 1496-049988, 1496-049989, 1496-049990, 1496-049991, 1496-049992, 1496-100566, 1496-100908, 1496-100909, 1496-100910, 1496-100911, 1496-100912, 1496-100913, 1496-100914, 1496-100915, 1496-100916, 1496-100969, 1496-100970, 1496-100971, 1496-101132, 1496-101156, 1496-101158, 1496-101159, 1496-101197, 1496-101198, 1496-101199, 1496-101200, 1496-101202, 1496-101203, 1496-101204, 1496-101205, 1496-101316, 1496-101317, 1496-101318, 1496-101319, 1496-101320, 1496-101321, 1496-101322, 1496-101323, 1496-101324, 1496-101325, 1496-101328, 1496-101329, 1496-101330, 1496-101331, 1496-101332, 1496-101333, 1496-101334, 1496-101335, 1496-101336, 1496-101337, 1496-101350, 1496-101387, 1496-101388, 1496-101390, 1496-101391, 1496-101392, 1496-101393, 1496-101394, 1496-101395, 1496-101396, 1496-101397, 1496-101398, 1496-101399, 1496-101400, 1496-101401, 1496-101402, 1496-101403, 1496-101404, 1496-101405, 1496-101406, 1496-101407, 1496-101408, 1496-101409, 1496-101410, 1496-101411, 1496-101412, 1496-101413, 1496-101414, 1496-101415, 1496-101416, 1496-101417, 1496-101418, 1496-101419, 1496-101420, 1496-101422, 1496-101423, 1496-101424, 1496-101425, 1496-101426, 1496-101427, 1496-101428, 1496-101429, 1496-101430, 1496-101431, 1496-101432, 1496-101433, 1496-101434, 1496-101435, 1496-101436, 1496-101437, 1496-101438, 1496-101439, 1496-101440, 1496-101441, 1496-101442, 1496-101443, 1496-101469, 1496-101470, 1496-101471, 1496-101473, 1496-101474, 1496-101475, 1496-101476, 1496-101477, 1496-101478, 1496-101479, 1496-101480, 1496-101566, 1496-101567, 1496-101568, 1496-101569, 1496-101570, 1496-101571, 1496-101572, 1496-101573, 1496-101574, 1496-101575, 1496-101576, 1496-101577, 1496-101578, 1496-101579, 1496-101580, 1496-101581, 1496-101582, 1496-101583, 1496-101584, 1496-101585, 1496-101586, 1496-101587, 1496-101588, 1496-101589, 1496-101590, 1496-101591, 1496-101592, 1496-101593, 1496-101594, 1496-101595, 1496-101596, 1496-101597, 1496-101598, 1496-101599, 1496-101600, 1496-101601, 1496-101602, 1496-101603, 1496-101604, 1496-101605, 1496-101606, 1496-101607, 1496-101608, 1496-101609, 1496-101610, 1496-101611, 1496-101612, 1496-101613, 1496-101614, 1496-101615, 1496-101616, 1496-101617, 1496-101618, 1496-101619, 1496-101620, 1496-101621, 1496-101622, 1496-101623, 1496-101624, 1496-101625, 1496-101626, 1496-101627, 1496-101628, 1496-101629, 1496-101630, 1496-101631, 1496-101632, 1496-101633, 1496-101634, 1496-101635, 1496-101636, 1496-101637, 1496-101638, 1496-101639, 1496-101640, 1496-101641, 1496-101642, 1496-101643, 1496-101644, 1496-101645, 1496-101646, 1496-101647, 1496-101648, 1496-101649, 1496-101650, 1496-101651, 1496-101652, 1496-101653, 1496-101654, 1496-101655, 1496-101656, 1496-101657, 1496-101658, 1496-101659, 1496-101660, 1496-101661, 1496-101662, 1496-101663, 1496-101664, 1496-101665, 1496-101666, 1496-101667, 1496-101668, 1496-101669, 1496-101670, 1496-101671, 1496-101672, 1496-101673, 1496-101674, 1496-101675, 1496-101676, 1496-101677, 1496-101678, 1496-101679, 1496-101680, 1496-101681, 1496-101682, 1496-101683, 1496-101684, 1496-101685, 1496-101686, 1496-101687, 1496-101688, 1496-101689, 1496-101690, 1496-101691, 1496-101692, 1496-101693, 1496-101694, 1496-101695, 1496-101696, 1496-101697, 1496-101698, 1496-101699, 1496-101700, 1496-101701, 1496-101702, 1496-101703, 1496-101704, 1496-101705, 1496-101706, 1496-101707, 1496-101709, 1496-101710, 1496-101711, 1496-101712, 1496-101713, 1496-101714, 1496-101715, 1496-101716, 1496-101717, 1496-101718, 1496-101719, 1496-101720, 1496-101721, 1496-101722, 1496-101723, 1496-101724, 1496-101725, 1496-101726, 1496-101727, 1496-101728, 1496-101729, 1496-101730, 1496-101731, 1496-101732, 1496-101733, 1496-101734, 1496-101735, 1496-101736, 1496-101737, 1496-101738, 1496-101739, 1496-101740, 1496-101741, 1496-101742, 1496-101743, 1496-101749, 1496-101750, 1496-101751, 1496-101752, 1496-101753, 1496-101754, 1496-101755, 1496-101756, 1496-101757, 1496-101758, 1496-101759, 1496-101760, 1496-101761, 1496-101762, 1496-101763, 1496-101764, 1496-101765, 1496-101766, 1496-101767, 1496-101768, 1496-101769, 1496-101770, 1496-101771, 1496-101772, 1496-101773, 1496-101774, 1496-101775, 1496-101776, 1496-101777, 1496-101778, 1496-101779, 1496-101780, 1496-101781, 1496-101782, 1496-101783, 1496-101784, 1496-101785, 1496-101786, 1496-101787, 1496-101788, 1496-101789, 1496-101790, 1496-101791, 1496-101792, 1496-101793, 1496-101794, 1496-101795, 1496-101796, 1496-101797, 1496-101798, 1496-101799, 1496-101850, 1496-101851, 1496-101852, 1496-101853, 1496-101854, 1496-101855, 1496-101856, 1496-101857, 1496-101858, 1496-101859, 1496-101860, 1496-101861, 1496-101862, 1496-101863, 1496-101864, 1496-101865, 1496-101866, 1496-101867, 1496-101868, 1496-101869, 1496-101870, 1496-101871, 1496-101872, 1496-101873, 1496-101874, 1496-101876, 1496-101877, 1496-101878, 1496-101879, 1496-101880, 1496-101881, 1496-101882, 1496-101883, 1496-101884, 1496-101885, 1496-101886, 1496-101887, 1496-101888, 1496-101889, 1496-101890, 1496-101891, 1496-101892, 1496-101893, 1496-101894, 1496-101895, 1496-101896, 1496-101897, 1496-101898, 1496-101899, 1496-101900, 1496-101901, 1496-101902, 1496-101903, 1496-101904, 1496-101905, 1496-101906, 1496-101907, 1496-101908, 1496-101909, 1496-101910, 1496-101911, 1496-101912, 1496-101913, 1496-101914, 1496-101915, 1496-101916, 1496-101917, 1496-101918, 1496-101919, 1496-101920, 1496-101921, 1496-101922, 1496-101923, 1496-101924, 1496-101925, 1496-101926, 1496-101927, 1496-101928, 1496-101929, 1496-101930, 1496-101931, 1496-101932, 1496-101933, 1496-101934, 1496-101935, 1496-101936, 1496-101937, 1496-101938, 1496-101939, 1496-101940, 1496-101941, 1496-101942, 1496-101943, 1496-101944, 1496-101945, 1496-101946, 1496-101947, 1496-101948, 1496-101949, 1496-101950, 1496-101951, 1496-101952, 1496-101953, 1496-101954, 1496-101955, 1496-101956, 1496-101957, 1496-101958, 1496-101959, 1496-101960, 1496-101961, 1496-101962, 1496-101963, 1496-101964, 1496-101965, 1496-101966, 1496-101967, 1496-101968, 1496-101969, 1496-101970, 1496-101971, 1496-101972, 1496-101973, 1496-101974, 1496-101975, 1496-101976, 1496-101977, 1496-101978, 1496-101979, 1496-101980, 1496-101981, 1496-101982, 1496-101983, 1496-101984, 1496-101985, 1496-101986, 1496-101987, 1496-101988, 1496-101989, 1496-101990, 1496-101991, 1496-101992, 