|
Class 2 Device Recall Solar 8000 and Transport Pro with Patient Data Module (Marketed as CARESCAPE Patient Data Module) |
|
Date Initiated by Firm |
April 03, 2013 |
Date Posted |
August 28, 2013 |
Recall Status1 |
Terminated 3 on July 25, 2016 |
Recall Number |
Z-2116-2013 |
Recall Event ID |
66016 |
510(K)Number |
K071073
|
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product |
SOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE (Marketed As GE Healthcare CARESCAPE Patient Data Module)
Product Usage: The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes. |
Code Information |
Mfg Lot or Serial # SA310290653GA SA310290655GA SA310290657GA SA310290659GA SA310290661GA SA310290748GA SA310290749GA SA310290750GA SA310290754GA SA310290755GA SA310290756GA SA310290757GA SA310290758GA SA310290759GA SA310361647GA SA310361877GA SA310372108GA SA310393017GA SA310393052GA SA310382648GA SA310382626GA SA310382631GA SA310382729GA SA310382761GA SA310393037GA SA310393054GA SA310393071GA SA310393073GA SA310351531GA SA310361792GA SA310361798GA SA310361801GA SA310361803GA SA310361804GA SA310361805GA SA310361807GA SA310361808GA SA310361815GA SA310361820GA SA310361838GA SA310361840GA SA310361841GA SA310393150GA SA310393155GA SA310393162GA SA310393165GA SA310393166GA SA310393167GA SA310393170GA SA310393171GA SA310393173GA SA310393175GA SA310321055GA SA310321062GA SA310361949GA SA310382418GA SA310382450GA SA310382499GA SA310382513GA SA310382608GA SA310382647GA SA310382649GA SA310382663GA SA310393156GA SA310393158GA SA310393164GA SA310393169GA SA310393177GA SA310259944GA SA310331337GA SA310331342GA SA310331343GA SA310331345GA SA310331349GA SA310331351GA SA310341373GA SA310341376GA SA310341387GA SA310341388GA SA310341399GA SA310341400GA SA310341401GA SA310341402GA SA310341403GA SA310341406GA SA310382386GA SA310382391GA SA310382472GA SA310382476GA SA310382477GA SA310382479GA SA310382484GA SA310382494GA SA310382618GA SA310382619GA SA310382634GA SA310382636GA SA310382645GA SA310382378GA SA310382449GA SA310382511GA SA310382606GA SA310382607GA SA310382610GA SA310382612GA SA310382613GA SA310382615GA SA310382616GA SA310382642GA SA310382652GA SA310382656GA SA310382664GA SA310423572GA SA310423573GA SA310290745GA SA310290753GA SA310413471GA SA310413476GA SA310423587GA SA310423607GA SA310413493GA SA310413498GA SA310413500GA SA310413507GA SA310413514GA SA310413524GA SA310413535GA SA310423604GA SA310423606GA SA310423611GA SA310423617GA SA310423619GA SA310423621GA SA310423626GA SA310423640GA SA310423649GA SA310423650GA SA310423730GA SA310423821GA SA310423825GA SA310423836GA SA310423838GA SA310423842GA SA310423864GA SA310423929GA SA310423935GA SA310423943GA SA310423959GA SA310423962GA SA310424000GA SA310424001GA SA310424017GA SA310413491GA SA310413499GA SA310413509GA SA310413511GA SA310413538GA SA310423613GA SA310423614GA SA310423618GA SA310423622GA SA310423628GA SA310423637GA SA310423639GA SA310423663GA SA310423709GA SA310423933GA SA310423937GA SA310423946GA SA310423950GA SA310423953GA SA310423956GA SA310423957GA SA310423960GA SA310423961GA SA310423967GA SA310423971GA SA310424002GA SA310424003GA SA310424004GA SA310424006GA SA310424008GA SA310424025GA SA310424027GAMKE SA310424033GA SA310424034GAMKE SA310424039GA SA310424046GA SA310434199GAMKE SA310434200GA SA310434204GAMKE SA310434205GA SA310434207GAMKE SA310434208GA SA310434210GA SA310434211GAMKE SA310434212GA SA310434213GA SA310434227GAMKE SA310434238GA SA310444247GAMKE SA310444253GAMKE SA310372179GA SA310382540GA SA310382541GA SA310382542GA SA310382544GA SA310382550GA SA310382553GA SA310382554GA SA310382557GA SA310382558GA SA310382562GA SA310382563GA SA310382575GA SA310382580GA SA310382593GA SA310382599GA SA310382602GA SA310382604GA SA310382693GA SA310403363GA SA310403364GA SA310403372GA SA310403433GA SA310423758GA SA310423793GA SA310423896GA SA310423914GA SA310423917GA SA310423919GA SA310423979GA SA310423994GA SA310434070GA SA310423777GA SA310423779GA SA310423787GA SA310423819GA SA310423871GA SA310444296GA SA310310864GA SA310382832GA SA310392903GA SA310392904GA SA310392943GA SA310392946GA SA310392997GA SA310392999GA SA310393006GA SA310393007GA SA310393008GA SA310393010GA SA310393011GA SA310393012GA SA310393015GA SA310393023GA SA310393026GA SA310393029GA SA310393035GA SA310393036GA SA310393048GA SA310393059GA SA310393061GA SA310393064GA SA310393065GA SA310393074GA SA310393075GA SA310393188GA SA310393190GA SA310393197GA SA310393200GA SA310393213GA SA310393219GA SA310393229GA SA310393249GA SA310393250GA SA310393252GA SA310393254GA SA310393260GA SA310393261GA SA310393268GA SA310393269GA SA310393274GA SA310393281GA SA310393283GA SA310393294GA SA310393304GA SA310423888GA SA310423989GA SA310434119GA SA310434120GA SA310434132GA SA310434136GA SA310434140GA SA310434151GA SA310434160GA SA310434170GA SA310434185GA SA310434186GA SA310434188GA SA310341368GA SA310341391GA SA310392912GA SA310393212GA SA310393216GA SA310393149GA SA310393151GA SA310393152GA SA310393154GA SA310393168GA SA310393172GA SA310393174GA SA310393178GA SA310403341GA SA310403348GA SA310403379GA SA310403384GA SA310403390GA SA310403398GA SA310403405GA SA310403414GA SA310403415GA SA310403418GA SA310403436GA SA310403442GA SA310423780GA SA310423785GA SA310423800GA SA310423811GA SA310423916GA SA310423974GA SA310434063GA SA310434064GA SA310434066GA SA310434071GA SA310434077GA SA310434084GA SA310434090GA SA310434091GA SA310434094GA SA310434097GA SA310434106GA SA310434107GA SA310434108GA SA310434110GA SA310434114GA SA310434118GA SA310434125GA SA310434134GA SA310434148GA SA310434173GA SA310434194GA SA310434195GA SA310444266GA SA310434089GA SA310260359GA SA310331162GA SA310351499GA SA310361741GA SA310361742GA SA310361743GA SA310361745GA SA310361746GA SA310361747GA SA310361748GA SA310361749GA SA310361750GA SA310361751GA SA310361753GA SA310361758GA SA310361763GA SA310361766GA SA310361770GA SA310361776GA SA310361810GA SA310444496GA SA310444504GA SA310455031GA SA310393080GA SA310393270GA SA310393280GA SA310393295GA SA310393309GA SA310393310GA SA310423766GA SA310341398GA SA310341404GA SA310455048GA SA310455056GA SA310455170GA SA310444369GA SA310444503GA SA310444513GA SA310444531GA SA310424044GA SA310424031GA SA310454982GA SA310423862GA SA310424010GA SA310454976GA SA310454981GA SA310454986GA SA310454996GA SA310455000GA SA310455004GA SA310455016GA SA310455018GA SA310455030GA SA310455058GA SA310455061GA SA310455062GA SA310455156GA SA310455161GA SA310455164GA SA310465352GA SA310465366GA SA310465394GA SA310423651GA SA310423694GA SA310423600GA SA310455006GA SA310455026GA SA310455057GA SA310382387GA SA310382413GA SA310382415GA SA310310974GA SA310310978GA SA310321007GA SA310321019GA SA310321041GA SA310321066GA SA310403396GA SA310403420GA SA310423683GA SA310423738GA SA310423921GA SA310423986GA SA310423988GA SA310455256GA SA310455275GA SA310465314GA SA310475525GA SA310475537GA SA310475611GA SA310485808GA SA310444499GA SA310444509GA SA310444839GA SA310444840GA SA310444842GA SA310444856GA SA310444859GA SA310444893GA SA310454928GA SA310454939GA SA310455259GA SA310403435GA SA310485856GA SA310444259GA SA310392895GA SA310393106GA SA310485867GA SA310475566GA SA310475585GA SA310475610GA SA310475628GA SA310475629GA SA310475630GA SA310475634GA SA310475637GA SA310475684GA SA310475685GA SA310475686GA SA310475696GA SA310475697GA SA310475701GA SA310475707GA SA310475728GA SA310475734GA SA310485786GA SA310485794GA SA310485800GA SA310485801GA SA310485804GA SA310485807GA SA310485814GA SA310485822GA SA310485833GA SA310423799GA SA310495953GA SA310495955GA SA310475682GA SA310485788GA SA310485792GA SA310485795GA SA310485811GA SA310392897GA SA310392911GA SA310393086GA SA310393094GA SA310393108GA SA310393109GA SA310393114GA SA310393131GA SA310393222GA SA310393233GA SA310393262GA SA310444364GA SA310455055GA SA310455091GA SA310465367GA SA310290679GA SA310321046GA SA310361952GA SA310392949GA SA310393138GA SA310393139GA SA310393140GA SA310393142GA SA310393143GA SA310393145GA SA310423680GA SA310423684GA SA310423882GA SA310423886GA SA310423906GA SA310434081GA SA310434082GA SA310434115GA SA310485851GA SA310485881GA SA310393237GA SA310393255GA SA310393296GA SA310403319GA SA310403320GA SA310403321GA SA310403322GA SA310403325GA SA310403332GA SA310413451GA SA310423666GA SA310423667GA SA310423670GA SA310423671GA SA310423672GA SA310423674GA SA310423675GA SA310423677GA SA310423681GA SA310423688GA SA310423696GA SA310423698GA SA310423701GA SA310423704GA SA310423708GA SA310423712GA SA310423713GA SA310423715GA SA310423720GA SA310423721GA SA310423741GA SA310423745GA SA310423753GA SA310423769GA SA310423822GA SA310423839GA SA310423841GA SA310423940GA SA310423949GA SA310423951GA SA310444851GA SA310444863GA SA310454938GA SA310454989GA SA310455015GA SA310455154GA SA310475481GA SA310475485GA SA310475489GA SA310485882GA SA310495919GA SA310495970GA SA310331081GA SA310393083GA SA310393130GA SA310403311GA SA310423748GA SA310423759GA SA310423866GA SA310423883GA SA310423887GA SA310423904GA SA310423912GA SA310423985GA SA310434048GA SA310434059GA SA310434069GA SA310434098GA SA310434139GA SA310434180GA SA310434183GA SA310454915GA SA310454946GA SA310454984GA SA310455009GA SA310455034GA SA310455065GA SA310455066GA SA310455069GA SA310455118GA SA310455120GA SA310455122GA SA310455125GA SA310455127GA SA310455131GA SA310455142GA SA310455227GA SA310455228GA SA310455232GA SA310455233GA SA310455248GA SA310455251GA SA310455257GA SA310455264GA SA310455266GA SA310455268GA SA310455272GA SA310455278GA SA310455282GA SA310455286GA SA310455289GA SA310465298GA SA310465310GA SA310465312GA SA310465317GA SA310372302GA SA310423697GA SA310506841GA SA310507174GA SA310507176GA SA310507183GA SA310507201GA SA310507205GA SA310507237GA SA310507292GA SA310507298GA SA310507301GA SA310507304GA SA310507309GA SA310507310GA SA310507319GA SA310507326GA SA310507328GA SA310507329GA SA310507332GA SA310507336GA SA310507337GA SA310507356GA SA310507366GA SA310507367GA SA310507371GA SA310507372GA SA310507377GA SA310507379GA SA310507381GA SA310507382GA SA310507383GA SA310507384GA SA310507385GA SA310507386GA SA310507387GA SA310507499GA SA310517583GA SA310361707GA SA310372294GA SA310403328GA SA310403426GA SA310413544GA SA310423668GA SA310423736GA SA310423755GA SA310423778GA SA310423818GA SA310423873GA SA310423874GA SA310423893GA SA310423895GA SA310423907GA SA310423910GA SA310423958GA SA310423978GA SA310423980GA SA310423987GA SA310454918GA SA310454925GA SA310454927GA SA310454944GA SA310454947GA SA310454948GA SA310454949GA SA310454950GA SA310454952GA SA310454953GA SA310454990GA SA310454991GA SA310454992GA SA310454995GA SA310455028GA SA310455047GA SA310455146GA SA310455147GA SA310455149GA SA310455150GA SA310455152GA SA310455157GA SA310475464GA SA310475536GA SA310475575GA SA310475579GA SA310475581GA SA310475582GA SA310475592GA SA310475598GA SA310475600GA SA310475608GA SA310475649GA SA310475663GA SA310475709GA SA310475719GA SA310475730GA SA310485785GA SA310485819GA SA310485824GA SA310485828GA SA310485843GA SA310485848GA SA310485853GA SA310485859GA SA310506749GA SA310507219GA SA310517588GA SA310517706GA SA310517789GA SA310517790GA SA310518161GA SA310518162GA SA310518205GA SA310351558GA SA310423652GA SA310423657GA SA310423903GA SA310423973GA SA310434074GA SA310434076GA SA310434078GA SA310434092GA SA310434124GA SA310434157GA SA310434182GA SA310455084GA SA310455123GA SA310455130GA SA310465350GA SA310465359GA SA310465383GA SA310465396GA SA310475501GA SA310496650GA SA310506747GA SA310517641GA SA310518223GA SA310403409GA SA310495945GA SA310496375GA SA310496376GA SA310496383GA SA310496394GA SA310496415GA SA310496418GA SA310496450GA SA310496624GA SA310506711GA SA310506725GA SA310506784GA SA310506785GA SA310506787GA SA310506788GA SA310506801GA SA310506808GA SA310506810GA SA310506813GA SA310506814GA SA310506821GA SA310506822GA SA310506838GA SA310506860GA SA310506869GA SA310506870GA SA310506882GA SA310506895GA SA310506989GA SA310507036GA SA310507038GA SA310507067GA SA310507082GA SA310507149GA SA310507156GA SA310507157GA SA310507163GA SA310507196GA SA310507197GA SA310507199GA SA310507200GA SA310507202GA SA310507203GA SA310507212GA SA310507216GA SA310507230GA SA310507262GA SA310507307GA SA310507357GA SA310507380GA SA310517708GA SA310517721GA SA310517987GA SA310518014GA SA310518152GA SA310518160GA SA310518164GA SA310518217GA SA310518225GA SA310518232GA SA310518237GA SA310403383GA SA310413556GA SA310423593GA SA310423845GA SA310423942GA SA310423944GA SA310454929GA SA310454935GA SA310454936GA SA310454937GA SA310454940GA SA310454975GA SA310454987GA SA310454988GA SA310454994GA SA310455012GA SA310455024GA SA310455025GA SA310455046GA SA310455052GA SA310455053GA SA310455148GA SA310455151GA SA310455153GA SA310455155GA SA310455159GA SA310455160GA SA310455163GA SA310455173GA SA310455174GA SA310496156GA SA310496470GA SA310496483GA SA310496513GA SA310496517GA SA310496630GA SA310496638GA SA310506665GA SA310506692GA SA310506732GA SA310506859GA SA310506874GA SA310506881GA SA310507062GA SA310507092GA SA310507093GA SA310507095GA SA310507107GA SA310507114GA SA310507116GA SA310507122GA SA310507406GA SA310507413GA SA310507417GA SA310507456GA SA310507481GA SA310507482GA SA310507483GA SA310507485GA SA310507486GA SA310507490GA SA310507492GA SA310507493GA SA310507494GA SA310507497GA SA310507504GA SA310507507GA SA310507509GA SA310507511GA SA310507512GA SA310507516GA SA310517585GA SA310517591GA SA310423690GA SA310506843GA SA310507215GA SA310507242GA SA310507317GA SA310517645GA SA310517646GA SA310517651GA SA310517652GA SA310517664GA SA310517678GA SA310517680GA SA310517683GA SA310518168GA SA310518169GA SA310518170GA SA310518176GA SA310518319GA SA310392914GA SA310393088GA SA310393192GA SA310393196GA SA310393230GA SA310393235GA SA310393291GA SA310393303GA SA310475710GA SA310495890GA SA310496368GA SA310496378GA SA310496384GA SA310496386GA SA310496387GA SA310496388GA SA310496393GA SA310496401GA SA310496428GA SA310496469GA SA310496529GA SA310496618GA SA310496626GA SA310496631GA SA310506658GA SA310506682GA SA310506706GA SA310506812GA SA310506836GA SA310506853GA SA310506877GA SA310507000GA SA310507012GA SA310507016GA SA310507024GA SA310507025GA SA310507027GA SA310507050GA SA310507069GA SA310507075GA SA310507127GA SA310507154GA SA310507162GA SA310507171GA SA310507267GA SA310507270GA SA310507272GA SA310507276GA SA310507288GA SA310507296GA SA310507338GA SA310507344GA SA310507346GA SA310507375GA SA310507399GA SA310507464GA SA310507519GA SA310517560GA SA310517866GA SA310260482GA SA310361684GA SA310361708GA SA310393077GA SA310393284GA SA310434053GA SA310434085GA SA310434138GA SA310434181GA SA310455230GA SA310506654GA SA310506672GA SA310506703GA SA310506708GA SA310506717GA SA310506726GA SA310506729GA SA310506990GA SA310506999GA SA310507005GA SA310507089GA SA310507090GA SA310507104GA SA310507408GA SA310507430GA SA310507443GA SA310507445GA SA310507461GA SA310507463GA SA310507473GA SA310507476GA SA310507477GA SA310507478GA SA310507488GA SA310507508GA SA310507518GA SA310507523GA SA310507547GA SA310507550GA SA310507552GA SA310507555GA SA310507556GA SA310517558GA SA310517559GA SA310517563GA SA310517566GA SA310517570GA SA310517574GA SA310517596GA SA310517745GA SA310517817GA SA310517869GA SA310517875GA SA310518242GA SA310518275GA SA310518335GA SA310518337GA SA310518350GA SA310518361GA SA310518414GA SA310393234GA SA310403350GA SA310403413GA SA310403423GA SA310403427GA SA310403440GA SA310423742GA SA310423744GA SA310423756GA SA310423761GA SA310423770GA SA310423771GA SA310423773GA SA310423794GA SA310423806GA SA310423807GA SA310423809GA SA310423812GA SA310423814GA SA310423816GA SA310423872GA SA310423879GA SA310423900GA SA310424037GA SA310455126GA SA310455254GA SA310465293GA SA310465294GA SA310465305GA SA310465307GA SA310465313GA SA310465315GA SA310465318GA SA310475510GA SA310475557GA SA310475609GA SA310475626GA SA310475631GA SA310475639GA SA310475695GA SA310475732GA SA310475733GA SA310485768GA SA310485782GA SA310485789GA SA310485790GA SA310485793GA SA310485796GA SA310485806GA SA310485810GA SA310485812GA SA310485816GA SA310485818GA SA310485834GA SA310485836GA SA310485847GA SA310485863GA SA310506674GA SA310506683GA SA310506713GA SA310506736GA SA310506741GA SA310507080GA SA310507083GA SA310507084GA SA310507110GA SA310507131GA SA310507141GA SA310507146GA SA310507161GA SA310507238GA SA310507303GA SA310507308GA SA310507320GA SA310507330GA SA310507396GA SA310507400GA SA310507416GA SA310507421GA SA310507426GA SA310507441GA SA310507447GA SA310507455GA SA310507491GA SA310507505GA SA310507506GA SA310507517GA SA310517587GA SA310517602GA SA310517605GA SA310517606GA SA310517607GA SA310517608GA SA310517612GA SA310517613GA SA310517614GA SA310517616GA SA310517619GA SA310517633GA SA310517635GA SA310517636GA SA310517638GA SA310517650GA SA310517657GA SA310517660GA SA310517661GA SA310517670GA SA310517697GA SA310517704GA SA310517711GA SA310517712GA SA310517717GA SA310517719GA SA310517728GA SA310517729GA SA310517731GA SA310517738GA SA310517742GA SA310517744GA SA310517751GA SA310517754GA SA310517758GA SA310517761GA SA310517768GA SA310517769GA SA310517770GA SA310517774GA SA310517776GA SA310517777GA SA310517778GA SA310517780GA SA310517783GA SA310517785GA SA310517786GA SA310517792GA SA310518147GA SA310518149GA SA310518156GA SA310518157GA SA310518163GA SA310518179GA SA310518185GA SA310518187GA SA310518190GA SA310518204GA SA310518209GA SA310518210GA SA310518214GA SA310518230GA SA310518241GA SA310518252GA SA310518259GA SA310518263GA SA310518265GA SA310518266GA SA310518269GA SA310518270GA SA310518280GA SA310518282GA SA310518286GA SA310518289GA SA310518304GA SA310518305GA SA310423705GA SA310454924GA SA310454926GA SA310454934GA SA310454943GA SA310454945GA SA310423685GA SA310424019GA SA310444277GA SA310444304GA SA310454983GA SA310496331GA SA310496615GA SA310507043GA SA310507395GA SA310507553GA SA310507554GA SA310517696GA SA310518261GA SA310518291GA SA310518299GA SA310518332GA SA310518352GA SA310518356GA SA310518364GA SA310518365GA SA310518384GA SA310518394GA SA310518466GA SA310518706GA SA310518715GA SA310518719GA SA310518720GA SA310518721GA SA310310918GA SA310496514GA SA310496653GA SA310506746GA SA310506831GA SA310506832GA SA310507022GA SA310507045GA SA310507228GA SA310507410GA SA310507415GA SA310507419GA SA310507420GA SA310507427GA SA310507431GA SA310507433GA SA310507439GA SA310507442GA SA310507454GA SA310507521GA SA310507525GA SA310507526GA SA310507543GA SA310507546GA SA310517564GA SA310517565GA SA310517595GA SA310517597GA SA310517656GA SA310517760GAMKE SA310517762GA SA310517791GA SA310518154GA SA310518194GA SA310518228GA SA310518235GA SA310518256GA SA310518271GA SA310518274GA SA310518292GA SA310518294GA SA310518303GA SA310518307GA SA310518312GA SA310518320GA SA310518343GA SA310518345GA SA310518366GA SA310518415GA SA310518437GA SA310518444GA SA310518457GA SA310518458GA SA310518537GA SA310518554GA SA310518617GA SA310518618GA SA310518633GA SA310518639GA SA310518640GA SA310518649GA SA310518651GAMKE SA310518659GA SA310518672GA SA310518690GA SA310518717GA SA310518718GA SA310529071GA SA310529081GA SA310529096GA SA310393186GA SA310393267GA SA310423673GA SA310423678GA SA310423682GA SA310423947GA SA310496643GA SA310506868GA SA310507128GA SA310507145GA SA310507392GA SA310507402GA SA310507422GA SA310507428GA SA310507460GA SA310507522GA SA310517610GA SA310517634GA SA310517723GA SA310517724GA SA310517736GA SA310517746GA SA310517747GA SA310517752GA SA310517756GA SA310517763GA SA310517764GA SA310517765GA SA310517766GA SA310517767GA SA310517771GA SA310517772GA SA310517773GA SA310517782GA SA310517823GA SA310517830GA SA310517877GA SA310517982GA SA310517983GA SA310517985GA SA310517990GA SA310517991GA SA310517993GA SA310517994GA SA310517995GA SA310517996GA SA310518001GA SA310518002GA SA310518007GA SA310518010GA SA310518159GA SA310518247GA SA310518251GA SA310518255GA SA310518257GA SA310518258GA SA310518268GA SA310518296GA SA310518297GA SA310518324GA SA310518325GA SA310518326GA SA310518330GA SA310518333GA SA310518336GA SA310518341GA SA310518342GA SA310518349GA SA310518351GA SA310518353GA SA310518358GA SA310518359GA SA310518360GA SA310518367GA SA310518368GA SA310518370GA SA310518371GA SA310518376GA SA310518385GA SA310518386GA SA310518393GA SA310518397GA SA310518398GA SA310518401GA SA310518402GA SA310518403GA SA310518408GA SA310518409GA SA310518410GA SA310518416GA SA310518417GA SA310518439GA SA310518446GA SA310518447GA SA310518449GA SA310518459GA SA310518461GA SA310518472GA SA310518475GA SA310518485GA SA310518615GA SA310518621GA SA310518624GA SA310518625GA SA310518626GA SA310518627GA SA310518631GA SA310518632GA SA310518634GA SA310518636GA SA310518637GA SA310518638GA SA310518642GA SA310518643GA SA310518645GA SA310518648GA SA310518652GA SA310518657GA SA310518658GA SA310518662GA SA310518670GA SA310518674GA SA310518687GA SA310518688GA SA310518691GA SA310518692GA SA310518693GA SA310518695GA SA310518696GA SA310518699GA SA310518700GA SA310518702GA SA310518703GA SA310518704GA SA310518710GA SA310518712GA SA310518713GA SA310518716GA SA310518723GA SA310528895GA SA310528899GA SA310528962GA SA310528963GA SA310528979GA SA310528985GA SA310528992GA SA310528998GA SA310529007GA SA310529010GA SA310529011GA SA310517753GA SA310517755GA SA310517757GA SA310518630GA SA310518635GA SA310518654GA SA310518655GA SA310518660GA SA310518661GA SA310518665GA SA310518666GA SA310518697GA SA310518698GA SA310528961GA SA310528981GA SA310529088GA SA310529103GA SA310529106GA SA310529112GA SA310518433GA SA310518487GA SA310528943GA SA310290668GA SA310361887GA SA310382754GA SA310424028GA SA310424038GA SA310455017GA SA310506820GA SA310518465GA SA310518694GA SA310528901GA SA310528906GA SA310528936GA SA310528940GA SA310528944GA SA310529163GA SA310529222GA SA310518391GA SA310528849GA SA310528932GA SA310528947GA SA310529060GA SA310529087GA SA310529095GA SA310529097GA SA310529114GA SA310529122GA SA310434088GA SA310496324GA SA310496340GA SA310496349GA SA310506655GA SA310518009GA SA310518276GA SA310518369GA SA310518424GA SA310518434GA SA310518436GA SA310518451GA SA310518473GA SA310518503GA SA310518528GA SA310518543GA SA310518548GA SA310518561GA SA310518673GA SA310518686GA SA310528865GA SA310528870GA SA310528878GA SA310528911GA SA310528949GA SA310528951GA SA310528953GA SA310528957GA SA310529001GA SA310529003GA SA310529031GA SA310529032GA SA310529033GA SA310529042GA SA310529043GA SA310529157GA SA310529161GA SA310529164GA SA310529199GA SA310529200GA SA310529202GA SA310529203GA SA310529206GA SA310529211GA SA310529214GA SA310529219GA SA311029364GA SA311029365GA SA310507480GA SA310528749GA SA310529156GA SA310382839GA SA310518390GA SA310518423GA SA310518477GA SA310518676GA SA310528937GA SA310528939GA SA310529034GA SA310529041GA SA310529160GA SA310529218GA SA310518452GA SA310518453GA SA310518463GA SA310518479GA SA310518492GA SA310518493GA SA310518501GA SA310518521GA SA310528775GA SA310528778GA SA310528792GA SA310528813GA SA310528842GA SA310528891GA SA310528914GA SA310528915GA SA310528918GA SA310528925GA SA310528952GA SA310528959GA SA310528960GA SA310260320GA SA310260383GA SA310260402GA SA310341381GA SA310361785GA SA310382838GA SA310392886GA SA310393202GA SA310434122GA SA310455119GA SA310455121GA SA310507218GA SA310518198GA SA310518203GA SA310518208GA SA310518262GA SA310518396GA SA310518399GA SA310518483GA SA310518495GA SA310518509GA SA310518533GA SA310528733GA SA310528795GA SA310528928GA SA310528955GA SA310529061GA SA310529069GA SA310529102GA SA310539239GA SA310539311GA SA310361695GA SA310393144GA SA310496211GA SA310496212GA SA310496622GA SA310507014GA SA310507302GA SA310507311GA SA310518348GA SA310518373GA SA310518388GA SA310518407GA SA310518411GA SA310518412GA SA310518450GA SA310518476GA SA310528801GA SA310528802GA SA310528803GA SA310528912GA SA310539306GA SA311039518GA SA311039529GA SA311039551GA SA310393246GA SA310403349GA SA310413495GA SA310518430GA SA310518460GA SA310518462GA SA310518471GA SA310518474GA SA310518480GA SA310518482GA SA310518484GA SA310518491GA SA310528736GA SA310528744GA SA310528773GA SA310528787GA SA310528922GA SA310528930GA SA311039382GA SA310413458GA SA310475591GA SA310518678GA SA310518679GA SA310518685GA SA310518724GA SA310528856GA SA310528858GA SA310528861GA SA310528863GA SA310528864GA SA310528869GA SA310528967GA SA310528994GA SA310529006GA SA310529013GA SA310529028GA SA310529030GA SA310529035GA SA310529038GA SA310529040GA SA310529044GA SA310529047GA SA310529150GA SA310529159GA SA310529197GA SA310529208GA SA310529216GA SA310529220GA SA310413545GA SA310496329GA SA310496614GA SA310518675GA SA310518711GA SA310518714GA SA310528862GA SA310528868GA SA310528879GA SA311039387GA SA311039389GA SA311039393GA SA311039409GA SA311039416GA SA311039432GA SA311039449GA SA311039453GA SA311039458GA SA311039459GA SA311039463GA SA311039472GA SA311039473GA SA311039475GA SA310351554GA SA310434099GA SA310528854GA SA310528898GA SA310528999GA SA310529027GA SA311039370GA SA311039372GA SA311039442GA SA310507064GA SA310507074GA SA310507078GA SA310507094GA SA310507096GA SA310507113GA SA310507125GA SA310507135GA SA310507143GA SA310507389GA SA310507398GA SA310507401GA SA310507404GA SA310507407GA SA310507414GA SA310507458GA SA310507475GA SA310517732GA SA310528859GA SA310528897GA SA310529029GA SA310529046GA SA310529165GA SA310529201GA SA310423583GA SA310444494GA SA310444864GA SA310496214GA SA310506656GA SA310506707GA SA310507021GA SA310528972GA SA311039391GA SA311039392GA SA311039488GA SA310518663GA SA310528968GA SA310260553GA SA310310926GA SA310382635GA SA310507040GA SA310507072GA SA310507073GA SA310507108GA SA310507129GA SA310507132GA SA310507397GA SA310507436GA SA310507484GA SA310517662GA SA310517828GA SA310517839GA SA310517986GA SA310517998GA SA310518003GA SA310528969GA SA311059870GA SA311059909GA SA310259884GA SA310310883GA SA310331249GA SA310382657GA SA310413473GA SA310413559GA SA310413560GA SA310423635GA SA310423662GA SA310424012GA SA310434242GA SA310444272GA SA310455033GA SA310475540GA SA310475681GA SA310496226GA SA310496471GA SA310496472GA SA310507070GA SA310507077GA SA310507100GA SA310507106GA SA310507111GA SA310507121GA SA310507388GA SA310507412GA SA310507444GA SA310507448GA SA310507451GA SA310507496GA SA310507503GA SA310507527GA SA310507548GA SA310517720GA SA310517737GA SA310517739GA SA310517818GA SA310517821GA SA310517838GA SA310517997GA SA310518677GA SA310528883GA SA310528887GA SA310528888GA SA310528889GA SA310528975GA SA310528980GA SA310528990GA SA310528997GA SA310529004GA SA310529005GA SA310529008GA SA310529048GA SA310529049GA SA310529050GA SA310529053GA SA310529057GA SA310529158GA SA310539321GA SA311059871GA SA311059872GA SA311059873GA SA311059874GA SA311059875GA SA311059876GA SA311059877GA SA311059878GA SA311059879GA SA311059880GA SA311059881GA SA311059882GA SA311059883GA SA311059884GA SA311059885GA SA311059886GA SA311059887GA SA311059888GA SA311059889GA SA311059890GA SA311059891GA SA311059892GA SA311059893GA SA311059894GA SA311059897GA SA311059898GA SA311059900GA SA311059901GA SA311059903GA SA311059904GA SA311059905GA SA311059906GA SA311059907GA SA311059908GA SA311069606GA SA311069726GA SA310423577GA SA310496623GA SA310506854GA SA310507138GA SA310517825GA SA310518004GA SA310528852GA SA310528987GA SA310529000GA SA310539349GA SA311059895GA SA311059896GA SA311059899GA SA311059902GA SA311060134GA SA311060135GA SA310260040GA SA310321001GA SA310351501GA SA310444289GA SA310444521GA SA310475613GA SA310507056GA SA310507541GA SA310517865GA SA310517988GA SA310518628GA SA310518653GA SA310518656GA SA310518664GA SA310518671GA SA311049576GA SA311060108GA SA311069619GA SA311069709GA SA311069710GA SA311069711GA SA311060129GA SA311060151GA SA311060217GA SA310413534GA SA310444248GA SA310444849GA SA310475703GA SA310496148GA SA310496288GA SA310496295GA SA310496310GA SA310496462GA SA310517841GA SA310517878GA SA310528871GA SA310539347GA SA310539351GA SA311039377GA SA311039450GA SA311039464GA SA311039467GA SA311039483GA SA311039485GA SA311060143GA SA311060153GA SA311060160GA SA311060169GA SA311069677GA SA311060075GA SA311060220GA SA310310993GA SA310518400GA SA310529204GA SA310529207GA SA311059592GA SA311060165GA SA311060194GA SA311060209GA SA311060215GA SA311060218GA SA311060309GA SA311069780GA SA310496306GA SA310496314GA SA310496322GA SA310507018GA SA311049578GA SA311060070GA SA311060112GA SA311060115GA SA311060133GA SA311060137GA SA311060234GA SA311060239GA SA311060241GA SA311060242GA SA311060244GA SA311060247GA SA311060248GA SA311060249GA SA311060251GA SA311060261GA SA311060263GA SA311060264GA SA311060268GA SA311060275GA SA311060301GA SA311069730GA SA310351562GA SA310528923GA SA310528933GA SA310528935GA SA311059601GA SA311069657GA SA311069694GA SA311069697GA SA311069745GA SA311060142GA SA311060179GA SA311060184GA SA311060210GA SA311060211GA SA311069721GA SA310361698GA SA310496215GA SA310496220GA SA310496221GA SA310496232GA SA310496296GA SA310496371GA SA310496456GA SA310496457GA SA310496467GA SA310496474GA SA310496489GA SA310496490GA SA310496492GA SA310496499GA SA311060084GA SA311060096GA SA311060098GA SA311060381GA SA311070547GA SA311059581GA SA311060072GA SA311080605GA SA310518272GA SA310518278GA SA310518288GA SA310528880GA SA311060119GA SA311060124GA SA311060139GA SA311060141GA SA311060150GA