|
Class 2 Device Recall Chorus Mobile Patient Lifter |
|
Date Initiated by Firm |
June 29, 2012 |
Date Posted |
July 06, 2012 |
Recall Status1 |
Terminated 3 on March 13, 2018 |
Recall Number |
Z-1948-2012 |
Recall Event ID |
62116 |
Product Classification |
Lift, patient, non-ac-powered - Product Code FSA
|
Product |
Chorus Mobile Patient Lifter; manufactured by Arjo Ltd., St. Catherine Street, Gloucester, GL1 2SL, England, U.K.; Models KKB5060-21US, KKB5062-21US and KKB5080-21US.
The Chorus is designed to be used as a standing and raising aid, it is useful for short transfers from one sitting position to another, and to elevate a patient for toileting, repositioning, changing of incontinence pads or wound dressings, standing practice etc. it is not intended for long periods of suspension or transportation. |
Code Information |
Model KKB5060-21US, serial numbers 050186430602, 050186430608, 050186430610, 060186430704, 080186430908, 080186430909, 090186431001, 090186431002, 090186431004, 090186431006, 090186431008, 090186431009, 090186431010, 100186431102, 100186431103, 100186431104, 100186431105, 100186431106, 100186431107, 100186431108, 100186431110, 100186431207, 100186431209, 120186815003, 120186815004, 120186815006, 120186815007, 140186815107, 140186815108, 140186815109, 140186815110, 140186820201, 140186820202, 140186820204, 140186820205, 140186820207, 140186820208, 140186820209, 140186820210, 140186820301, 140186820302, 140186820303, 140186820304, 140186820305, 140186820306, 140186820308, 140186820309, 190084658701, 190084658702, 190084658703, 190084658704, 190084658705, 190084658706, 190084658708, 190084658709, 190084658710, 200084658808, 200084658809, 200084658810, 210084711101, 210084711102, 210084711103, 210084711104, 210084711105, 210084711106, 210084711107, 230084817403, 230084817404, 230084817405, 230084817406, 230084817407, 230084817408, 230084817409, 230084817410, 230084817501, 230084817502, 230084817503, 230084817504, 230084817505, 230084817506, 230084817507, 230084817508, 230084817509, 240084817601, 240084817602, 240084817603, 240084817605, 240084817606, 240084817607, 240084817608, 240084817609, 240084817610, 240084883501, 240084883502, 240084883503, 240084883504, 240084883505, 240084883506, 240084883507, 240084883508, 240084883509, 240084883510, 250084883601, 250084883602, 250084883603, 250084883604, 250084883605, 250084883606, 250084883607, 250084883608, 250084883609, 260084883706, 260084883707, 260084883708, 260084883709, 260084883710, 280084890507, 280084890508, 280084890509, 290085094401, 290085094402, 290085094403, 290085094404, 290085094405, 290085094406, 290085094407, 290085094408, 290085094409, 290085094410, 290085097401, 290085097403, 290085097404, 290085097405, 290085097406, 290085097407, 290085097408, 290085097409, 290085097410, 300085097501, 300085097503, 390085458901, 390085458903, 390085458904, 390085458905, 390085458907, 390085458909, 390085458910, 390085532401, 390085532402, 390085532403, 390085532405, 390085532406, 390085532407, 390085532408, 390085532409, 390085532410, 390085532501, 390085532502, 390085532503, 390085532504, 390085532505, 390085532506, 390085532507, 390085532508, 390085532509, 390085532510, 410085535901, 410085535902, 410085535903, 410085535904, 410085535905, 410085535906, 410085535907, 410085535908, 410085535909, 410085535910, 420085536002, 420085536004, 420085536005, 420085536006, 420085536007, 420085536008, 420085536009, 420085536010, 430085536101, 430085536102, 430085536103, 430085536104, 430085536105, 470085863205, 470085863206, 470085863207, 500085888204, 500085888205, 500085888208, 500085888209, 500085888210, 500085908001, 500085908002, 500085908003, 500085908004, 500085908005, 510085961909, 510085976601, 510085976602, 510085976603, 510085976605, 510085976606, 510085976608, 510085976702, 510085976704, 510085976705, 