1496-101993, 1496-101994, 1496-101995, 1496-101996, 1496-101997, 1496-101998, 1496-101999, 1496-102000, 1496-102002, 1496-102003, 1496-102004, 1496-102005, 1496-102006, 1496-102007, 1496-102008, 1496-102009, 1496-102010, 1496-102011, 1496-102012, 1496-102013, 1496-102014, 1496-102015, 1496-102016, 1496-102017, 1496-102018, 1496-102019, 1496-102020, 1496-102021, 1496-102022, 1496-102023, 1496-102024, 1496-102025, 1496-102026, 1496-102027, 1496-102028, 1496-102029, 1496-102030, 1496-102031, 1496-102032, 1496-102033, 1496-102034, 1496-102035, 1496-102036, 1496-102037, 1496-102038, 1496-102039, 1496-102040, 1496-102041, 1496-102042, 1496-102043, 1496-102044, 1496-102045, 1496-102046, 1496-102047, 1496-102048, 1496-102049, 1496-102050, 1496-102051, 1496-102052, 1496-102053, 1496-102054, 1496-102055, 1496-102056, 1496-102057, 1496-102058, 1496-102059, 1496-102060, 1496-102061, 1496-102062, 1496-102063, 1496-102064, 1496-102065, 1496-102066, 1496-102067, 1496-102067, 1496-102068, 1496-102069, 1496-102070, 1496-102071, 1496-102072, 1496-102073, 1496-102074, 1496-102075, 1496-102076, 1496-102077, 1496-102078, 1496-102079, 1496-102080, 1496-102081, 1496-102082, 1496-102083, 1496-102084, 1496-102085, 1496-102086, 1496-102087, 1496-102088, 1496-102089, 1496-102090, 1496-102091, 1496-102092, 1496-102093, 1496-102094, 1496-102095, 1496-102096, 1496-102097, 1496-102098, 1496-102099, 1496-102100, 1496-102101, 1496-102102, 1496-102103, 1496-102104, 1496-102105, 1496-102106, 1496-102107, 1496-102108, 1496-102109, 1496-102110, 1496-102111, 1496-102112, 1496-102113, 1496-102114, 1496-102115, 1496-102116, 1496-102117, 1496-102118, 1496-102119, 1496-102120, 1496-102121, 1496-102122, 1496-102123, 1496-102124, 1496-102125, 1496-102126, 1496-102127, 1496-102128, 1496-102129, 1496-102130, 1496-102131, 1496-102132, 1496-102133, 1496-102134, 1496-102135, 1496-102136, 1496-102137, 1496-102138, 1496-102139, 1496-102140, 1496-102141, 1496-102142, 1496-102143, 1496-102144, 1496-102145, 1496-102146, 1496-102147, 1496-102148, 1496-102149, 1496-102150, 1496-102151, 1496-102152, 1496-102153, 1496-102154, 1496-102155, 1496-102156, 1496-102157, 1496-102158, 1496-102159, 1496-102160, 1496-102161, 1496-102162, 1496-102163, 1496-102164, 1496-102165, 1496-102166, 1496-102167, 1496-102168, 1496-102169, 1496-102170, 1496-102171, 1496-102172, 1496-102173, 1496-102174, 1496-102175, 1496-102176, 1496-102177, 1496-102178, 1496-102179, 1496-102180, 1496-102181, 1496-102182, 1496-102183, 1496-102184, 1496-102185, 1496-102186, 1496-102187, 1496-102188, 1496-102189, 1496-102190, 1496-102191, 1496-102192, 1496-102193, 1496-102194, 1496-102195, 1496-102196, 1496-102197, 1496-102198, 1496-102199, 1496-102200, 1496-102200, 1496-102201, 1496-102202, 1496-102203, 1496-102204, 1496-102205, 1496-102206, 1496-102207, 1496-102208, 1496-102209, 1496-102210, 1496-102212, 1496-102213, 1496-102214, 1496-102215, 1496-102221, 1496-102222, 1496-102223, 1496-102224, 1496-102225, 1496-102226, 1496-102227, 1496-102228, 1496-102229, 1496-102230, 1496-102231, 1496-102232, 1496-102233, 1496-102234, 1496-102235, 1496-102236, 1496-102237, 1496-102238, 1496-102239, 1496-102240, 1496-102241, 1496-102242, 1496-102243, 1496-102244, 1496-102245, 1496-102246, 1496-102247, 1496-102248, 1496-102249, 1496-102250, 1496-102251, 1496-102252, 1496-102253, 1496-102254, 1496-102255, 1496-102256, 1496-102257, 1496-102258, 1496-102259, 1496-102260, 1496-102261, 1496-102262, 1496-102263, 1496-102264, 1496-102265, 1496-102266, 1496-102267, 1496-102268, 1496-102269, 1496-102270, 1496-102271, 1496-102272, 1496-102273, 1496-102274, 1496-102275, 1496-102276, 1496-102277, 1496-102278, 1496-102279, 1496-102280, 1496-102281, 1496-102282, 1496-102283, 1496-102284, 1496-102285, 1496-102286, 1496-102287, 1496-102288, 1496-102289, 1496-102290, 1496-102291, 1496-102292, 1496-102293, 1496-102294, 1496-102295, 1496-102296, 1496-102297, 1496-102298, 1496-102299, 1496-102300, 1496-102301, 1496-102302, 1496-102303, 1496-102304, 1496-102305, 1496-102306, 1496-102307, 1496-102308, 1496-102309, 1496-102310, 1496-102311, 1496-102312, 1496-102313, 1496-102314, 1496-102315, 1496-102316, 1496-102317, 1496-102318, 1496-102319, 1496-102320, 1496-102321, 1496-102322, 1496-102323, 1496-102324, 1496-102325, 1496-102326, 1496-102327, 1496-102328, 1496-102329, 1496-102330, 1496-102331, 1496-102332, 1496-102334, 1496-102335, 1496-102336, 1496-102337, 1496-102339, 1496-102340, 1496-102341, 1496-102342, 1496-102343, 1496-102344, 1496-102345, 1496-102346, 1496-102347, 1496-102348, 1496-102349, 1496-102350, 1496-102351, 1496-102352, 1496-102353, 1496-102354, 1496-102355, 1496-102356, 1496-102357, 1496-102358, 1496-102359, 1496-102360, 1496-102361, 1496-102362, 1496-102363, 1496-102364, 1496-102365, 1496-102366, 1496-102367, 1496-102368, 1496-102369, 1496-102370, 1496-102371, 1496-102372, 1496-102373, 1496-102374, 1496-102375, 1496-102376, 1496-102377, 1496-102378, 1496-102379, 1496-102380, 1496-102381, 1496-102382, 1496-102383, 1496-102384, 1496-102385, 1496-102386, 1496-102387, 1496-102388, 1496-102389, 1496-102390, 1496-102391, 1496-102392, 1496-102393, 1496-102394, 1496-102395, 1496-102396, 1496-102397, 1496-102398, 1496-102399, 1496-102400, 1496-102401, 1496-102402, 1496-102403, 1496-102404, 1496-102405, 1496-102406, 1496-102407, 1496-102408, 1496-102409, 1496-102410, 1496-102412, 1496-102413, 1496-102414, 1496-102415, 1496-102416, 1496-102417, 1496-102418, 1496-102419, 1496-102420, 1496-102421, 1496-102422, 1496-102423, 1496-102424, 1496-102425, 1496-102426, 1496-102427, 1496-102428, 1496-102429, 1496-102430, 1496-102431, 1496-102432, 1496-102433, 1496-102434, 1496-102435, 1496-102436, 1496-102437, 1496-102438, 1496-102439, 1496-102440, 1496-102441, 1496-102442, 1496-102443, 1496-102444, 1496-102445, 1496-102446, 1496-102447, 1496-102448, 1496-102449, 1496-102450, 1496-102451, 1496-102452, 1496-102453, 1496-102454, 1496-102455, 1496-102456, 1496-102457, 1496-102458, 1496-102459, 1496-102460, 1496-102461, 1496-102462, 1496-102463, 1496-102464, 1496-102465, 1496-102466, 1496-102467, 1496-102468, 1496-102469, 1496-102470, 1496-102471, 1496-102472, 1496-102473, 1496-102474, 1496-102475, 1496-102476, 1496-102477, 1496-102478, 1496-102479, 1496-102480, 1496-102481, 1496-102482, 1496-102483, 1496-102484, 1496-102485, 1496-102486, 