SA311060152GA SA311060158GA SA311060176GA SA311060182GA SA311060199GA SA311060206GA SA311060231GA SA311060278GA SA311060285GA SA311060296GA SA311060346GA SA311060358GA SA311060363GA SA311069714GA SA311069729GA SA311070523GA SA311070537GA SA311070542GA SA311070574GA SA311070588GA SA311070593GA SA311080599GA SA311080632GA SA311080639GA SA311080704GA SA311090814GA SA311090826GA SA311090832GA SA310528777GA SA311059585GA SA311059588GA SA311059593GA SA311059596GA SA311060077GA SA311060085GA SA311060086GA SA311060104GA SA311060284GA SA311069819GA SA311070474GA SA311070538GA SA311070575GA SA311070579GA SA311070597GA SA311090893GA SA311070521GA SA311070576GA SA311070595GA SA311090834GA SA310518681GA SA311060166GA SA311060192GA SA311060283GA SA311060293GA SA311060294GA SA311060300GA SA311060304GA SA311060318GA SA311060325GA SA311060337GA SA311060349GA SA311060351GA SA311060354GA SA311060359GA SA311060368GA SA311060370GA SA311060389GA SA311080634GA SA311080636GA SA311080640GA SA311080643GA SA311080662GA SA311080710GA SA311080723GA SA311080726GA SA311080741GA SA311090813GA SA311090815GA SA311090819GA SA311090820GA SA310361945GA SA310496276GA SA310496317GA SA311060122GA SA311060127GA SA311060128GA SA311060130GA SA311060131GA SA311060132GA SA311060175GA SA311060177GA SA311060178GA SA311060187GA SA311060190GA SA311060191GA SA311060222GA SA311060223GA SA311060226GA SA311060272GA SA311060276GA SA311060286GA SA311060287GA SA311060288GA SA311060289GA SA311060290GA SA311060292GA SA311060306GA SA311060307GA SA311060312GA SA311060313GA SA311060314GA SA311060315GA SA311060319GA SA311060321GA SA311060344GA SA311060348GA SA311060357GA SA311060361GA SA311060362GA SA311060367GA SA311060390GA SA311069712GA SA311069727GA SA311070540GA SA311070544GA SA311070545GA SA311070546GA SA311070550GA SA311070551GA SA311070552GA SA311070554GA SA311070561GA SA311070563GA SA311070565GA SA311070566GA SA311070570GA SA311070582GA SA311070594GA SA311080600GA SA311080601GA SA311080603GA SA311080608GA SA311080610GA SA311080611GA SA311080614GA SA311080617GA SA311080618GA SA311080619GA SA311080620GA SA311080621GA SA311080622GA SA311080623GA SA311080624GA SA311080625GA SA311080637GA SA311080641GA SA311080649GA SA311080655GA SA310496140GA SA311039510GA SA311039540GA SA311039547GA SA311059597GA SA311060069GA SA311060103GA SA311060188GA SA311060195GA SA311060200GA SA311060203GA SA311060216GA SA311060279GA SA311060282GA SA311060295GA SA311070524GA SA311070596GA SA311080645GA SA311080646GA SA311090775GA SA310528855GA SA311039520GA SA311039526GA SA311039528GA SA311039539GA SA311039545GA SA311060004GA SA311060007GA SA311060023GA SA311060024GA SA311060028GA SA311060035GA SA311060041GA SA311060236GA SA311060302GA SA311069816GA SA311069817GA SA311070518GA SA311070527GA SA311070529GA SA311070532GA SA311070533GA SA311070555GA SA311070572GA SA311070577GA SA311070578GA SA311070581GA SA311080627GA SA311080631GA SA311080656GA SA311080657GA SA311080712GA SA311080733GA SA311080739GA SA311080740GA SA311080742GA SA311090759GA SA311090766GA SA311090771GA SA311090774GA SA311090782GA SA311090785GA SA311090788GA SA311090791GA SA311090796GA SA311090801GA SA311090811GA SA311090812GA SA311090817GA SA311090818GA SA311090825GA SA311090829GA SA311090833GA SA311090836GA SA311090839GA SA311090840GA SA311090841GA SA311090844GA SA311090847GA SA311090848GA SA311090858GA SA311090867GA SA311090880GA SA311090881GA SA310507305GA SA311039537GA SA311060015GA SA311060037GA SA311060067GA SA311060088GA SA311060092GA SA311060106GA SA311070443GA SA311080671GA SA311080699GA SA311080702GA SA311090776GA SA311090781GA SA311070528GA SA311070534GA SA311070539GA SA311070569GA SA311090822GA SA311090824GA SA311101216GA SA310403374GA SA310423869GA SA310434153GA SA310506762GA SA310507164GA SA310507266GA SA310518181GA SA310528746GA SA311039494GA SA311049577GA SA311059580GA SA311059589GA SA311059600GA SA311060027GA SA311060074GA SA311069627GA SA311069638GA SA311069649GA SA311069658GA SA311069758GA SA311080667GA SA311080668GA SA311080688GA SA311080689GA SA311080692GA SA311090753GA SA311090763GA SA311090790GA SA311090794GA SA311090883GA SA311090885GA SA311101174GA SA311101175GA SA311101179GA SA311101192GA SA311101195GA SA311101198GA SA311101199GA SA311101202GA SA311101203GA SA311101214GA SA311101218GA SA311101236GA SA311101237GA SA311101238GA SA311101239GA SA311101242GA SA311101269GA SA310423723GA SA310496283GA SA310496366GA SA310507015GA SA311090806GA SA310434146GA SA311090986GA SA310506771GA SA311059587GA SA311059602GA SA311060017GA SA311060038GA SA311069625GA SA311069639GA SA311069641GA SA311069648GA SA311069651GA SA311069655GA SA311069656GA SA311069659GA SA311069674GA SA311069678GA SA311069785GA SA311069814GA SA311069820GA SA311069828GA SA311069835GA SA311069841GA SA311070476GA SA311080664GA SA311080665GA SA311080666GA SA311080669GA SA311080680GA SA311080684GA SA311080690GA SA311080698GA SA311080700GA SA311080701GA SA311090895GA SA311090897GA SA311090902GA SA311091009GA SA311091023GA SA311091026GA SA311091030GA SA311091037GA SA311091043GA SA311091045GA SA311091050GA SA311091054GA SA311091055GA SA311091058GA SA311091059GA SA311091060GA SA311101063GA SA311101070GA SA311101083GA SA311101280GA SA311101308GA SA311101319GA SA311101321GA SA311101340GA SA311101351GA SA311101359GA SA311101374GA SA311101376GA SA311101381GA SA310382849GA SA310495916GA SA310506786GA SA311039552GA SA311060008GA SA311060011GA SA311060012GA SA311060013GA SA311069653GA SA311069679GA SA311069692GA SA311069699GA SA311069753GA SA311069755GA SA311069762GA SA311069763GA SA311069765GA SA311069767GA SA311069769GA SA311069771GA SA311069772GA SA311069774GA SA311069777GA SA311069781GA SA311069782GA SA311069786GA SA311069787GA SA311069788GA SA311069803GA SA311069827GA SA311069840GA SA311069857GA SA311080663GA SA311080683GA SA311080685GA SA311080694GA SA311080719GA SA311080720GA SA311080730GA SA311080738GA SA311090747GA SA311090750GA SA311090757GA SA311090758GA SA311090772GA SA311090875GA SA311090884GA SA311090890GA SA311090937GA SA311090977GA SA311090978GA SA311090982GA SA311090983GA SA311090984GA SA311090985GA SA311091001GA SA311091018GA SA311091019GA SA311091022GA SA311091024GA SA311091057GA SA311101064GA SA311101065GA SA311101067GA SA311101069GA SA311101071GA SA311101082GA SA311101124GA SA311101311GA SA311101316GA SA311101317GA SA311101320GA SA311101323GA SA311101324GA SA311101325GA SA311101326GA SA311101327GA SA311101328GA SA311101330GA SA311101331GA SA311101337GA SA311101341GA SA311101348GA SA311101353GA SA311101356GA SA311101361GA SA311101365GA SA311101366GA SA311101369GA SA311101370GA SA311101375GA SA311101379GA SA311101430GA SA311101469GA SA311101475GA SA311060323GA SA311060353GA SA311060377GA SA311069665GA SA311090764GA SA311090808GA SA311090874GA SA311101204GA SA311101240GA SA311101253GA SA311101256GA SA311101284GA SA311101285GA SA311101289GA SA311101291GA SA311101298GA SA310403312GA SA310403346GA SA310529054GA SA310539316GA SA311069637GA SA311069644GA SA311090966GA SA311091014GA SA311091015GA SA311090852GA SA311090945GA SA310423772GA SA310518441GA SA311090969GA SA311091011GA SA311091013GA SA311101102GA SA311101119GA SA311101261GA SA311101262GA SA311101263GA SA311090786GA SA311091041GA SA311091051GA SA311132384GA SA311132386GA SA311132435GA SA311132465GA SA311132470GA SA311132507GA SA310517659GA SA310517667GA SA310517679GA SA311090865GA SA311090871GA SA311090889GA SA311101387GA SA311101411GA SA311101477GA SA311101480GA SA311101483GA SA311101494GA SA311101496GA SA311101502GA SA311101503GA SA311101517GA SA311101520GA SA311101521GA SA311101523GA SA311101556GA SA310413497GA SA310413537GA SA310496229GA SA310496303GA SA310507119GA SA311060116GA SA311060117GA SA311060118GA SA311060120GA SA311060123GA SA311060125GA SA311060144GA SA311060145GA SA311060159GA SA311060245GA SA311060253GA SA311060257GA SA311060258GA SA311060259GA SA311060262GA SA311069713GA SA311069715GA SA311069716GA SA311069718GA SA311069720GA SA311069725GA SA311069733GA SA311069735GA SA311059598GA SA311059599GA SA311069744GA SA311080613GA SA311090873GA SA311090963GA SA311091002GA SA311091056GA SA311101066GA SA311101068GA SA311101096GA SA311101104GA SA311101116GA SA311101117GA SA311060393GA SA311070520GA SA311070568GA SA311080652GA SA311080731GA SA311101111GA SA311101180GA SA311101187GA SA311101190GA SA311101228GA SA311101229GA SA311101234GA SA311101244GA SA311101245GA SA311101250GA SA311101264GA SA311101281GA SA311101419GA SA311101421GA SA311070522GA SA311080628GA SA311080711GA SA311080718GA SA311080721GA SA311080724GA SA311080725GA SA311080729GA SA311101304GA SA311101425GA SA311090882GA SA311101388GA SA311101401GA SA311101402GA SA311101391GA SA311101488GA SA311101495GA SA311101506GA SA311101507GA SA311101513GA SA311101515GA SA311101527GA SA311101528GA SA311101531GA SA311101535GA SA311101536GA SA311101544GA SA311111559GA SA311111562GA SA311111568GA SA311111574GA SA311111604GA SA311111610GA SA311101522GA SA311090762GA SA311090765GA SA311090854GA SA311101184GA SA311101217GA SA311101482GA SA311101499GA SA311121714GA SA311121718GA SA311070567GA SA311080635GA SA311080717GA SA311090748GA SA311090755GA SA311090777GA SA311090784GA SA311090807GA SA311090810GA SA311090816GA SA311090868GA SA311101212GA SA311101448GA SA311090894GA SA311090899GA SA311090904GA SA311090959GA SA311090993GA SA311090994GA SA311101074GA SA311101079GA SA311101107GA SA311101163GA SA311101428GA SA311101436GA SA311101446GA SA311101454GA SA311101478GA SA311101505GA SA311111623GA SA311111626GA SA311111629GA SA311121700GA SA311121703GA SA311121749GA SA311132422GA SA311132425GA SA311132455GA SA311132461GA SA311132464GA SA311132490GA SA311132495GA SA311132502GA SA311132505GA SA311132515GA SA311132517GA SA311090954GA SA311090991GA SA311090996GA SA311091000GA SA311101125GA SA311101130GA SA311101168GA SA311111567GA SA311111600GA SA311111609GA SA311121698GA SA311121740GA SA311121741GA SA311121746GA SA311132278GA SA311132441GA SA311132444GA SA311132452GA SA311132460GA SA311132463GA SA311132488GA SA311132491GA SA311132494GA SA311132500GA SA311132508GA SA311132510GA SA311132522GA SA311132527GA SA311060329GA SA311060366GA SA311060374GA SA311080714GA SA311080735GA SA311080743GA SA311080744GA SA311090752GA SA311090756GA SA311090809GA SA311090866GA SA311101519GA SA311101543GA SA311111572GA SA311111575GA SA311111580GA SA311111582GA SA311111594GA SA311111603GA SA311122216GA SA311101383GA SA311101405GA SA311101486GA SA311101532GA SA311101537GA SA311101546GA SA311101547GA SA311111566GA SA311111581GA SA311111593GA SA311070473GA SA311090953GA SA311090971GA SA311101429GA SA311101435GA SA311101473GA SA311101479GA SA311101497GA SA311101514GA SA311101529GA SA311101540GA SA311111601GA SA311121708GA SA311121725GA SA311121742GA SA311121743GA SA311122127GA SA311122128GA SA311132413GA SA311132430GA SA311132438GA SA311132440GA SA311132449GA SA311132454GA SA311132499GA SA311101134GA SA311101334GA SA311101427GA SA311101459GA SA311101498GA SA311101504GA SA311101525GA SA311101538GA SA311101542GA SA311101545GA SA311101549GA SA311101554GA SA311111560GA SA311111571GA SA311111578GA SA311132496GA SA311090908GA SA311090958GA SA311132327GA SA311132521GA SA311132523GA SA311132421GA SA311132509GA SA310507139GA SA311060392GA SA311070477GA SA311070480GA SA311090926GA SA311101177GA SA311101201GA SA311101230GA SA311101303GA SA311101306GA SA311101358GA SA311101372GA SA311101382GA SA311101407GA SA311101412GA SA311101415GA SA311101491GA SA311101493GA SA311101511GA SA311101512GA SA311101516GA SA311101541GA SA311101555GA SA311111558GA SA311111564GA SA311122204GA SA310372076GA SA310382364GA SA310382814GA SA311059584GA SA311060395GA SA311069630GA SA311069645GA SA311069652GA SA311069660GA SA311069663GA SA311069666GA SA311069667GA SA311069688GA SA311069695GA SA311069696GA SA311069701GA SA311069704GA SA311069705GA SA311069706GA SA311090769GA SA311101182GA SA311101200GA SA311101219GA SA311101232GA SA311101233GA SA311101246GA SA311101277GA SA311101290GA SA311101301GA SA311101307GA SA311101416GA SA311101417GA SA311101476GA SA311101485GA SA311101487GA SA311101489GA SA311101510GA SA311101526GA SA311101534GA SA311101539GA SA311111563GA SA311121744GA SA311070509GA SA311090909GA SA311090919GA SA311090923GA SA311090925GA SA311090939GA SA311090947GA SA311090952GA SA311090970GA SA311101349GA SA311101350GA SA311101367GA SA311101371GA SA311101377GA SA311121729GA SA311121730GA SA311122168GA SA311122214GA SA311122215GA SA311132282GA SA311090910GA SA311090931GA SA311090936GA SA311090956GA SA311090957GA SA311090974GA SA311101081GA SA311101122GA SA311101128GA SA311101145GA SA311101149GA SA311101150GA SA311101167GA SA311101171GA SA311101332GA SA311101433GA SA311111624GA SA311111633GA SA311122109GA SA311122117GA SA311122161GA SA311132417GA SA311132492GA SA311080626GA SA311122012GA SA311122074GA SA311122080GA SA311122136GA SA311122153GA SA311122171GA SA311122172GA SA311122182GA SA311122184GA SA311122201GA SA311122208GA SA311122209GA SA311122225GA SA311122239GA SA311122256GA SA311122260GA SA311132303GA SA311132308GA SA311132311GA SA311132312GA SA311132323GA SA311132328GA SA311132332GA SA311132333GA SA311132336GA SA311132338GA SA311132339GA SA311132341GA SA311132346GA SA311132347GA SA311132355GA SA311132356GA SA311132357GA SA311132359GA SA311132361GA SA311132364GA SA311132365GA SA311132390GA SA311132394GA SA311132403GA SA311069770GA SA311069830GA SA311090988GA SA311090990GA SA311091005GA SA311101148GA SA311101346GA SA311121707GA SA311121727GA SA311121728GA SA311121733GA SA311121735GA SA311121738GA SA311121851GA SA311121876GA SA311122056GA SA311122061GA SA311122067GA SA311122124GA SA311122159GA SA311122163GA SA311122178GA SA311122181GA SA311122183GA SA311122202GA SA311122210GA SA311122213GA SA311122217GA SA311122218GA SA311122223GA SA311122228GA SA311122233GA SA311122243GA SA311122246GA SA311122250GA SA311122253GA SA311122255GA SA311122261GA SA311122263GA SA311122264GA SA311122266GA SA311122268GA SA311122269GA SA311122273GA SA311132285GA SA311132289GA SA311132292GA SA311132293GA SA311132304GA SA311132305GA SA311132306GA SA311132329GA SA311132337GA SA311132342GA SA311132345GA SA311132351GA SA311132353GA SA311132354GA SA311132362GA SA311132392GA SA311132400GA SA311132506GA SA311132526GA SA311069654GA SA311069668GA SA311069675GA SA311069683GA SA311069689GA SA311069702GA SA311069708GA SA311069759GA SA311069839GA SA311069849GA SA311069851GA SA311069854GA SA311069856GA SA311069860GA SA311090903GA SA311090980GA SA311091021GA SA311091028GA SA311091031GA SA311091032GA SA311091042GA SA311091044GA SA311091048GA SA311091061GA SA311091062GA SA311101131GA SA311101309GA SA311101310GA SA311101318GA SA311101322GA SA311101335GA SA311101339GA SA311101343GA SA311101344GA SA311101352GA SA311101363GA SA311101380GA SA311101426GA SA311101437GA SA311101438GA SA311101447GA SA311101462GA SA311101471GA SA311101472GA SA311101474GA SA311121798GA SA311121804GA SA311121847GA SA311122108GA SA311122132GA SA311122138GA SA311122155GA SA311122167GA SA311132382GA SA311132557GA SA311132589GA SA311070478GA SA311070479GA SA311070482GA SA311090965GA SA311090997GA SA311091008GA SA311091010GA SA311091016GA SA311091017GA SA311091020GA SA311091025GA SA311101313GA SA311101333GA SA311101338GA SA311101342GA SA311101345GA SA311101347GA SA311101354GA SA311101355GA SA311101357GA SA311101360GA SA311101364GA SA311111608GA SA311122231GA SA311122237GA SA311122240GA SA311122241GA SA311122242GA SA311122247GA SA311122249GA SA311122254GA SA311132300GA SA311132309GA SA311132310GA SA311132315GA SA311132316GA SA311132317GA SA311132319GA SA311132320GA SA311132334GA SA311132335GA SA311132349GA SA311132350GA SA311132363GA SA311132368GA SA311132402GA SA311132407GA SA311132411GA SA311059595GA SA311069822GA SA311069824GA SA311069842GA SA311069847GA SA311090928GA SA311111687GA SA311121712GA SA311121723GA SA311121734GA SA311121755GA SA311121758GA SA311121766GA SA311121767GA SA311121800GA SA311121802GA SA311121803GA SA311121807GA SA311121808GA SA311121821GA SA311121828GA SA311121831GA SA311121835GA SA311121848GA SA311121946GA SA311121999GA SA311122001GA SA311122060GA SA311122113GA SA311122131GA SA311122185GA SA311122234GA SA311122245GA SA311122252GA SA311122265GA SA311122271GA SA311122272GA SA311132276GA SA311132296GA SA311132298GA SA311132301GA SA311132307GA SA311132331GA SA311132343GA SA311132344GA SA311132358GA SA311132360GA SA311132369GA SA311132370GA SA311132374GA SA311132375GA SA311132376GA SA311132378GA SA311132381GA SA311132383GA SA311132385GA SA311132391GA SA311132393GA SA311132395GA SA311132396GA SA311132397GA SA311132401GA SA311132405GA SA311132408GA SA311132409GA SA311132410GA SA311132434GA SA311132457GA SA311132458GA SA311132459GA SA311132467GA SA311132468GA SA311132472GA SA311132476GA SA311132479GA SA311132481GA SA311132482GA SA311132533GA SA311132534GA SA311132537GA SA311132549GA SA311132550GA SA311132552GA SA311132553GA SA311132555GA SA311132570GA SA311132581GA SA311132582GA SA311132585GA SA311132587GA SA311132591GA SA311132595GA SA311132597GA SA310528937GA SA311070475GA SA311070481GA SA311090896GA SA311090905GA SA311090906GA SA311090912GA SA311090917GA SA311090929GA SA311090930GA SA311090942GA SA311090943GA SA311090961GA SA311090962GA SA311101095GA SA311101110GA SA311101112GA SA311111689GA SA311121768GA SA311121773GA SA311121777GA SA311121782GA SA311121783GA SA311121792GA SA311121806GA SA311121809GA SA311121813GA SA311121816GA SA311121817GA SA311121827GA SA311122020GA SA311122024GA SA311132418GA SA311132433GA SA311132439GA SA311132447GA SA311132448GA SA310392898GA SA310434058GA SA310434123GA SA310434137GA SA310465295GA SA310496410GA SA310496411GA SA310507316GA SA310528810GA SA310528825GA SA310529115GA SA311060020GA SA311069676GA SA311069681GA SA311069682GA SA311070409GA SA311080682GA SA311080687GA SA311080695GA SA311090940GA SA311090941GA SA311090960GA SA311122118GA SA311122119GA SA311122123GA SA311101302GA SA311122034GA SA311142754GA SA311142758GA SA311142760GA SA311142763GA SA311142765GA SA311142768GA SA311142769GA SA311142775GA SA311142779GA SA311142780GA SA311142782GA SA311142784GA SA311142786GA SA311142788GA SA311142759GA SA311142767GA SA311101283GA SA311121867GA SA311121877GA SA311122017GA SA311122051GA SA311122211GA SA311142973GA SA311101432GA SA311142748GA SA311142749GA SA311142756GA SA311142783GA SA311142791GA SA311142793GA SA311142941GA SA311142979GA SA311101143GA SA311101155GA SA311101441GA SA311101442GA SA311101451GA SA311101456GA SA311101461GA SA311101463GA SA311101464GA SA311101465GA SA311111659GA SA311101078GA SA311101091GA SA311101113GA SA311101126GA SA311101136GA SA311101158GA SA311101164GA SA311101165GA SA311101450GA SA311101467GA SA311101468GA SA311111631GA SA311111680GA SA311111681GA SA311121717GA SA311121748GA SA311121765GA SA311122014GA SA311122022GA SA311122032GA SA311122033GA SA311122035GA SA311122050GA SA311122133GA SA311122135GA SA311122149GA SA311069844GA SA311069845GA SA311090901GA SA311090932GA SA311090975GA SA311090981GA SA311111688GA SA311111694GA SA311121775GA SA311121780GA SA311121822GA SA311121823GA SA311121830GA SA311121899GA SA311122104GA SA311132498GA SA311132504GA SA311132512GA SA311132514GA SA311132558GA SA311132561GA SA311132563GA SA311132565GA SA311132573GA SA311132574GA SA311132577GA SA311132586GA SA311142764GA SA311142792GA SA311142873GA SA311142876GA SA311142877GA SA311142882GA SA311142892GA SA311142896GA SA311142896GA SA311142898GA SA311142901GA SA311142904GA SA311142922GA SA311142944GA SA311142945GA SA311142969GA SA311142976GA SA311111662GA SA311111663GA SA311111667GA SA311111671GA SA311111674GA SA311111676GA SA311111677GA SA311111679GA SA311111685GA SA311111686GA SA311111695GA SA311111696GA SA311101086GA SA311101087GA SA311101109GA SA311101458GA SA311111588GA SA311111589GA SA311111597GA SA311111598GA SA311111605GA SA311111613GA SA311111625GA SA311111637GA SA311121721GA SA311122065GA SA311122079GA SA311122082GA SA311142890GA SA311101147GA SA311111570GA SA311111576GA SA311111585GA SA311111587GA SA311111590GA SA311111591GA SA311111592GA SA311111596GA SA311111607GA SA311111612GA SA311121858GA SA311121868GA SA311121880GA SA311121892GA SA311122025GA SA311132578GA SA311142874GA SA311142883GA SA311142886GA SA311142900GA SA311142905GA SA311101431GA SA311121724GA SA311121820GA SA311121992GA SA311122003GA SA311122027GA SA311122042GA SA311122049GA SA311122084GA SA311122088GA SA311122098GA SA311122114GA SA311122116GA SA311142921GA SA311142943GA SA311153110GA SA311121890GA SA311132415GA SA311132572GA SA311132579GA SA311132592GA SA311132594GA SA311142789GA SA311142879GA SA311142880GA SA311142891GA SA311153124GA SA311142700GA SA311153101GA SA311153149GA SA311153161GA SA311153164GA SA311153206GA SA311153209GA SA311153210GA SA311101101GA SA311142707GA SA311142823GA SA311142910GA SA311142913GA SA311142954GA SA311142963GA SA311142971GA SA311142972GA SA311142982GA SA311153112GA SA311153172GA SA311090892GA SA311090898GA SA311121751GA SA311121770GA SA311121776GA SA311121833GA SA311122085GA SA311122095GA SA311122105GA SA311122120GA SA311122150GA SA311122156GA SA311142802GA SA311060386GA SA311090878GA SA311090888GA SA311121856GA SA311121885GA SA311121891GA SA311121897GA SA311121909GA SA311122011GA SA311122175GA SA311122186GA SA311122200GA SA311122205GA SA311132325GA SA311132453GA SA311142790GA SA311142931GA SA311142974GA SA311153080GA SA311153144GA SA311153154GA SA311153157GA SA311153163GA SA311153192GA SA311153213GA SA311153217GA SA311153223GA SA311153225GA SA310475517GA SA310506675GA SA310506696GA SA310506700GA SA310506731GA SA310506733GA SA310506734GA SA310506739GA SA310506744GA SA310506754GA SA311132562GA SA311142785GA SA311142920GA SA311153084GA SA311153090GA SA311153115GA SA311153165GA SA311153169GA SA311153176GA SA311142679GA SA311142689GA SA311142696GA SA311132416GA SA311132427GA SA311069684GA SA311111654GA SA311111655GA SA311111657GA SA311121981GA SA310528939GA SA311142984GA SA311039377GA SA311101453GA SA311101470GA SA311090838GA SA311132606GA SA311132612GA SA311132623GA SA311132636GA SA311142781GA SA311142795GA SA311132590GA SA311069796GA SA311153068GA SA311153230GA SA311153236GA SA311153238GA SA311153240GA SA311153243GA SA311153245GA SA311153248GA SA311153249GA SA311153250GA SA311153275GA SA311153276GA SA311153281GA SA311153287GA SA311153288GA SA311153293GA SA311153294GA SA311153295GA SA311153297GA SA311153300GA SA311039495GA SA311059590GA SA311059603GA SA311069650GA SA311069747GA SA311069760GA SA311069764GA SA311069768GA SA311069779GA SA311069792GA SA311069793GA SA311069794GA SA311069795GA SA311069798GA SA311069799GA SA311069801GA SA311069804GA SA311069813GA SA311069833GA SA311069843GA SA311069853GA SA311121799GA SA311121843GA SA311153043GA SA311060396GA SA311153216GA SA311163393GA SA311142719GA SA311142907GA SA311142911GA SA311153096GA SA311153104GA SA311153111GA SA311153118GA SA311153136GA SA311153170GA SA311153188GA SA311153190GA SA311153194GA SA311153195GA SA311153196GA SA311153197GA SA311153219GA SA311153266GA SA311111692GA SA311121788GA SA311121790GA SA311121796GA SA311121801GA SA311121814GA SA311121815GA SA311121819GA SA311121829GA SA311121832GA SA311121838GA SA311121839GA SA311121840GA SA311121844GA SA311121845GA SA311132372GA SA311132389GA SA311132529GA SA311132616GA SA311132630GA SA311142713GA SA311142716GA SA311142718GA SA311142745GA SA311142752GA SA311142798GA SA311142804GA SA311142806GA SA311142903GA SA311142929GA SA311142964GA SA311142966GA SA311153088GA SA311153089GA SA311153094GA SA311153108GA SA311153114GA SA311153117GA SA311153119GA SA311153120GA SA311153129GA SA311153132GA SA311153138GA SA311153139GA SA311153146GA SA311153150GA SA311153162GA SA311153181GA SA311153189GA SA311153226GA SA311153256GA SA311153350GA SA311153351GA SA311153352GA SA311163376GA SA311163398GA SA311163884GA SA311163899GA SA311163981GA SA311101273GA SA311111569GA SA311121701GA SA311122016GA SA311122023GA SA311122039GA SA311122041GA SA311122048GA SA311142885GA SA311142894GA SA311142935GA SA311142936GA SA311142937GA SA311142938GA SA311142940GA SA311142942GA SA311142946GA SA311142958GA SA311163412GA SA311163873GA SA311163900GA SA311163923GA SA311163926GA SA311163928GA SA311163941GA SA311163948GA SA311163950GA SA311163977GA SA311163989GA SA311163995GA SA311163996GA SA311142720GA SA311153317GA SA311163783GA SA311153036GA SA311111665GA SA311111666GA SA311111668GA SA311111670GA SA311111678GA SA311111682GA SA311111683GA SA311111684GA SA311111690GA SA311111693GA SA311121752GA SA311121759GA SA311121763GA SA311121769GA SA311121772GA SA311121785GA SA311121791GA SA311121793GA SA311121794GA SA311121795GA SA311121824GA SA311121825GA SA311132483GA SA311132610GA SA311132620GA SA311153065GA SA311163872GA SA311163905GA SA311163968GA SA311142678GA SA311142717GA SA311142727GA SA311142739GA SA311142824GA SA311142852GA SA311153241GA SA311153247GA SA311153257GA SA311153264GA SA311153272GA SA311153299GA SA311163901GA SA311163902GA SA311163903GA SA311163904GA SA311163916GA SA311163919GA SA311163949GA SA311163970GA SA311163986GA SA311163991GA SA311163992GA SA311163390GA SA311163862GA SA311132604GA SA311142721GA SA311153032GA SA311153047GA SA311153048GA SA311153052GA SA311153060GA SA311163403GA SA311163416GA SA311163858GA SA311163878GA SA311163891GA SA311163896GA SA311163906GA SA311163907GA SA311163927GA SA311163957GA SA311163959GA SA311195079GA SA311205456GA SA311205486GA SA311205501GA SA311163401GA SA311163410GA SA311163857GA SA311163860GA SA311163868GA SA311163871GA SA311163876GA SA311163880GA SA311163881GA SA311163882GA SA311163886GA SA311163894GA SA311163911GA SA311163913GA SA311163924GA SA311163938GA SA311163942GA SA311163952GA SA311163953GA SA311163954GA SA311163963GA SA311163965GA SA311163966GA SA311163967GA SA311163969GA SA311163982GA SA311163985GA SA311163990GA SA311256694GA SA311163369GA SA311163406GA SA311195056GA SA311205463GA SA311205480GA SA311163628GA SA311163639GA SA311184454GA SA311163782GA SA311163846GA SA311163936GA SA311163937GA SA311174166GA SA311174181GA SA311184355GA SA311184437GA SA311163915GA SA311163922GA SA311163932GA SA311174158GA SA311174165GA SA311174180GA SA311174186GA SA311174187GA SA311174189GA SA311174195GA SA311174196GA SA311174200GA SA311174205GA SA311174208GA SA311174213GA SA311174214GA SA311174215GA SA311174216GA SA311174217GA SA311174219GA SA311174225GA SA311174226GA SA311174227GA SA311174229GA SA311174231GA SA311174232GA SA311174233GA SA311174246GA SA311174247GA SA311174248GA SA311174249GA SA311174250GA SA311174251GA SA311174253GA SA311184336GA SA311184343GA SA311184346GA SA311184350GA SA311184351GA SA311184352GA SA311184354GA SA311184356GA SA311184357GA SA311184358GA SA311184361GA SA311184365GA SA311184368GA SA311184369GA SA311184374GA SA311184378GA SA311184381GA SA311184382GA SA311184385GA SA311184389GA SA311184395GA SA311184403GA SA311184404GA SA311184405GA SA311184408GA SA311184409GA SA311184411GA SA311184540GA SA311184550GA SA311184555GA SA311184561GA SA311184580GA SA311184586GA SA311184596GA SA311287202GA SA311153055GA SA311153319GA SA311153333GA SA311153342GA SA311153347GA SA311163430GA SA311163435GA SA311163436GA SA311163459GA SA311163460GA SA311163466GA SA311163469GA SA311163483GA SA311163490GA SA311163567GA SA311163696GA SA311163710GA SA311163723GA SA311163724GA SA311163726GA SA311163784GA SA311163799GA SA311163800GA SA311163917GA SA311174004GA SA311174016GA SA311174019GA SA311174035GA SA311174073GA SA311174080GA SA311174191GA SA311174244GA SA311184407GA SA311184415GA SA311184416GA SA311184418GA SA311184420GA SA311184425GA SA311184426GA SA311184432GA SA311184434GA SA311184435GA SA311184436GA SA311184441GA SA311184444GA SA311184446GA SA311184447GA SA311184448GA SA311184452GA SA311184455GA SA311184458GA SA311184459GA SA311184462GA SA311184522GA SA311184523GA SA311184531GA SA311184548GA SA311184549GA SA311184563GA SA311184601GA SA311184602GA SA311184605GA SA311184609GA SA311121995GA SA311142722GA SA311142848GA SA311143015GA SA311153056GA SA311153072GA SA311153237GA SA311153318GA SA311153335GA SA311153340GA SA311163446GA SA311163453GA SA311163606GA SA311194950GA SA311194961GA SA311194965GA SA311194977GA SA311194986GA SA311194996GA SA311194998GA SA311195000GA SA311195003GA SA311195016GA SA311195037GA SA311195038GA SA311195040GA SA311205290GA SA311205291GA SA311205292GA SA311205294GA SA311205295GA SA311205296GA SA311205297GA SA311205298GA SA311205299GA SA311205300GA SA311205301GA SA311205302GA SA311205303GA SA311205307GA SA311205308GA SA311205309GA SA311205312GA SA311205314GA SA311205316GA SA311205319GA SA311205320GA SA311205321GA SA311205324GA SA311205326GA SA311205327GA SA311205330GA SA311205332GA SA311205333GA SA311205334GA SA311205335GA SA311205336GA SA311205338GA SA311132639GA SA311142714GA SA311142821GA SA311142854GA SA311142990GA SA311143001GA SA311143004GA SA311153044GA SA311153050GA SA311153059GA SA311153063GA SA311174133GA SA311174164GA SA311174168GA SA311174172GA SA311174194GA SA311174221GA SA311174222GA SA311184359GA SA311184360GA SA311184364GA SA311184366GA SA311184367GA SA311184392GA SA311184394GA SA311184396GA SA311184414GA SA311225993GA SA311256687GA SA311256695GA SA311256788GA SA311153228GA SA311153233GA SA311153246GA SA311153251GA SA311153252GA SA311153253GA SA311153261GA SA311153263GA SA311153267GA SA311153269GA SA311153270GA SA311153274GA