 Class 1 Recall
OEC 6800 Miniview
see related information
Date Posted July 24, 2012
Recall Status1 Terminated on May 13, 2013
Recall Number Z-2040-2012
Recall Event ID 62382
Premarket Notification
510(K) Number
K992506 
Product Classification Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Code OXO
Product OEC 6800 Miniview, Image-intensified fluoroscopic x-ray system used during diagnostic, surgical and interventional procedures. Designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imagine. The device is not intended for whole-body pediatric imaging.
Code Information 86-1834 , 86-1175 , 86-0238 , 86-0585 , 86-1230 , 86-1992 , 86-0184 , 86-1033 , 86-1480 , 86-1369 , 86-0579 , 86-1678 , 86-1059 , 86-0639 , 86-0389 & 86-0390 , 86-0775 , 86-0850 & 86-1703 , 86-0151 & 86-1311 , 86-1883 , 86-0048 & 86-2015 , 86-0915, 86-1626 & 86-1631 , 86-2118 , 86-0949 , 86-0937 , 86-0071 & 86-1015 , 86-0352 , 86-0527 , 86-0140 , 86-1697 , 86-0697 , 86-1162 , 86-1424 , 86-0795 , 86-0260 , 86-0077 , 86-2002 , 86-0811 , 86-2148 , 86-1582 , 86-1633 , 86-0454 , 86-0185, 86-1119 & 86-1885 , 86-0668 , 86-1026 , 86-0529 , 86-1754 , 86-0901 , 86-0228 , 86-1555 , 86-1212 , 86-1095 , 86-0997 , 86-1475 , 86-2137 , 86-0829 , 86-1737 , 86-1259 , 86-0013 , 86-1139 , 86-0058 , 86-0327 , 86-1346 , 86-0419 & 86-2047 , 86-1892 , 86-0359 & 86-1070 , 86-0331 , 86-1255 , 86-0785 , 86-1043 , 86-1854 , 86-0148 & 86-1036 , 86-0427 & 86-2068 , 86-1988 , 86-0092 , 86-1283 , 86-1955 , 86-0711 , 86-2035 , 86-2017 , 86-0770 & 86-2044 , 86-0716 , 86-0692 , 86-1610 , 86-0154 & 86-1822 , 86-1727 , 86-1209 & 86-1210 , 86-0046 , 86-1303 , 86-0468 , 86-1274 , 86-0113 , 86-0035 , 86-1481 , 86-1808 , 86-1332 , 86-0618 , 86-1860 , 86-0425 , 86-0912 , 86-0925 , 86-0306 & 86-0664 , 86-0987 , 86-2170 , 86-0308 , 86-1302 , 86-0399 , 86-1205 , 86-1615 , 86-2128 , 86-0162 , 86-0166 , 86-0105 , 86-1687 & 86-1844 , 86-0865 , 86-1045 , 86-1360 , 86-0903 , 86-1456 , 86-1213 , 86-1326 , 86-0745 , 86-0517 , 86-2140 , 86-0625 & 86-1399 , 86-0923 , 86-1319 & 86-2133 , 86-1499 , 86-1823 , 86-0401 , 86-0157 , 86-0685 , 86-0570 , 86-1446 , 86-1606 , 86-2030 , 86-0086 , 86-0075 & 86-0076 , 86-0435 , 86-1780 , 86-0247-RC , 86-1478 , 86-1894 , 86-0896 , 86-1249 , 86-0601, 86-0914 & 86-1248 , 86-1250 , 86-1836 , 86-0805, 86-0806 & 86-0972 , 86-1619 , 86-1576 , 86-0544 , 86-1886 , 86-1782 , 86-0475 , 86-0321 , 86-0511 , 86-1949 , 86-1150 , 86-1065 , 86-1614 , 86-1907 , 86-1625 , 86-0757 & 86-0786 , 86-2092 , 86-0501 , 86-2117 , 86-1879 , 86-1714 , 86-0416 , 86-1506 , 86-1732 , 86-0977 , 86-0964 , 86-1531 , 86-0994 & 86-1909 , 86-0091 , 86-0115 , 86-1154 , 86-2171 , 86-0739 , 86-1621 & 86-1721 , 86-1647 , 86-0589 , 86-1490 , 86-0485 , 86-0995 , 86-0657 , 86-0700 , 86-1620 , 86-2032 , 86-0487 & 86-0494 , 86-1343 , 86-0102 , 86-2005 , 86-0504 , 86-1178 , 86-0329 , 86-0777 & 86-1273 , 86-1187 , 86-0519 , 86-1536 , 86-1448 , 86-0572 , 86-1223 , 86-2199 , 86-1880 , 86-0827 , 86-0068, 86-0155 & 86-0239 , 86-0492 , 