|
Class 2 Device Recall FLEXSTENT Femoropopliteal SelfExpanding Stent System (EU Commercial 80 cm Length) |
|
Date Initiated by Firm |
April 23, 2013 |
Date Posted |
June 06, 2013 |
Recall Status1 |
Terminated 3 on June 27, 2014 |
Recall Number |
Z-1510-2013 |
Recall Event ID |
65182 |
Product Classification |
Stent, superficial femoral artery - Product Code NIP
|
Product |
FLEXSTENT Femoropopliteal Self-Expanding Stent System.
Product Usage: The FSS FLEXSTENT¿ Femoropopliteal SES clinical product is indicated for treatment of atherosclerotic de novo native superficial femoral artery lesions and proximal popliteal lesions.
|
Code Information |
32 Catalog Numbers FLX-50030-V1, FLX-60030-V1, FLX-70030-V1, FLX-80030-V1, FLX-50040-V1, FLX-60040-V1, FLX-70040-V1, FLX-80040-V1, FLX-50060-V1, FLX-60060-V1, FLX-70060-V1, FLX-80060-V1, FLX-50080-V1, FLX-60080-V1, FLX-70080-V1, FLX-80080-V1, FLX-50100-V1, FLX-60100-V1, FLX-70100-V1, FLX-80100-V1, FLX-50120-V1, FLX-60120-V1, FLX-70120-V1, FLX-80120-V1, FLX-50150-V1, FLX-60150-V1, FLX-70150-V1, FLX-80150-V1, FLX-50200-V1, FLX-60200-V1, FLX-70200-V1, FLX-80200-V1 |
Recalling Firm/ Manufacturer |
Cordis Corporation 14201 Nw 60th Ave Miami Lakes FL 33014-2802
|
For Additional Information Contact |
Ivenette Guzman 786-313-2000
|
Manufacturer Reason for Recall |
Holes and/or tears can be introduced in the FLEXSTENT¿ inner pouch under some conditions, and to a lesser degree, in the outer pouch.
|
FDA Determined Cause 2 |
Package design/selection |
Action |
Cordis sent an Urgent Medical Device Correction and Potential Removal letter dated April 24, 2013 e-mail to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately identify and set aside all affected products to prevent further use of the affected product, until inspected. ¿ Use only product identified with a blue dot next to the lot number on the carton face label, carton end label, and outer pouch label, or product with upgraded sterile barrier system (not yet available, as of the date of this letter), review, complete, sign and return the enclosed Acknowledgement Form and maintain a copy of this notice with the affected product. For questions related to the issue, please contact our FSS/Cordis
representative, or please contact Cordis Direct at (800) 781-0282, Monday
through Friday from 9:00 AM to 5:00 PM EDT. |
Quantity in Commerce |
250 units |
Distribution |
Worldwide Distribution - USA (nationwide) and the states of: AZ, CA, CT, FL, GA, IA, IL, LA, MD, MI, MS, NC, NJ, NY, OH, PA, RI, SC, TX, WA, WI, WV, and the countries of: Czech Republic, Italy, Netherlands, and New Zealand |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|