|
Class 2 Device Recall CARESCAPE Monitor B850 |
|
Date Initiated by Firm |
December 20, 2013 |
Date Posted |
February 04, 2014 |
Recall Status1 |
Terminated 3 on May 02, 2016 |
Recall Number |
Z-0927-2014 |
Recall Event ID |
66946 |
510(K)Number |
K092027
|
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product |
CARESCAPE Monitor B850, a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility.
Product Usage: The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 system is indicated for monitoring of Hemodynamic (including ECG, ST Segment, Arrhythmia Detection, ECG Diagnostic Analysis and Measurement, Invasive Pressure, Noninvasive Blood Pressure, Pulse Oximetry, Cardiac Output, Temperature, Impedance Respiration and SvO2 (Mixed Venous Oxygen Saturation)), Airway Gases (Fi/Et C02, 02, N20 and Anesthetic Agent), Spirometry, Gas Exchange (02 Consumption (VO2), C02 production (VCO2), energy expenditure (BE), and respiratory quotient (RQ)) and neurophysiological (including electroencephalography (EEG), Entropy, Bispectral Index (BIS) and Neuromuscular Transmission (NMT) Monitoring) status. The CARESCAPE Monitor B850 provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for bed to bed viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. In addition to the healthcare practitioner, the CARESCAPE Monitor B850 is designed to provide configuration and troubleshooting capabilities to qualified service personnel. |
Code Information |
serial numbers: SEG09270517HX, SEG09270549HX, SEG09270551HX, SEG09270553HX, SEG09270554HX, SEG09270557HX, SEG09270559HX, SEG09270561HX, SEG09270562HX, SEG09270563HX, SEG09270564HX, SEG09270565HX, SEG09270566HX, SEG09270567HX, SEG09270568HX, SEG09270569HX, SEG09270583HX, SEG09270584HX, SEG09270585HX, SEG09270587HX, SEG09270588HX, SEG09270589HX, SEG09270591HX, SEG09270592HX, SEG09280596HX, SEG09280597HX, SEG09280598HX, SEG09280599HX, SEG09280600HX, SEG09280601HX, SEG09280602HX, SEG09280603HX, SEG09280604HX, SEG09280605HX, SEG09280606HX, SEG09280607HX, SEG09280608HX, SEG09280609HX, SEG09280610HX, SEG09280611HX, SEG09280612HX, SEG09280613HX, SEG09280614HX, SEG09280615HX, SEG09280616HX, SEG09280617HX, SEG09280618HX, SEG09280619HX, SEG09280620HX, SEG09280621HX, SEG09280622HX, SEG09280623HX, SEG09280624HX, SEG09280625HX, SEG09310656HX, SEG09310657HX, SEG09310658HX, SEG09310659HX, SEG09310660HX, SEG09310661HX, SEG09310662HX, SEG09310663HX, SEG09310664HX, SEG09310665HX, SEG09310666HX, SEG09310667HX, SEG09310668HX, SEG09310669HX, SEG09310670HX, SEG09310671HX, SEG09310672HX, SEG09310673HX, SEG09310674HX, SEG09310675HX, SEG09310696HX, SEG09310697HX, SEG09310698HX, SEG09310699HX, SEG09310700HX, SEG09310701HX, SEG09310702HX, SEG09310703HX, SEG09310704HX, SEG09310705HX, SEG09310706HX, SEG09310707HX, SEG09310708HX, SEG09310709HX, SEG09310710HX, SEG09310711HX, SEG09310712HX, SEG09310713HX, SEG09310714HX, SEG09310715HX, SEG09310716HX, SEG09310717HX, SEG09310718HX, SEG09310719HX, SEG09310720HX, SEG09310721HX, SEG09310722HX, SEG09310723HX, SEG09310724HX, SEG09310725HX, SEG09310726HX, SEG09310727HX, SEG09310728HX, SEG09310729HX, SEG09310730HX, SEG09310731HX, SEG09310732HX, SEG09310733HX, SEG09310734HX, SEG09310735HX, SEG09310736HX, SEG09310737HX, SEG09310738HX, SEG09310739HX, SEG09310740HX, SEG09310741HX, SEG09310742HX, SEG09310743HX, SEG09310744HX, SEG09310745HX, SEG09320797HX, SEG09320798HX, SEG09320799HX, SEG09320800HX, SEG09320801HX, SEG09320802HX, SEG09320803HX, SEG09320804HX, SEG09320805HX, SEG09320806HX, SEG09320807HX, SEG09320808HX, SEG09320809HX, SEG09320810HX, SEG09320811HX, SEG09320812HX, SEG09320813HX, SEG09320814HX, SEG09320815HX, SEG09320816HX, SEG09320817HX, SEG09320818HX, SEG09320819HX, SEG09320820HX, SEG09320821HX, SEG09320822HX, SEG09320823HX, SEG09320824HX, SEG09320825HX, SEG09320826HX, SEG09320827HX, SEG09320828HX, SEG09320829HX, SEG09320830HX, SEG09320831HX, SEG09320832HX, SEG09320833HX, SEG09320834HX, SEG09320835HX, SEG09320836HX, SEG09320837HX, SEG09320838HX, SEG09320839HX, SEG09320840HX, SEG09320841HX, SEG09320842HX, SEG09320843HX, SEG09320844HX, SEG09320845HX, SEG09320846HX, SEG09320847HX, SEG09320848HX, SEG09320849HX, SEG09320850HX, SEG09320851HX, SEG09320852HX, SEG09320853HX, SEG09320854HX, SEG09320855HX, SEG09320856HX, SEG09320857HX, SEG09320858HX, SEG09320859HX, SEG09320860HX, SEG09320861HX, SEG09320862HX, SEG09320863HX, SEG09320864HX, SEG09320865HX, SEG09320866HX, SEG09320867HX, SEG09320868HX, SEG09320869HX, SEG09320870HX, SEG09320871HX, SEG09320872HX, SEG09320873HX, SEG09320874HX, SEG09320875HX, SEG09320876HX, SEG09320877HX, SEG09320878HX, SEG09320879HX, SEG09320880HX, SEG09320881HX, SEG09320882HX, SEG09320883HX, SEG09320884HX, SEG09320885HX, SEG09320886HX, SEG09320887HX, SEG09320888HX, SEG09320889HX, SEG09320890HX, SEG09320891HX, SEG09320892HX, SEG09320893HX, SEG09320894HX, SEG09320895HX, SEG09320896HX, SEG09360898HX, SEG09360899HX, SEG09360900HX, SEG09360901HX, SEG09360902HX, SEG09360903HX, SEG09360904HX, SEG09360905HX, SEG09360906HX, SEG09360907HX, SEG09360908HX, SEG09360909HX, SEG09360910HX, SEG09360911HX, SEG09360912HX, SEG09360913HX, SEG09360914HX, SEG09360915HX, SEG09360916HX, SEG09360917HX, SEG09360918HX, SEG09360919HX, SEG09360920HX, SEG09360921HX, SEG09360922HX, SEG09360923HX, SEG09360924HX, SEG09360925HX, SEG09360926HX, SEG09360927HX, SEG09360928HX, SEG09360929HX, SEG09360930HX, SEG09360931HX, SEG09360932HX, SEG09360933HX, SEG09360934HX, SEG09360935HX, SEG09360936HX, SEG09360937HX, SEG09360938HX, SEG09360939HX, SEG09360940HX, SEG09360941HX, SEG09360942HX, SEG09360943HX, SEG09360944HX, SEG09360945HX, SEG09360946HX, SEG09360947HX, SEG09370948HX, SEG09370949HX, SEG09370950HX, SEG09370951HX, SEG09370952HX, SEG09370953HX, SEG09370954HX, SEG09370955HX, SEG09370956HX, SEG09370957HX, SEG09370958HX, SEG09370959HX, SEG09370960HX, SEG09370961HX, SEG09370962HX, SEG09370963HX, SEG09370964HX, SEG09370965HX, SEG09370966HX, SEG09370967HX, SEG09370968HX, SEG09370969HX, SEG09370970HX, SEG09370971HX, SEG09370972HX, SEG09370973HX, SEG09370974HX, SEG09370975HX, SEG09370976HX, SEG09370977HX, SEG09370978HX, SEG09370979HX, SEG09370980HX, SEG09370981HX, SEG09370982HX, SEG09370983HX, SEG09370984HX, SEG09370985HX, SEG09370986HX, SEG09370987HX, SEG09370988HX, SEG09370989HX, SEG09370990HX, SEG09370991HX, SEG09370992HX, SEG09370993HX, SEG09370994HX, SEG09370995HX, SEG09370996HX, SEG09370997HX, SEG09370998HX, SEG09370999HX, SEG09371000HX, SEG09371001HX, SEG09371002HX, SEG09371003HX, SEG09371004HX, SEG09371005HX, SEG09371006HX, SEG09371007HX, SEG09371008HX, SEG09371009HX, SEG09371010HX, SEG09371011HX, SEG09371012HX, SEG09371013HX, SEG09371014HX, SEG09371015HX, SEG09371016HX, SEG09371017HX, SEG09371018HX, SEG09371019HX, SEG09371020HX, SEG09371021HX, SEG09371022HX, SEG09371023HX, SEG09371024HX, SEG09371025HX, SEG09371026HX, SEG09371027HX, SEG09371028HX, SEG09371029HX, SEG09371030HX, SEG09371031HX, SEG09371032HX, SEG09371033HX, SEG09371034HX, SEG09371035HX, SEG09371036HX, SEG09371037HX, SEG09371038HX, SEG09371039HX, SEG09371040HX, SEG09371041HX, SEG09371042HX, SEG09371043HX, SEG09371044HX, SEG09371045HX, SEG09371046HX, SEG09371047HX, SEG09371048HX, SEG09371049HX, SEG09371050HX, SEG09371051HX, SEG09371052HX, SEG09371053HX, SEG09371054HX, SEG09371055HX, SEG09371056HX, SEG09371057HX, SEG09371058HX, SEG09371059HX, SEG09371060HX, SEG09371061HX, SEG09371062HX, SEG09371063HX, SEG09371064HX, SEG09371065HX, SEG09371066HX, SEG09371067HX, SEG09381068HX, SEG09381069HX, SEG09381070HX, SEG09381071HX, SEG09381072HX, SEG09381073HX, SEG09381074HX, SEG09381075HX, SEG09381076HX, SEG09381077HX, SEG09381078HX, SEG09381079HX, SEG09381080HX, SEG09381081HX, SEG09381082HX, SEG09381083HX, SEG09381084HX, SEG09381085HX, SEG09381086HX, SEG09381087HX, SEG09381088HX, SEG09381089HX, SEG09381090HX, SEG09381091HX, SEG09381092HX, SEG09381093HX, SEG09381094HX, SEG09381095HX, SEG09381096HX, SEG09381097HX, SEG09381098HX, SEG09381099HX, SEG09381100HX, SEG09381101HX, SEG09381102HX, SEG09381103HX, SEG09381104HX, SEG09381105HX, SEG09381106HX, SEG09381107HX, SEG09381108HX, SEG09381109HX, SEG09381110HX, SEG09381111HX, SEG09381112HX, SEG09381113HX, SEG09381114HX, SEG09381115HX, SEG09381116HX, SEG09381117HX, SEG09381118HX, SEG09381119HX, SEG09381120HX, SEG09381121HX, SEG09381122HX, SEG09381123HX, SEG09381124HX, SEG09381125HX, SEG09381126HX, SEG09381127HX, SEG09381128HX, SEG09381129HX, SEG09381130HX, SEG09381131HX, SEG09381132HX, SEG09381133HX, SEG09381134HX, SEG09381135HX, SEG09381136HX, SEG09381137HX, SEG09381138HX, SEG09381139HX, SEG09381140HX, SEG09381141HX, SEG09381142HX, SEG09381143HX, SEG09381144HX, SEG09381145HX, SEG09381146HX, SEG09381147HX, SEG09381148HX, SEG09381149HX, SEG09381150HX, SEG09381151HX, SEG09381152HX, SEG09381153HX, SEG09381154HX, SEG09381155HX, SEG09381156HX, SEG09381157HX, SEG09381158HX, SEG09381159HX, SEG09381160HX, SEG09381161HX, SEG09381162HX, SEG09381163HX, SEG09381164HX, SEG09381165HX, SEG09381166HX, SEG09381167HX, SEG09411368HX, SEG09411369HX, SEG09411370HX, SEG09411371HX, SEG09411372HX, SEG09411373HX, SEG09411374HX, SEG09411375HX, SEG09411376HX, SEG09411377HX, SEG09411378HX, SEG09411379HX, SEG09411380HX, SEG09411381HX, SEG09411382HX, SEG09411383HX, SEG09411384HX, SEG09411385HX, SEG09411386HX, SEG09411387HX, SEG09411388HX, SEG09411389HX, SEG09411390HX, SEG09411391HX, SEG09411393HX, SEG09411394HX, SEG09411395HX, SEG09411396HX, SEG09411397HX, SEG09411398HX, SEG09411399HX, SEG09411400HX, SEG09411401HX, SEG09411402HX, SEG09411403HX, SEG09411404HX, SEG09411405HX, SEG09411406HX, SEG09411407HX, SEG09411408HX, SEG09411409HX, SEG09411410HX, SEG09411411HX, SEG09411412HX, SEG09411413HX, SEG09411414HX, SEG09411415HX, SEG09411416HX, SEG09411417HX, SEG09411418HX, SEG09411419HX, SEG09411420HX, SEG09411421HX, SEG09411422HX, SEG09411423HX, SEG09411424HX, SEG09411425HX, SEG09411426HX, SEG09411427HX, SEG09411428HX, SEG09411429HX, SEG09411430HX, SEG09411431HX, SEG09411432HX, SEG09411433HX, SEG09411434HX, SEG09411435HX, SEG09411436HX, SEG09411437HX, SEG09411438HX, SEG09411439HX, SEG09411440HX, SEG09411441HX, SEG09411442HX, SEG09411443HX, SEG09411444HX, SEG09411445HX, SEG09411446HX, SEG09411447HX, SEG09411448HX, SEG09411449HX, SEG09411450HX, SEG09411451HX, SEG09411452HX, SEG09411453HX, SEG09411454HX, SEG09411455HX, SEG09411456HX, SEG09411457HX, SEG09411458HX, SEG09411459HX, SEG09411460HX, SEG09411461HX, SEG09411462HX, SEG09411463HX, SEG09411464HX, SEG09411465HX, SEG09411466HX, SEG09411467HX, SEG09411468HX, SEG09411469HX, SEG09411470HX, SEG09411471HX, SEG09411472HX, SEG09411473HX, SEG09411474HX, SEG09411475HX, SEG09411476HX, SEG09411477HX, SEG09411478HX, SEG09411479HX, SEG09411480HX, SEG09411481HX, SEG09411482HX, SEG09411483HX, SEG09411484HX, SEG09411485HX, SEG09411486HX, SEG09411487HX, SEG09411488HX, SEG09411489HX, SEG09411490HX, SEG09411491HX, SEG09411492HX, SEG09411493HX, SEG09411494HX, SEG09411495HX, SEG09411496HX, SEG09411497HX, SEG09411498HX, SEG09411499HX, SEG09411500HX, SEG09411501HX, SEG09411502HX, SEG09411503HX, SEG09411504HX, SEG09411505HX, SEG09411506HX, SEG09411507HX, SEG09411508HX, SEG09411509HX, SEG09411510HX, SEG09411511HX, SEG09411512HX, SEG09411513HX, SEG09411514HX, SEG09411515HX, SEG09411516HX, SEG09411517HX, SEG09411518HX, SEG09411519HX, SEG09411520HX, SEG09411521HX, SEG09411522HX, SEG09411523HX, SEG09411524HX, SEG09411525HX, SEG09411526HX, SEG09411527HX, SEG09411528HX, SEG09411529HX, SEG09411530HX, SEG09411531HX, SEG09411532HX, SEG09411533HX, SEG09411534HX, SEG09411535HX, SEG09411536HX, SEG09411537HX, SEG09411538HX, SEG09411539HX, SEG09411540HX, SEG09411541HX, SEG09411542HX, SEG09411543HX, SEG09411544HX, SEG09411545HX, SEG09411546HX, SEG09411547HX, SEG09411548HX, SEG09411549HX, SEG09411550HX, SEG09411551HX, SEG09411552HX, SEG09411553HX, SEG09411554HX, SEG09411555HX, SEG09411556HX, SEG09411557HX, SEG09411558HX, SEG09411559HX, SEG09411560HX, SEG09411561HX, SEG09411562HX, SEG09411563HX, SEG09411564HX, SEG09411565HX, SEG09411566HX, SEG09411567HX, SEG09411568HX, SEG09421619HX, SEG09421620HX, SEG09421621HX, SEG09421622HX, SEG09421623HX, SEG09421624HX, SEG09421625HX, SEG09421626HX, SEG09421627HX, SEG09421628HX, SEG09421629HX, SEG09421630HX, SEG09421631HX, SEG09421632HX, SEG09421633HX, SEG09421634HX, SEG09421635HX, SEG09421636HX, SEG09421637HX, SEG09421638HX, SEG09421639HX, SEG09421640HX, SEG09421641HX, SEG09421642HX, SEG09421643HX, SEG09421644HX, SEG09421645HX, SEG09421646HX, SEG09421647HX, SEG09421648HX, SEG09421649HX, SEG09421650HX, SEG09421651HX, SEG09421652HX, SEG09421653HX, SEG09421654HX, SEG09421655HX, SEG09421656HX, SEG09421657HX, SEG09421658HX, SEG09421659HX, SEG09421660HX, SEG09421661HX, SEG09421662HX, SEG09421663HX, SEG09421664HX, SEG09421665HX, SEG09421666HX, SEG09421667HX, SEG09421668HX, SEG09421669HX, SEG09421670HX, SEG09421671HX, SEG09421672HX, SEG09421673HX, SEG09421674HX, SEG09421675HX, SEG09421676HX, SEG09421677HX, SEG09421678HX, SEG09421679HX, SEG09421680HX, SEG09421681HX, SEG09421682HX, SEG09421683HX, SEG09421684HX, SEG09421685HX, SEG09421686HX, SEG09421687HX, SEG09421688HX, SEG09421689HX, SEG09421690HX, SEG09421691HX, SEG09421692HX, SEG09421