1496-102487, 1496-102488, 1496-102489, 1496-102490, 1496-102491, 1496-102492, 1496-102493, 1496-102494, 1496-102495, 1496-102496, 1496-102497, 1496-102498, 1496-102499, 1496-102500, 1496-102501, 1496-102502, 1496-102503, 1496-102504, 1496-102505, 1496-102506, 1496-102507, 1496-102508, 1496-102509, 1496-102510, 1496-102511, 1496-102512, 1496-102513, 1496-102514, 1496-102515, 1496-102516, 1496-102517, 1496-102518, 1496-102519, 1496-102520, 1496-102521, 1496-102522, 1496-102523, 1496-102524, 1496-102525, 1496-102526, 1496-102527, 1496-102528, 1496-102529, 1496-102530, 1496-102531, 1496-102532, 1496-102533, 1496-102534, 1496-102535, 1496-102536, 1496-102537, 1496-102538, 1496-102539, 1496-102540, 1496-102541, 1496-102542, 1496-102543, 1496-102544, 1496-102545, 1496-102546, 1496-102547, 1496-102548, 1496-102549, 1496-102550, 1496-102551, 1496-102552, 1496-102553, 1496-102554, 1496-102555, 1496-102556, 1496-102557, 1496-102558, 1496-102559, 1496-102560, 1496-102561, 1496-102562, 1496-102563, 1496-102564, 1496-102565, 1496-102566, 1496-102567, 1496-102568, 1496-102569, 1496-102570, 1496-102571, 1496-102572, 1496-102573, 1496-102574, 1496-102575, 1496-102576, 1496-102577, 1496-102578, 1496-102579, 1496-102580, 1496-102581, 1496-102582, 1496-102583, 1496-102584, 1496-102585, 1496-102586, 1496-102587, 1496-102588, 1496-102589, 1496-102590, 1496-102591, 1496-102592, 1496-102593, 1496-102594, 1496-102595, 1496-102596, 1496-102597, 1496-102598, 1496-102599, 1496-102600, 1496-102601, 1496-102602, 1496-102603, 1496-102604, 1496-102605, 1496-102606, 1496-102607, 1496-102608, 1496-102609, 1496-102610, 1496-102611, 1496-102612, 1496-102613, 1496-102614, 1496-102615, 1496-102616, 1496-102617, 1496-102618, 1496-102619, 1496-102620, 1496-102621, 1496-102622, 1496-102623, 1496-102624, 1496-102625, 1496-102626, 1496-102627, 1496-102628, 1496-102629, 1496-102630, 1496-102631, 1496-102632, 1496-102633, 1496-102634, 1496-102635, 1496-102636, 1496-102637, 1496-102638, 1496-102639, 1496-102640, 1496-102641, 1496-102642, 1496-102655, 1496-102656, 1496-102657, 1496-102658, 1496-102659, 1496-102673, 1496-102674, 1496-102675, 1496-102676, 1496-102677, 1496-102678, 1496-102679, 1496-102680, 1496-102681, 1496-102682, 1496-102683, 1496-102684, 1496-102685, 1496-102686, 1496-102687, 1496-102688, 1496-102689, 1496-102690, 1496-102691, 1496-102692, 1496-102693, 1496-102694, 1496-102695, 1496-102696, 1496-102697, 1496-102698, 1496-102699, 1496-102700, 1496-102701, 1496-102702, 1496-102703, 1496-102704, 1496-102705, 1496-102706, 1496-102707, 1496-102708, 1496-102709, 1496-102710, 1496-102711, 1496-102712, 1496-102713, 1496-102714, 1496-102715, 1496-102716, 1496-102717, 1496-102718, 1496-102719, 1496-102720, 1496-102721, 1496-102722, 1496-102723, 1496-102724, 1496-102725, 1496-102726, 1496-102727, 1496-102728, 1496-102729, 1496-102730, 1496-102731, 1496-102732, 1496-102733, 1496-102734, 1496-102735, 1496-102736, 1496-102737, 1496-102738, 1496-102739, 1496-102740, 1496-102741, 1496-102742, 1496-102743, 1496-102744, 1496-102745, 1496-102746, 1496-102747, 1496-102748, 1496-102749, 1496-102750, 1496-102751, 1496-102752, 1496-102753, 1496-102754, 1496-102755, 1496-102756, 1496-102757, 1496-102758, 1496-102759, 1496-102760, 1496-102761, 1496-102762, 1496-102763, 1496-102764, 1496-102765, 1496-102766, 1496-102767, 1496-102768, 1496-102770, 1496-102771, 1496-102772, 1496-102773, 1496-102774, 1496-102775, 1496-102776, 1496-102777, 1496-102778, 1496-102779, 1496-102780, 1496-102781, 1496-102782, 1496-102783, 1496-102784, 1496-102785, 1496-102786, 1496-102787, 1496-102788, 1496-102789, 1496-102790, 1496-102791, 1496-102792, 1496-102793, 1496-102794, 1496-102795, 1496-102796, 1496-102797, 1496-102798, 1496-102799, 1496-102800, 1496-102801, 1496-102802, 1496-102803, 1496-102804, 1496-102805, 1496-102806, 1496-102807, 1496-102808, 1496-102809, 1496-102810, 1496-102811, 1496-102812, 1496-102813, 1496-102814, 1496-102815, 1496-102816, 1496-102817, 1496-102818, 1496-102819, 1496-102820, 1496-102821, 1496-102822, 1496-102823, 1496-102824, 1496-102825, 1496-102826, 1496-102827, 1496-102828, 1496-102829, 1496-102830, 1496-102831, 1496-102832, 1496-102833, 1496-102834, 1496-102835, 1496-102836, 1496-102837, 1496-102838, 1496-102839, 1496-102840, 1496-102841, 1496-102842, 1496-102843, 1496-102844, 1496-102845, 1496-102846, 1496-102847, 1496-102848, 1496-102849, 1496-102850, 1496-102851, 1496-102852, 1496-102853, 1496-102854, 1496-102855, 1496-102856, 1496-102857, 1496-102858, 1496-102859, 1496-102860, 1496-102861, 1496-102862, 1496-102863, 1496-102864, 1496-102865, 1496-102865, 1496-102867, 1496-102868, 1496-102869, 1496-102870, 1496-102871, 1496-102872, 1496-102873, 1496-102874, 1496-102875, 1496-102876, 1496-102877, 1496-102878, 1496-102879, 1496-102880, 1496-102881, 1496-102882, 1496-102883, 1496-102884, 1496-102885, 1496-102886, 1496-102887, 1496-102888, 1496-102889, 1496-102890, 1496-102891, 1496-102892, 1496-102893, 1496-102894, 1496-102895, 1496-102896, 1496-102897, 1496-102898, 1496-102899, 1496-102900, 1496-102901, 1496-102902, 1496-102903, 1496-102904, 1496-102905, 1496-102906, 1496-102907, 1496-102908, 1496-102909, 1496-102910, 1496-102911, 1496-102912, 1496-102913, 1496-102914, 1496-102915, 1496-102916, 1496-102917, 1496-102918, 1496-102919, 1496-102920, 1496-102921, 1496-102922, 1496-102923, 1496-102924, 1496-102925, 1496-102926, 1496-102927, 1496-102928, 1496-102930, 1496-102931, 1496-102932, 1496-102933, 1496-102934, 1496-102935, 1496-102936, 1496-102937, 1496-102938, 1496-102939, 1496-102940, 1496-102941, 1496-102942, 1496-102943, 1496-102944, 1496-102945, 1496-102946, 1496-102947, 1496-102948, 1496-102949, 1496-102950, 1496-102951, 1496-102952, 1496-102953, 1496-102954, 1496-102955, 1496-102956, 1496-102957, 1496-102958, 1496-102959, 1496-102960, 1496-102961, 1496-102962, 1496-102963, 1496-102964, 1496-102965, 1496-102966, 1496-102967, 1496-102969, 1496-102970, 1496-102971, 1496-102972, 1496-102973, 1496-102974, 1496-102975, 1496-102976, 1496-102977, 1496-102978, 1496-102979, 1496-102980, 1496-102981, 1496-102985, 1496-102986, 1496-102987, 1496-102988, 1496-102989, 1496-102990, 1496-102991, 1496-102992, 1496-102993, 1496-102994, 1496-102995, 1496-102996, 1496-102997, 1496-102998, 1496-102999, 1496-103000, 1496-103001, 1496-103002, 1496-103003, 1496-103004, 