| Date Posted |
March 14, 2012 |
| Recall Number |
Z-1206-2012 |
| Product |
Smiths Medical, Medfusion® 3500 Syringe Pumps, Software version 5.0.0, Rx Only.
These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes. |
| Code Information |
M26722 M27294 M27864 M27960 M45874
M26723 M27295 M27865 M27961 M45875
M26724 M27296 M27866 M27962 M45876
M26725 M27297 M27867 M27963 M45877
M26727 M27298 M27868 M27964 M45878
M26728 M27299 M27871 M27966 M45879
M26729 M27300 M27872 M27967 M45880
M26730 M27301 M27873 M27968 M45881
M26731 M27302 M27874 M27969 M45882
M26732 M27303 M27875 M27970 M45883
M26733 M27304 M27876 M27971 M45884
M26734 M27305 M27877 M27973 M45885
M26735 M27306 M27878 M27972 M45886
M26736 M27307 M27879 M27974 M45887
M26737 M27308 M27880 M27975 M45888
M26738 M27309 M27882 M27976 M45889
M26739 M27310 M27883 M27977 M47924
M26740 M27311 M27884 M27978 M47925
M26741 M27312 M27885 M27979 M47926
M26742 M27313 M27886 M27980 M47927
M26743 M27314 M27887 M27981 M47928
M26744 M27315 M27888 M27982 M28016
M26745 M27316 M27889 M27983 M28017
M26746 M27317 M27891 M27984 M28018
M26747 M27318 M27892 M27985 M28019
M26749 M27319 M27893 M27986 M28020
M26750 M27320 M27894 M36853 M28021
M26751 M27321 M27895 M36854 M28022
M26753 M27382 M27896 M36855 M28023
M26754 M27383 M27897 M36856 M28024
M26755 M27384 M27899 M36857 M28025
M26756 M27385 M27900 M36858 M28026
M26757 M27386 M27901 M36859 M28027
M26758 M27387 M27902 M36860 M28028
M26760 M27388 M27903 M36861 M28029
M26761 M27389 M27904 M36862 M28030
M26762 M27390 M27905 M36863 M28031
M26763 M27391 M27951 M45865 M28032
M26764 M27392 M27952 M45866 M28033
M26766 M27393 M27953 M45867 M28034
M26767 M27438 M27954 M45868 M28035
M26768 M27439 M27955 M45869 M28036
M26769 M27440 M27956 M45870 M28037
M27291 M27441 M27957 M45871 M28038
M27292 M27862 M27958 M45872 M28039
M27293 M27863 M27959 M45873 M28040
M28041 M28091 M27270 M27423 M27940
M28042 M28092 M27271 M27424 M27941
M28043 M28093 M27272 M27425 M27942
M28044 M28094 M27273 M27427 M27943
M28046 M28095 M27274 M27428 M27944
M28047 M28096 M27275 M27429 M27945
M28048 M28097 M27276 M27430 M27946
M28049 M36839 M27277 M27431 M27947
M28050 M36840 M27279 M27432 M27948
M28051 M36841 M27280 M27433 M27949
M28052 M36842 M27281 M27434 M27950
M28053 M36843 M27282 M27435 M27995
M28054 M36844 M27285 M27436 M27996
M28055 M36845 M27286 M27437 M27997
M28056 M36846 M27287 M27906 M27998
M28057 M77871 M27288 M27907 M27999
M28058 M77912 M27289 M27908 M28000
M28059 M77915 M27290 M27909 M28001
M28060 M77930 M27394 M27910 M28002
M28061 M77943 M27395 M27912 M28003
M28062 M77982 M27396 M27913 M28004
M28063 M77988 M27397 M27914 M28005
M28064 M26770 M27398 M27915 M28006
M28065 M26771 M27399 M27916 M28007
M28066 M26772 M27400 M27917 M28008
M28067 M26773 M27401 M27918 M28009
M28068 M26774 M27402 M27919 M28010
M28069 M26775 M27403 M27920 M28011
M28071 M26777 M27404 M27921 M28012
M28072 M26778 M27405 M27922 M28013