693HX, SEG09421694HX, SEG09421695HX, SEG09421696HX, SEG09421697HX, SEG09421698HX, SEG09421699HX, SEG09421700HX, SEG09421701HX, SEG09421702HX, SEG09421703HX, SEG09421704HX, SEG09421705HX, SEG09421706HX, SEG09421707HX, SEG09421708HX, SEG09421709HX, SEG09421710HX, SEG09421711HX, SEG09421712HX, SEG09421713HX, SEG09421714HX, SEG09421715HX, SEG09421716HX, SEG09421717HX, SEG09421718HX, SEG09431720HX, SEG09431721HX, SEG09431722HX, SEG09431723HX, SEG09431724HX, SEG09431725HX, SEG09431726HX, SEG09431727HX, SEG09431728HX, SEG09431729HX, SEG09431730HX, SEG09431731HX, SEG09431732HX, SEG09431733HX, SEG09431734HX, SEG09431735HX, SEG09431736HX, SEG09431737HX, SEG09431738HX, SEG09431739HX, SEG09431740HX, SEG09431741HX, SEG09431742HX, SEG09431743HX, SEG09431744HX, SEG09431745HX, SEG09431746HX, SEG09431747HX, SEG09431748HX, SEG09431749HX, SEG09431750HX, SEG09431751HX, SEG09431752HX, SEG09431753HX, SEG09431754HX, SEG09431755HX, SEG09431756HX, SEG09431757HX, SEG09431758HX, SEG09431759HX, SEG09431760HX, SEG09431761HX, SEG09431762HX, SEG09431763HX, SEG09431764HX, SEG09431765HX, SEG09431766HX, SEG09431767HX, SEG09431768HX, SEG09431769HX, SEG09431770HX, SEG09431771HX, SEG09431772HX, SEG09431773HX, SEG09431774HX, SEG09431775HX, SEG09431776HX, SEG09431777HX, SEG09431778HX, SEG09431779HX, SEG09431780HX, SEG09431781HX, SEG09431782HX, SEG09431783HX, SEG09431784HX, SEG09431785HX, SEG09431786HX, SEG09431787HX, SEG09431788HX, SEG09431789HX, SEG09431790HX, SEG09431791HX, SEG09431792HX, SEG09431793HX, SEG09431794HX, SEG09431795HX, SEG09431796HX, SEG09431797HX, SEG09431798HX, SEG09431799HX, SEG09431800HX, SEG09431801HX, SEG09431802HX, SEG09431803HX, SEG09431804HX, SEG09431805HX, SEG09431806HX, SEG09431807HX, SEG09431808HX, SEG09431809HX, SEG09431810HX, SEG09431811HX, SEG09431812HX, SEG09431813HX, SEG09431814HX, SEG09431815HX, SEG09431816HX, SEG09431817HX, SEG09431818HX, SEG09441819HX, SEG09441820HX, SEG09441821HX, SEG09441822HX, SEG09441823HX, SEG09441824HX, SEG09441825HX, SEG09441826HX, SEG09441827HX, SEG09441828HX, SEG09451879HX, SEG09451880HX, SEG09451881HX, SEG09451882HX, SEG09451883HX, SEG09451884HX, SEG09451885HX, SEG09451886HX, SEG09451887HX, SEG09451888HX, SEG09451889HX, SEG09451890HX, SEG09451891HX, SEG09451892HX, SEG09451893HX, SEG09451894HX, SEG09451895HX, SEG09451896HX, SEG09451897HX, SEG09451898HX, SEG09451899HX, SEG09451900HX, SEG09451901HX, SEG09451902HX, SEG09451903HX, SEG09451904HX, SEG09451905HX, SEG09451906HX, SEG09451907HX, SEG09451908HX, SEG09451909HX, SEG09451910HX, SEG09451911HX, SEG09451912HX, SEG09451913HX, SEG09451914HX, SEG09451915HX, SEG09451916HX, SEG09451917HX, SEG09451918HX, SEG09451919HX, SEG09451920HX, SEG09451921HX, SEG09451922HX, SEG09451923HX, SEG09451924HX, SEG09451925HX, SEG09451926HX, SEG09451927HX, SEG09451928HX, SEG09451929HX, SEG09451930HX, SEG09451931HX, SEG09451932HX, SEG09451933HX, SEG09451934HX, SEG09451935HX, SEG09451936HX, SEG09451937HX, SEG09451938HX, SEG09451939HX, SEG09451940HX, SEG09451941HX, SEG09451942HX, SEG09451943HX, SEG09451944HX, SEG09451945HX, SEG09451946HX, SEG09451947HX, SEG09451948HX, SEG09451949HX, SEG09451950HX, SEG09451951HX, SEG09451952HX, SEG09451953HX, SEG09451954HX, SEG09451955HX, SEG09451956HX, SEG09451957HX, SEG09451958HX, SEG09451959HX, SEG09451960HX, SEG09451961HX, SEG09451962HX, SEG09451963HX, SEG09451964HX, SEG09451965HX, SEG09451966HX, SEG09451967HX, SEG09451968HX, SEG09451969HX, SEG09451970HX, SEG09451971HX, SEG09451972HX, SEG09451973HX, SEG09451974HX, SEG09451975HX, SEG09451976HX, SEG09451977HX, SEG10012104HX, SEG10012105HX, SEG10012106HX, SEG10012107HX, SEG10012108HX, SEG10012109HX, SEG10012110HX, SEG10012111HX, SEG10012112HX, SEG10012113HX, SEG10012114HX, SEG10012115HX, SEG10012116HX, SEG10012117HX, SEG10012118HX, SEG10012119HX, SEG10012120HX, SEG10012121HX, SEG10012122HX, SEG10012123HX, SEG10012124HX, SEG10012125HX, SEG10012126HX, SEG10012127HX, SEG10012128HX, SEG10012129HX, SEG10012130HX, SEG10012131HX, SEG10012132HX, SEG10012133HX, SEG10012134HX, SEG10012135HX, SEG10012136HX, SEG10012137HX, SEG10012138HX, SEG10012139HX, SEG10012140HX, SEG10012141HX, SEG10012143HX, SEG10012144HX, SEG10012145HX, SEG10012146HX, SEG10012147HX, SEG10012148HX, SEG10012149HX, SEG10012150HX, SEG10012151HX, SEG10012152HX, SEG10012153HX, SEG10012154HX, SEG10012155HX, SEG10012156HX, SEG10012157HX, SEG10012158HX, SEG10012159HX, SEG10012160HX, SEG10012161HX, SEG10012162HX, SEG10012163HX, SEG10012164HX, SEG10012165HX, SEG10012166HX, SEG10012167HX, SEG10012168HX, SEG10012169HX, SEG10012170HX, SEG10012171HX, SEG10012172HX, SEG10012173HX, SEG10012174HX, SEG10012175HX, SEG10012176HX, SEG10012177HX, SEG10012178HX, SEG10012179HX, SEG10012180HX, SEG10012181HX, SEG10012182HX, SEG10012183HX, SEG10012184HX, SEG10012185HX, SEG10012186HX, SEG10012187HX, SEG10012188HX, SEG10012189HX, SEG10012190HX, SEG10012191HX, SEG10012192HX, SEG10012193HX, SEG10012194HX, SEG10012195HX, SEG10012196HX, SEG10012197HX, SEG10012198HX, SEG10012199HX, SEG10012200HX, SEG10012201HX, SEG10012202HX, SEG10012203HX, SEG10022264HX, SEG10022265HX, SEG10022266HX, SEG10022267HX, SEG10022268HX, SEG10022269HX, SEG10022270HX, SEG10022271HX, SEG10022272HX, SEG10022273HX, SEG10022274HX, SEG10022275HX, SEG10022276HX, SEG10022277HX, SEG10022278HX, SEG10022279HX, SEG10022280HX, SEG10022281HX, SEG10022282HX, SEG10022283HX, SEG10022284HX, SEG10022285HX, SEG10022286HX, SEG10022287HX, SEG10022288HX, SEG10022289HX, SEG10022290HX, SEG10022291HX, SEG10022292HX, SEG10022293HX, SEG10022294HX, SEG10022295HX, SEG10022296HX, SEG10022297HX, SEG10022298HX, SEG10022299HX, SEG10022300HX, SEG10022301HX, SEG10022302HX, SEG10022303HX, SEG10022304HX, SEG10022305HX, SEG10022306HX, SEG10022307HX, SEG10022308HX, SEG10022309HX, SEG10022310HX, SEG10022311HX, SEG10022312HX, SEG10022313HX, SEG10022314HX, SEG10022315HX, SEG10022316HX, SEG10022317HX, SEG10022318HX, SEG10022319HX, SEG10022320HX, SEG10022321HX, SEG10022322HX, SEG10022323HX, SEG10022324HX, SEG10022325HX, SEG10022326HX, SEG10022327HX, SEG10022328HX, SEG10022329HX, SEG10022330HX, SEG10022331HX, SEG10022332HX, SEG10022333HX, SEG10022334HX, SEG10022335HX, SEG10022336HX, SEG10022337HX, SEG10022338HX, SEG10022339HX, SEG10022340HX, SEG10022341HX, SEG10022342HX, SEG10022343HX, SEG10022344HX, SEG10022345HX, SEG10022346HX, SEG10022347HX, SEG10022348HX, SEG10022349HX, SEG10022350HX, SEG10022351HX, SEG10022352HX, SEG10022353HX, SEG10022354HX, SEG10022355HX, SEG10022356HX, SEG10022357HX, SEG10022358HX, SEG10022359HX, SEG10022360HX, SEG10022361HX, SEG10022362HX, SEG10022363HX, SEG10032414HX, SEG10032415HX, SEG10032416HX, SEG10032417HX, SEG10032418HX, SEG10032419HX, SEG10032420HX, SEG10032421HX, SEG10032422HX, SEG10032423HX, SEG10032424HX, SEG10032425HX, SEG10032426HX, SEG10032427HX, SEG10032428HX, SEG10032429HX, SEG10032430HX, SEG10032431HX, SEG10032432HX, SEG10032433HX, SEG10032434HX, SEG10032435HX, SEG10032436HX, SEG10032437HX, SEG10032438HX, SEG10032439HX, SEG10032440HX, SEG10032441HX, SEG10032443HX, SEG10032444HX, SEG10032445HX, SEG10032446HX, SEG10032447HX, SEG10032448HX, SEG10032449HX, SEG10032450HX, SEG10032451HX, SEG10032452HX, SEG10032453HX, SEG10032454HX, SEG10032455HX, SEG10032456HX, SEG10032457HX, SEG10032458HX, SEG10032459HX, SEG10032460HX, SEG10032461HX, SEG10032462HX, SEG10032463HX, SEG10032464HX, SEG10032465HX, SEG10032466HX, SEG10032467HX, SEG10032468HX, SEG10032469HX, SEG10032470HX, SEG10032471HX, SEG10032472HX, SEG10032473HX, SEG10032474HX, SEG10032475HX, SEG10032476HX, SEG10032477HX, SEG10032478HX, SEG10032479HX, SEG10032480HX, SEG10032481HX, SEG10032482HX, SEG10032483HX, SEG10032484HX, SEG10032485HX, SEG10032486HX, SEG10032487HX, SEG10032488HX, SEG10032489HX, SEG10032490HX, SEG10032491HX, SEG10032492HX, SEG10032493HX, SEG10032494HX, SEG10032495HX, SEG10032496HX, SEG10032497HX, SEG10032498HX, SEG10032499HX, SEG10032500HX, SEG10032501HX, SEG10032502HX, SEG10032503HX, SEG10032504HX, SEG10032505HX, SEG10032506HX, SEG10032507HX, SEG10032508HX, SEG10032509HX, SEG10032510HX, SEG10032511HX, SEG10032512HX, SEG10032513HX, SEG10042514HX, SEG10042515HX, SEG10042516HX, SEG10042517HX, SEG10042518HX, SEG10042519HX, SEG10042520HX, SEG10042521HX, SEG10042522HX, SEG10042523HX, SEG10042524HX, SEG10042525HX, SEG10042526HX, SEG10042527HX, SEG10042528HX, SEG10042529HX, SEG10042530HX, SEG10042531HX, SEG10042532HX, SEG10042533HX, SEG10042534HX, SEG10042535HX, SEG10042536HX, SEG10042537HX, SEG10042538HX, SEG10042539HX, SEG10042540HX, SEG10042541HX, SEG10042542HX, SEG10042543HX, SEG10042544HX, SEG10042545HX, SEG10042546HX, SEG10042547HX, SEG10042548HX, SEG10042549HX, SEG10042550HX, SEG10042551HX, SEG10042552HX, SEG10042553HX, SEG10042554HX, SEG10042555HX, SEG10042556HX, SEG10042557HX, SEG10042558HX, SEG10042559HX, SEG10042560HX, SEG10042561HX, SEG10042562HX, SEG10042563HX, SEG10042564HX, SEG10042565HX, SEG10042566HX, SEG10042567HX, SEG10042568HX, SEG10042569HX, SEG10042570HX, SEG10042571HX, SEG10042572HX, SEG10042573HX, SEG10042574HX, SEG10042575HX, SEG10042576HX, SEG10042577HX, SEG10042578HX, SEG10042579HX, SEG10042580HX, SEG10042581HX, SEG10042582HX, SEG10042583HX, SEG10042584HX, SEG10042587HX, SEG10042588HX, SEG10042589HX, SEG10042590HX, SEG10042591HX, SEG10042592HX, SEG10042593HX, SEG10042594HX, SEG10042596HX, SEG10042597HX, SEG10042598HX, SEG10042599HX, SEG10042600HX, SEG10042601HX, SEG10042602HX, SEG10042603HX, SEG10042609HX, SEG10052618HX, SEG10052619HX, SEG10052622HX, SEG10052625HX, SEG10052626HX, SEG10052627HX, SEG10052628HX, SEG10052629HX, SEG10052630HX, SEG10052631HX, SEG10052632HX, SEG10052633HX, SEG10052634HX, SEG10052635HX, SEG10052636HX, SEG10052637HX, SEG10052638HX, SEG10052639HX, SEG10052640HX, SEG10052641HX, SEG10052642HX, SEG10052643HX, SEG10052644HX, SEG10052645HX, SEG10052646HX, SEG10052647HX, SEG10052648HX, SEG10052649HX, SEG10052650HX, SEG10052651HX, SEG10052652HX, SEG10052653HX, SEG10052654HX, SEG10052655HX, SEG10052656HX, SEG10052657HX, SEG10052658HX, SEG10052659HX, SEG10052660HX, SEG10052661HX, SEG10052662HX, SEG10052663HX, SEG10062714HX, SEG10062715HX, SEG10062716HX, SEG10062717HX, SEG10062718HX, SEG10062719HX, SEG10062720HX, SEG10062721HX, SEG10062722HX, SEG10062723HX, SEG10062724HX, SEG10062725HX, SEG10062726HX, SEG10062727HX, SEG10062728HX, SEG10062729HX, SEG10062730HX, SEG10062731HX, SEG10062732HX, SEG10062733HX, SEG10062734HX, SEG10062735HX, SEG10062736HX, SEG10062737HX, SEG10062738HX, SEG10062739HX, SEG10062740HX, SEG10062741HX, SEG10062742HX, SEG10062743HX, SEG10062744HX, SEG10062745HX, SEG10062746HX, SEG10062747HX, SEG10062748HX, SEG10062749HX, SEG10062750HX, SEG10062751HX, SEG10062752HX, SEG10062753HX, SEG10062754HX, SEG10062755HX, SEG10062756HX, SEG10062757HX, SEG10062758HX, SEG10062759HX, SEG10062760HX, SEG10062761HX, SEG10062762HX, SEG10062763HX, SEG10072764HX, SEG10072765HX, SEG10072766HX, SEG10072767HX, SEG10072768HX, SEG10072769HX, SEG10072770HX, SEG10072771HX, SEG10072772HX, SEG10072773HX, SEG10072774HX, SEG10072775HX, SEG10072776HX, SEG10072777HX, SEG10072778HX, SEG10072779HX, SEG10072780HX, SEG10072781HX, SEG10072782HX, SEG10072783HX, SEG10072784HX, SEG10072785HX, SEG10072786HX, SEG10072787HX, SEG10072788HX, SEG10072789HX, SEG10072790HX, SEG10072791HX, SEG10072792HX, SEG10072793HX, SEG10072794HX, SEG10072795HX, SEG10072796HX, SEG10072797HX, SEG10072798HX, SEG10072799HX, SEG10072800HX, SEG10072801HX, SEG10072802HX, SEG10072803HX, SEG10072804HX, SEG10072805HX, SEG10072806HX, SEG10072807HX, SEG10072808HX, SEG10072809HX, SEG10072810HX, SEG10072811HX, SEG10072812HX, SEG10072813HX, SEG10072814HX, SEG10072815HX, SEG10072816HX, SEG10072817HX, SEG10072818HX, SEG10072819HX, SEG10072820HX, SEG10072821HX, SEG10072822HX, SEG10072823HX, SEG10082844HX, SEG10082845HX, SEG10082846HX, SEG10082847HX, SEG10082848HX, SEG10082849HX, SEG10082850HX, SEG10082851HX, SEG10082852HX, SEG10082853HX, SEG10082854HX, SEG10082855HX, SEG10082856HX, SEG10082857HX, SEG10082858HX, SEG10082859HX, SEG10082860HX, SEG10082861HX, SEG10082862HX, SEG10082863HX, SEG10082864HX, SEG10082865HX, SEG10082866HX, SEG10082867HX, SEG10082868HX, SEG10082869HX, SEG10082870HX, SEG10082871HX, SEG10082872HX, SEG10082873HX, SEG10082874HX, SEG10082875HX, SEG10082876HX, SEG10082877HX, SEG10082878HX, SEG10082879HX, SEG10082880HX, SEG10082881HX, SEG10082882HX, SEG10082883HX, SEG10162884HX, SEG10162885HX, SEG10162886HX, SEG10162887HX, SEG10162888HX, SEG10162889HX, SEG10162890HX, SEG10162891HX, SEG10162892HX, SEG10162893HX, SEG10162894HX, SEG10162895HX, SEG10162896HX, SEG10162897HX, SEG10162898HX, SEG10162899HX, SEG10162900HX, SEG10162901HX, SEG10162902HX, SEG10162903HX, SEG10162904HX, SEG10162905HX, SEG10162906HX, SEG10162907HX, SEG10162908HX, SEG10162909HX, SEG10162910HX, SEG10162911HX, SEG10162912HX, SEG10162913HX, SEG10162914HX, SEG10162915HX, SEG10162916HX, SEG10162917HX, SEG10162918HX, SEG10162919HX, SEG10162920HX, SEG10162921HX, SEG10162922HX, SEG10162923HX, SEG10162924HX, SEG10162925HX, SEG10162926HX, SEG10162927HX, SEG10162928HX, SEG10162929HX, SEG10162930HX, SEG10162931HX, SEG10162932HX, SEG10162933HX, SEG10162934HX, SEG10162935HX, SEG10162936HX, SEG10162937HX, SEG10162938HX, SEG10162939HX, SEG10162940HX, SEG10162941HX, SEG10162942HX, SEG10162943HX, SEG10162944HX, SEG10162945HX, SEG10162946HX, SEG10162947HX, SEG10162948HX, SEG10162949HX, SEG10162950HX, SEG10162951HX, SEG10162952HX, SEG10162953HX, SEG10162954HX, SEG10162955HX, SEG10162956HX, SEG10162957HX, SEG10162958HX, SEG10162959HX, SEG10162960HX, SEG10162961HX, SEG10162962HX, SEG10162963HX, SEG10162964HX, SEG10162965HX, SEG10162966HX, SEG10162967HX, SEG10162968HX, SEG10162969HX, SEG10162970HX, SEG10162971HX, SEG10162972HX, SEG10162973HX, SEG10162974HX, SEG10162975HX, SEG10162976HX, SEG10162977HX, SEG10162978HX, SEG10162979HX, SEG10162980HX, SEG10162981HX, SEG10162982HX, SEG10162983HX, SEG10172984HX, SEG10172985HX, SEG10172986HX, SEG10172987HX, SEG10172988HX, SEG10172989HX, SEG10172990HX, SEG10172991HX, SEG10172992HX, SEG10172993HX, SEG10172994HX, SEG10172995HX, SEG10172996HX, SEG10172997HX, SEG10172998HX, SEG10172999HX, SEG10173000HX, SEG10173001HX, SEG10173002HX, SEG10173003HX, SEG10173004HX, SEG10173005HX, SEG10173006HX, SEG10173007HX, SEG10173008HX, SEG10173009HX, SEG10173010HX, SEG10173011HX, SEG10173012HX, SEG10173013HX, SEG10173014HX, SEG10173015HX, SEG10173016HX, SEG10173017HX, SEG10173018HX, SEG10173019HX, SEG10173020HX, SEG10173021HX, SEG10173022HX, SEG10173023HX, SEG10173024HX, SEG10173025HX, SEG10173026HX, SEG10173027HX, SEG10173028HX, SEG10173030HX, SEG10173031HX, SEG10173034HX, SEG10173036HX, SEG10173037HX, SEG10173039HX, SEG10173041HX, SEG10173043HX, SEG10183044HX, SEG10183045HX, SEG10183047HX, SEG10183051HX, SEG10183052HX, SEG10183053HX, SEG10183054HX, SEG10183056HX, SEG10183058HX, SEG10183059HX, SEG10183060HX, SEG10183061HX, SEG10183062HX, SEG10183063HX, SEG10183064HX, SEG10183065HX, SEG10183066HX, SEG10183067HX, SEG10183068HX, SEG10183069HX, SEG10183070HX, SEG10183071HX, SEG10183072HX, SEG10183073HX, SEG10183075HX, SEG10183076HX, SEG10183078HX, SEG10183079HX, SEG10183080HX, SEG10183081HX, SEG10183082HX, SEG10183083HX, SEG10183085HX, SEG10183086HX, SEG10183087HX, SEG10183088HX, SEG10183089HX, SEG10183090HX, SEG10183091HX, SEG10183092HX, SEG10183093HX, SEG10233095HX, SEG10233096HX, SEG10233097HX, SEG10233098HX, SEG10233099HX, SEG10233100HX, SEG10233101HX, SEG10233102HX, SEG10233103HX, SEG10233104HX, SEG10233105HX, SEG10233106HX, SEG10233107HX, SEG10233108HX, SEG10233109HX, SEG10233110HX, SEG10233111HX, SEG10233112HX, SEG10233113HX, SEG10233114HX, SEG10233115HX, SEG10233116HX, SEG10233117HX, SEG10233118HX, SEG10233119HX, SEG10233120HX, SEG10233121HX, SEG10233122HX, SEG10233123HX, SEG10233124HX, SEG10233125HX, SEG10233126HX, SEG10233127HX, SEG10233128HX, SEG10233129HX, SEG10233130HX, SEG10233131HX, SEG10233132HX, SEG10233133HX, SEG10233134HX, SEG10233135HX, SEG10233136HX, SEG10233137HX, SEG10233138HX, SEG10233139HX, SEG10233141HX, SEG10233142HX, SEG10233143HX, SEG10233144HX, SEG10233145HX, SEG10233146HX, SEG10233147HX, SEG10233148HX, SEG10233149HX, SEG10233150HX, SEG10233151HX, SEG10233152HX, SEG10233153HX, SEG10233154HX, SEG10233155HX, SEG10233156HX, SEG10233157HX, SEG10233158HX, SEG10233159HX, SEG10233160HX, SEG10233161HX, SEG10233162HX, SEG10233163HX, SEG10233164HX, SEG10233165HX, SEG10233166HX, SEG10233167HX, SEG10233168HX, SEG10233169HX, SEG10233170HX, SEG10233171HX, SEG10233172HX, SEG10233173HX, SEG10233174HX, SEG10233175HX, SEG10233176HX, SEG10233177HX, SEG10233178HX, SEG10233179HX, SEG10233180HX, SEG10233181HX, SEG10233182HX, SEG10233183HX, SEG10233184HX, SEG10233185HX, SEG10233186HX, SEG10233187HX, SEG10233188HX, SEG10233189HX, SEG10233190HX, SEG10233191HX, SEG10233192HX, SEG10233193HX, SEG10233194HX, SEG10233296HX, SEG10233297HX, SEG10233298HX, SEG10233299HX, SEG10233300HX, SEG10233301HX, SEG10233302HX, SEG10233303HX, SEG10233304HX, SEG10233305HX, SEG10233306HX, SEG10233307HX, SEG10233308HX, SEG10233309HX, SEG10233310HX, SEG10233311HX, SEG10233312HX, SEG10233313HX, SEG10233314HX, SEG10233315HX, SEG10233316HX, SEG10233317HX, SEG10233318HX, SEG10233319HX, SEG10233320HX, SEG10233321HX, SEG10233322HX, SEG10233323HX, SEG10233324HX, SEG10233325HX, SEG10233326HX, SEG10233327HX, SEG10233328HX, SEG10233329HX, SEG10233330HX, SEG10233331HX, SEG10233332HX, SEG10233333HX, SEG10233334HX, SEG10233335HX, SEG10233336HX, SEG10233337HX, SEG10233338HX, SEG10233339HX, SEG10233340HX, SEG10233341HX, SEG10233342HX, SEG10233343HX, SEG10233344HX, SEG10233345HX, SEG10233346HX, SEG10233347HX, SEG10233348HX, SEG10233349HX, SEG10233350HX, SEG10233351HX, SEG10233352HX, SEG10233353HX, SEG10233354HX, SEG10233355HX, SEG10233356HX, SEG10233357HX, SEG10233358HX, SEG10233359HX, SEG10233360HX, SEG10233361HX, SEG10233362HX, SEG10233363HX, SEG10233364HX, SEG10233365HX, SEG10233366HX, SEG10233367HX, SEG10233368HX, SEG10233369HX, SEG10233370HX, SEG10233372HX, SEG10233373HX, SEG10233374HX, SEG10233375HX, SEG10233376HX, SEG10233377HX, SEG10233378HX, SEG10233379HX, SEG10233380HX, SEG10233381HX, SEG10233382HX, SEG10233383HX, SEG10233384HX, SEG10233385HX, SEG10233386HX, SEG10233387HX, SEG10233388HX, SEG10233389HX, SEG10233390HX, SEG10233391HX, SEG10233392HX, SEG10233393HX, SEG10233394HX, SEG10233395HX, SEG10233396HX, SEG10233397HX, SEG10233398HX, SEG10233399HX, SEG10233400HX, SEG10233401HX, SEG10233402HX, SEG10233403HX, SEG10233404HX, SEG10233405HX, SEG10233406HX, SEG10233407HX, SEG10233408HX, SEG10233409HX, SEG10233410HX, SEG10233411HX, SEG10233412HX, SEG10233413HX, SEG10233414HX, SEG10233415HX, SEG10233416HX, SEG10233417HX, SEG10233418HX, SEG10233419HX, SEG10233420HX, SEG10233421HX, SEG10233422HX, SEG10233423HX, SEG10233424HX, SEG10233425HX, SEG10233426HX, SEG10233427HX, SEG10233428HX, SEG10233429HX, SEG10233430HX, SEG10233431HX, SEG10233432HX, SEG10233433HX, SEG10233434HX, SEG10233435HX, SEG10233436HX, SEG10233437HX, SEG10233438HX, SEG10233439HX, SEG10233440HX, SEG10233441HX, SEG10233442HX, SEG10233443HX, SEG10233444HX, SEG10233445HX, SEG10273535HX, SEG10273536HX, SEG10273537HX, SEG10273538HX, SEG10273539HX, SEG10273540HX, SEG10273541HX, SEG10273542HX, SEG10273543HX, SEG10273544HX, SEG10273545HX, SEG10273546HX, SEG10273547HX, SEG10273548HX, SEG10273549HX, SEG10273550HX, SEG10273551HX, SEG10273552HX, SEG10273553HX, SEG10273554HX, SEG10273555HX, SEG10273556HX, SEG10273557HX, SEG10273558HX, SEG10273559HX, SEG10273560HX, SEG10273561HX, SEG10273562HX, SEG10273563HX, SEG10273564HX, SEG10273565HX, SEG10273566HX, SEG10273567HX, SEG10273568HX, SEG10273569HX, SEG10273570HX, SEG10273571HX, SEG10273572HX, SEG10273573HX, SEG10273574HX, SEG10273575HX, SEG10273576HX, SEG10273577HX, SEG10273578HX, SEG10273579HX, SEG10273580HX, SEG10273581HX, SEG10273582HX, SEG10273583HX, SEG10273584HX, SEG10273585HX, SEG10273586HX, SEG10273587HX, SEG10273588HX, SEG10273589HX, SEG10273590HX, SEG10273591HX, SEG10273592HX, SEG10273593HX, SEG10273594HX, SEG10273595HX, SEG10273596HX, SEG10273597HX, SEG10273598HX, SEG10273599HX, SEG10273600HX, SEG10273601HX, SEG10273602HX, SEG10273603HX, SEG10273604HX, SEG10273605HX, SEG10273606HX, SEG10273607HX, SEG10273608HX, SEG10273609HX, SEG10273610HX, SEG10273611HX, SEG10273612HX, SEG10273613HX, SEG10273614HX, SEG10273615HX, SEG10273616HX, SEG10273617HX, SEG10273618HX, SEG10273619HX, SEG10273620HX, SEG10273621HX, SEG10273622HX, SEG10273623HX, SEG10273624HX, SEG10273625HX, SEG10273626HX, SEG10273627HX, SEG10273628HX, SEG10273629HX, SEG10273630HX, SEG10273631HX, SEG10273632HX, SEG10273633HX, SEG10273634HX, SEG10293861HX, SEG10293862HX, SEG10293863HX, SEG10293864HX, SEG10293865HX, SEG10293866HX, SEG10293867HX, SEG10293868HX, SEG10293869HX, SEG10293870HX, SEG10293871HX, SEG10293872HX, SEG10293873HX, SEG10293874HX, SEG10293875HX, SEG10293876HX, SEG10293877HX, SEG10293878HX, SEG10293879HX, SEG10293880HX, SEG10293881HX, SEG10293882HX, SEG10293883HX, SEG10293884HX, SEG10293885HX, SEG10293886HX, SEG10293887HX, SEG10293888HX, SEG10293889HX, SEG10293890HX, SEG10293891HX, SEG10293892HX, SEG10293893HX, SEG10293894HX, SEG10293895HX, SEG10293896HX, SEG10293897HX, SEG10293898HX, SEG10293899HX, SEG10293900HX, SEG10293901HX, SEG10293902HX, SEG10293903HX, SEG10293904HX, SEG10293905HX, SEG10293906HX, SEG10293907HX, SEG10293908HX, SEG10293909HX, SEG10293910HX, SEG10293911HX, SEG10293912HX, SEG10293913HX, SEG10293914HX, SEG10293915HX, SEG10293916HX, SEG10293917HX, SEG10293918HX, SEG10293919HX, SEG10293920HX, SEG10293921HX, SEG10293922HX, SEG10293923HX, SEG10293924HX, SEG10293925HX, SEG10293926HX, SEG10293927HX, SEG10293928HX, SEG10293929HX, SEG10293930HX, SEG10293931HX, SEG10293932HX, SEG10293933HX, SEG10293934HX, SEG10293935HX, SEG10293936HX, SEG10293938HX, SEG10293939HX, SEG10293940HX, SEG10293941HX, SEG10293943HX, SEG10293944HX, SEG10293945HX, SEG10293946HX, SEG10293947HX, SEG10293948HX, SEG10293949HX, SEG10293950HX, SEG10293951HX, SEG10293952HX, SEG10293953HX, SEG10293954HX, SEG10293955HX, SEG10293956HX, SEG10293957HX, SEG10293958HX, SEG10293959HX, SEG10293960HX, SEG10304063HX, SEG10304064HX, SEG10304065HX, SEG10304066HX, SEG10304067HX, SEG10304068HX, SEG10304069HX, SEG10304070HX, SEG10304071HX, SEG10304072HX, SEG10304073HX, SEG10304074HX, SEG10304075HX, SEG10304076HX, SEG10304077HX, SEG10304078HX, SEG10304079HX, SEG10304080HX, SEG10304081HX, SEG10304082HX, SEG10304083HX, SEG10304084HX, SEG10304085HX, SEG10304086HX, SEG10304087HX, SEG10304088HX, SEG10304089HX, SEG10304090HX, SEG10304091HX, SEG10304092HX, SEG10304093HX, SEG10304094HX, SEG10304095HX, SEG10304096HX, SEG10304097HX, SEG10304098HX, SEG10304099HX, SEG10304100HX, SEG10304101HX, SEG10304102HX, SEG10304103HX, SEG10304104HX, SEG10304105HX, SEG10304106HX, SEG10304107HX, SEG10304108HX, SEG10304109HX, SEG10304110HX, SEG10304111HX, SEG10304112HX, SEG10304113HX, SEG10304114HX, SEG10304115HX, SEG10304116HX, SEG10304117HX, SEG10304118HX, SEG10304119HX, SEG10304120HX, SEG10304121HX, SEG10304122HX, SEG10304123HX, SEG10304124HX, SEG10304125HX, SEG10304126HX, SEG10304127HX, SEG10304128HX, SEG10304129HX, SEG10304130HX, SEG10304131HX, SEG10304132HX, SEG10304133HX, SEG10304134HX, SEG10304135HX, SEG10304136HX, SEG10304137HX, SEG10304138HX, SEG10304139HX, SEG10304140HX, SEG10304141HX, SEG10304142HX, SEG10304143HX, SEG10304144HX, SEG10304145HX, SEG10304146HX, SEG10304147HX, SEG10304148HX, SEG10304149HX, SEG10304150HX, SEG10304151HX, SEG10304152HX, SEG10304153HX, SEG10304154HX, SEG10304155HX, SEG10304156HX, SEG10304157HX, SEG10304158HX, SEG10304159HX, SEG10304160HX, SEG10304161HX, SEG10304162HX, SEG10334284HX, SEG10334285HX, SEG10334286HX, SEG10334287HX, SEG10334288HX, SEG10334289HX, SEG10334290HX, SEG10334291HX, SEG10334292HX, SEG10334293HX, SEG10334294HX, SEG10334295HX, SEG10334296HX, SEG10334297HX, SEG10334298HX, SEG10334299HX, SEG10334300HX, SEG10334301HX, SEG10334302HX, SEG10334303HX, SEG10334304HX, SEG10334305HX, SEG10334306HX, SEG10334307HX, SEG10334308HX, SEG10334309HX, SEG10334310HX, SEG10334311HX, SEG10334312HX, SEG10334313HX, SEG10334314HX, SEG10334315HX, SEG10334316HX, SEG10334317HX, SEG10334318HX, SEG10334319HX, SEG10334320HX, SEG10334321HX, SEG10334322HX, SEG10334323HX, SEG10334324HX, SEG10334325HX, SEG10334326HX, SEG10334327HX, SEG10334328HX, SEG10334329HX, SEG10334330HX, SEG10334331HX, SEG10334332HX, SEG10334333HX, SEG10334334HX, SEG10334335HX, SEG10334336HX, SEG10334337HX, SEG10334338HX, SEG10334339HX, SEG10334340HX, SEG10334341HX, SEG10334342HX, SEG10334343HX, SEG10334344HX, SEG10334345HX, SEG10334346HX, SEG10334347HX, SEG10334348HX, SEG10334349HX, SEG10334350HX, SEG10334351HX, SEG10334352HX, SEG10334353HX, SEG10334354HX, SEG10334355HX, SEG10334356HX, SEG10334357HX, SEG10334358HX, SEG10334359HX, SEG10334360HX, SEG10334361HX, SEG10334362HX, SEG10334363HX, SEG10334364HX, SEG10334365HX, SEG10334366HX, SEG10334367HX, SEG10334368HX, SEG10334369HX, SEG10334370HX, SEG10334371HX, SEG10334372HX, SEG10334373HX, SEG10334374HX, SEG10334375HX, SEG10334376HX, SEG10334377HX, SEG10334378HX, SEG10334379HX, SEG10334380HX, SEG10334381HX, SEG10334382HX, SEG10334383HX, SEG10344384HX, SEG10344385HX, SEG10344386HX, SEG10344387HX, SEG10344388HX, SEG10344389HX, SEG10344390HX, SEG10344391HX, SEG10344392HX, SEG10344393HX, SEG10344394HX, SEG10344395HX, SEG10344396HX, SEG10344397HX, SEG10344398HX, SEG10344399HX, SEG10344400HX, SEG10344401HX, SEG10344402HX, SEG10344403HX, SEG10344404HX, SEG10344405HX, SEG10344406HX, SEG10344407HX, SEG10344408HX, SEG10344409HX, SEG10344410HX, SEG10344411HX, SEG10344412HX, SEG10344413HX, SEG10344414HX, SEG10344415HX, SEG10344416HX, SEG10344417HX, SEG10344418HX, SEG10344419HX, SEG10344420HX, SEG10344421HX, SEG10344422HX, SEG10344423HX, SEG10344424HX, SEG10344425HX, SEG10344426HX, SEG10344427HX, SEG10344428HX, SEG10344429HX, SEG10344430HX, SEG10344431HX, SEG10344432HX, SEG10344433HX, SEG10344434HX, SEG10344435HX, SEG10344436HX, SEG10344437HX, SEG10344438HX, SEG10344439HX, SEG10344440HX, SEG10344441HX, SEG10344442HX, SEG10344443HX, SEG10344444HX, SEG10344445HX, SEG10344446HX, SEG10344447HX, SEG10344448HX, SEG10344449HX, SEG10344450HX, SEG10344451HX, SEG10344452HX, SEG10344453HX, SEG10344454HX, SEG10344455HX, SEG10344456HX, SEG10344457HX, SEG10344458HX, SEG10344459HX, SEG10344460HX, SEG10344461HX, SEG10344462HX, SEG10344463HX, SEG10344464HX, SEG10344465HX, SEG10344466HX, SEG10344467HX, SEG10344468HX, SEG10344469HX, SEG10344470HX, SEG10344471HX, SEG10344472HX, SEG10344473HX, SEG10344474HX, SEG10344475HX, SEG10344476HX, SEG10344477HX, SEG10344478HX, SEG10344479HX, SEG10344480HX, SEG10344481HX, SEG10344482HX, SEG10344483HX, SEG10354585HX, SEG10354586HX, SEG10354587HX, SEG10354588HX, SEG10354589HX, SEG10354590HX, SEG10354591HX, SEG10354592HX, SEG10354593HX, SEG10354594HX, SEG10354595HX, SEG10354596HX, SEG10354597HX, SEG10354598HX, SEG10354599HX, SEG10354600HX, SEG10354601HX, SEG10354602HX, SEG10354603HX, SEG10354604HX, SEG10354605HX, SEG10354607HX, SEG10354608HX, SEG10354609HX, SEG10354610HX, SEG10354611HX, SEG10354612HX, SEG10354613HX, SEG10354614HX, SEG10354615HX, SEG10354616HX, SEG10354617HX, SEG10354618HX, SEG10354619HX, SEG10354620HX, SEG10354621HX, SEG10354622HX, SEG10354623HX, SEG10354624HX, SEG10354625HX, SEG10354626HX, SEG10354627HX, SEG10354628HX, SEG10354629HX, SEG10354630HX, SEG10354631HX, SEG10354632HX, SEG10354633HX, SEG10354634HX, SEG10354635HX, SEG10354636HX, SEG10354637HX, SEG10354638HX, SEG10354639HX, SEG10354640HX, SEG10354641HX, SEG10354642HX, SEG10354643HX, SEG10354644HX, SEG10354645HX, SEG10354646HX, SEG10354647HX, SEG10354649HX, SEG10354650HX, SEG10354651HX, SEG10354652HX, SEG10354653HX, SEG10354654HX, SEG10354655HX, SEG10354656HX, SEG10354657HX, SEG10354658HX, SEG10354659HX, SEG10354660HX, SEG10354661HX, SEG10354662HX, SEG10354663HX, SEG10354664HX, SEG10354665HX, SEG10354666HX, SEG10354667HX, SEG10354668HX, SEG10354669HX, SEG10354670HX, SEG10354671HX, SEG10354672HX, SEG10354673HX, SEG10354674HX, SEG10354675HX, SEG10354676HX, SEG10354677HX, SEG10354678HX, SEG10354679HX, SEG10354680HX, SEG10354681HX, SEG10354682HX, SEG10354683HX, SEG10354684HX, SEG10364785HX, SEG10364786HX, SEG10364787HX, SEG10364788HX, SEG10364789HX, SEG10364790HX, SEG10364791HX, SEG10364792HX, SEG10364793HX, SEG10364794HX, SEG10364795HX, SEG10364796HX, SEG10364797HX, SEG10364798HX, SEG10364799HX, SEG10364800HX, SEG10364801HX, SEG10364802HX, SEG10364803HX, SEG10364804HX, SEG10364805HX, SEG10364806HX, SEG10364807HX, SEG10364808HX, SEG10364809HX, SEG10364810HX, SEG10364811HX, SEG10364812HX, SEG10364813HX, SEG10364814HX, SEG10364815HX, SEG10364816HX, SEG10364817HX, SEG10364818HX, SEG10364819HX, SEG10364820HX, SEG10364821HX, SEG10364822HX, SEG10364823HX, SEG10364824HX, SEG10364825HX, SEG10364826HX, SEG10364827HX, SEG10364828HX, SEG10364829HX, SEG10364830HX, SEG10364831HX, SEG10364832HX, SEG10364833HX, SEG10364834HX, SEG10364835HX, SEG10364836HX, SEG10364837HX, SEG10364838HX, SEG10364839HX, SEG10364840HX, SEG10364841HX, SEG10364842HX, SEG10364843HX, SEG10364844HX, SEG10364845HX, SEG10364846HX, SEG10364847HX, SEG10364848HX, SEG10364849HX, SEG10364851HX, SEG10364852HX, SEG10364853HX, SEG10364854HX, SEG10364855HX, SEG10364856HX, SEG10364857HX, SEG10364858HX, SEG10364861HX, SEG10364862HX, SEG10364863HX, SEG10364864HX, SEG10364865HX, SEG10364866HX, SEG10364867HX, SEG10364868HX, SEG10364869HX, SEG10364870HX, SEG10364871HX, SEG10364872HX, SEG10364873HX, SEG10364874HX, SEG10364875HX, SEG10364876HX, SEG10364877HX, SEG10364878HX, SEG10364879HX, SEG10364880HX, SEG10364881HX, SEG10364882HX, SEG10364883HX, SEG10364884HX, SEG10364885HX, SEG10364886HX, SEG10364887HX, SEG10364888HX, SEG10364889HX, SEG10364890HX, SEG10364891HX, SEG10364892HX, SEG10364893HX, SEG10364894HX, SEG10364895HX, SEG10364896HX, SEG10364897HX, SEG10364898HX, SEG10364899HX, SEG10364900HX, SEG10364901HX, SEG10364902HX, SEG10364903HX, SEG10364904HX, SEG10364905HX, SEG10364906HX, SEG10364907HX, SEG10364908HX, SEG10364909HX, SEG10364910HX, SEG10364911HX, SEG10364912HX, SEG10364913HX, SEG10364914HX, SEG10364915HX, SEG10364916HX, SEG10364917HX, SEG10364918HX, SEG10364919HX, SEG10364920HX, SEG10364921HX, SEG10364922HX, SEG10364923HX, SEG10364924HX, SEG10364925HX, SEG10364926HX, SEG10364927HX, SEG10364928HX, SEG10364929HX, SEG10364930HX, SEG10364931HX, SEG10364932HX, SEG10364933HX, SEG10364934HX, SEG10364935HX, SEG10364936HX, SEG10364937HX, SEG10364938HX, SEG10364939HX, SEG10364940HX, SEG10364941HX, SEG10364942HX, SEG10364943HX, SEG10364944HX, SEG10364945HX, SEG10364946HX, SEG10364947HX, SEG10364948HX, SEG10364949HX, SEG10364950HX, SEG10364951HX, SEG10364952HX, SEG10364953HX, SEG10364954HX, SEG10364955HX, SEG10364956HX, SEG10364957HX, SEG10364958HX, SEG10364959HX, SEG10364960HX, SEG10364961HX, SEG10364962HX, SEG10364963HX, SEG10364964HX, SEG10364965HX, SEG10364966HX, SEG10364967HX, SEG10364968HX, SEG10364969HX, SEG10364970HX, SEG10364971HX, SEG10364972HX, SEG10364973HX, SEG10364974HX, SEG10364975HX, SEG10364976HX, SEG10364977HX, SEG10364978HX, SEG10364980HX, SEG10364981HX, SEG10364982HX, SEG10364983HX, SEG10364984HX, SEG10375001HX, SEG10375002HX, SEG10375003HX, SEG10375004HX, SEG10375005HX, SEG10375006HX, SEG10375007HX, SEG10375008HX, SEG10375009HX, SEG10375010HX, SEG10375011HX, SEG10375012HX, SEG10375013HX, SEG10375014HX, SEG10375015HX, SEG10375016HX, SEG10375017HX, SEG10375018HX, SEG10375019HX, SEG10375020HX, SEG10375021HX, SEG10375022HX, SEG10375023HX, SEG10375024HX, SEG10375025HX, SEG10375026HX, SEG10375027HX, SEG10375028HX, SEG10375029HX, SEG10375030HX, SEG10375031HX, SEG10375032HX, SEG10375033HX, SEG10375034HX, SEG10375035HX, SEG10375036HX, SEG10375037HX, SEG10375038HX, SEG10375039HX, SEG10375040HX, SEG10375041HX, SEG10375042HX, SEG10375043HX, SEG10375044HX, SEG10375045HX, SEG10375046HX, SEG10375047HX, SEG10375048HX, SEG10375049HX, SEG10375050HX, SEG10375051HX, SEG10375052HX, SEG10375053HX, SEG10375054HX, SEG10375055HX, SEG10375056HX, SEG10375057HX, SEG10375058HX, SEG10375059HX, SEG10375060HX, SEG10375061HX, SEG10375062HX, SEG10375063HX, SEG10375064HX, SEG10375065HX, SEG10375066HX, SEG10375067HX, SEG10375068HX, SEG10375069HX, SEG10375070HX, SEG10375071HX, SEG10375072HX, SEG10375073HX, SEG10375074HX, SEG10375075HX, SEG10375076HX, SEG10375077HX, SEG10375078HX, SEG10375079HX, SEG10375080HX, SEG10375081HX, SEG10375082HX, SEG10375083HX, SEG10375084HX, SEG10375085HX, SEG10375086HX, SEG10375087HX, SEG10375088HX, SEG10375089HX, SEG10375090HX, SEG10375091HX, SEG10375092HX, SEG10375093HX, SEG10375094HX, SEG10375095HX, SEG10375096HX, SEG10375097HX, SEG10375098HX, SEG10375099HX, SEG10385100HX, SEG10385101HX, SEG10385102HX, SEG10385103HX, SEG10385104HX, SEG10385105HX, SEG10385106HX, SEG10385107HX, SEG10385108HX, SEG10385109HX, SEG10385110HX, SEG10385111HX, SEG10385112HX, SEG10385113HX, SEG10385114HX, SEG10385115HX, SEG10385116HX, SEG10385117HX, SEG10385118HX, SEG10385119HX, SEG10385120HX, SEG10385121HX, SEG10385122HX, SEG10385123HX, SEG10385124HX, SEG10385125HX, SEG10385126HX, SEG10385127HX, SEG10385128HX, SEG10385129HX, SEG10385130HX, SEG10385131HX, SEG10385132HX, SEG10385133HX, SEG10385134HX, SEG10385135HX, SEG10385136HX, SEG10385137HX, SEG10385138HX, SEG10385139HX, SEG10385140HX, SEG10385141HX, SEG10385142HX, SEG10385143HX, SEG10385144HX, SEG10385145HX, SEG10385146HX, SEG10385147HX, SEG10385148HX, SEG10385149HX, SEG10385150HX, SEG10385151HX, SEG10385152HX, SEG10385153HX, SEG10385154HX, SEG10385155HX, SEG10385156HX, SEG10385157HX, SEG10385158HX, SEG10385159HX, SEG10385160HX, SEG10385161HX, SEG10385162HX, SEG10385163HX, SEG10385164HX, SEG10385165HX, SEG10385166HX, SEG10385167HX, SEG10385168HX, SEG10385169HX, SEG10385170HX, SEG10385171HX, SEG10385172HX, SEG10385173HX, SEG10385174HX, SEG10385175HX, SEG10385176HX, SEG10385177HX, SEG10385178HX, SEG10385179HX, SEG10385180HX, SEG10385181HX, SEG10385182HX, SEG10385183HX, SEG10385184HX, SEG10385185HX, SEG10385186HX, SEG10385187HX, SEG10385188HX, SEG10385189HX, SEG10385190HX, SEG10385191HX, SEG10385192HX, SEG10385193HX, SEG10385194HX, SEG10385195HX, SEG10385196HX, SEG10385197HX, SEG10385198HX, SEG10385199HX, SEG10395321HX, SEG10395322HX, SEG10395323HX, SEG10395324HX, SEG10395325HX, SEG10395326HX, SEG10395327HX, SEG10395329HX, SEG10395330HX, SEG10395331HX, SEG10395332HX, SEG10395333HX, SEG10395334HX, SEG10395335HX, SEG10395336HX, SEG10395337HX, SEG10395338HX, SEG10395339HX, SEG10395340HX, SEG10395341HX, SEG10395342HX, SEG10395343HX, SEG10395344HX, SEG10395345HX, SEG10395346HX, SEG10395347HX, SEG10395348HX, SEG10395349HX, SEG10395350HX, SEG10395351HX, SEG10395352HX, SEG10395353HX, SEG10395354HX, SEG10395355HX, SEG10395356HX, SEG10395357HX, SEG10395358HX, SEG10395359HX, SEG10395360HX, SEG10395361HX, SEG10395362HX, SEG10395363HX, SEG10395364HX, SEG10395365HX, SEG10395366HX, SEG10395367HX, SEG10395368HX, SEG10395369HX, SEG10395370HX, SEG10395371HX, SEG10395372HX, SEG10395373HX, SEG10395374HX, SEG10395375HX, SEG10395376HX, SEG10395377HX, SEG10395378HX, SEG10395379HX, SEG10395380HX, SEG10395381HX, SEG10395382HX, SEG10395383HX, SEG10395384HX, SEG10395385HX, SEG10395386HX, SEG10395387HX, SEG10395388HX, SEG10395389HX, SEG10395390HX, SEG10395391HX, SEG10395392HX, SEG10395393HX, SEG10395394HX, SEG10395395HX, SEG10395396HX, SEG10395397HX, SEG10395398HX, SEG10395399HX, SEG10395400HX, SEG10395401HX, SEG10395402HX, SEG10395403HX, SEG10395404HX, SEG10395405HX, SEG10395406HX, SEG10395407HX, SEG10395408HX, SEG10395409HX, SEG10395410HX, SEG10395411HX, SEG10395412HX, SEG10395413HX, SEG10395414HX, SEG10395415HX, SEG10395416HX, SEG10395417HX, SEG10395418HX, SEG10395419HX, SEG10395420HX, SEG10405741HX, SEG10405742HX, SEG10405743HX, SEG10405744HX, SEG10405745HX, SEG10405746HX, SEG10405747HX, SEG10405748HX, SEG10405749HX, SEG10405750HX, SEG10405751HX, SEG10405752HX, SEG10405753HX, SEG10405754HX, SEG10405755HX, SEG10405756HX, SEG10405757HX, SEG10405758HX, SEG10405759HX, SEG10405760HX, SEG10405761HX, SEG10405762HX, SEG10405763HX, SEG10405764HX, SEG10405765HX, SEG10405766HX, SEG10405767HX, SEG10405768HX, SEG10405769HX, SEG10405770HX, SEG10405771HX, SEG10405772HX, SEG10405773HX, SEG10405774HX, SEG10405775HX, SEG10405776HX, SEG10405777HX, SEG10405778HX, SEG10405779HX, SEG10405780HX, SEG10405781HX, SEG10405782HX, SEG10405783HX, SEG10405784HX, SEG10405785HX, SEG10405786HX, SEG10405787HX, SEG10405788HX, SEG10405789HX, SEG10405790HX, SEG10405791HX, SEG10405792HX, SEG10405793HX, SEG10405794HX, SEG10405795HX, SEG10405796HX, SEG10405797HX, SEG10405798HX, SEG10405799HX, SEG10405800HX, SEG10405801HX, SEG10405802HX, SEG10405803HX, SEG10405805HX, SEG10405806HX, SEG10405807HX, SEG10405808HX, SEG10405809HX, SEG10405810HX, SEG10405811HX, SEG10405812HX, SEG10405813HX, SEG10405814HX, SEG10405815HX, SEG10405816HX, SEG10405817HX, SEG10405818HX, SEG10405819HX, SEG10405820HX, SEG10405821HX, SEG10405822HX, SEG10405823HX, SEG10405824HX, SEG10405826HX, SEG10405827HX, SEG10405828HX, SEG10405829HX, SEG10405830HX, SEG10405831HX, SEG10405832HX, SEG10405833HX, SEG10405834HX, SEG10405835HX, SEG10405836HX, SEG10405837HX, SEG10405838HX, SEG10405839HX, SEG10405840HX, SEG10405841HX, SEG10405842HX, SEG10405843HX, SEG10405844HX, SEG10405845HX, SEG10405846HX, SEG10405847HX, SEG10405848HX, SEG10405850HX, SEG10405851HX, SEG10405852HX, SEG10405854HX, SEG10405855HX, SEG10405856HX, SEG10405857HX, SEG10405858HX, SEG10405859HX, SEG10405860HX, SEG10405861HX, SEG10405862HX, SEG10405863HX, SEG10405864HX, SEG10405865HX, SEG10405866HX, SEG10405867HX, SEG10405869HX, SEG10405870HX, SEG10405871HX, SEG10405872HX, SEG10405873HX, SEG10405874HX, SEG10405875HX, SEG10405876HX, SEG10405877HX, SEG10405878HX, SEG10405879HX, SEG10405880HX, SEG10405881HX, SEG10405882HX, SEG10405883HX, SEG10405884HX, SEG10405885HX, SEG10405886HX, SEG10405887HX, SEG10405888HX, SEG10405889HX, SEG10405890HX, SEG10405891HX, SEG10405892HX, SEG10405893HX, SEG10405894HX, SEG10405895HX, SEG10405896HX, SEG10405897HX, SEG10405898HX, SEG10405899HX, SEG10405900HX, SEG10405901HX, SEG10405902HX, SEG10405903HX, SEG10405904HX, SEG10405905HX, SEG10405906HX, SEG10405907HX, SEG10405908HX, SEG10405909HX, SEG10405910HX, SEG10405911HX, SEG10405913HX, SEG10405914HX, SEG10405915HX, SEG10405916HX, SEG10405917HX, SEG10405918HX, SEG10405919HX, SEG10405920HX, SEG10405921HX, SEG10405922HX, SEG10405923HX, SEG10405924HX, SEG10405925HX, SEG10405926HX, SEG10405927HX, SEG10405928HX, SEG10405929HX, SEG10405930HX, SEG10405931HX, SEG10405932HX, SEG10405933HX, SEG10405934HX, SEG10405935HX, SEG10405936HX, SEG10405937HX, SEG10405938HX, SEG10405939HX, SEG10405940HX, SEG10415941HX, SEG10415942HX, SEG10415943HX, SEG10415945HX, SEG10415947HX, SEG10415948HX, SEG10415949HX, SEG10415950HX, SEG10446507HX, SEG10446508HX, SEG10446509HX, SEG10446510HX, SEG10446511HX, SEG10446512HX, SEG10446513HX, SEG10446514HX, SEG10446515HX, SEG10446516HX, SEG10446517HX, SEG10446518HX, SEG10446519HX, SEG10446520HX, SEG10446521HX, SEG10446522HX, SEG10446523HX, SEG10446524HX, SEG10446525HX, SEG10446526HX, SEG10446527HX, SEG10446528HX, SEG10446529HX, SEG10446530HX, SEG10446531HX, SEG10446532HX, SEG10446533HX, SEG10446534HX, SEG10446535HX, SEG10446536HX, SEG10446537HX, SEG10446538HX, SEG10446539HX, SEG10446540HX, SEG10446541HX, SEG10446542HX, SEG10446543HX, SEG10446544HX, SEG10446545HX, SEG10446546HX, SEG10446547HX, SEG10446548HX, SEG10446549HX, SEG10446550HX, SEG10446551HX, SEG10446552HX, SEG10446553HX, SEG10446554HX, SEG10446555HX, SEG10446556HX, SEG10456607HX, SEG10456608HX, SEG10456609HX, SEG10456610HX, SEG10456611HX, SEG10456612HX, SEG10