510085976707, 510085976708, 510085976710, 520086075301, 520086075302 and 520086075303; Model KKB5062-21US, serial numbers 010391015510, 020493522101, 020493522102, 020493522103, 020493522104, 020493522105, 020493522106, 030391048708, 030391048709, 040391133401, 040391133402, 040391133403, 040391133404, 040391133408, 040391133409, 050288822704, 050288822705, 050288822706, 050288822708, 050288822709, 050288822710, 050288913701, 050288913702, 050288913703, 050288913704, 050288913705, 050288913706, 050288913707, 050288913708, 050391133508, 050391133509, 050391133510, 050391146201, 050391146202, 050391146203, 050391146204, 050391146205, 050391146206, 060391240510, 060391261107, 060391261108, 060391261110, 070391292701, 080288978202, 080288978203, 080288978204, 080289003502, 080289003503, 080289003504, 080289003505, 080289003506, 080289003507, 080289003508, 080289003509, 080289003510, 090186431007, 090289044001, 090289044009, 090391328804, 090391328805, 090391328806, 090391328807, 090391328808, 100186431101, 100289064101, 100289064103, 100289064104, 100289064106, 100289064107, 100289064108, 100289064109, 100289064110, 110289152001, 110289152002, 110289152003, 110289152008, 110289152009, 110289152010, 110289188501, 110289188502, 110289188504, 110289188505, 110289188506, 110289188507, 110289188508, 110289188509, 110289188510, 110391404204, 110391404206, 110391404207, 110391404208, 110391404209, 110391458406, 110391458407, 110391458408, 110391458409, 110391458410, 120186815008, 120186815009, 120186815010, 120289227201, 130289251810, 130391473301, 130391473310, 140186815101, 140186815102, 140186815104, 140186815105, 140186815106, 140186820307, 150186993502, 150186993503, 150186993504, 150186993505, 150186993506, 150186993507, 150186993509, 150186993510, 150186994001, 150186994002, 150186994003, 150186994004, 150186994005, 150186994006, 150186994007, 150289304601, 150289304602, 150289304610, 150391556801, 150391556804, 150391556805, 150391556806, 150391556807, 150391556808, 150391556809, 150391556810, 150391706601, 150391706602, 150391706603, 150391706702, 150391706703, 150391706704, 150391706705, 150391706706, 150391706707, 160186994108, 160186994109, 160186994110, 160186994201, 160186994202, 160186994203, 160186994204, 160186994205, 160186994206, 160186994207, 160186994208, 160186994209, 160186994210, 160289410007, 160289410009, 160289410010, 160289431301, 160289431302, 160391734809, 160391734810, 170187000001, 170187000002, 170187000003, 170187000004, 170187000005, 170187000006, 170187000007, 170187000008, 170187000009, 170187000010, 170187108201, 170187108202, 170187108203, 170187108204, 170187108205, 170187108206, 170187108207, 170187108208, 170187108209, 170187108210, 170187108301, 170187108302, 170187108303, 170187108304, 170187108309, 170187108310, 170289470006, 180187108401, 180187108402, 180187108403, 180187108404, 180187108405, 180187108406, 180187108407, 180187108408, 180187108409, 180187108410, 180187108501, 180187108502, 180187108503, 180187108504, 180187108505, 180187108506, 180187108507, 180187108508, 180187108509, 180187108510, 180187108601, 180187108602, 180187108603, 180187108604, 180187108605, 180187108606, 180187108607, 180187108608, 180187108609, 180187108610, 180187110401, 180187110402, 180187110403, 180187110404, 180187110405, 180187110406, 180187110407, 180187110408, 180187110410, 180289541404, 180289541405, 180289541406, 180289541407, 180289541408, 180289541409, 180289541410, 180391776801, 180391776802, 180391776803, 180391776804, 180391776805, 190391790510, 190391812701, 190391812703, 190391812704, 190391812707, 190391812708, 200187210601, 200187210602, 200187210605, 200187210606, 200187210607, 200187210608, 200187210609, 200187210610, 200187210701, 200187210702, 200187210703, 200187210704, 200187210705, 200187210706, 200187210707, 200187210708, 