Class 2 Device Recall Philips Medical System Extended Brilliance Workspace (EBW), Computed Tomography

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
 Class 2 Recall
Philips Medical System Extended Brilliance Workspace (EBW), Computed Tomography
see related information
Date Posted March 05, 2013
Recall Status1 Terminated on December 26, 2013
Recall Number Z-0906-2013
Recall Event ID 64259
Premarket Notification
510(K) Number
K961464 
Product Classification System, X-Ray, Tomography, Computed - Product Code JAK
Product Philips Medical System Extended Brilliance Workspace (EBW), Computed Tomography X-Ray System. The Extended Brilliance Workspace (EBW) is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Code Information Model 728260, Serial #, 3467,5610,6151,6204,6209,6221,6437,6483,6543,6636,6658,6717,6788,6865,7154,7186,7253,7470, 7507,7576,7579,7639,7690,7692,7708,7835,7977,8099,8255,8332,8443,8756,8757,8890,8922,9079, 9201,9395,9422,10085,10086,10090,10102,10138,10155,10180,10197,10216,10240,10261,10264,10271, 10312, 10324,10328,10346,10355,10359,10378,10383,10389,10422,10427,10431,10435,10437,10442, 10455,10456,10470,10476,10492,10537,10547,10561,10584,10598,10604,10606,10622,10630,10646, 10672,10686,10699,10722,10727,10738,10745,10747,10770,10776,10778,10796,10798,10817,10821, 10822,10823,10838,10841,10874,10889,10890,10897,10952,10953,10965,10980,10985,10988,10998, 11007,11010,11025,11036,11037,11043,11051,11054,11058,11062,11065,11080,11083,11085,11086, 11088,11089,11105,11107,11113,11121,11127,11128,11140,11141,11147,11148,11149,11151,11152, 11153,11166,11169,11173,11174,11208,11221,11230,11232,11241,11243,11260,11262,11267,11274, 11278,11280,11282,11291,11314,11317,11324,11334,11339,11341,11557,11561,11564,11567,11572, 11577,11581,11585,11603,11610,11612,11620,11622,11630,11631,11633,11644,11649,11650,11656, 11657,11659,11660,11662,11663,11667,11668,11670,11673,11674,11684,11685,11692,11694,11696, 11699,11700,11703,11706,11708,11711,11714,11717,11718,11720,11729,11739,11740,11742,11743, 11744,11745,11747,11748,11749,11750,11751,11753,11754,11755,11756,11758,11760,11762,11765, 11766,11768,11769,11770,11772,11773,11774,11776,11779,11780,11781,11782,11784,11785,11788, 11790,11791,11793,11794,11795,11796,11797,11798,11800,11801,11802,11803,11804,11805,11806, 11807,11808,11809,11810,11812,11813,11814,11816,11817,11818,11819,11822,11823,11830,11831, 11832,11833,11834,11838,11839,11840,11841,11842,11843,11844,11845,11846,11847,11848,11850, 11851,11852,11854,11856,11858,11859,11862,11863,11864,11865,11866,11868,11869,11870,11873, 11874,11877,11878,11879,11880,11881,11882,11883,11884,11885,11886,11888,11889,11890,11891, 11892,11893,11896,11897,11898,11899,11900,11901,11902,11903,11904,11905,11906,11907,11911, 11912,11913,11916,11917,11918,11920,11923,11925,11927,11928,11929,11930,11932,11933,11934, 11935,11936,11937,11938,11939,11942,11943,11944,11947,11976,11977,11978,11979,11980,12049, 12051,12100,12100,12102,12103,12104,12105,12106,12107,12108,12109,12110,12111,12112,12115, 12117,12118,12119,12120,12121,12122,12123,12124,12125,12126,12128,12129,12130,12132,12133, 12136,12138,12139,12140,12141,12142,12144,12145,12146,12147,12152,12154,12155,12157,12158, 12159,12160,12162,12163,12164,12167,12170,12171,12172,12173,12174,12175,12176,12178,12179, 12180,12181,12182,12183,12200,12201,12202,12203,12204,12205,12206,12207,12208,12209,12210, 12211,12215,12216,12217,12218,12219,12220,12222,12223,12224,12225,12226,12227,12228,12229, 12230,12231,12236,12237,12239,12240,12241,12242,12243,12244,12248,12249,12250,12271,12273, 12274,12302,12303,12304,12305,12306,12309,12310,12311,12312,12314,12315,12316,12317,12318, 12321,12322,12323,12324,12325,12326,12327,12328,12329,12330,12331,12332,12333,12336,12337, 