1496-103005, 1496-103006, 1496-103007, 1496-103008, 1496-103009, 1496-103010, 1496-103011, 1496-103012, 1496-103013, 1496-103014, 1496-103015, 1496-103016, 1496-103017, 1496-103018, 1496-103019, 1496-103020, 1496-103021, 1496-103022, 1496-103023, 1496-103024, 1496-103025, 1496-103026, 1496-103027, 1496-103028, 1496-103029, 1496-103030, 1496-103031, 1496-103032, 1496-103033, 1496-103034, 1496-103035, 1496-103036, 1496-103037, 1496-103038, 1496-103039, 1496-103040, 1496-103041, 1496-103042, 1496-103043, 1496-103044, 1496-103045, 1496-103046, 1496-103047, 1496-103048, 1496-103049, 1496-103050, 1496-103051, 1496-103052, 1496-103053, 1496-103054, 1496-103055, 1496-103056, 1496-103057, 1496-103058, 1496-103059, 1496-103060, 1496-103061, 1496-103062, 1496-103063, 1496-103064, 1496-103065, 1496-103066, 1496-103067, 1496-103068, 1496-103069, 1496-103070, 1496-103071, 1496-103072, 1496-103073, 1496-103074, 1496-103075, 1496-103076, 1496-103077, 1496-103078, 1496-103079, 1496-103080, 1496-103081, 1496-103082, 1496-103083, 1496-103084, 1496-103085, 1496-103086, 1496-103087, 1496-103088, 1496-103090, 1496-103091, 1496-103092, 1496-103093, 1496-103094, 1496-103095, 1496-103096, 1496-103097, 1496-103098, 1496-103099, 1496-103100, 1496-103101, 1496-103102, 1496-103103, 1496-103104, 1496-103105, 1496-103106, 1496-103107, 1496-103108, 1496-103109, 1496-103110, 1496-103111, 1496-103112, 1496-103113, 1496-103114, 1496-103115, 1496-103116, 1496-103117, 1496-103118, 1496-103119, 1496-103120, 1496-103121, 1496-103122, 1496-103123, 1496-103124, 1496-103125, 1496-103126, 1496-103127, 1496-103128, 1496-103129, 1496-103130, 1496-103131, 1496-103132, 1496-103133, 1496-103134, 1496-103135, 1496-103136, 1496-103137, 1496-103138, 1496-103139, 1496-103140, 1496-103141, 1496-103142, 1496-103143, 1496-103144, 1496-103145, 1496-103146, 1496-103147, 1496-103148, 1496-103149, 1496-103150, 1496-103151, 1496-103152, 1496-103153, 1496-103154, 1496-103155, 1496-103156, 1496-103157, 1496-103158, 1496-103159, 1496-103160, 1496-103161, 1496-103162, 1496-103163, 1496-103164, 1496-103165, 1496-103166, 1496-103167, 1496-103168, 1496-103169, 1496-103170, 1496-103171, 1496-103172, 1496-103173, 1496-103174, 1496-103176, 1496-103177, 1496-103178, 1496-103179, 1496-103180, 1496-103181, 1496-103182, 1496-103183, 1496-103184, 1496-103185, 1496-103186, 1496-103187, 1496-103188, 1496-103189, 1496-103190, 1496-103191, 1496-103192, 1496-103193, 1496-103194, 1496-103195, 1496-103196, 1496-103197, 1496-103198, 1496-103199, 1496-103200, 1496-103201, 1496-103202, 1496-103203, 1496-103204, 1496-103205, 1496-103206, 1496-103207, 1496-103208, 1496-103209, 1496-103210, 1496-103211, 1496-103212, 1496-103213, 1496-103214, 1496-103215, 1496-103216, 1496-103217, 1496-103218, 1496-103219, 1496-103220, 1496-103221, 1496-103222, 1496-103223, 1496-103224, 1496-103225, 1496-103226, 1496-103227, 1496-103228, 1496-103229, 1496-103230, 1496-103231, 1496-103232, 1496-103233, 1496-103234, 1496-103235, 1496-103236, 1496-103237, 1496-103238, 1496-103239, 1496-103240, 1496-103241, 1496-103242, 1496-103243, 1496-103244, 1496-103245, 1496-103246, 1496-103247, 1496-103248, 1496-103249, 1496-103250, 1496-103251, 1496-103252, 1496-103253, 1496-103254, 1496-103255, 1496-103256, 1496-103257, 1496-103258, 1496-103259, 1496-103260, 1496-103261, 1496-103262, 1496-103263, 1496-103264, 1496-103265, 1496-103266, 1496-103267, 1496-103268, 1496-103269, 1496-103270, 1496-103272, 1496-103273, 1496-103274, 1496-103275, 1496-103276, 1496-103277, 1496-103278, 1496-103279, 1496-103280, 1496-103281, 1496-103282, 1496-103283, 1496-103284, 1496-103285, 1496-103286, 1496-103287, 1496-103288, 1496-103289, 1496-103290, 1496-103291, 1496-103292, 1496-103293, 1496-103294, 1496-103295, 1496-103296, 1496-103297, 1496-103298, 1496-103299, 1496-103300, 1496-103301, 1496-103302, 1496-103303, 1496-103304, 1496-103305, 1496-103306, 1496-103307, 1496-103308, 1496-103309, 1496-103310, 1496-103311, 1496-103312, 1496-103313, 1496-103314, 1496-103315, 1496-103316, 1496-103317, 1496-103318, 1496-103319, 1496-103320, 1496-103321, 1496-103322, 1496-103323, 1496-103324, 1496-103325, 1496-103326, 1496-103327, 1496-103328, 1496-103329, 1496-103330, 1496-103331, 1496-103335, 1496-103336, 1496-103337, 1496-103338, 1496-103339, 1496-103340, 1496-103341, 1496-103342, 1496-103343, 1496-103344, 1496-103345, 1496-103346, 1496-103347, 1496-103348, 1496-103349, 1496-103350, 1496-103351, 1496-103352, 1496-103353, 1496-103354, 1496-103355, 1496-103356, 1496-103357, 1496-103358, 1496-103359, 1496-103360, 1496-103361, 1496-103362, 1496-103363, 1496-103364, 1496-103365, 1496-103366, 1496-103367, 1496-103368, 1496-103369, 1496-103370, 1496-103371, 1496-103372, 1496-103373, 1496-103374, 1496-103375, 1496-103376, 1496-103377, 1496-103378, 1496-103379, 1496-103380, 1496-103381, 1496-103382, 1496-103383, 1496-103384, 1496-103385, 1496-103386, 1496-103387, 1496-103388, 1496-103389, 1496-103390, 1496-103391, 1496-103392, 1496-103393, 1496-103394, 1496-103395, 1496-103396, 1496-103397, 1496-103398, 1496-103399, 1496-103400, 1496-103401, 1496-103402, 1496-103403, 1496-103404, 1496-103405, 1496-103406, 1496-103407, 1496-103408, 1496-103409, 1496-103410, 1496-103411, 1496-103412, 1496-103413, 1496-103414, 1496-103415, 1496-103416, 1496-103417, 1496-103418, 1496-103419, 1496-103420, 1496-103421, 1496-103422, 1496-103423, 1496-103424, 1496-103425, 1496-103426, 1496-103427, 1496-103428, 1496-103429, 1496-103430, 1496-103431, 1496-103432, 1496-103433, 1496-103434, 1496-103435, 1496-103436, 1496-103437, 1496-103438, 1496-103439, 1496-103440, 1496-103441, 1496-103442, 1496-103443, 1496-103444, 1496-103445, 1496-103446, 1496-103447, 1496-103448, 1496-103449, 1496-103450, 1496-103451, 1496-103452, 1496-103453, 1496-103454, 1496-103455, 1496-103456, 1496-103457, 1496-103458, 1496-103459, 1496-103460, 1496-103461, 1496-103462, 1496-103463, 1496-103464, 1496-103465, 1496-103466, 1496-103467, 1496-103468, 1496-103469, 1496-103470, 1496-103471, 1496-103472, 1496-103473, 1496-103474, 1496-103475, 1496-103476, 1496-103477, 1496-103478, 1496-103479, 1496-103480, 1496-103481, 1496-103482, 1496-103483, 1496-103484, 1496-103485, 1496-103486, 1496-103487, 1496-103488, 1496-103489, 1496-103490, 1496-103491, 1496-103492, 1496-103493, 1496-103494, 1496-103495, 1496-103496, 1496-103497, 