SA311153277GA SA311153279GA SA311153285GA SA311153291GA SA311153296GA SA311153304GA SA311163698GA SA311184417GA SA311163479GA SA311174122GA SA311184421GA SA311184424GA SA311184428GA SA311184431GA SA311184440GA SA311184450GA SA311184479GA SA311184495GA SA311184663GA SA311195070GA SA311195081GA SA311195082GA SA311195083GA SA311195086GA SA311195088GA SA311195090GA SA311195097GA SA311195099GA SA311195100GA SA311195104GA SA311195105GA SA311195107GA SA311195110GA SA311195111GA SA311195112GA SA311195113GA SA311195119GA SA311195120GA SA311195122GA SA311195259GA SA311195260GA SA311195262GA SA311195263GA SA311195279GA SA311195282GA SA311205310GA SA311236076GA SA311236085GA SA311236092GA SA311256709GA SA311256766GA SA311256767GA SA311256768GA SA311256771GA SA311256773GA SA311256775GA SA311256778GA SA311256783GA SA311256789GA SA311256791GA SA311256792GA SA311256795GA SA311256796GA SA311256800GA SA311256801GA SA311266899GA SA311266952GA SA311266963GA SA311266977GA SA311266980GA SA311266991GA SA311267009GA SA311060328GA SA311060355GA SA311142996GA SA311163417GA SA311174075GA SA311174240GA SA311174254GA SA311184397GA SA311184412GA SA311184487GA SA311184665GA SA311184666GA SA311184667GA SA311195053GA SA311195055GA SA311195057GA SA311195058GA SA311195059GA SA311195073GA SA311195077GA SA311246546GA SA311246562GA SA311256786GA SA311266959GA SA311174066GA SA311184456GA SA311184463GA SA311184467GA SA311184469GA SA311184472GA SA311184475GA SA311184478GA SA311184481GA SA311184499GA SA311184503GA SA311184509GA SA311195092GA SA311195096GA SA311195108GA SA311195115GA SA311195117GA SA311195151GA SA311205448GA SA311205450GA SA311205451GA SA311205455GA SA311205461GA SA311205465GA SA311205471GA SA311205473GA SA311205477GA SA311205478GA SA311205483GA SA311205484GA SA311205499GA SA311205500GA SA311121983GA SA311122008GA SA311142997GA SA311143007GA SA311153301GA SA311153346GA SA311163440GA SA311163561GA SA311163585GA SA311174072GA SA311174084GA SA311174144GA SA311184429GA SA311184439GA SA311184466GA SA311184483GA SA311184494GA SA311184496GA SA311246517GA SA311246529GA SA311246534GA SA311246542GA SA311266955GA SA311266972GA SA311266973GA SA311266974GA SA311266976GA SA311266978GA SA311266979GA SA311266982GA SA311266983GA SA311266985GA SA311266988GA SA311266990GA SA311266997GA SA311267011GA SA310382768GA SA311060376GA SA311070519GA SA311121837GA SA311122072GA SA311142856GA SA311153069GA SA311153231GA SA311153234GA SA311153280GA SA311153290GA SA311153292GA SA311153298GA SA311153303GA SA311153326GA SA311153357GA SA311153365GA SA311163428GA SA311163431GA SA311163437GA SA311163439GA SA311163442GA SA311163447GA SA311163449GA SA311163464GA SA311163468GA SA311163473GA SA311163480GA SA311163487GA SA311163545GA SA311163552GA SA311163560GA SA311163580GA SA311163613GA SA311163623GA SA311163624GA SA311163645GA SA311184390GA SA311184430GA SA311184445GA SA311184465GA SA311184489GA SA311194976GA SA311195008GA SA311195050GA SA311195098GA SA311195102GA SA311195125GA SA311195126GA SA311195128GA SA311195129GA SA311195132GA SA311195137GA SA311195139GA SA311195141GA SA311195143GA SA311195144GA SA311195149GA SA311195162GA SA311195163GA SA311195166GA SA311195185GA SA311195217GA SA311195220GA SA311195222GA SA311195224GA SA311195225GA SA311195226GA SA311195228GA SA311195229GA SA311195232GA SA311195237GA SA311195239GA SA311195241GA SA311195246GA SA311195253GA SA311195273GA SA311195280GA SA311195281GA SA311195286GA SA311205341GA SA311205343GA SA311205348GA SA311205353GA SA311205355GA SA311205356GA SA311205357GA SA311205359GA SA311205363GA SA311205364GA SA311205449GA SA311205452GA SA311205453GA SA311205457GA SA311205458GA SA311205469GA SA311205474GA SA311205479GA SA311205481GA SA311205482GA SA311205487GA SA311205488GA SA311205490GA SA311205494GA SA311205495GA SA311225976GA SA311225987GA SA311226002GA SA311236082GA SA311236098GA SA311236104GA SA311236116GA SA311236286GA SA311236288GA SA311236293GA SA311236337GA SA311246443GA SA311246516GA SA311246518GA SA311246520GA SA311246522GA SA311246523GA SA311246524GA SA311246525GA SA311246527GA SA311246528GA SA311246531GA SA311246537GA SA311246543GA SA311246545GA SA311246547GA SA311246551GA SA311246552GA SA311246559GA SA311246560GA SA311246566GA SA311246568GA SA311246570GA SA311246571GA SA311246573GA SA311246576GA SA311246581GA SA311256769GA SA311256787GA SA311256797GA SA311256799GA SA311266953GA SA311266954GA SA311266956GA SA311266957GA SA311266958GA SA311266960GA SA311266965GA SA311266967GA SA311266968GA SA311266969GA SA311266970GA SA311266971GA SA311266975GA SA311266981GA SA311266984GA SA311266987GA SA311266989GA SA311267000GA SA311267003GA SA311267004GA SA311267006GA SA311267007GA SA311267010GA SA311142816GA SA311153037GA SA311153039GA SA311153054GA SA311153071GA SA311153255GA SA311153282GA SA311153309GA SA311153311GA SA311153356GA SA311153358GA SA311163433GA SA311163441GA SA311163443GA SA311163445GA SA311163470GA SA311163484GA SA311163584GA SA311163587GA SA311163592GA SA311163617GA SA311163649GA SA311184493GA SA311184688GA SA311194940GA SA311194966GA SA311194971GA SA311194997GA SA311195011GA SA311195012GA SA311195013GA SA311195014GA SA311195039GA SA311195041GA SA311195173GA SA311195176GA SA311195192GA SA311195193GA SA311195195GA SA311195197GA SA311195202GA SA311195216GA SA311195219GA SA311195221GA SA311195234GA SA311195244GA SA311236296GA SA311236297GA SA311236312GA SA311236319GA SA311236334GA SA311246398GA SA311246418GA SA311246526GA SA311246530GA SA311246549GA SA311246567GA SA311163448GA SA311163729GA SA311163801GA SA311174053GA SA311174060GA SA311174062GA SA311184419GA SA311184427GA SA311184433GA SA311184443GA SA311184451GA SA311184484GA SA311184484GAMKE SA311184492GA SA311184500GA SA311184502GA SA311184506GA SA311184671GA SA311184674GA SA311184680GA SA311184683GA SA311184685GA SA311184687GA SA311184703GA SA311184704GA SA311184705GA SA311184707GA SA311184708GA SA311184712GA SA311184713GA SA311184715GA SA311195227GA SA311195230GA SA311195231GA SA311195233GA SA311195242GA SA311205443GA SA311205446GA SA311205460GA SA311205467GA SA311205492GA SA311236292GA SA311236323GA SA311236326GA SA311236332GA SA311236338GA SA311163489GA SA311163542GA SA311174149GA SA311195064GA SA311195089GA SA311195091GA SA311195101GA SA311195106GA SA311195118GA SA311195123GA SA311195124GA SA311195133GA SA311195135GA SA311195140GA SA311195142GA SA311195145GA SA311195148GA SA311195157GA SA311195160GA SA311195161GA SA311195165GA SA311195168GA SA311195169GA SA311205445GA SA311205464GA SA311236079GA SA311236187GA SA311236244GA SA311236276GA SA311236344GA SA311256721GA SA311142686GA SA311142730GA SA311142801GA SA311142815GA SA311142853GA SA311174085GA SA311174131GA SA311184480GA SA311184507GA SA311184513GA SA311184643GA SA311184675GA SA311184676GA SA311184698GA SA311184702GA SA311184709GA SA311184710GA SA311205362GA SA311225980GA SA311226000GA SA311236075GA SA311236081GA SA311236087GA SA311236090GA SA311236094GA SA311236097GA SA311236100GA SA311236101GA SA311236102GA SA311236110GA SA311236111GA SA311236294GA SA311236341GA SA311111638GA SA311111640GA SA311111641GA SA311111669GA SA311111675GA SA311121747GA SA311121750GA SA311121760GA SA311121761GA SA311121764GA SA311121789GA SA311122069GA SA311122125GA SA311122130GA SA311142723GA SA311142746GA SA311163971GA SA311174176GA SA311184485GA SA311184628GA SA311184630GA SA311184631GA SA311184640GA SA311184657GA SA311184662GA SA311195067GA SA311195080GA SA311236077GA SA311236285GA SA311236290GA SA311236291GA SA311236322GA SA311236327GA SA311236335GA SA311236340GA SA311163815GA SA311174156GA SA311184529GA SA311184535GA SA311184624GA SA311184627GA SA311184629GA SA311184644GA SA311184654GA SA311184659GA SA311184695GA SA311184714GA SA311195215GA SA311153244GA SA311153278GA SA311153302GA SA311153320GA SA311153323GA SA311153327GA SA311153329GA SA311153339GA SA311153341GA SA311153343GA SA311153353GA SA311153359GA SA311153362GA SA311163562GA SA311163586GA SA311163625GA SA311163640GA SA311163641GA SA311174059GA SA311174064GA SA311174126GA SA311184468GA SA311184504GA SA311184505GA SA311184679GA SA311184681GA SA311184692GA SA311184717GA SA311184718GA SA311184740GA SA311184746GA SA311184748GA SA311184781GA SA311184801GA SA311184821GA SA311236325GA SA311163934GA SA311174159GA SA311184658GA SA311236330GA SA311236336GA SA311246411GA SA311184588GA SA311184697GA SA311184796GA SA311205408GA SA311205429GA SA311205434GA SA311153334GA SA311184795GA SA311184815GA SA311184820GA SA311184827GA SA311184828GA SA311184862GA SA311236035GA SA311236233GA SA311236234GA SA311236235GA SA311236239GA SA311236240GA SA311236241GA SA311236242GA SA311236245GA SA311236246GA SA311236248GA SA311236260GA SA311236261GA SA311236262GA SA311236263GA SA311236264GA SA311236265GA SA311236269GA SA311236270GA SA311236273GA SA311236274GA SA311236275GA SA311236277GA SA311236278GA SA311236279GA SA311163935GA SA311184626GA SA311184642GA SA311184656GA SA311184771GA SA311184778GA SA311184783GA SA311184818GA SA311195054GA SA311195094GA SA311266962GA SA311195164GA SA311153242GA SA311153273GA SA311153321GA SA311163559GA SA311163620GA SA311163817GA SA311163818GA SA311163824GA SA311163840GA SA311163943GA SA311174002GA SA311174007GA SA311174024GA SA311174027GA SA311174031GA SA311174038GA SA311174040GA SA311174049GA SA311174051GA SA311174052GA SA311174076GA SA311174078GA SA311174081GA SA311174083GA SA311174087GA SA311174088GA SA311174089GA SA311174091GA SA311174093GA SA311174095GA SA311174096GA SA311174097GA SA311174098GA SA311174100GA SA311174103GA SA311174107GA SA311174119GA SA311174120GA SA311174121GA SA311174123GA SA311174130GA SA311174140GA SA311174141GA SA311174142GA SA311174143GA SA311174147GA SA311174150GA SA311174151GA SA311174152GA SA311174153GA SA311184460GA SA311184749GA SA311195066GA SA311163574GA SA311163630GA SA311174071GA SA311184461GA SA311184491GA SA311184574GA SA311184668GA SA311184669GA SA311184706GA SA311184734GA SA311184756GA SA311195206GA SA311205391GA SA311205392GA SA311205427GA SA311184616GA SA311184655GA SA311236315GA SA311236317GA SA311184637GA SA311195175GA SA311195203GA SA311184647GA SA311195261GA SA311195269GA SA311184744GA SA311194955GA SA311194960GA SA311194969GA SA311195010GA SA311195019GA SA311195174GA SA311195178GA SA311195183GA SA311195186GA SA311195187GA SA311195189GA SA311195207GA SA311205440GA SA311256774GA SA311266992GA SA311287109GA SA311287113GA SA311287123GA SA311287134GAMKE SA311287171GA SA311287197GA SA311287216GA SA311287217GA SA311184638GA SA311184652GA SA311205373GA SA311205376GA SA311205379GA SA311205491GA SA311236064GA SA311287112GA SA311287115GA SA311287131GA SA311287132GA SA311287166GA SA311287170GA SA311287178GA SA311287182GA SA311287211GA SA311297260GA SA311297348GA SA311297356GA SA311297364GA SA311317709GA SA311153073GA SA311184422GA SA311184453GA SA311205304GA SA311205305GA SA311205318GA SA311205409GA SA311184569GA SA311184579GA SA311184633GA SA311163533GA SA311163543GA SA311163565GA SA311163579GA SA311163627GA SA311163629GA SA311163633GA SA311163790GA SA311174009GA SA311184423GA SA311184793GA SA311184811GA SA311184876GA SA311184877GA SA311194880GA SA311194891GA SA311194926GA SA311195002GA SA311195005GA SA311195020GA SA311246679GA SA311287120GA SA311287126GA SA311287128GA SA311287139GA SA311287142GA SA311287148GA SA311287155GA SA311287156GA SA311287162GA SA311287167GA SA311287168GA SA311287169GA SA311287174GA SA311287175GA SA311287177GA SA311287180GA SA311287183GA SA311287184GA SA311287185GA SA311287186GA SA311287187GA SA311287188GA SA311287189GA SA311287191GA SA311287192GA SA311287193GA SA311287194GA SA311287196GA SA311287198GA SA311287199GA SA311287200GA SA311287201GA SA311287203GA SA311287205GA SA311287206GA SA311287207GA SA311287208GA SA311287209GA SA311287210GA SA311287212GA SA311287213GA SA311287214GA SA311287215GA SA311297255GA SA311297256GA SA311297257GA SA311297258GA SA311297263GA SA311297265GA SA311297267GA SA311297270GA SA311297276GA SA311297277GA SA311297278GA SA311297280GA SA311297281GA SA311297285GA SA311297286GA SA311297287GA SA311297289GA SA311297329GA SA311297333GA SA311297336GA SA311297337GA SA311297338GA SA311297339GA SA311297343GA SA311297344GA SA311297345GA SA311297347GA SA311297355GA SA311297358GA SA311297359GA SA311297361GA SA311297363GA SA311153316GA SA311153328GA SA311153330GA SA311153344GA SA311163734GA SA311205390GA SA311236343GA SA311246384GA SA311246553GA SA311246558GA SA311246564GA SA311246572GA SA311256706GA SA311256708GA SA311256714GA SA311256716GA SA311256717GA SA311256719GA SA311256720GA SA311266886GA SA311287106GA SA311287114GA SA311287116GA SA311287117GA SA311287118GA SA311287119GA SA311287130GA SA311287135GA SA311287140GA SA311287141GA SA311287143GA SA311287144GA SA311132651GA SA311163612GA SA311163776GA SA311163830GA SA311163838GA SA311174032GA SA311174041GA SA311184477GA SA311184672GA SA311184673GA SA311184677GA SA311184678GA SA311184682GA SA311184689GA SA311184691GA SA311184693GA SA311184694GA SA311184699GA SA311184700GA SA311184701GA SA311184720GA SA311184726GA SA311184727GA SA311184728GA SA311184731GA SA311184741GA SA311184742GA SA311184743GA SA311184745GA SA311184755GA SA311184757GA SA311184759GA SA311184766GA SA311184767GA SA311184772GA SA311184773GA SA311184774GA SA311184775GA SA311184777GA SA311184780GA SA311184785GA SA311184787GA SA311184794GA SA311184802GA SA311184804GA SA311184806GA SA311184807GA SA311184814GA SA311184817GA SA311184860GA SA311184878GA SA311194892GA SA311194895GA SA311194898GA SA311194900GA SA311194902GA SA311194903GA SA311194908GA SA311194910GA SA311194912GA SA311194914GA SA311194919GA SA311194928GA SA311194936GA SA311194945GA SA311194949GA SA311194951GA SA311194963GA SA311194964GA SA311194967GA SA311194970GA SA311194982GA SA311194988GA SA311195179GA SA311195180GA SA311195188GA SA311195190GA SA311195194GA SA311195196GA SA311195200GA SA311195201GA SA311195205GA SA311195210GA SA311195211GA SA311195212GA SA311195213GA SA311195214GA SA311195240GA SA311195247GA SA311205397GA SA311205398GA SA311205402GA SA311205412GA SA311205414GA SA311205424GA SA311205425GA SA311205432GA SA311215845GA SA311215853GA SA311256755GA SA311266925GA SA311184686GA SA311184782GA SA311184812GA SA311184834GA SA311184836GA SA311184837GA SA311194901GA SA311194906GA SA311194907GA SA311194909GA SA311194947GA SA311194953GA SA311194979GA SA311194981GA SA311194992GA SA311205395GA SA311205399GA SA311205400GA SA311205410GA SA311205411GA SA311205439GA SA311205573GA SA311215835GA SA311215836GA SA311215839GA SA311215861GA SA311225906GA SA311256751GA SA311297351GA SA311297353GA SA311327836GA SA311337973GA SA311236295GA SA311225900GA SA311142740GA SA311184516GA SA311184819GA SA311184822GA SA311184839GA SA311184840GA SA311194879GA SA311194881GA SA311194887GA SA311194894GA SA311194922GA SA311194937GA SA311194946GA SA311194954GA SA311194958GA SA311194968GA SA311194972GA SA311194975GA SA311194985GA SA311194990GA SA311194999GA SA311195018GA SA311195191GA SA311205350GA SA311215724GA SA311215725GA SA311215735GA SA311215737GA SA311215738GA SA311215741GA SA311236066GA SA311236067GA SA311236068GA SA311236069GA SA311236074GA SA311236078GA SA311266878GA SA311266879GA SA311266880GA SA311266881GA SA311266882GA SA311266883GA SA311266889GA SA311266890GA SA311266894GA SA311266895GA SA311266896GA SA311266898GA SA311266901GA SA311266903GA SA311266906GA SA311266907GA SA311317656GA SA311317658GA SA311317664GA SA311317670GA SA311101132GA SA311142712GA SA311153035GA SA311163461GA SA311163598GA SA311163789GA SA311174128GA SA311184747GA SA311184758GA SA311184789GA SA311184800GA SA311184809GA SA311184813GA SA311184824GA SA311184846GA SA311194882GA SA311194884GA SA311194886GA SA311194888GA SA311194905GA SA311194915GA SA311194916GA SA311194917GA SA311194920GA SA311194925GA SA311194929GA SA311194930GA SA311194943GA SA311194952GA SA311194956GA SA311194974GA SA311194983GA SA311194989GA SA311194991GA SA311205360GA SA311205381GA SA311205466GA SA311215691GA SA311225909GA SA311184723GA SA311184830GA SA311184835GA SA311184859GA SA311194883GA SA311194889GA SA311194897GA SA311194913GA SA311194921GA SA311194941GA SA311195204GA SA311205574GA SA311205597GA SA311205607GA SA311205617GA SA311205618GA SA311205619GA SA311194885GA SA311163576GA SA311163647GA SA311163650GA SA311163718GA SA311163832GA SA311163844GA SA311174125GA SA311184867GA SA311195045GA SA311195046GA SA311195181GA SA311195209GA SA311205592GA SA311205601GA SA311205615GA SA311215684GA SA311215687GA SA311215688GA SA311215692GA SA311215697GA SA311215718GA SA311215719GA SA311215721GA SA311215728GA SA311236106GA SA311184738GA SA311184826GA SA311184829GA SA311184833GA SA311184850GA SA311184864GA SA311184870GA SA311184875GA SA311194893GA SA311194899GA SA311194927GA SA311194938GA SA311194942GA SA311194980GA SA311194994GA SA311194995GA SA311195001GA SA311195006GA SA311195017GA SA311195021GA SA311195182GA SA311205345GA SA311205367GA SA311205580GA SA311205614GA SA311215802GA SA311215803GA SA311215805GA SA311215807GA SA311215808GA SA311215810GA SA311215811GA SA311215812GA SA311215814GA SA311215834GA SA311215837GA SA311215850GA SA311215851GA SA311215852GA SA311215854GA SA311215855GA SA311215856GA SA311215859GA SA311215860GA SA311225908GA SA311225911GA SA311225912GA SA311225945GA SA311287111GA SA311287179GA SA311287195GA SA311297327GA SA311297330GA SA311297335GA SA311297341GA SA311297362GA SA311307440GA SA311307446GA SA311307450GA SA311215713GA SA311236016GA SA311236032GA SA311236036GA SA311236043GA SA311236047GA SA311236236GA SA311246450GA SA311256804GA SA311266922GA SA310518623GA SA311039486GA SA311236023GA SA311236025GA SA311236050GA SA311236060GA SA311236363GA SA311297273GA SA311327878GA SA311317662GA SA311317671GA SA311317672GA SA311327807GA SA311327809GA SA311327824GA SA311327881GA SA311327882GA SA311039451GA SA311215736GA SA311225966GA SA311236015GA SA311236028GA SA311236228GA SA311297279GA SA311297334GA SA311236040GA SA311163601GA SA311163607GA SA311163638GA SA311163823GA SA311236369GA SA311266938GA SA310528965GA SA310529058GA SA311029363GA SA311039423GA SA311039424GA SA311039425GA SA311039441GA SA311039452GA SA311184536GA SA311184620GA SA311215682GA SA311215689GA SA311215703GA SA311215704GA SA311215707GA SA311215708GA SA311215709GA SA311215710GA SA311215711GA SA311215712GA SA311215716GA SA311215720GA SA311215722GA SA311215726GA SA311215730GA SA311215732GA SA311287190GA SA311297254GA SA311307410GA SA311307411GA SA311307412GA SA311307418GA SA311307421GA SA311307424GA SA311307427GA SA311307428GA SA311307437GA SA311307438GA SA311307441GA SA311307443GA SA311307448GA SA311307454GA SA311307512GA SA311307529GA SA311070517GA SA311070585GA SA311070586GA SA311070587GA SA311080606GA SA311080642GA SA311080658GA SA311080661GA SA311080703GA SA311080705GA SA311080706GA SA311080707GA SA311080709GA SA311080716GA SA311080722GA SA311090793GA SA311101191GA SA311101193GA SA311101197GA SA311101206GA SA311101210GA SA311101222GA SA311101231GA SA311101247GA SA311101249GA SA311101265GA SA311101266GA SA311101267GA SA311101268GA SA311101276GA SA311236185GA SA311236186GA SA311236281GA SA311236282GA SA311256822GA SA311256829GA SA311256840GA SA311266964GA SA311287021GA SA311287042GA SA311287046GA SA311287051GA SA311287057GA SA311287100GA SA311287103GA SA310454954GA SA311184724GA SA311184733GA SA311184786GA SA311184841GA SA311184842GA SA311184861GA SA311184873GA SA311184874GA SA311194932GA SA311194939GA SA311194944GA SA311194948GA SA311195184GA SA311205420GA SA311205423GA SA311205562GA SA311205587GA SA311205588GA SA311205604GA SA311205605GA SA311205608GA SA311205609GA SA311205611GA SA311205613GA SA311205616GA SA311215727GA SA311215806GA SA311215857GA SA311225892GA SA311225896GA SA311225902GA SA311225903GA SA311225904GA SA311225905GA SA311225919GA SA311225921GA SA311246485GA SA311317707GA SA311327884GA SA311327888GA SA311327889GA SA311327893GA SA311327895GA SA311327898GA SA311428162GA SA311458224GA SA311458241GA SA311468409GA SA311205475GA SA311236206GA SA311246639GA SA311246660GA SA311256805GA SA311256819GA SA311256847GA SA311256856GA SA311287012GA SA311287020GA SA311287023GA SA311287028GA SA311287033GA SA311287037GA SA311287038GA SA311287044GA SA311287045GA SA311287056GA SA311287058GA SA311287059GA SA311287061GA SA311287077GA SA311287097GA SA311297259GA SA311307419GA SA311307425GA SA311307436GA SA311307444GA SA311307445GA SA311307449GA SA311205315GA SA311225893GA SA311225942GA SA311225952GA SA311236029GA SA311236216GA SA311236355GA SA311236376GA SA311246452GA SA311246461GA SA311246470GA SA311287016GA SA311307456GA SA311327816GA SA311338000GA SA311468377GA SA311287025GA SA311287071GA SA311287078GA SA311287102GA SA311287154GA SA311287158GA SA311287161GA SA311458217GA SA311458221GA SA311458226GA SA311458229GA SA311458354GA SA311458374GA SA311468375GA SA311039419GA SA311184449GA SA311297366GA SA311428157GA SA311458223GA SA311458240GA SA311458244GA SA311458245GA SA311458253GA SA311458341GA SA311458344GA SA311458348GA SA311468402GA SA311468404GA SA311236231GA SA311246618GA SA311256727GA SA311256729GA SA311256739GA SA311256744GA SA311256747GA SA311256750GA SA311256761GA SA311287019GA SA311287093GA SA311287099GA SA311327875GA SA311337983GA SA311428017GA SA311458222GA SA311458237GA SA311458239GA SA311458350GA SA311468396GA SA311468397GA SA310529108GA SA311039413GA SA311039414GA SA311039469GA SA311039479GA SA311039484GA SA311060299GA SA311132560GA SA311142909GA SA311205472GA SA311236210GA SA311236249GA SA311236256GA SA311256726GA SA311256731GA SA311256732GA SA311256734GA SA311256737GA SA311256740GA SA311256756GA SA311256758GA SA311256759GA SA311256760GA SA311256763GA SA311448177GA SA311458345GA SA311458351GA SA311458358GA SA311458362GA SA311468380GA SA311468383GA SA311468388GA SA310496651GA SA310528875GA SA310529009GA SA311236088GA SA311236227GA SA311236271GA SA311236329GA SA311246383GA SA311246386GA SA311246388GA SA311246390GA SA311246392GA SA311246393GA SA311246397GA SA311246399GA SA311246412GA SA311246419GA SA311246425GA SA311246426GA SA311246429GA SA311246436GA SA311246440GA SA311246442GA SA311246444GA SA311246509GA SA311246586GA SA311246597GA SA311246603GA SA311246611GA SA311246620GA SA311246621GA SA311246638GA SA311246664GA SA311256748GA SA311317718GA SA311317736GA SA311327808GA SA311327963GA SA311428093GA SA311428095GA SA311428097GA SA311428098GA SA311428099GA SA311428100GA SA311428101GA SA311428102GA SA311428103GA SA311428104GA SA311428106GA SA311428107GA SA311428108GA SA311428109GA SA311428110GA SA311428111GA SA311428113GA SA311428114GA SA311428117GA SA311428118GA SA311428120GA SA311428122GA SA311428123GA SA311428126GA SA311428127GA SA311428132GA SA311428133GA SA311428135GA SA311428138GA SA311428140GA SA311428141GA SA311428143GA SA311428147GA SA311428149GA SA311428150GA SA311428152GA SA311428153GA SA311428155GA SA311428156GA SA311428160GA SA311428161GA SA311428163GA SA311428165GA SA311428166GA SA311428167GA SA311428168GA SA311428169GA SA311428170GA SA311448189GA SA311448190GA SA311458352GA SA311184858GA SA311256718GA SA311287022GA SA311287030GA SA311287053GA SA311287104GA SA311458272GA SA311458282GA SA311458292GA SA311039480GA SA311101208GA SA311215701GA SA311236348GA SA311236349GA SA311246463GA SA311246466GA SA311246493GA SA311246565GA SA311256723GA SA311256742GA SA311256776GA SA311256790GA SA311256831GA SA311266902GA SA311297293GA SA311297299GA SA311297373GA SA311297382GA SA311297399GA SA311297400GA SA311307423GA SA311307453GA SA311307501GA SA311307507GA SA311307582GA SA311307586GA SA311307588GA SA311307590GA SA311307592GA SA311307593GA SA311307596GA SA311307597GA SA311317648GA SA311317651GA SA311317733GA SA311327814GA SA311327830GA SA311327887GA SA311327891GA SA311327945GA SA311327946GA SA311327947GA SA311428092GA SA311428116GA SA311428128GA SA311428146GA SA311428151GA SA311458258GA SA311458260GA SA311468395GA SA310413561GA SA310518705GA SA310518725GA SA310528896GA SA310528984GA SA311029366GA SA311039405GA SA311039412GA SA311039417GA SA311039418GA SA311039421GA SA311039422GA SA311039430GA SA311039439GA SA311256785GA SA311307500GA SA311307502GA SA311307503GA SA311307506GA SA311307508GA SA311307510GA SA311307511GA SA311307581GA SA311307585GA SA311307587GA SA311307589GA SA311317643GA SA311317647GA SA311317652GA SA311317657GA SA311317663GA SA311317667GA SA311317703GA SA311317704GA SA311317708GA SA311317716GA SA311317720GA SA311317724GA SA311317729GA SA311317734GA SA311317735GA SA311317740GA SA311327810GA SA311327827GA SA311327866GA SA311327879GA SA311327896GA SA311327897GA SA311327935GA SA311327939GA SA311327941GA SA311327944GA SA311327948GA SA311327949GA SA311327952GA SA311327954GA SA311327955GA SA311327957GA SA311327961GA SA311327965GA SA311327967GA SA310393097GA SA310528844GA SA310528917GA SA310528920GA SA310529107GA SA311060076GA SA311060102GA SA311246622GA SA311246650GA SA311246651GA SA311246655GA SA311246662GA SA311246663GA SA311246670GA SA311256733GA SA311256738GA SA311256752GA SA311317730GA SA311317732GA SA310393243GA SA310496370GA SA310528845GA SA310528902GA SA310528977GA SA310529116GA SA310529120GA SA311039468GA SA311195238GA SA311205569GA SA311236061GA SA311236182GA SA311236212GA SA311236222GA SA311236238GA SA311236247GA SA311236346GA SA311236356GA SA311236360GA SA311236361GA SA311236371GA SA311236380GA SA311246472GA SA311246492GA SA311246497GA SA311246500GA SA311246515GA SA311246595GA SA311246598GA SA311246631GA SA311256728GA SA311256757GA SA311256762GA SA311256807GA SA311256826GA SA311256870GA SA311287024GA SA311287027GA SA311287029GA SA311287036GA SA311287048GA SA311287050GA SA311287072GA SA311287082GA SA311287084GA SA311287085GA SA311287087GA SA311287101GA SA311287151GA SA311297354GA SA311307544GA SA311317600GA SA311327817GA SA311327819GA SA311327829GA SA311327943GA SA311508588GA SA311236347GA SA311246462GA SA311246583GA SA311246585GA SA311246587GA SA311246594GA SA311246600GA SA311246600GAMKE SA311246602GA SA311246615GA SA311246619GA SA311246625GA SA311246628GA SA311246632GA SA311246642GA SA311246645GA SA311246653GA SA311246656GA SA311246666GA SA311246674GA SA311246675GA SA311246677GA SA311287026GA SA311287034GA SA311287108GA SA311297293GA SA311327820GA SA311327821GA SA311327832GA SA311468407GA SA310444287GA SA310539322GA SA311039410GA SA311039477GA SA311163427GA SA311184639GA SA311205476GA SA311215699GA SA311215705GA SA311215706GA SA311246404GA SA311266872GA SA311266911GA SA311287136GA SA311307514GA SA311307531GA SA311327956GA SA311327960GA SA311327964GA SA311327968GA SA311327970GA SA311448188GA SA311458230GA SA311508572GA SA311508589GA SA311246480GA SA311246588GA SA311246606GA SA311246624GA SA311246640GA SA311246649GA SA311246652GA SA311246654GA SA311246658GA SA311246659GA SA311246665GA SA311246669GA SA311246673GA SA311246681GA SA311256725GA SA311256730GA SA311256745GA SA311256823GA SA311256825GA SA311256832GA SA311307561GA SA311327909GA SA311327933GA SA311508561GA SA311508590GA SA311508591GA SA311039385GA SA311039429GA SA311163563GA SA311174077GA SA311236251GA SA311236372GA SA311246456GA SA311246460GA SA311246465GA SA311246468GA SA311246483GA SA311246484GA SA311246490GA SA311246499GA SA311246502GA SA311246506GA SA311246507GA SA311246513GA SA311246591GA SA311246592GA SA311246593GA SA311246596GA SA311246599GA SA311246601GA SA311246604GA SA311246605GA SA311246607GA SA311246608GA SA311246609GA SA311246613GA SA311246614GA SA311246616GA SA311246617GA SA311246626GA SA311246627GA SA311246633GA SA311246634GA SA311246635GA SA311246637GA SA311246641GA SA311246643GA SA311246644GA SA311246646GA SA311246647GA SA311246648GA SA311246672GA SA311246676GA SA311256809GA SA311256814GA SA311256815GA SA311256830GA SA311256833GA SA311256834GAMKE SA311256835GA SA311256838GA SA311256839GA SA311256841GA SA311256855GA SA311256868GA SA311287013GA SA311287015GA SA311287031GA SA311287039GA SA311287040GA SA311287049GA SA311287054GA SA311287055GA SA311287075GA SA311287079GA SA311287083GA SA311287088GA SA311287094GA SA311297223GA SA311297228GA SA311297250GA SA311297292GA SA311297298GA SA311297301GA SA311297325GA SA311297340GA SA311297368GA SA311297370GA SA311297371GA SA311297375GA SA311297376GA SA311297377GA SA311297378GA SA311297379GA SA311297380GA SA311297381GA SA311297383GA SA311297384GA SA311297385GA SA311297386GA SA311297387GA SA311297393GA SA311297394GA SA311297395GA SA311297396GA SA311297397GA SA311297398GA SA311297402GA SA311297403GA SA311297404GA SA311297405GA SA311297407GA SA311307463GA SA311307464GA SA311307466GA SA311307471GA