12338,12339,12341,12342,12343,12345,12346,12347,12348,12349,12350,12351,12352,12353,12354, 12355,12356,12357,12359,12360,12362,12363,12364,12366,12369,12370,12373,12376,12377,12379, 12380,12382,12383,12386,12388,12389,12391,12392,12394,12396,12397,12398,12399,12400,12403, 12405,12406,12407,12408,12409,12410,12411,12412,12413,12414,12415,12416,12417,12418,12419, 12420,12421,12422,12423,12424,12425,12428,12429,12430,12431,12432,12434,12435,12436,12438, 12439,12440,12441,12443,12444,12445,12446,12447,12471,12472,12473,12474,12475,12476,12477, 12492,12493,12494,12495,12496,12497,12498,12500,12501,12503,12504,12505,12508,12510,12513, 12514,12516,12517,12521,12522,12526,12527,12528,12530,12531,12538,12539,12541,12542,12543, 12544,12545,12546,12547,12548,12549,12550,12551,12552,12553,12554,12555,12556,12557,12558, 12559,12561,12562,12563,12565,12567,12568,12569,12570,12573,12574,12575,12577,12579,12580, 12581,12583,12584,12586,12587,12588,12589,12591,12592,12594,12595,12596,12597,12598,12601, 12602,12603,12604,12605,12606,12611,12612,12614,12617,12618,12619,12620,12621,12622,12623, 12625,12626,12627,12628,12629,12631,12632,12633,12634,12636,12638,12640,12641,12643,12645, 12646,12647,12648,12649,12650,12651,12652,12653,12654,12656,12657,12658,12662,12663,12664, 12665,12669,12670,12671,12672,12673,12674,12675,12676,12677,12678,12679,12680,12681,12700, 12701,12702,12703,12704,12705,12706,12707,12708,12709,12710,12711,12712,12713,12716,12719, 12720,12722,12723,12724,12725,12726,12727,12730,12731,12732,12733,12734,12735,12740,12741, 12742,12743,12744,12745,12747,12748,12752,12753,12756,12757,12761,12763,12765,12767,12768, 12769,12770,12771,12772,12773,12774,12776,12777,12778,12780,12783,12784,12786,12791,12792, 12797,12800,12801,12802,12803,12805,12806,12807,12808,12809,12814,12817,12818,12819,12820, 12821,12822,12823,12826,12827,12828,12829,12830,12831,12832,12833,12834,12835,12836,12837, 12838,12839,12840,12844,12845,12846,12847,12848,12849,12850,12851,12861,12865,12874,12875, 12877,12878,12879,12900,12901,12902,12904,12905,12907,12908,12909,12911,12913,12917,12919, 12920,12922,12923,12925,12926,12927,12928,12929,12930,12931,12932,12935,12937,12938,12939, 12940,12941,12942,12944,12947,12949,12953,12956,12957,12958,12960,12961,12962,12963,12964, 12965,12966,12967,12968,12969,12970,12971,12972,12973,12974,12976,12978,12979,12980,12983, 12985,12986,12987,12988,12992,12994,12996,12997,12999,13000,13002,13003,13004,13005,13007, 13008,13009,13010,13013,13014,13015,13016,13017,13018,13019,13020,13022,13023,13024,13025, 13026,13027,13028,13029,13031,13034,13035,13036,13039,13041,13042,13043,13044,13046,13047, 13060,13061,13062,13063,13064,13065,13066,13067,13068,13069,13070,13071,13072,13073,13074, 13075,13076,13077,13078,13079,13080,13081,13082,13083,13084,13086,13088,13089,13101,13102, 13103,13104,13105,13109,13110,13113,13114,13115,13116,13117,13118,13119,13120,13121,13122, 13123,13125,13128,13129,13131,13132,13133,13134,13135,13136,13137,13138,13139,13142,13148, 13149,13150,13151,13152,13153,13154,13155,13156,13157,13158,13159,13160,13163,13164,13165, 13166,13167,13168,13169,13170,13171,13173,13177,13178,13179,13180,13181,13182,13183,13184, 13185,13186,13188,13189,13190,13191,13192,13194,13195,13196,13197,13198,13199,13201,13202, 13203,13204,13206,13207,13208,13213,13214,13215,13217,13218,13219,13220,13221,13223,13224, 13225,13226,13227,13228,13229,13230,13231,13232,13233,13234,13238,13239,13240,13241,13242, 13244,13245,13246,13249,13263,13264,13265,13266,13272,13273,13274,13277,13278,13279,13298, 13299,13301,13302,13303,13304,13305,13306,13307,13308,13309,13310,13312,13313,13314,13315, 13316,13317,13318,13320,13321,13322,13323,13324,13325,13326,13327,13328,13329,13330,13331, 