456613HX, SEG10456614HX, SEG10456615HX, SEG10456616HX, SEG10456617HX, SEG10456618HX, SEG10456619HX, SEG10456620HX, SEG10456621HX, SEG10456622HX, SEG10456623HX, SEG10456624HX, SEG10456625HX, SEG10456626HX, SEG10456627HX, SEG10456628HX, SEG10456629HX, SEG10456630HX, SEG10456631HX, SEG10456632HX, SEG10456633HX, SEG10456634HX, SEG10456635HX, SEG10456636HX, SEG10456637HX, SEG10456638HX, SEG10456639HX, SEG10456640HX, SEG10456641HX, SEG10456642HX, SEG10456643HX, SEG10456644HX, SEG10456645HX, SEG10456646HX, SEG10456647HX, SEG10456648HX, SEG10456649HX, SEG10456650HX, SEG10456651HX, SEG10456652HX, SEG10456653HX, SEG10456654HX, SEG10456655HX, SEG10456656HX, SEG10456657HX, SEG10456658HX, SEG10456659HX, SEG10456660HX, SEG10456661HX, SEG10456662HX, SEG10456663HX, SEG10456664HX, SEG10456665HX, SEG10456666HX, SEG10456667HX, SEG10456668HX, SEG10456669HX, SEG10456670HX, SEG10456671HX, SEG10456672HX, SEG10456673HX, SEG10456674HX, SEG10456675HX, SEG10456676HX, SEG10456677HX, SEG10456678HX, SEG10456679HX, SEG10456680HX, SEG10456681HX, SEG10456682HX, SEG10456683HX, SEG10456684HX, SEG10456685HX, SEG10456686HX, SEG10456687HX, SEG10456688HX, SEG10456689HX, SEG10456690HX, SEG10456691HX, SEG10456692HX, SEG10456693HX, SEG10456694HX, SEG10456695HX, SEG10456696HX, SEG10456697HX, SEG10456698HX, SEG10456699HX, SEG10456700HX, SEG10456701HX, SEG10456702HX, SEG10456703HX, SEG10456704HX, SEG10456705HX, SEG10456706HX, SEG10466887HX, SEG10466888HX, SEG10466889HX, SEG10466890HX, SEG10466891HX, SEG10466892HX, SEG10466893HX, SEG10466894HX, SEG10466895HX, SEG10466896HX, SEG10466897HX, SEG10466898HX, SEG10466899HX, SEG10466900HX, SEG10466901HX, SEG10466902HX, SEG10466903HX, SEG10466904HX, SEG10466905HX, SEG10466906HX, SEG10466907HX, SEG10466908HX, SEG10466909HX, SEG10466910HX, SEG10466911HX, SEG10466912HX, SEG10466913HX, SEG10466914HX, SEG10466915HX, SEG10466916HX, SEG10466917HX, SEG10466918HX, SEG10466919HX, SEG10466920HX, SEG10466921HX, SEG10466922HX, SEG10466923HX, SEG10466924HX, SEG10466925HX, SEG10466926HX, SEG10466927HX, SEG10466928HX, SEG10466929HX, SEG10466930HX, SEG10466931HX, SEG10466932HX, SEG10466933HX, SEG10466934HX, SEG10466935HX, SEG10466936HX, SEG10466937HX, SEG10466938HX, SEG10466939HX, SEG10466940HX, SEG10466941HX, SEG10466942HX, SEG10466943HX, SEG10466944HX, SEG10466945HX, SEG10466946HX, SEG10466947HX, SEG10466948HX, SEG10466949HX, SEG10466950HX, SEG10466951HX, SEG10466952HX, SEG10466953HX, SEG10466954HX, SEG10466955HX, SEG10466956HX, SEG10466957HX, SEG10466958HX, SEG10466959HX, SEG10466960HX, SEG10466961HX, SEG10466962HX, SEG10466963HX, SEG10466964HX, SEG10466965HX, SEG10466966HX, SEG10466967HX, SEG10466968HX, SEG10466969HX, SEG10466970HX, SEG10466971HX, SEG10466972HX, SEG10466973HX, SEG10466974HX, SEG10466975HX, SEG10466976HX, SEG10466977HX, SEG10466978HX, SEG10466979HX, SEG10466980HX, SEG10466981HX, SEG10466982HX, SEG10466983HX, SEG10466984HX, SEG10466985HX, SEG10466986HX, SEG10477387HX, SEG10477388HX, SEG10477389HX, SEG10477390HX, SEG10477391HX, SEG10477392HX, SEG10477393HX, SEG10477394HX, SEG10477395HX, SEG10477396HX, SEG10477397HX, SEG10477398HX, SEG10477399HX, SEG10477400HX, SEG10477401HX, SEG10477402HX, SEG10477403HX, SEG10477404HX, SEG10477405HX, SEG10477406HX, SEG10477407HX, SEG10477408HX, SEG10477409HX, SEG10477410HX, SEG10477411HX, SEG10477412HX, SEG10477423HX, SEG10477424HX, SEG10477425HX, SEG10477426HX, SEG10477427HX, SEG10477428HX, SEG10477429HX, SEG10477430HX, SEG10477431HX, SEG10477432HX, SEG10477433HX, SEG10477434HX, SEG10477435HX, SEG10477436HX, SEG10477437HX, SEG10477438HX, SEG10477439HX, SEG10477440HX, SEG10477441HX, SEG10477442HX, SEG10477443HX, SEG10477444HX, SEG10477445HX, SEG10477446HX, SEG10477447HX, SEG10477448HX, SEG10477449HX, SEG10477450HX, SEG10477451HX, SEG10477452HX, SEG10477453HX, SEG10477454HX, SEG10477455HX, SEG10477456HX, SEG10477457HX, SEG10477458HX, SEG10477459HX, SEG10477460HX, SEG10477461HX, SEG10477462HX, SEG10477463HX, SEG10477464HX, SEG10477465HX, SEG10477466HX, SEG10477467HX, SEG10477468HX, SEG10477469HX, SEG10477470HX, SEG10477471HX, SEG10477472HX, SEG10477473HX, SEG10477474HX, SEG10477475HX, SEG10477476HX, SEG10477477HX, SEG10477478HX, SEG10477479HX, SEG10477480HX, SEG10477481HX, SEG10477482HX, SEG10477483HX, SEG10477484HX, SEG10477485HX, SEG10477486HX, SEG10507687HX, SEG10507688HX, SEG10507689HX, SEG10507690HX, SEG10507691HX, SEG10507692HX, SEG10507693HX, SEG10507694HX, SEG10507695HX, SEG10507696HX, SEG10507697HX, SEG10507698HX, SEG10507699HX, SEG10507700HX, SEG10507701HX, SEG10507702HX, SEG10507703HX, SEG10507704HX, SEG10507705HX, SEG10507706HX, SEG10507707HX, SEG10507708HX, SEG10507709HX, SEG10507710HX, SEG10507711HX, SEG10507712HX, SEG10507713HX, SEG10507714HX, SEG10507715HX, SEG10507716HX, SEG10507717HX, SEG10507718HX, SEG10507719HX, SEG10507720HX, SEG10507721HX, SEG10507722HX, SEG10507723HX, SEG10507724HX, SEG10507725HX, SEG10507726HX, SEG10507727HX, SEG10507728HX, SEG10507729HX, SEG10507730HX, SEG10507731HX, SEG10507732HX, SEG10507733HX, SEG10507734HX, SEG10507735HX, SEG10507736HX, SEG10507737HX, SEG10507738HX, SEG10507739HX, SEG10507740HX, SEG10507741HX, SEG10507742HX, SEG10507743HX, SEG10507744HX, SEG10507745HX, SEG10507746HX, SEG10507747HX, SEG10507748HX, SEG10507749HX, SEG10507750HX, SEG10507751HX, SEG10507752HX, SEG10507753HX, SEG10507754HX, SEG10507755HX, SEG10507756HX, SEG10507757HX, SEG10507758HX, SEG10507759HX, SEG10507760HX, SEG10507761HX, SEG10507762HX, SEG10507763HX, SEG10507764HX, SEG10507765HX, SEG10507766HX, SEG10507767HX, SEG10507768HX, SEG10507769HX, SEG10507770HX, SEG10507771HX, SEG10507772HX, SEG10507773HX, SEG10507774HX, SEG10507775HX, SEG10507776HX, SEG10507777HX, SEG10507778HX, SEG10507779HX, SEG10507780HX, SEG10507781HX, SEG10507782HX, SEG10507783HX, SEG10507784HX, SEG10507785HX, SEG10507786HX, SEG10507887HX, SEG10507888HX, SEG10507889HX, SEG10507890HX, SEG10507891HX, SEG10507892HX, SEG10507893HX, SEG10507894HX, SEG10507895HX, SEG10507896HX, SEG10507897HX, SEG10507898HX, SEG10507899HX, SEG10507900HX, SEG10507901HX, SEG10507902HX, SEG10507903HX, SEG10507904HX, SEG10507905HX, SEG10507906HX, SEG10507907HX, SEG10507908HX, SEG10507909HX, SEG10507910HX, SEG10507911HX, SEG10507912HX, SEG10507913HX, SEG10507914HX, SEG10507915HX, SEG10507916HX, SEG10507917HX, SEG10507918HX, SEG10507919HX, SEG10507920HX, SEG10507921HX, SEG10507922HX, SEG10507923HX, SEG10507924HX, SEG10507925HX, SEG10507926HX, SEG10507927HX, SEG10507928HX, SEG10507929HX, SEG10507930HX, SEG10507931HX, SEG10507932HX, SEG10507933HX, SEG10507934HX, SEG10507935HX, SEG10507936HX, SEG10518037HX, SEG10518038HX, SEG10518039HX, SEG10518040HX, SEG10518041HX, SEG10518042HX, SEG10518043HX, SEG10518044HX, SEG10518045HX, SEG10518046HX, SEG10518047HX, SEG10518048HX, SEG10518049HX, SEG10518050HX, SEG10518051HX, SEG10518052HX, SEG10518053HX, SEG10518054HX, SEG10518055HX, SEG10518056HX, SEG10528157HX, SEG10528158HX, SEG10528159HX, SEG10528160HX, SEG10528161HX, SEG10528162HX, SEG10528163HX, SEG10528164HX, SEG10528165HX, SEG10528166HX, SEG10528167HX, SEG10528168HX, SEG10528169HX, SEG10528170HX, SEG10528171HX, SEG10528172HX, SEG10528173HX, SEG10528174HX, SEG10528175HX, SEG10528176HX, SEG10528177HX, SEG10528178HX, SEG10528179HX, SEG10528180HX, SEG10528181HX, SEG10528182HX, SEG10528183HX, SEG10528184HX, SEG10528185HX, SEG10528186HX, SEG10528187HX, SEG10528188HX, SEG10528189HX, SEG10528190HX, SEG10528191HX, SEG10528192HX, SEG10528193HX, SEG10528194HX, SEG10528195HX, SEG10528196HX, SEG10528257HX, SEG10528258HX, SEG10528259HX, SEG10528260HX, SEG10528261HX, SEG10528262HX, SEG10528263HX, SEG10528264HX, SEG10528265HX, SEG10528266HX, SEG10528267HX, SEG10528268HX, SEG10528269HX, SEG10528270HX, SEG10528271HX, SEG10528272HX, SEG10528273HX, SEG10528274HX, SEG10528275HX, SEG10528276HX, SEG10528281HX, SEG10528282HX, SEG10528283HX, SEG10528284HX, SEG10528285HX, SEG10528286HX, SEG10528287HX, SEG10528288HX, SEG10528289HX, SEG10528290HX, SEG10528291HX, SEG10528292HX, SEG10528293HX, SEG10528294HX, SEG10528295HX, SEG10528296HX, SEG10528297HX, SEG10528298HX, SEG10528299HX, SEG10528300HX, SEG10528301HX, SEG10528302HX, SEG10528303HX, SEG10528304HX, SEG10528305HX, SEG10528306HX, SEG10528307HX, SEG10528308HX, SEG10528309HX, SEG10528316HX, SEG10528317HX, SEG10528318HX, SEG10528319HX, SEG10528320HX, SEG10528321HX, SEG10528322HX, SEG10528323HX, SEG10528324HX, SEG10528325HX, SEG11018331HX, SEG11018332HX, SEG11018333HX, SEG11018334HX, SEG11018335HX, SEG11018336HX, SEG11018337HX, SEG11018338HX, SEG11018339HX, SEG11018340HX, SEG11018341HX, SEG11018342HX, SEG11018343HX, SEG11018344HX, SEG11018345HX, SEG11018346HX, SEG11018347HX, SEG11018348HX, SEG11018349HX, SEG11018350HX, SEG11018351HX, SEG11018352HX, SEG11018353HX, SEG11018354HX, SEG11018355HX, SEG11018356HX, SEG11018357HX, SEG11018358HX, SEG11018359HX, SEG11018360HX, SEG11018361HX, SEG11018362HX, SEG11018363HX, SEG11018364HX, SEG11018365HX, SEG11018366HX, SEG11018367HX, SEG11018368HX, SEG11018369HX, SEG11018370HX, SEG11018371HX, SEG11018372HX, SEG11018373HX, SEG11018374HX, SEG11018375HX, SEG11018376HX, SEG11018377HX, SEG11018378HX, SEG11018379HX, SEG11018380HX, SEG11018381HX, SEG11018382HX, SEG11018383HX, SEG11018384HX, SEG11018385HX, SEG11018386HX, SEG11018387HX, SEG11018388HX, SEG11018389HX, SEG11018390HX, SEG11018391HX, SEG11018392HX, SEG11018393HX, SEG11018394HX, SEG11018395HX, SEG11018396HX, SEG11018397HX, SEG11018398HX, SEG11018399HX, SEG11018400HX, SEG11018401HX, SEG11018402HX, SEG11018404HX, SEG11018405HX, SEG11018406HX, SEG11018407HX, SEG11018408HX, SEG11018409HX, SEG11018410HX, SEG11018411HX, SEG11018412HX, SEG11018413HX, SEG11018414HX, SEG11018415HX, SEG11018416HX, SEG11018417HX, SEG11018418HX, SEG11018419HX, SEG11018420HX, SEG11018421HX, SEG11018422HX, SEG11018423HX, SEG11018424HX, SEG11018425HX, SEG11018426HX, SEG11018427HX, SEG11018428HX, SEG11018429HX, SEG11018430HX, SEG11028751HX, SEG11028752HX, SEG11028753HX, SEG11028754HX, SEG11028755HX, SEG11028756HX, SEG11028757HX, SEG11028758HX, SEG11028759HX, SEG11028760HX, SEG11028761HX, SEG11028762HX, SEG11028763HX, SEG11028764HX, SEG11028765HX, SEG11028766HX, SEG11028767HX, SEG11028768HX, SEG11028769HX, SEG11028770HX, SEG11028771HX, SEG11028772HX, SEG11028773HX, SEG11028774HX, SEG11028775HX, SEG11028776HX, SEG11028777HX, SEG11028778HX, SEG11028779HX, SEG11028780HX, SEG11028781HX, SEG11028782HX, SEG11028783HX, SEG11028784HX, SEG11028785HX, SEG11028786HX, SEG11028787HX, SEG11028788HX, SEG11028789HX, SEG11028790HX, SEG11028791HX, SEG11028792HX, SEG11028793HX, SEG11028794HX, SEG11028795HX, SEG11028796HX, SEG11028797HX, SEG11028798HX, SEG11028799HX, SEG11028800HX, SEG11028801HX, SEG11028802HX, SEG11028803HX, SEG11028804HX, SEG11028805HX, SEG11028806HX, SEG11028807HX, SEG11028808HX, SEG11028809HX, SEG11028810HX, SEG11028811HX, SEG11028812HX, SEG11028813HX, SEG11028814HX, SEG11028815HX, SEG11028816HX, SEG11028817HX, SEG11028818HX, SEG11028819HX, SEG11028820HX, SEG11028821HX, SEG11028822HX, SEG11028823HX, SEG11028824HX, SEG11028825HX, SEG11028826HX, SEG11028827HX, SEG11028828HX, SEG11028829HX, SEG11028830HX, SEG11028831HX, SEG11028832HX, SEG11028833HX, SEG11028834HX, SEG11028835HX, SEG11028836HX, SEG11028837HX, SEG11028838HX, SEG11028839HX, SEG11028840HX, SEG11028841HX, SEG11028842HX, SEG11028843HX, SEG11028844HX, SEG11028845HX, SEG11028846HX, SEG11028847HX, SEG11028848HX, SEG11028849HX, SEG11028850HX, SEG11048951HX, SEG11048952HX, SEG11048953HX, SEG11048954HX, SEG11048955HX, SEG11048956HX, SEG11048957HX, SEG11048958HX, SEG11048959HX, SEG11048960HX, SEG11048961HX, SEG11048962HX, SEG11048963HX, SEG11048964HX, SEG11048965HX, SEG11048966HX, SEG11048967HX, SEG11048968HX, SEG11048969HX, SEG11048970HX, SEG11048971HX, SEG11048972HX, SEG11048973HX, SEG11048974HX, SEG11048975HX, SEG11048976HX, SEG11048977HX, SEG11048978HX, SEG11048979HX, SEG11048980HX, SEG11048981HX, SEG11048982HX, SEG11048983HX, SEG11048984HX, SEG11048985HX, SEG11048986HX, SEG11048987HX, SEG11048988HX, SEG11048989HX, SEG11048990HX, SEG11048991HX, SEG11048992HX, SEG11048993HX, SEG11048994HX, SEG11048995HX, SEG11048996HX, SEG11048997HX, SEG11048998HX, SEG11048999HX, SEG11049000HX, SEG11049001HX, SEG11049002HX, SEG11049003HX, SEG11049004HX, SEG11049005HX, SEG11049006HX, SEG11049007HX, SEG11049008HX, SEG11049009HX, SEG11049010HX, SEG11049011HX, SEG11049012HX, SEG11049013HX, SEG11049014HX, SEG11049015HX, SEG11049016HX, SEG11049017HX, SEG11049018HX, SEG11049019HX, SEG11049020HX, SEG11049021HX, SEG11049022HX, SEG11049023HX, SEG11049024HX, SEG11049025HX, SEG11049026HX, SEG11049027HX, SEG11049028HX, SEG11049029HX, SEG11049030HX, SEG11049031HX, SEG11049032HX, SEG11049033HX, SEG11049034HX, SEG11049035HX, SEG11049036HX, SEG11049037HX, SEG11049038HX, SEG11049039HX, SEG11049040HX, SEG11049041HX, SEG11049042HX, SEG11049043HX, SEG11049044HX, SEG11049045HX, SEG11049046HX, SEG11049047HX, SEG11049048HX, SEG11049049HX, SEG11049050HX, SEG11059251HX, SEG11059252HX, SEG11059253HX, SEG11059254HX, SEG11059255HX, SEG11059256HX, SEG11059257HX, SEG11059258HX, SEG11059259HX, SEG11059260HX, SEG11059261HX, SEG11059262HX, SEG11059263HX, SEG11059264HX, SEG11059265HX, SEG11059