SA311307472GA SA311307475GA SA311307479GA SA311307481GA SA311307482GA SA311307484GA SA311307491GA SA311307492GA SA311307493GA SA311307494GA SA311307495GA SA311307496GA SA311307499GA SA311307541GA SA311307542GA SA311307543GA SA311307545GA SA311307548GA SA311307549GA SA311307550GA SA311307551GA SA311307552GA SA311307555GA SA311307556GA SA311307557GA SA311307559GA SA311307560GA SA311307563GA SA311307564GA SA311307565GA SA311307566GA SA311307567GA SA311307571GA SA311307572GA SA311307575GA SA311307577GA SA311307578GA SA311307579GA SA311317598GA SA311317606GA SA311317608GA SA311317613GA SA311317618GA SA311317620GA SA311317621GA SA311317627GA SA311317628GA SA311317635GA SA311317638GA SA311317640GA SA311317690GA SA311327837GA SA311327840GA SA311327841GA SA311327846GA SA311327851GA SA311327901GA SA311327902GA SA311327903GA SA311327904GA SA311327906GA SA311327907GA SA311327908GA SA311327912GA SA311327916GA SA311327917GA SA311327918GA SA311327921GA SA311327922GA SA311327923GA SA311327924GA SA311327926GA SA311327928GA SA311327929GA SA311327930GA SA311327931GA SA311337971GA SA311337972GA SA311337974GA SA311337977GA SA311337979GA SA311337982GA SA311337984GA SA311337987GA SA311337988GA SA311337990GA SA311337991GA SA311337992GA SA311337994GA SA311337995GA SA311338001GA SA311338002GA SA311338005GA SA311508566GA SA311508593GA SA311236184GA SA311317755GA SA311317772GA SA311317773GA SA311327927GA SA311327937GA SA311327950GA SA311327951GA SA311508640GA SA311508643GA SA311508646GA SA311508648GA SA311508650GA SA311508652GA SA311508655GA SA311508656GA SA311508664GA SA311508665GA SA311508668GA SA311508673GA SA311508674GA SA311508675GA SA311236367GA SA311256743GA SA311256753GA SA311468394GA SA311508560GA SA311508563GA SA311508564GA SA311508565GA SA311508567GA SA311508568GA SA311508569GA SA311508570GA SA311508573GA SA311508574GA SA311508575GA SA311508578GA SA311508581GA SA311508582GA SA311508584GA SA311508585GA SA311508586GA SA311508592GA SA311508596GA SA311508597GA SA311508598GA SA311508599GA SA311508614GA SA311508623GA SA311508624GA SA311508663GA SA311518703GA SA311518707GA SA311518710GA SA311518712GA SA311518713GA SA311518714GA SA311518715GA SA311518716GA SA311518718GA SA311246474GA SA311287065GA SA311287069GA SA311287070GA SA311287073GA SA311287089GA SA311287150GA SA311297219GA SA311317614GA SA311317693GA SA311317765GA SA311518683GA SA311518695GA SA311518701GA SA311518702GA SA311518727GA SA311518736GA SA311287047GA SA311287060GA SA311287064GA SA311287066GA SA311287090GA SA311287153GA SA311307476GA SA311307574GA SA311327843GA SA311518699GA SA311195060GA SA311195078GA SA311195109GA SA311195121GA SA311195127GA SA311287159GA SA311287165GA SA311287218GA SA311297221GA SA311297222GA SA311297225GA SA311297226GA SA311297227GA SA311297229GA SA311297232GA SA311297234GA SA311297236GA SA311297238GA SA311297243GA SA311297246GA SA311297249GA SA311518681GA SA311518686GA SA311518704GA SA311518705GA SA311518706GA SA311518709GA SA311518719GA SA311518721GA SA311518722GA SA311518723GA SA311518724GA SA311518730GA SA311518731GA SA311518732GA SA311518734GA SA311528737GA SA311528738GA SA311528739GA SA311528741GA SA311528742GA SA311528743GA SA311528744GA SA311039478GA SA311039490GA SA311132545GA SA311132600GA SA311184784GA SA311184823GA SA311184851GA SA311184865GA SA311236309GA SA311287124GA SA311297220GA SA311297231GA SA311297237GA SA311297239GA SA311297240GA SA311297242GA SA311297245GA SA311297253GA SA311297275GA SA311297302GA SA311297303GA SA311297306GA SA311297314GA SA311297316GA SA311297326GA SA311297367GA SA311297369GA SA311297372GA SA311297374GA SA311297388GA SA311297389GA SA311297391GA SA311297392GA SA311297401GA SA311297406GA SA311297408GA SA311297409GA SA311307468GA SA311317599GA SA311317605GA SA311317614GA SA311317619GA SA311317679GA SA311317681GA SA311317697GA SA311317699GA SA311317700GA SA311317701GA SA311317702GA SA311428130GA SA311428136GA SA311428137GA SA311428139GA SA311428164GA SA311428171GA SA311508576GA SA311518687GA SA311518689GA SA311518693GA SA311518697GA SA311518700GA SA311287074GA SA311287080GA SA311184527GA SA311195069GA SA311236303GA SA311236307GA SA311236314GA SA311236328GA SA311236339GA SA311236342GA SA311327873GA SA311327899GA SA311518708GA SA312018746GA SA312018754GA SA312018755GA SA311508654GA SA312018851GA SA312018856GA SA312018857GA SA312018861GA SA312018862GA SA312018863GA SA312018865GA SA312018867GA SA312018868GA SA312018869GA SA312018870GA SA312018871GA SA311287076GA SA311297304GA SA311297311GA SA311297324GA SA311317754GA SA311488477GA SA311488478GA SA311297322GA SA312018847GA SA311317692GA SA311246678GA SA311508647GA SA311518735GA SA312018841GA SA312018844GA SA312018845GA SA312018846GA SA312018848GA SA312018866GA SA312018897GA SA312028910GA SA312028914GAMKE SA312028915GA SA312028916GA SA312028917GA SA312028919GA SA312028921GA SA312028922GA SA312028926GA SA312028928GA SA312028929GA SA312028932GA SA312028933GA SA312028936GA SA312018850GA SA312018852GA SA312018887GA SA312018898GA SA312018902GA SA312029014GA SA312029069GA SA312028979GA SA312028991GA SA312029020GA SA312029023GA SA312029034GA SA311317685GA SA311458252GA SA311518717GA SA311528740GA SA312028948GA SA312028982GA SA312028984GA SA312028986GA SA312028987GA SA312028988GA SA312028989GA SA312028992GA SA312028993GA SA312028994GA SA312028995GA SA312029001GA SA312029005GA SA312029007GA SA312029009GA SA312029011GA SA312029026GA SA312029039GA SA312029062GA SA312029065GA SA311508636GA SA311518690GA SA311518720GA SA312028966GA SA312028997GA SA312028998GA SA312029003GA SA312029004GA SA312029006GA SA312029010GA SA312029012GA SA312029017GA SA312029028GA SA312029032GA SA312029035GA SA312029036GA SA312029038GA SA312029040GA SA312029043GA SA312029045GA SA312029047GA SA312029050GA SA312029052GA SA312029055GA SA312029058GA SA312029063GA SA312029066GA SA312039084GA SA312039086GA SA312039093GA SA311468433GA SA311468445GA SA311468446GA SA311488468GA SA311488469GA SA311488470GA SA311488485GA SA312018808GA SA312029057GA SA311236218GA SA311297294GA SA311297390GA SA311317676GA SA311317678GA SA311317680GA SA311317683GA SA311317687GA SA311317696GA SA311327835GA SA311327838GA SA311327845GA SA311327849GA SA311327850GA SA311327855GA SA311327858GA SA311327861GA SA311327863GA SA311327864GA SA311337998GA SA311468420GA SA311468428GA SA311468438GA SA311488466GA SA311488467GA SA311317602GA SA311317615GA SA311317641GA SA311458367GA SA312039140GA SA311327842GA SA311327853GA SA311327857GA SA311428058GA SA311428071GA SA311468424GA SA311468432GA SA311468443GA SA311488456GA SA311488458GA SA311488462GA SA311488464GA SA311488471GA SA311488472GA SA311488473GA SA311488475GA SA311488476GA SA311488480GA SA311488482GA SA311488484GA SA311488486GA SA311488487GA SA311488488GA SA311163936GA SA311174166GA SA311163937GA SA311174181GA SA311184711GA SA311195177GA SA311195218GA SA311195223GA SA311195236GA SA311195243GA SA311195245GA SA311236013GA SA311256821GA SA311266947GA SA311287063GA SA311287067GA SA311297321GA SA311317629GA SA311327839GA SA311428006GA SA311428007GA SA311428008GA SA311428009GA SA311428010GA SA311428014GA SA311428016GA SA311428019GA SA311428021GA SA311428022GA SA311428024GA SA311428027GA SA311428029GA SA311428030GA SA311428034GA SA311428036GA SA311428037GA SA311428038GA SA311428039GA SA311428040GA SA311428041GA SA311428042GA SA311428044GA SA311428045GA SA311428046GA SA311428047GA SA311428048GA SA311428049GA SA311428051GA SA311428052GA SA311428053GA SA311428054GA SA311428067GA SA311428074GA SA311428075GA SA311428076GA SA311428080GA SA311428082GA SA311428083GA SA311428085GA SA311428087GA SA311428089GA SA311458369GA SA311468414GA SA311468419GA SA311488459GA SA312028980GA SA312039083GA SA312039087GA SA312039089GA SA312039102GA SA312039103GA SA311428025GA SA311428055GA SA311458370GA SA311458372GA SA311458373GA SA311468408GA SA311468448GA SA311488474GA SA312028978GA SA312029002GA SA312029031GA SA312039076GA SA312039077GA SA312039080GA SA312039081GA SA312039085GA SA312039090GA SA312039094GA SA312039098GA SA312039101GA SA312039107GA SA312039131GA SA312039133GA SA312039136GA SA312039179GA SA312039205GA SA312039220GA SA312039221GA SA312049241GA SA312049244GA SA312049246GA SA312049247GA SA312049248GA SA312049250GA SA312049251GA SA312049252GA SA312049253GA SA312049256GA SA312049258GA SA312049262GA SA312049264GA SA312049268GA SA311468415GA SA311488465GA SA311488481GA SA311458347GA SA311468384GA SA312028950GA SA312039108GA SA312039109GA SA312039111GA SA312039112GA SA312039113GA SA312039118GA SA312039120GA SA312039122GA SA312039128GA SA312039129GA SA312039142GA SA312039143GA SA312039144GA SA312039149GA SA312039151GA SA312039152GA SA312039154GA SA312039157GA SA312039158GA SA312039159GA SA312039162GA SA312039164GA SA312039168GA SA312039180GA SA312039223GA SA312039225GA SA312039228GA SA312039230GA SA312039231GA SA312039232GA SA312049235GA SA312049236GA SA312049238GA SA312049249GA SA312049254GA SA312049255GA SA312049259GA SA312049261GA SA312049263GA SA312049267GA SA312049272GA SA312049276GA SA312049287GA SA312049293GA SA312049299GA SA312049370GA SA312049378GA SA312049391GA SA312049396GA SA312049400GA SA312059507GA SA312059510GA SA312059512GA SA312059522GA SA312059525GA SA311317623GA SA311317637GA SA311317639GA SA311458249GA SA311508600GA SA312029024GA SA312029027GA SA312029033GA SA312029060GA SA312039078GA SA312039082GA SA312039092GA SA312039097GA SA312039100GA SA312039105GA SA312039106GA SA312039114GA SA312039117GA SA312039125GA SA312039130GA SA312039137GA SA312049242GA SA312049243GA SA312049260GA SA312049265GA SA311458333GA SA311488454GA SA311488455GA SA311488460GA SA311488479GA SA312018809GA SA312018810GA SA312018811GA SA312018812GA SA312018813GA SA312018814GA SA312018815GA SA312018817GA SA312018819GA SA312018820GA SA312018822GA SA312018823GA SA312018824GA SA312018826GA SA312018827GA SA312018828GA SA312018830GA SA312018831GA SA312018835GA SA312018836GA SA312018837GA SA312018839GA SA312018840GA SA312018843GA SA312018881GA SA312018886GA SA312018889GA SA312018893GA SA312018895GA SA312018904GA SA312029016GA SA312029021GA SA312029022GA SA312029025GA SA312029030GA SA312029046GA SA312029054GA SA312029056GA SA312029059GA SA312029061GA SA312029064GA SA312039190GA SA312049270GA SA312049271GA SA312049279GA SA312049281GA SA312049289GA SA312049292GA SA312049294GA SA312049296GA SA312049297GA SA312049298GA SA312049300GA SA312049302GA SA312049305GA SA312049307GA SA312049309GA SA312049312GA SA312049315GA SA312049316GA SA312049317GA SA312049426GA SA312049432GA SA312049459GA SA312039196GA SA312049342GA SA312049371GA SA312049390GA SA312049430GA SA312049454GA SA312049468GA SA312059471GA SA312059528GA SA312059530GA SA312059533GA SA312059534GA SA312049372GA SA312049373GA SA312049374GA SA312049377GA SA312049383GA SA312049385GA SA312049388GA SA312049389GA SA312049392GA SA312049393GA SA312049395GA SA312049399GA SA312049405GA SA312049407GA SA312049408GA SA312049411GA SA312049414GA SA312049418GA SA312049419GA SA312049437GA SA312059500GA SA312059502GA SA312059503GA SA312059515GA SA312059516GA SA312059520GA SA312049318GA SA312049321GA SA312049329GA SA312049331GA SA312049332GA SA312049339GA SA312049341GA SA312049346GA SA312049348GA SA312049350GA SA312049353GA SA312049354GA SA312049359GA SA312049362GA SA312049363GA SA312049423GA SA312049436GA SA312049449GA SA312049452GA SA312059489GA SA312079899GA SA312079900GA SA312079902GA SA312079918GA SA312079919GA SA312079922GA SA312079924GA SA312079938GA SA312049275GA SA312049386GA SA312069818GA SA312069821GA SA312069822GA SA312069837GA SA312069850GA SA312069855GA SA312069857GA SA312069859GA SA312069860GA SA312069861GA SA312069862GA SA312069863GA SA312069864GA SA312069865GA SA312069868GA SA312069869GA SA312069881GA SA312049278GA SA312049283GA SA312049288GA SA312049368GA SA312049387GA SA312049394GA SA312049401GA SA312049402GA SA312049403GA SA312049409GA SA312049410GA SA312049416GA SA312059513GA SA312059514GA SA312059517GA SA312059518GA SA312059519GA SA312059521GA SA312059523GA SA312059524GA SA312059526GA SA312059638GA SA312069646GA SA312069652GA SA312069654GA SA312069657GA SA312069659GA SA312069834GA SA312069875GA SA312079887GA SA312079908GA SA312079915GA SA312079917GA SA312079926GA SA312079927GA SA312079928GA SA312079933GA SA312079934GA SA312079935GA SA312079936GA SA312079937GA SA312079944GA SA312079948GA SA312079966GA SA312079970GA SA312079972GA SA312079974GA SA312079975GA SA312079992GA SA312079995GA SA312079998GA SA312059609GA SA312059619GA SA312059620GA SA312070006GA SA312049347GA SA312049365GA SA312049427GA SA312049429GA SA312049433GA SA312049442GA SA312049443GA SA312049447GA SA312049461GA SA312049466GA SA312059480GA SA312059483GA SA312059494GA SA312059496GA SA312079940GA SA312079941GA SA312079950GA SA312079953GA SA312079956GA SA312079958GA SA312079961GA SA312079977GA SA312079980GA SA312079987GA SA312079988GA SA312079994GA SA312079996GA SA311458371GA SA312018875GA SA312039123GA SA312039126GA SA312039134GA SA312039138GA SA312039147GA SA312039148GA SA312039153GA SA312039161GA SA312039165GA SA312039166GA SA312039169GA SA312039172GA SA312039173GA SA312039174GA SA312039175GA SA312039177GA SA312039183GA SA312039193GA SA312059474GA SA312059537GA SA312059539GA SA312059543GA SA312059552GA SA312059553GA SA312059559GA SA312059562GA SA312059563GA SA312059612GA SA312059615GA SA312059631GA SA312059636GA SA312039091GA SA312039127GA SA312039163GA SA312039181GA SA312039185GA SA312039194GA SA312039197GA SA312039198GA SA312039199GA SA312039201GA SA312039207GA SA312039211GA SA312069698GA SA312080398GA SA312080401GA SA312069715GA SA312080402GA SA312080467GA SA312080512GA SA312080523GA SA312059608GA SA312059610GA SA312059611GA SA312059613GA SA312059614GA SA312059616GA SA312059617GA SA312059618GA SA312059625GA SA312059632GA SA312059633GA SA312059635GA SA312059637GA SA312059639GA SA312069640GA SA312069643GA SA312069656GA SA312069663GA SA312069667GA SA312069668GA SA312069669GA SA312069672GA SA312069675GA SA312069688GA SA312069707GA SA312069720GA SA312080397GA SA312080528GA SA311327808GA SA312049324GA SA312049336GA SA312049361GA SA312059557GA SA312069648GA SA312069749GA SA312069760GA SA312080386GA SA312080390GA SA312080416GA SA312080417GA SA312080419GA SA312080422GA SA312080423GA SA312080425GA SA312080432GA SA312080433GA SA312080465GA SA312080488GA SA312080494GA SA312080496GA SA312080498GA SA312080500GA SA312080505GA SA312080506GA SA312080509GA SA312080514GA SA312080515GA SA312080526GA SA312080532GA SA312080497GA SA312080510GA SA312059511GA SA312070024GA SA312079943GA SA312079945GA SA312079963GA SA312079985GA SA312080462GA SA312080420GA SA312080489GA SA312080531GA SA312080533GA SA312080539GA SA312080555GA SA312080556GA SA312080558GA SA312080561GA SA312080571GA SA312080577GA SA312080584GA SA312080587GA SA312080590GA SA312080596GA SA312080604GA SA312080608GA SA312080617GA SA312080622GA SA312080623GA SA312080624GA SA312080625GA SA312080626GA SA312080631GA SA312080645GA SA312080673GA SA312080674GA SA312080679GA SA312080686GA SA312080688GA SA312080690GA SA312080692GA SA312049358GA SA312049360GA SA312049428GA SA312059475GA SA312059481GA SA312059535GA SA312059536GA SA312059538GA SA312059540GA SA312059542GA SA312059547GA SA312059549GA SA312059550GA SA312059551GA SA312059554GA SA312059555GA SA312059556GA SA312059558GA SA312059561GA SA312059564GA SA312059566GA SA312059569GA SA312080389GA SA312080403GA SA312080410GA SA312080415GA SA312080424GA SA312080427GA SA312080429GA SA312080431GA SA312080434GA SA312080463GA SA312080464GA SA312080466GA SA312080471GA SA312080472GA SA312080473GA SA312080476GA SA312080477GA SA312080478GA SA312080479GA SA312080486GA SA312080487GA SA312080508GA SA312080513GA SA312080517GA SA312080520GA SA312080522GA SA312080525GA SA312080535GA SA312080536GA SA312080537GA SA312080538GA SA312080540GA SA312080541GA SA312080543GA SA312080545GA SA312080548GA SA312080549GA SA312080552GA SA312080559GA SA312080560GA SA312080566GA SA312080573GA SA312080574GA SA312080575GA SA312080579GA SA312080594GA SA312070002GA SA312070023GA SA312070055GA SA312070058GA SA312070060GA SA312070085GA SA312080388GA SA312080396GA SA312080399GA SA312080635GA SA312070005GA SA312070020GA SA312070032GA SA312070033GA SA312070056GA SA312070061GA SA312070090GA SA312070092GA SA312070096GA SA312070111GA SA312070136GA SA312080534GA SA312080649GA SA312080656GA SA312080659GA SA312080661GA SA312080662GA SA312080667GA SA312080678GA SA312080695GA SA312080701GA SA312080706GA SA312080708GA SA312080719GA SA312080744GA SA312080763GA SA312080769GA SA312080779GA SA312080783GA SA312080791GA SA312080792GA SA312080797GA SA312080802GA SA312080814GA SA312080840GA SA312080903GA SA312080909GA SA312049438GA SA312059472GA SA312070086GA SA312070097GA SA312070221GA SA312070232GA SA312080663GA SA312080677GA SA312080707GA SA312080724GA SA312080761GA SA312080765GA SA312080775GA SA312069693GA SA312069694GA SA312069717GA SA312069725GA SA312069787GA SA312070017GA SA312070053GA SA312070082GA SA312070083GA SA312070087GA SA312070091GA SA312070093GA SA312070094GA SA312080670GA SA312080830GA SA312080848GA SA312080852GA SA312080871GA SA312080873GA SA312080885GA SA312080900GA SA312080901GA SA312080905GA SA312070062GA SA312070084GA SA312070095GA SA312080409GA SA312080412GA SA312080653GA SA312080691GA SA312049320GA SA312049326GA SA312049344GA SA312049434GA SA312059479GA SA312059486GA SA312059490GA SA312059493GA SA312070219GA SA312070004GA SA312070007GA SA312070008GA SA312070009GA SA312070010GA SA312070014GA SA312070018GA SA312070019GA SA312070026GA SA312070027GA SA312070035GA SA312070037GA SA312070038GA SA312070040GA SA312079979GA SA312079982GA SA312079991GA SA312080550GA SA312080591GA SA312080629GA SA312080648GA SA312080655GA SA312080683GA SA312080700GA SA312069713GA SA312070043GA SA312070069GA SA312070074GA SA312070076GA SA312070080GA SA312070081GA SA312070124GA SA312070132GA SA312070135GA SA312080542GA SA312080610GA SA312080666GA SA312080689GA SA312080693GA SA312080720GA SA312080729GA SA312080731GA SA312080749GA SA312080751GA SA312080781GA SA312080800GA SA312080803GA SA312080824GA SA312080826GA SA312080833GA SA312080872GA SA312080874GA SA312070001GA SA312070247GA SA312070294GA SA312080570GA SA312080681GA SA312080817GA SA312080819GA SA312080860GA SA312080876GA SA312080884GA SA312070064GA SA312070079GA SA312070100GA SA312070106GA SA312070110GA SA312070115GA SA312070123GA SA312070155GA SA312070163GA SA312070175GA SA312070182GA SA312070184GA SA312070191GA SA312070216GA SA312070220GA SA312070258GA SA312070259GA SA312070262GA SA312070264GA SA312070266GA SA312070269GA SA312070272GA SA312070275GA SA312070276GA SA312070278GA SA312070279GA SA312070281GA SA312070285GA SA312070288GA SA312070290GA SA312070314GA SA312070317GAMKE SA312080563GA SA312080564GA SA312080597GA SA312080612GA SA312080647GA SA312080711GA SA312080717GA SA312080721GA SA312080727GA SA312080753GA SA312080764GA SA312080796GA SA312080798GA SA312080804GA SA312080806GA SA312080808GA SA312080810GA SA312080811GA SA312080813GA SA312080878GA SA312080882GA SA312080894GA SA312070187GA SA312070193GA SA312070195GA SA312070200GA SA312070209GA SA312070280GA SA312070287GA SA312070291GA SA312070318GA SA312070321GA SA312080684GA SA312080716GA SA312080725GA SA312080754GA SA312080758GA SA312080771GA SA312080777GA SA312080827GA SA312080834GA SA312080839GA SA312080858GA SA312080892GA SA312080893GA SA312049280GA SA312049369GA SA312070034GA SA312070046GA SA312070048GA SA312070052GA SA312070054GA SA312070070GA SA312070071GA SA312070073GA SA312070101GA SA312070102GA SA312070107GA SA312070109GA SA312070113GA SA312070114GA SA312070116GA SA312070120GA SA312070126GA SA312070133GA SA312070151GA SA312079952GA SA312079957GA SA312079959GA SA312079962GA SA312079967GA SA312079968GA SA312079989GA SA312080426GA SA312080646GA SA312080704GA SA312080714GA SA312080759GA SA312080794GA SA312080820GA SA312080845GA SA312080877GA SA312080881GA SA312080890GA SA312080912GA SA312080916GA SA312091019GA SA312091020GA SA312091021GA SA312091022GA SA312091027GA SA312091033GA SA312091042GA SA312091053GA SA312070044GA SA312070065GA SA312070197GA SA312070204GA SA312070211GA SA312070231GA SA312070243GA SA312070248GA SA312070250GA SA312070253GA SA312070271GA SA312070283GA SA312070307GA SA312070308GA SA312070320GA SA312080529GA SA312080544GA SA312080553GA SA312080554GA SA312080598GA SA312080630GA SA312080705GA SA312080713GA SA312080748GA SA312080770GA SA312080787GA SA312080801GA SA312080805GA SA312080809GA SA312080812GA SA312080837GA SA312080886GA SA312080899GA SA312080914GA SA312091018GA SA312091024GA SA312091025GA SA312091028GA SA312091030GA SA312091031GA SA312091032GA SA312091034GA SA312091035GA SA312091041GA SA312091044GA SA312091045GA SA312091048GA SA312091058GA SA312091060GA SA312091062GA SA312091067GA SA312091070GA SA312091071GA SA312091074GA SA312091075GA SA312091076GA SA312091082GA SA312091109GA SA312091116GA SA312049356GA SA312049441GA SA312049446GA SA312049451GA SA312049453GA SA312049457GA SA312049462GA SA312049463GA SA312049467GA SA312049469GA SA312059473GA SA312069642GA SA312069651GA SA312069653GA SA312069661GA SA312069664GA SA312069665GA SA312070042GA SA312070049GA SA312070059GA SA312070067GA SA312070075GA SA312070112GA SA312070131GA SA312070142GA SA312070150GA SA312070152GA SA312070156GA SA312070158GA SA312070199GA SA312070217GA SA312070226GA SA312070324GA SA312070347GA SA312070354GA SA312070361GA SA312070373GA SA312080668GA SA312080671GA SA312080675GA SA312080728GA SA312080732GA SA312091017GA SA312091023GA SA312091026GA SA312091037GA SA312091038GA SA312091039GA SA312091040GA SA312091050GA SA312091051GA SA312091052GA SA312091054GA SA312091057GA SA312091061GA SA312091064GA SA312091069GA SA312091077GA SA312091079GA SA312091083GA SA312070030GA SA312070057GA SA312070068GA SA312070139GA SA312070143GA SA312070164GA SA312070198GA SA312070201GA SA312070215GA SA312079942GA SA312091043GA SA312091059GA SA312069804GA SA312070039GA SA312070130GA SA312070141GA SA312070145GA SA312070146GA SA312070147GA SA312070148GA SA312070149GA SA312070153GA SA312070154GA SA312070157GA SA312070159GA SA312070161GA SA312070162GA SA312070165GA SA312070166GA SA312070167GA SA312070168GA SA312070169GA SA312070170GA SA312070171GA SA312070172GA SA312070173GA SA312070176GA SA312070177GA SA312070178GA SA312070179GA SA312070183GA SA312070188GA SA312070190GA SA312070196GA SA312070202GA SA312070206GA SA312070208GA SA312070210GA SA312070212GA SA312070225GA SA312070227GA SA312070230GA SA312070233GA SA312070236GA SA312070237GA SA312070241GA SA312070242GA SA312070244GA SA312070245GA SA312070246GA SA312070249GA SA312070254GA SA312070255GA SA312070256GA SA312070257GA SA312070260GA SA312070261GA SA312070263GA SA312070265GA SA312070267GA SA312070270GA SA312070273GA SA312070274GA SA312070282GA SA312070284GA SA312070286GA SA312070289GA SA312070292GA SA312070293GA SA312070295GA SA312070296GA SA312070297GA SA312070298GA SA312070299GA SA312070300GA SA312070301GA SA312070303GA SA312070305GA SA312070306GA SA312070309GA SA312070310GA SA312070311GA SA312070312GA SA312070313GA SA312070315GA SA312070319GA SA312070322GA SA312070323GA SA312070325GA SA312070326GA SA312070327GA SA312070328GA SA312070329GA SA312070330GA SA312070332GA SA312070333GA SA312070334GA SA312070335GA SA312070336GA SA312070338GA SA312070339GA SA312070342GA SA312070343GA SA312070346GA SA312070349GA SA312070350GA SA312070353GA SA312070355GA SA312070356GA SA312070357GA SA312070359GA SA312070363GA SA312070364GA SA312070365GA SA312070366GA SA312070367GA SA312070368GA SA312070369GA SA312070370GA SA312070372GA SA312070374GA SA312070375GA SA312070376GA SA312070377GA SA312070378GA SA312070380GA SA312070381GA SA312070384GA SA312070385GA SA312090917GA SA312090921GA SA312090934GA SA312090942GA SA312090948GA SA312090966GA SA312090967GA SA312090969GA SA312090970GA SA312090972GA SA312090974GA SA312090981GA SA312090982GA SA312090983GA SA312090985GA SA312090987GA SA312090988GA SA312090989GA SA312090990GA SA312090993GA SA312090994GA SA312090995GA SA312090996GA SA312090997GA SA312090998GA SA312090999GA SA312091000GA SA312091001GA SA312091002GA SA312091003GA SA312091004GA SA312091006GA SA312091011GA SA312091046GA SA312091066GA SA312091094GA SA312091095GA SA312091104GA SA312091108GA SA312070304GA SA312070331GA SA312070351GA SA312070371GA SA312070382GA SA312091007GA SA312101406GA SA312091085GA SA312091113GA SA312069743GA SA312101453GA SA312101455GA SA312111529GA SA312091036GA SA312091063GA SA312091072GA SA312091078GA SA312091081GA SA312091084GA SA312091086GA SA312091087GA SA312091088GA SA312091090GA SA312091091GA SA312091096GA SA312091110GA SA312091111GA SA312080491GA SA312101389GA SA312101448GA SA312111510GA SA312111517GA SA312111518GA SA312111558GA SA312111560GA SA312111566GA SA312111572GA SA312111582GA SA312111584GA SA312111587GA SA312111591GA SA312069754GA SA312069766GA SA312090927GA SA312091055GA SA312091073GA SA312091092GA SA312091103GA SA312091107GA SA312091245GA SA312101446GA SA312101449GA SA312101450GA SA312101452GA SA312111521GA SA312111540GA SA312111549GA SA312111552GA SA312111565GA SA312111569GA SA312111575GA SA312111580GA SA312111585GA SA312111586GA SA312111588GA SA312111590GA SA312069678GA SA312069734GA SA312069742GA SA312069755GA SA312069761GA SA312069767GA SA312091080GA SA312091114GA SA312101445GA SA312101454GA SA312101456GA SA312101457GA SA312111546GA SA312049237GA SA312111524GA SA312111536GA SA312111567GA SA312111523GA SA312111528GA SA312111545GA SA312111561GA SA312111573GA SA312121753GA SA312091190GA SA312091202GA SA312091219GA SA312091244GA SA312091246GA SA312111512GA SA312111514GA SA312111519GA SA312111522GA SA312111525GA SA312111530GA SA312111531GA SA312111533GA SA312111534GA SA312111535GA SA312111537GA SA312111538GA SA312111541GA SA312111542GA SA312111543GA SA312111544GA SA312111548GA SA312111554GA SA312111556GA SA312111557GA SA312111559GA SA312111563GA SA312111564GA SA312111568GA SA312111571GA SA312111577GA SA312111578GA SA312111581GA SA312111513GA SA312069775GA SA312069779GA SA312069790GA SA312069792GA SA312069793GA SA312069794GA SA312069795GA SA312069796GA SA312069799GA SA312069801GA SA312069802GA SA312069803GA SA312069810GA SA312069823GA SA312069844GA SA312069846GA SA312069849GA SA312090978GA SA312091125GA SA312091145GA SA312091146GA SA312091148GA SA312091154GA SA312091166GA SA312091178GA SA312091196GA SA312091201GA SA312091204GA SA312091206GA SA312091207GA SA312091211GA SA312091218GA SA312091220GA SA312091222GA SA312091232GA SA312091234GA SA312091241GA SA312091243GA SA312091254GA SA312091257GA SA312111652GA SA312111666GA SA312111685GA SA312111688GA SA312111690GA SA312121695GA SA312121698GA SA312121703GA SA312121709GA SA312121713GA SA312121739GA SA312121740GA SA312121751GA SA312121755GA SA312121794GA SA312121798GA SA312091226GA SA312121700GA SA312121706GA SA312121707GA SA312121716GA SA312121729GA SA312121757GA SA312121759GA SA312121767GA SA312121784GA SA312121786GA SA312121787GA SA312121796GA SA312121797GA SA312121800GA SA312121804GA SA312121721GA SA312091252GA SA312091253GA SA312091269GA SA312091277GA SA312091278GA SA312091280GA SA312091282GA SA312091283GA SA312091284GA SA312111654GA SA312121813GA SA312080551GA SA312091239GA SA312091275GA SA312091287GA SA312091289GA SA312091293GA SA312091294GA SA312091297GA SA312091335GA SA312121708GA SA312121768GA SA312121771GA SA312121772GA SA312121778GA SA312121790GA SA312121811GA SA312121812GA SA312121814GA SA312121817GA SA312121821GA SA312121827GA SA312121831GA SA312121833GA SA312121913GA SA312091265GA SA312091298GA SA312091302GA SA312091355GA SA312101428GA SA312111486GA SA312111669GA SA312121705GA SA312121715GA SA312121717GA SA312121718GA SA312121719GA SA312121799GA SA312121818GA SA312091274GA SA312091279GA SA312091304GA SA312091306GA SA312091309GA SA312091311GA SA312091319GA SA312091322GA SA312091330GA SA312091336GA SA312091340GA SA312091345GA SA312091347GA SA312091349GA SA312121829GA SA312121834GA SA312121873GA SA312121887GA SA312121892GA SA312121903GA SA312121997GA SA312121837GA SA312121838GA SA312121849GA SA312121897GA SA312121910GA SA312121920GA SA312121926GA SA312121928GA SA312121939GA SA312121953GA SA312121955GA SA312090928GA SA312091197GA SA312091215GA SA312091221GA SA312091223GA SA312091224GA SA312091225GA SA312091227GA SA312091229GA SA312091230GA SA312091231GA SA312091233GA SA312091236GA SA312091237GA SA312091238GA SA312091240GA SA312091242GA SA312091247GA SA312091248GA SA312091258GA SA312091259GA SA312091260GA SA312091261GA SA312091266GA SA312091268GA SA312091271GA SA312091272GA SA312091276GA SA312069816GA SA312069826GA SA312069838GA SA312070041GA SA312070105GA SA312090923GA SA312091009GA SA312091010GA SA312091013GA SA312091014GA SA312091137GA SA312091153GA SA312091310GA SA312091332GA SA312091337GA SA312091365GA SA312101375GA SA312101386GA SA312101416GA SA312101418GA SA312101434GA SA312101441GA SA312101443GA SA312122026GA SA312069865GA SA312070089GA SA312121762GA SA312121855GA SA312121859GA SA312121929GA SA312121936GA SA312121980GA SA312121982GA SA312121985GA SA312101451GA SA312121839GA SA312121894GA SA312121933GA SA312121940GA SA312121941GA SA312121944GA SA312121946GA SA312121951GA SA312121952GA SA312121960GA SA312121973GA SA312121976GA SA312121984GA SA312122014GA SA312122055GA SA312122057GA SA312069842GA SA312070063GA SA312070088GA SA312070137GA SA312121930GA SA312121932GA SA312121934GA SA312121935GA SA312121958GA SA312121966GA SA312121988GA SA312121990GA SA312122003GA SA312122004GA SA312122007GA SA312122016GA SA312122037GA SA312122038GA SA312122041GA SA312122043GA SA312122052GA SA312122054GA SA312122059GA SA312122062GA SA312122064GA SA312122069GA SA312122071GA SA312132090GA SA312132093GA SA312132095GA SA312132102GA SA312132109GA SA312132110GA SA312132113GA SA312132130GA SA312132132GA SA312132138GA SA312132140GA SA312132142GA SA312132173GA SA312132177GA SA312132178GA SA312069666GA SA312069674GA SA312069677GA SA312069679GA SA312069683GA SA312069685GA SA312069689GA SA312069709GA SA312069712GA SA312122018GA SA312049386GA SA312069818GA SA312069861GA SA312069862GA SA312069881GA SA312121841GA SA312121957GA SA312122025GA SA312122028GA SA312132089GA SA312132100GA SA312132103GA SA312132118GA SA312132136GA SA312132175GA