13332,13333,13334,13336,13337,13341,13342,13343,13344,13348,13351,13353,13355,13359,13362, 13363,13365,13366,13367,13368,13370,13371,13373,13374,13375,13376,13377,13378,13379,13381, 13382,13385,13386,13389,13391,13393,13397,13399,13400,13401,13402,13403,13404,13406,13407, 13408,13410,13411,13412,13415,13416,13417,13420,13421,13423,13425,13427,13428,13429,13430, 13431,13438,13439,13440,13443,13444,13445,13447,13448,13450,13451,13453,13454,13455,13456, 13458,13459,13463,13465,13466,13467,13471,13472,13473,13474,13475,13476,13477,13478,13500, 13501,13502,13508,13509,13510,13511,13512,13513,13514,13515,13516,13517,13518,13519,13520, 13521,13522,13523,13524,13525,13526,13527,13528,13529,13530,13531,13532,13533,13534,13535, 13536,13537,13538,13539,13540,13541,13542,13543,13544,13545,13546,13547,13548,13549,13550, 13552,13553,13556,13557,13558,13560,13561,13562,13563,13564,13566,13567,13568,13569,13570, 13571,13572,13574,13575,13578,13579,13580,13581,13582,13584,13585,13586,13587,13588,13589, 13590,13591,13592,13593,13594,13595,13596,13598,13599,13600,13602,13603,13604,13605,13606, 13607,13608,13609,13610,13611,13612,13613,13614,13616,13617,13619,13622,13624,13625,13626, 13627,13628,13629,13634,13637,13638,13639,13640,13641,13642,13643,13644,13645,13648,13649, 13651,13652,13653,13654,13655,13656,13657,13658,13659,13661,13664,13666,13667,13672,13675, 13679,13680,13681,13684,13686,13690,13693,13694,13696,13697,13698,13715,13716,13718,13727, 13739,13750,13751,13752,13753,13754,13755,13756,13757,13758,13759,13760,13761,13762,13763, 13764,13765,13766,13767,13768,13769,13770,13771,13772,13773,13774,13775,13776,13777,13779, 13784,13785,13786,13787,13788,13789,13790,13791,13792,13793,13794,13796,13797,13798,13799, 13803,13804,13805,13806,13808,13809,13810,13811,13814,13816,13817,13823,13824,13825,13846, 13847,13848,13849,13850,13851,13852,13853,13856,13857,13859,13862,13863,13864,13867,13868, 13869,13870,13871,13872,13875,13876,13877,13878,13880,13883,13884,13886,13887,13892,13893, 13894,13895,13896,13897,13899,13900,13901,13902,13903,13904,13905,13906,13907,13908,13909, 13910,13911,13912,13913,13914,13915,13916,13917,13918,13919,13920,13921,13923,13925,13926,13927,13928, 13930,13931,13950,13951,13952,13953,13954,13955,13957,13959,13962,13963,13966,13967,13968,13969,13970, 13971,13972,13974,13976,13977,13978,13980,13982,13983,13984,13986,13987,13988,13989,13990,13992,13993,13995,13997,13998,13998,13999,14000,14005,14006,14007,14008,14009,14010,14011,14012,14013,14014,14015,14016,14017,14018,14019,14020,14021,14022,14023,14024,14025,14026,14027,14028,14029,14030,14031,14032,14033,14034,14035,14036,14037,14038,14039,14040,14050,14054,14055,14058,14059,14061,14062,14063,14064,14065,14068,14069,14070,14071,14072,14073,14074,14075,14076,14077,14078,14079,14080,14081,14082,14083,14084,14085,14086,14087,14088,14089,14090,14091,14095,14096,14097,14098,14100,14101,14102,14103,14105,14106,14107,14108,14109,14110,14111,14113,14114,14115,14117,14118,14119,14120,14121,14122,14123,14125,14126,14129,14130,14131,14132,14133,14135,14136,14137,14138,14139,14140,14141,14142,14147,14148,14149,14152,14157,14158,14159,14162,14163,14164,14165,14167,14170,14174,14219,14221,14223,14224,14248,14250,14254,14255,14256,14257,14258,14259,14260,14261,14262,14263,14264,14265,14266,14267,14268,14269,14269,14270,14271,14274,14275,14276,14277,14278,14279,14282,14284,14285,14286,14287,14288,14289,14290,14291,14293,14295,14296,14297,14298,14299,14300,14301,14302,14303,14304,14305,14306,14307,14309,14311,14313,14314,14315,14316,14317,14318,14319,14320,14321,14322,14323,14324,14325,14327,14328,14329,14330,14331,14332,14333,14334,14335,14336,14337,14338,14339,14340,14341,14342,14343,14344,14345,14346,14347,14348,14349,14350,14351,14352,14353,14354,14355,14356