1496-103498, 1496-103499, 1496-103500, 1496-103501, 1496-103502, 1496-103503, 1496-103504, 1496-103505, 1496-103506, 1496-103507, 1496-103508, 1496-103509, 1496-103510, 1496-103511, 1496-103512, 1496-103513, 1496-103514, 1496-103515, 1496-103516, 1496-103517, 1496-103518, 1496-103519, 1496-103520, 1496-103521, 1496-103522, 1496-103523, 1496-103524, 1496-103525, 1496-103526, 1496-103527, 1496-103528, 1496-103529, 1496-103530, 1496-103531, 1496-103532, 1496-103533, 1496-103534, 1496-103535, 1496-103536, 1496-103537, 1496-103538, 1496-103539, 1496-103540, 1496-103541, 1496-103542, 1496-103543, 1496-103544, 1496-103545, 1496-103546, 1496-103547, 1496-103548, 1496-103549, 1496-103550, 1496-103551, 1496-103552, 1496-103553, 1496-103554, 1496-103555, 1496-103556, 1496-103557, 1496-103558, 1496-103559, 1496-103560, 1496-103561, 1496-103562, 1496-103563, 1496-103564, 1496-103565, 1496-103566, 1496-103567, 1496-103568, 1496-103569, 1496-103570, 1496-103571, 1496-103572, 1496-103573, 1496-103574, 1496-103575, 1496-103576, 1496-103577, 1496-103578, 1496-103579, 1496-103580, 1496-103581, 1496-103582, 1496-103583, 1496-103584, 1496-103585, 1496-103586, 1496-103587, 1496-103588, 1496-103589, 1496-103590, 1496-103591, 1496-103592, 1496-103593, 1496-103594, 1496-103595, 1496-103596, 1496-103597, 1496-103598, 1496-103599, 1496-103600, 1496-103601, 1496-103602, 1496-103603, 1496-103604, 1496-103605, 1496-103606, 1496-103607, 1496-103608, 1496-103609, 1496-103610, 1496-103611, 1496-103612, 1496-103613, 1496-103614, 1496-103615, 1496-103616, 1496-103617, 1496-103618, 1496-103619, 1496-103620, 1496-103621, 1496-103622, 1496-103623, 1496-103624, 1496-103625, 1496-103626, 1496-103627, 1496-103628, 1496-103629, 1496-103630, 1496-103631, 1496-103632, 1496-103633, 1496-103634, 1496-103635, 1496-103636, 1496-103637, 1496-103638, 1496-103639, 1496-103640, 1496-103641, 1496-103642, 1496-103643, 1496-103644, 1496-103645, 1496-103646, 1496-103647, 1496-103648, 1496-103649, 1496-103650, 1496-103651, 1496-103652, 1496-103653, 1496-103654, 1496-103655, 1496-103656, 1496-103657, 1496-103658, 1496-103659, 1496-103660, 1496-103661, 1496-103662, 1496-103663, 1496-103664, 1496-103665, 1496-103666, 1496-103667, 1496-103668, 1496-103669, 1496-103670, 1496-103671, 1496-103672, 1496-103673, 1496-103674, 1496-103675, 1496-103676, 1496-103677, 1496-103678, 1496-103679, 1496-103680, 1496-103681, 1496-103682, 1496-103683, 1496-103684, 1496-103685, 1496-103686, 1496-103687, 1496-103688, 1496-103689, 1496-103690, 1496-103691, 1496-103692, 1496-103693, 1496-103694, 1496-103695, 1496-103696, 1496-103697, 1496-103698, 1496-103699, 1496-103700, 1496-103701, 1496-103702, 1496-103703, 1496-103704, 1496-103705, 1496-103706, 1496-103707, 1496-103708, 1496-103709, 1496-103710, 1496-103711, 1496-103712, 1496-103713, 1496-103714, 1496-103715, 1496-103716, 1496-103717, 1496-103718, 1496-103719, 1496-103720, 1496-103721, 1496-103722, 1496-103723, 1496-103724, 1496-103725, 1496-103726, 1496-103727, 1496-103728, 1496-103729, 1496-103730, 1496-103731, 1496-103732, 1496-103733, 1496-103734, 1496-103735, 1496-103736, 1496-103737, 1496-103738, 1496-103739, 1496-103740, 1496-103741, 1496-103743, 1496-103744, 1496-103745, 1496-103746, 1496-103747, 1496-103748, 1496-103749, 1496-103750, 1496-103751, 1496-103752, 1496-103753, 1496-103754, 1496-103755, 1496-103756, 1496-103757, 1496-103758, 1496-103759, 1496-103760, 1496-103761, 1496-103762, 1496-103763, 1496-103764, 1496-103765, 1496-103766, 1496-103767, 1496-103768, 1496-103769, 1496-103770, 1496-103771, 1496-103772, 1496-103773, 1496-103774, 1496-103775, 1496-103776, 1496-103777, 1496-103778, 1496-103779, 1496-103780, 1496-103781, 1496-103782, 1496-103783, 1496-103784, 1496-103785, 1496-103786, 1496-103787, 1496-103788, 1496-103789, 1496-103790, 1496-103791, 1496-103792, 1496-103793, 1496-103794, 1496-103795, 1496-103796, 1496-103797, 1496-103798, 1496-103799, 1496-103800, 1496-103801, 1496-103802, 1496-103803, 1496-103804, 1496-103805, 1496-103806, 1496-103807, 1496-103808, 1496-103809, 1496-103810, 1496-103811, 1496-103812, 1496-103813, 1496-103814, 1496-103815, 1496-103816, 1496-103817, 1496-103818, 1496-103819, 1496-103820, 1496-103821, 1496-103822, 1496-103823, 1496-103824, 1496-103825, 1496-103826, 1496-103827, 1496-103828, 1496-103829, 1496-103830, 1496-103831, 1496-103832, 1496-103833, 1496-103834, 1496-103835, 1496-103836, 1496-103837, 1496-103838, 1496-103839, 1496-103840, 1496-103841, 1496-103842, 1496-103843, 1496-103844, 1496-103845, 1496-103846, 1496-103847, 1496-103848, 1496-103849, 1496-103850, 1496-103851, 1496-103852, 1496-103853, 1496-103854, 1496-103855, 1496-103856, 1496-103857, 1496-103858, 1496-103859, 1496-103860, 1496-103861, 1496-103862, 1496-103863, 1496-103864, 1496-103865, 1496-103866, 1496-103867, 1496-103868, 1496-103869, 1496-103870, 1496-103871, 1496-103872, 1496-103873, 1496-103874, 1496-103875, 1496-103876, 1496-103877, 1496-103878, 1496-103879, 1496-103880, 1496-103881, 1496-103882, 1496-103883, 1496-103884, 1496-103885, 1496-103886, 1496-103887, 1496-103888, 1496-103889, 1496-103890, 1496-103891, 1496-103892, 1496-103893, 1496-103894, 1496-103895, 1496-103896, 1496-103897, 1496-103898, 1496-103899, 1496-103900, 1496-103901, 1496-103902, 1496-103903, 1496-103904, 1496-103905, 1496-103906, 1496-103907, 1496-103908, 1496-103909, 1496-103910, 1496-103911, 1496-103912, 1496-103913, 1496-103914, 1496-103915, 1496-103916, 1496-103917, 1496-103918, 1496-103919, 1496-103920, 1496-103921, 1496-103922, 1496-103923, 1496-103924, 1496-103925, 1496-103926, 1496-103927, 1496-103928, 1496-103929, 1496-103930, 1496-103931, 1496-103932, 1496-103933, 1496-103934, 1496-103935, 1496-103936, 1496-103937, 1496-103938, 1496-103939, 1496-103940, 1496-103941, 1496-103942, 1496-103943, 1496-103944, 1496-103945, 1496-103947, 1496-103948, 1496-103949, 1496-103950, 1496-103951, 1496-103952, 1496-103953, 1496-103954, 1496-103955, 1496-103956, 1496-103957, 1496-103958, 1496-103959, 1496-103960, 1496-103961, 1496-103962, 1496-103963, 1496-103964, 1496-103965, 1496-103966, 1496-103967, 1496-103968, 1496-103969, 1496-103970, 1496-103971, 1496-103972, 1496-103973, 1496-103974, 1496-103975, 1496-103976, 1496-103977, 1496-103978, 1496-103979, 1496-103980, 1496-103981, 1496-103982, 1496-103983, 1496-103984, 1496-103985, 