SA312152417GA SA311488466GA SA311488467GA SA312049275GA SA312069692GA SA312069700GA SA312069716GA SA312069721GA SA312069722GA SA312069821GA SA312069855GA SA312069857GA SA312069859GA SA312069860GA SA312069864GA SA312069868GA SA312070022GA SA312132101GA SA312121891GA SA312121899GA SA312121956GA SA312121962GA SA312121967GA SA312121974GA SA312121983GA SA312122031GA SA312122079GA SA312122080GA SA312122087GA SA312132107GA SA312132126GA SA312132135GA SA312152371GA SA312152372GA SA312152373GA SA312152374GA SA312152375GA SA312152376GA SA312152377GA SA312152378GA SA312152379GA SA312152380GA SA312152382GA SA312152386GA SA312152387GA SA312152388GA SA312152390GA SA312152395GA SA312152399GA SA312152400GA SA312152404GA SA312152406GA SA312152409GA SA312152413GA SA312152415GA SA312152418GA SA312152419GA SA312152422GA SA312152424GA SA312152425GA SA312152431GA SA312152433GA SA312152434GA SA312152436GA SA312070214GA SA312091250GA SA312091251GA SA312091256GA SA312091262GA SA312111509GA SA312121732GA SA312121805GA SA312121959GA SA312121964GA SA312122042GA SA312122051GA SA312122058GA SA312122063GA SA312122076GA SA312122082GA SA312122084GA SA312122085GA SA312132105GA SA312132106GA SA312132114GA SA312132115GA SA312132116GA SA312132119GA SA312132123GA SA312132124GA SA312132128GA SA312132139GA SA312132141GA SA312132143GA SA312132144GA SA312132145GA SA312132146GA SA312132147GA SA312132150GA SA312132164GA SA312132166GA SA312132168GA SA312132172GA SA312152311GA SA312152313GA SA312152341GA SA312152353GA SA312152363GA SA312152366GA SA312152414GA SA312152432GA SA312152444GA SA312152448GA SA312152451GA SA312152452GA SA312152461GA SA312152463GA SA312152466GA SA312152467GAMKE SA312192565GA SA312192566GA SA312192567GA SA312192568GA SA312192569GA SA312192570GA SA312192571GA SA312192572GA SA312192573GA SA312192574GA SA312192575GA SA312192576GA SA312192577GA SA312192578GA SA312192579GA SA312192580GAMKE SA312192581GA SA312192582GA SA312192583GA SA312192584GA SA312192585GA SA312192586GA SA312192587GA SA312192589GA SA312192591GA SA312192592GA SA312192593GA SA312192594GA SA312192596GA SA312192598GA SA312192601GA SA312192602GA SA312192606GA SA312192607GA SA312192608GA SA312192609GA SA312192610GA SA312192611GA SA312192612GA SA312192613GA SA312192614GA SA312192615GA SA312192616GA SA312192618GA SA312192619GA SA312192621GA SA312192622GA SA312192624GA SA312192625GA SA312192626GA SA312192627GA SA312192628GA SA312192629GA SA312192630GA SA312192634GA SA312192635GA SA312192636GA SA312192638GA SA312192640GA SA312192641GA SA312192642GAMKE SA312192643GA SA312192647GA SA312192649GA SA312152391GA SA312111682GA SA312121802GA SA312121938GA SA312121949GA SA312122030GA SA312122036GA SA312122039GA SA312122040GA SA312122048GA SA312122050GA SA312122060GA SA312122065GA SA312122067GA SA312122068GA SA312132134GA SA312132167GA SA312132170GA SA312132174GA SA312132176GA SA312152323GA SA312152338GA SA312152340GA SA312152342GA SA312152348GA SA312152354GA SA312152357GA SA312152408GA SA312152411GA SA312152429GA SA312059591GA SA312069714GA SA312121774GA SA312121937GA SA312121994GA SA312122009GA SA312152334GA SA312152355GA SA312069837GA SA312069850GA SA312069863GA SA312122001GA SA312152471GA SA312069869GA SA311428062GA SA311090830GA SA311153186GA SA311195136GA SA311236331GA SA311246391GA SA311246394GA SA311246396GA SA311246401GA SA311246402GA SA311246403GA SA311246405GA SA311246406GA SA311246407GA SA311246413GA SA311246414GA SA311246415GA SA311246428GA SA311246431GA SA311246434GA SA311246448GA SA311297331GA SA311297332GA SA311307420GA SA311307433GA SA311307442GA SA311428096GA SA311428105GA SA311428112GA SA311428115GA SA311428119GA SA311428121GA SA311428124GA SA311428125GA SA311428129GA SA311428131GA SA311428134GA SA311428142GA SA311428144GA SA311428145GA SA311428148GA SA311428154GA SA311428158GA SA311428159GA SA312049234GA SA312080392GA SA312080685GA SA312080702GA SA312080715GA SA312080723GA SA312080730GA SA312080736GA SA312080739GA SA312080740GA SA312080741GA SA312080743GA SA312080746GA SA312080750GA SA312080752GA SA312080755GA SA312080757GA SA312080760GA SA312080762GA SA312080780GA SA312080785GA SA312080788GA SA312080795GA SA310393176GA SA310260540GA SA310290637GA SA310423596GA SA310444256GA SA310455011GA SA310455022GA SA310475465GA SA310475531GA SA310475553GA SA310475555GA SA310475558GA SA310475562GA SA310475564GA SA310475578GA SA310475602GA SA310475616GA SA310475618GA SA310475620GA SA310475687GA SA310475700GA SA310475704GA SA310475705GA SA310475706GA SA310475714GA SA310475717GA SA310475718GA SA310485761GA SA310485763GA SA310485766GA SA310485770GA SA310485772GA SA310485791GA SA310485813GA SA310485820GA SA310485832GA SA310485839GA SA310485842GA SA310485845GA SA310485858GA SA310485862GA SA310485873GA SA310485880GA SA310496236GA SA310496315GA SA310496319GA SA310496325GA SA310496350GA SA310496369GA SA310496466GA SA310496475GA SA310496478GA SA310496486GA SA310496488GA SA310496493GA SA310506816GA SA310506842GA SA310507124GA SA310507213GA SA310507249GA SA310507252GA SA310507253GA SA310507339GA SA310507340GA SA310507347GA SA310507418GA SA310507424GA SA310507437GA SA310507459GA SA310507462GA SA310507470GA SA310507474GA SA310496389GA SA310496440GA SA310506755GA SA310506768GA SA310506774GA SA310506783GA SA310506799GA SA310506803GA SA310506807GA SA310507178GA SA310507195GA SA310507204GA SA310507222GA SA310507226GA SA310507233GA SA310507255GA SA310507257GA SA310507258GA SA310507280GA SA310507284GA SA310507289GA SA310507293GA SA310507294GA SA310507297GA SA310507299GA SA310507300GA SA310507306GA SA310507312GA SA310507315GA SA310507318GA SA310507321GA SA310507322GA SA310507323GA SA310507327GA SA310507331GA SA310507333GA SA310507334GA SA310507335GA SA310507342GA SA310507343GA SA310507345GA SA310507349GA SA310507352GA SA310507354GA SA310507355GA SA310507360GA SA310507361GA SA310507362GA SA310507363GA SA310507378GA SA310517599GA SA310517601GA SA310517611GA SA310517617GA SA310517623GA SA310517631GA SA310517639GA SA310517649GA SA310517675GA SA310517677GA SA310517694GA SA310518158GA SA310518165GA SA310518189GA SA310518206GA SA310518211GA SA310382840GA SA311039509GA SA311059583GA SA311069740GA SA311069757GA SA311080676GA SA311090914GA SA311090915GA SA311090916GA SA311090921GA SA311090922GA SA311090924GA SA311091036GA SA311091046GA SA311091052GA SA311090746GA SA311090749GA SA311121731GA SA311121732GA SA311121878GA SA311121881GA SA311122169GA SA311122173GA SA311122174GA SA311122177GA SA311122179GA SA311122187GA SA311122188GA SA311122190GA SA311122192GA SA311122193GA SA311122196GA SA311122206GA SA311122207GA SA311122212GA SA311122221GA SA311122224GA SA311122226GA SA311122227GA SA311122235GA SA311122236GA SA311122238GA SA311122257GA SA311122258GA SA311122259GA SA311122262GA SA311122267GA SA311132277GA SA311132280GA SA311132290GA SA311132291GA SA311132295GA SA311132297GA SA311132318GA SA311132321GA SA311132330GA SA311132340GA SA311132348GA SA311132367GA SA311132399GA SA311132404GA SA311132406GA SA311122143GA SA311132423GA SA311132432GA SA311132436GA SA311132451GA SA311142914GA SA311142923GA SA311142928GA SA311142956GA SA311153081GA SA311153085GA SA311153087GA SA311153103GA SA311153107GA SA311153109GA SA311153122GA SA311153123GA SA311153130GA SA311153133GA SA311153137GA SA311153142GA SA311153160GA SA311153173GA SA311153182GA SA311153207GA SA311153214GA SA311153218GA SA311153220GA SA311153222GA SA311132380GA SA311142881GA SA311205454GA SA311184542GA SA311142888GA SA311215690GA SA311215695GA SA311215698GA SA311215715GA SA311215731GA SA311215740GA SA311307524GA SA311307538GA SA311307539GA SA311468376GA SA311468389GA SA311468406GA SA311468410GA SA311256808GA SA311256837GA SA311256858GA SA311246487GA SA311518733GA SA312080405GA SA312080407GA SA312080418GA SA312080428GA SA312080568GA SA312080582GA SA312080585GA SA312080600GA SA312080605GA SA312080607GA SA312080613GA SA312080616GA SA312080620GA SA312080638GA SA312080639GA SA312080640GA SA312080652GA SA312080657GA SA312080660GA SA312080664GA SA312080696GA SA312080710GA SA312080733GA SA312080734GA SA312080735GA SA312080430GA SA312080599GA SA312080601GA SA312070016GA SA312070029GA SA312091366GA SA312121908GA SA312101436GA SA312111461GA SA312111463GA SA312111465GA SA312111471GA SA312111506GA SA312111507GA SA311184614GA SA311101295GA SA311111650GA SA311121762GA SA311297288GA SA311246420GA SA311246427GA SA311246432GA SA311246435GA SA311246439GA SA311246441GA SA311297274GA SA311307413GA SA311307414GA SA311307422GA SA311307426GA SA311307429GA SA311307430GA SA311307434GA SA311307435GA SA311307583GA SA311307584GA SA311307591GA SA311458214GA SA311458218GA SA311458219GA SA311458225GA SA311458231GA SA311458243GA SA311458256GA SA311458257GA SA311458262GA SA311458263GA SA311458343GA SA311458355GA SA311458356GA SA311458360GA SA311458364GA SA311458365GA SA311458368GA SA311518728GA SA311518729GA SA312080546GA SA312080557GA SA312080578GA SA312080490GA SA312080634GA SA312080867GA SA312080868GA SA312080869GA SA312080883GA SA312080910GA SA312101447GA SA312111515GA SA312111551GA SA312111562GA SA312111570GA SA312111657GA SA312111658GA SA312111659GA SA312111661GA SA312111662GA SA312111665GA SA312111668GA SA312111670GA SA312111671GA SA312111672GA SA312111674GA SA312111675GA SA312111676GA SA312111678GA SA312111679GA SA312111680GA SA312111681GA SA312111683GA SA312111684GA SA312111686GA SA312111687GA SA312111691GA SA312121692GA SA312121693GA SA312121696GA SA312121697GA SA312121699GA SA312121710GA SA312121714GA SA312121720GA SA312121722GA SA312121723GA SA312121724GA SA312121725GA SA312121728GA SA312121730GA SA312121731GA SA312121733GA SA312121734GA SA312121735GA SA312121737GA SA312121738GA SA312121743GA SA312121745GA SA312121747GA SA312111527GA SA312111660GA SA312121727GA SA312121736GA SA310392930GA SA310393112GA SA310393136GA SA310393263GA SA310393278GA SA310393293GA SA310393300GA SA310393308GA SA310392893GA SA310392901GA SA310392907GA SA310392909GA SA310392917GA SA310393089GA SA310393096GA SA310393185GA SA310393287GA SA310393297GA SA310393298GA SA310393302GA SA310393305GA SA310424042GA SA310413505GA SA310423576GA SA310423702GA SA310423829GA SA310423834GA SA310423952GA SA310423970GA SA310444506GA SA310444510GA SA310454997GA SA310455078GA SA310465371GA SA310465376GA SA310465379GA SA310444846GA SA310496159GA SA310496326GA SA310496330GA SA310496341GA SA310496358GA SA310496362GA SA310496363GA SA310496372GA SA310496454GA SA310496461GA SA310496479GA SA310496481GA SA310496485GA SA310496496GA SA310496500GA SA310496501GA SA310496509GA SA310496511GA SA310496518GA SA310496519GA SA310496520GA SA310496522GA SA310496524GA SA310496649GA SA310506849GA SA310506852GA SA310506867GA SA310507041GA SA310507047GA SA310507076GA SA310507115GA SA310434145GA SA310496617GA SA310496636GA SA310507061GA SA310507513GA SA310507514GA SA310517718GA SA310506776GA SA310506833GA SA310506839GA SA310507165GA SA310507167GA SA310507211GA SA310507214GA SA310507251GA SA310507261GA SA310507264GA SA310507265GA SA310507269GA SA310507273GA SA310507274GA SA310507275GA SA310507277GA SA310507282GA SA310507285GA SA310507341GA SA310507350GA SA310507358GA SA310507359GA SA310507364GA SA310507365GA SA310507369GA SA310517598GA SA310517603GA SA310517609GA SA310517615GA SA310517620GA SA310517628GA SA310517630GA SA310517655GA SA310518151GA SA310518155GA SA310518183GA SA310518188GA SA310518199GA SA310518207GA SA310506767GA SA310506798GA SA310506802GA SA310507147GA SA310507150GA SA310507192GA SA310507207GA SA310496321GA SA310507542GA SA310517561GA SA310517571GA SA310517579GA SA310517580GA SA310517582GA SA310517584GA SA310517586GA SA310517716GA SA310517748GA SA310517749GA SA310517750GA SA310517779GA SA310517816GA SA310517824GA SA310506793GA SA310506815GA SA310506823GA SA310506825GA SA310506829GA SA310507172GA SA310507173GA SA310507191GA SA310507209GA SA310507210GA SA310507229GA SA310507234GA SA310507245GA SA310507283GA SA310517647GA SA310517658GA SA310518173GA SA310518239GA SA310413522GA SA310496127GA SA310496309GA SA310528966GA SA310528988GA SA310528993GA SA311039481GA SA311039489GA SA311039474GA SA310260005GA SA310361657GA SA310423851GA SA310539350GA SA311039435GA SA311039455GA SA311039465GA SA310321023GA SA310382456GA SA310413479GA SA310413481GA SA310413518GA SA310423658GA SA310475726GA SA310507054GA SA310507394GA SA310528860GA SA310528876GA SA310528882GA SA310528964GA SA310528974GA SA310528991GA SA310528996GA SA311039371GA SA310423831GA SA310444311GA SA310455166GA SA310475617GA SA311039404GA SA310382846GA SA311060014GA SA311060045GA SA311069673GA SA310496143GA SA310496218GA SA310496278GA SA310496282GA SA310496285GA SA310496291GA SA311060126GA SA311060154GA SA311060157GA SA311060163GA SA311060198GA SA311069723GA SA311060172GA SA311070536GA SA311060148GA SA310393285GA SA311059586GA SA311059594GA SA311060033GA SA311069626GA SA311069628GA SA311069629GA SA311069631GA SA311069632GA SA311069633GA SA311069634GA SA311069635GA SA311069642GA SA311069643GA SA311069647GA SA311069664GA SA311069686GA SA311069698GA SA311069834GA SA311080672GA SA311080673GA SA311080675GA SA311080681GA SA311080686GA SA311080691GA SA311080693GA SA311080696GA SA311080697GA SA311060010GA SA311101220GA SA311101274GA SA311101300GA SA311101272GA SA311101279GA SA311101424GA SA311090938GA SA311101098GA SA311090792GA SA311122141GA SA311121850GA SA311121854GA SA311121863GA SA311122018GA SA311122031GA SA311122046GA SA311122054GA SA311122229GA SA311122230GA SA311122248GA SA311122251GA SA311122270GA SA311122275GA SA311132287GA SA311132313GA SA311132314GA SA311132322GA SA311132326GA SA311132352GA SA311132398GA SA311132556GA SA311132588GA SA311142750GA SA311142753GA SA311142755GA SA311142761GA SA311142762GA SA311142766GA SA311142770GA SA311142771GA SA311142774GA SA311142776GA SA311142777GA SA311142778GA SA311142794GA SA311142796GA SA311090934GA SA311090998GA SA311091007GA SA311101403GA SA311121737GA SA311121779GA SA311121849GA SA311121853GA SA311121855GA SA311121857GA SA311121862GA SA311121870GA SA311121879GA SA311121894GA SA311121912GA SA311122013GA SA311122015GA SA311122021GA SA311122026GA SA311122028GA SA311122029GA SA311122044GA SA311122047GA SA311122052GA SA311122053GA SA311122055GA SA311122058GA SA311132593GA SA311132598GA SA311142757GA SA311142773GA SA311111614GA SA311121852GA SA311121859GA SA311121860GA SA311121864GA SA311121865GA SA311121866GA SA311121871GA SA311121873GA SA311121874GA SA311121875GA SA311121882GA SA311121883GA SA311121884GA SA311121886GA SA311121887GA SA311121893GA SA311121896GA SA311121898GA SA311121915GA SA311121916GA SA311132554GA SA311142947GA SA311132377GA SA311132387GA SA311121943GA SA311121975GA SA311122007GA SA311122076GA SA311122087GA SA311122089GA SA311122129GA SA311132371GA SA311132518GA SA311142724GA SA311153046GA SA311153067GA SA311153147GA SA311153185GA SA311142918GA SA311153141GA SA311153202GA SA311163944GA SA311163383GA SA311163404GA SA311163856GA SA311163865GA SA311163869GA SA311163888GA SA311163893GA SA311163947GA SA311163951GA SA311163955GA SA311163962GA SA311163980GA SA311163890GA SA311163914GA SA311195052GA SA311195062GA SA311195074GA SA311195076GA SA311195085GA SA311195087GA SA311163930GA SA311174161GA SA311174170GA SA311184363GA SA311184373GA SA311184376GA SA311184379GA SA311184383GA SA311174134GA SA311184457GA SA311184470GA SA311184476GA SA311184486GA SA311184488GA SA311184490GA SA311184501GA SA311184508GA SA311184510GA SA311163809GA SA311174045GA SA311174127GA SA311184473GA SA311184482GA SA311184512GA SA311195007GA SA311195042GA SA311195043GA SA311195047GA SA311195048GA SA311195049GA SA311195051GA SA311195170GA SA311195171GA SA311195172GA SA311246519GA SA311246532GA SA311246535GA SA311246536GA SA311246538GA SA311246539GA SA311246540GA SA311246548GA SA311246550GA SA311246554GA SA311246555GA SA311246556GA SA311246557GA SA311246563GA SA311246579GA SA311153033GA SA311153077GA SA311153258GA SA311163438GA SA311163531GA SA311163535GA SA311163550GA SA311163566GA SA311163568GA SA311163588GA SA311163597GA SA311163699GA SA311184464GA SA311184471GA SA311184511GA SA311163450GA SA311163452GA SA311184521GA SA311153260GA SA311153348GA SA311163610GA SA311163835GA SA311174056GA SA311163648GA SA311163792GA SA311163794GA SA311163814GA SA311163821GA SA311163833GA SA311163854GA SA311174047GA SA311174065GA SA311184722GA SA311184730GA SA311184735GA SA311184736GA SA311184739GA SA311184751GA SA311184752GA SA311184762GA SA311184764GA SA311184768GA SA311184769GA SA311184770GA SA311184779GA SA311184788GA SA311184790GA SA311184791GA SA311184792GA SA311184797GA SA311184805GA SA311184808GA SA311184869GA SA311184871GA SA311205313GA SA311205317GA SA311205322GA SA311205331GA SA311205393GA SA311205404GA SA311205405GA SA311205406GA SA311205415GA SA311205417GA SA311205421GA SA311205422GA SA311205428GA SA311205430GA SA311205431GA SA311205435GA SA311205438GA SA311256811GA SA311256846GA SA311256851GA SA311256863GA SA311163805GA SA311184719GA SA311184725GA SA311184729GA SA311184732GA SA311184753GA SA311184776GA SA311184798GA SA311184799GA SA311132546GA SA311153307GA SA311174058GA SA311184690GA SA311184721GA SA311184737GA SA311184763GA SA311184816GA SA311184825GA SA311194962GA SA311194987GA SA311205328GA SA311205339GA SA311205403GA SA311205407GA SA311205413GA SA311205419GA SA311205433GA SA311205570GA SA311205586GA SA311205594GA SA311205595GA SA311205596GA SA311205598GA SA311205599GA SA311205600GA SA311205603GA SA311205612GA SA311205620GA SA311225897GA SA311307452GA SA311317650GA SA311039373GA SA311060232GA SA311060274GA SA311266877GA SA311266885GA SA311266887GA SA311266888GA SA311266904GA SA311266905GA SA311266908GA SA311205311GA SA311225928GA SA311225929GA SA311225930GA SA311225951GA SA311225957GA SA311236041GA SA311236044GA SA311225923GA SA311225924GA SA311225927GA SA311225933GA SA311225944GA SA311225947GA SA311225955GA SA311225960GA SA311225970GA SA311236054GA SA311287014GA SA311287041GA SA311297272GA SA311307415GA SA311307417GA SA311307432GA SA311307532GA SA311060274GA SA311236373GA SA311246451GA SA311246455GA SA311246458GA SA311246464GA SA311246475GA SA311246505GA SA311246508GA SA311246590GA SA311246667GA SA311256722GA SA311256724GA SA311256735GA SA311256736GA SA311256741GA SA311256820GA SA311256824GA SA311256853GA SA311297307GA SA311297310GA SA311184838GA SA311236217GA SA311236353GA SA311236354GA SA311236370GA SA311236375GA SA311236378GA SA311236381GA SA311246491GA SA311518726GA SA312018906GA SA311195235GA SA311195248GA SA311297297GA SA312028949GA SA312028967GA SA312028990GA SA312028996GA SA312029015GA SA312029053GA SA312029068GA SA312029072GA SA311468412GA SA311468426GA SA311468439GA SA311194931GA SA311194973GA SA311205576GA SA311205583GA SA311205606GA SA311307490GA SA311458307GA SA311458308GA SA311458309GA SA311458311GA SA311458312GA SA311458313GA SA311458314GA SA311458316GA SA311458319GA SA311458331GA SA311458334GA SA311468413GA SA311468416GA SA311468418GA SA311468421GA SA311468422GA SA311468425GA SA311468427GA SA311468429GA SA311468431GA SA311468436GA SA311468437GA SA311468440GA SA311468441GA SA311468442GA SA311468444GA SA311468449GA SA311488463GA SA312039115GA SA312039188GA SA312049323GA SA312049325GA SA312049333GA SA312049334GA SA312049345GA SA312049357GA SA312049366GA SA312049424GA SA312049422GA SA312069824GA SA312069827GA SA312069835GA SA312079889GA SA312079893GA SA312079896GA SA312079898GA SA312079901GA SA312079904GA SA312079905GA SA312079906GA SA312079909GA SA312079910GA SA312079911GA SA312079925GA SA312049458GA SA312059476GA SA312059629GA SA312059630GA SA312049444GA SA312059478GA SA312079949GA SA312080468GA SA312080470GA SA312080475GA SA312080518GA SA312080519GA SA312080527GA SA312080643GA SA312080480GA SA312080699GA SA312080786GA SA312080789GA SA312080825GA SA312080841GA SA312080850GA SA312080854GA SA312080889GA SA312080895GA SA312080906GA SA312080907GA SA312080887GA SA312070078GA SA312080861GA SA312080902GA SA312080904GA SA312080908GA SA312091029GA SA312091047GA SA312091065GA SA312070268GA SA312091049GA SA312069788GA SA312069791GA SA312069800GA SA312069830GA SA312069840GA SA312069845GA SA312069854GA SA312091138GA SA312091140GA SA312091150GA SA312111677GA SA312121744GA SA312069819GA SA312091270GA SA312091281GA SA312091290GAMKE SA312091300GA SA312091307GA SA312091308GA SA312091312GA SA312091315GA SA312091316GA SA312091320GA SA312091323GA SA312091325GA SA312091326GA SA312091329GA SA312091331GA SA312091334GA SA312091338GA SA312091339GA SA312091341GA SA312091342GA SA312091343GA SA312091346GA SA312091348GA SA312091356GA SA312091358GA SA312091359GA SA312091360GA SA312091362GA SA312091364GA SA312121999GA SA312121942GA SA312132108GA SA312121927GA SA312121947GA SA312121995GA SA312122029GA SA312122046GA SA312069705GA SA312091318GA SA312091351GA SA312101367GA SA312101368GA SA312101371GA SA312101373GA SA312101381GA SA312101382GA SA312101383GA SA312101384GA SA312101385GA SA312101387GA SA312101388GA SA312101391GA SA312101392GA SA312101393GA SA312101394GA SA312122000GA SA312122005GA SA311184515GA SA311184533GA SA311448205GA SA310444286GA SA310506715GA SA310507112GA SA310507403GA SA310507453GA SA310507457GA SA310507520GA SA310507544GA SA311090964GA SA311101088GA SA311101090GA SA311101135GA SA311101139GA SA311101157GA SA311101457GA SA311101466GA SA311111620GA SA311111634GA SA311111635GA SA311111639GA SA311111643GA SA311111644GA SA311111645GA SA311111646GA SA311111647GA SA311111648GA SA311111651GA SA311111652GA SA311111653GA SA311111656GA SA311111658GA SA311111660GA SA311111664GA SA311121980GA SA311122068GA SA311122070GA SA311122075GA SA311122078GA SA311122081GA SA311122083GA SA311122090GA SA311122091GA SA311122092GA SA311122097GA SA311122110GA SA311122121GA SA311122122GA SA311122137GA SA311122139GA SA311122144GA SA311122146GA SA311122151GA SA311122152GA SA311122157GA SA311122158GA SA311122162GA SA311122165GA SA311132428GA SA311132493GA SA311132538GA SA311132542GA SA311132544GA SA311132601GA SA311132602GA SA311132608GA SA311132609GA SA311132611GA SA311132621GA SA311132622GA SA311132624GA SA311132625GA SA311132626GA SA311132627GA SA311132628GA SA311132629GA SA311132632GA SA311132633GA SA311132637GA SA311132638GA SA311132640GA SA311132642GA SA311132643GA SA311132644GA SA311132645GA SA311132647GA SA311132654GA SA311132656GA SA311132657GA SA311132661GA SA311132664GA SA311132665GA SA311132670GA SA311132671GA SA311132672GA SA311132673GA SA311142687GA SA311142699GA SA311142843GA SA311163402GA SA311163861GA SA311163864GA SA311163866GA SA311163875GA SA311163879GA SA311163895GA SA311163910GA SA311205293GA SA311225953GA SA311236006GA SA311236007GA SA311236008GA SA311236014GA SA311236020GA SA311236021GA SA311236024GA SA311236042GA SA311236046GA SA311236052GA SA311236056GA SA311236165GA SA311236188GA SA311236200GA SA311236202GA SA311236204GA SA311236205GA SA311236208GA SA311236213GA SA311236214GA SA311236220GA SA311236224GA SA311236226GA SA311236230GA SA311236232GA SA311236253GA SA311236255GA SA311236258GA SA311236280GA SA311236364GA SA311246449GA SA311246476GA SA311246478GA SA311246495GA SA311246498GA SA311246501GA SA311256802GA SA311256806GA SA311256813GA SA311256849GA SA311256861GA SA311256864GA SA311256867GA SA311266913GA SA311266916GA SA311266917GA SA311266918GA SA311266920GA SA311266921GA SA311266924GA SA311266926GA SA311266927GA SA311266929GA SA311266930GA SA311266931GA SA311266932GA SA311266933GA SA311266934GA SA311266936GA SA311266937GA SA311266939GA SA311266940GA SA311266943GA SA311266944GA SA311266945GA SA311266948GA SA311266949GA SA311266950GA SA311307576GA SA311327915GA SA311184866GA SA311236189GA SA311256854GA SA311468430GA SA311468435GA SA311468447GA SA311488453GA SA312079885GA SA312079886GA SA312079913GA SA312079931GA SA312079947GA SA312079976GA SA312079981GA SA312079984GA SA312049367GA SA312049375GA SA312049376GA SA312049379GA SA312049381GA SA312049384GA SA312049413GA SA312049420GA SA311195061GA SA311195068GA SA311236065GA SA311256798GA SA312018864GA SA310423750GA SA311069707GA SA311132540GA SA311195199GA SA311132299GA SA311428072GA SA310496633GA SA310496634GA SA310506676GA SA310506704GA SA310506710GA SA310506719GA SA310507495GA SA310528921GA SA310475483GA SA311132617GA SA311132631GA SA311174199GA SA311174224GA SA311163711GA SA311163716GA SA311256812GA SA311256816GA SA311256869GA SA310361689GA SA310372009GA SA310372028GA SA310372032GA SA310372034GA SA310361903GA SA310361913GA SA310382668GA SA310382677GA SA310382842GA SA310392855GA SA310392872GA SA310392891GA SA310403316GA SA310423774GA SA310475466GA SA310475482GA SA310475488GA SA310475519GA SA310475527GA SA310475529GA SA310475542GA SA310475543GA SA310475551GA SA310475586GA SA310475597GA SA310475612GA SA310475638GA SA310475644GA SA310475673GA SA310475702GA SA310485805GA SA310485809GA SA310485829GA SA310485838GA SA310485840GA SA310485861GA SA310485870GA SA310485883GA SA310485889GA SA310495910GA SA310495969GA SA310495971GA SA310496233GA SA310496237GA SA310496238GA SA310496484GA SA310403355GA SA310495895GA SA310495898GA SA310495902GA SA310495905GA SA310495907GA SA310495914GA SA310495915GA SA310495917GA SA310495937GA SA310495944GA SA310495947GA SA310495948GA SA310495949GA SA310495951GA SA310495952GA SA310495958GA SA310495960GA SA310495961GA SA310495972GA SA310495974GA SA310495975GA SA310496142GA SA310496274GA SA310496284GA SA310496290GA SA310496292GA SA310496294GA SA310496299GA SA310496304GA SA310496332GA SA310496347GA SA310496352GA SA310496353GA SA310496354GA SA310496360GA SA310496361GA SA310496364GA SA310496373GA SA310496464GA SA310496476GA SA310496504GA SA310496526GA SA310496531GA SA310496532GA SA310496533GA SA310434062GA SA310475477GA SA310475490GA SA310475526GA SA310475528GA SA310475589GA SA310475596GA SA310496132GA SA310496154GA SA310496208GA SA310496348GA SA310496367GA SA310496523GA SA310496289GA SA310496334GA SA310496377GA SA310496379GA SA310496400GA SA310496403GA SA310496413GA SA310496430GA SA310496507GA SA310496516GA SA310496619GA SA310496635GA SA310496639GA SA310496640GA SA310496647GA SA310496652GA SA310506660GA SA310506698GA SA310506720GA SA310506730GA SA310506758GA SA310506764GA SA310506769GA SA310506777GA SA310506779GA SA310506794GA SA310506797GA SA310506804GA SA310506811GA SA310506866GA SA310506898GA SA310507159GA SA310507180GA SA310507189GA SA310507194GA SA310507198GA SA310507208GA SA310507502GA SA310496147GA SA310496342GA SA310496359GA SA310496502GA SA310507026GA SA310507035GA SA310507039GA SA310507091GA SA310517734GA SA310517743GA SA310518221GA SA310518226GA SA310518227GA SA310518231GA SA310518309GA SA310506668GA SA310506721GA SA310506740GA SA310506751GA SA310506856GA SA310506858GA SA310517710GA SA310517787GA SA310517833GA SA310507434GA SA310507438GA SA310507467GA SA310507468GA SA310507557GA SA310517642GA SA310517643GA SA310517644GA SA310517648GA SA310517653GA SA310517684GA SA310518145GA SA310518153GA SA310518167GA SA310518171GA SA310518172GA SA310518174GA SA310518175GA SA310518178GA SA310506709GA SA310507087GA SA310517822GA SA310517834GA SA310517835GA SA310517843GA SA310517879GA SA310517984GA SA310517569GA SA310517733GA SA310517781GA SA310517819GA SA310517820GA SA310517829GA SA310517989GA SA310517992GA SA310517999GA SA310518006GA SA310518008GA SA310518011GA SA310528877GA SA310496327GA SA310496337GA SA310496345GA SA310518616GA SA310528851GA SA310528857GA SA310528873GA SA310528893GA SA310528900GA SA310528970GA SA310528971GA SA310528973GA SA310528978GA SA310528983GA SA310528986GA SA310529012GA SA310529146GA SA310529149GA SA311039369GA SA311039400GA SA311039411GA SA311039415GA SA310496335GA SA310528866GA SA310528874GA SA310528885GA SA311039394GA SA311039398GA SA311039407GA SA311039471GA SA310393069GA SA310444495GA SA310465377GA SA310260394GA SA310382363GA SA310382833GA SA310393215GA SA310403347GA SA310465309GA SA310465319GA SA310518260GA SA310518354GA SA310518425GA SA310518428GA SA310518497GA SA310518500GA SA310518504GA SA310518514GA SA310518517GA SA310518518GA SA310518520GA SA310518524GA SA310518562GA SA310528784GA SA310528786GA SA310528788GA SA310528789GA SA310528796GA SA310528797GA SA310528811GA SA310528841GA SA311039503GA SA311039506GA SA311039512GA SA311039514GA SA311039515GA SA311039517GA SA311039522GA SA310403314GA SA311060018GA SA311060039GA SA311060042GA SA311060078GA SA311060214GA SA311069646GA SA311069670GA SA311069671GA SA311069693GA SA311069743GA SA311069775GA SA311069776GA SA311069778GA SA311069791GA SA311069818GA SA311069852GA SA311060155GA SA311060171GA SA311060235GA SA311060238GA SA311060240GA SA311060254GA SA311060265GA SA311069719GA SA311060255GA SA311060270GA SA311060271GA SA311060297GA SA311060298GA SA311060303GA SA311060308GA SA311060343GA SA311060385GA SA311060397GA SA310528824GA SA311060087GA SA311060094GA SA311060109GA SA311060068GA SA311060089GA SA311060107GA SA311080678GA SA311080670GA SA311080679GA SA311101297GA SA311090849GA SA311090860GA SA311090870GA SA311090877GA SA311101241GA SA311101278GA SA311091033GA SA311091047GA SA311080615GA SA311101077GA SA311101080GA SA311101085GA SA311122199GA SA311101140GA SA311101141GA SA311132475GA SA311121913GA SA311142968GA SA311142953GA SA311121781GA SA311121786GA SA311121805GA SA311122142GA SA311122164GA SA311122100GA SA311069783GA SA311069809GA SA311069812GA SA311090944GA SA311091003GA SA311091006GA SA311101094GA SA311101105GA SA311101106GA SA311121753GA SA311121970GA SA311121977GA SA311121985GA SA311132373GA