|
Class 2 Device Recall Allura Xray system |
|
Date Initiated by Firm |
April 02, 2012 |
Date Posted |
April 25, 2012 |
Recall Status1 |
Terminated 3 on May 09, 2014 |
Recall Number |
Z-1462-2012 |
Recall Event ID |
61514 |
510(K)Number |
K031333 K033737
|
Product Classification |
System, x-ray, angiographic - Product Code IZI
|
Product |
Allura X-ray System.
The Allura X-ray System is an angiographic X-ray system. |
Code Information |
722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722001, 722003, 722008 |
Recalling Firm/ Manufacturer |
Philips Healthcare Inc. 3000 Minuteman Road Andover MA 01810
|
For Additional Information Contact |
Dominic Siewko 978-687-1501
|
Manufacturer Reason for Recall |
The monitor Ceiling suspension may fail, which may cause the monitor to drop.
|
FDA Determined Cause 2 |
Process control |
Action |
Philips sent an "URGENT-FIELD SAFETY NOTICE" dated March 27, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provides information for continued safe and proper use of the equipment. Contact your local Philips representative at 866-767-2822 for further support regarding this issue. Recall expanded 5-2013 to additional 460 units distributed worldwide. |
Quantity in Commerce |
145 |
Distribution |
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Algeria, Argentina, Belgium, Brazil, Canada, Czech Rep, Chili, China, Columbia, Denmark, Egypt, France, Germany, India, Indonesia, Italy, Japan, Malaysia, Mexico, Morocco, Netherlands, Oman, Russia, Saudi Arabia, Singapore, South Korea, Spain, Thailand, United Kingdom, and Vietnam. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IZI and Original Applicant = PHILIPS MEDICAL SYSTEMS NORTH AMERICA, INC.
|
|
|
|