• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall SYMBIQ OneChannel and TwoChannel Infusers

  • Print
  • Share
  • E-mail
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections

New Search Back to Search Results
 Class 2 Recall
SYMBIQ OneChannel and TwoChannel Infusers
see related information
Date Posted November 20, 2013
Recall Status1 Open
Recall Number Z-0372-2014
Recall Event ID 66263
Premarket Notification
510(K) Number
Product Classification Pump, Infusion - Product Code FRN
Product *** 1) SYMBIQ One-Channel Infuser; *** 2) SYMBIQ Two-Channel Infuser; *** E09 batteries can be identified by the "E09" that is printed beneath the bar code and at the end of the serial number along the bottom of the label found on each battery; *** E10 batteries can be identified by the "E10" that is printed beneath the bar code and at the end of the serial number along the bottom of the label found on each battery. Symbiq is a general-purpose infuser designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral, enteral, intra-arterial, subcutaneous, or irrigation routes of administration.
Code Information *** 1) SYMBIQ One-Channel Infuser, List Number: 16026-04-01/02/03/04/25/26/43/44/45/46/51/52/53/54/55/56/57/58/65/66/69/70/77/78/79/80/81/82/83/84/87/88/89/90/93/94; *** 2) SYMBIQ Two- Channel Infuser, List Number: 16027-04-01/02/03/04/25/26/43/44/45/46/51/52/53/54/55/56/57/58/65/66/69/70/77/78/79/80/81/82/83/84/87/88/89/90/ 93/94; *** X) Battery Pack E10 Part Number: HSP3566-E10; *** X) Battery Pack E09 Part Number: HSP3566-E09
Recalling Firm/
Hospira Inc.
275 N Field Dr
Lake Forest, Illinois 60045-2579
For Additional Information Contact Hospira Global Complaint Management
Manufacturer Reason
for Recall
Hospira shipped the incorrect replaceable battery pack (Part No: HSP3566-E10) instead of the correct E09 battery pack (Part No. HSP3566-E09) for the Symbiq Single and Dual Channel Infusers.
FDA Determined
Cause 2
TRAINING: Employee Error
Action The firm, Hospira, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated August 28, 2013 to its customers. Hospira contracted with Stericycle, Inc., in Indianapolis, IN for customer notification, response, tracking, returns, effectiveness checks and destruction of returned product starting on August 28, 2013. The letter described the product, problem and actions to be taken. Customers were instructed to: 1) complete and return the attached reply form via fax to 1-866-470-3291 or e-mail to Hospira2921@stericycle.com, even if you do not currently have the affected batteries; 2) contact Stericycle at 1-877-779-7810 (M-F, 8am - 5pm ET) if they require additional response forms; and, 3) return any affected batteries upon receipt of the new (correct) replacement batteries to Stericycle. Customers are also instructed to forward the recall information to any accounts that they may have shipped the affected batteries to and have them contact Stericycle to recieve a reply form and return label. For further inquiries, please contact Hospira Advanced Knowledge Center at 1-800-241-4002, option 6, availabel 24 hours a day/7 days per week.
Quantity in Commerce 59 batteries
Distribution US Distribution in states of: : KY, MN, ND, NJ, NY, PA, and SD.
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = FRN and Original Applicant = HOSPIRA, INC.