• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
 Class 2 Recall
GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes
see related information
Date Posted August 22, 2012
Recall Status1 Terminated on August 22, 2012
Recall Number Z-2241-2012
Recall Event ID 62729
Premarket Notification
510(K) Numbers
K021049  K022069  K024012  K041932 
Product Classification Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Code OXO
Product GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes Snubber Boards are a PCB with a 70 amp fuse to provide fused protection of internal system components from the high voltage system components, such as the X-ray tube or High Voltage Transformer. System model number A349855, Snubber Board number 00-880405-03 rev 5. This product is used for image intensified fluoroscopic x-rays for interventional radiologic procedures.
Code Information Serial numbers: 69-0128, 82-0157, 82-0158, 82-0185, 82-0191, 82-0210, 82-0215, 82-0223, 82-0229, 82-0266, 82-0324, 82-0328, 82-0464, 82-0477, 82-0484, 82-0505-RC, 82-0507, 82-0542, 82-0544, 82-0546, 82-0551, 82-0604, 82-0648, 82-0673, 82-0680, 82-0714, 82-0733, 82-0779, 82-0858, 82-0869, 82-0884, 82-0906, 82-0918, 82-0965, 82-1004, 82-1009, 82-1057-RC, 82-1070, 82-1139, 82-1191, 82-1194, 82-1225, 82-1247, 82-1303, 82-1405, 82-1434, 82-1458, 82-1480, 82-1534, 82-1547-RC, 82-1564, 82-1585, 82-1588, 82-1601, 82-1604, 82-1621, 82-1694, 82-1695, 82-1699, 82-1797, 82-1798, 82-1842, 82-1850, 82-1921, 82-1959, 82-2015, 82-2120, 82-2192, 82-2245, 82-2247, 82-2274, 82-2277, 82-2306, 82-2386, 82-2487, 82-2661, 82-2662, 82-2716, 82-2721, 82-2741, 82-2745, 82-2797, 82-2846, 82-2905, 82-2965, 82-3102, 82-3107, 82-3122, 82-3135, 82-3151, 82-3351, 82-3366, 82-3403, 82-3431, 82-3492, 82-3498, 82-3563, 82-3570, 82-3614, 82-3634, 82-3762, 82-3815, 82-3819, 82-3824, 82-3839, 82-3851, 82-3882, 82-3884, 82-3887, 82-3888, 82-3890, 82-3894, 82-3895, 82-3896, 82-3898, 82-3899, 82-3901, 82-3903, 82-3904, 82-3910, 82-3911, 82-3912, 82-3913, 82-3914, 82-3915, 82-3917, 82-3919, 82-3920, 82-3921, 82-3922, 82-3923, 82-3925, 82-3928, 82-3929, 82-3930, 82-3931, 82-3933, 82-3934, 82-3935, 82-3936, 82-3937, 82-3938, 82-3939, 82-3941, 82-3942, 82-3943, 82-3944, 82-7011-MH, 82-7024-MH, 82-7154-CMH, 82-7189-MH, 82-7190-MH, 82-7191-MH, 82-7197-MH, 82-7200-MH, 82-7206-MH, 89-0100, 89-0128, 89-0135-RC, 89-0210, 89-0248, 89-0315, 89-0362, 89-0385, 89-0401, 89-0416, 89-0457, 89-0464, 89-0467, 89-0505, 89-0506, 89-0537, 89-0556, 89-0559, 89-0577, 89-0612, 89-0643, 89-0660, 89-0662, 89-0729, 89-0738, 89-0780, 89-0781, 89-0832, 89-0853, 89-0868, 89-0876, 89-0882, 89-0922, 89-0954-RC, 89-0963, 89-1040, 89-1071, 89-1172, 89-1195, 89-1228, 89-1245, 89-1248, 89-1255, 89-1262, 89-1268, 89-1271, 89-1274, 89-1286, 89-1295, 89-1308, 89-1343, 89-1373, 89-1400-RC, 89-1408, 89-1426, 89-1494, 89-1531, 89-1576, 89-1598, 89-1606, 89-1612, 89-1640, 89-1663, 89-1673, 89-1721, 89-1729, 89-1739, 89-1749, 89-1868, 89-2121, 89-2259, 89-2264, 89-2275, 89-2309, 89-2315, 89-2350, 89-2354, 89-2456, 89-2464, 89-2500, 89-2680, 89-2866, 89-3211, 89-3294, 