200187210709, 200187210710, 210187210801, 210187210802, 210187210803, 210187210804, 210187210805, 210187210806, 210187210807, 210187210808, 210187210809, 210187210810, 210187210901, 210187210902, 210187210903, 210187210904, 210187210905, 210187210906, 210187210908, 210187210909, 210187210910, 210289591401, 210289591402, 210289591403, 210289591405, 210289591406, 210289591407, 210289591408, 210289591409, 210289591410, 220187211001, 220187211003, 220187211004, 220187211005, 220187211006, 220187211007, 220187211008, 220187211009, 220187211010, 220187211102, 220187214701, 220187214702, 220187214703, 220187214704, 220187214705, 220289597201, 220289597202, 220289597203, 220289597204, 220289597209, 220289597210, 230084817402, 230391993804, 230391993805, 230391993806, 230391993807, 230391993808, 230391993809, 230391993810, 230392023502, 230392023503, 230392023504, 230392023505, 230392023509, 230392023510, 240187214903, 240187214904, 240187214905, 240187214906, 240187214908, 240187214909, 240289597301, 240289597302, 240289597303, 240289597304, 240289597305, 240289597306, 240289597307, 240289597308, 240289597309, 240289597310, 250187225801, 250187225802, 250187225803, 250187225804, 250187225805, 250187225806, 250187225807, 250187225810, 250187396201, 250289597401, 250289597402, 250289597403, 250289597404, 250289597405, 250289597406, 250289597407, 250289597408, 250289597409, 250289597410, 250289597504, 250289597505, 250289597506, 250289597507, 250289597508, 250289597509, 250289597510, 250392023601, 250392023602, 250392023603, 250392023604, 250392112607, 250392112608, 250392112610, 260187243301, 260187243302, 260187243303, 260187243304, 260187243307, 260187243308, 260187243309, 260187243310, 260187313001, 260187313002, 260187313003, 260187313004, 260187313005, 260187313006, 260187313007, 260187313008, 260187313009, 260187313010, 260187313101, 260187313102, 260187313103, 260187313108, 260187313109, 260187313110, 260289812301, 260289812302, 260289812303, 260289812304, 260289812305, 260289812306, 260289812307, 260289812308, 260289812309, 260289812310, 260289828201, 260289828202, 260289828203, 260289828204, 260289828205, 260289828206, 260289828207, 260289828208, 260289828209, 260289828210, 260289828301, 260289828302, 260289828303, 260391256509, 260392156501, 260392156502, 260392156503, 260392156504, 260392156505, 260392156506, 260392156507, 260392156508, 260392156510, 260392184701, 260392184702, 260392184703, 260392184704, 260392184705, 260392184706, 260392184707, 260392184708, 260392184709, 260392184710, 260392184801, 260392184802, 260392184803, 260392184804, 260392184805, 270187366502, 270187366503, 270187366504, 270187366505, 270187366507, 270187366508, 270187366510, 270392229509, 270392229510, 280187395901, 280187395902, 280187395903, 280187395904, 280187395905, 280187395906, 280187395907, 280187395909, 280187395910, 280187533601, 280187533602, 280187533603, 280187533604, 280187533605, 280187533606, 280187533608, 280187533609, 280187533610, 280289874902, 280289874903, 280289874904, 280289874905, 280289874906, 280289874907, 280289874908, 280289874909, 280289874910, 280392229601, 280392229602, 280392229603, 280392229604, 280392229605, 280392229606, 280392229607, 280787395906, 290187565601, 290187565602, 290187565603, 290187565604, 290187565605, 290187565607, 290187565609, 290187565610, 290289947701, 290289947702, 290289947703, 290289947704, 290289947705, 290289947706, 290289947707, 290289947708, 290289947709, 290289947710, 290289981101, 290289981102, 290289981103, 290289981104, 290289981105, 290289981106, 290289981107, 290289981108, 290289981109, 290289981110, 290289987201, 290289987206, 290289987210, 300187624701, 300187624702, 300187624703, 300187624704, 300187624705, 310187669705, 310187669710, 310187699707, 310289996301, 310289996302, 310289996303, 310289996304, 310289996305, 310289996306, 