1496-103986, 1496-103987, 1496-103988, 1496-103989, 1496-103990, 1496-103991, 1496-103992, 1496-103993, 1496-103994, 1496-103995, 1496-103996, 1496-103997, 1496-103998, 1496-103999, 1496-104000, 1496-104001, 1496-104002, 1496-104003, 1496-104004, 1496-104005, 1496-104006, 1496-104007, 1496-104008, 1496-104009, 1496-104010, 1496-104011, 1496-104012, 1496-104013, 1496-104014, 1496-104015, 1496-104016, 1496-104017, 1496-104018, 1496-104019, 1496-104020, 1496-104021, 1496-104022, 1496-104023, 1496-104024, 1496-104025, 1496-104026, 1496-104027, 1496-104028, 1496-104029, 1496-104030, 1496-104031, 1496-104032, 1496-104033, 1496-104034, 1496-104035, 1496-104036, 1496-104037, 1496-104038, 1496-104039, 1496-104040, 1496-104041, 1496-104042, 1496-104043, 1496-104044, 1496-104045, 1496-104046, 1496-104047, 1496-104048, 1496-104049, 1496-104050, 1496-104051, 1496-104052, 1496-104053, 1496-104054, 1496-104055, 1496-104056, 1496-104057, 1496-104058, 1496-104059, 1496-104060, 1496-104061, 1496-104062, 1496-104063, 1496-104064, 1496-104065, 1496-104066, 1496-104067, 1496-104068, 1496-104069, 1496-104070, 1496-104071, 1496-104072, 1496-104073, 1496-104074, 1496-104075, 1496-104076, 1496-104077, 1496-104078, 1496-104079, 1496-104080, 1496-104081, 1496-104082, 1496-104083, 1496-104084, 1496-104085, 1496-104086, 1496-104087, 1496-104088, 1496-104089, 1496-104090, 1496-104091, 1496-104092, 1496-104093, 1496-104094, 1496-104095, 1496-104096, 1496-104097, 1496-104098, 1496-104099, 1496-104100, 1496-104101, 1496-104102, 1496-104103, 1496-104104, 1496-104105, 1496-104106, 1496-104107, 1496-104108, 1496-104109, 1496-104110, 1496-104111, 1496-104112, 1496-104113, 1496-104114, 1496-104115, 1496-104116, 1496-104117, 1496-104118, 1496-104119, 1496-104120, 1496-104121, 1496-104122, 1496-104123, 1496-104124, 1496-104125, 1496-104126, 1496-104127, 1496-104128, 1496-104129, 1496-104130, 1496-104131, 1496-104132, 1496-104133, 1496-104134, 1496-104135, 1496-104136, 1496-104137, 1496-104138, 1496-104139, 1496-104140, 1496-104141, 1496-104142, 1496-104143, 1496-104144, 1496-104145, 1496-104146, 1496-104147, 1496-104148, 1496-104149, 1496-104150, 1496-104151, 1496-104152, 1496-104153, 1496-104154, 1496-104155, 1496-104156, 1496-104157, 1496-104158, 1496-104159, 1496-104160, 1496-104161, 1496-104162, 1496-104163, 1496-104164, 1496-104165, 1496-104166, 1496-104167, 1496-104168, 1496-104169, 1496-104170, 1496-104171, 1496-104172, 1496-104173, 1496-104174, 1496-104175, 1496-104176, 1496-104177, 1496-104178, 1496-104179, 1496-104180, 1496-104181, 1496-104182, 1496-104183, 1496-104184, 1496-104185, 1496-104186, 1496-104187, 1496-104188, 1496-104189, 1496-104190, 1496-104191, 1496-104192, 1496-104193, 1496-104194, 1496-104195, 1496-104196, 1496-104197, 1496-104198, 1496-104199, 1496-104200, 1496-104201, 1496-104202, 1496-104203, 1496-104204, 1496-104205, 1496-104206, 1496-104207, 1496-104208, 1496-104209, 1496-104210, 1496-104211, 1496-104212, 1496-104213, 1496-104214, 1496-104215, 1496-104216, 1496-104217, 1496-104218, 1496-104219, 1496-104220, 1496-104221, 1496-104222, 1496-104223, 1496-104224, 1496-104225, 1496-104226, 1496-104227, 1496-104228, 1496-104229, 1496-104230, 1496-104231, 1496-104232, 1496-104233, 1496-104234, 1496-104235, 1496-104236, 1496-104237, 1496-104238, 1496-104239, 1496-104240, 1496-104241, 1496-104242, 1496-104243, 1496-104244, 1496-104245, 1496-104246, 1496-104247, 1496-104248, 1496-104249, 1496-104250, 1496-104251, 1496-104252, 1496-104253, 1496-104254, 1496-104255, 1496-104256, 1496-104257, 1496-104258, 1496-104259, 1496-104260, 1496-104261, 1496-104262, 1496-104263, 1496-104264, 1496-104265, 1496-104266, 1496-104267, 1496-104268, 1496-104269, 1496-104270, 1496-104271, 1496-104272, 1496-104273, 1496-104274, 1496-104275, 1496-104276, 1496-104277, 1496-104278, 1496-104279, 1496-104280, 1496-104281, 1496-104282, 1496-104283, 1496-104284, 1496-104285, 1496-104286, 1496-104287, 1496-104288, 1496-104289, 1496-104290, 1496-104291, 1496-104292, 1496-104293, 1496-104294, 1496-104295, 1496-104295, 1496-104296, 1496-104297, 1496-104298, 1496-104299, 1496-104300, 1496-104301, 1496-104302, 1496-104303, 1496-104304, 1496-104305, 1496-104306, 1496-104307, 1496-104308, 1496-104309, 1496-104310, 1496-104311, 1496-104312, 1496-104313, 1496-104314, 1496-104315, 1496-104316, 1496-104317, 1496-104318, 1496-104319, 1496-104320, 1496-104321, 1496-104322, 1496-104323, 1496-104324, 1496-104325, 1496-104326, 1496-104327, 1496-104328, 1496-104329, 1496-104330, 1496-104331, 1496-104332, 1496-104333, 1496-104334, 1496-104335, 1496-104336, 1496-104337, 1496-104338, 1496-104339, 1496-104340, 1496-104341, 1496-104342, 1496-104343, 1496-104344, 1496-104345, 1496-104346, 1496-104347, 1496-104348, 1496-104349, 1496-104350, 1496-104351, 1496-104352, 1496-104353, 1496-104354, 1496-104355, 1496-104356, 1496-104357, 1496-104358, 1496-104359, 1496-104360, 1496-104361, 1496-104362, 1496-104363, 1496-104364, 1496-104365, 1496-104366, 1496-104367, 1496-104368, 1496-104369, 1496-104370, 1496-104371, 1496-104372, 1496-104373, 1496-104374, 1496-104375, 1496-104376, 1496-104377, 1496-104378, 1496-104379, 1496-104380, 1496-104381, 1496-104382, 1496-104383, 1496-104384, 1496-104385, 1496-104386, 1496-104387, 1496-104388, 1496-104389, 1496-104390, 1496-104391, 1496-104392, 1496-104393, 1496-104394, 1496-104395, 1496-104396, 1496-104397, 1496-104398, 1496-104399, 1496-104400, 1496-104401, 1496-104402, 1496-104403, 1496-104404, 1496-104405, 1496-104406, 1496-104407, 1496-104408, 1496-104409, 1496-104410, 1496-104411, 1496-104412, 1496-104413, 1496-104414, 1496-104415, 1496-104416, 1496-104417, 1496-104418, 1496-104419, 1496-104420, 1496-104422, 1496-104423, 1496-104424, 1496-104425, 1496-104426, 1496-104427, 1496-104428, 1496-104429, 1496-104430, 1496-104431, 1496-104432, 1496-104433, 1496-104434, 1496-104435, 1496-104436, 1496-104437, 1496-104438, 1496-104439, 1496-104440, 1496-104441, 1496-104442, 1496-104443, 1496-104444, 1496-104445, 1496-104446, 1496-104447, 1496-104448, 1496-104450, 1496-104451, 1496-104452, 1496-104453, 1496-104454, 1496-104455, 1496-104456, 1496-104457, 1496-104458, 1496-104459, 1496-104460, 1496-104461, 1496-104462, 1496-104463, 1496-104464, 1496-104465, 1496-104466, 1496-104467, 1496-104468, 1496-104469, 1496-104470, 1496-104471, 1496-104472, 1496-104473, 