266HX, SEG11059267HX, SEG11059268HX, SEG11059269HX, SEG11059270HX, SEG11059271HX, SEG11059272HX, SEG11059273HX, SEG11059274HX, SEG11059275HX, SEG11059276HX, SEG11059277HX, SEG11059278HX, SEG11059279HX, SEG11059280HX, SEG11059281HX, SEG11059282HX, SEG11059283HX, SEG11059284HX, SEG11059285HX, SEG11059286HX, SEG11059287HX, SEG11059288HX, SEG11059289HX, SEG11059290HX, SEG11059291HX, SEG11059292HX, SEG11059293HX, SEG11059294HX, SEG11059295HX, SEG11059296HX, SEG11059297HX, SEG11059298HX, SEG11059299HX, SEG11059300HX, SEG11059301HX, SEG11059302HX, SEG11059303HX, SEG11059304HX, SEG11059305HX, SEG11059306HX, SEG11059307HX, SEG11059308HX, SEG11059309HX, SEG11059310HX, SEG11059311HX, SEG11059312HX, SEG11059313HX, SEG11059314HX, SEG11059315HX, SEG11059316HX, SEG11059317HX, SEG11059318HX, SEG11059319HX, SEG11059320HX, SEG11059321HX, SEG11059322HX, SEG11059323HX, SEG11059324HX, SEG11059325HX, SEG11059326HX, SEG11059327HX, SEG11059328HX, SEG11059329HX, SEG11059330HX, SEG11059331HX, SEG11059332HX, SEG11059333HX, SEG11059334HX, SEG11059335HX, SEG11059336HX, SEG11059337HX, SEG11059338HX, SEG11059339HX, SEG11059340HX, SEG11059341HX, SEG11059342HX, SEG11059343HX, SEG11059344HX, SEG11059345HX, SEG11059346HX, SEG11059347HX, SEG11059348HX, SEG11059349HX, SEG11059350HX, SEG11079651HX, SEG11079652HX, SEG11079653HX, SEG11079654HX, SEG11079655HX, SEG11079656HX, SEG11079657HX, SEG11079658HX, SEG11079659HX, SEG11079660HX, SEG11079661HX, SEG11079662HX, SEG11079663HX, SEG11079664HX, SEG11079665HX, SEG11079666HX, SEG11079667HX, SEG11079668HX, SEG11079669HX, SEG11079670HX, SEG11079671HX, SEG11079672HX, SEG11079673HX, SEG11079674HX, SEG11079675HX, SEG11079676HX, SEG11079677HX, SEG11079678HX, SEG11079679HX, SEG11079680HX, SEG11079681HX, SEG11079682HX, SEG11079683HX, SEG11079684HX, SEG11079685HX, SEG11079686HX, SEG11079687HX, SEG11079688HX, SEG11079689HX, SEG11079690HX, SEG11079691HX, SEG11079692HX, SEG11079693HX, SEG11079694HX, SEG11079695HX, SEG11079696HX, SEG11079697HX, SEG11079698HX, SEG11079699HX, SEG11079700HX, SEG11079701HX, SEG11079702HX, SEG11079703HX, SEG11079704HX, SEG11079705HX, SEG11079706HX, SEG11079707HX, SEG11079708HX, SEG11079709HX, SEG11079710HX, SEG11079711HX, SEG11079712HX, SEG11079713HX, SEG11079714HX, SEG11079715HX, SEG11079716HX, SEG11079717HX, SEG11079718HX, SEG11079719HX, SEG11079720HX, SEG11079721HX, SEG11079722HX, SEG11079723HX, SEG11079724HX, SEG11079725HX, SEG11079726HX, SEG11079727HX, SEG11079728HX, SEG11079729HX, SEG11079730HX, SEG11079731HX, SEG11079732HX, SEG11079733HX, SEG11079734HX, SEG11079735HX, SEG11079736HX, SEG11079737HX, SEG11079738HX, SEG11079739HX, SEG11079740HX, SEG11079741HX, SEG11079742HX, SEG11079743HX, SEG11079744HX, SEG11079745HX, SEG11079746HX, SEG11079747HX, SEG11079748HX, SEG11079749HX, SEG11079750HX, SEG11089751HX, SEG11089752HX, SEG11089753HX, SEG11089754HX, SEG11089755HX, SEG11089756HX, SEG11089757HX, SEG11089758HX, SEG11089759HX, SEG11089760HX, SEG11089761HX, SEG11089762HX, SEG11089763HX, SEG11089764HX, SEG11089765HX, SEG11089766HX, SEG11089767HX, SEG11089768HX, SEG11089769HX, SEG11089770HX, SEG11089771HX, SEG11089772HX, SEG11089773HX, SEG11089774HX, SEG11089775HX, SEG11089776HX, SEG11089777HX, SEG11089778HX, SEG11089779HX, SEG11089780HX, SEG11089781HX, SEG11089782HX, SEG11089783HX, SEG11089784HX, SEG11089785HX, SEG11089786HX, SEG11089787HX, SEG11089788HX, SEG11089789HX, SEG11089790HX, SEG11089791HX, SEG11089792HX, SEG11089793HX, SEG11089794HX, SEG11089795HX, SEG11089796HX, SEG11089797HX, SEG11089798HX, SEG11089799HX, SEG11089800HX, SEG11089801HX, SEG11089802HX, SEG11089803HX, SEG11089804HX, SEG11089805HX, SEG11089806HX, SEG11089807HX, SEG11089808HX, SEG11089809HX, SEG11089810HX, SEG11089811HX, SEG11089812HX, SEG11089813HX, SEG11089814HX, SEG11089815HX, SEG11089816HX, SEG11089817HX, SEG11089818HX, SEG11089819HX, SEG11089820HX, SEG11089821HX, SEG11089822HX, SEG11089823HX, SEG11089824HX, SEG11089825HX, SEG11089826HX, SEG11089827HX, SEG11089828HX, SEG11089829HX, SEG11089830HX, SEG11089831HX, SEG11089832HX, SEG11089833HX, SEG11089834HX, SEG11089835HX, SEG11089836HX, SEG11089837HX, SEG11089838HX, SEG11089839HX, SEG11089840HX, SEG11089841HX, SEG11089842HX, SEG11089843HX, SEG11089844HX, SEG11089845HX, SEG11089846HX, SEG11089847HX, SEG11089848HX, SEG11089849HX, SEG11089850HX, SEG11100231HX, SEG11100232HX, SEG11100233HX, SEG11100234HX, SEG11100235HX, SEG11100236HX, SEG11100237HX, SEG11100238HX, SEG11100239HX, SEG11100240HX, SEG11100241HX, SEG11100242HX, SEG11100243HX, SEG11100244HX, SEG11100245HX, SEG11100246HX, SEG11100247HX, SEG11100248HX, SEG11100249HX, SEG11100250HX, SEG11100251HX, SEG11100252HX, SEG11100253HX, SEG11100254HX, SEG11100255HX, SEG11100256HX, SEG11100257HX, SEG11100258HX, SEG11100259HX, SEG11100260HX, SEG11100261HX, SEG11100262HX, SEG11100263HX, SEG11100264HX, SEG11100265HX, SEG11100266HX, SEG11100267HX, SEG11100268HX, SEG11100269HX, SEG11100270HX, SEG11100271HX, SEG11100272HX, SEG11100273HX, SEG11100274HX, SEG11100275HX, SEG11100276HX, SEG11100277HX, SEG11100278HX, SEG11100279HX, SEG11100280HX, SEG11100281HX, SEG11100282HX, SEG11100283HX, SEG11100284HX, SEG11100285HX, SEG11100286HX, SEG11100287HX, SEG11100288HX, SEG11100289HX, SEG11100290HX, SEG11100291HX, SEG11100292HX, SEG11100293HX, SEG11100294HX, SEG11100295HX, SEG11100296HX, SEG11100297HX, SEG11100298HX, SEG11100299HX, SEG11100300HX, SEG11100301HX, SEG11100302HX, SEG11100303HX, SEG11100304HX, SEG11100305HX, SEG11100306HX, SEG11100307HX, SEG11100308HX, SEG11100309HX, SEG11100310HX, SEG11100311HX, SEG11100312HX, SEG11100313HX, SEG11100314HX, SEG11100315HX, SEG11100316HX, SEG11100317HX, SEG11100318HX, SEG11100319HX, SEG11100320HX, SEG11100321HX, SEG11100322HX, SEG11100323HX, SEG11100324HX, SEG11100325HX, SEG11100326HX, SEG11100327HX, SEG11100328HX, SEG11100329HX, SEG11100330HX, SEG11110337HX, SEG11110338HX, SEG11110339HX, SEG11110340HX, SEG11110341HX, SEG11110342HX, SEG11110343HX, SEG11110344HX, SEG11110345HX, SEG11110346HX, SEG11110347HX, SEG11110348HX, SEG11110349HX, SEG11110350HX, SEG11110351HX, SEG11110352HX, SEG11110353HX, SEG11110354HX, SEG11110355HX, SEG11110356HX, SEG11110357HX, SEG11110358HX, SEG11110359HX, SEG11110360HX, SEG11110361HX, SEG11110362HX, SEG11110363HX, SEG11110364HX, SEG11110365HX, SEG11110366HX, SEG11110367HX, SEG11110368HX, SEG11110369HX, SEG11110370HX, SEG11110371HX, SEG11110372HX, SEG11110373HX, SEG11110374HX, SEG11110375HX, SEG11110376HX, SEG11110377HX, SEG11110378HX, SEG11110379HX, SEG11110380HX, SEG11110381HX, SEG11110382HX, SEG11110383HX, SEG11110384HX, SEG11110385HX, SEG11110386HX, SEG11110387HX, SEG11110388HX, SEG11110389HX, SEG11110390HX, SEG11110391HX, SEG11110392HX, SEG11110393HX, SEG11110394HX, SEG11110395HX, SEG11110396HX, SEG11110397HX, SEG11110398HX, SEG11110399HX, SEG11110400HX, SEG11110401HX, SEG11110402HX, SEG11110403HX, SEG11110404HX, SEG11110405HX, SEG11110406HX, SEG11110407HX, SEG11110408HX, SEG11110409HX, SEG11110410HX, SEG11110411HX, SEG11110412HX, SEG11110413HX, SEG11110414HX, SEG11110415HX, SEG11110416HX, SEG11110417HX, SEG11110418HX, SEG11110419HX, SEG11110420HX, SEG11110421HX, SEG11110422HX, SEG11110423HX, SEG11110424HX, SEG11110425HX, SEG11110426HX, SEG11110427HX, SEG11110428HX, SEG11110429HX, SEG11110430HX, SEG11110431HX, SEG11110432HX, SEG11110433HX, SEG11110434HX, SEG11110435HX, SEG11110436HX, SEG11120487HX, SEG11120488HX, SEG11120489HX, SEG11120490HX, SEG11120491HX, SEG11120492HX, SEG11120493HX, SEG11120494HX, SEG11120495HX, SEG11120496HX, SEG11120497HX, SEG11120498HX, SEG11120499HX, SEG11120500HX, SEG11120501HX, SEG11120502HX, SEG11120503HX, SEG11120504HX, SEG11120505HX, SEG11120506HX, SEG11120507HX, SEG11120508HX, SEG11120509HX, SEG11120510HX, SEG11120511HX, SEG11120512HX, SEG11120513HX, SEG11120514HX, SEG11120515HX, SEG11120516HX, SEG11120517HX, SEG11120518HX, SEG11120519HX, SEG11120520HX, SEG11120521HX, SEG11120522HX, SEG11120523HX, SEG11120524HX, SEG11120525HX, SEG11120526HX, SEG11120527HX, SEG11120528HX, SEG11120529HX, SEG11120530HX, SEG11120531HX, SEG11120532HX, SEG11120533HX, SEG11120534HX, SEG11120535HX, SEG11120536HX, SEG11120537HX, SEG11120538HX, SEG11120539HX, SEG11120540HX, SEG11120541HX, SEG11120542HX, SEG11120543HX, SEG11120544HX, SEG11120545HX, SEG11120546HX, SEG11120547HX, SEG11120548HX, SEG11120549HX, SEG11120550HX, SEG11120551HX, SEG11120552HX, SEG11120553HX, SEG11120554HX, SEG11120555HX, SEG11120556HX, SEG11120557HX, SEG11120558HX, SEG11120559HX, SEG11120560HX, SEG11120561HX, SEG11120562HX, SEG11120563HX, SEG11120564HX, SEG11120565HX, SEG11120566HX, SEG11120567HX, SEG11120568HX, SEG11120569HX, SEG11120570HX, SEG11120571HX, SEG11120572HX, SEG11120573HX, SEG11120574HX, SEG11120575HX, SEG11120576HX, SEG11120577HX, SEG11120578HX, SEG11120579HX, SEG11120580HX, SEG11120581HX, SEG11120582HX, SEG11120583HX, SEG11120584HX, SEG11120585HX, SEG11120586HX, SEG11120587HX, SEG11140888HX, SEG11140889HX, SEG11140890HX, SEG11140891HX, SEG11140892HX, SEG11140893HX, SEG11140894HX, SEG11140895HX, SEG11140896HX, SEG11140897HX, SEG11140898HX, SEG11140899HX, SEG11140900HX, SEG11140901HX, SEG11140902HX, SEG11140903HX, SEG11140904HX, SEG11140905HX, SEG11140906HX, SEG11140907HX, SEG11140908HX, SEG11140909HX, SEG11140910HX, SEG11140911HX, SEG11140912HX, SEG11140913HX, SEG11140914HX, SEG11140915HX, SEG11140916HX, SEG11140917HX, SEG11140918HX, SEG11140919HX, SEG11140920HX, SEG11140921HX, SEG11140922HX, SEG11140923HX, SEG11140924HX, SEG11140925HX, SEG11140926HX, SEG11140927HX, SEG11140928HX, SEG11140929HX, SEG11140930HX, SEG11140931HX, SEG11140932HX, SEG11140933HX, SEG11140934HX, SEG11140935HX, SEG11140936HX, SEG11140937HX, SEG11140938HX, SEG11140939HX, SEG11140940HX, SEG11140941HX, SEG11140942HX, SEG11140943HX, SEG11140944HX, SEG11140945HX, SEG11140946HX, SEG11140947HX, SEG11140948HX, SEG11140949HX, SEG11140950HX, SEG11140951HX, SEG11140952HX, SEG11140953HX, SEG11140954HX, SEG11140955HX, SEG11140956HX, SEG11140957HX, SEG11140958HX, SEG11140959HX, SEG11140960HX, SEG11140961HX, SEG11140962HX, SEG11140963HX, SEG11140964HX, SEG11140965HX, SEG11140966HX, SEG11140967HX, SEG11140968HX, SEG11140969HX, SEG11140970HX, SEG11140971HX, SEG11140972HX, SEG11140973HX, SEG11140974HX, SEG11140975HX, SEG11140976HX, SEG11140977HX, SEG11140978HX, SEG11140979HX, SEG11140980HX, SEG11140981HX, SEG11140982HX, SEG11140983HX, SEG11140984HX, SEG11140985HX, SEG11140986HX, SEG11140987HX, SEG11161338HX, SEG11161339HX, SEG11161340HX, SEG11161341HX, SEG11161342HX, SEG11161343HX, SEG11161344HX, SEG11161345HX, SEG11161346HX, SEG11161347HX, SEG11161348HX, SEG11161349HX, SEG11161350HX, SEG11161351HX, SEG11161352HX, SEG11161353HX, SEG11161354HX, SEG11161355HX, SEG11161356HX, SEG11161357HX, SEG11161358HX, SEG11161359HX, SEG11161360HX, SEG11161361HX, SEG11161362HX, SEG11161363HX, SEG11161364HX, SEG11161365HX, SEG11161366HX, SEG11161367HX, SEG11161368HX, SEG11161369HX, SEG11161370HX, SEG11161371HX, SEG11161372HX, SEG11161373HX, SEG11161374HX, SEG11161375HX, SEG11161376HX, SEG11161377HX, SEG11161378HX, SEG11161379HX, SEG11161380HX, SEG11161381HX, SEG11161382HX, SEG11161383HX, SEG11161384HX, SEG11161385HX, SEG11161386HX, SEG11161387HX, SEG11161388HX, SEG11161389HX, SEG11161390HX, SEG11161391HX, SEG11161392HX, SEG11161393HX, SEG11161394HX, SEG11161395HX, SEG11161396HX, SEG11161397HX, SEG11161398HX, SEG11161399HX, SEG11161400HX, SEG11161401HX, SEG11161402HX, SEG11161403HX, SEG11161404HX, SEG11161405HX, SEG11161406HX, SEG11161407HX, SEG11161408HX, SEG11161409HX, SEG11161410HX, SEG11161411HX, SEG11161412HX, SEG11161413HX, SEG11161414HX, SEG11161415HX, SEG11161416HX, SEG11161417HX, SEG11161418HX, SEG11161419HX, SEG11161420HX, SEG11161421HX, SEG11161422HX, SEG11161423HX, SEG11161424HX, SEG11161425HX, SEG11161426HX, SEG11161427HX, SEG11161428HX, SEG11161429HX, SEG11161430HX, SEG11161431HX, SEG11161432HX, SEG11161433HX, SEG11161434HX, SEG11161435HX, SEG11161436HX, SEG11161437HX, SEG11171542HX, SEG11171543HX, SEG11171544HX, SEG11171545HX, SEG11171546HX, SEG11171547HX, SEG11171548HX, SEG11171549HX, SEG11171550HX, SEG11171551HX, SEG11171552HX, SEG11171553HX, SEG11171554HX, SEG11171555HX, SEG11171556HX, SEG11171557HX, SEG11171558HX, SEG11171559HX, SEG11171560HX, SEG11171561HX, SEG11171562HX, SEG11171563HX, SEG11171564HX, SEG11171565HX, SEG11171566HX, SEG11171567HX, SEG11171568HX, SEG11171569HX, SEG11171570HX, SEG11171571HX, SEG11171572HX, SEG11171573HX, SEG11171574HX, SEG11171575HX, SEG11171576HX, SEG11171577HX, SEG11171578HX, SEG11171579HX, SEG11171580HX, SEG11171581HX, SEG11171582HX, SEG11171583HX, SEG11171584HX, SEG11171585HX, SEG11171586HX, SEG11171587HX, SEG11171588HX, SEG11171589HX, SEG11171590HX, SEG11171591HX, SEG11171592HX, SEG11171593HX, SEG11171594HX, SEG11171595HX, SEG11171596HX, SEG11171597HX, SEG11171598HX, SEG11171599HX, SEG11171600HX, SEG11171601HX, SEG11171602HX, SEG11171603HX, SEG11171604HX, SEG11171605HX, SEG11171606HX, SEG11171607HX, SEG11171608HX, SEG11171609HX, SEG11171610HX, SEG11171611HX, SEG11171612HX, SEG11171613HX, SEG11171614HX, SEG11171615HX, SEG11171616HX, SEG11171617HX, SEG11171618HX, SEG11171619HX, SEG11171620HX, SEG11171621HX, SEG11171622HX, SEG11171623