SA311132379GA SA311132388GA SA311132412GA SA311132474GA SA311132478GA SA311132484GA SA311132603GA SA311121797GA SA311121842GA SA311121924GA SA311121951GA SA311121978GA SA311121984GA SA311121991GA SA311121996GA SA311122000GA SA311122077GA SA311132424GA SA311132456GA SA311132469GA SA311132530GA SA311132532GA SA311132535GA SA311132536GA SA311132541GA SA311132547GA SA311132548GA SA311132615GA SA311132641GA SA311163993GA SA311163870GA SA311163908GA SA311163946GA SA311163961GA SA311163973GA SA311163974GA SA311163979GA SA311153051GA SA311174163GA SA311174197GA SA311163929GA SA311174171GA SA311174173GA SA311174174GA SA311174175GA SA311174183GA SA311174193GA SA311174204GA SA311184348GA SA311184546GA SA311184551GA SA311195075GA SA311174190GA SA311184375GA SA311184384GA SA311184388GA SA311184526GA SA311184573GA SA311184583GA SA311184587GA SA311184607GA SA311225982GA SA311226001GA SA311236071GA SA311236083GA SA311246389GA SA311246533GA SA311246541GA SA311246569GA SA311246574GA SA311246575GA SA311246577GA SA311246578GA SA311246580GA SA311246582GA SA311142694GA SA311142725GA SA311142731GA SA311142736GA SA311143010GA SA311143012GA SA311143022GA SA311153331GA SA311163429GA SA311163454GA SA311163455GA SA311163488GA SA311163532GA SA311163551GA SA311163594GA SA311163609GA SA311163697GA SA311163825GA SA311163836GA SA311163843GA SA311163850GA SA311184750GA SA311184765GA SA311184844GA SA311184857GA SA311205396GA SA311256844GA SA311256866GA SA311205306GA SA311205323GA SA311163556GA SA311163569GA SA311163571GA SA311163573GA SA311163575GA SA311163578GA SA311163590GA SA311163593GA SA311163600GA SA311163603GA SA311163604GA SA311163615GA SA311163616GA SA311163618GA SA311163619GA SA311163622GA SA311163631GA SA311163635GA SA311163637GA SA311163643GA SA311163644GA SA311163652GA SA311163654GA SA311163701GA SA311163725GA SA311163785GA SA311163786GA SA311163788GA SA311163793GA SA311163812GA SA311163813GA SA311163828GA SA311163851GA SA311163855GA SA311174001GA SA311174020GA SA311174030GA SA311174034GA SA311174039GA SA311174048GA SA311174050GA SA311174054GA SA311174061GA SA311174063GA SA311174068GA SA311174069GA SA311174070GA SA311174074GA SA311174079GA SA311174082GA SA311174086GA SA311174090GA SA311174092GA SA311174094GA SA311174099GA SA311174101GA SA311174102GA SA311174104GA SA311174105GA SA311174106GA SA311174109GA SA311174110GA SA311174111GA SA311174112GA SA311174113GA SA311174114GA SA311174115GA SA311174116GA SA311174117GA SA311174118GA SA311174124GA SA311174129GA SA311174135GA SA311174136GA SA311174137GA SA311174138GA SA311174139GA SA311174145GA SA311174146GA SA311174148GA SA311174154GA SA311174155GA SA311174157GA SA311184684GA SA311184760GA SA311184761GA SA311184810GA SA311266928GA SA311080677GA SA311132480GA SA311132503GA SA311163570GA SA311163596GA SA311163599GA SA311163605GA SA311163632GA SA311163634GA SA311163646GA SA311163653GA SA311163796GA SA311163807GA SA311266912GA SA311153041GA SA311163457GA SA311163539GA SA311163541GA SA311163557GA SA311163611GA SA311163651GA SA311163703GA SA311163706GA SA311163708GA SA311163709GA SA311163721GA SA311163727GA SA311163730GA SA311163731GA SA311163779GA SA311163787GA SA311163795GA SA311163797GA SA311163798GA SA311163802GA SA311163804GA SA311163810GA SA311163816GA SA311163819GA SA311163822GA SA311163826GA SA311163827GA SA311163831GA SA311163834GA SA311163842GA SA311173998GA SA311174008GA SA311174010GA SA311174011GA SA311174012GA SA311174013GA SA311174014GA SA311174022GA SA311174023GA SA311174025GA SA311174026GA SA311174028GA SA311174029GA SA311174036GA SA311174037GA SA311174042GA SA311174043GA SA311174046GA SA311174055GA SA311174057GA SA311317653GA SA311317654GA SA311317659GA SA311317714GA SA311163577GA SA311163608GA SA311163777GA SA311194904GA SA311194933GA SA311194934GA SA311194978GA SA311194984GA SA311194993GA SA311195004GA SA311205564GA SA311327826GA SA311327890GA SA310517784GA SA310529078GA SA311039368GA SA311039374GA SA311039376GA SA311039401GA SA311039406GA SA311039428GA SA311039431GA SA311039436GA SA311039443GA SA311039444GA SA311039444GAMKE SA311039445GA SA311039456GA SA311039462GA SA311039466GA SA311039476GA SA311039574GA SA311060228GA SA311142939GA SA311142977GA SA311215686GA SA311236070GA SA311225954GA SA311236004GA SA311236018GA SA311236022GA SA311236026GA SA311236031GA SA311236039GA SA311236045GA SA311236048GA SA311236053GA SA311236057GA SA311236123GA SA311236156GA SA311236157GA SA311236158GA SA311236162GA SA311236175GA SA311236176GA SA311236179GA SA311236180GA SA311236183GA SA311236192GA SA311236193GA SA311236194GA SA311236195GA SA311236196GA SA311236197GA SA311236198GA SA311236199GA SA311236201GA SA311236203GA SA311236207GA SA311236209GA SA311236215GA SA311236219GA SA311236221GA SA311236223GA SA311236225GA SA311236229GA SA311236257GA SA311236259GA SA311266914GA SA311266923GA SA311266942GA SA311287098GA SA311287176GA SA311307457GA SA311317682GA SA311317750GA SA311327847GA SA311337980GA SA311337985GA SA311337986GA SA311337989GA SA311337997GA SA311337999GA SA311338004GA SA311428013GA SA311428026GA SA311428028GA SA311448175GA SA311448176GA SA311448180GA SA311448196GA SA311448197GA SA311448198GA SA311448200GA SA311448203GA SA311448204GA SA311448206GA SA311448207GA SA311448208GA SA311448209GA SA311448210GA SA311448211GA SA311458267GA SA311458268GA SA311458269GA SA311458270GA SA311458271GA SA311458273GA SA311458276GA SA311458277GA SA311458278GA SA311458280GA SA311458281GA SA311458283GA SA311458284GA SA311458285GA SA311458286GA SA311458287GA SA311458288GA SA311458289GA SA311458290GA SA311458293GA SA311458294GA SA311458295GA SA311458318GA SA311458320GA SA311458323GA SA311458324GA SA311458325GA SA311458326GA SA311458327GA SA311458328GA SA311458330GA SA311458332GA SA311458337GA SA311458339GA SA311236345GA SA311246471GA SA311246473GA SA311287032GA SA311287092GA SA311297224GA SA310444298GA SA311205447GA SA311205489GA SA311307416GA SA311307451GA SA311317665GA SA311327894GA SA311142702GA SA311225948GA SA311236011GA SA311337975GA SA311337976GA SA311337981GA SA311428081GA SA311448172GA SA311448174GA SA311448193GA SA311448199GA SA311448201GA SA311458274GA SA311458291GA SA311458296GA SA311458315GA SA311458329GA SA311458335GA SA311458336GA SA311468378GA SA311468379GA SA311468385GA SA311468405GA SA311317721GA SA311317726GA SA311317731GA SA311256850GA SA311256842GA SA310528976GA SA311508571GA SA311508579GA SA311508580GA SA311508583GA SA311508594GA SA311508595GA SA311508601GA SA311508602GA SA311508603GA SA311508604GA SA311508605GA SA311508606GA SA311508607GA SA311508608GA SA311508609GA SA311508610GA SA311508611GA SA311508612GA SA311508613GA SA311508615GA SA311508617GA SA311508618GA SA311508620GA SA311508621GA SA311508622GA SA311508628GA SA311508629GA SA311508630GA SA311508631GA SA311508632GA SA311508633GA SA311508634GA SA311508635GA SA311508638GA SA311508639GA SA311508637GA SA311508653GA SA311508657GA SA311327877GA SA311508619GA SA311518691GA SA311518696GA SA311297233GA SA311297235GA SA311317757GA SA311317762GA SA311142682GA SA311142692GA SA311142728GA SA311142734GA SA311142735GA SA311317604GA SA311317601GA SA311317636GA SA312029067GA SA312039075GA SA312039079GA SA312039095GA SA311246481GA SA311307573GA SA311327844GA SA311327854GA SA311327859GA SA311327860GA SA311327862GA SA311458275GA SA311458310GA SA311458317GA SA311458322GA SA311468423GA SA311468434GA SA311488452GA SA311488461GA SA311488489GA SA312018821GA SA311142685GA SA312049445GA SA312059624GA SA312059628GA SA312069729GA SA312069773GA SA312069699GA SA312069702GA SA312069703GA SA312069704GA SA312069708GA SA312069710GA SA312069811GA SA312070138GA SA312080632GA SA312080642GA SA312080387GA SA312080404GA SA312080406GA SA312080413GA SA312080414GA SA312080474GA SA312080586GA SA312080588GA SA312080619GA SA312080637GA SA312080644GA SA312080650GA SA312080654GA SA312080665GA SA312080669GA SA312080672GA SA312080676GA SA312080697GA SA312080703GA SA312080738GA SA312069789GA SA312070077GA SA312070127GA SA312070128GA SA312070144GA SA312070160GA SA312070181GA SA312070185GA SA312070189GA SA312070192GA SA312070203GA SA312070205GA SA312070207GA SA312070213GA SA312070218GA SA312070223GA SA312070224GA SA312070228GA SA312070229GA SA312070234GA SA312070235GA SA312070238GA SA312070239GA SA312070240GA SA312069706GA SA312069724GA SA312069781GA SA312069783GA SA312069786GA SA312069797GA SA312069798GA SA312069807GA SA312069809GA SA312069813GA SA312070194GA SA312070222GA SA312070316GA SA312070337GA SA312070352GA SA312070358GA SA312070362GA SA312070379GA SA312070383GA SA312079965GA SA312091255GA SA312121901GA SA312121945GA SA312121954GA SA312122006GA SA312122019GA SA312122023GA SA312122024GA SA312122032GA SA312122035GA SA312122049GA SA312122061GA SA312122072GA SA312132120GA SA312121776GA SA312121991GA SA311508637GA SA311508653GA SA311508657GA SA312121890GA SA310444282GA SA310444367GA SA310444487GA SA310444489GA SA310444491GA SA310444493GA SA310444497GA SA310444500GA SA310444501GA SA310444502GA SA310444517GA SA310444519GA SA310444526GA SA310444532GA SA310444534GA SA310444843GA SA310444844GA SA310444845GA SA310444852GA SA310444857GA SA310444860GA SA310444862GA SA310444865GA SA310444890GA SA310444891GA SA310423881GA SA310424040GA SA310434095GA SA310454931GA SA310454951GA SA310331330GA SA310393238GA SA310403315GA SA310403323GA SA310403329GA SA310403339GA SA310403354GA SA310403356GA SA310403358GA SA310403359GA SA310403360GA SA310403361GA SA310403362GA SA310403367GA SA310403370GA SA310403378GA SA310403434GA SA310423743GA SA310423747GA SA310423751GA SA310423802GA SA310423899GA SA310423977GA SA310423991GA SA310434049GA SA310434057GA SA310444268GA SA310444271GA SA310444273GA SA310444274GA SA310444279GA SA310444280GA SA310444285GA SA310444288GA SA310444290GA SA310444293GA SA310444295GA SA310444297GA SA310444299GA SA310444300GA SA310444303GA SA310444307GA SA310444309GA SA310444310GA SA310444313GA SA310444315GA SA310444316GA SA310444322GA SA310444323GA SA310444366GA SA310444368GA SA310444490GA SA310444492GA SA310444514GA SA310444520GA SA310444522GA SA310444848GA SA310444869GA SA310413516GA SA310423584GA SA310423856GA SA310434230GA SA310444302GA SA310444533GA SA310444853GA SA310444854GA SA310454978GA SA310454980GA SA310454999GA SA310455002GA SA310455008GA SA310455013GA SA310455014GA SA310455019GA SA310455023GA SA310455027GA SA310455049GA SA310455054GA SA310455059GA SA310455064GA SA310455158GA SA310455162GA SA310351539GA SA310423580GA SA310423846GA SA310444511GA SA310444847GA SA310444861GA SA310454930GA SA310454932GA SA310454941GA SA310454942GA SA310454979GA SA310454998GA SA310455005GA SA310455007GA SA310455010GA SA310455020GA SA310455050GA SA310455063GA SA310455171GA SA310413453GA SA310413543GA SA310423693GA SA310423699GA SA310423867GA SA310423877GA SA310423889GA SA310423909GA SA310423911GA SA310423976GA SA310423993GA SA310434105GA SA310455070GA SA310455071GA SA310455072GA SA310455073GA SA310455074GA SA310455117GA SA310455129GA SA310455135GA SA310455137GA SA310455141GA SA310455225GA SA310455226GA SA310455229GA SA310455239GA SA310455240GA SA310455241GA SA310455242GA SA310455243GA SA310455244GA SA310455246GA SA310455250GA SA310455252GA SA310455255GA SA310455258GA SA310455269GA SA310455276GA SA310455281GA SA310455285GA SA310455287GA SA310455290GA SA310465296GA SA310465320GA SA310465336GA SA310465340GA SA310465345GA SA310465353GA SA310465356GA SA310465357GA SA310465360GA SA310465362GA SA310465363GA SA310465365GA SA310465388GA SA310465395GA SA310465399GA SA310465401GA SA310465402GA SA310465404GA SA310465405GA SA310465406GA SA310465407GA SA310465409GA SA310465413GA SA310465415GA SA310465416GA SA310465417GA SA310465418GA SA310465421GA SA310465422GA SA310465424GA SA310465425GA SA310465426GA SA310465428GA SA310465429GA SA310465430GA SA310465434GA SA310465440GA SA310465442GA SA310465444GA SA310465447GA SA310465448GA SA310475450GA SA310475454GA SA310475455GA SA310475456GA SA310475457GA SA310475459GA SA310475462GA SA310475463GA SA310475469GA SA310475472GA SA310475475GA SA310475476GA SA310475478GA SA310475486GA SA310475491GA SA310475492GA SA310475502GA SA310475503GA SA310475506GA SA310475507GA SA310475512GA SA310475514GA SA310403382GA SA310403387GA SA310423732GA SA310424035GA SA310434068GA SA310455001GA SA310455021GA SA310455032GA SA310455051GA SA310455089GA SA310455167GA SA310455169GA SA310465351GA SA310465355GA SA310465358GA SA310465364GA SA310465372GA SA310465373GA SA310465375GA SA310465378GA SA310465382GA SA310465384GA SA310465387GA SA310465389GA SA310465390GA SA310465392GA SA310465393GA SA310424043GA SA310475461GA SA310475521GA SA310475453GA SA310506679GA SA310518244GA SA310518253GA SA310518254GA SA310518328GA SA310518331GA SA310518338GA SA310518339GA SA310518340GA SA310518344GA SA310518405GA SA310518469GA SA310506673GA SA310506680GA SA310506684GA SA310506687GA SA310506694GA SA310506695GA SA310506714GA SA310506716GA SA310506724GA SA310507071GA SA310507079GA SA310507390GA SA310507423GA SA310517573GA SA310517698GA SA310517699GA SA310517701GA SA310517702GA SA310517705GA SA310517707GA SA310517713GA SA310517722GA SA310517725GA SA310517827GA SA310517867GA SA310517868GA SA310518012GA SA310518646GA SA310518650GA SA310518667GA SA310518668GA SA310518669GA SA310518707GA SA310518708GA SA310518709GA SA310518722GA SA310496336GA SA310507066GA SA310507081GA SA310507101GA SA310507136GA SA310517590GA SA310517826GA SA310518005GA SA310517832GA SA310517874GA SA310518248GA SA310518281GA SA310518284GA SA310518287GA SA310518290GA SA310518302GA SA310518316GA SA310518329GA SA310507450GA SA310518701GA SA311039383GA SA311039426GA SA311039492GA SA311039427GA SA311039487GA SA310434065GA SA310529077GA SA311121812GA SA311132283GA SA311121713GA SA311121917GA SA311163405GA SA311163859GA SA311163863GA SA311163867GA SA311163877GA SA311163883GA SA311163889GA SA311163898GA SA311163909GA SA311163918GA SA311163920GA SA311163940GA SA311163945GA SA311163958GA SA311163960GA SA311163964GA SA311163972GA SA311163976GA SA311163978GA SA311163983GA SA311163984GA SA311163987GA SA311163988GA SA311163994GA SA311163997GA SA311153049GA SA311174177GA SA311174207GA SA311174210GA SA311174211GA SA311184625GA SA311184872GA SA312049421GA SA312111547GA SA312111550GA SA312111553GA SA312111555GA SA312111574GA SA312111576GA SA312111579GA SA312111583GA SA310392884GA SA310393182GA SA310393183GA SA310393208GA SA310393221GA SA310423665GA SA310423669GA SA310423679GA SA310423692GA SA310423700GA SA310423703GA SA310423706GA SA310423707GA SA310423710GA SA310423711GA SA310423714GA SA310423716GA SA310423718GA SA310423719GA SA310423722GA SA310423727GA SA310423927GA SA310423941GA SA310423954GA SA310423964GA SA310423969GA SA310424014GA SA310434087GA SA310434121GA SA310434147GA SA310475449GA SA310475458GA SA310475468GA SA310475470GA SA310475473GA SA310475479GA SA310475484GA SA310475499GA SA310475504GA SA310475505GA SA310475508GA SA310475509GA SA310475511GA SA310475522GA SA310475524GA SA310475533GA SA310475480GA SA310475523GA SA310496328GA SA310496365GA SA310496459GA SA310496482GA SA310496338GA SA310496339GA SA310496455GA SA310496505GA SA310496521GA SA310496525GA SA310496621GA SA310496627GA SA310496637GA SA310496641GA SA310496642GA SA310496644GA SA310507109GA SA310423792GA SA310496224GA SA310496395GA SA310496407GA SA310496409GA SA310496420GA SA310496424GA SA310496425GA SA310496429GA SA310496444GA SA310496445GA SA310496449GA SA310496451GA SA310506727GA SA310506735GA SA310506763GA SA310506765GA SA310506770GA SA310507059GA SA310507063GA SA310507133GA SA310507137GA SA310507151GA SA310507155GA SA310507160GA SA310507193GA SA310507217GA SA310506760GA SA310506780GA SA310506795GA SA310506805GA SA310506834GA SA310507168GA SA310507169GA SA310507184GA SA310507186GA SA310507239GA SA310507248GA SA310507250GA SA310507256GA SA310507286GA SA310507295GA SA310507368GA SA310507370GA SA310506857GA SA310518180GA SA310507435GA SA310507531GA SA310506688GA SA310506690GA SA310507046GA SA310507144GA SA310507393GA SA310518285GA SA310518300GA SA310518308GA SA310518315GA SA310518317GA SA310331340GA SA310403380GA SA310423975GA SA310507177GA SA310507291GA SA310517622GA SA310517627GA SA310517640GA SA310517654GA SA310517665GA SA310517666GA SA310517668GA SA310517669GA SA310517671GA SA310517674GA SA310518148GA SA310518166GA SA310518196GA SA310518201GA SA310518215GA SA310518218GA SA310518224GA SA310518238GA SA310518243GA SA310518246GA SA310518249GA SA310518250GA SA310518264GA SA310518267GA SA310518273GA SA310518279GA SA310518293GA SA310518295GA SA310518301GA SA310518310GA SA310518313GA SA310518314GA SA310518318GA SA310518321GA SA310518327GA SA310518334GA SA310518619GA SA310518346GA SA310518387GA SA310518389GA SA310518429GA SA310518435GA SA310518438GA SA310528945GA SA310529152GA SA310529221GA SA311039388GA SA311039399GA SA311039408GA SA311039420GA SA311039437GA SA311039447GA SA311039454GA SA311070513GA SA311070541GA SA311070562GA SA311070564GA SA311080728GA SA311080732GA SA311080734GA SA311090773GA SA311090778GA SA311090779GA SA311090780GA SA311090800GA SA311090804GA SA311090835GA SA311090842GA SA311090850GA SA311090853GA SA311090856GA SA311090861GA SA310455168GA SA311090887GA SA311111661GA SA311111672GA SA311122134GA SA311132525GA SA311142978GA SA311142961GA SA311122057GA SA311142747GA SA311142934GA SA311153204GA SA311266994GA SA311142711GA SA311163377GA SA311163385GA SA311163396GA SA311195063GA SA311195065GA SA311195114GA SA311256701GA SA311184518GA SA311184519GA SA311184544GA SA311184554GA SA311184558GA SA311184559GA SA311184560GA SA311184564GA SA311184571GA SA311184572GA SA311184581GA SA311184582GA SA311184584GA SA311184590GA SA311184592GA SA311184608GA SA311184610GA SA311184611GA SA311184612GA SA311174238GA SA311174242GA SA311236093GA SA311236096GA SA311184362GA SA311184532GA SA311184545GA SA311184556GA SA311184565GA SA311184567GA SA311184568GA SA311184591GA SA311184597GA SA311184599GA SA311184603GA SA311184606GA SA311195198GA SA311205470GA SA311121969GA SA311121972GA SA311121974GA SA311121976GA SA311121988GA SA311122006GA SA311132539GA SA311132607GA SA311132668GA SA311142807GA SA311163458GA SA311163467GA SA311163485GA SA311163553GA SA311174067GA SA311184438GA SA311184474GA SA311184498GA SA311205459GA SA311205468GA SA311205485GA SA311205493GA SA311205496GA SA311205497GA SA311225985GA SA311236289GA SA311236311GA SA311236333GA SA311174167GA SA311184832GA SA311266884GA SA311205394GA SA311205418GA SA311236051GA SA311236352GA SA311236357GA SA311236359GA SA311236366GA SA311236368GA SA311236377GA SA311246469GA SA311246511GA SA311317644GA SA311327870GA SA311327871GA SA311327876GA SA311236019GA SA311236027GA SA311236033GA SA311236059GA SA311236173GA SA311236181GA SA311236266GA SA311236272GA SA311246504GA SA311266935GA SA311266941GA SA311327880GA SA311307439GA SA311458213GA SA311458259GA SA311458342GA SA311458349GA SA311458359GA SA311468386GA SA311468390GA SA311468400GA SA311246421GA SA311307431GA SA311307509GA SA311307580GA SA311307594GA SA311317666GA SA311317706GA SA311317711GA SA311317715GA SA311317717GA SA311317719GA SA311317727GA SA311317728GA SA311317739GA SA311317741GA SA311317742GA SA311256818GA SA312018759GA SA311297241GA SA311317763GA SA311297313GA SA311317673GA SA311317688GA SA311327900GA SA311327911GA SA311327913GA SA311327932GA SA311337993GA SA311337996GA SA311338003GA SA312029048GA SA312028911GA SA311428032GA SA312039088GA SA312039132GA SA311142681GA SA311142691GA SA311142701GA SA311142704GA SA311142715GA SA311153057GA SA312069820GA SA312069825GA SA312069831GA SA312069833GA SA312069848GA SA312069851GA SA312069853GA SA312049322GA SA312049337GA SA312049338GA SA312049343GA SA312049349GA SA312049352GA SA312049455GA SA312059482GA SA312059485GA SA312059488GA SA312059499GA SA312080621GA SA312079951GA SA312080562GA SA312080565GA SA312080651GA SA312080737GA SA312080745GA SA312080774GA SA312080776GA SA312080784GA SA312080799GA SA312080815GA SA312080823GA SA312080828GA SA312080838GA SA312080847GA SA312080849GA SA312080851GA SA312080853GA SA312080857GA SA312080859GA SA312080862GA SA312080863GA SA312080866GA SA312080897GA SA312080915GA SA312070118GA SA312070134GA SA312080865GA SA312091100GA SA312069645GA SA312069658GA SA312069681GA SA312069727GA SA312069735GA SA312069738GA SA312069741GA SA312069747GA SA312069748GA SA312069752GA SA312069756GA SA312069763GA SA312069774GA SA312091093GA SA312091101GA SA312069744GA SA312069771GA SA312121760GA SA312121761GA SA312121763GA SA312121766GA SA312121769GA SA312121770GA SA312121773GA SA312121775GA SA312121780GA SA312121795GA SA312121808GA SA312121815GA SA312121816GA SA312121820GA SA312121823GA SA312121850GA SA312121852GA SA312091264GA SA312070340GA SA312122011GA SA312122021GA SA310382574GA SA310361790GA SA310361872GA SA310372074GA SA310372081GA SA310372083GA SA310372089GA SA310372102GA SA310372129GA SA310372170GA SA310372172GA SA310372191GA SA310382361GA SA310382362GA SA310372092GA SA310372094GA SA310382546GA SA310382582GA SA310382589GA SA310382691GA SA310382697GA SA310382698GA SA310382651GA SA310403437GA SA310423813GA SA310403317GA SA310403385GA SA310403401GA SA310403438GA SA310403445GA SA310423995GA SA310423995GAMKE SA310434102GA SA310434168GA SA310455067GA SA310455134GA SA310475452GA SA310475471GA SA310475520GA SA310475535GA SA310475538GA SA310475541GA SA310475546GA SA310475547GA SA310475548GA SA310475549GA SA310475550GA SA310475552GA SA310475556GA SA310475565GA SA310475567GA SA310475568GA SA310475569GA SA310475571GA SA310475574GA SA310475576GA SA310475577GA SA310475580GA SA310475584GA SA310475587GA SA310475588GA SA310475590GA SA310475593GA SA310475594GA SA310475595GA SA310403344GA SA310403345GA SA310403376GA SA310403386GA SA310403400GA SA310403406GA SA310403425GA SA310403446GA SA310423760GA SA310423762GA SA310423764GA SA310423765GA SA310423775GA SA310423783GA SA310423915GA SA310423981GA SA310423983GA SA310331079GA SA310331088GA SA310361775GA SA310361782GA SA310361793GA SA310361806GA SA310361849GA SA310361856GA SA310361883GA SA310361885GA SA310372071GA SA310372086GA SA310372109GA SA310372119GA SA310372127GA SA310372128GA SA310372131GA SA310372181GA SA310372211GA SA310382318GA SA310382320GA SA310382324GA SA310382328GA SA310382337GA SA310382367GA SA310382373GA SA310382525GA SA310382526GA SA310382549GA SA310382551GA SA310382564GA SA310382567GA SA310382594GA SA310382595GA SA310382597GA SA310382666GA SA310382669GA SA310382676GA SA310382678GA SA310382683GA SA310382688GA SA310382694GA SA310382702GA SA310382703GA SA310382772GA SA310382781GA SA310382788GA SA310382789GA SA310382794GA SA310382799GA SA310382804GA SA310382807GA SA310382810GA SA310382811GA SA310382815GA SA310382841GA SA310392851GA SA310392852GA SA310392854GA SA310392856GA SA310392857GA SA310392860GA SA310392862GA SA310392865GA SA310392867GA SA310392875GA SA310392877GA SA310392878GA SA310392880GA SA310392881GA SA310392883GA SA310392885GA SA310392888GA SA310392890GA SA310392892GA SA310392896GA SA310392906GA SA310392908GA SA310392913GA SA310392916GA SA310392918GA SA310392920GA SA310392921GA SA310392922GA SA310392926GA SA310392927GA SA310392928GA SA310392929GA SA310392932GA SA310393078GA SA310393079GA SA310393081GA SA310393082GA SA310393085GA SA310393087GA SA310393091GA SA310393093GA SA310393098GA SA310393099GA SA310393100GA SA310393101GA SA310393102GA SA310393103GA SA310393104GA SA310393105GA SA310393107GA SA310393110GA SA310393111GA SA310393113GA SA310393115GA SA310393127GA SA310393132GA SA310393133GA SA310393134GA SA310393135GA SA310393137GA SA310393181GA SA310393184GA SA310393189GA SA310393199GA SA310393201GA SA310393204GA SA310393205GA SA310393206GA SA310393207GA SA310393209GA SA310393210GA SA310393211GA SA310393214GA SA310393218GA SA310393220GA SA310393223GA SA310393225GA SA310393226GA SA310393228GA SA310393232GA SA310393240GA SA310393241GA SA310393244GA SA310393245GA SA310393247GA SA310393253GA SA310393256GA SA310393257GA SA310393264GA SA310393266GA SA310393273GA SA310393286GA SA310393289GA SA310393301GA SA310393307GA SA310260291GA SA310403388GA SA310403392GA SA310403403GA SA310403419GA SA310423897GA SA310423901GA SA310423902GA SA310423918GA SA310423982GA SA310423996GA SA310434050GA SA310434051GA SA310496225GA SA310496458GA SA310496468GA SA310496497GA SA310496527GA SA310496528GA SA310506848GA SA310506878GA SA310506994GA SA310506995GA SA310507030GA SA310507034GA SA310507042GA SA310507052GA SA310507053GA SA310507058GA SA310507065GA SA310507126GA SA310382711GA SA310506996GA SA310507044GA SA310507060GA SA310507068GA SA310507098GA SA310507099GA SA310507103GA SA310507142GA SA310507411GA SA310496374GA SA310496385GA SA310506773GA SA310506809GA SA310506828GA SA310506835GA SA310506855GA SA310507224GA SA310506796GA SA310506817GA SA310506818GA SA310506824GA SA310506826GA SA310506837GA SA310506840GA SA310506844GA SA310507181GA SA310507225GA SA310507235GA SA310507240GA SA310507241GA SA310507243GA SA310507244GA SA310507246GA SA310507279GA SA310507281GA SA310507314GA SA310507351GA SA310507373GA SA310507374GA SA310434131GA SA310434174GA SA310434190GA SA310434192GA SA310506757GA SA310506772GA SA310506778GA SA310506782GA SA310507152GA SA310507153GA SA310507166GA SA310507170GA SA310507175GA SA310507179GA SA310507182GA SA310507247GA SA310506827GA SA310507188GA SA310507220GA SA310507271GA SA310507278GA SA310517663GA SA310518200GA SA310518213GA SA310506819GA SA310506830GA SA310507185GA SA310507190GA SA310507227GA SA310507231GA SA310507232GA SA310507236GA SA310507260GA SA310518454GA SA310528812GA SA310518550GA SA310528772GA SA310528822GA SA310528904GA SA310382529GA SA310392899GA SA310403394GA SA310495913GA SA310507223GA SA310518381GA SA310518419GA SA310518421GA SA310518443GA SA310518467GA SA310518498GA SA310518507GA SA310518508GA SA310518510GA SA310518511GA SA310518536GA SA310518544GA