86-0558 & 86-1966 , 86-0347 & 86-0411 , 86-1324 , 86-0191 , 86-1166 , 86-0701 , 86-2008 , 86-1219 , 86-0737 , 86-0974 , 86-2095 , 86-1460 , 86-0381 , 86-1234 , 86-0890 , 86-0099 , 86-1420 , 86-0750 , 86-0259 , 86-1226 , 86-0204 , 86-0482 , 86-0378 , 86-2194 , 86-1293 , 86-0966 , 86-1111 , 86-1775 & 86-1965 , 86-0069 , 86-0137, 86-0428 & 86-0432 , 86-1813 , 86-0978 , 86-0609 , 86-0047 , 86-0070 , 86-1646 & 86-1838 , 86-0906 , 86-1432 , 86-1391 , 86-1431 , 86-1266 & 86-2238 , 86-2239 , 86-1375 , 86-0560 , 86-1686 , 86-1218 , 86-1308 , 86-0277 , 86-0755 & 86-1959 , 86-1179 , 86-0245 & 86-0253 , 86-2056 , 86-1151 , 86-2043 , 86-1192 & 86-1632 , 86-0840 , 86-1728 , 86-0025 , 86-1022 , 86-0871 , 86-0163 , 86-0271 , 86-1962 , 86-0153 & 86-1267 , 86-1523 & 86-1560 , 86-1611 , 86-0732 , 86-1540 , 86-1094 , 86-0573 & 86-1211 , 86-1494 , 86-0198 , 86-0388 , 86-0200 & 86-1381 , 86-2087 , 86-0981 & 86-1713 , 86-0120 , 86-1914 , 86-1972 , 86-1918 , 86-0106 , 86-1858 & 86-1889 , 86-1869 , 86-1583 , 86-1085 , 86-0020 , 86-0913 & 86-2126 , 86-0674-RC , 86-1652 , 86-0520 , 86-0131 , 86-1542 , 86-0036 , 86-0577 , 86-1544 , 86-0206 , 86-0731 , 86-2115 , 86-1056 , 86-0237 , 86-0358 , 86-0361 , 86-0870 , 86-0541 & 86-0980 , 86-1123 , 86-2193 , 86-2026 , 86-0298 , 86-0214 , 86-2011 , 86-0042 , 86-0918 , 86-0441 , 86-2147 , 86-1509 , 86-1428 & 86-2057 , 86-2077 , 86-1607 , 86-0299 , 86-0973 , 86-0583 & 86-0584 , 86-0575 , 86-0346, 86-0556 & 86-2113 , 86-2018 , 86-0552 , 86-2110 , 86-0624 , 86-0403 , 86-1443 , 86-0736 , 86-1579 & 86-1580 , 86-1029 , 86-1142 , 86-1009 , 86-2075 , 86-0804 , 86-1784 , 86-1513 , 86-0567 , 86-1807 , 86-0032 , 86-0255 , 86-1398 , 86-1067 , 86-1258 , 86-1960 , 86-2121 , 86-0832 & 86-2210 , 86-1317 , 86-0364 & 86-1565 , 86-1690 , 86-0064 , 86-1884 , 86-2165 , 86-1891 , 86-0594 , 86-1134 , 86-1068 & 86-1963 , 86-0452 , 86-0703 , 86-0790 , 86-2227 , 86-0838 , 86-1115 , 86-1099 , 86-0112 & 86-0715 , 86-1521 , 86-1736 , 86-2006 , 86-0996 , 86-1541 & 86-1812 , 86-0460 , 86-1133 , 86-0291 , 86-0720 , 86-0828 , 86-1489 , 86-0845 , 86-1849 , 86-0549, 86-0673 & 86-1575 , 86-1393 , 86-1388 , 86-2039 , 86-0695 , 86-0138 , 86-2157 , 86-0641 , 86-0443 , 86-0365 , 86-0430 , 86-0722 , 86-1113 , 86-0897 , 86-0109 & 86-1778 , 86-0173 , 86-0798 , 86-0262 , 86-1505 , 86-0917 , 86-0680 , 86-1841 , 86-1404 , 86-1217 , 86-0951 , 86-1831 , 86-0458 , 86-2119 , 86-0215, 86-1720 & 86-1920 , 86-0275 , 86-1698 , 86-1818 , 86-2187 , 86-1278 , 86-1657 , 86-0791 , 86-0478 , 86-1373 , 86-0328 , 86-0891, 86-0893 & 86-1338 , 86-0895 , 86-0894 , 86-1791 , 86-0554 , 86-1495 , 86-2101 , 86-0776 , 86-0107, 86-0167, 86-0766 & 86-1176 , 86-0121 , 86-0767 , 86-1550 , 86-2226 , 86-0424 , 86-0164 , 86-0844 , 86-0753 , 86-1125 , 86-1018 , 86-1126, 86-1268 & 86-1269 , 86-1156 , 86-0413 , 86-0282 , 86-1000 , 86-1037 , 86-0886 , 86-0406 , 86-0159 , 86-0332 , 86-0873 , 86-0586 , 86-1430 , 86-2205 , 86-0521 , 86-0348 , 86-0778 , 86-0402 , 86-0836 , 86-0887 , 86-0792 , 86-1551 , 86-2139 , 