Class 2 Device Recall 3.2 MM IMHS(R) CP SHORT BRAD POINT DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall 3.2 MM IMHS(R) CP SHORT BRAD POINT DRILL see related information
Date Initiated by Firm April 13, 2012
Date Posting Updated June 06, 2012
Recall Status1 Terminated 3 on September 22, 2015
Recall Number Z-1714-2012
Recall Event ID 61767
Product Classification Orthopedic manual surgical instrument - Product Code LXH
Product 3.2 MM IMHS(R) CP SHORT BRAD POINT DRILL, REF 71687001, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116

Orthopedic surgical device.
Code Information Lot Numbers: 06HHU0020E, 07HM04821A, 07JHU0008, 08AM10972A, 08KM16567A, 09FM00898

09FM00899, 09FM00900, 09FM00901, 09FM00902, 09FM00903, 09FM03558, 10AM01791, 10AM01796, 10AM01799, 10AM01803, 10BM04399, 10CM18678, 10CM18680, 10CM18682, 10HM07692, 10HM07789, 10JM07318, 10JM07320, 11AM15222, 11BM09186, 11BM09205, 11DM02807, 11DM02828, 11FM04982, 11FM04985, 11HM18980, 11HM18991, 11KM09348, 11KM09350, 11KM15076, 11KM15077, 12BM08098, 475163-13A, 475163-30B, 475843-15B, 475843-19A, 475843-1B, 475843C, 483936-1B, 485275-4B.
Recalling Firm/
Manufacturer
Smith & Nephew Inc
1450 E. Brooks Rd
Memphis TN 38116
For Additional Information Contact Joe Metzger
978-749-1330
Manufacturer Reason
for Recall
Plastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, voiding sterility.
FDA Determined
Cause 2
Packaging
Action Smith & Nephew Sales Representatives and Smith & Nephew International Distributors were notified of the problem with the affected product and the recall via email on April 13, 2012. All affected direct consignees were notified via Fed Ex on April 13, 2012. The Urgent Medical Device Recall letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to cease use and distribution of the product and quarantine them immediately for return. Customers were instructed to complete the Inventory Return Certification Form included with the letter. Customers should fax or email the form to 901-566-7975 or fieldactions@smith-nephew.com. The affected product should be returned to: Smith & Nephew Attn: Global Field Actions Safety Affairs 1450 Brooks Road Memphis, Tennessee 38116 If the affected product was distributed further customers were instructed to provide a copy of the recall letter to these customers.
Quantity in Commerce 3435 units
Distribution Worldwide Distribution - USA (nationwide) and the countries of Australia, Malaysia, Germany, Japan, Portugal, United Arab Emirates, Spain, France, South Africa, Norway, Sweden, Denmark, Switzerland, Belgium, China, The Netherlands, Finland, Mexico, Italy, UK, Venezuela, Colombia, Thailand, Chile, South Korea, India, and Turkey
Total Product Life Cycle TPLC Device Report