,14357,14358,14359,14360,14361,14362,14363,14365,14366,14367,14368,14369,14370,14373,14374,14375,14376,14377,14378,14379,14380,14381,14382,14383,14384,14385,14386,14388,14389,14390,14391,14393,14394,14395,14396,14398,14399,14400,14402,14403,14450,14451,14452,14453,14454,14455,14456,14457,14458,14459,14460,14461,14462,14463,14464,14465,14466,14467,14468,14469,14470,14471,14472,14473,14474,14475,14476,14477,14478,14479,14481,14482,14483,14486,14489,14490,14491,14492,14493,14494,14495,14496,14497,14498,14499,14500,14504,14534,14535,14536,14537,14538,14539,14540,14541,14558,14559,14560,14561,14562,14563,14564,14566,14567,14568,14569,14570,14571,14572,14586,14587,14588,14595,14596,14597,14598,14599,14600,14603,14604,14605,14606,14607,14609,14610,14611,14613,14614,14617,14619,14620,14621,14622,14625,14627,14628,14629,14630,14631,14632,14633,14634,14635,14637,14638,14639,14640,14642,14644,14645,14647,14648,14649,14650,14651,14652,14653,14654,14655,14656,14657,14658,14659,14661,14662,14663,14664,14665,14667,14668,14669,14670,14671,14673,14674,14675,14676,14678,14679,14680,14683,14685,14686,14687,14688,14690,14691,14693,14696,14697,14698,14699,14701,14704,14706,14708,14709,14711,14712,14714,14715,14717,14719,14720,14721,14722,14723,14724,14725,14729,14731,14732,14733,14736,14737,14739,14740,14742,14743,14744,14745,14746,14747,14748,14749,14750,14753,14754,14756,14757,14759,14760,14761,14762,14763,14764,14765,14766,14767,14768,14770,14771,14773,14774,14775,14776,14777,14778,14779,14780,14782,14783,14785,14786,14787,14788,14789,14790,14791,14792,14793,14794,14795,14796,14797,14798,14799,14800,14801,14802,14803,14804,14806,14807,14808,14809,14810,14811,14812,14813,14815,14816,14819,14820,14821,14823,14826,14828,14829,14831,14832,14833,14834,14835,14837,14838,14839,14842,14843,14844,14845,14846,14847,14848,14849,14850,14851,14852,14853,14854,14855,14856,14857,14859,14860,14863,14864,14865,14867,14868,14869,14870,14871,14872,14873,14874,14875,14876,14878,14880,14881,14882,14883,14884,14885,14886,14888,14889,14890,14891,14892,14893,14894,14895,14896,14897,14898,14899,14900,14901,14902,14903,14904,14905,14906,14907,14908,14909,14910,14911,14912,14913,14914,14915,14916,14918,14919,14920,14922,14923,14924,14925,14927,14928,14931,14933,14934,14935,14937,14938,14941,14943,14944,14945,14946,14948,14949,14950,14951,14952,14953,14954,14955,14956,14957,14958,14959,14960,14961,14965,14966,14971,14972,14973,14974,14975,14976,14977,14979,14981,14982,14983,14984,14985,14986,14987,14988,14989,14990,14991,14992,14993,14994,14995,14996,14997,14999,15000,15001,15003,15004,15005,15007,15009,15010,15011,15013,15016,15018,15019,15023,15025,15026,15027,15028,15029,15030,15031,15032,15033,15035,15036,15040,15043,15044,15153,15154,15155,15156,15157,15158,15159,15160,15161,15162,15163,15164,15165,15166,15167,15168,15169,15170,15171,15172,15173,15174,15175,15176,15177,15178,15179,15180,15181,15182,15183,15185,15186,15188,15189,15196,15198,15199,15201,15203,15204,15207,15208,15209,15210,15211,15212,15213,15214,15215,15216,15222,15223,15224,15226,15228,15229,15230,15231,15254,15255,15256,15259,15260,15261,15263,15273,15274,15275,15276,15278,15279,15280,15283,15284,15288,15289,15290,15291,15292,15293,15294,15315,15316,15317,15320,15321,15322,15325,15326,15328,15330,15331,15332,15333,15334,15335, 15336,15338,15339,15340,15341,15343,15344,15347,15348,15349,15350,55525,55529,55533,10032,10175, 10192,10361,10372,10449, 10192, 10266,10356, 10372, 10449, 10690, 11016, 11167, 11229, 11320, 11611, 