86-0085 & 86-2097 , 86-1999 , 86-0339 , 86-0864 , 86-0822 , 86-0506 , 86-0270-RC , 86-2003 , 86-1746 & 86-1749 , 86-1057 , 86-0802 , 86-0212 , 86-2122 , 86-0783 , 86-2135 , 86-0338, 86-0553, 86-0862 & 86-1900 , 86-1157 , 86-2107 , 86-1316 , 86-1627 , 86-0279 , 86-1587 , 86-0702 , 86-0935-RC , 86-1915 , 86-1368 , 86-1590 , 86-0831 , 86-0746 , 86-0257 , 86-2111 , 86-0311 , 86-0136 , 86-1702 , 86-0334 & 86-1733 , 86-0596, 86-0613, 86-0614, 86-0868 & 86-1724 , 86-0309 , 86-0921 , 86-1855 , 86-1527 , 86-1216 , 86-1385 , 86-1695 , 86-0826 , 86-1529 , 86-1515 , 86-0375 , 86-1761 & 86-1764 , 86-0616 , 86-0859 , 86-0843 , 86-1819 , 86-0263 , 86-1982 , 86-0649 , 86-0227 , 86-0742 & 86-2004 , 86-0663 , 86-1638 , 86-0170 , 86-0097 & 86-0100 , 86-1581 , 86-2168 & 86-2198 , 86-0223 , 86-0683 , 86-0201 , 86-0203 & 86-1688 , 86-0769 , 86-0156 , 86-2232 & 86-2234 , 86-0057 , 86-1201 , 86-1287 , 86-1856 , 86-1027 , 86-1772 , 86-1260 , 86-1310 , 86-0817 , 86-0984 , 86-1670 , 86-0957 , 86-0119 , 86-2203 , 86-0333 , 86-1926 , 86-1704 , 86-0067 , 86-1870 , 86-0610 , 86-0597 , 86-0171, 86-0763 & 86-0883 , 86-0030 , 86-1138 , 86-2136 , 86-0061, 86-0453, 86-1097 & 86-1357 , 86-1300 , 86-0985 , 86-1342 , 86-1304 , 86-1204 , 86-2176 & 86-2177 , 86-0867 , 86-1659 & 86-0547 , 86-1184 & 86-1487 , 86-1055 , 86-1106 , 86-0623 & 86-1693 , 86-1003 , 86-0125 , 86-2197 , 86-1636 , 86-1532 , 86-0360 , 86-1158 , 86-0869 , 86-0533 , 86-1671 , 86-0765 , 86-0310 & 86-1630 , 86-1305 , 86-1307 , 86-0369 , 86-0652 , 86-1314 , 86-1829 , 86-1155 , 86-0190 , 86-0307 , 86-1572 , 86-0513 , 86-0250 , 86-1107 , 86-1588 , 86-0486 , 86-1207 , 86-0124 , 86-0098 , 86-0193 , 86-0049 & 86-0648 , 86-1906 , 86-1277 , 86-1121 , 86-0095 , 86-1124 , 86-1117 , 86-2055 , 86-1421 & 86-1445 , 86-0861 , 86-0660 , 86-1923 , 86-1140 , 86-2222 , 86-0525 , 86-0599 , 86-1722 , 86-1779 , 86-0962 & 86-1851 , 86-1423 , 86-1806 & 86-1957 , 86-0706 , 86-1696 , 86-1866 , 86-1425 , 86-2099 , 86-1500 & 86-2207 , 86-1345 , 86-0412 , 86-0932 , 86-1561, 86-1876 & 86-1878 , 86-0684 , 86-2164 , 86-1719 , 86-0383 & 86-1589 , 86-0114 , 86-1102 , 86-0322 , 86-0983 , 86-1998 , 86-1832 , 86-0396 , 86-0342 , 86-1815 , 86-1110, 86-1933 & 86-1986 , 86-0283 , 86-1637, 86-2038 & 86-2060 , 86-1491 , 86-1108 , 86-1927 , 86-2191 & 86-2192 , 86-0080 , 86-2130 , 86-0699 & 86-1241 , 86-0374, 86-1603, 86-1604 & 86-1605 , 86-1752 , 86-1514 , 86-0074 & 86-1147 , 86-1821 , 86-1781 , 86-0251 , 86-0229 , 86-0857 , 86-0920 , 86-1974 , 86-0559 & 86-0719 , 86-0366 , 86-1001 , 86-1845 , 86-1661 , 86-1881 , 86-0122 , 86-1076 , 86-0916 & 86-1435 , 86-1052 , 86-0384, 86-0385, 86-0636 & 86-1066 , 86-0281 , 86-1379 , 86-1377 , 86-1867 , 86-0631 , 86-0417 & 86-1503 , 86-1221 , 86-1383 , 86-2209 , 86-2208 , 86-0244 , 86-0408 & 86-1562 , 86-0490 , 86-1916 & 86-1917 , 86-0910-RC , 86-0797 , 86-1047, 86-1786 & 86-1940 , 86-0545 & 86-0735 , 86-1535 , 86-1852 , 86-1622 , 86-2007 , 86-2014 , 86-0199 , 86-1941 , 86-0650 , 86-0789, 86-1612 & 86-1613 , 86-1476 , 