M28073 M26779 M27406 M27923 M28014
M28075 M26780 M27407 M27924 M28015
M28077 M26781 M27408 M27925 M27987
M28078 M27084 M27409 M27926 M27988
M28079 M27258 M27410 M27927 M27989
M28080 M27259 M27411 M27928 M27990
M28081 M27260 M27412 M27929 M27991
M28082 M27261 M27413 M27930 M27992
M28083 M27262 M27415 M27931 M27993
M28084 M27263 M27416 M27932 M27994
M28085 M27264 M27417 M27933 M36847
M28086 M27265 M27418 M27934 M36848
M28087 M27266 M27419 M27935 M36849
M28088 M27267 M27420 M27936 M36850
M28089 M27268 M27421 M27938 M36851
M28090 M27269 M27422 M27939 M36852
M47929 M63913 M75871
M47930 M63914 M75872
M47931 M63915 M75873
M56249 M63916 M75874
M56211 M63917 M75875
M56212 M63918 M75876
M56213 M63919 M75878
M56214 M63920 M75879
M56215 M63921 M75882
M56216 M63922 M75883
M56217 M63923 M75884
M56218 M63924 M75885
M56219 M63925 M75886
M56220 M63926 M75888
M56221 M63927 M75889
M56222 M63928 M75890
M56223 M63929 M75891
M56224 M63930 M75895
M56225 M63931 M75896
M56226 M63932 M75897
M56227 M63933 M76001
M56228 M63934 M76003
M56229 M63935 M76004
M56230 M63936 M76005
M56231 M63937 M76006
M56232 M63938 M76007
M56233 M63939 M76008
M56234 M63940 M76009
M56235 M63941 M76010
M63896 M63942 M76011
M63897 M63943 M76409
M63898 M63944 M76410
M63899 M63945 M76412
M63900 M63946 M76416
M63901 M66848 M76417
M63902 M66849 M77870
M63903 M75858 M77879
M63904 M75859 M77914
M63905 M75860 M77916
M63906 M75861 M77935
M63907 M75862 M77949
M63908 M75863 M77983
M63909 M75865 M77998
M63910 M75867
M63911 M75868
M63912 M75869
M74556
M74559
M69797
M69376
M69369
M69370
M69371
M69372
M69373
M69368
M69375
M69365
M69377
M69378
M69379
M69374
M69380
M69366
M69383
M69364
M69363
M69362
M69361
M69360
M69359
M69358
M69357
M69356
M69367
M69393
M69406
M69405
M69404
M69403
M69402
M69401
M69400
M69399
M69398
M69397
M69396
M69381
M69394
M69355
M69392
M69391
M69390
M69389
M69388
M69387
M69386
M69385
M69384
M69352
M69382
M69395
M69313
M69354
M69325
M69324
M69323
M69322
M69321
M69320
M69319
M69318
M69317
M69316
M69327
M69314
M69328
M69312
M69311
M69310
M69309
M69308
M69307
M69306
M69305
M69304
M69303
M69302
M69315
M69340
M69353
M69410
M69351
M69350
M69349
M69348
M69347
M69346
M69345
M69344
M69343
M69326
M69341
M69407
M69339
M69338
M69337
M69336
M69335
M69334
M69333
M69332
M69331
M69330
M69329
M69342
M69674
M69686
M69685
M69684
M69683
M69682
M69681
M69680
M69679
M69678
M69677
M69408
M69675
M69689
M69673
M69672
M69671
M69670
M69669
M69668
M69667
M69666
M69665
M69664
M69663
M69676
M69700
M69713
M69712
M69711
M69710
M69709
M69708
M69707
M69706
M69705
M69704
M69703
M69687
M69701
M69688
M69699
M69698
M69697
M69696
M69695
M69694
M69693
M69692
M69691
M69690
M69660
M69702
M69620
M69662
M69632
M69631
M69630
M69629
M69628
M69627
M69626
M69625
M69624
M69623
M69634
M69621
M69635
M69619
M69618
M69617
M69616
M69615
M69614
M69613
M69612
M69411
M69298
M69409
M69622
M69647
M69301
M69659
M69658
M69657
M69656
M69655
M69654
M69653
M69652
M69651