310289996307, 310289996308, 310289996309, 310289996310, 320290038301, 320290038302, 320290038303, 320290038304, 320290038305, 320290062907, 320290062908, 320290062909, 320290062910, 320392387304, 320392387305, 320392387308, 320392387309, 320392387310, 330187717002, 330187717003, 330187717004, 330187717005, 330187717006, 330187717008, 330187717009, 330187717010, 330187717101, 330187717102, 330187717104, 330187717106, 330187717107, 330187717108, 340187757707, 340187757708, 340187757709, 340187757710, 340290156201, 340290156202, 340290156203, 340290156204, 340290156207, 340290156208, 340290156209, 340290156210, 340290156301, 340290156302, 340290156303, 340290156304, 340290156305, 340290156306, 340290156307, 340290156308, 340290156309, 340290156310, 340290156401, 340290156402, 340290156403, 340290156404, 340290156405, 340290156406, 340290156407, 340290156408, 340290156410, 360187811002, 360187811003, 360187811004, 360187811005, 360187811006, 360187811007, 360392506701, 360392548203, 360392548204, 360392548205, 360392548206, 360392548207, 360392548208, 380188054805, 380188054806, 380188054807, 380188054808, 380188054809, 380188054810, 380188118101, 380188118102, 380188118104, 380188118105, 380188118106, 380188118109, 380290193001, 380290193002, 380290193003, 380290193004, 380290193005, 380290193006, 380290193007, 380290193008, 380290193009, 380290276503, 380290276504, 380290276505, 380290276506, 380290276507, 380290276508, 380290276509, 380290276510, 390085532404, 390188118202, 390188118203, 390188118204, 390188118205, 390188118206, 390188118207, 390188118208, 390188118209, 390188118210, 390188118302, 390188118303, 390188118307, 390188118308, 390188118309, 390188118310, 390188138801, 390188138802, 390188138803, 390188138804, 390188138805, 390392550403, 390392550404, 390392550405, 390392550406, 390392550407, 390392550408, 390392550409, 400290311101, 400290311102, 400392720106, 400392720107, 400392720108, 400392720109, 400392720110, 420188254502, 420188254503, 420188254506, 420188254509, 420188254510, 420290416604, 420290416605, 420290416606, 420290416607, 420290416608, 420290416609, 420290416610, 420392743201, 420392743202, 420392743203, 420392743208, 420392743209, 420392743210, 430188323302, 430188323303, 430188323304, 430188323305, 430188323306, 430188323307, 430188323308, 430188323309, 430188323310, 430188323401, 430188323402, 430188323403, 430188323404, 430188323501, 430188323502, 430188323503, 430188323504, 430188323507, 430188323508, 430188323509, 430188323510, 430392874901, 430392874902, 430392874903, 440290459401, 440290459402, 440290459403, 440290459404, 440290459405, 440290459406, 440290459407, 440290459408, 440290459409, 440290459504, 440290459506, 440290459507, 440290459508, 440290459509, 440290459510, 440392875001, 440392875002, 440392875003, 440392875004, 440392875005, 440392875006, 440392875007, 440392875008, 440392875009, 440392875010, 450290490201, 450290490202, 450290490203, 450290490204, 450290490205, 450290490206, 450290490207, 450290490208, 450290490209, 450290490210, 450290545601, 450290545602, 450290545603, 450290545607, 450290545608, 450290545609, 450290545610, 460188423001, 460188423002, 460188423003, 460188423004, 460188423005, 460188423006, 460188423007, 460188423009, 460188423010, 460188423101, 460188423102, 460188423103, 460188423104, 460188423105, 460188423106, 460188423107, 460188423108, 460188423109, 460188423110, 460188427901, 460188427902, 460188427903, 460188427904, 460188427905, 460188427906, 460188427907, 460188427908, 460188427909, 460188428003, 460188428004, 460188428005, 460188428006, 460188428007, 460188428008, 460188428009, 460188428010, 460290576901, 460290576902, 460290576906, 460290576907, 460290576908, 460290576909, 460290576910, 460393044301, 460393044302, 470188446201, 470188446202, 470188446203, 470188446204, 470188446205, 