86-1742 , 86-2146 , 86-0540 , 86-1014 , 86-0038 , 86-0420, 86-0421 & 86-1566 , 86-1069 , 86-1079, 86-1082 & 86-1098 , 86-1143 , 86-1188 , 86-1510 , 86-1112 , 86-0470 , 86-1465 , 86-1231 , 86-1041 , 86-1297 , 86-0813, 86-0880 & 86-1251 , 86-0236 , 86-0976 , 86-0855 , 86-0084 & 86-0698 , 86-1429 , 86-0093 , 86-1654 , 86-0226, 86-0979 & 86-1382 , 86-1729 , 86-0550 , 86-0264 , 86-1888 , 86-1785 , 86-0444 , 86-1252 , 86-1595 & 86-1596 , 86-2237 , 86-0542 , 86-0999 , 86-1279 , 86-1694 , 86-0415, 86-0833 & 86-0900 , 86-0841 , 86-0543 , 86-0174 , 86-1214 , 86-0219 , 86-1053 , 86-0508 , 86-0904 , 86-1093 , 86-0394 , 86-0371 , 86-0576 , 86-1196 , 86-1669 , 86-0286 , 86-2138 , 86-1634 , 86-0644 , 86-0738 , 86-0688 & 86-1790 , 86-1903 , 86-0712 , 86-1136 , 86-0955 , 86-0687 & 86-2034 , 86-1182 , 86-1286 , 86-0571 , 86-1254 , 86-0507 , 86-0045 , 86-1402 & 86-1537 , 86-0192 , 86-0370 & 86-0620 , 86-1276 , 86-0284 , 86-1450 , 86-1451 & 86-1452 , 86-1816 , 86-0637 , 86-0989 , 86-1597 , 86-0592 , 86-1890 , 86-0600 & 86-1272 , 86-1953 , 86-1993 , 86-0608 , 86-2215, 86-2216, 86-1810, 86-2000, 86-2188 & 86-2189 , 86-1864 & 86-1939 , 86-0866 , 86-1058 & 86-2083 , 86-0642 , 86-1242 , 86-2025 , 86-0988 , 86-1422 , 86-0393 , 86-0662 , 86-1987 , 86-0149 , 86-1483 , 86-0014 & 86-0017 , 86-0581, 86-0678 & 86-0679 , 86-0222 & 86-1725 , 86-1739 , 86-0040 , 86-0852 , 86-0343 , 86-0029 , 86-0691 , 86-0033 , 86-0764 , 86-1570 & 86-1574 , 86-1013 , 86-0548 , 86-0349 , 86-0462 , 86-1600 , 86-0555 , 86-0677 , 86-0605 , 86-1616 & 86-1618 , 86-0304 & 86-1677 , 86-1061 , 86-1449 , 86-0382 , 86-2229 , 86-0909 , 86-0782 & 86-2112 , 86-1454 , 86-0846 , 86-1682 , 86-1734 , 86-0380 , 86-1392 , 86-2149 , 86-2163 , 86-1240 , 86-0952 , 86-0211 & 86-0690 , 86-0647 & 86-1712 , 86-0875 , 86-1715 , 86-1372 , 86-1034 , 86-1945 , 86-0147 , 86-0761 & 86-2167 , 86-0325 , 86-2127 , 86-1904 , 86-1771 , 86-1760 & 86-1847 , 86-0655 & 86-1135 , 86-0276 , 86-0477 , 86-1011 , 86-1183 , 86-1320 , 86-0233 , 86-0320 , 86-1060 , 86-0863 , 86-0718 , 86-0231 & 86-0457 , 86-0422 , 86-0748 , 86-0078 & 86-1330 , 86-0693 , 86-0874 , 86-1050 , 86-1051 , 86-0788 , 86-0858 , 86-1679 , 86-1556 , 86-0016 & 86-1144 , 86-0150 , 86-1042 & 86-1783 , 86-1441 , 86-1077 , 86-0741 , 86-1709 , 86-1173 & 86-2103 , 86-0269 , 86-0465 , 86-0146 & 86-1376 , 86-1224 & 86-1271 , 86-0431 & 86-1768 , 86-1374 , 86-1658 , 86-1922 , 86-0986 , 86-0034 , 86-0230 , 86-1899 , 86-2028 , 86-2081 , 86-1169 , 86-0168 , 86-0704 , 86-1651 , 86-2041 , 86-0261 , 86-1924 , 86-1171 & 86-1501 , 86-0337 , 86-0953 , 86-1968 , 86-0450 , 86-1868 , 86-2230 , 86-0161 , 86-0357 , 86-0301, 86-0303 & 86-1186 , 86-0177 , 86-1983 , 86-0787 , 86-2131 , 86-1502 , 86-2067 , 86-2145 , 86-1208 , 86-1895 , 86-0669 & 86-1664 , 86-0653 , 86-0796 , 86-1012 , 86-0646 , 86-0423 & 86-2072 , 86-0397 , 86-0039 , 86-0128 , 86-2049 , 86-0178 , 86-0265 , 86-1763 , 86-2225 , 86-0931 , 86-0258 , 86-1457 , 86-1608 , 86-1081 , 86-0754 , 86-1358 , 86-0670 , 86-1396 , 86-2090 , 86-1774 , 86-1996 & 86-1997 , 86-1617 , 86-1181 , 86-1875 , 86-1717 , 86-0196 & 86-1681 , 86-0661 , 86-0825 , 86-0217 , 86-1528 , 86-0111 & 86-1517 , 86-1623 , 86-0537 , 86-1710 , 86-1261 , 86-0123 & 86-0433 , 86-0530 , 86-0116-RC , 86-0617 , 86-1141 , 86-1990 , 86-0235 , 86-1675 , 86-2061 , 86-0368 , 86-1447 , 86-2223 , 86-0793 , 86-0024 , 86-1835 , 86-1573 , 86-1039 , 86-1757 , 86-1554 , 86-0290 , 86-1306 , 86-0856 , 86-1913 , 86-0050 , 86-0946 , 86-0392, 86-1473 & 86-2042 , 86-1750 , 86-1723 , 86-0141 , 86-2098 , 86-0090 , 86-2109 , 86-1378 , 86-0445 , 86-1127 & 86-1361 , 86-1485 , 86-1078 , 86-1753 , 86-1100 , 86-0934 , 86-1969 , 86-1281 , 86-2123 , 86-2089 , 86-0336 , 86-2162 , 86-2202 , 86-0216 , 86-1486 , 86-0240 , 86-0466 , 86-1031 , 86-1648 , 86-1801 , 86-0942 , 86-1685 , 86-0129 , 86-0410 , 86-1200 , 86-1017 , 86-0768 , 86-1534 , 86-2086 , 86-0241 , 86-1301 , 86-1642 , 86-1025 , 86-0967 & 86-2022 , 86-1718 , 86-0363 & 86-1970 , 86-0289 & 86-1526 , 86-1165 , 86-1882 , 86-1132 , 86-0407 , 86-1893 , 86-0595-RC , 86-0254 & 86-1518 , 86-2040 , 86-0574 , 86-0028 , 86-1943 , 86-0096 , 86-1545 , 86-1253 , 86-0179 & 86-1461 , 86-2108 , 86-1512 , 86-1020 , 86-0249 , 86-0205 , 86-0502 , 86-1777 , 86-1474 , 86-1956 , 86-0800 , 86-1359 , 86-0635 , 86-0094 , 86-2106 , 86-1751 , 86-0654 & 86-1105 , 86-1233 , 86-0801 , 86-0892 , 86-0377 & 86-0484 , 86-1599 , 86-2220 , 86-0835 , 86-1340 , 86-1384 , 86-1951 , 86-0021 , 86-0528 & 86-0773 , 86-0051 , 86-1436 , 86-1814 , 86-1564 , 86-0062 , 86-1770 , 86-0588-RC , 86-0820 , 86-1160 , 86-1086 , 86-1048 , 86-0326 , 86-1859 , 86-1689 , 86-2033 , 86-0376 , 86-1901 , 86-1530 , 86-0524 , 86-0651 , 86-0312 , 86-0442 , 86-1023 , 86-0760 , 86-0593 , 86-1109 , 86-2012 , 86-0056 , 86-0627 , 86-1089 , 86-0340 , 86-0993 & 86-0998 , 86-0645 , 86-1504 , 86-1643 , 86-1309 , 86-1994 , 86-0902-RC , 86-1400 , 86-0752 , 86-2062 , 86-0975 , 86-1837 , 86-1180 , 86-0089 , 86-0451 , 86-1655 , 86-0073 , 86-1440 & 86-1442 , 86-0256 , 86-0330 , 86-0354 , 86-1295 , 86-1769 , 86-0954 , 86-0437 , 86-1850 & 86-2154 , 86-0418 , 86-1159 , 86-2114 , 86-0355 , 86-1862 , 86-2214 , 86-1129 , 86-0103 & 86-1394 , 86-0313 & 86-0400 , 86-0602 , 86-1232 , 86-0041 , 86-0335 , 86-1558 & 86-1563 , 86-0837 , 86-0088 , 86-2058 , 86-0707 , 86-1185 , 86-2069 , 86-0638 , 86-1280 , 86-0936 , 86-0449 & 86-0479 , 86-0234 & 86-1104 , 86-2213 , 86-0278 & 86-2212 , 86-1363 , 86-1789 , 86-0749 , 86-0152 , 86-0438 , 86-0172 & 86-1824 , 86-0072, 86-0087, 86-0434, 86-0938 & 86-1672 , 86-1758 , 86-0395 , 86-1674 , 86-0818 & 86-0819 , 86-0580 & 86-1971 , 86-2173 , 86-0022 , 86-1936 , 86-0296 , 86-0027 & 86-0812 , 86-0656 , 86-0207 , 86-1087 , 86-0183 , 86-0367 , 86-0604 , 86-0710 , 86-1049 , 86-1128 , 86-1479 , 86-1549 , 86-1755 , 86-0037 & 86-1458 , 86-0500 , 86-1275 , 86-0842 , 86-0175 , 86-0672 & 86-1327 , 86-0834 , 86-1364 , 86-0740 & 86-0965 , 86-0884 & 86-0885 , 86-1569 , 86-0830 , 86-0023 , 86-1692 , 86-0391 , 86-0118 , 86-0681 , 86-1028 , 86-0911 , 86-0982 , 