SA310518545GA SA310518558GA SA310528734GA SA310528765GA SA310528766GA SA310528790GA SA310528905GA SA310528907GA SA310528909GA SA310528910GA SA310528913GA SA310528919GA SA310528926GA SA310528929GA SA310528931GA SA310528934GA SA310528938GA SA310528946GA SA310528950GA SA310528958GA SA310529064GA SA310529068GA SA310529070GA SA310529072GA SA310529075GA SA310529076GA SA310529082GA SA310529084GA SA310529091GA SA310529092GA SA310529100GA SA310529117GA SA310529118GA SA310529119GA SA310518375GA SA310518413GA SA310518418GA SA310518456GA SA310518549GA SA310528726GA SA310528751GA SA310528754GA SA310528769GA SA310528806GA SA310528916GA SA310529062GA SA310529063GA SA310529066GA SA310529074GA SA310529080GA SA310529083GA SA310529085GA SA310529109GA SA310529110GA SA310260227GA SA310506876GA SA310507471GA SA310507540GA SA310528881GA SA310528982GA SA310528989GA SA310260317GA SA310260356GA SA310413563GA SA310424013GA SA310434201GA SA310475698GA SA310496279GA SA310496302GA SA310496648GA SA310507013GA SA310507532GA SA310518620GA SA310518680GA SA310518682GA SA310528853GA SA310529026GA SA310529055GA SA310529056GA SA310529147GA SA310529155GA SA311039375GA SA311039438GA SA311039440GA SA311039448GA SA311039470GA SA311060005GA SA311060066GA SA311060164GA SA311060207GA SA311060001GA SA311060003GA SA311060006GA SA311060040GA SA311060043GA SA311060044GA SA311060099GA SA311069636GA SA311069685GA SA311069687GA SA311069742GA SA311069746GA SA311069749GA SA311069750GA SA311069751GA SA311060185GA SA311060189GA SA311060205GA SA311060340GA SA311060347GA SA311060373GA SA311060388GA SA311070514GA SA311070543GA SA311070558GA SA311080604GA SA311080630GA SA311080650GA SA311080654GA SA311080659GA SA311060149GA SA311060227GA SA311060267GA SA311060291GA SA311080602GA SA311060016GA SA311060071GA SA311060093GA SA310382837GA SA310434072GA SA311060095GA SA311060110GA SA311091012GA SA310528829GA SA311069669GA SA311069680GA SA311069739GA SA311069741GA SA311069766GA SA311069815GA SA311069859GA SA311069862GA SA311069864GA SA311069865GA SA311069866GA SA311090948GA SA311090973GA SA311091029GA SA311101312GA SA311101315GA SA311101329GA SA311101373GA SA311101378GA SA311060111GA SA311090913GA SA310507528GA SA311069821GA SA311069823GA SA311069825GA SA311069855GA SA311069868GA SA311132466GA SA311132473GA SA311132524GA SA311121888GA SA311101400GA SA311142872GA SA311142949GA SA311142951GA SA311142952GA SA311142959GA SA311142960GA SA311142962GA SA311142967GA SA311142975GA SA311111599GA SA311111606GA SA311132284GA SA311142970GA SA311153082GA SA311153121GA SA311153128GA SA311153131GA SA311153140GA SA311153151GA SA311163368GA SA311163370GA SA311163371GA SA311163389GA SA311163394GA SA311163397GA SA311163420GA SA311163897GA SA311142826GA SA311121919GA SA311121949GA SA311121957GA SA311121968GA SA311122106GA SA311132414GA SA311142695GA SA311142697GA SA311142706GA SA311153312GA SA311153325GA SA311153336GA SA311153354GA SA311153355GA SA311153360GA SA311163547GA SA311163583GA SA311163642GA SA311163707GA SA311163713GA SA311163715GA SA311163728GA SA311163733GA SA311163781GA SA311163808GA SA311174160GA SA311174162GA SA311174178GA SA311174179GA SA311174182GA SA311174198GA SA311174206GA SA311122244GA SA311132596GA SA311174201GA SA311174203GA SA311174235GA SA311174236GA SA311174237GA SA311174239GA SA311174241GA SA311174243GA SA311174245GA SA311174252GA SA311184371GA SA311184387GA SA311184399GA SA311184400GA SA311184401GA SA311184402GA SA311184406GA SA311184410GA SA311184413GA SA311184524GA SA311195250GA SA311195256GA SA311195258GA SA311195264GA SA311195266GA SA311195267GA SA311195268GA SA311195270GA SA311195277GA SA311195278GA SA311195283GA SA311195284GA SA311195285GA SA311195287GA SA311195288GA SA311205358GA SA311225977GA SA311225979GA SA311225981GA SA311225986GA SA311225988GA SA311225989GA SA311225991GA SA311225992GA SA311225994GA SA311225995GA SA311225997GA SA311225999GA SA311236080GA SA311236084GA SA311236086GA SA311236115GA SA311236119GA SA311236121GA SA311236122GA SA311163407GA SA311163409GA SA311163415GA SA311246521GA SA311246561GA SA311256682GA SA311256683GA SA311256684GA SA311256685GA SA311256686GA SA311256688GA SA311256690GA SA311256691GA SA311256692GA SA311256693GA SA311256696GA SA311256697GA SA311256698GA SA311256699GA SA311256700GA SA311256702GA SA311256704GA SA311256710GA SA311256712GA SA311256715GA SA311256764GA SA311256772GA SA311256777GA SA311256779GA SA311256782GA SA311256784GA SA311256794GA SA311132516GA SA311184632GA SA311195130GA SA311195134GA SA311195147GA SA311195156GA SA311195254GA SA311184520GA SA311184618GA SA311184621GA SA311184623GA SA311184635GA SA311184497GA SA311205325GA SA311174108GA SA311184754GA SA311184803GA SA311205329GA SA311205337GA SA311205416GA SA311205436GA SA311205441GA SA311287173GA SA311297262GA SA311297264GA SA311195131GA SA311195255GA SA311195271GA SA311297266GA SA311297271GA SA311297291GA SA311297328GA SA311297342GA SA311297349GA SA311297350GA SA311297352GA SA311297365GA SA311307523GA SA311205621GA SA311215858GA SA311225895GA SA311215683GA SA311215685GA SA311215693GA SA311215696GA SA311215700GA SA311215702GA SA311215714GA SA311215717GA SA311215723GA SA311215739GA SA311236302GA SA311327883GA SA311317661GA SA311317669GA SA311184847GA SA311184868GA SA311194918GA SA311205563GA SA311205565GA SA311205567GA SA311205568GA SA311205572GA SA311205575GA SA311205577GA SA311205579GA SA311205581GA SA311205582GA SA311205584GA SA311205585GA SA311205591GA SA311205593GA SA311205602GA SA311215809GA SA311215838GA SA311215694GA SA311236038GA SA311246494GA SA311225938GA SA311225974GA SA311236003GA SA311236034GA SA311327934GA SA311428011GA SA311428057GA SA311448173GA SA311448179GA SA311448181GA SA311448192GA SA311448195GA SA311448202GA SA311184670GA SA311184831GA SA311194923GA SA311194935GA SA311194957GA SA311194959GA SA311195009GA SA311205610GA SA311246459GA SA311246477GA SA311246486GA SA311246488GA SA311246489GA SA311246514GA SA311256746GA SA311256749GA SA311256754GA SA311256810GA SA311256827GA SA311256828GA SA311256845GA SA311256848GA SA311256852GA SA311256859GA SA311256860GA SA311256862GA SA311256817GA SA311246589GA SA311246610GA SA311246612GA SA311246630GA SA311246636GA SA311246661GA SA311246671GA SA311307458GA SA311307473GA SA311428090GA SA311448194GA SA311246446GA SA311246584GA SA311246623GA SA311246657GA SA311246668GA SA311287152GA SA311317691GA SA311317698GA SA311317759GA SA311317766GA SA311317774GA SA311327910GA SA311327920GA SA311428073GA SA311428094GA SA311458212GA SA311458220GA SA311458236GA SA311458353GA SA311458361GA SA311458366GA SA311468399GA SA311287160GA SA312018777GA SA312049273GA SA312049274GA SA312049380GA SA312049398GA SA312049404GA SA312049412GA SA312059501GA SA312059504GA SA312059505GA SA312059506GA SA312059508GA SA312059509GA SA312059529GA SA312059532GA SA312059578GA SA312059581GA SA312059584GA SA312059586GA SA312059587GA SA312059595GA SA312059599GA SA312069815GA SA312069817GA SA312069829GA SA312069847GA SA312069852GA SA312049382GA SA312069856GA SA312069858GA SA312069866GA SA312069867GA SA312069870GA SA312069871GA SA312069872GA SA312069873GA SA312069874GA SA312069876GA SA312069877GA SA312069878GA SA312069879GA SA312069880GA SA312069882GA SA312069883GA SA312069884GA SA312079888GA SA312079890GA SA312079891GA SA312079892GA SA312079894GA SA312079895GA SA312079897GA SA312079903GA SA312079907GA SA312079912GA SA312079914GA SA312079916GA SA312079921GA SA312079929GA SA312079930GA SA312079932GA SA312049425GA SA312049435GA SA312049439GA SA312049460GA SA312059477GA SA312059484GA SA312059487GA SA312059491GA SA312059492GA SA312059495GA SA312059497GA SA312059498GA SA312080658GA SA312080698GA SA312091056GA SA312069670GA SA312069680GA SA312069684GA SA312069686GA SA312069687GA SA312069726GA SA312069728GA SA312069730GA SA312069731GA SA312069732GA SA312069733GA SA312069736GA SA312069737GA SA312069739GA SA312069740GA SA312069745GA SA312069746GA SA312069750GA SA312069751GA SA312069753GA SA312069757GA SA312069758GA SA312069759GA SA312069762GA SA312069764GA SA312069765GA SA312069768GA SA312069769GA SA312069770GA SA312069772GA SA312091099GA SA312091102GA SA312091115GA SA312101458GA SA312069690GA SA312069696GA SA312069719GA SA312069776GA SA312069780GA SA312069782GA SA312069785GA SA312069805GA SA312069806GA SA312069808GA SA312069812GA SA312069814GA SA312070302GA SA312070344GA SA312090931GA SA312091216GA SA312111629GA SA312111640GA SA312111663GA SA312121701GA SA312121702GA SA312121741GA SA312121749GA SA312121750GA SA312121806GA SA312121848GA SA312069832GA SA312101396GA SA312101401GA SA312101402GA SA312101404GA SA312101405GA SA312101407GA SA312101411GA SA312101412GA SA312101414GA SA312101421GA SA312101433GA SA312101438GA SA312121809GA SA312121819GA SA312121871GA SA312121872GA SA312121875GA SA312121879GA SA312121895GA SA312121898GA SA312121963GA SA312121971GA SA312121978GA SA312121979GA SA312121986GA SA312121987GA SA312091291GA SA312091296GA SA312091299GA SA312091313GA SA312091314GA SA312091317GA SA312091321GA SA312091327GA SA312091328GA SA312091344GA SA312091350GA SA312091352GA SA312091353GA SA312091361GA SA312101369GA SA312101372GA SA312101374GA SA312101376GA SA312101377GA SA312101415GA SA312091305GA SA312101378GA SA312101379GA SA312101380GA SA312101390GA SA312101408GA SA312101410GA SA312111466GA SA312111470GA SA312111472GA SA312111484GA SA312111485GA SA310382623GA SA310382641GA SA310382705GA SA310382731GA SA310382735GA SA310382738GA SA310382753GA SA310382758GA SA310382767GA SA310382770GA SA310393141GA SA310413517GA SA310423646GA SA310423654GA SA310424026GA SA310475539GA SA310475573GA SA310475583GA SA310475615GA SA310495892GA SA310495903GA SA310495909GA SA310495912GA SA310495933GA SA310495934GA SA310495941GA SA310495946GA SA310495950GA SA310495954GA SA310495957GA SA310495973GA SA310495896GA SA310495897GA SA310495901GA SA310495904GA SA310495906GA SA310495932GA SA310495943GA SA310341375GA SA310495908GA SA310496381GA SA310496382GA SA310496391GA SA310496392GA SA310496397GA SA310496406GA SA310496408GA SA310496414GA SA310496419GA SA310496421GA SA310496422GA SA310496423GA SA310496426GA SA310496432GA SA310496448GA SA310496452GA SA310495911GA SA310495956GA SA310495959GA SA310496131GA SA310496145GA SA310496152GA SA310496209GA SA310496216GA SA310496280GA SA310496286GA SA310496301GA SA310496316GA SA310496356GA SA310496477GA SA310496530GA SA310495891GA SA310495894GA SA310496128GA SA310496380GA SA310496390GA SA310496398GA SA310496412GA SA310496416GA SA310496417GA SA310496427GA SA310496431GA SA310496433GA SA310506670GA SA310506689GA SA310506697GA SA310506722GA SA310506873GA SA310507118GA SA310506738GA SA310496297GA SA310496305GA SA310496307GA SA310496308GA SA310496311GA SA310496313GA SA310496343GA SA310496506GA SA310517604GA SA310517618GA SA310517621GA SA310517625GA SA310517626GA SA310517629GA SA310517632GA SA310517637GA SA310496134GA SA310496135GA SA310496144GA SA310496146GA SA310496210GA SA310496277GA SA310496293GA SA310496351GA SA310496446GA SA310496503GA SA310496510GA SA310496632GA SA310496646GA SA310506750GA SA310506753GA SA310506766GA SA310506800GA SA310506991GA SA310507003GA SA310507011GA SA310495893GA SA310495899GA SA310495900GA SA310495918GA SA310495942GA SA310507313GA SA310507324GA SA310507325GA SA310507353GA SA310507376GA SA310517681GA SA310518146GA
SA310518177GA SA310518182GA SA310518184GA SA310518186GA SA310518191GA SA310518192GA SA310518193GA SA310518195GA SA310518197GA SA310423805GA SA310517673GA SA310517676GA SA310517682GA SA310518306GA SA310518311GA SA310517844GA SA310517870GA SA310517871GA SA310518000GA SA310518150GA SA310518202GA SA310518216GA SA310518219GA SA310518220GA SA310518222GA SA310518229GA SA310518233GA SA310518234GA SA310518236GA SA310518240GA SA310496399GA SA310518486GA SA310518629GA SA310434052GA SA310507524GA SA310507538GA SA310507549GA SA310528791GA SA310528867GA SA310528872GA SA310528884GA SA311039378GA SA311039380GA SA311039384GA SA311039533GA SA311039544GA SA311060305GA SA311060310GA SA311060168GA SA311060213GA SA311060224GA SA311060229GA SA311060233GA SA311060252GA SA311060260GA SA311069724GA SA311069738GA SA311060326GA SA311060338GA SA311060345GA SA311060365GA SA311060369GA SA311060383GA SA311060398GA SA311060399GA SA311070553GA SA311080616GA SA311060331GA SA311060332GA SA311060335GA SA311060356GA SA311060371GA SA311060372GA SA311060375GA SA311060379GA SA311060380GA SA311060281GA SA311070556GA SA311080598GA SA311080651GA SA311080660GA SA311080713GA SA311080715GA SA311039532GA SA311060101GA SA311060105GA SA311060114GA SA311069790GA SA311069797GA SA311069805GA SA311069806GA SA311069808GA SA311069810GA SA311069829GA SA311069831GA SA311069832GA SA311069848GA SA311070557GA SA311070571GA SA311070573GA SA311070583GA SA311070584GA SA311070590GA SA311080647GA SA311080727GA SA311080736GA SA311080737GA SA311090799GA SA311090805GA SA311090831GA SA311090843GA SA311090845GA SA311090846GA SA311090855GA SA310454933GA SA310507102GA SA311060334GA SA311080607GA SA311080609GA SA311080629GA SA311080638GA SA311080648GA SA311080653GA SA311080745GA SA311090751GA SA311090754GA SA311090760GA SA311090770GA SA311090789GA SA311090802GA SA311090803GA SA311090869GA SA311090872GA SA311090876GA SA311090886GA SA311101097GA SA311101176GA SA311101183GA SA311101185GA SA311101186GA SA311101188GA SA311101194GA SA311101196GA SA311101205GA SA311101207GA SA311101209GA SA311101211GA SA311101215GA SA311101223GA SA311101255GA SA311101258GA SA311101259GA SA311101271GA SA311101287GA SA311101288GA SA311101296GA SA311101305GA SA311101397GA SA311070525GA SA311101270GA SA311101385GA SA311101390GA SA311101395GA SA311101404GA SA311101422GA SA311101553GA SA311111565GA SA311111579GA SA311122036GA SA311121869GA SA311121872GA SA311121895GA SA311122219GA SA311122222GA SA311142916GA SA311142932GA SA311153098GA SA311153102GA SA311153105GA SA311153125GA SA311153145GA SA311122043GA SA311122045GA SA311122059GA SA311122180GA SA311122195GA SA311132366GA SA311142930GA SA311153083GA SA311153106GA SA311153113GA SA311153116GA SA311153126GA SA311153127GA SA311153134GA SA311153143GA SA311153148GA SA311153158GA SA311153205GA SA311153211GA SA311153212GA SA311121699GA SA311121705GA SA311121709GA SA311121711GA SA311121720GA SA311121726GA SA311132566GA SA311142915GA SA311142919GA SA311142933GA SA311142955GA SA311142980GA SA311153135GA SA311121861GA SA311121889GA SA311121911GA SA311142908GA SA311153097GA SA311153099GA SA311153175GA SA311153227GA SA311163975GA SA311225996GA SA311225998GA SA311236108GA SA311236114GA SA311236117GA SA311236120GA SA311246410GA SA311205442GA SA311205444GA SA311205462GA SA311184386GA SA311184530GA SA311184553GA SA311184576GA SA311184641GA SA311184646GA SA311184648GA SA311184649GA SA311184651GA SA311184660GA SA311184661GA SA311195146GA SA311195153GA SA311195154GA SA311195158GA SA311195167GA SA311195252GA SA311195257GA SA311205366GA SA311266897GA SA311266900GA SA311307505GA SA311317725GA SA311174185GA SA311184604GA SA311184650GA SA311195095GA SA311195138GA SA311195152GA SA311195155GA SA311195159GA SA311195249GA SA311195251GA SA311195265GA SA311205340GA SA311205342GA SA311205344GA SA311205346GA SA311205347GA SA311205349GA SA311205351GA SA311205352GA SA311205354GA SA311205361GA SA311205365GA SA311205368GA SA311205374GA SA311205375GA SA311205377GA SA311205380GA SA311205382GA SA311205383GA SA311205384GA SA311205385GA SA311205387GA SA311205388GA SA311205389GA SA311246544GA SA311256689GA SA311256707GA SA311256711GA SA311256713GA SA311256781GA SA311256793GA SA311266873GA SA311266876GA SA311266892GA SA311266910GA SA311266993GA SA311266995GA SA311266996GA SA311266998GA SA311266999GA SA311267001GA SA311267002GA SA311267005GA SA311287110GA SA311307447GA SA311317649GA SA311327833GA SA311256765GA SA311256770GA SA311287105GA SA311287107GA SA311287121GA SA311287122GA SA311287125GA SA311287127GA SA311287129GA SA311287133GA SA311287137GA SA311287138GA SA311287146GA SA311287147GA SA311287149GA SA311287172GA SA311287181GA SA311297261GA SA311297268GA SA311297269GA SA311297282GA SA311297283GA SA311297284GA SA311297290GA SA311205401GA SA311215804GA SA311215846GA SA311225898GA SA311225901GA SA311225910GA SA311225913GA SA311225914GA SA311225915GA SA311225916GA SA311225917GA SA311225918GA SA311225920GA SA311225922GA SA311225925GA SA311225926GA SA311225931GA SA311225932GA SA311225934GA SA311225935GA SA311225936GA SA311225937GA SA311225939GA SA311225940GA SA311225941GA SA311225943GA SA311225946GA SA311225949GA SA311225950GA SA311225956GA SA311225958GA SA311225959GA SA311225961GA SA311225962GA SA311225963GA SA311225964GA SA311225965GA SA311225967GA SA311225968GA SA311225969GA SA311225971GA SA311225972GA SA311225973GA SA311225975GA SA311236005GA SA311236009GA SA311236010GA SA311236012GA SA311236017GA SA311236030GA SA311236037GA SA311236049GA SA311236055GA SA311236058GA SA311236062GA SA311236190GA SA311236191GA SA311236237GA SA311236243GA SA311236250GA SA311236252GA SA311236254GA SA311236267GA SA311236268GA SA311236283GA SA311236284GA SA311236350GA SA311236358GA SA311236362GA SA311236365GA SA311236374GA SA311236379GA SA311246453GA SA311246457GA SA311246479GA SA311246510GA SA311246512GA SA311256871GA SA311266915GA SA311266919GA SA311266946GA SA311266951GA SA311508641GA SA311508642GA SA311508644GA SA311508649GA SA311508651GA SA311508666GA SA311508671GA SA311508645GA SA311508658GA SA311508660GA SA311508661GA SA311508667GA SA311508670GA SA311508676GA SA311508677GA SA311508678GA SA311508679GA SA311508680GA SA311518682GA SA311518684GA SA311518685GA SA311518688GA SA311518692GA SA311518694GA SA312018745GA SA312018747GA SA312018752GA SA312018753GA SA312018756GA SA312018757GA SA312018763GA SA312018764GA SA312018765GA SA312018767GA SA312018773GA SA312018774GA SA312018775GA SA312018776GA SA312018838GA SA312018842GA SA312018849GA SA312018882GA SA312018891GA SA312018892GA SA312018900GA SA312018903GA SA312018907GA SA311508669GA SA312028999GA SA312029008GA SA312029018GA SA312029019GA SA312029051GA SA312029071GA SA311287164GA SA311327848GA SA311327852GA SA311428018GA SA311428043GA SA312018816GA SA312018825GA SA312018833GA SA312039214GA SA312039216GA SA312039217GA SA312039218GA SA312039219GA SA312039222GA SA312039226GA SA312039227GA SA312039229GA SA312069701GA SA312069718GA SA312080394GA SA312080627GA SA312080391GA SA312080393GA SA312080400GA SA312080408GA SA312080411GA SA312080421GA SA312080435GA SA312080492GA SA312080495GA SA312080499GA SA312080501GA SA312080502GA SA312080503GA SA312080504GA SA312080507GA SA312080511GA SA312080618GA SA312080628GA SA312080636GA SA312080641GA SA312080583GA SA312080603GA SA312069691GA SA312070045GA SA312070047GA SA312070051GA SA312070066GA SA312070072GA SA312070098GA SA312070103GA SA312070104GA SA312070108GA SA312070117GA SA312070119GA SA312070121GA SA312070122GA SA312070125GA SA312070129GA SA312070140GA SA312080835GA SA312080818GA SA312080589GA SA312080680GA SA312080709GA SA312080712GA SA312080718GA SA312080726GA SA312080821GA SA312080846GA SA312080855GA SA312069777GA SA312070341GA SA312070345GA SA312070348GA SA312070360GA SA312111511GA SA312111532GA SA312091267GA SA312091273GA SA312091292GA SA312111653GA SA312121756GA SA312121765GA SA312121801GA SA312121807GA SA312121832GA SA312121846GA SA312121858GA SA312121896GA SA312121996GA SA312122010GA SA312122012GA SA312122020GA SA312122022GA SA312122027GA SA312132117GA SA312132133GA SA312121868GA SA310310963GA SA310506742GA SA310507498GA SA310517578GA SA310517700GA SA310517703GA SA310517715GA SA310517759GA SA310517775GA SA310517788GA SA312049330GA SA312049335GA SA312049364GA SA312080595GA SA312080611GA SA312121919GA SA310310959GA SA310444314GA SA310321060GA SA310444498GA SA310517709GA SA310528892GA SA311039379GA SA311039461GA SA311070580GA SA311090767GA SA311090823GA SA311090828GA SA311090857GA SA311101252GA SA311101254GA SA311101257GA SA311101260GA SA311101248GA SA311132575GA SA311142751GA SA311132551GA SA311163885GA SA311184600GA SA311184398GA SA311184514GA SA311184575GA SA311184593GA SA311184595GA SA311163912GA SA311163874GA SA311163921GA SA311184372GA SA311184534GA SA311297357GA SA311468381GA SA311307536GA SA311468392GA SA311317722GA SA311508562GA SA312018872GA SA311246467GA SA312049351GA SA312049328GA SA312049340GA SA312049355GA SA312059470GA SA312059545GA SA312080493GA SA312080516GA SA312080633GA SA312080682GA SA310423797GA SA310434083GA SA310434143GA SA310434162GA SA310361959GA SA310382398GA SA310382502GA SA310423595GA SA310423676GA SA310423695GA SA310434209GA SA310454985GA SA310455035GA SA310465321GA SA310465322GA SA310465326GA SA310465327GA SA310465328GA SA310465330GA SA310465331GA SA310465332GA SA310465333GA SA310465334GA SA310465338GA SA310465339GA SA310465342GA SA310465343GA SA310465346GA SA310465347GA SA310465380GA SA310465381GA SA310465397GA SA310465398GA SA310465403GA SA310465410GA SA310465411GA SA310465412GA SA310465420GA SA310465427GA SA310465431GA SA310465432GA SA310465433GA SA310465435GA SA310465438GA SA310465439GA SA310465441GA SA310465445GA SA310465446GA SA310465323GA SA310465325GA SA310465329GA SA310465341GA SA310465419GA SA310465443GA SA310475487GA SA310475515GA SA310475516GA SA310475518GA SA310475544GA SA310475545GA SA310475554GA SA310475572GA SA310475599GA SA310475601GA SA310475603GA SA310475604GA SA310475606GA SA310475607GA SA310475614GA SA310475619GA SA310475621GA SA310475622GA SA310475623GA SA310475624GA SA310475625GA SA310475627GA SA310475633GA SA310475635GA SA310475636GA SA310475640GA SA310475641GA SA310475642GA SA310475643GA SA310475645GA SA310475646GA SA310475647GA SA310475648GA SA310475650GA SA310475651GA SA310475652GA SA310475653GA SA310475654GA SA310475655GA SA310475656GA SA310475657GA SA310475658GA SA310475660GA SA310475661GA SA310475662GA SA310475664GA SA310475665GA SA310475666GA SA310475667GA SA310475668GA SA310475669GA SA310475670GA SA310475671GA SA310475672GA SA310475674GA SA310475675GA SA310475676GA SA310475677GA SA310475678GA SA310475679GA SA310475680GA SA310475683GA SA310475688GA SA310475689GA SA310475690GA SA310475691GA SA310475692GA SA310475693GA SA310475694GA SA310475699GA SA310475708GA SA310475713GA SA310475715GA SA310475716GA SA310475720GA SA310475721GA SA310475722GA SA310475723GA SA310475724GA SA310475725GA SA310475727GA SA310475729GA SA310475731GA SA310475735GA SA310475736GA SA310475737GA SA310475738GA SA310475739GA SA310475740GA SA310475741GA SA310475742GA SA310475743GA SA310475744GA SA310475745GA SA310475746GA SA310475747GA SA310475748GA SA310475749GA SA310475750GA SA310475751GA SA310475752GA SA310475753GA SA310475754GA SA310485755GA SA310485756GA SA310485757GA SA310485758GA
SA310485759GA SA310485760GA SA310485762GA SA310485764GA SA310485765GA SA310485767GA SA310485771GA SA310485773GA SA310485774GA SA310485775GA SA310485776GA SA310485777GA SA310485778GA SA310485779GA SA310485780GA SA310485781GA SA310485783GA SA310485784GA SA310485787GA SA310485797GA SA310485798GA SA310485799GA SA310485802GA SA310485815GA SA310485821GA SA310485826GA SA310485835GA SA310485841GA SA310485844GA SA310485846GA SA310485850GA SA310485857GA SA310485860GA SA310485868GA SA310485869GA SA310485871GA SA310485874GA SA310485878GA SA310485884GA SA310485885GA SA310485888GA SA310403342GA SA310403373GA SA310403375GA SA310403397GA SA310403404GA SA310403408GA SA310403416GA SA310403417GA SA310403424GA SA310403444GA SA310423757GA SA310423776GA SA310423786GA SA310423789GA SA310423810GA SA310423815GA SA310423817GA SA310423890GA SA310434135GA SA310434156GA SA310434172GA SA310434187GA SA310496138GA SA310496149GA SA310496151GA SA310496206GA SA310496207GA SA310496357GA SA310496616GA SA310496620GA SA310496629GA SA310496300GA SA310496346GA SA310496355GA SA310496508GA SA310507134GA SA310507254GA SA310507465GA SA310507501GA SA310517714GA SA310517726GA SA310517730GA SA310517735GA SA310496155GA SA310496645GA SA310507105GA SA310507515GA SA310507545GA SA311039381GA SA311039386GA SA311039395GA SA311039396GA SA311039397GA SA311039402GA SA311039403GA SA311101282GA SA311101178GA SA311101181GA SA311101189GA SA311101213GA SA311101243GA SA311101386GA SA311101389GA SA311101408GA SA311101410GA SA311101418GA SA311101384GA SA311101393GA SA311101394GA SA311101396GA SA311101420GA SA311174169GA SA311174184GA SA311174188GA SA311174218GA SA311174220GA SA311174228GA SA311174234GA SA311184517GA SA311184538GA SA311184539GA SA311184547GA SA311184562GA SA311184552GA SA311184589GA SA311184615GA SA311184617GA SA311039491GA SA311060147GA SA311060173GA SA311060204GA SA311060280GA SA311327872GA SA311039460GA SA311039482GA SA311142948GA SA311327822GA SA311327868GA SA311327886GA SA311327892GA SA312018771GA SA311307540GA SA311307554GA SA311307562GA SA311317617GA SA311317622GA SA311317633GA SA311317675GA SA312039150GA SA312039167GA SA312039171GA SA312039182GA SA312039186GA SA312039187GA SA312039189GA SA312039192GA SA312039200GA SA312039203GA SA312039209GA SA312039210GA SA312039212GA SA311060021GA SA311060050GA SA311225894GA SA311225899GA SA311225907GA SA311236351GA SA311246496GA SA310392900GA SA310392902GA SA310392924GA SA310393227GA SA310393242GA SA310393271GA SA310393282GA SA310393299GA SA310424036GA SA310455124GA SA310455139GA SA310455234GA SA310413536GA SA310528776GA SA310528780GA SA310528782GA SA310528783GA SA310528793GA SA310382817GA SA310393248GA SA310403351GA SA310403366GA SA310434079GA SA310507187GA SA310507259GA SA311039523GA SA311039527GA SA311069754GA SA311069773GA SA311069784GA SA311070591GA SA311060327GA SA311060364GA SA311060378GA SA311070515GA SA311069826GA SA311069858GA SA311069861GA SA311101314GA SA311101336GA SA311101362GA SA311101368GA SA311080612GA SA311080644GA SA311090851GA SA311090879GA SA311101251GA SA311101286GA SA311101293GA SA311122019GA SA311122030GA SA311122040GA SA311069846GA SA311111691GA SA311121754GA SA311121756GA SA311121757GA SA311121771GA SA311121774GA SA311121778GA SA311121784GA SA311121787GA SA311121810GA SA311121811GA SA311121818GA SA311121826GA SA311121834GA SA311121836GA SA311142772GA SA311101089GA SA311101093GA SA311101121GA SA311101169GA SA311101170GA SA311101173GA SA311101439GA SA311101443GA SA311101444GA SA311101452GA SA311111673GA SA311132569GA SA311132571GA SA311132580GA SA311132584GA SA311142875GA SA311142878GA SA311142887GA SA311111583GA SA311111584GA SA311111586GA SA311111595GA SA311111602GA SA311111615GA SA311163892GA SA311163719GA SA311163700GA SA311184442GA SA311184543GA SA311266966GA SA311267008GA SA311163626GA SA311163705GA SA311163778GA SA311205498GA SA311327867GA SA311468411GA SA311236211GA SA311327811GA SA311458357GA SA311468391GA SA311184696GA SA311297244GA SA312018860GA SA311317689GA SA311428015GA SA311428050GA SA311428084GA SA311428088GA SA312039178GA SA312039184GA SA312039191GA SA312039195GA SA312039206GA SA312039208GA SA312039215GA SA312039224GA SA312039233GA SA312049245GA SA312049257GA SA312049431GA SA312049450GA SA312049319GA SA312049440GA SA312069644GA SA312069647GA SA312069649GA SA312069650GA SA312069660GA SA312069662GA SA312049327GA SA312080521GA SA312080524GA SA312080569GA SA312080581GA SA312080593GA SA312080606GA SA312080609GA SA312080592GA SA312080602GA SA312080614GA SA312080615GA SA312059544GA SA312059546GA SA312059560GA SA312059567GA SA312059568GA SA312059605GA SA312059606GA SA312059607GA SA312059621GA SA312059622GA SA312059623GA SA312059626GA SA312059627GA SA312059634GA SA312091097GA SA312091098GA SA312091105GA SA312121791GA SA312091285GA SA312091295GA SA312091301GA SA312091303GA SA312091357GA SA312121830GA SA312121866GA SA312121878GA SA312121914GA SA312121916GA SA312121993GA SA312059531GA SA312070180GA SA312091015GA SA312091235GA SA312111539GA SA312122015GA SA312122017GA SA312122033GA SA312091354GA SA312121900GA SA312070252GA SA312111460GA SA312111467GA SA312121931GA SA312132104GA SA310413521GA SA310413526GA SA310413529GA SA310413530GA SA310413551GA SA310423598GA SA310423642GA SA310423643GA SA310423644GA SA310423648GA SA310423655GA SA310423717GA SA310423726GA SA310423733GA SA310423824GA SA310423826GA SA310423827GA SA310423835GA SA310423837GA SA310423840GA SA310423847GA SA310423850GA SA310423853GA SA310423922GA SA310423923GA SA310423924GA SA310423926GA SA310423930GA SA310423945GA SA310423963GA SA310423966GA SA310424009GA SA310413570GA SA310454993GA SA311060174GA SA311060183GA SA311060196GA SA311060320GA SA311060322GA SA311060339GA SA311070535GA SA311070559GA SA311070560GA SA311070589GA SA311080708GA SA311090768GA SA311090797GA SA311090821GA SA311090827GA SA311070531GA SA311060193GA SA311060186GA SA311060212GA SA311060219GA SA311060250GA SA311060256GA SA311060311GA SA311060316GA SA311060317GA SA311060341GA SA311060350GA SA311070516GA SA311070548GA SA311070592GA SA311121702GA SA311121736GA SA311121739GA SA311121745GA SA311121697GA SA311121719GA SA311121722GA SA310475467GA SA310506748GA SA311142950GA SA311142957GA SA311142965GA SA311142981GA SA311142983GA SA311142985GA SA311153153GA SA311153155GA SA311153167GA SA311153168GA SA311153174GA SA311153177GA SA311153178GA SA311153184GA SA311153193GA SA311153203GA SA311153183GA SA311153187GA SA311153198GA SA311153221GA SA311153224GA SA311215729GA SA311215733GA SA311215734GA SA311297251GA SA312049456GA SA312049465GA SA312069655GA SA312069676GA SA312080896GA SA311142787GA SA310528850GA SA311317624GA SA311317677GA SA311317684GA SA311317746GA SA311428031GA SA311448178GA SA311069748GA SA311069761GA SA311069838GA SA311132559GA SA311132567GA SA311132576GA SA311142889GA SA311142895GA SA311142899GA SA312080395GA SA312080772GA SA310496137GA SA310496160GA SA310496287GA SA310496318GA SA310496344GA SA310496473GA SA310496487GA SA310496491GA SA310496494GA SA310496495GA SA310496463GA SA310506663GA SA310506745GA SA310506845GA SA310506851GA SA310507019GA SA310507029GA SA310507051GA SA310507057GA SA310507097GA SA310507117GA SA310507148GA SA311132583GA SA312080842GA SA312080875GA SA310517831GA SA310517842GA SA310517876GA SA311060146GA SA311060181GA SA311195116GA SA311195150GA SA311236287GA SA311236316GA SA311236321GA SA311236324GA SA311039446GA SA311508659GA SA311508662GA SA311508672GA SA311327818GA SA312018874GA SA312018883GA SA312018884GA SA312018888GA SA312018890GA SA312018896GA SA312018899GA SA312018905GA SA312028908GA SA312028909GA SA312028918GA SA312028927GA SA312028935GA SA312028938GA SA312028940GA SA312028941GA SA312039213GA SA312059548GA SA312049464GA SA312080687GA SA312080694GA SA312080782GA SA312080807GA SA312080567GA SA312080742GA SA312080778GA SA312080790GA SA312080836GA SA312080843GA SA312080844GA SA312080856GA SA312080864GA SA312080880GA SA312080891GA SA312080898GA SA312080913GA SA312121865GA SA310465361GA SA310496628GA SA310506657GA SA310506661GA SA310506678GA SA310506702GA SA310506728GA SA310506864GA SA310506865GA SA310506872GA SA310506879GA SA310506880GA SA310506892GA SA310506893GA SA310506894GA SA310506896GA SA310506897GA SA310506992GA SA310506998GA SA310507001GA SA310507002GA SA310507004GA SA310507006GA SA310507008GA SA310507009GA SA310507010GA SA310507017GA SA310507020GA SA310507023GA SA310507032GA SA310507048GA SA310506846GA SA310506862GA SA310506871GA SA310506993GA SA310506997GA SA310507007GA SA310507028GA SA310507031GA SA310507037GA SA310507049GA SA310507055GA SA310506875GA SA310506737GA SA311090976GA SA312018748GA SA312018751GA SA312018758GA SA312018762GA SA312018768GA SA312018769GA SA310259942GA SA310382820GA SA310392853GA SA310392864GA SA310392868GA SA310392870GA SA310392871GA SA310392873GA SA310392879GA SA310392915GA SA310392919GA SA310392923GA SA310403391GA SA310403393GA SA310403395GA SA310403399GA SA310423763GA SA310423781GA SA310423808GA SA310423875GA SA310423880GA SA310423913GA SA310434142GA SA310434150GA SA310434154GA SA310434155GA SA310434158GA SA310434166GA SA310434178GA SA310434189GA SA310403421GA SA310403429GA SA310403443GA SA310403447GA SA310423892GA SA310423908GA SA310424045GA SA310434056GA SA310434104GA SA310434109GA SA310434111GA SA310434113GA SA310434179GA SA310518432GA SA310528942GA SA310372085GA SA310393084GA SA310393292GA SA310403352GA SA310403410GA SA310403432GA SA310403448GA SA310403450GA SA310434054GA SA310518404GA SA310518470GA SA310518519GA SA310518526GA SA310518532GA SA310528731GA SA310528741GA SA310528774GA SA310529052GA SA311287062GA SA311287068GA SA312152460GA SA312152465GA SA310475451GA SA310475460GA SA310475530GA SA310475532GA SA310475534GA SA311184578GA SA310260487GA SA310393002GA SA310424029GA SA310444262GA SA310455003GA SA310507086GA SA310507479GA SA310517577GA SA310518683GA SA310528894GA SA310529145GA SA311039390GA SA311060161GA SA311060162GA SA311060170GA SA311069722GA SA311069734GA SA311153086GA SA311153152GA SA311153156GA SA311153159GA SA311153166GA SA311153171GA SA311153179GA SA311153180GA SA311153191GA SA311153199GA SA311153200GA SA311153201GA SA311163372GA SA311163414GA SA311163422GA SA311163424GA SA311163425GA SA311163426GA SA311184594GA SA311184664GA SA311256705GA SA311184636GA SA311184528GA SA311266875GA SA311266891GA SA311266893GA SA311317645GA SA311297360GA SA311317646GA SA311317655GA SA311060197GA SA311060269GA SA311307513GA SA311307519GA SA311307521GA SA311307530GA SA311458340GA SA311458346GA SA311327936GA SA311327942GA SA311327812GA SA311458234GA SA312018749GA SA312018750GA SA312018760GA SA312018761GA SA312018766GA SA312018770GA SA312018772GA SA312018901GA SA312039074GA SA312039202GA SA312039204GA SA310372132GA SA310372171GA SA310372205GA SA310382533GA SA310382600GA SA310382785GA SA310382826GA SA310382850GA SA310403381GA SA310403389GA SA310403402GA SA310403407GA SA310423784GA SA310423788GA SA310423868GA SA310423878GA SA310434067GA SA310434096GA SA310434152GA SA310455115GA SA310455236GA SA310455260GA SA310455271GA SA310465297GA SA310465306GA SA310465311GA SA310465316GA SA311287017GA SA311287018GA SA311287035GA SA311287043GA SA311287052GA SA311287081GA SA311287086GA SA311287091GA SA311287096GA SA312079923GA SA312079939GA SA312079955GA SA312070186GA SA312070251GA SA311184370GA