M69650
M69633
M69648
M69661
M69646
M69645
M69644
M69643
M69642
M69641
M69640
M69639
M69638
M69637
M69636
M69649
M69097
M69085
M69109
M69108
M69107
M69106
M69105
M69104
M69103
M69102
M69101
M69100
M69111
M69098
M69126
M69096
M69095
M69094
M69093
M69092
M69091
M69090
M69089
M69088
M69087
M69152
M69099
M69138
M69300
M69150
M69149
M69148
M69147
M69146
M69145
M69144
M69143
M69142
M69141
M69110
M69139
M69084
M69137
M69136
M69135
M69134
M69133
M69132
M69131
M69130
M69129
M69128
M69127
M69140
M69043
M69086
M69055
M69054
M69053
M69052
M69051
M69050
M69049
M69048
M69047
M69046
M69057
M69044
M69058
M69042
M69041
M69040
M69039
M69038
M69037
M69036
M69035
M69034
M69033
M69032
M69045
M69070
M69083
M69082
M69081
M69080
M69079
M69078
M69077
M69076
M69075
M69074
M69073
M69056
M69071
M69153
M69069
M69068
M69067
M69066
M69065
M69064
M69063
M69062
M69061
M69060
M69059
M69072
M69259
M69207
M69271
M69270
M69269
M69268
M69267
M69266
M69265
M69264
M69263
M69262
M69273
M69260
M69274
M69258
M69257
M69256
M69255
M69254
M69253
M69252
M69211
M69210
M69209
M69151
M69261
M69286
M69299
M69717
M69297
M69296
M69295
M69294
M69293
M69292
M69291
M69290
M69289
M69272
M69287
M69206
M69285
M69284
M69283
M69282
M69281
M69280
M69279
M69278
M69277
M69276
M69275
M69288
M69165
M69208
M69177
M69176
M69175
M69174
M69173
M69172
M69171
M69170
M69169
M69168
M69179
M69166
M69180
M69164
M69163
M69162
M69161
M69160
M69159
M69158
M69157
M69156
M69155
M69154
M69167
M69192
M69205
M69204
M69203
M69202
M69201
M69200
M69199
M69198
M69197
M69196
M69195
M69178
M69193
M69714
M69191
M69190
M69189
M69188
M69187
M69186
M69185
M69184
M69183
M69182
M69181
M69194
M87249
M87333
M87332
M87331
M87330
M87257
M87256
M87255
M87254
M87253
M87252
M69715
M87250
M87336
M87248
M87247
M87246
M87245
M87244
M87243
M87242
M87241
M87240
M87239
M87238
M87251
M87347
M87360
M87359
M87358
M87357
M87356
M87355
M87354
M87353
M87352
M87351
M87350
M87334
M87348
M87335
M87346
M87345
M87344
M87343
M87342
M87341
M87340
M87339
M87338
M87337
M87235
M87349
M77698
M87237
M78795
M78794
M78793
M78792
M78791
M78790
M78789
M78787
M78786
M78785
M78797
M78780
M78798
M77680
M76545
M76543
M76542
M76541
M76540
M76539
M76538
M76537
M76535
M76534
M78784
M79069
M87363
M87234
M87233
M87232
M87231
M87230
M87229
M87197
M86959
M86909
M79077
M78796
M79073
M87236
M79066
M79053
M78881
M78810
M78809
M78807
M78806
M78805
M78804
M78802
M78799
M79076
M88215
M88203
M88227
M88226
M88225
M88224
M88223
M88222
M88221
M88220
M88219
M88218
M88229
M88216
M88230
M88214
M88213
M88212
M88211
M88210
M88209
M88208
M88207
M88206
M88205
M87361
M88217
M88243
M88260
M88259
M88258
M88257
M88255
M88254
M88253
M88252
M88249
M88248
M88246
M88228
M88244
M88202
M88242
M88241
M88240
M88238
M88237
M88236
M88235
M88234
M88233
M88232
M88231
M88245
M87374
M88204
M87386
M87385
M87384
M87383
M87382
M87381
M87380
M87379
M87378
M87377
M87388
M87375
M87389
M87373
M87372
M87371
M87370
M87369
M87368
M87367
