| | Class 2 Recall Capsule Neuron Docking Station |  |
| Date Posted |
September 13, 2012 |
| Recall Number |
Z-2377-2012 |
| Product |
Capsule Neuron Docking Station, a component in the DataCaptor Connectivity System
Catalog Number DC-NU-DS-CK
The Capsule Neuron Docking Station is a component in the DataCaptor Connectivity System. The Capsule Neuron Docking Station provides connection for the Capsule Neuron UMPC which is a bedside device that enables automatic collection of vital signs data that can connect to the hospital network through a wired Ethernet connection or via a standard 802.11 b/g/n wireless network. |
| Code Information |
Serial Numbers
TS9801371 -TS9801650;
TS9900670 - TS9900679;
TS9B00413 - TS9B00722;
TS9B01753 - TS9B02041;
TSA301587 - TSA302186;
TSA600246 - TSA600845
TSA801222 - TSA801821;
TSAB00031 - TSAB00629;
TSB300410 - TSB300912;
TSB302698 - TSB303201;
TSB401570 - TSB402012;
TSB500633 - TSB501136;
TSB502359 - TSB502861;
TSB600767 - TSB601270;
TSB700368 - TSB700871;
TSB801147 - TSB803555;
TSB902081 - TSB902585;
TSBA03569 - TSBA03595;
TSBB00027 - TSBB00537;
TSBC00200 - TSBC01321;
TSC100319 - TSC101318;
TSC101972 - TSC103113;
TSC301294 - TSC301299;
TSC301389 - TSC303425;
TSC401639 - TSC401995;
TSC402020 - TSC402087;
TSC402136 - TSC402346
|
Recalling Firm/ Manufacturer |
Capsule Tech Inc. 300 Brickstone Sq Suite 203 Andover, Massachusetts 01810-1492 |
| Consumer Instructions |
Contact the recalling firm for information |
| For Additional Information Contact |
Peter Kelley 978-482-2309
|
Reason for Recall |
Capsule Neuron Docking Station may experience a lost connection, resulting in loss of data or loss of power to the Caspsule Neuron
|
| Action |
The firm, Capsule Tech, sent a recall letter dated August 10,2012 to its customers. The letter described the product, problem and actions to be taken. Capsule is sending field correction kits to customers with instructions. Capsule's Professional Services team will contact the customers to schedule shipment of their kit. Final shipments of the field correction kits will be completed by Oct 2012.
If you have any questions or concerns, contact your professional services representative and/or call in US: 978-482-2340 or France: +33 153-34-14-00. |
| Quantity in Commerce |
12,448 units |
| Distribution |
Worldwide distribution: USA (nationwide) and countries of: Belgium, Canada, Denmark, France, Germany, Netherlands and UK. |
| |
|
|
|
|