86-0947 & 86-2174 , 86-0948 , 86-0682 , 86-0341 , 86-1193 , 86-1571 , 86-1322 , 86-0344 , 86-1063 , 86-0387 , 86-0619 & 86-0459 , 86-1389 , 86-1522 , 86-0079 & 86-1591 , 86-0130 , 86-0733 , 86-0958 , 86-0569 , 86-0724 , 86-0532, 86-0603 & 86-1177 , 86-1857 , 86-1592, 86-1601 & 86-1602 , 86-0436 , 86-0590 , 86-1477 , 86-1663 , 86-0694 & 86-1559 , 86-0615 , 86-1929 , 86-1825 , 86-0043 & 86-1439 , 86-1170 , 86-1773 , 86-0483 , 86-0345 , 86-0879 , 86-1318 & 86-1547 , 86-1593 , 86-0496 , 86-0509 , 86-1662 , 86-0971 , 86-1270 , 86-0243 , 86-0285 & 86-2104 , 86-1932 , 86-2102 , 86-0940 , 86-1103 , 86-1735 , 86-1942 , 86-0031 & 86-1116 , 86-0568 & 86-1337 , 86-1245 , 86-1743 , 86-2169 , 86-0467 , 86-0405 , 86-0562 , 86-1683 , 86-0629 , 86-1937 , 86-1676 , 86-1977 , 86-2124 , 86-0063 , 86-0246-RC , 86-1064 , 86-1328 , 86-2190 , 86-2027 , 86-0429 , 86-0882 , 86-1244 & 86-2175 , 86-0824 , 86-0696 & 86-2233 , 86-1482 , 86-0725 , 86-0065 , 86-1496 , 86-0232 & 86-1229 , 86-0780 , 86-1080 , 86-1054 , 86-0379 , 86-1707 & 86-1708 , 86-1594 , 86-1705 , 86-1578 , 86-1265 , 86-1024 , 86-2059 , 86-0751 , 86-1030 , 86-0248 , 86-1700 , 86-2228 , 86-2120 , 86-0762 , 86-1800 , 86-1455 , 86-0512, 86-0860 & 86-1684 , 86-1842 , 86-1313 , 86-2036 , 86-1902 , 86-0714 , 86-1291 , 86-0474 , 86-0705 & 86-0794 , 86-1538 , 86-1131 , 86-0643 , 86-0561 , 86-2009 , 86-1344 , 86-0881 , 86-0872 , 86-0019 , 86-1691 , 86-2218 , 86-1152 , 86-1492 , 86-0839 & 86-1557 , 86-1325 , 86-0591 , 86-1756 , 86-0242 , 86-1741 & 86-2088 , 86-1323 , 86-2204 , 86-0404 , 86-1497 , 86-0823 & 86-1995 , 86-0110 , 86-1865 , 86-1641 , 86-0213 , 86-1146 , 86-0426 , 86-1032 , 86-2079 , 86-0101 , 86-1964 , 86-1908 , 86-1706 , 86-2132 , 86-0456 , 86-0469, 86-0611 & 86-1887 , 86-0104 & 86-1227 , 86-0816 & 86-1228 , 86-2023 & 86-2024 , 86-0158 , 86-0582 , 86-1148 , 86-2206 , 86-2074 , 86-0734 & 86-1928 , 86-0758 & 86-0759 , 86-1362 , 86-1438 , 86-1730 , 86-0922 , 86-1312 , 86-1493 , 86-1021, 86-1161 & 86-1976 , 86-1975 , 86-0939 , 86-0060 , 86-0026, 86-0489, 86-1130 & 86-1329 , 86-1699 , 86-0165 , 86-1114 & 86-1437 , 86-1246, 86-1339 & 86-1826 , 86-1586 , 86-1759 , 86-0372 , 86-1804 , 86-0188 , 86-2096 , 86-0743 , 86-1628 , 86-2217 & 86-2221 , 86-1820 , 86-1673 , 86-1958 , 86-1084 , 86-0956 , 86-0252 & 86-1805 , 86-0526 , 86-0493 , 86-1366 , 86-1827 , 86-1853 , 86-1905 , 86-1910 , 86-1944 , 86-0963 , 86-1921 , 86-2219 , 86-1453 , 86-0686 , 86-0044 , 86-0626-RC , 86-0854 , 86-0578 , 86-0409 , 86-1367, 86-1433 & 86-1981 , 86-0495, 86-0498 & 86-1985 , 86-2076 , 86-2029 , 86-0139 , 86-1660 , 86-1846 , 86-1243 , 86-1434 , 86-0557 , 86-1650 , 86-1863 & 86-2001 , 86-2082 , 86-0774 & 86-1016 , 86-1802 , 86-1292 , 86-2020, 86-2021 & 86-2093 , 86-0108 & 86-0536 , 86-0186 , 86-1498 , 86-1083 , 86-0015-R , 86-1282 , 86-0160 , 86-0566 & 86-1843 , 86-0300 , 86-1989 , 86-1649 , 86-2129 , 86-0055 , 86-1711 , 86-1222 , 86-0781 , 86-1738 , 86-2016 , 86-0747 , 86-0676 , 86-1002 , 86-1315 , 86-1680 , 86-1787 , 86-1321 , 86-0905 & 