11618,11682,11683,11688,11698,11712,11763,11783,11792,11811,11815,11820,11825,11849,11853,11855,11860,11871,11872,11894,11895,11908,11909,11910,11914,11915,11921,11922,12062,12063,12064,12065,12113,12116,12134,12143,12148,12149,12150,12151,12161,12165,12166,12168,12214,12221,12238,12245,12246,12247,12267,12300,12301,12307,12308,12319,12320,12358,12365,12367,12374,12381,12385,12390,12427,12448,12449,12506,12507,12509,12515,12519,12525,12529,12560,12564,12566,12571,12572,12576,12585,12590,12593,12608,12609,12616,12637,12659,12660,12661,12666,12714,12739,12749,12750,12751,12754,12755,12758,12759,12760,12764,12787,12799,12804,12811,12812,12824,12841,12842,12876,12912,12914,12915,12921,12924,12933,12934,12936,12946,12951,12952,12954,12955,12959,12981,12989,12990,12991,12995,12998,13001,13006,13012,13021,13030,13032,13033,13038,13048,13049,13100,13106,13108,13111,13112,13124,13140,13141,13143,13145,13146,13147,13161,13162,13172,13187,13222,13235,13236,13237,13243,13247,13248,13276,13300,13311,13335,13352,13357,13358,13361,13372,13380,13383,13384,13388,13390,13398,13414,13419,13422,13424,13432,13433,13434,13446,13457,13468,13503,13504,13505,13506,13551,13559,13565,13583,13615,13620,13623,13630,13631,13632,13635,13646,13647,13650,13660,13662,13663,13665,13683,13699,13719,13720,13738,13740,13795,13807,13813,13815,13839,13854,13855,13860,13861,13865,13866,13873,13882,13885,13889,13890,13891,13898,13922,13924,13956,13960,13961,13973,13975,13979,13981,13985,13994,13996,14001,14002,14003,14051,14052,14053,14057,14060,14066,14067,140802,14092,14093,14094,14099,14104,14112,14116,14124,14143,14145,14155,14156,14160, 14161,14169,14171,14172,14173,14199,14200,14220,14222,14251,14252,14253,14273,14280,14281,14283, 14292,14294,14308,14310,14371,14372,14387,14397,14480,14624,14666,14677,14682,14694,14695,14702,14707,14710,14726,14727,14728,14730,14741,14769,14781,14784,14805,14814,14817,14818,14824,14827,14840,14926,14930,14932,14940,14967,14970,15002,15008,15015,15017,15020,15038,15039,15184,15187,15193,15194,15195,15197,15205,15206,15218,15219,15220,15221,15227,15281,15287,15295,15296,15297,15298,15299,15345,41968,41980,55553 & 72762.
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Rd
Cleveland, Ohio 44143-2131
For Additional Information Contact Kumudini J. Carter
440-483-3032
Manufacturer Reason
for Recall
Incorrect values may be displayed. Philips is issuing a software update (version 4.5.5) to the Extended Brilliance Workspace (EBW) systems currently running software version(s) 4.0.2, V4.5.2 or V4.5.3, addressing a number of non-conformances.
FDA Determined
Cause 2
DESIGN: Software Design (Manufacturing Process)
Action Philips Healthcare sent a Urgent Medical Device Correction letter dated September 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. This Field Safety notification is to inform the actions that should be taken by the customer/user to prevent risks for patient or users, and the actions planned by Philips to correct the problem. If you need further information or supplies please contact your local Philips Healthcare office. For North American and Canada contact the Customer Care Solutions Center ( 1-800 722-9377) option 5 enter site or follow the prompts.
Quantity in Commerce 2,580 units
Distribution Worldwide Distribution - USA including the states of AL, AT, AZ, BAH, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NJ, NV, NY, OH, OR, PA, PR, PR, SC, TN, TX, VA, VT, WA, WI & WV. and the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican, Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Greece, Guadeloupe, Guinea, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, LAO People's Dem, Rep., Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Malawi, Malaysia, Malta, Martinique, Mexico, Netherlands, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela & Viet Nam.
Total Product Life Cycle TPLC Device Report

 
1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = JAK and Original Applicant = ELSCINT, INC.
-
-