89-3312, 89-3349, 89-3352, 89-3355, 89-3377, 89-3420, 89-3433, 89-3461, 89-3479, 89-3483, 89-3488, 89-3490, 89-3494, 89-3495, 89-3499, 89-3501, 89-3504, 89-3508, 89-3509, 89-3511, 89-3512, 89-3513, 89-3514, 89-3515, 89-3519, 89-3520, 89-3522, 89-3524, 89-3527, 89-3529, 89-3530, 89-3533, 89-3537, 89-3538, 89-3539, 89-3543, 89-3544, 89-3545, 89-3546, 89-3547, 89-3548, 89-3549, 89-3551, 89-3552, 89-3553, 89-3554, 89-3555, 89-3556, 89-3557, 89-3558, 89-3559, 89-3560, 89-3561, 89-3562, 89-3563, 89-3564, 89-3565, 89-3567, 89-3568, 89-3569, 89-3570, 89-3571, 89-3572, 89-3573, 89-3574, 89-3578, 89-3579, 8S-0027, 8S-0076-RC, 8S-0086, 8S-0121-C, 8S-0158, 8S-0165, 8S-0245, 8S-0291, 8S-0301-C, 8S-0315, 8S-0413-C, 8S-0494, 8S-0615-C, 8S-0642, 8S-0691, 8S-0707, 8S-0711, 8S-0735, 8S-0747, 8S-0776, 8S-0840, 8S-0844-RC, 8S-0858, 8S-0867, 8S-0872, 8S-0921, 8S-0974, 8S-1013-C, 8S-1046, 8S-1056, 8S-1058, 8S-1065-C, 8S-1174, 8S-1203, 8S-1224-C-RC, 8S-1285-RC, 8S-1360, 8S-1445, 8S-1644, 8S-1856, 8S-1930-N, 8S-1954, 8S-1959, 8S-1961, 8S-2052, 8S-2053, 8S-2063, 8S-2063, 8S-2113, 8S-2145, 8S-2164, 8S-2170, 8S-2207, 8S-2256, 8S-2329, 8S-2439, 8S-2451, 8S-2452, 8S-2452, 8S-2462, 8S-2532, 8S-2622, 8S-2865, 8S-2956, 8S-2985, 8S-3002, 8S-3108, 8S-3111, 8S-3112, 8S-3130, 8S-3179, 8S-3183, 8S-3197, 8S-3205, 8S-3207, 8S-3211-CH, 8S-3216, 8S-3222, 8s-3231, 8S-3233, 8S-3243, 8S-3245, 8S-3246, 8S-3247, 8S-3248, 8S-3249, 8S-3253, 8S-3254, 8S-3255, 8S-3256-H, 8S-3257, 8S-3258-CH, 8S-3259, 8S-3260, 8S-3261, 8S-3262, 8S-3263, 8S-3264, 8S-3265, 8S-3266, 8S-3267, 8S-3268, 8S-3269-H, 8S-3270-H, 8S-3271-H, 8S-3272-H, 8S-3273-CH, 8S-3274-CH, 8S-3275, 8S-3276, 8S-3278, 8S-3279, 8S-3280, 8S-3281, 8S-3282, 8S-3283, 8S-3285, 8S-3286, 8S-3287, 8S-3288-H, 8S-3289-H, 8S-3291, 8S-3292, 8S-3296, 8S-3299, 8S-3301-H, 8S-3302, 8S-3305, 8S-3306, 8S-3307, 8S-3308, 8S-3309, 8S-3310-H, 8S-3311, 8S-3315, 8S-7094-CMH, 8S-7228-CMH, 8S-7229-CMH, 8S-7230-CMH, 8S-7231-MH, 8S-7232-MH, 8S-7233-MH, E2-0001, E2-0015, E2-0026, E2-0028, E2-0034, E2-0078, E2-0084, E2-0085, E2-0108, E2-0110, E2-0111, E2-0117, E2-0120, E2-0122, E2-0130, E2-0142, E2-0143, E2-0148, E2-0151, E2-0152, E2-0153, E2-0154, E2-0155, E2-0156, E2-0157, E2-0158, E2-0159, E2-0160, E2-0161, E2-0162, E2-0163, E2-0164, E2-0165, E2-0167, E2-0168, E2-0169, E2-0171, E2-0172, E2-0173, E2-0174, E2-0175, E2-0176, E2-0177, E2-0178, E2-0179, E2-0180, E2-0181, E2-0182, E2-0183, E2-0184, E2-0185, E2-0186, E2-0187, E2-0188, E2-0189, E2-0190, E2-0191, E2-0192, E2-0193, E2-0194, E2-0195, E2-0196, E2-0197, E2-0198, E2-0199, E2-0200, E2-0201, E2-0202, E2-0203, E2-0204, E2-0205, E2-0206, E2-0207, E2-0208, E2-0209, E2-0210, E2-0211, E2-0212, E2-0213, E2-0215, E2-0216, E2-0217, E2-0222, E2-0224, E2-0225, E2-0230, E2-0231, E2-0232, E2-0233, E2-0234, E2-0235-C, E2-0237, E2-0238, E2-0239, E2-0240, E2-0241, E2-0242, E2-0243, E2-0244, E2-0245, E2-0246, E2-0247, E2-0248, E2-0250, E2-0251, E2-0252, E2-0253, E2-0254, E2-0255, E2-0256, E2-0257, E2-0258, E2-0259, E2-0260, E2-0261, E2-0262, E2-0263, E2-0264, E2-0265, E2-0266, E2-0267, E2-0268, E2-0269, E2-0270, E2-0271, E2-0272, E2-0