Class 2 Device Recall Medtronic, DLP, DLP Malleable Single Stage Venous Cannulae

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medtronic, DLP, DLP Malleable Single Stage Venous Cannulae see related information
Date Posted January 18, 2012
Recall Status1 Terminated on May 22, 2014
Recall Number Z-0811-2012
Recall Event ID 60720
510(K)Number K854487 
Product Classification Catheter, cannula and tubing, vascular, cardiopulmonary bypass - Product Code DWF
Product Medtronic, DLP, DLP Malleable Single Stage Venous CannulaeRx Only, Sterilized Using Ethylene Oxide, Nonpyrogenic.
Model # 91037.

These cannulae are intended for collection of venous blood from the right side of the heart via the superior and inferior vena cava during cardiopulmonary bypass surgery up to six hours or less.
Code Information Model # / Lot #
91037

2008120758
2008121114
2009020137
2009040853
2009050547
2009061436
2010051435
2010101499
2011011022
2011030416
2011041113
Recalling Firm/
Manufacturer
Medtronic Inc. Cardiac Rhythm Disease Management
8200 Coral Sea St. N.E.
Saint Paul MN 55112
For Additional Information Contact
763-526-6000
Manufacturer Reason
for Recall
Instructions for Use for the DLP Single Stage Venous Cannulae are in the process of being updated. The current precaution statement referencing ventricular assist procedures will be identified as a contraindication in the updated Instructions for Use. DO NOT USE THIS PRODUCT FOR EXTENDED TERMS SUCH AS VENTRICULAR ASSIST PROCEDURES. Medtronic has received two reports of patient deaths due to p
FDA Determined
Cause 2
Labeling design
Action Medtronic sent an "URGENT MEDICAL DEVICE INFORMATION" letter dated December 2011 to all affected customers. The letter identifies the product, problem, and the action to be taken by the customers. Contract Medtronic CardioVascular LifeLine Technical Support at 1-877-526-7890 for questions related to this issue.
Quantity in Commerce 883
Distribution Worldwide Distribution-USA (nationwide) including Puerto Rico and the Virgin Islands and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Barbados, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macau, Macedonia, Malaysia, Mauritius, Mexico, Montenegro, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Viet Nam, and Yemen.