86-1731 , 86-1896 , 86-1149 , 86-0675 , 86-2125 , 86-1744 , 86-0294, 86-2053 & 86-2054 , 86-2019 , 86-1386 & 86-1387 , 86-0455 & 86-0689 , 86-1062, 86-1198, 86-1347, 86-1484, 86-2166 & 86-2172 , 86-1256 , 86-0851 , 86-0218 & 86-0853 , 86-1459 , 86-2116 , 86-1163 & 86-1164 , 86-1035 , 86-0464, 86-1194 & 86-1195 , 86-0814 , 86-0353 , 86-1046 , 86-1828 , 86-1044 , 86-0176 , 86-2236 , 86-0505 , 86-1225 , 86-0709 , 86-1740 & 86-1950 , 86-0531 , 86-0356 , 86-0744 , 86-1911 & 86-1912 , 86-0132 , 86-0809 & 86-1553 , 86-0499 & 86-1809 , 86-0926 , 86-0471 , 86-1947 , 86-0472-RC, , 86-0018 , 86-0052 , 86-0053 , 86-0054 , 86-0059 , 86-0066 , 86-0081 , 86-0083 , 86-0126 , 86-0127 , 86-0133 , 86-0134 , 86-0135 , 86-0142 , 86-0143 , 86-0144 , 86-0145 , 86-0169 , 86-0180 , 86-0181 , 86-0187 , 86-0195 , 86-0197 , 86-0202 , 86-0208 , 86-0209 , 86-0210 , 86-0220 , 86-0221 , 86-0225 , 86-0266 , 86-0267 , 86-0268 , 86-0272 , 86-0273 , 86-0274 , 86-0280 , 86-0287 , 86-0288 , 86-0292 , 86-0293 , 86-0295 , 86-0297 , 86-0302 , 86-0305 , 86-0314 , 86-0315 , 86-0316 , 86-0317 , 86-0318 , 86-0319 , 86-0323 , 86-0324 , 86-0350 , 86-0351 , 86-0362 , 86-0373 , 86-0386 , 86-0398 , 86-0414 , 86-0439 , 86-0440 , 86-0446 , 86-0447 , 86-0448 , 86-0461 , 86-0463 , 86-0473 , 86-0476 , 86-0480 , 86-0481 , 86-0488 , 86-0491 , 86-0497 , 86-0503 , 86-0510 , 86-0514 , 86-0516 , 86-0518 , 86-0522 , 86-0523 , 86-0534 , 86-0535 , 86-0538 , 86-0539 , 86-0563 , 86-0564 , 86-0565 , 86-0587 , 86-0598 , 86-0606 , 86-0607 , 86-0612 , 86-0621 , 86-0622 , 86-0630 , 86-0632 , 86-0633 , 86-0634 , 86-0659 , 86-0665 , 86-0666 , 86-0667 , 86-0671 , 86-0708 , 86-0713 , 86-0717 , 86-0721 , 86-0723 , 86-0726 , 86-0727 , 86-0728 , 86-0729 , 86-0730 , 86-0756 , 86-0771 , 86-0772 , 86-0779 , 86-0784 , 86-0803 , 86-0807 , 86-0808 , 86-0810 , 86-0815 , 86-0821 , 86-0847 , 86-0848 , 86-0849 , 86-0876 , 86-0877 , 86-0878 , 86-0888 , 86-0889 , 86-0898 , 86-0899 , 86-0907 , 86-0908 , 86-0919 , 86-0924 , 86-0927 , 86-0928 , 86-0929 , 86-0930 , 86-0933 , 86-0941 , 86-0943 , 86-0944 , 86-0945 , 86-0950 , 86-0959 , 86-0960 , 86-0961 , 86-0968 , 86-0969 , 86-0970 , 86-0990 , 86-0991 , 86-0992 , 86-1004 , 86-1005 , 86-1006 , 86-1007 , 86-1008 , 86-1010 , 86-1019 , 86-1038 , 86-1040 , 86-1071 , 86-1072 , 86-1073 , 86-1074 , 86-1075 , 86-1088 , 86-1090 , 86-1091 , 86-1092 , 86-1096 , 86-1101 , 86-1118 , 86-1120 , 86-1122 , 86-1137 , 86-1145 , 86-1153 , 86-1168 , 86-1172 , 86-1174 , 86-1189 , 86-1190 , 86-1191 , 86-1197 , 86-1199 , 86-1202 , 86-1203 , 86-1206 , 86-1215 , 86-1220 , 86-1235 , 86-1236 , 86-1237 , 86-1238 , 86-1239 , 86-1247 , 86-1257 , 86-1262 , 86-1263 , 86-1264 , 86-1284 , 86-1285 , 86-1288 , 86-1289 , 86-1290 , 86-1294 , 86-1296 , 86-1298 , 86-1299 , 86-1333 , 86-1334 , 86-1335 , 86-1336 , 86-1341 , 86-1348 , 86-1349 , 86-1350 , 86-1351 , 86-1352 , 86-1353 , 86-1354 , 86-1355 , 86-1356 , 86-1365 , 86-1370 , 86-1371 , 86-1380 , 86-1390 , 86-1395 , 86-1397 , 86-1401 , 86-1403 , 86-1405 , 86-1406 , 86-1407 , 86-1408 , 86-1409 , 86-1410 , 86-1411 , 86-1412 , 86-1413 , 86-1414 , 86-1415 , 86-1416 , 86-1417 , 86-1418 , 86-1419 , 86-1426 , 86-1427 , 86-1444 , 86-1462 , 86-1463 , 86-1464 , 86-1466 , 86-1467 , 86-1468 , 86-1469 , 86-1470 , 86-1471 , 86-1472 , 86-1507 , 86-1508 , 86-1511 , 86-1516 , 86-1519 , 86-1520 , 86-1524 , 86-1525 , 86-1533 , 86-1543 , 86-1546 , 86-1552 , 86-1567 , 86-1568 , 86-1577 , 86-1584 , 86-1585 , 86-1598 , 86-1624 , 86-1629 , 86-1639 , 86-1644 , 86-1645 , 86-1653 , 86-1656 , 86-1665 , 86-1666 , 86-1667 , 86-1668 , 86-1716 , 86-1726 , 86-1745 , 86-1747 , 86-1748 , 86-1762 , 86-1765 , 86-1766 , 86-1767 , 86-1776 , 86-1788 , 86-1792 , 86-1793 , 86-1803 , 86-1811 , 86-1817 , 86-1833 , 86-1848 , 86-1861 , 86-1871 , 86-1872 , 86-1873 , 86-1877 , 86-1897 , 86-1898 , 86-1919 , 86-1930 , 86-1931 , 86-1934 , 86-1935 , 86-1948 , 86-1952 , 86-1961 , 86-1967 , 86-1973 , 86-1978 , 86-1979 , 86-1980 , 86-1991 , 86-2010 , 86-2013 , 86-2037 , 86-2045 , 86-2046 , 86-2048 , 86-2050 , 86-2051 , 86-2052 , 86-2063 , 86-2064 , 86-2066 , 86-2091 , 86-2094 , 86-2105 , 86-2134 , 86-2141 , 86-2142 , 86-2143 , 86-2144 , 86-2150 , 86-2151 , 86-2152 , 86-2153 , 86-2155 , 86-2156 , 86-2158 , 86-2159 , 86-2160 , 86-2161 , 86-2195 , 86-2196 , 86-2200 , 86-2201 , 86-2231 , and 86-2235.
Recalling Firm/
Manufacturer
GE OEC Medical Systems, Inc
384 Wright Brothers Dr
Salt Lake City, Utah 84116-2862
For Additional Information Contact Deena Pease
801-536-4516
Manufacturer Reason
for Recall
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The OEC 6800 Miniview was recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lock-up, incorrect image display, incorrect dose area calculated air kerma in "film mo
FDA Determined
Cause 2
DESIGN: Device Design
Action Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. On 11/20/2006, two written notifications were sent to 1513 identified consignees, addressed to the Facility Administrator and the Radiology Administrator for a combined total of 3026 letters. The Urgent Recall Notice, dated 11/8/2006, advised customers that GE Healthcare had identified several intermittent potential safety issues that may occur with the identified affected product. The notification also provided interim or permanent solutions for each of the problems described. For the issues that did not have permanent solutions at the time, the firm was to contact customers and remedy the issues free of charge. For the issues that did have a permanent fix at the time, the GEHC OEC service team was to contact customers to install the correction. If customers had questions or concerns, they were to contact the service team for further information at 800-874-7378 option 8. On 04/23, 2007, a second written notification was sent via certified mail to the same 1513 consignees providing further clarification on the issue of "incorrect Image Display" and a temporary remedy to the issue.
Quantity in Commerce 1,933 units
Distribution Worldwide Distribution