|
Class 2 Device Recall Cardiopulmonary bypass temperature controllers |
|
Date Initiated by Firm |
February 03, 2012 |
Date Posted |
May 24, 2012 |
Recall Status1 |
Terminated 3 on February 26, 2013 |
Recall Number |
Z-1656-2012 |
Recall Event ID |
61754 |
Product Classification |
System, thermal regulating - Product Code DWJ
|
Product |
ECMO Heaters Model 333W, 115V - Part #86135, and 230V & 240V - Part #86136, Cardiopulmonary by-pass temperature controllers. The ECMO Heaters are used in conjunction with a heat exchanger to warm the fluid temperature. The device is composed of a heater, a circulating pump, fan, safety high limit, and a microprocessor board. Water is heated and pumped from the unit to the heat exchanger. The device can be set so that the operator can monitor the actual water temperature or, with the use of a 400 series probe, the fluid (external) temperature.
This device is intended to provide temperature-controlled warm water flow to a blood heat exchanger in order to maintain a patient's blood temperature. It can be used during extracorporeal circulation procedures. It is intended to use with commercially available heat exchangers. |
Code Information |
Serial Numbers: 032-E5201 Through 091-E6690 are subject to recall. The Serial Numbers for the 115V units are as follows: 032-E5201 032-E5202 032-E5203 032-E5204 032-E5205 032-E5226 032-E5227 032-E5228 032-E5229 032-E5230 033-E5286 033-E5287 033-E5288 033-E5289 033-E5290 033-E5306 033-E5307 033-E5308 033-E5309 033-E5310 033-E5311 033-E5312 033-E5313 033-E5314 033-E5315 034-E5327 034-E5328 034-E5329 034-E5330 034-E5331 034-E5332 034-E5333 034-E5334 034-E5335 034-E5336 034-E5352 034-E5353 034-E5354 034-E5355 034-E5356 034-E5367 034-E5388 034-E5369 034-E5370 034-E5371 041-E5422 041-E5423 041-E5424 041-E5425 041-E5426 041-E5442 041-E5443 041-E5444 041-E5445 041-E5446 042-E5502 042-E5503 042-E5504 042-E5505 042-E5506 042-E5507 042-E5508 042-E5509 042-E5510 042-E5511 Q44-E5572 044-E5573 044-E5574 Q44-E5575 Q44-E5591 Q44-E5S92 Q44-E5593 Q44-E5594 051-E5605 052-E5646 052-ES647 052-E5648 052-E5664 052-E5665 052-E5692 052-E5693 052-E5694 052-E5710 052-E5711 052-E5712 052-E5718 053-E5724 053-E5725 053-E5731 053-E5732 053-E5733 053-E5734 053-E5740 053-E5741 053-E5757 053-E5758 053-E5759 053-E5775 053-E5776 053-E5777 053-E5778 053-E5779 054-E5785 054-E5811 054-E5812 054-E5828 054-E5829 054-E5830 054-E5831 054-E5832 054-E5833 054-E5834 054-E5835 054-E5836 054-E5837 054-E5838 054-E5839 054-E5840 061-E5881 061-E5882 061-E5883 061-E5884 061-E5885 061E5901 061-E5902 061-E5903 061-E5904 061-E5905 061-E5927 061-E5928 061-E5904 061-E5930 061-E5931 062-E5952 062-E5953 062-E5954 062-E5955 062-E5956 062-E5957 062-E5958 062-E5959 062-E5960 062-E5961 063-E6003 063-E6004 063-E6005 063-E6006 063-E6007 063-E6013 063-E6014 063-E6015 063-E6016 063-E6017 063-E6018 063-E6019 063-E6020 063-E6021 063-E6022 063-E6028 063-E6029 063-E6030 063-E6031 063-E6032 063-E6033 063-E6034 063-E6035 063-E6036 063-E6037 063-E6038 063-E6039 064-E6040 064-E6041 064-E6042 064-E6043 064-E6044 064-E6089 064-E6090 064-E6091 064-E6092 064-E6093 064-E6119 064-E6120 064-E6121 064-E6122 064-E6123 071-E6129 071-E6130 071-E6131 071-E6132 071-E6133 071-E6134 071-E6150 071-E6151 071-E6152 071-E6153 071-E6154 071-E6160 071-E6161 071-E6162 071-E6163 071-E6164 072-E6180 072-E6181 072-E6182 072-E6183 072-E6184 072-E6196 072-E6197 072-E6198 072-E6199 072-E6200 072-E6206 072-E6207 072-E6208 072-E6209 073-E6277 073-E6278 073-E6279 073-E6280 073-E6281 073-E6292 073-E6293 073-E6294 073-E6295 073-E6296 074-E6297 074-E6298 074"E6299 074-E6300 074-E6301 074-E6318 074-E6319 074-E6320 074-E6327 074-E6328 074-E6329 074-E6330 074-E6331 081-E6363 081-E6364 081-E6365 081-E6366 081-E6367 081-E6378 081-E6379 081-E6360 081-E6381 081-E6382 081-E6393 081-E6394 081-E6395 081-E6396 081-E6397 081-E6429 081-E6430 081-E6431 081-E6432 081-E6433 082-E6454 082-E6455 082-E6456 082-E6457 082-E6458 082-E6494 082-E6495 082-E6496 082-E6497 082-E6498 082-E6519 082-E6520 082-E6521 082-E6522 082-E6523 083-E6614 083-E6615 083-E6616 083-E6617 083-E6618 083-E6619 083-E6620 084-E6621 084-E6622 084-E6623 084-E6624 084-E6625 091-E6651 091-E6652 091-E6653 091-E6654 091-E6655 091-E6686 091-E6687 091-E6668 091-E6689, and 091-E6690. The Serial Numbers for the 230V units are as follows: 052-E5666 230V, 061-E5911 230V, and 081-E6408 230V. |
Recalling Firm/ Manufacturer |
Cincinnati Sub-Zero Products Inc 12011 Mosteller Rd Cincinnati OH 45241-1528
|
For Additional Information Contact |
Beatrice Washington 513-772-8810
|
Manufacturer Reason for Recall |
Design error. The probe jack contacts (terminals on certain of the firm's ECMO Heater devices are located too close to the solid state relay contacts (terminals) which can lead to the terminals coming into contact with one another. If this situation occurs, a Service Technician, Operator/User, and/or Patient could be exposed to electric shock during device servicing or use. This situation was ori
|
FDA Determined Cause 2 |
Device Design |
Action |
Cincinnati Sub-Zero (CSZ) issued an 'URGENT Medical Device Field Action Letter to their customers dated January 31, 2012. The letter was mailed out starting on February 3, 2012. The letter includes: a description of the product and problem, describes the affected Serial Number Range subject to recall/correction and instructs the customers to discontinue using the affected serial-numbered units.The customers are instructed to return the product to their Service Department or Biomedical Engineering Dept. for correction and to complete and mail to CSZ the attached 'Field Action Form.' For additional information, the firm can be reached on a toll-free # 1-800-989-7373. |
Quantity in Commerce |
298 units (115V=295 and 230V/240V=3) |
Distribution |
AL, AR. AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL,K IN,, KS, LA, MA, MD, MI, MN, MO, NC, NE, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, and WI |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|