Class 2 Device Recall Integra Camino 1104 Intracranial Pressure (ICP) Monitoring |
|
Date Initiated by Firm |
September 10, 2013 |
Date Posted |
October 30, 2013 |
Recall Status1 |
Terminated 3 on June 04, 2014 |
Recall Number |
Z-0117-2014 |
Recall Event ID |
66243 |
510(K)Number |
K102875 K102875
|
Product Classification |
Device, monitoring, intracranial pressure - Product Code GWM
|
Product |
Brand Names: Integra¿ Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively.
Common Name: Intracranial Pressure and Pressure-Temperature Monitoring Kits Classification Name: Device, Monitoring, Intracranial Pressure 510(k) Number: K102875
The Camino 110-4 Intracranial Pressure monitoring catheters are sterile transducer-tipped pressure monitoring catheters that are used by neurosurgeons to rapidly determine and continuously monitor intracranial pressure or pressure/temperature. The Camino 110-4B catheter is used for Intracranial Pressure Monitoring in the parenchyma or subarachnoid space. The Camino 110-4G catheter is used for Post Craniotomy Subdural Pressure Monitoring. The Camino 110-4H catheter is intended for Intracranial Pressure Monitoring and Cerebrospinal fluid drainage. The Camino 110-4L catheter is intended for use with the Licox¿ Brain Oxygen Monitoring System to measure intracranial pressure in the parenchyma. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. |
Code Information |
Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. Lot Numbers: Catalog Number 110-4B: 3050RX195484, 3050RX198473, 305000205426, 3050RX197641, 3050RX201421, 305000205648, 305000205652, 3050RX199149, 3050RX201407, 3050RX199850, 305000205976, 3050RY189979, 3050RX200862, 305000206178, 3050RY193648, 3050RX200486, 3050RX181026, 3050RX185115, 3050RY195484, 305000207061, 305000207062, 3050RY198473, 305000207330, 305000207334, 3050RY197641, 3050RY200862, 305000207580, 305000207892, 3050RY200486, 3050RY181026, 3050RY185115, 3050RX207330, 3050RX207334, 3050RX207580, 305000208008, 305000208510, 3050RY207892, 305000210060, 3050RX208008, 3050RX208510, 3050RX205976, 3050RX206178, 305000210563, 305000211723, 3050RX210060, 305000212616, 305000213007, 305000213191, 30500X210563, 3050RX207061, 305000214590, 3050RX205426, 30500X215502, 3050RX207062, 305000215531, 30500X216410, 3050RY199850, 305000216132, 30500X216627, 30500X217119, 3050RY217112, 305000216718, 3050RX217751, 3050RX217840, 3050RY217837, 305000217858, 30500X218369, 30500X218391, 30500X219054, 3050RY206178, 305000220708, 305000218578, 30500X220430, 305000219261, 30500X221221, 305000221226, 305000219989, 30500X222569, 305000221953, 305000221955, 30500X223153, 30500X223709, 305000223275, 305000223280, 305000222640, 30500X224142, 305000223867, 305000223866, 30500X224692, 305000224352, 305000224353, 30500X225386, 305000225045, 305000225044, 30500X225992, 305000225632, 305000225633, 30500X226540, 30500X227059, 305000226844, 305000226846, 305000226127, 30500X227564, 305000226836, 30500X227820, 305000227127, 305000227128, 30500X228194, 305000227790, 305000227789, 30500X228787, 305000228522, 305000228521, 305000228971, 305000228972, 30500X229560, 305000229328, 305000229329, 30500X229766, 30500X230087, 305000229825, 305000229826, 30500X230565, 305000230303, 305000230302, 30500X231053, 305000230780, 305000230781, 30500X231786, 305000231461, 305000231462, 30500X232331, 305000232098, 305000232099, 30500X232960, 305000232605, 305000232604, 305000233245, 305000233244, 305000233243, 305000233242, 305000233241, 305000233240, 30500X233415, 30500X234183, 305000233607, 305000233608, 30500X234806, 305000234303, 305000234304, 305000235022, 305000235017, 30500X235732, 305000235598, 30500X236328, 305000236105, 305000236104, 305000236655, 305000236653, 30500Y237087, 30500X237090, 305000237232, 30500X237761, 305000237824, 305000237825, 30500X238273, 305000238263, 305000238264, 30500X239095, 305000238818, 305000238819, 30500X239497, 30500X239748, 30500X239958, 30500X240334, 30500X240668, 30500X241368, 305000241516, 305000242236, 305000242792, 305000242795, 305000242796, 305000242794, 