470290643201, 470290643202, 470290643203, 470290643204, 470290643205, 470290643206, 470290643207, 470290643208, 470290643209, 470290643210, 480290643301, 480290643302, 480290643303, 480290643304, 480290643305, 480393053806, 480393053807, 480393053808, 480393069401, 480393069402, 480393069403, 480393069404, 480393069405, 480393069409, 480393069410, 500188537402, 500188537403, 500188537404, 500188537405, 500188537406, 500188537407, 500188537408, 500188537409, 500188537410, 500290831504, 500290831505, 500290831506, 500290831507, 500290831508, 500290831509, 500290831510, 500290831601, 500290831602, 500290831603, 500290831604, 500290831605, 500290831606, 500290831607, 500290831608, 510085976604, 510188636401, 510188636402, 510188636403, 510188636404, 510188636405, 510188636406, 510188636407, 510188636408, 510188636409, 510188636410, 510188669901, 510290870101, 510290870104, 510290870105, 510290870106, 510290870107, 510290877503, 510290877504, 510290877505, 510290877506, 510290877507, 510290877508, 510393181203, 510393181204, 510393181205, 510393181206, 530290929802, 530290929803, 530290929804, 530290929805, 530290929806, 530290929807, 530290929808, 530290929809, 530290929810, 530393461707, 530393461708, 530393461709, 530393461710, 530393490701, 530393490702, 530393490703, 530393490704, 630391261106, GB060402522T, GB060402523T, GB080405381T, GB080405382T, GB080405383T, GB080405384T, GB080405795T, GB080405796T, GB080405797T, GB080405798T, GB090407337T, GB090407338T, GB090407339T, GB100534821T, GB100534822T, GB100534823T, GB120414646T, GB120414647T, GB120414648T, GB120414649T, GB120414650T, GB120414651T, GB120414652T, GB120414653T, GB120539316T, GB120539317T, GB120539318T, GB120539319T, GB120539320T, GB120539321T, GB120539322T, GB170424021T, GB170424022T, GB170424023T, GB170424028T, GB230435450T, GB230435451T, GB250439156T, GB250439157T, GB250439160T, GB260441754T, GB260441755T, GB260441756T, GB260441758T, GB260441760T, GB260443105T, GB280446487T, GB280446489T, GB280466488T, GB300448385T, GB300448387T, GB300448388T, GB300448389T, GB300448390T, GB370458261T, GB370458262T, GB370458263T, GB370458264T, GB390461960T, GB390461961T, GB390461962T, GB390461963T, GB390461964T, GB390461965T, GB390461966T, GB390461967T, GB390461968T, GB390461969T, GB400465379T, GB400465380T, GB400465381T, GB400465382T, GB400465383T, GB400465384T, GB410467120T, GB410467121T, GB410467123T, GB410467126T, GB450476122T, GB450476123T, GB450476124T, GB450476125T, GB450476126T, GB490408657T, GB490408658T, GB490408659T, GB490408660T, GB490408661T, GB490408662T, GB490408663T, GB490408664T, GB490408665T, GB490408666T and GB490408667T; and Model KKB5080-21US, serial numbers 010390930401 010390930403, 010390930404, 010390930405, 010390930406, 010390930407, 010390930408, 010390930409, 010390930410, 030391005101, 040391134707, 040391134708, 040391134709, 040391134710, 050391134801, 050391134802, 050391134803, 050391134804, 050391134805, 050391134806, 050391134807, 050391134808, 050391134809, 050391134810, 050391134901, 050391134904, 050391134905, 050391134906, 050391134907, 050391134908, 050391134909, 050391134910, 060391201101, 060391201102, 060391201103, 060391201104, 060391201106, 060391201108, 060391201109, 060391201110, 090391293401, 090391293402, 090391293403, 090391293404, 090391293405, 090391293406, 090391293407, 110391375201, 110391375202, 110391375203, 110391375204, 110391375205, 110391375206, 110391375207, 110391375208, 110391375209, 110391375210, 130289044105, 130289044106, 130289044107, 130289044108, 130289044109, 130289044110, 150289328101, 150289328102, 150289328103, 150289328104, 150289328105, 150289328107, 150289328108, 150289328109, 150289328110, 150391568501, 150391568502, 150391568503, 150391568504, 150391568505, 150391568506, 150391568507, 150391568508, 150391568509, 150391568510, 160289328209, 