1496-104474, 1496-104475, 1496-104476, 1496-104477, 1496-104478, 1496-104479, 1496-104480, 1496-104481, 1496-104482, 1496-104483, 1496-104484, 1496-104485, 1496-104486, 1496-104487, 1496-104488, 1496-104489, 1496-104490, 1496-104491, 1496-104492, 1496-104493, 1496-104494, 1496-104495, 1496-104496, 1496-104497, 1496-104498, 1496-104499, 1496-104500, 1496-104501, 1496-104502, 1496-104503, 1496-104504, 1496-104505, 1496-104506, 1496-104507, 1496-104508, 1496-104509, 1496-104510, 1496-104511, 1496-104512, 1496-104513, 1496-104514, 1496-104515, 1496-104516, 1496-104517, 1496-104518, 1496-104519, 1496-104520, 1496-104521, 1496-104522, 1496-104523, 1496-104524, 1496-104525, 1496-104526, 1496-104527, 1496-104528, 1496-104529, 1496-104530, 1496-104531, 1496-104532, 1496-104533, 1496-104534, 1496-104535, 1496-104536, 1496-104537, 1496-104538, 1496-104539, 1496-104540, 1496-104541, 1496-104542, 1496-104543, 1496-104544, 1496-104545, 1496-104546, 1496-104547, 1496-104548, 1496-104549, 1496-104550, 1496-104551, 1496-104552, 1496-104553, 1496-104554, 1496-104555, 1496-104556, 1496-104557, 1496-104558, 1496-104559, 1496-104560, 1496-104561, 1496-104562, 1496-104563, 1496-104564, 1496-104565, 1496-104566, 1496-104567, 1496-104568, 1496-104569, 1496-104570, 1496-104571, 1496-104572, 1496-104573, 1496-104574, 1496-104575, 1496-104576, 1496-104577, 1496-104578, 1496-104579, 1496-104580, 1496-104581, 1496-104582, 1496-104583, 1496-104584, 1496-104585, 1496-104586, 1496-104587, 1496-104588, 1496-104589, 1496-104590, 1496-104591, 1496-104592, 1496-104593, 1496-104594, 1496-104595, 1496-104596, 1496-104597, 1496-104598, 1496-104599, 1496-104600, 1496-104601, 1496-104602, 1496-104603, 1496-104604, 1496-104605, 1496-104606, 1496-104607, 1496-104608, 1496-104609, 1496-104610, 1496-104611, 1496-104612, 1496-104613, 1496-104614, 1496-104615, 1496-104616, 1496-104617, 1496-104618, 1496-104619, 1496-104620, 1496-104621, 1496-104622, 1496-104623, 1496-104624, 1496-104625, 1496-104626, 1496-104627, 1496-104628, 1496-104629, 1496-104630, 1496-104631, 1496-104632, 1496-104633, 1496-104634, 1496-104635, 1496-104636, 1496-104637, 1496-104638, 1496-104639, 1496-104640, 1496-104641, 1496-104642, 1496-104643, 1496-104644, 1496-104645, 1496-104646, 1496-104647, 1496-104648, 1496-104649, 1496-104650, 1496-104651, 1496-104652, 1496-104653, 1496-104654, 1496-104655, 1496-104656, 1496-104657, 1496-104658, 1496-104659, 1496-104660, 1496-104661, 1496-104662, 1496-104664, 1496-104665, 1496-104666, 1496-104667, 1496-104668, 1496-104669, 1496-104670, 1496-104671, 1496-104672, 1496-104673, 1496-104674, 1496-104675, 1496-104676, 1496-104677, 1496-104678, 1496-104679, 1496-104680, 1496-104681, 1496-104682, 1496-104683, 1496-104684, 1496-104685, 1496-104686, 1496-104687, 1496-104688, 1496-104689, 1496-104690, 1496-104691, 1496-104692, 1496-104693, 1496-104694, 1496-104695, 1496-104696, 1496-104697, 1496-104698, 1496-104699, 1496-104700, 1496-104701, 1496-104702, 1496-104703, 1496-104704, 1496-104705, 1496-104706, 1496-104707, 1496-104708, 1496-104709, 1496-104710, 1496-104711, 1496-104712, 1496-104713, 1496-104714, 1496-104715, 1496-104716, 1496-104717, 1496-104718, 1496-104719, 1496-104720, 1496-104721, 1496-104722, 1496-104723, 1496-104724, 1496-104725, 1496-104726, 1496-104727, 1496-104728, 1496-104729, 1496-104730, 1496-104731, 1496-104732, 1496-104733, 1496-104735, 1496-104740, 1496-104741, 1496-104742, 1496-104743, 1496-104744, 1496-104745, 1496-104746, 1496-104747, 1496-104748, 1496-104749, 1496-104750, 1496-104751, 1496-104752, 1496-104753, 1496-104754, 1496-104755, 1496-104756, 1496-104757, 1496-104758, 1496-104759, 1496-104760, 1496-104761, 1496-104762, 1496-104763, 1496-104764, 1496-104765, 1496-104766, 1496-104767, 1496-104768, 1496-104769, 1496-104770, 1496-104771, 1496-104772, 1496-104773, 1496-104774, 1496-104775, 1496-104776, 1496-104777, 1496-104778, 1496-104779, 1496-104780, 1496-104781, 1496-104782, 1496-104783, 1496-104784, 1496-104785, 1496-104786, 1496-104787, 1496-104788, 1496-104789, 1496-104790, 1496-104791, 1496-104792, 1496-104793, 1496-104794, 1496-104795, 1496-104796, 1496-104797, 1496-104798, 1496-104799, 1496-104800, 1496-104801, 1496-104802, 1496-104803, 1496-104804, 1496-104805, 1496-104806, 1496-104807, 1496-104808, 1496-104809, 1496-104810, 1496-104811, 1496-104812, 1496-104813, 1496-104814, 1496-104815, 1496-104816, 1496-104817, 1496-104818, 1496-104819, 1496-104820, 1496-104821, 1496-104822, 1496-104823, 1496-104824, 1496-104825, 1496-104826, 1496-104827, 1496-104828, 1496-104829, 1496-104830, 1496-104831, 1496-104832, 1496-104833, 1496-104834, 1496-104835, 1496-104836, 1496-104837, 1496-104838, 1496-104839, 1496-104840, 1496-104841, 1496-104842, 1496-104843, 1496-104844, 1496-104845, 1496-104846, 1496-104847, 1496-104848, 1496-104849, 1496-104850, 1496-104851, 1496-104852, 1496-104853, 1496-104854, 1496-104855, 1496-104856, 1496-104857, 1496-104858, 1496-104859, 1496-104860, 1496-104861, 1496-104868, 1496-104869, 1496-104870, 1496-104871, 1496-104872, 1496-104873, 1496-104874, 1496-104875, 1496-104876, 1496-104877, 1496-104878, 1496-104879, 1496-104880, 1496-104881, 1496-104882, 1496-104883, 1496-104884, 1496-104885, 1496-104886, 1496-104887, 1496-104888, 1496-104889, 1496-104890, 1496-104891, 1496-104892, 1496-104893, 1496-104894, 1496-104895, 1496-104896, 1496-104897, 1496-104898, 1496-104899, 1496-104900, 1496-104901, 1496-104902, 1496-104903, 1496-104904, 1496-104905, 1496-104906, 1496-104907, 1496-104908, 1496-104909, 1496-104910, 1496-104911, 1496-104912, 1496-104913, 1496-104914, 1496-104915, 1496-104916, 1496-104917, 1496-104918, 1496-104919, 1496-104920, 1496-104921, 1496-104922, 1496-104923, 1496-104924, 1496-104925, 1496-104926, 1496-104927, 1496-104928, 1496-104929, 1496-104930, 1496-104931, 1496-104932, 1496-104933, 1496-104934, 1496-104935, 1496-104936, 1496-104937, 1496-104938, 1496-104939, 1496-104940, 1496-104941, 1496-104943, 1496-104944, 1496-104945, 1496-104946, 1496-104947, 1496-104948, 1496-104949, 1496-104950, 1496-104951, 1496-104952, 1496-104953, 1496-104954, 1496-104955, 1496-104956, 1496-104957, 1496-104958, 1496-104959, 1496-104960, 1496-104961, 1496-104962, 1496-104963, 1496-104964, 1496-104965, 1496-104966, 1496-104967, 1496-104968, 1496-104969, 1496-104970, 1496-104971, 1496-104972, 1496-104973, 