|
Class 2 Device Recall Siemens Healthcare |
|
Date Initiated by Firm |
September 17, 2012 |
Date Posted |
October 29, 2013 |
Recall Status1 |
Terminated 3 on April 01, 2015 |
Recall Number |
Z-0103-2014 |
Recall Event ID |
66166 |
510(K)Number |
K110277
|
Product Classification |
Electrode measurement, blood-gases (pco2, po2) and blood ph - Product Code CHL
|
Product |
Siemens RAPIDPoint¿¿ 405 system blood gas analyzer
Product PUsage: These systems are intended for near-patient and laboratory testing of blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The systems are capable of determining the following parameters: RAPIDPoint 400 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, Hct RAPIDPoint 405 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili |
Code Information |
All Serial Numbers |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics Inc 2 Edgewater Dr Norwood MA 02062-4637
|
For Additional Information Contact |
781-269-3000
|
Manufacturer Reason for Recall |
RAPIDPoint¿ 400 Series, RAPIDPoint¿ 500, or
RAPIDLab¿ 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample
|
FDA Determined Cause 2 |
Software Manufacturing/Software Deployment |
Action |
Siemens Healthcare sent an Urgent Field Safety Notification letter dated September 2013 via Federal Express, to affected Siemens domestic customers. Effected Siemens customers outside the US was provided a copy of the Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. Previously Siemens Healthcare had issued a Customer Notification on August 23, 2012. The letter identified the affected product, issue, information to mitigate the issue and actions to be taken. The firm will issue an upgrades software to users. For questions contact your local Customer Care Center - Technical Support Representative at 1-877-229-3711. |
Quantity in Commerce |
4494 units |
Distribution |
Worldwide Distribution - USA Nationwide and in the countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Brazil, Brunei, Bulgaria, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Fren.Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Mayotte, Netherlands, New Caledonia, New Zealand, Norway, P.R. China, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, U.A.E., United Kingdom, Vatikancity, Venezuela, Vietnam, |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = CHL and Original Applicant = SIEMENS HEALTHCARE DIAGNOSTICS
|
|
|
|