160289328210, 160391682201, 160391682202, 160391682203, 160391682204, 160391682205, 160391682206, 160391682207, 160391682208, 160391682209, 160391682210, 170289431501, 170289431502, 170289431503, 170289431504, 170289431505, 170289431506, 170289431508, 170289431509, 170289431510, 180391707501, 180391707502, 180391707503, 180391707504, 180391707505, 180391707506, 180391707507, 180391707508, 200289620801, 200289620802, 200289620803, 200289620804, 200289620805, 200289620806, 200289620808, 200289620809, 200289620810, 200289620901, 200289620902, 200289620903, 200289620904, 200289620905, 200289620906, 200289620907, 200289620908, 200289620910, 200289621001, 200289621002, 200289621003, 200289621004, 200289621005, 200289621006, 200289621007, 200289621008, 200289621009, 200289621010, 200289621101, 200289621102, 200289621103, 200289621104, 200289621105, 200289621106, 200289621107, 200289621108, 200289621109, 200289621110, 200391940410, 202896208007, 210370000002, 210391967101, 210391967102, 210391967103, 210391967104, 210391967105, 210391967106, 210391967107, 210391967108, 210391967109, 220391858001, 220391858002, 220391858003, 220391858004, 220391858005, 220391875701, 220391875702, 220391875703, 220391875704, 230392039101, 230392039102, 230392039103, 230392039104, 230392039105, 230392039106, 230392039107, 230392039108, 230392039109, 230392039110, 240289716501, 240289716502, 240289716505, 240289716506, 240289716507, 240289716508, 240289716509, 240392039201, 240392039202, 240392039203, 240392039204, 240392039205, 240392039206, 240392039207, 240392039208, 240392039209, 240392039210, 250392039301, 250392039302, 250392039303, 250392039304, 280392230401, 280392230402, 280392230403, 280392230404, 280392230405, 280392230406, 280392230407, 280392230408, 280392230409, 280392230410, 280392230501, 280392230502, 280392230503, 280392230504, 280392230505, 280392230506, 280392230507, 280392230508, 280392230509, 280392230510, 280392230601, 280392230602, 280392230603, 280392230604, 280392230605, 280392230606, 280392230607, 280392230608, 280392230609, 280392230610, 290392230701, 290392230702, 290392230703, 290392230704, 290392230705, 290392230706, 290392230707, 290392230708, 290392230709, 290392230710, 290392230801, 290392230802, 290392230803, 290392230804, 290392230805, 290392230806, 290392230807, 290392230808, 290392230809, 290392230810, 300392275901, 300392275902, 300392275903, 300392275904, 300392275905, 300392275906, 300392275907, 300392275908, 300392275909, 300392275910, 310392300901, 310392300902, 310392300903, 310392300904, 310392300906, 310392300907, 310392300908, 310392300909, 310392300910, 310392349001, 310392349002, 310392349003, 310392349004, 310392349005, 310392349006, 310392349007, 310392349008, 310392349009, 310392349010, 320392349101, 320392349102, 320392349103, 320392349104, 320392349105, 320392349106, 320392349107, 320392349108, 320392349109, 320392349110, 330392405503, 330392405504, 330392405505, 330392405506, 330392405507, 330392405508, 330392405509, 330392405510, 340392300905, 370290170403, 370290170404, 370290170405, 370290170406, 370290170407, 370392555201, 370392555202, 370392555203, 370392555204, 370392555205, 370392555206, 370392555207, 370392555208, 370392555209, 370392555210, 390290276903, 390290276904, 390290276905, 390290276906, 390290276907, 390290276908, 390290276909, 390290276910, 400290311501, 400290311502, 400290311503, 400290311504, 400290311505, 400290311506, 400290311507, 400290311508, 400392597401, 400392597402, 400392597403, 400392597404, 400392597405, 400392597406, 400392597407, 400392597408, 400392597409, 420392720703, 420392720704, 420392720705, 420392720706, 420392720707, 420392720708, 420392720709, 420392720710, 430290354803, 430290354804, 430290354805, 430290354806, 430290354808, 430290354809, 430290354810, 430392834701, 430392834702, 430392834703, 