M87366
M87365
M87364
M76528
M87376
M88097
M88201
M88200
M88199
M88198
M88197
M88196
M88195
M88103
M88102
M88101
M88100
M87387
M88098
M87362
M88096
M88095
M88094
M88093
M88092
M88091
M88089
M87393
M87392
M87391
M87390
M88099
M69783
M69771
M74540
M74539
M72168
M72167
M72166
M72165
M72164
M69788
M69787
M69786
M74545
M69784
M74547
M69782
M69781
M69780
M69779
M69778
M69777
M69776
M69775
M69774
M69773
M74574
M69785
M74560
M76532
M74572
M74571
M74570
M74569
M74568
M74567
M74566
M74565
M74564
M74563
M74544
M74561
M69770
M74559
M74558
M74557
M74556
M74555
M74554
M74553
M74552
M74550
M74549
M74548
M74562
M69728
M69772
M69740
M69739
M69738
M69737
M69736
M69735
M69734
M69733
M69732
M69731
M69742
M69729
M69743
M69727
M69726
M69725
M69724
M69723
M69722
M69721
M69720
M69719
M69718
M69716
M69730
M69756
M69769
M69768
M69767
M69766
M69765
M69764
M69763
M69762
M69761
M69760
M69759
M69741
M69757
M74575
M69754
M69753
M69752
M69751
M69750
M69749
M69748
M69747
M69746
M69745
M69744
M69758
M76480
M76464
M76493
M76492
M76491
M76490
M76489
M76488
M76487
M76486
M76485
M76484
M76497
M76481
M76498
M76479
M76478
M76477
M76476
M76475
M76472
M76471
M76469
M76468
M76467
M74573
M76483
M76510
M69018
M76526
M76524
M76521
M76520
M76519
M76518
M76517
M76516
M76515
M76513
M76495
M76511
M76463
M76509
M76508
M76507
M76506
M76505
M76504
M76503
M76502
M76501
M76500
M76499
M76512
M74589
M76466
M76304
M76303
M76298
M76297
M76240
M76216
M74595
M74594
M74593
M74592
M76429
M74590
M76432
M74587
M74586
M74584
M74583
M74582
M74581
M74580
M74579
M74578
M74577
M74576
M74591
M76445
M76461
M76460
M76459
M76458
M76457
M76456
M76455
M76454
M76452
M76451
M76449
M76308
M76446
M76529
M76444
M76443
M76442
M76440
M76439
M76438
M76437
M76436
M76435
M76434
M76433
M76448
|
Recalling Firm/ Manufacturer |
Smiths Medical ASD, Inc. 1265 Grey Fox Rd Saint Paul, Minnesota 55112-6929 |
| Consumer Instructions |
Contact the recalling firm for information |
Reason for Recall |
Pumps rendered inoperable at power up (displaying a blank screen and alarm).
|
| Action |
Smiths Medical ASD, Inc. contacted two consignees by telephone in November 2010 and January 2011. Smiths Medical worked with one of the two customers to correct 540 of the 595 devices (91%) while on-site in January 2011. The remaining devices were corrected by hospital personnel.
The other customer was also made aware of the issue, root cause and corrective action. This customer reported to Smiths Medical that they understood the issue, but did not want Smiths Medical to perform the correction on their pumps at the time (January 2011). They preferred to send their pumps to Smiths Medical for repair, as and when needed. As of December 2011, The hospital has returned 67 devices to Smiths Medical.
For questions regarding this recall call 651-628-7477. |
| Quantity in Commerce |
595 (WI )+ 864 (DC) = 1459 |
| Distribution |
Nationwide Distribution including WI and DC |
| |
|