305000242793, 305000242791, 3050RY243869, 305000243251, 305000243882, 305000243881, 305000243888, 305000243889, 305000243883, 30500X244281, 30500X244880, 305000244513,30500X245306, 305000244481, 305000244480, 305000244934, 30500X245937, 30500X246067, 305000245408, 30500X246872, 305000245835, 30500X246951, 30500X247608, 305000246460, 30500X248284, 305000247007, 3050RX227127, 3050RX234303, 305000247717, 305000249072, 3050RX242794, 3050RX242793, 3050RX237825, 3050RX241516, 30500X249488, 305000248479, 30500X250423, 305000249743, 30500X250964, 305000250456, 30500X251060, 305000251055, 30500X251560, 30500X252133, 3050RY252136, 305000251453, 30500X252771, 305000252058, 3050RX244513, 305000252646, 305000253785, 30500X254929, 3050RY255145, 30500X255141, 305000254589, 305000255549, 305000255551, 305000255038, 30500X256614, 305000256192, 305000257281, 305000256709, 305000257280, 305000258048, 30500X259385, 305000258379, 305000258921, 305000258922, 30500X259557, 305000259431, 305000259432, 30500X260159, 305000259966, 305000259960, 305000260514, 305000260527, 30500Y261748, 3050RX253785, 305000261325, 305000261326, 305000261737, 305000261736, 3050RX256192, 30500X262441, 305000262328, 305000262327, 305000262978, 305000262977, 3050RY263508, 30500X263559, 305000263504, 305000264073, 305000264074, 305000264625, 305000264624, 3050RY265348, 30500X265422, 30500X265976, 3050RY265975, 305000265530, 305000265527, 30500X266142, 305000266134, 305000266135, 30500X266789, 305000267193, 305000267194, 3050RX261325, 30500X267468, 305000267421, 305000267422, 30500X267990, 305000267673, 305000267674, 305000268294, 305000268293, 305000266784, 305000268914, 305000268915, 3050RX261736, 3050RY269955, 30500X269963, 305000269220, 305000269221, 305000270081, 3050RY271045, 30500X271047, 305000271042, 305000271716, 305000271715, 30500X272367, 305000272273, 305000272274, 305000272852, 305000272851, 30500X274204, 305000273466, 305000273463, 305000274200, 305000274201, 30500X275297, 30500X275393, 305000274816, 30500X276594, 3050RX268914, 305000277761, 3050RY282196, 305000282304, 30500X282815, 3050RY283371, 3050RX283315, 3050RX283314 Catalog Number 110-4G: 3050RX195650, 3050RX198690, 3050RY193969, 3050RY197837, 3050RX199151, 3050RX199942, 3050RX201552, 3050RY185259, 3050RX200538, 305000206177, 3050RX200989, 305000206556, 305000206557, 3050RX200990, 3050RY199151, 3050RY199942, 305000207465, 3050RY200538, 3050RY200989, 305000208007, 3050RY200990, 3050RX207465, 3050RX206557, 3050RX206556, 305000209417, 305000209105, 3050RX208007, 305000210891, 305000211419, 3050RX206177, 30500X209105, 3050RX209417, 305000212298, 30500X210891, 305000213240, 305000214103, 30500X215501, 30500X215577, 3050RX202201, 305000215628, 3050RY195650, 3050RY198690, 30500X216621, 305000216246,3050RY217113, 30500X217118, 305000217140, 30500X217835, 3050RY217839, 305000217464, 30500X218393, 305000218009, 30500X219053, 305000218225, 305000219118, 30500X220001, 30500X220429, 305000219990, 30500X221225, 305000221227, 30500X222567, 305000222642, 305000223279, 30500X223707, 305000223868, 30500X224685, 305000224354, 30500X225383, 305000225046, 305000225634, 30500X226000, 30500X226541, 305000226731, 305000226869, 305000226867, 30500X227567, 305000227129, 30500X228784, 305000228523, 305000228973, 30500X229561, 305000229330, 30500X230084, 305000229827, 305000230304, 30500X231054,305000230782, 30500X231787, 305000231463, 30500X232333, 305000232101, 30500X232776, 305000232606, 30500X233419, 305000233246, 305000233609, 30500X234690, 30500X234807, 305000234306, 305000235566, 30500X236329, 30500Y237088, 30500X237767, 305000237233, 305000238820, 30500X239496, 305000239476, 30500X240336, 30500X241357, 305000241518, 30500X242172, 305000242249, 305000243252, 305000244364, 30500X245304, 305000244935, 305000245409, 305000245838, 30500X246522, 305000246308, 30500X246873, 30500X247609, 30500X248281, 305000249073, 3050RX241518, 3050RX233246, 3050RX242249, 3050RX228973, 30500X251061, 305000251056, 30500X251563, 3050RY252137, 305000252059, 305000252648, 305000254591, 30500X254931, 