430392834704, 430392834708, 430392834709, 430392834710, 450393044601, 450393044602, 450393044603, 450393044604, 450393044605, 450393044606, 450393044607, 450393044608, 450393044609, 450393044610, 450393044701, 450393044705, 450393044710, 460290546201, 460290546202, 460290546203, 460290546204, 460290546205, 460290546206, 460290546207, 460290546208, 460290546209, 460290546210, 460393044801, 460393044802, 460393044803, 460393044804, 460393044806, 460393044808, 460393044809, 460393044810, 470290591901, 470290591902, 470290591903, 470290591904, 470290591905,, 470290591906, 470393044901, 470393044902, 470393044904, 470393044905, 470393044906, 470393044907, 470393044908, 480290819202, 480290819203, 480290819209, 480290819210, 480290832001, 480290832002, 480290832003, 480290832004, 480290832005, 480290832007, 480290832008, 480290832009, 480290832010, 480393045001, 480393045002, 480393045003, 480393045004, 480393045005, 480393045006, 480393045007, 480393045008, 480393045009, 480393045010, 482029083204, 490290832101, 490290832103, 490290832104, 490290832105, 490290832106, 490290832107, 490290832108, 490290832109, 490290832110, 490393069503, 490393069504, 490393069505, 490393069506, 490393069507, 490393069508, 490393069509, 490393069510, 490393099501, 490393099502, 490393099503, 490393099504, 490393099505, 490393099506, 490393099507, 490393099508, 490393099509, 490393099510, 500393118801, 500393118802, 500393118803, 500393118804, 500393118805, 500393118806, 500393118807, 500393118808, 500393118809, 500393118810, 500393118901, 500393118902, 500393118903, 500393118904, 500393118905, 500393118906, 510290832202, 510290832203, 510290832204, 510290832205, 510290832206, 510290832207, 510290832208, 510290832209, 510290832210, 510290889401, 510290889402, 510290889403, 510290889404, 510290889406, 510290889407, 510290889408, 510290889409, 510290899405, GB060402327T, GB060402328T, GB060402329T, GB060402330T, GB060402331T, GB060402332T, GB060402333T, GB110537656T, GB110537657T, GB110537658T, GB110537659T, GB110537660T, GB110537661T, GB110537662T, GB110537663T, GB110537664T, GB110537665T, GB120540046T, GB120540047T, GB120540048T, GB120540049T, GB120540050T, GB120540051T, GB120540052T, GB120540053T, GB120540054T, GB120540055T, GB130417812T, GB130417813T, GB130417814T, GB130417817T, GB130417818T, GB130417819T, GB130417820T, GB130417821T, GB130541300T, GB130541301T, GB130541302T, GB130541303T, GB130541304T, GB130541305T, GB130541306T, GB130541307T, GB150420741T, GB150420742T, GB150420743T, GB150420744T, GB150420745T, GB150420746T, GB150420747T, GB150420748T, GB150420749T, GB150420750T, GB160422810T, GB160422811T, GB160422812T, GB170425490T, GB170425491T, GB170425492T, GB170425494T, GB170425495T, GB170425496T, GB170425497T, GB170425498T, GB170425499T, GB190428732T, GB190428733T, GB190428734T, GB190428735T, GB190428736T, GB190428738T, GB190428739T, GB190428740T, GB190428741T, GB220434081T, GB220434082T, GB220434083T, GB220434084T, GB220434085T, GB220434086T, GB220434087T, GB220434088T, GB220434089T, GB220434090T, GB220434599T, GB220434600T, GB220434601T, GB220434602T, GB220434603T, GB220434604T, GB220434605T, GB220434606T, GB220434607T, GB220434608T, GB230435903T, GB230435904T, GB230435905T, GB230435906T, GB230435907T, GB230435908T, GB230435909T, GB230435910T, GB230435911T, GB230435912T, GB250440700T, GB250440701T, GB250440702T, GB250440703T, GB250440704T, GB250440705T, GB250440706T, GB250440707T, GB250440708T, GB260441366T, GB260441368T, GB260441369T, GB260441370T, GB260441371T, GB260441372T, GB260441373T, GB260441374T, GB260441375T, GB270444953T, GB270444954T, GB270444955T, GB270444956T, GB270444958T, GB270444959T, GB300448820T, GB300448823T, GB300448824T, GB300448825T, GB300448826T, GB300448827T, GB300448828T, GB300448829T, GB320452306T, GB320452307T, GB320452308T, GB320452309T, GB320452310T, GB320452