273, E2-0274, E2-0275, E2-7014-MH, E2-7020-MH, E2-7048-MH, E2-7050-MH, E2-7051-MH, E2-7057-MH, E2-7058-MH, E2-7059-MH, E2-7060-MH, E2-7061-MH, E2-7062-MH, E2-7063-MH, E2-7064-MH, E2-7065-MH, E2-7066-MH, E2-7067-MH, E2-7068-MH, E2-7069-CMH, E2-7070-MH, E2-7071-MH, E2-7072-MH, E2-7074-MH, E2-7075-MH, E2-7076-MH, E2-7077-CMH, E2-7078-MH, E2-7079-MH, E9-0008, E9-0025, E9-0031, E9-0035-G, E9-0037, E9-0050, E9-0083, E9-0085, E9-0086, E9-0088, E9-0089-G, E9-0090, E9-0091, E9-0092, E9-0093, E9-0094, E9-0095, E9-0098, E9-0099, E9-0100, E9-0101, E9-0102, E9-0103, E9-0108, E9-0109, E9-0110, E9-0111, E9-0112, E9-0113, E9-0114, E9-0115, E9-0116, E9-0118, E9-0119, E9-0120, E9-0121, E9-0122, E9-0123, E9-0124, E9-0125, E9-0126, E9-0127, E9-0128, E9-0129, E9-0129, E9-0130, E9-0139, E9-0141, E9-0142, E9-0143, E9-0144, E9-0145, E9-0146, E9-0147, E9-0148, E9-0149, E9-0150, E9-0151, E9-0152, E9-0153, E9-0155, E9-0157, E9-0158, E9-0159, E9-0162, E9-0163, E9-0164, E9-0165, E9-0166, E9-0169, E9-0170, E9-0171, E9-0172, E9-0173, E9-0177, E9-0178, E9-0179, E9-0180, E9-0181, E9-0183, E9-0184, E9-0185, E9-0188, E9-0249, E9-0253, ES-0005, ES-0006, ES-0007-C, ES-0022-CH, ES-0029-CH, ES-0034-G, ES-0056, ES-0069, ES-0071, ES-0072, ES-0076, ES-0078, ES-0084-G, ES-0086-G, ES-0087, ES-0088, ES-0089, ES-0090, ES-0091-H, ES-0092-CH, ES-0096, ES-0097, ES-0098, ES-0099-CH, ES-0101-CH, ES-0102-CH, ES-0103, ES-0104, ES-0105, ES-0106, ES-0107, ES-0108, ES-0109, ES-0110, ES-0111, ES-0112, ES-0113, ES-0114, ES-0115, ES-0116, ES-0117, ES-0118, ES-0119, ES-0120, ES-0121, ES-0122, ES-0123, ES-0124, ES-0125, ES-0126, ES-0127, ES-0128, ES-0129, ES-0130, ES-0131-H, ES-0132, ES-0133, ES-0134, ES-0135, ES-0136, ES-0138, ES-0139, ES-0140, ES-0141, ES-0142, ES-0147, ES-0152, ES-0155-H, ES-0156, ES-0158, ES-0159, ES-0160, ES-0163-G, ES-0169-CH, ES-0170, ES-0171, ES-0172, ES-0176, ES-0178, ES-0180, ES-0181, ES-0182, ES-0184-CH, ES-0185-H, ES-0186-H, ES-0187-H, ES-0190-CH, ES-0191-H, ES-0192-H, ES-7010-CMH, ES-7015-MH, ES-7018-MH, ES-7021-MH, ES-7028-CMH, ES-7029-MH, ES-7030-CMH, ES-7031-CMH, ES-7033-CMH, ES-7034-CMH, ES-7035-CMH, ES-7036-CMH, ES-7037-CMH, ES-7039-MH.
Recalling Firm/
Manufacturer
GE OEC Medical Systems, Inc
384 Wright Brothers Dr
Salt Lake City, Utah 84116-2862
Consumer Instructions No consumer action necessary
For Additional Information Contact Ms. Deena Pease
801-536-4952
Manufacturer Reason
for Recall
GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.
FDA Determined
Cause 2
CHANGE CONTROL (GMP - GOOD MANUFACTURING PRACTICE): Component Change Control
Action No customer communication took place. On April 26, 2010, GE began opening service cases for field engineers to begin replacing the snubber boards. This action has been completed as of October 29, 2010.
Quantity in Commerce 845 units
Distribution Worldwide and Nationwide (USA) distribution.
Total Product Life Cycle TPLC Device Report

 
1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = OXO and Original Applicant = GE OEC MEDICAL SYSTEMS
510(K)s with Product Code = OXO and Original Applicant = GE OEC MEDICAL SYSTEMS, INC.
-
-