30500X255143, 305000255039, 305000255548, 305000256193, 305000256737, 305000257244, 305000257246, 305000257921, 305000257920, 305000258303, 305000258290, 30500X259383, 305000258793, 305000259573, 30500X260163, 305000259958, 305000259959, 3050RX254591, 3050RX255039, 305000260513, 305000260512, 30500Y261771, 3050RX257244, 3050RX257920, 305000261159, 305000261160, 3050RY262444, 30500X262443, 305000261734, 305000261732, 305000262325, 305000262326, 3050RX258303, 305000262976, 305000262975, 3050RY263510, 30500X263560, 305000263507, 305000263506, 305000264075, 305000264076, 305000264627, 305000264628, 30500X265424, 305000265525, 305000265524, 30500X265978, 30500X266143, 305000266063, 305000266064, 30500X266787, 305000266638, 305000266637, 305000267191, 30500X267470, 305000267423, 305000267675, 3050RY269957, 30500X274206 Catalog Number 110-4H: 3050RX198281, 3050RY194144, 3050RY198281, 3050RY200861, 30500X223154, 305000220948, 305000223622, 30500X225385, 30500X233417, 305000231714, 305000241350, 30500X243874, 305000250959, 30500X254932, 305000263557 Catalog Number 110-4L: 305000221383, 30500X222566, 305000222762, 30500X223971,305000223969, 305000225078, 30500X226539, 305000226221, 30500X227058, 305000226848, 305000226749, 305000226847, 305000226871, 305000226,870, 30500X228783, 305000229450, 305000229909, 305000230871 30500X232775, 305000232199, 30500X232961, 305000233311, 305000234438, 30500X235731, 305000235181, 305000236201, 305000237368, 305000237369, 305000237799, 305000238539, 305000239102, 30500X240245, 30500X240246, 30500X240247, 305000239736, 305000239737, 30500X241365, 30500X241522, 305000241082, 30500X242170, 30500X245765, 30500X246521, 3050RX237799, 3050RX226749, 3050RX229450, 3050RX241082, 3050RX232199, 3050RY252775, 3050RX240446, 3050RY255148, 305000259553, 305000260626, 305000261835, 305000262436, 3050RX259553, 305000263146, 3050RX260626, 305000264077, 3050RX262436, 3050RX261835, 305000264175, 3050RY265351, 3050RY265766, 305000265531, 30500X266144, 3050RY266146, 30500X266522, 305000266062, 305000267297, 30500X267473, 30500X268133, 305000267784, 305000268412, 3050RX265531, 305000268916, 3050RY271046, 305000271872, 30500X272374, 305000272591, 30500X274210, 30500X274963, 305000281820 Catalog Number 110-4BC: 176044, 177018, 178013, 178898, 179757, 180139, 180955, 182279, 182908, 184079, 184687, 185284, 186277, 187293, 190452, 187780, 190451, 191891, 193414, 194074, 195161, 195608, 196987, 197492, 198189, 198512, 200069, 200490, 199609, 201039, 201595, 202423, 203565, 207077, 208532 Catalog Number 110-4HC: 179758, 180956, 183071, 184686, 186278, 187781, 190450, 192246, 194993, 195450, 198191, 199708, 201040, 201596, 202009 Catalog Number 110-4LC: 162523, 167450, 167786, 175720, 177188, 178547, 182130, 187495, 195245, 197493. |
Recalling Firm/ Manufacturer |
Integra Limited 311 Enterprise Dr Plainsboro NJ 08536-3344
|
Manufacturer Reason for Recall |
Integra LifeSciences has identified through an internal evaluation that in order to comply with US FDA
requirements, certain Camino¿ Intracranial Pressure Monitoring Kits require additional labelling to indicate
they are UNSAFE in an MR environment.
Only the following Camino¿ Intracranial Pressure Monitoring Kits are affected:
110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Integra sent an Urgent Medical Device Correction Field Safety Notice dted September 10, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory and determine if they have any of the affected product. If so, customers were asked to ensure that the information contained in the letter be distributed to the appropriate personnel. Complete the attached form return it by email or fax as indicated on the form. Customers with questions were instructed to contact Customer Service at 1-855-532-1723.
For questions regarding this recall call 1-800-997-4868, outside of the US 888-980-7742. |
Quantity in Commerce |
79,887 |
Distribution |
Nationwide Distribution |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = GWM and Original Applicant = INTEGRA LIFESCIENCES CORPORATION
|
|
|
|