SA311184525GA SA311184557GA SA311184613GA SA310403343GA SA310403353GA SA310403431GA SA310403439GA SA310423795GA SA310423876GA SA312069671GA SA312069673GA SA312069682GA SA311246454GA SA311246482GA SA311246503GA SA311256803GA SA311256836GA SA311256857GA SA311256865GA SA311317761GA SA310260002GA SA310413452GA SA310413460GA SA310413461GA SA310413467GA SA310413468GA SA310413470GA SA310413472GA SA310413474GA SA310413475GA SA310413477GA SA310413478GA SA310413480GA SA310413482GA SA310413483GA SA310413484GA SA310413487GA SA310413488GA SA310413489GA SA310413490GA SA310413492GA SA310413494GA SA310413496GA SA310413501GA SA310413502GA SA310413503GA SA310413506GA SA310413510GA SA310413512GA SA310413513GA SA310413515GA SA310413519GA SA310413520GA SA310413525GA SA310413527GA SA310413528GA SA310413531GA SA310413532GA SA310413533GA SA310413539GA SA310413540GA SA310413546GA SA310413547GA SA310413550GA SA310413552GA SA310413553GA SA310413554GA SA310413562GA SA310413566GA SA310413567GA SA310413568GA SA310413569GA SA310423574GA SA310423575GA SA310423578GA SA310423579GA SA310423590GA SA310423594GA SA310423597GA SA310423599GA SA310423601GA SA310423603GA SA310423605GA SA310423612GA SA310423620GA SA310423624GA SA310423625GA SA310423629GA SA310423630GA SA310423631GA SA310423632GA SA310423633GA SA310423634GA SA310423636GA SA310423638GA SA310423641GA SA310423645GA SA310423647GA SA310423660GA SA310423724GA SA310423725GA SA310423728GA SA310423729GA SA310423731GA SA310423734GA SA310423735GA SA310423823GA SA310423828GA SA310423830GA SA310423832GA SA310423833GA SA310423844GA SA310423848GA SA310423849GA SA310423854GA SA310423855GA SA310423857GA SA310423858GA SA310423859GA SA310423860GA SA310423861GA SA310423920GA SA310423925GA SA310423931GA SA310423932GA SA310423934GA SA310423936GA SA310423938GA SA310423939GA SA310423999GA SA310423990GA SA310434133GA SA310434141GA SA310434171GA SA310403441GA SA310455283GA SA310465304GA SA310465308GA SA310517815GA SA310517873GA SA310518277GA SA310518283GA SA310423581GA SA310423582GA SA310423852GA SA310423863GA SA310423865GA SA310423997GA SA310434206GA SA310434214GA SA310434215GA SA310434216GA SA310434217GA SA310434218GA SA310434219GA SA310434220GA SA310434221GA SA310434222GA SA310434223GA SA310434224GA SA310434225GA SA310434228GA SA310434229GA SA310434231GA SA310434232GA SA310434233GA SA310434234GA SA310434235GA SA310434236GA SA310434237GA SA310434239GA SA310434240GA SA310434241GA SA310434243GA SA310434244GA SA310434245GA SA310444246GA SA310444249GA SA310444250GA SA310444252GA SA310444254GA SA310444255GA SA310444257GA SA310444260GA SA310444261GA SA310444263GA SA310444264GA SA310444265GA SA310444276GA SA310444281GA SA310444283GA SA310444284GA SA310444305GA SA310444308GA SA310455082GA SA310455087GA SA310455107GA SA310455145GA SA310455165GA SA310455172GA SA310455175GA SA310465324GA SA310465335GA SA310465337GA SA310465348GA SA310465349GA SA310465354GA SA310465368GA SA310465369GA SA310465370GA SA310465374GA SA310465385GA SA310465391GA SA310465400GA SA310465408GA SA310465414GA SA310465423GA SA310465437GA SA310518420GA SA310518426GA SA310518427GA SA310528809GA SA310518378GA SA310403449GA SA310434055GA SA310434075GA SA310434103GA SA311060002GA SA311060009GA SA310496126GA SA311101299GA SA311069690GA SA311069700GA SA311069703GA SA311090911GA SA311090949GA SA311090968GA SA311090979GA SA311090992GA SA311090999GA SA311101144GA SA311101156GA SA311101159GA SA311121997GA SA311122066GA SA311122111GA SA311132419GA SA311132420GA SA311132426GA SA311132429GA SA311132431GA SA311132437GA SA311132445GA SA311132446GA SA311132450GA SA311132462GA SA311132486GA SA311132487GA SA311132513GA SA311132520GA SA311132528GA SA311090935GA SA311090955GA SA311090972GA SA311090995GA SA311091040GA SA311101072GA SA311101073GA SA311101075GA SA311101092GA SA311101100GA SA311101103GA SA311101153GA SA311101160GA SA311111649GA SA311142677GA SA311142680GA SA311121841GA SA311121846GA SA311121971GA SA311121973GA SA311121979GA SA311121982GA SA311121986GA SA311121987GA SA311121989GA SA311121993GA SA311121994GA SA311121998GA SA311122002GA SA311122005GA SA311122009GA SA311122010GA SA311122071GA SA311122073GA SA311122093GA SA311122094GA SA311122101GA SA311122102GA SA311122112GA SA311122115GA SA311132485GA SA311132543GA SA311132605GA SA311132613GA SA311132614GA SA311132619GA SA311153232GA SA311153239GA SA311153268GA SA311153283GA SA311153284GA SA311153289GA SA311153314GA SA311163474GA SA311163481GA SA311163482GA SA311142683GA SA311142684GA SA311142688GA SA311142693GA SA311142698GA SA311142703GA SA311142729GA SA311142737GA SA311142741GA SA311142742GA SA311142744GA SA311142797GA SA311142855GA SA311143011GA SA311143017GA SA311143018GA SA311143020GA SA311153030GA SA311153038GA SA311153053GA SA311153306GA SA311153322GA SA311163432GA SA311163434GA SA311163444GA SA311163451GA SA311163456GA SA311163462GA SA311163463GA SA311163465GA SA311163471GA SA311163472GA SA311163475GA SA311163476GA SA311163477GA SA311163478GA SA311163486GA SA311163534GA SA311163536GA SA311163537GA SA311163538GA SA311163540GA SA311163544GA SA311163546GA SA311163548GA SA311163549GA SA311163554GA SA311163555GA SA311163558GA SA311163564GA SA311163581GA SA311163589GA SA311163591GA SA311163595GA SA311163614GA SA311163621GA SA311163636GA SA311163655GA SA311163702GA SA311163704GA SA311163712GA SA311163714GA SA311163717GA SA311163720GA SA311163722GA SA311163732GA SA311163780GA SA311163791GA SA311163803GA SA311163806GA SA311163811GA SA311163820GA SA311163829GA SA311163837GA SA311163839GA SA311163841GA SA311163848GA SA311163853GA SA311173999GA SA311174000GA SA311174003GA SA311174005GA SA311174006GA SA311174015GA SA311174017GA SA311174018GA SA311174021GA SA311174033GA SA311174044GA SA311205426GA SA311205437GA SA310529148GA SA310529153GA SA310539348GA SA311039433GA SA311060202GA SA311060221GA SA311060266GA SA311060273GA SA311060277GA SA311287204GA SA311297346GA SA311307515GA SA311307516GA SA311307517GA SA311307518GA SA311307520GA SA311307522GA SA311307525GA SA311307526GA SA311307527GA SA311307528GA SA311307533GA SA311307534GA SA311307535GA SA311307537GA SA311307595GA SA311317660GA SA311317668GA SA311317737GA SA311327823GA SA311327825GA SA311327874GA SA311448182GA SA311448183GA SA311448184GA SA311448185GA SA311448186GA SA311448187GA SA311448191GA SA311458215GA SA311458216GA SA311458227GA SA311458228GA SA311458232GA SA311458233GA SA311458235GA SA311458238GA SA311458242GA SA311458246GA SA311458247GA SA311458248GA SA311458250GA SA311458251GA SA311458254GA SA311458255GA SA311458261GA SA311458264GA SA311458265GA SA311458266GA SA311184843GA SA311184845GA SA311194890GA SA311194911GA SA311194924GA SA311246680GA SA311184716GA SA311468387GA SA311508577GA SA311518698GA SA311518711GA SA311518725GA SA312018873GA SA312018876GA SA312018877GA SA312018878GA SA312018879GA SA312018880GA SA312018885GA SA312018894GA SA312028912GA SA312028913GA SA312028920GA SA312028923GA SA312028924GA SA312028925GA SA312028930GA SA312028931GA SA312028934GA SA312028937GA SA312028939GA SA312028942GA SA312028943GA SA312028944GA SA312028945GA SA312028946GA SA312028947GA SA312028951GA SA312028952GA SA312028953GA SA312028954GA SA312028955GA SA312028956GA SA312028957GA SA312028958GA SA312028959GA SA312028960GA SA312028961GA SA312028962GA SA312028963GA SA312028964GA SA312028965GA SA312028968GA SA312028969GA SA312028970GA SA312028971GA SA312028972GA SA312028973GA SA312028974GA SA312028975GA SA312028976GA SA312028977GA SA312028981GA SA312028983GA SA312028985GA SA312029000GA SA312029013GA SA312029029GA SA312029037GA SA312029041GA SA312029042GA SA312029044GA SA312029049GA SA312029070GA SA312080870GA SA312080879GA SA312069784GA SA312079978GA SA312090918GA SA312090919GA SA312090920GA SA312090922GA SA312090924GA SA312090925GA SA312090926GA SA312090929GA SA312090930GA SA312090932GA SA312090933GA SA312090935GA SA312090936GA SA312090937GA SA312090938GA SA312090939GA SA312090940GA SA312090941GA SA312090943GA SA312090944GA SA312090945GA SA312090946GA SA312090947GA SA312090949GA SA312090950GA SA312090951GA SA312090952GA SA312090953GA SA312090954GA SA312090955GA SA312090956GA SA312090957GA SA312090958GA SA312090959GA SA312090960GA SA312090961GA SA312090962GA SA312090963GA SA312090964GA SA312090965GA SA312090968GA SA312090971GA SA312090973GA SA312090975GA SA312090976GA SA312090977GA SA312090979GA SA312090980GA SA312090984GA SA312090986GA SA312090991GA SA312090992GA SA312091008GA SA312091012GA SA312091016GA SA312070277GA SA312091005GA SA312101432GA SA312111464GA SA312111474GA SA312111475GA SA312111477GA SA312111478GA SA312111480GA SA312111481GA SA312111490GA SA312111494GA SA312111497GA SA312111498GA SA312111501GA SA312111592GA SA312111593GA SA312111594GA SA312111595GA SA312111596GA SA312111597GA SA312111599GA SA312111600GA SA312111601GA SA312111602GA SA312111603GA SA312111604GA SA312111605GA SA312111606GA SA312111607GA SA312111608GA SA312111609GA SA312111610GA SA312111611GA SA312111612GA SA312111613GA SA312111614GA SA312111615GA SA312111616GA SA312111617GA SA312111618GA SA312111619GA SA312111620GA SA312111621GA SA312111622GA SA312111623GA SA312111624GA SA312111625GA SA312111626GA SA312111627GA SA312111628GA SA312111634GA SA312111635GA SA312111637GA SA312111638GA SA312111639GA SA312111641GA SA310434126GA SA310434128GA SA310434129GA SA310260398GA SA310341378GA SA310423891GA SA310424041GA SA310434061GA SA310434080GA SA310434116GA SA310434184GA SA310434191GA SA310455136GA SA310455238GA SA310455253GA SA310455270GA SA310455274GA SA310403318GA SA310423754GA SA310423782GA SA310423870GA SA310423884GA SA310423894GA SA310423898GA SA310423905GA SA310423972GA SA310423984GA SA310423992GA SA310434047GA SA310434073GA SA310434086GA SA310434127GA SA310434130GA SA310434144GA SA310434149GA SA310434159GA SA310434161GA SA310434163GA SA310434164GA SA310434165GA SA310434167GA SA310434169GA SA310434175GA SA310434176GA SA310434177GA SA310434193GA SA310455068GA SA310455116GA SA310455231GA SA310455288GA SA310506756GA SA310506759GA SA310506761GA SA310506775GA SA310506781GA SA310506791GA SA310506792GA SA310507158GA SA310507221GA SA310507263GA SA310507268GA SA310507287GA SA310507290GA SA310507088GA SA310507391GA SA310507405GA SA310507425GA SA310507440GA SA310507446GA SA310507466GA SA310507472GA SA310507530GA SA310517727GA SA310517740GA SA310517741GA SA310518455GA SA310518478GA SA310506806GA SA310507206GA SA310518322GA SA311059579GA SA311059591GA SA311059604GA SA311069662GA SA311069672GA SA311069691GA SA310475513GA SA310475570GA SA311060121GA SA311060246GA SA311039521GA SA311069752GA SA311069756GA SA311070526GA SA311070530GA SA311122189GA SA311122198GA SA311122203GA SA311122220GA SA311132279GA SA311132286GA SA311101133GA SA311101172GA SA311111628GA SA311122096GA SA311132471GA SA311132497GA SA311132501GA SA311121910GA SA311132564GA SA311132568GA SA311142884GA SA311142893GA SA311142897GA SA311142902GA SA311142906GA SA311184380GA SA311184391GA SA311184393GA SA311195071GA SA311256703GA SA311266909GA SA311194896GA SA311195208GA SA311184570GA SA311205378GA SA311205386GA SA311508616GA SA312018853GA SA312018854GA SA312018855GA SA312018858GA SA312018859GA SA312049239GA SA312049240GA SA312049266GA SA312080469GA SA312080530GA SA312080547GA SA312080572GA SA312080576GA SA312080580GA SA312080767GA
SA312080829GA SA312121803GA SA310413485GA SA310413486GA SA310423653GA SA310423843GA SA310423928GA SA310423948GA SA310423955GA SA310423965GA SA310423968GA SA310424007GA SA310424011GA SA310424015GA SA310424016GA SA310424021GA SA310424024GA SA310321010GA SA310423687GA SA310423998GA SA310424005GA SA310424018GA SA310424020GA SA310424022GA SA310424023GA SA310260256GA SA310361677GA SA310372305GA SA310413523GA SA310434226GA SA310444267GA SA310444317GA SA310444318GA SA310444319GA SA310444320GA SA310444321GA SA310444356GA SA310444357GA SA310444359GA SA310444363GA SA310444488GA SA310444505GA SA310444508GA SA310444512GA SA310444515GA SA310444516GA SA310444525GA SA310444528GA SA310444529GA SA310444530GA SA310444535GA SA310444841GA SA310444855GA SA310444858GA SA310444866GA SA310475493GA SA310475500GA SA310475559GA SA310475560GA SA310475561GA SA310475563GA SA310475605GA SA310475659GA SA310475712GA SA310485769GA SA310485803GA SA310485817GA SA310485823GA SA310485831GA SA310485849GA SA310485852GA SA310485864GA SA310485865GA SA310485866GA SA310485872GA SA310485875GA SA310485887GA SA310454977GA SA310465344GA SA310475632GA SA310475711GA SA310485825GA SA310485827GA SA310485830GA SA310485837GA SA310485854GA SA310485855GA SA310485879GA SA310485886GA SA310496213GA SA310496217GA SA310496219GA SA310496234GA SA310496235GA SA310496239GA SA310496240GA SA310506659GA SA310506664GA SA310506666GA SA310506667GA SA310506671GA SA310506699GA SA310506705GA SA310506712GA SA310506847GA SA310506891GA SA310506669GA SA310506752GA SA310506850GA SA310506863GA SA310507033GA SA310496130GA SA310496133GA SA310496136GA SA310496150GA SA310496153GA SA310496158GA SA310496312GA SA310496320GA SA310506662GA SA310506677GA SA310506685GA SA310506686GA SA310506693GA SA310506701GA SA310506718GA SA310506723GA SA310507085GA SA310517562GA SA310517572GA SA310517575GA SA310517576GA SA310507510GA SA310507500GA SA310507120GA SA310507487GA SA310382624GA SA310434202GA SA310518622GA SA310518641GA SA310518644GA SA310529143GA SA310529144GA SA310529151GA SA310529154GA SA310529209GA SA310403377GA SA310403412GA SA310403430GA SA310423791GA SA310434060GA SA310434100GA SA310434101GA SA310434112GA SA310434117GA SA310259904GA SA310507348GA SA310518422GA SA310518489GA SA310518494GA SA310518496GA SA310518506GA SA310518525GA SA310539289GA SA310496298GA SA311060136GA SA311060138GA SA311060140GA SA311060167GA SA311060208GA SA311060230GA SA311069728GA SA311069731GA SA311069732GA SA311101275GA SA311069863GA SA311090900GA SA311090907GA SA311090918GA SA311090920GA SA311090933GA SA311091034GA SA311091035GA SA311091049GA SA311091053GA SA311101099GA SA311101114GA SA311101123GA SA311080633GA SA311091004GA SA311132288GA SA311122037GA SA311122038GA SA311090783GA SA311153208GA SA311163375GA SA311163379GA SA311163380GA SA311163381GA SA311163384GA SA311163391GA SA311163392GA SA311163395GA SA311163399GA SA311163413GA SA311163418GA SA311163419GA SA311163421GA SA311163887GA SA311163956GA SA311060156GA SA311060225GA SA311468393GA SA311468398GA SA311468401GA SA311468403GA SA310539359GA SA311327969GA SA311039434GA SA311458279GA SA311287095GA SA311287163GA SA311297230GA SA311297248GA SA311297252GA SA311297295GA SA311297296GA SA311297305GA SA311297308GA SA311297309GA SA311297312GA SA311297315GA SA311297317GA SA311297318GA SA311297319GA SA311297320GA SA311297323GA SA311307459GA SA311307477GA SA311307546GA SA311317609GA SA311317632GA SA311317674GA SA311317686GA SA311317744GA SA311317753GA SA311317756GA SA311317760GA SA311327856GA SA311327905GA SA311327914GA SA311327919GA SA311327925GA SA311337978GA SA311428023GA SA312039073GA SA312039116GA SA312039119GA SA312039124GA SA312039145GA SA312039146GA SA312039160GA SA312039170GA SA312039176GA SA312049417GA SA312070021GA SA312069723GA SA312079997GA SA312080832GA SA312070174GA SA312080722GA SA312080747GA SA312080766GA SA312080773GA SA312080793GA SA312080816GA SA312069828GA SA312091159GA SA312091175GA SA312091176GA SA312091177GA SA312091179GA SA312091188GA SA312091189GA SA312091193GA SA312091195GA SA312091198GA SA312091199GA SA312091205GA SA312091212GA SA312091214GA SA312091217GA SA312091228GA SA312070003GA SA312070012GA SA312070099GA SA312079954GA SA312079969GA SA311090891GA SA311297300GA SA311080674GA SA312079964GA SA312079973GA SA312079986GA SA312079990GA SA310393236GA SA310393259GA SA310393272GA SA310393275GA SA310393277GA SA310393288GA SA310403336GA SA310403337GA SA310413454GA SA310413455GA SA310413456GA SA310413457GA SA310413459GA SA310413462GA SA310413463GA SA310413464GA SA310413465GA SA310413466GA SA310413469GA SA310413508GA SA310413541GA SA310413542GA SA310413548GA SA310413549GA SA310413555GA SA310413557GA SA310413558GA SA310413564GA SA310413565GA SA310423585GA SA310423586GA SA310423588GA SA310423589GA SA310423591GA SA310423592GA SA310423602GA SA310423608GA SA310423609GA SA310423610GA SA310423615GA SA310423616GA SA310423623GA SA310423656GA SA310423659GA SA310423661GA SA310423664GA SA310423691GA SA310423740GA SA310423746GA SA310423752GA SA310423767GA SA310423768GA SA310496139GA SA310496141GA SA310496460GA SA310496465GA SA310496498GA SA310496515GA SA310518647GA SA310496157GA SA311101481GA SA311101484GA SA311101490GA SA311101492GA SA311101500GA SA311101501GA SA311101508GA SA311101509GA SA311101518GA SA311101524GA SA311101530GA SA311101533GA SA311101548GA SA311101551GA SA311111611GA SA311121704GA SA311121706GA SA311121710GA SA311121715GA SA311121716GA SA311069811GA SA311174209GA SA311174223GA SA311184377GA SA311184541GA SA311184577GA SA311184585GA SA311225978GA SA311225983GA SA311225984GA SA311225990GA SA311236063GA SA311236072GA SA311236073GA SA311236089GA SA311236091GA SA311236095GA SA311236099GA SA311236105GA SA311236107GA SA311236109GA SA311236112GA SA311236113GA SA311236118GA SA311266961GA SA311266986GA SA311184537GA SA311184645GA SA312059541GA SA312069641GA SA311184634GA SA311287145GA SA311307504GA SA311317705GA SA311317710GA SA311317712GA SA311317713GA SA311317723GA SA311317738GA SA311327805GA SA311327806GA SA311327813GA SA311327815GA SA311327828GA SA311327831GA SA311508587GA SA310393279GA SA310506691GA SA310507123GA SA310507140GA SA310507409GA SA310507429GA SA310507432GA SA310507449GA SA310507452GA SA310507469GA SA310507489GA SA310507529GA SA310507533GA SA310507534GA SA310507535GA SA310507536GA SA310507537GA SA310507539GA SA310507551GA SA310517567GA SA310517568GA SA310517581GA SA310517589GA SA310424032GA SA310444251GA SA310444258GA SA310444269GA SA310444270GA SA310444275GA SA310444278GA SA310444291GA SA310444292GA SA310444294GA SA310444301GA SA310444306GA SA310444312GA SA310444324GA SA310444325GA SA310444326GA SA310517624GA SA310518464GA SA310518499GA SA310518523GA SA310518527GA SA310518530GA SA310518535GA SA310518539GA SA310518541GA SA310518542GA SA310518551GA SA310518552GA SA310518553GA SA310518555GA SA310518557GA SA310529073GA SA310529079GA SA310529086GA SA310529090GA SA310529098GA SA310529099GA SA310529101GA SA310529113GA SA310361850GA SA310529093GA SA311090837GA SA311059582GA SA311060097GA SA311069640GA SA311069789GA SA311069807GA SA311069836GA SA311069837GA SA311069850GA SA311090787GA SA311090795GA SA309375462GA SA309385568GA SA309385595GA SA309385614GA SA309385630GA SA309385666GA SA309395719GA SA309395749GA SA309395771GA SA309395789GA SA311070549GA SA311101392GA SA311101398GA SA311101399GA SA311101406GA SA311101409GA SA311101413GA SA311101414GA SA311101423GA SA311101550GA SA311101552GA SA311111557GA SA311111561GA SA311111573GA SA311111577GA SA311090927GA SA311090946GA SA311090950GA SA311090951GA SA311090967GA SA311090989GA SA311091038GA SA311101076GA SA311101108GA SA311101115GA SA311101118GA SA311101120GA SA311122170GA SA311122176GA SA311122191GA SA311122194GA SA311122197GA SA311122232GA SA311122274GA SA311132281GA SA311132294GA SA311132302GA SA311132324GA SA309405831GA SA309405835GA SA311090987GA SA311069661GAMKE SA311069800GAMKE SA311069802GAMKE SA311101084GAMKE SA311101127GAMKE SA311101129GAMKE SA311101161GAMKE SA311101440GAMKE SA311101445GAMKE SA311101449GAMKE SA311101455GAMKE SA311122062GAMKE SA311122064GAMKE SA311122099GAMKE SA311122126GAMKE SA311122140GAMKE SA311122145GAMKE SA311122148GAMKE SA311122154GAMKE SA311122160GAMKE SA308398742GR SA308511625GR SA309092523GR SA309143016GR SA309283997GR SA309283999GR SA309304262GR SA309304272GR SA309365227GR SA309365232GR SA309365347GR SA309517572GR SA310028114GR SA310028140GR SA310149355GR SA310259951GR SA310260386GR SA310372203GR SA310382325GR SA310382326GR SA310393186GR SA310423753GR SA310423978GR SA310507320GR SA309283983GR SA309324548GR SA309334756GR SA309365225GR SA309365229GR SA309365237GR SA309375364GR SA309416005GR SA309476874GR SA309476893GR SA309497149GR SA309517559GR SA309517853GR SA310028149GR SA310260303GR SA310260308GR SA310260359GR SA310372167GR SA310382347GR SA310393263GR SA310517610GR SA310518235GR SA310518390GR SA310518423GR SA311091027GA SA308511487GR SA309304179GR SA309395794GR SA309395812GR SA309517830GR SA309527946GR SA310260522GR SA310454991GR SA310475637GR SA310518637GR SA311060244GR SA311060295GR SA309375454GR SA310028280GR SA309355023GR SA309365271GR SA309375473GR SA309385528GR SA309395746GR SA309436297GR SA309527923GR SA310139066GR SA310139244GR SA310517721GR SA309344865GR SA309375471GR SA309375482GR SA309436270GR SA309446441GR SA309446459GR SA309517639GR SA310290663GR SA310351552GR SA310392937GR SA309365234GR SA309477028GR SA308511483GR SA308511540GR SA309395705GR SA308388468GR SA308398817GR SA308409032GR SA308491168GR SA308501421GR SA308511464GR SA308511501GR SA308511512GR SA308511551GR SA309294116GR SA309304201GR SA309324342GR SA309354967GR SA309375456GR SA309395717GR SA309395754GR SA309405851GR SA309426121GR SA309446472GR SA309446487GR SA309446503GR SA309446505GR SA309476957GR SA309476976GR SA309517679GR SA309517707GR SA309517796GR SA309517816GR SA309517817GR SA309517823GR SA310139061GR SA310199565GR SA310219839GR SA310393139GR SA310413473GR SA310423680GR SA310423684GR SA311090776GR SA311142984GR SA308459843GR SA308511481GR SA309324411GR SA309354926GR SA309385561GR SA309395792GR SA309405845GR SA309446532GR SA309456540GR SA310485793GR SA311090781GR SA311468382GR SA310517646GR SA309446468GR SA309456546GR SA309456552GR SA310259884GR SA310260553GR SA310331249GR SA310496653GR SA310507400GR SA310517824GR SA311090882GR SA311132350GR SA311205492GR SA307220890GR SA307220904GR SA307220931GR SA307221013GR SA307221026GR SA307221168GR SA307241668GR SA307241677GR SA307241691GR SA307241708GR SA307251710GR SA307251729GR SA308186433GR SA308511496GR SA308511520GR SA309375441GR SA309405821GR SA309446484GR SA309446488GR SA309476970GR SA309517533GR SA310310883GR SA310382657GR SA310393073GR SA310423635GR SA310424012GR SA310444272GR SA310507110GR SA310507131GR SA310507396GR SA310507416GR SA310507426GR SA310507441GR SA310507447GR SA310517707GR SA311101219GR SA311205467GR SA309426039GR SA307251757GR SA310507146GR SA307190745GR SA307220898GR SA307220901GR SA307220983GR SA307221096GR SA307221114GR SA307221154GR SA307221183GR SA307231366GR SA307231476GR SA307241674GR SA307251713GR SA307251782GR SA307362392GR SA308084673GR SA308186372GR SA310506731GR SA310506734GR SA311111566GR SA311142752GR SA311142903GR SA311153188GR SA307030411GR SA307110560GR SA307241690GR SA307251735GR SA308459744GR
SA309122640GR SA309304244GR SA309365226GR SA309365230GR SA309365233GR SA309365235GR SA309365348GR SA309476885GR SA309476916GR SA309477022GR SA310372168GR SA310382331GR SA310434088GR SA310507275GR SA310517645GR SA310518168GR SA310518169GR SA310518170GR SA310518179GR SA310518207GR SA310518225GR SA310518477GR SA310529069GR SA311142802GR SA311163435GR SA311163460GR SA311163466GR SA311163726GR SA311184444GR SA307210819GR SA307220900GR SA307231376GR SA307231444GR SA307241586GR SA307251727GR SA307251743GR SA307291955GR SA310518715GR SA311236085GR SA311236092GR SA311266955GR SA311266976GR SA311266979GR SA311266982GR SA311266983GR SA311266991GR SA311267010GR SA310382839GR SA310507215GR SA311184448GR SA311184452GR SA311184455GR SA312091112GA SA312121826GA SA312122044GA SA312132129GA SA312152330GA SA312152394GA SA312152412GA SA312152456GA SA312152470GA SA312152483GA SA312152484GA SA312192599GA SA312192604GA SA312192617GA SA312192645GA SA312192650GA SA312192652GA SA312192653GA SA312192655GA SA312192658GA SA312192659GA SA312192660GA SA312192667GA SA312192714GA SA312192717GA SA312122083GA SA312132091GA SA312132148GA SA312132149GA SA312132151GA SA312132165GA SA312152319GA SA312152332GA SA312152335GA SA312152344GA SA312152347GA SA312152350GA SA312152361GA SA312152367GA SA312192654GA SA312192663GA SA312192669GA SA312192687GA SA307070450GR SA307200777GR SA307221081GR SA307221104GR SA307231370GR SA307251721GR SA307251747GR SA308166102GR SA309395710GR SA310393145GR SA310475537GR SA311266957GR SA311266974GR SA311266984GR SA311267003GR SA312079960GA SA312091208GA SA312121943GA SA312142250GA SA312152327GA SA312152345GA SA312152346GA SA312152351GA SA312152398GA SA312152410GA SA312152416GA SA312152420GA SA312152421GA SA312152426GA SA312152427GA SA312152430GA SA312152439GA SA312152440GA SA312152441GA SA312152445GA SA312152446GA SA312152447GA SA312152449GA SA312152450GA SA312152453GA SA312152454GA SA312152455GA SA312152457GA SA312152458GA SA312152459GA SA312152462GA SA312152464GA SA312152468GA SA312192670GA SA312192680GA SA312192683GA SA312192697GA SA312192698GA SA312069697GA SA312069711GA SA312070011GA SA312070025GA SA312070031GA SA312070036GA SA312079920GA SA312079946GA SA312079971GA SA312079983GA SA312079993GA SA312111673GA SA312121961GA SA312121965GA SA312121968GA SA312121969GA SA312121970GA SA312121977GA SA312122034GA SA312122045GA SA312122047GA SA312122053GA SA312122066GA SA312122070GA SA312122073GA SA312122074GA SA312122075GA SA312122077GA SA312122078GA SA312122088GA SA312132096GA SA312132099GA SA312132111GA SA312132112GA SA312132137GA SA312132169GA SA312152312GA SA312152316GA SA312152318GA SA312152329GA SA312152356GA SA312152360GA SA312152364GA SA312152368GA SA312152472GA SA312152481GA SA312152482GA SA312152496GA SA312152508GA SA312152515GA SA312152525GA SA312152533GA SA312152543GA SA312152544GA SA312152550GA SA312152556GA SA312152559GA SA312192656GA SA312192657GA SA312192664GA SA312192665GA SA312192668GA SA312192673GA SA312192675GA SA312192677GA SA312192685GA SA312192688GA SA312192689GA SA312192690GA SA312192691GA SA312192692GA SA312192693GA SA312192694GA SA312192699GA SA312192700GA SA312192702GA SA312192703GA SA312192704GA SA312192705GA SA312192706GA SA312192707GA SA312192708GA SA312192709GA SA312192710GA SA312192711GA SA312192712GA SA312192715GA SA312192716GA SA312192718GA SA312192719GA SA312192720GA SA312192721GA SA312192722GA SA312192725GA SA312192727GA SA312192728GA SA312192729GA SA312192731GA SA307221156GR SA307251726GR SA308084724GR SA309426041GR SA309446404GR SA309517567GR SA310506843GR SA310517652GR SA311132356GR SA311508572GR SA311508589GR SA312091068GA SA312101429GA SA312111500GA SA312111505GA SA312152478GA SA312152479GA SA312152485GA SA312152488GA SA312152493GA SA312152505GA SA312152511GA SA312152512GA SA312152513GA SA312152519GA SA312152520GA SA312152523GA SA312152526GA SA312152529GA SA312152530GA SA312152536GA SA312152538GA SA312152539GA SA312152545GA SA312152546GA SA312152554GA SA312192661GA SA312192662GA SA312192666GA SA312192672GA SA312192676GA SA312192678GA SA312192679GA SA312232826GA SA312232831GA SA312232834GA SA312232835GA |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
800-558-7044
|
Manufacturer Reason for Recall |
GE Healthcare has recently become aware of a potential safety issue due to intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG connector on your DASH, PDM, and/or Tram Module.
ECG and Respiration signal noise caused by DASH, PDM, and/or Tram ECG cable connector movement may result in reduced ECG, Respiration and arrhythmia detection performance.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
GE Healthcare "Urgent Medical Device Correction" letter dated April 3, 2013 to all affected customers. The letter was addressed to Healthcare Administrator / Risk manger, Chief of Nursing and Director of Biomedical Engineering. The letter described the Safety Issue, Product, Safety Instructions, Affected Product Details, and Product Correction. For questions contact Technical Support at 1-800-558-7044. |
Quantity in Commerce |
16,462 (8497 US, 7307 OUS, & 658 unlocated) |
Distribution |
Worldwide Distribution - USA Nationwide including DC and PR except ND and the countries of: ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOSNIA-HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, CHANA, GREECE, HONDURAS, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, NEPAL, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAK IA, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TUNSIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, BOLIVIA, HONG KONG, JAMAICA, PHILIPPINES, SOUTH KOREA, TRINIDAD TOBAGO, URUGUAY, VENEZUELA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MHX and Original Applicant = GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES
|
|
|
|