1496-104974, 1496-104975, 1496-104976, 1496-104977, 1496-104978, 1496-104979, 1496-104980, 1496-104981, 1496-104982, 1496-104983, 1496-104984, 1496-104985, 1496-104986, 1496-104987, 1496-104988, 1496-104989, 1496-104990, 1496-104991, 1496-104992, 1496-104993, 1496-104994, 1496-104995, 1496-104996, 1496-104997, 1496-104998, 1496-104999, 1496-105000, 1496-105001, 1496-105002, 1496-105003, 1496-105004, 1496-105005, 1496-105006, 1496-105007, 1496-105008, 1496-105009, 1496-105010, 1496-105011, 1496-105012, 1496-105013, 1496-105014, 1496-105015, 1496-105020, 1496-105021, 1496-105022, 1496-105023, 1496-105024, 1496-105025, 1496-105026, 1496-105027, 1496-105028, 1496-105029, 1496-105030, 1496-105031, 1496-105032, 1496-105033, 1496-105034, 1496-105035, 1496-105036, 1496-105037, 1496-105038, 1496-105039, 1496-105040, 1496-105041, 1496-105042, 1496-105043, 1496-105044, 1496-105045, 1496-105046, 1496-105047, 1496-105048, 1496-105049, 1496-105050, 1496-105051, 1496-105052, 1496-105053, 1496-105054, 1496-105055, 1496-105056, 1496-105057, 1496-105058, 1496-105059, 1496-105060, 1496-105061, 1496-105062, 1496-105063, 1496-105064, 1496-105065, 1496-105066, 1496-105067, 1496-105068, 1496-105069, 1496-105070.1496-105071, 1496-105072, 1496-105073, 1496-105074, 1496-105075, 1496-105076, 1496-105077, 1496-105078, 1496-105079, 1496-105080, 1496-105081, 1496-105082, 1496-105083, 1496-105084, 1496-105085, 1496-105091, 1496-105092, 1496-105093, 1496-105094, 1496-105095, 1496-105096, 1496-105097, 1496-105098, 1496-105099, 1496-105100, 1496-105101, 1496-105102, 1496-105103, 1496-105104, 1496-105105, 1496-105106, 1496-105107, 1496-105108, 1496-105109, 1496-105110, 1496-105111, 1496-105112, 1496-105113, 1496-105115, 1496-105116, 1496-105117, 1496-105118, 1496-105119, 1496-105120, 1496-105121, 1496-105122, 1496-105123, 1496-105124, 1496-105125, 1496-105126, 1496-105127, 1496-105128, 1496-105129, 1496-105130, 1496-105131, 1496-105132, 1496-105133, 1496-105134, 1496-105135, 1496-105136, 1496-105137, 1496-105138, 1496-105139, 1496-105140, 1496-105141, 1496-105142, 1496-105143, 1496-105144, 1496-105145, 1496-105146, 1496-105147, 1496-105148, 1496-105149, 1496-105150, 1496-105151, 1496-105152, 1496-105153, 1496-105154, 1496-105155, 1496-105156, 1496-105157, 1496-105158, 1496-105159, 1496-105160, 1496-105161, 1496-105162, 1496-105163, 1496-105164, 1496-105165, 1496-105166, 1496-105167, 1496-105168, 1496-105169, 1496-105170, 1496-105171, 1496-105172, 1496-105173, 1496-105174, 1496-105175, 1496-105176, 1496-105177, 1496-105178, 1496-105179, 1496-105180, 1496-105181, 1496-105182, 1496-105184, 1496-105185, 1496-105186, 1496-105187, 1496-105188, 1496-105189, 1496-105190, 1496-105191, 1496-105192, 1496-105193, 1496-105194, 1496-105195, 1496-105196, 1496-105197, 1496-105198, 1496-105199, 1496-105200, 1496-105201, 1496-105202, 1496-105203, 1496-105204, 1496-105205, 1496-105206, 1496-105207, 1496-105208, 1496-105209, 1496-105210, 1496-105211, 1496-105212, 1496-105213, 1496-105214, 1496-105215, 1496-105216, 1496-105217, 1496-105218, 1496-105219, 1496-105220, 1496-105221, 1496-105222, 1496-105223, 1496-105224, 1496-105225, 1496-105226, 1496-105227, 1496-105228, 1496-105229, 1496-105230, 1496-105231, 1496-105232, 1496-105233, 1496-105234, 1496-105235, 1496-105236, 1496-105237, 1496-105238, 1496-105239, 1496-105240, 1496-105241, 1496-105242, 1496-105243, 1496-105248, 1496-105249, 1496-105250, 1496-105251, 1496-105252, 1496-105253, 1496-105254, 1496-105255, 1496-105256, 1496-105257, 1496-105258, 1496-105259, 1496-105260, 1496-105261, 1496-105262, 1496-105263, 1496-105264, 1496-105265, 1496-105266, 1496-105267, 1496-105268, 1496-105269, 1496-105270, 1496-105271, 1496-105272, 1496-105273, 1496-105274, 1496-105275, 1496-105276, 1496-105277, 1496-105278, 1496-105279, 1496-105280, 1496-105281, 1496-105282, 1496-105283, 1496-105284, 1496-105285, 1496-105286, 1496-105287, 1496-105288, 1496-105289, 1496-105290, 1496-105291, 1496-105292, 1496-105293, 1496-105294, 1496-105295, 1496-105296, 1496-105297, 1496-105298, 1496-105299, 1496-105300, 1496-105301, 1496-105302, 1496-105303, 1496-105305, 1496-105306, 1496-105307, 1496-105308, 1496-105309, 1496-105310, 1496-105311, 1496-105312, 1496-105313, 1496-105314, 1496-105315, 1496-105317, 1496-105318, 1496-105319, 1496-105320, 1496-105321, 1496-105322, 1496-105323, 1496-105324, 1496-105325, 1496-105326, 1496-105327, 1496-105328, 1496-105329, 1496-105330, 1496-105331, 1496-105333, 1496-105334, 1496-105335, 1496-105336, 1496-105337, 1496-105338, 1496-105339, 1496-105340, 1496-105343, 1496-105344, 1496-105346, 1496-105349, 1496-105350, 1496-105355. |
Recalling Firm/ Manufacturer |
Spacelabs Healthcare, Llc 5150 220th Ave Se Issaquah WA 98029-6834
|
For Additional Information Contact |
Technical Support 425-657-7200
|
Manufacturer Reason for Recall |
NIBP pump inside the Ultraview SL Command Module, Model 91496, became loose. The pump is suspended by the air hoses and electrical connections, but there can be enough movement to
damage internal components and cause the module to fail.
|
FDA Determined Cause 2 |
Process change control |
Action |
SpaceLabs Healthcare sent a Urgent Field Safety Notice dated November 28, 2011, return receipt requested to all US customers. The letter identified the product the problem and the action needed to be taken by customer.
On December 6, 2011, SpaceLabs will send a customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Records of notification will be maintained to ensure complete notification.
Customers were instructed If you hear an unusual rattling noise, tag the module, set it aside and contact your Spacelabs field service engineer. Your field service engineer will be contacting you early in 2012 to schedule a mutually agreeable time to upgrade all of your affected modules. Upgrades will be performed on-site and at no cost to you.
For any questions about this corrective action program, please call Spacelabs Medical at 1-800-522-7025 and select 2 for Technical Support or, outside the United States, 1-425-657-7200 x5089. |
Quantity in Commerce |
6722 units worldwide (3598 units in US and 3124 units outside US) |
Distribution |
Worldwide Distribution -- USA (nationwide) |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DSI and Original Applicant = SPACELABS HEALTHCARE, INC.
|
|
|
|