311T, GB320452312T, GB320452313T, GB320452314T, GB320452315T, GB340454664T, GB340454665T, GB340454666T, GB340454667T, GB340454668T, GB340454671T, GB340454672T, GB340454673T, GB390461245T, GB390461246T, GB390461247T, GB390461248T, GB390461249T, GB390461250T, GB390461251T, GB390461252T, GB390461253T, GB390461254T, GB400464772T, GB400464773T, GB400464774T, GB400464775T, GB400464776T, GB400464777T, GB400464778T, GB400464779T, GB400464780T, GB400464781T, GB410465843T, GB410465844T, GB410465845T, GB410465846T, GB410465847T, GB410465848T, GB410465849T, GB410465850T, GB420468355T, GB420468356T, GB420468357T, GB420468358T, GB420468359T, GB420468360T, GB420468361T, GB420468362T, GB420468363T, GB420468364T, GB420468902T, GB420468903T, GB420468904T, GB420468905T, GB420468906T, GB420468907T, GB420468908T, GB420468909T, GB420468910T, GB420468911T, GB470479419T, GB470479420T, GB470479421T, GB470479422T, GB470479423T, GB470479425T, GB470479426T, GB470479427T, GB480406353T, GB480406354T, GB480406355T, GB480406356T, GB480406357T, GB480406358T, GB480406359T, GB480406360T, GB480406361T, GB480406362T, GB480406744T, GB480406745T, GB480406746T, GB480406747T, GB480406748T, GB480406749T, GB480406750T, GB480406751T, GB480406752T, GB480406753T, GB480407131T, GB480407132T, GB480407133T, GB480407134T, GB480407135T, GB480407136T, GB480407137T, GB480407138T, GB480407139T, GB480407140T, GB480407350T, GB480407351T, GB480407352T, GB480407353T, GB480407354T, GB480407355T, GB480407356T, GB480407357T, GB480407358T, GB480407359T, GB490408079T, GB490408080T, GB490408081T, GB490408082T, GB490408083T, GB490408084T, GB490408085T, GB490408086T, GB490408087T, GB490408088T, GB520413764T, GB520413765T, GB520413766T, GB520413767T, GB520413768T, GB520413769T, GB520413770T, GB520413771T, GB520413772T, GB520413773T, GB520414012T, GB520414013T, GB520414014T, GB520414015T, GB520414016T, GB520414017T, GB520414018T, GB520414019T and GB520414020T, |
Recalling Firm/ Manufacturer |
Arjo, Inc. dba ArjoHuntleigh 2349 W Lake St Addison IL 60101-6183
|
For Additional Information Contact |
800-323-1245
|
Manufacturer Reason for Recall |
ArjoHuntleigh has received a limited number of customer reports concerning injuries related to an exposed metal support bracket when the proactive knee pad assembly is removed.
|
FDA Determined Cause 2 |
Labeling False and Misleading |
Action |
ArjoHuntleigh sent an "URGENT FIELD CORRECTION RECALL" notice dated June 25, 2012, to all affected customers via UPS Ground on June 29, 2012. The notice informed the accounts of the potential for severe patient injury if the proactive knee pad support is removed from the lift without the removal of the knee pad support bracket. The accounts were reminded that the entire knee pad and support bracket assembly must be removed from the lift when required to do so for exercise. An additional warning statement with diagrams was included in the notice. The accounts were requested to disseminate the information to all users of the affected lifts, to include a copy of the notice in the Instruction for Use Manuals for their lifts, and to complete the Customer Response Form online, using the enclosed card's instructions. Questions were directed to the ArjoHuntleigh at 1-800-323-1245, ext 57985. |
Quantity in Commerce |
1,989 units in the U.S., 1,626 units in the rest of the world |
Distribution |
Worldwide Distribution-USA (nationwide) and the countries of Australia, Austria, Belgium, Bermuda, Canada, Cyprus, Czech Republic, Denmark, Dubai, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Hong Kong, Iceland, Ireland, Israel, Italy, Japan, Kenya, Korea, Luxembourg, the Netherlands, New Zealand, Northern Ireland, Norway, Poland, Portugal, Quatar, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland and Turkey |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|