HX, SEG11171624HX, SEG11171625HX, SEG11171626HX, SEG11171627HX, SEG11171628HX, SEG11171629HX, SEG11171630HX, SEG11171631HX, SEG11171632HX, SEG11171633HX, SEG11171634HX, SEG11171635HX, SEG11171636HX, SEG11171637HX, SEG11171638HX, SEG11171639HX, SEG11171640HX, SEG11171641HX, SEG11181652HX, SEG11181653HX, SEG11181654HX, SEG11181655HX, SEG11181656HX, SEG11181657HX, SEG11181658HX, SEG11181659HX, SEG11181660HX, SEG11181661HX, SEG11181662HX, SEG11181663HX, SEG11181664HX, SEG11181665HX, SEG11181666HX, SEG11181667HX, SEG11181668HX, SEG11181669HX, SEG11181670HX, SEG11181671HX, SEG11181672HX, SEG11181673HX, SEG11181674HX, SEG11181675HX, SEG11181676HX, SEG11181677HX, SEG11181678HX, SEG11181679HX, SEG11181680HX, SEG11181681HX, SEG11181682HX, SEG11181683HX, SEG11181684HX, SEG11181685HX, SEG11181686HX, SEG11181687HX, SEG11181688HX, SEG11181689HX, SEG11181690HX, SEG11181691HX, SEG11181692HX, SEG11181693HX, SEG11181694HX, SEG11181695HX, SEG11181696HX, SEG11181697HX, SEG11181698HX, SEG11181699HX, SEG11181700HX, SEG11181701HX, SEG11181702HX, SEG11181703HX, SEG11181704HX, SEG11181705HX, SEG11181706HX, SEG11181707HX, SEG11181708HX, SEG11181709HX, SEG11181710HX, SEG11181711HX, SEG11181712HX, SEG11181713HX, SEG11181714HX, SEG11181715HX, SEG11181716HX, SEG11181717HX, SEG11181718HX, SEG11181719HX, SEG11181720HX, SEG11181721HX, SEG11181722HX, SEG11181723HX, SEG11181724HX, SEG11181725HX, SEG11181726HX, SEG11181727HX, SEG11181728HX, SEG11181729HX, SEG11181730HX, SEG11181731HX, SEG11181732HX, SEG11181733HX, SEG11181734HX, SEG11181735HX, SEG11181736HX, SEG11181737HX, SEG11181738HX, SEG11181739HX, SEG11181740HX, SEG11181741HX, SEG11181742HX, SEG11181743HX, SEG11181744HX, SEG11181745HX, SEG11181746HX, SEG11181747HX, SEG11181748HX, SEG11181749HX, SEG11181750HX, SEG11181751HX, SEG11181952HX, SEG11181953HX, SEG11181954HX, SEG11181955HX, SEG11181956HX, SEG11181957HX, SEG11181958HX, SEG11181959HX, SEG11181960HX, SEG11181961HX, SEG11181962HX, SEG11181963HX, SEG11181964HX, SEG11181965HX, SEG11181966HX, SEG11181967HX, SEG11181968HX, SEG11181969HX, SEG11181970HX, SEG11181971HX, SEG11181972HX, SEG11181973HX, SEG11181974HX, SEG11181975HX, SEG11181976HX, SEG11181977HX, SEG11181978HX, SEG11181979HX, SEG11181980HX, SEG11181981HX, SEG11181982HX, SEG11181983HX, SEG11181984HX, SEG11181985HX, SEG11181986HX, SEG11181987HX, SEG11181988HX, SEG11181989HX, SEG11181990HX, SEG11181991HX, SEG11181992HX, SEG11181993HX, SEG11181994HX, SEG11181995HX, SEG11181996HX, SEG11181997HX, SEG11181998HX, SEG11181999HX, SEG11182000HX, SEG11182001HX, SEG11182002HX, SEG11182003HX, SEG11182004HX, SEG11182005HX, SEG11182006HX, SEG11182007HX, SEG11182008HX, SEG11182009HX, SEG11182010HX, SEG11182011HX, SEG11182012HX, SEG11182013HX, SEG11182014HX, SEG11182015HX, SEG11182016HX, SEG11182017HX, SEG11182018HX, SEG11182019HX, SEG11182020HX, SEG11182021HX, SEG11182022HX, SEG11182023HX, SEG11182024HX, SEG11182025HX, SEG11182026HX, SEG11182027HX, SEG11182028HX, SEG11182029HX, SEG11182030HX, SEG11182031HX, SEG11182032HX, SEG11182033HX, SEG11182034HX, SEG11182035HX, SEG11182036HX, SEG11182037HX, SEG11182038HX, SEG11182039HX, SEG11182040HX, SEG11182041HX, SEG11182042HX, SEG11182043HX, SEG11182044HX, SEG11182045HX, SEG11182046HX, SEG11182047HX, SEG11182048HX, SEG11182049HX, SEG11182050HX, SEG11182051HX, SEG11182052HX, SEG11182053HX, SEG11182054HX, SEG11182055HX, SEG11182056HX, SEG11182057HX, SEG11182058HX, SEG11182059HX, SEG11182060HX, SEG11182061HX, SEG11182062HX, SEG11182063HX, SEG11182064HX, SEG11182065HX, SEG11182066HX, SEG11182067HX, SEG11182068HX, SEG11182069HX, SEG11182070HX, SEG11182071HX, SEG11182072HX, SEG11182073HX, SEG11182074HX, SEG11182075HX, SEG11182076HX, SEG11182077HX, SEG11182078HX, SEG11182079HX, SEG11182080HX, SEG11182081HX, SEG11182082HX, SEG11182083HX, SEG11182084HX, SEG11182085HX, SEG11182086HX, SEG11182087HX, SEG11182088HX, SEG11182089HX, SEG11182090HX, SEG11182091HX, SEG11182092HX, SEG11182093HX, SEG11182094HX, SEG11182095HX, SEG11182096HX, SEG11182097HX, SEG11182098HX, SEG11182099HX, SEG11182100HX, SEG11182101HX, SEG11182102HX, SEG11182103HX, SEG11182104HX, SEG11182105HX, SEG11182106HX, SEG11182107HX, SEG11182108HX, SEG11182109HX, SEG11182110HX, SEG11182111HX, SEG11182112HX, SEG11182113HX, SEG11182114HX, SEG11182115HX, SEG11182116HX, SEG11182117HX, SEG11182118HX, SEG11182119HX, SEG11182120HX, SEG11182121HX, SEG11182122HX, SEG11182123HX, SEG11182124HX, SEG11182125HX, SEG11182126HX, SEG11182127HX, SEG11182128HX, SEG11182129HX, SEG11182130HX, SEG11182131HX, SEG11182132HX, SEG11182133HX, SEG11182134HX, SEG11182135HX, SEG11182136HX, SEG11182137HX, SEG11182138HX, SEG11182139HX, SEG11182140HX, SEG11182141HX, SEG11182142HX, SEG11182143HX, SEG11182144HX, SEG11182145HX, SEG11182146HX, SEG11182147HX, SEG11182148HX, SEG11182149HX, SEG11182150HX, SEG11182151HX, SEG11202482HX, SEG11202483HX, SEG11202484HX, SEG11202485HX, SEG11202486HX, SEG11202487HX, SEG11202488HX, SEG11202489HX, SEG11202490HX, SEG11202491HX, SEG11202492HX, SEG11202493HX, SEG11202494HX, SEG11202495HX, SEG11202496HX, SEG11202497HX, SEG11202498HX, SEG11202499HX, SEG11202500HX, SEG11202501HX, SEG11202502HX, SEG11202503HX, SEG11202504HX, SEG11202505HX, SEG11202506HX, SEG11202507HX, SEG11202508HX, SEG11202509HX, SEG11202510HX, SEG11202511HX, SEG11202512HX, SEG11202513HX, SEG11202514HX, SEG11202515HX, SEG11202516HX, SEG11202517HX, SEG11202518HX, SEG11202519HX, SEG11202520HX, SEG11202521HX, SEG11202522HX, SEG11202523HX, SEG11202524HX, SEG11202525HX, SEG11202526HX, SEG11202527HX, SEG11202528HX, SEG11202529HX, SEG11202530HX, SEG11202531HX, SEG11202532HX, SEG11202533HX, SEG11202534HX, SEG11202535HX, SEG11202536HX, SEG11202537HX, SEG11202538HX, SEG11202539HX, SEG11202540HX, SEG11202541HX, SEG11202542HX, SEG11202543HX, SEG11202544HX, SEG11202545HX, SEG11202546HX, SEG11202547HX, SEG11202548HX, SEG11202549HX, SEG11202550HX, SEG11202551HX, SEG11202552HX, SEG11202553HX, SEG11202554HX, SEG11202555HX, SEG11202556HX, SEG11202557HX, SEG11202558HX, SEG11202559HX, SEG11202560HX, SEG11202561HX, SEG11202562HX, SEG11202563HX, SEG11202564HX, SEG11202565HX, SEG11202566HX, SEG11202567HX, SEG11202568HX, SEG11202569HX, SEG11202570HX, SEG11202571HX, SEG11202572HX, SEG11202573HX, SEG11202574HX, SEG11202575HX, SEG11202576HX, SEG11202577HX, SEG11202578HX, SEG11202579HX, SEG11202580HX, SEG11202581HX, SEG11222897HX, SEG11222898HX, SEG11222899HX, SEG11222900HX, SEG11222901HX, SEG11222902HX, SEG11222903HX, SEG11222904HX, SEG11222905HX, SEG11222906HX, SEG11222907HX, SEG11222908HX, SEG11222909HX, SEG11222910HX, SEG11222911HX, SEG11222912HX, SEG11222913HX, SEG11222914HX, SEG11222915HX, SEG11222916HX, SEG11222917HX, SEG11222918HX, SEG11222919HX, SEG11222920HX, SEG11222921HX, SEG11222922HX, SEG11222923HX, SEG11222924HX, SEG11222925HX, SEG11222926HX, SEG11222927HX, SEG11222928HX, SEG11222929HX, SEG11222930HX, SEG11222931HX, SEG11222932HX, SEG11222933HX, SEG11222934HX, SEG11222935HX, SEG11222936HX, SEG11222937HX, SEG11222938HX, SEG11222939HX, SEG11222940HX, SEG11222941HX, SEG11222942HX, SEG11222943HX, SEG11222944HX, SEG11222945HX, SEG11222946HX, SEG11222947HX, SEG11222948HX, SEG11222949HX, SEG11222950HX, SEG11222951HX, SEG11222952HX, SEG11222953HX, SEG11222954HX, SEG11222955HX, SEG11222956HX, SEG11222957HX, SEG11222958HX, SEG11222959HX, SEG11222960HX, SEG11222961HX, SEG11222962HX, SEG11222963HX, SEG11222964HX, SEG11222965HX, SEG11222966HX, SEG11222967HX, SEG11222968HX, SEG11222969HX, SEG11222970HX, SEG11222971HX, SEG11222972HX, SEG11222973HX, SEG11222974HX, SEG11222975HX, SEG11222976HX, SEG11222977HX, SEG11222978HX, SEG11222979HX, SEG11222980HX, SEG11222981HX, SEG11222982HX, SEG11222983HX, SEG11222984HX, SEG11222985HX, SEG11222986HX, SEG11222987HX, SEG11222988HX, SEG11222989HX, SEG11222990HX, SEG11222991HX, SEG11222992HX, SEG11222993HX, SEG11222994HX, SEG11222995HX, SEG11222996HX, SEG11233147HX, SEG11233148HX, SEG11233149HX, SEG11233150HX, SEG11233151HX, SEG11233152HX, SEG11233153HX, SEG11233154HX, SEG11233155HX, SEG11233156HX, SEG11233157HX, SEG11233158HX, SEG11233159HX, SEG11233160HX, SEG11233161HX, SEG11233162HX, SEG11233163HX, SEG11233164HX, SEG11233165HX, SEG11233166HX, SEG11233167HX, SEG11233168HX, SEG11233172HX, SEG11233174HX, SEG11233175HX, SEG11233176HX, SEG11233177HX, SEG11233178HX, SEG11233179HX, SEG11233180HX, SEG11233181HX, SEG11233182HX, SEG11233183HX, SEG11233184HX, SEG11233185HX, SEG11233186HX, SEG11233187HX, SEG11233188HX, SEG11233189HX, SEG11233190HX, SEG11233191HX, SEG11233192HX, SEG11233193HX, SEG11233194HX, SEG11233195HX, SEG11233196HX, SEG11233197HX, SEG11233198HX, SEG11233199HX, SEG11233200HX, SEG11233201HX, SEG11233202HX, SEG11233203HX, SEG11233204HX, SEG11233205HX, SEG11233206HX, SEG11233207HX, SEG11233208HX, SEG11233209HX, SEG11233210HX, SEG11233211HX, SEG11233212HX, SEG11233213HX, SEG11233214HX, SEG11233215HX, SEG11233216HX, SEG11233217HX, SEG11233218HX, SEG11233219HX, SEG11233220HX, SEG11233221HX, SEG11233222HX, SEG11233223HX, SEG11233224HX, SEG11233225HX, SEG11233226HX, SEG11233227HX, SEG11233228HX, SEG11233229HX, SEG11233230HX, SEG11233231HX, SEG11233232HX, SEG11233233HX, SEG11233234HX, SEG11233235HX, SEG11233236HX, SEG11233237HX, SEG11233238HX, SEG11233239HX, SEG11233240HX, SEG11233241HX, SEG11233242HX, SEG11233243HX, SEG11233244HX, SEG11233245HX, SEG11233246HX, SEG11253678HX, SEG11253679HX, SEG11253680HX, SEG11253681HX, SEG11253682HX, SEG11253683HX, SEG11253684HX, SEG11253685HX, SEG11253686HX, SEG11253687HX, SEG11253688HX, SEG11253689HX, SEG11253690HX, SEG11253691HX, SEG11253692HX, SEG11253693HX, SEG11253694HX, SEG11253695HX, SEG11253696HX, SEG11253697HX, SEG11253698HX, SEG11253699HX, SEG11253700HX, SEG11253701HX, SEG11253702HX, SEG11253703HX, SEG11253704HX, SEG11253705HX, SEG11253706HX, SEG11253707HX, SEG11253708HX, SEG11253709HX, SEG11253710HX, SEG11253711HX, SEG11253712HX, SEG11253713HX, SEG11253714HX, SEG11253715HX, SEG11253716HX, SEG11253717HX, SEG11253718HX, SEG11253719HX, SEG11253720HX, SEG11253721HX, SEG11253722HX, SEG11253723HX, SEG11253724HX, SEG11253725HX, SEG11253726HX, SEG11253727HX, SEG11253728HX, SEG11253729HX, SEG11253730HX, SEG11253731HX, SEG11253732HX, SEG11253733HX, SEG11253734HX, SEG11253735HX, SEG11253736HX, SEG11253737HX, SEG11253738HX, SEG11253739HX, SEG11253740HX, SEG11253741HX, SEG11253742HX, SEG11253743HX, SEG11253744HX, SEG11253745HX, SEG11253746HX, SEG11253747HX, SEG11253748HX, SEG11253749HX, SEG11253750HX, SEG11253751HX, SEG11253752HX, SEG11253753HX, SEG11253754HX, SEG11253755HX, SEG11253756HX, SEG11253757HX, SEG11253758HX, SEG11253759HX, SEG11253760HX, SEG11253761HX, SEG11253762HX, SEG11253763HX, SEG11253764HX, SEG11253765HX, SEG11253766HX, SEG11253767HX, SEG11253768HX, SEG11253769HX, SEG11253770HX, SEG11253771HX, SEG11253772HX, SEG11253773HX, SEG11253774HX, SEG11253775HX, SEG11253776HX, SEG11253777HX, SEG11263815HX, SEG11263816HX, SEG11263817HX, SEG11263818HX, SEG11263819HX, SEG11263820HX, SEG11263821HX, SEG11263822HX, SEG11263823HX, SEG11263824HX, SEG11263825HX, SEG11263826HX, SEG11263827HX, SEG11263828HX, SEG11263829HX, SEG11263830HX, SEG11263831HX, SEG11263832HX, SEG11263833HX, SEG11263834HX, SEG11263835HX, SEG11263836HX, SEG11263837HX, SEG11263838HX, SEG11263839HX, SEG11263840HX, SEG11263841HX, SEG11263842HX, SEG11263843HX, SEG11263844HX, SEG11263845HX, SEG11263846HX, SEG11263847HX, SEG11263848HX, SEG11263849HX, SEG11263850HX, SEG11263851HX, SEG11263852HX, SEG11263853HX, SEG11263854HX, SEG11263855HX, SEG11263856HX, SEG11263857HX, SEG11263858HX, SEG11263859HX, SEG11263860HX, SEG11263861HX, SEG11263862HX, SEG11263863HX, SEG11263864HX, SEG11314107HX, SEG11314108HX, SEG11314109HX, SEG11314110HX, SEG11314111HX, SEG11314112HX, SEG11314113HX, SEG11314114HX, SEG11314115HX, SEG11314116HX, SEG11314117HX, SEG11314118HX, SEG11314119HX, SEG11314120HX, SEG11314121HX, SEG11314122HX, SEG11314123HX, SEG11314124HX, SEG11314125HX, SEG11314126HX, SEG11314127HX, SEG11314128HX, SEG11314129HX, SEG11314130HX, SEG11314131HX, SEG11314132HX, SEG11314133HX, SEG11314134HX, SEG11314135HX, SEG11314136HX, SEG11314137HX, SEG11314138HX, SEG11314139HX, SEG11314140HX, SEG11314141HX, SEG11314142HX, SEG11314143HX, SEG11314144HX, SEG11314145HX, SEG11314146HX, SEG11314147HX, SEG11314148HX, SEG11314149HX, SEG11314150HX, SEG11314151HX, SEG11314152HX, SEG11314153HX, SEG11314154HX, SEG11314155HX, SEG11314156HX, SEG11324157HX, SEG11324158HX, SEG11324159HX, SEG11324160HX, SEG11324161HX, SEG11324162HX, SEG11324163HX, SEG11324164HX, SEG11324165HX, SEG11324166HX, SEG11324167HX, SEG11324168HX, SEG11324169HX, SEG11324170HX, SEG11324171HX, SEG11324172HX, SEG11324173HX, SEG11324174HX, SEG11324175HX, SEG11324176HX, SEG11324177HX, SEG11324178HX, SEG11324179HX, SEG11324180HX, SEG11324181HX, SEG11324182HX, SEG11324183HX, SEG11324184HX, SEG11324185HX, SEG11324186HX, SEG11324187HX, SEG11324188HX, SEG11324189HX, SEG11324190HX, SEG11324191HX, SEG11324192HX, SEG11324193HX, SEG11324194HX, SEG11324195HX, SEG11324196HX, SEG11324197HX, SEG11324198HX, SEG11324199HX, SEG11324200HX, SEG11324201HX, SEG11324202HX, SEG11324203HX, SEG11324204HX, SEG11324205HX, SEG11324206HX, SEG11324417HX, SEG11324418HX, SEG11324419HX, SEG11324420HX, SEG11324421HX, SEG11324422HX, SEG11324423HX, SEG11324424HX, SEG11324425HX, SEG11324426HX, SEG11324427HX, SEG11324428HX, SEG11324429HX, SEG11324430HX, SEG11324431HX, SEG11324432HX, SEG11324433HX, SEG11324434HX, SEG11324435HX, SEG11324436HX, SEG11324437HX, SEG11324438HX, SEG11324439HX, SEG11324440HX, SEG11324441HX, SEG11324442HX, SEG11324443HX, SEG11324444HX, SEG11324445HX, SEG11324446HX, SEG11324447HX, SEG11324448HX, SEG11324449HX, SEG11324450HX, SEG11324451HX, SEG11324452HX, SEG11324453HX, SEG11324454HX, SEG11324455HX, SEG11324456HX, SEG11324457HX, SEG11324458HX, SEG11324459HX, SEG11324460HX, SEG11324461HX, SEG11324462HX, SEG11324463HX, SEG11324464HX, SEG11324465HX, SEG11324466HX, SEG11334477HX, SEG11334478HX, SEG11334479HX, SEG11334480HX, SEG11334481HX, SEG11334482HX, SEG11334483HX, SEG11334484HX, SEG11334485HX, SEG11334486HX, SEG11334487HX, SEG11334488HX, SEG11334489HX, SEG11334491HX, SEG11334492HX, SEG11334493HX, SEG11334494HX, SEG11334495HX, SEG11334496HX, SEG11334497HX, SEG11334498HX, SEG11334499HX, SEG11334500HX, SEG11334501HX, SEG11334502HX, SEG11334503HX, SEG11334504HX, SEG11334505HX, SEG11334506HX, SEG11334507HX, SEG11334508HX, SEG11334509HX, SEG11334510HX, SEG11334511HX, SEG11334512HX, SEG11334513HX, SEG11334514HX, SEG11334515HX, SEG11334516HX, SEG11334517HX, SEG11334518HX, SEG11334519HX, SEG11334520HX, SEG11334521HX, SEG11334522HX, SEG11334523HX, SEG11334524HX, SEG11334525HX, SEG11334526HX, SEG11334757HX, SEG11334758HX, SEG11334759HX, SEG11334760HX, SEG11334761HX, SEG11334762HX, SEG11334763HX, SEG11334764HX, SEG11334765HX, SEG11334766HX, SEG11334767HX, SEG11334768HX, SEG11334769HX, SEG11334770HX, SEG11334771HX, SEG11334772HX, SEG11334773HX, SEG11334774HX, SEG11334775HX, SEG11334776HX, SEG11334777HX, SEG11334778HX, SEG11334779HX, SEG11334780HX, SEG11334781HX, SEG11334782HX, SEG11334783HX, SEG11334784HX, SEG11334785HX, SEG11334786HX, SEG11334787HX, SEG11334788HX, SEG11334789HX, SEG11334790HX, SEG11334791HX, SEG11334792HX, SEG11334793HX, SEG11334794HX, SEG11334795HX, SEG11334796HX, SEG11334797HX, SEG11334798HX, SEG11334799HX, SEG11334800HX, SEG11334801HX, SEG11334802HX, SEG11334803HX, SEG11334804HX, SEG11334805HX, SEG11334806HX, SEG11334807HX, SEG11334808HX, SEG11334809HX, SEG11334810HX, SEG11334811HX, SEG11334812HX, SEG11334813HX, SEG11334814HX, SEG11334815HX, SEG11334816HX, SEG11334817HX, SEG11334818HX, SEG11334819HX, SEG11334820HX, SEG11334821HX, SEG11334822HX, SEG11334823HX, SEG11334824HX, SEG11334825HX, SEG11334826HX, SEG11344837HX, SEG11344838HX, SEG11344839HX, SEG11344840HX, SEG11344841HX, SEG11344842HX, SEG11344843HX, SEG11344844HX, SEG11344845HX, SEG11344846HX, SEG11344847HX, SEG11344848HX, SEG11344849HX, SEG11344850HX, SEG11344851HX, SEG11344852HX, SEG11344853HX, SEG11344854HX, SEG11344855HX, SEG11344856HX, SEG11344857HX, SEG11344858HX, SEG11344859HX, SEG11344860HX, SEG11344861HX, SEG11344862HX, SEG11344863HX, SEG11344864HX, SEG11344865HX, SEG11405635HX, SEG11405636HX, SEG11405637HX, SEG11405638HX, SEG11405639HX, SEG11405640HX, SEG11405641HX, SEG11405642HX, SEG11405643HX, SEG11405644HX, SEG11405645HX, SEG11405646HX, SEG11405647HX, SEG11405648HX, SEG11405649HX, SEG11405650HX, SEG11405651HX, SEG11405652HX, SEG11405653HX, SEG11405654HX, SEG11405655HX, SEG11405656HX, SEG11405657HX, SEG11405658HX, SEG11405659HX, SEG11405660HX, SEG11405661HX, SEG11405662HX, SEG11405663HX, SEG11405664HX, SEG11405665HX, SEG11405666HX, SEG11405667HX, SEG11405668HX, SEG11405669HX, SEG11405670HX, SEG11405671HX, SEG11405672HX, SEG11405673HX, SEG11405674HX, SEG11405675HX, SEG11405676HX, SEG11405677HX, SEG11405678HX, SEG11405679HX, SEG11405680HX, SEG11405681HX, SEG11405682HX, SEG11405683HX, SEG11405684HX, SEG11405685HX, SEG11405686HX, SEG11405687HX, SEG11405688HX, SEG11405689HX, SEG11405690HX, SEG11405691HX, SEG11405692HX, SEG11405693HX, SEG11405694HX, SEG11405695HX, SEG11405696HX, SEG11405697HX, SEG11405698HX, SEG11405699HX, SEG11405700HX, SEG11405701HX, SEG11405702HX, SEG11405703HX, SEG11405704HX, SEG11405705HX, SEG11405706HX, SEG11405707HX, SEG11405708HX, SEG11405709HX, SEG11405710HX, SEG11405711HX, SEG11405712HX, SEG11405713HX, SEG11405714HX, SEG11405715HX, SEG11405716HX, SEG11405717HX, SEG11405718HX, SEG11405719HX, SEG11405720HX, SEG11405721HX, SEG11405722HX, SEG11405723HX, SEG11405724HX, SEG11405725HX, SEG11405726HX, SEG11405727HX, SEG11405728HX, SEG11405729HX, SEG11405730HX, SEG11405731HX, SEG11405732HX, SEG11405733HX, SEG11405734HX, SEG11405735HX, SEG11405736HX, SEG11405737HX, SEG11405738HX, SEG11405739HX, SEG11405740HX, SEG11405741HX, SEG11405742HX, SEG11405743HX, SEG11405744HX, SEG11405745HX, SEG11405746HX, SEG11405747HX, SEG11405748HX, SEG11405749HX, SEG11405750HX, SEG11405751HX, SEG11446379HX, SEG11446380HX, SEG11446381HX, SEG11446382HX, SEG11446383HX, SEG11446384HX, SEG11446385HX, SEG11446386HX, SEG11446387HX, SEG11446388HX, SEG11446389HX, SEG11446390HX, SEG11446391HX, SEG11446392HX, SEG11446393HX, SEG11446394HX, SEG11446395HX, SEG11446396HX, SEG11446397HX, SEG11446398HX, SEG11446399HX, SEG11446400HX, SEG11446401HX, SEG11446402HX, SEG11446403HX, SEG11446404HX, SEG11446405HX, SEG11446406HX, SEG11446407HX, SEG11446408HX, SEG11446409HX, SEG11446410HX, SEG11446411HX, SEG11446412HX, SEG11446413HX, SEG11446414HX, SEG11446415HX, SEG11446416HX, SEG11446417HX, SEG11446418HX, SEG11446419HX, SEG11446420HX, SEG11446421HX, SEG11446422HX, SEG11446423HX, SEG11446424HX, SEG11446425HX, SEG11446426HX, SEG11446427HX, SEG11446428HX, SEG11456550HX, SEG11456551HX, SEG11456552HX, SEG11456553HX, SEG11456554HX, SEG11456555HX, SEG11456560HX, SEG11456561HX, SEG11456562HX, SEG11456563HX, SEG11456564HX, SEG11456565HX, SEG11456567HX, SEG11456568HX, SEG11456569HX, SEG11456575HX, SEG11456576HX, SEG11456577HX, SEG11456578HX, SEG11456579HX, SEG11456580HX, SEG11456581HX, SEG11456582HX, SEG11456583HX, SEG11456584HX, SEG11456585HX, SEG11456586HX, SEG11456587HX, SEG11456588HX, SEG11456589HX, SEG11456593HX, SEG11456594HX, SEG11456595HX, SEG11456596HX, SEG11456597HX, SEG11456598HX, SEG11456599HX, SEG11456600HX, SEG11456601HX, SEG11456602HX, SEG11456603HX, SEG11456604HX, SEG11456605HX, SEG11456606HX, SEG11456607HX, SEG11456608HX, SEG11456609HX, SEG11456610HX, SEG11456611HX, SEG11456612HX, SEG11456613HX, SEG11456614HX, SEG11456615HX, SEG11456616HX, SEG11456617HX, SEG11456618HX, SEG11456619HX, SEG11456620HX, SEG11456621HX, SEG11456622HX, SEG11456623HX, SEG11456624HX, SEG11456625HX, SEG11456626HX, SEG11456627HX, SEG11456628HX, SEG11456629HX, SEG11456630HX, SEG11456631HX, SEG11456632HX, SEG11456633HX, SEG11456634HX, SEG11456635HX, SEG11456636HX, SEG11456637HX, SEG11456638HX, SEG11456639HX, SEG11456640HX, SEG11456641HX, SEG11456642HX, SEG11456643HX, SEG11456644HX, SEG11456645HX, SEG11456646HX, SEG11456647HX, SEG11456648HX, SEG11456649HX, SEG11456650HX, SEG11456651HX, SEG11456652HX, SEG11456653HX, SEG11456654HX, SEG11456655HX, SEG11456656HX, SEG11456657HX, SEG11456658HX, SEG11456659HX, SEG11456660HX, SEG11456661HX, SEG11456662HX, SEG11456663HX, SEG11456664HX, SEG11456665HX, SEG11456666HX, SEG11456667HX, SEG11456668HX, SEG11456669HX, SEG11456670HX, SEG11456671HX, SEG11456672HX, SEG11456673HX, SEG11456674HX, SEG11456675HX, SEG11456676HX, SEG11456677HX, SEG11456678HX, SEG11456679HX, SEG11466851HX, SEG11466852HX, SEG11466853HX, SEG11466854HX, SEG11466855HX, SEG11466856HX, SEG11466857HX, SEG11466858HX, SEG11466859HX, SEG11466860HX, SEG11466861HX, SEG11466862HX, SEG11466863HX, SEG11466864HX, SEG11466865HX, SEG11466866HX, SEG11466867HX, SEG11466868HX, SEG11466869HX, SEG11466870HX, SEG11466871HX, SEG11466872HX, SEG11466873HX, SEG11466874HX, SEG11466875HX, SEG11466876HX, SEG11466877HX, SEG11466878HX, SEG11466879HX, SEG11466880HX, SEG11466881HX, SEG11466882HX, SEG11466883HX, SEG11466884HX, SEG11466885HX, SEG11466886HX, SEG11466887HX, SEG11466888HX, SEG11466889HX, SEG11466890HX, SEG11466891HX, SEG11466892HX, SEG11466893HX, SEG11466894HX, SEG11466895HX, SEG11466896HX, SEG11466897HX, SEG11466898HX, SEG11466899HX, SEG11466900HX, SEG11466901HX, SEG11466902HX, SEG11466903HX, SEG11466904HX, SEG11466905HX, SEG11466906HX, SEG11466907HX, SEG11466908HX, SEG11466909HX, SEG11466910HX, SEG11466911HX, SEG11466912HX, SEG11466913HX, SEG11466914HX, SEG11466915HX, SEG11466916HX, SEG11466917HX, SEG11466918HX, SEG11466919HX, SEG11466920HX, SEG11466921HX, SEG11466922HX, SEG11466923HX, SEG11466924HX, SEG11466925HX, SEG11466926HX, SEG11466927HX, SEG11466928HX, SEG11466929HX, SEG11466930HX, SEG11487136HX, SEG11487137HX, SEG11487138HX, SEG11487139HX, SEG11487140HX, SEG11487141HX, SEG11487142HX, SEG11487143HX, SEG11487144HX, SEG11487145HX, SEG11487146HX, SEG11487147HX, SEG11487148HX, SEG11487149HX, SEG11487150HX, SEG11487151HX, SEG11487152HX, SEG11487153HX, SEG11487154HX, SEG11487155HX, SEG11487156HX, SEG11487157HX, SEG11487158HX, SEG11487159HX, SEG11487160HX, SEG11487161HX, SEG11487162HX, SEG11487163HX, SEG11487164HX, SEG11487165HX, SEG11487166HX, SEG11487167HX, SEG11487168HX, SEG11487169HX, SEG11487170HX, SEG11487171HX, SEG11487172HX, SEG11487173HX, SEG11487174HX, SEG11487175HX, SEG11487176HX, SEG11487177HX, SEG11487178HX, SEG11487179HX, SEG11487180HX, SEG11487181HX, SEG11487182HX, SEG11487183HX, SEG11487184HX, SEG11487185HX, SEG11487186HX, SEG11487187HX, SEG11487188HX, SEG11487189HX, SEG11487190HX, SEG11487191HX, SEG11487192HX, SEG11487193HX, SEG11487194HX, SEG11487195HX, SEG11487196HX, SEG11487197HX, SEG11487198HX, SEG11487199HX, SEG11487200HX, SEG11487201HX, SEG11487202HX, SEG11487203HX, SEG11487204HX, SEG11487205HX, SEG11487206HX, SEG11487207HX, SEG11487208HX, SEG11487209HX, SEG11487210HX, SEG11487211HX, SEG11487212HX, SEG11487213HX, SEG11487214HX, SEG11487215HX, SEG11487216HX, SEG11487217HX, SEG11487218HX, SEG11487219HX, SEG11487220HX, SEG11487221HX, SEG11487222HX, SEG11487223HX, SEG11487224HX, SEG11487225HX, SEG11487226HX, SEG11487227HX, SEG11487228HX, SEG11487229HX, SEG11487230HX, SEG11487231HX, SEG11487232HX, SEG11487233HX, SEG11487234HX, SEG11487235HX, SEG11487352HX, SEG11487353HX, SEG11487354HX, SEG11487355HX, SEG11487356HX, SEG11487357HX, SEG11487358HX, SEG11487359HX, SEG11487360HX, SEG11487361HX, SEG11487362HX, SEG11487363HX, SEG11487364HX, SEG11487365HX, SEG11487366HX, SEG11487367HX, SEG11487368HX, SEG11487369HX, SEG11487370HX, SEG11487371HX, SEG11487372HX, SEG11487373HX, SEG11487374HX, SEG11487375HX, SEG11487376HX, SEG11487377HX, SEG11487378HX, SEG11487379HX, SEG11487380HX, SEG11487381HX, SEG11487382HX, SEG11487383HX, SEG11487384HX, SEG11487385HX, SEG11487386HX, SEG11487387HX, SEG11487388HX, SEG11487389HX, SEG11487390HX, SEG11487391HX, SEG11487392HX, SEG11487393HX, SEG11487394HX, SEG11487395HX, SEG11487396HX, SEG11487397HX, SEG11487398HX, SEG11487399HX, SEG11487400HX, SEG11487401HX, SEG11487402HX, SEG11487403HX, SEG11487404HX, SEG11487405HX, SEG11487406HX, SEG11487407HX, SEG11487408HX, SEG11487409HX, SEG11487410HX, SEG11487411HX, SEG11487412HX, SEG11487413HX, SEG11487414HX, SEG11487415HX, SEG11487416HX, SEG11487417HX, SEG11487418HX, SEG11487419HX, SEG11487420HX, SEG11487421HX, SEG11487422HX, SEG11487423HX, SEG11487424HX, SEG11487425HX, SEG11487426HX, SEG11487427HX, SEG11487428HX, SEG11487429HX, SEG11487430HX, SEG11487431HX, SEG11487432HX, SEG11487433HX, SEG11487434HX, SEG11487435HX, SEG11487436HX, SEG11487437HX, SEG11487438HX, SEG11487439HX, SEG11487440HX, SEG11487441HX, SEG11487442HX, SEG11487443HX, SEG11487444HX, SEG11487445HX, SEG11487446HX, SEG11487447HX, SEG11487448HX, SEG11487449HX, SEG11487450HX, SEG11487451HX, SEG11507613HX, SEG11507614HX, SEG11507615HX, SEG11507616HX, SEG11507617HX, SEG11507618HX, SEG11507619HX, SEG11507620HX, SEG11507621HX, SEG11507622HX, SEG11507623HX, SEG11507624HX, SEG11507625HX, SEG11507626HX, SEG11507627HX, SEG11507628HX, SEG11507629HX, SEG11507630HX, SEG11507631HX, SEG11507632HX, SEG11507633HX, SEG11507634HX, SEG11507635HX, SEG11507636HX, SEG11507637HX, SEG11507638HX, SEG11507639HX, SEG11507640HX, SEG11507641HX, SEG11507642HX, SEG11507643HX, SEG11507644HX, SEG11507645HX, SEG11507646HX, SEG11507647HX, SEG11507648HX, SEG11507649HX, SEG11507650HX, SEG11507651HX, SEG11507652HX, SEG11507653HX, SEG11507654HX, SEG11507655HX, SEG11507656HX, SEG11507657HX, SEG11507658HX, SEG11507659HX, SEG11507660HX, SEG11507661HX, SEG11507662HX, SEG11507663HX, SEG11507664HX, SEG11507665HX, SEG11507666HX, SEG11507667HX, SEG11507668HX, SEG11507669HX, SEG11507670HX, SEG11507671HX, SEG11507672HX, SEG11507673HX, SEG11507674HX, SEG11507675HX, SEG11507676HX, SEG11507677HX, SEG11507678HX, SEG11507679HX, SEG11507680HX, SEG11507681HX, SEG11507682HX, SEG11507683HX, SEG11507684HX, SEG11507685HX, SEG11507686HX, SEG11507687HX, SEG12018188HX, SEG12018189HX, SEG12018190HX, SEG12018191HX, SEG12018192HX, SEG12018193HX, SEG12018194HX, SEG12018195HX, SEG12018196HX, SEG12018197HX, SEG12018198HX, SEG12018199HX, SEG12018200HX, SEG12018201HX, SEG12018202HX, SEG12018203HX, SEG12018204HX, SEG12018205HX, SEG12018206HX, SEG12018207HX, SEG12018208HX, SEG12018209HX, SEG12018210HX, SEG12018211HX, SEG12018212HX, SEG12018213HX, SEG12018214HX, SEG12018215HX, SEG12018216HX, SEG12018217HX, SEG12018218HX, SEG12018219HX, SEG12018220HX, SEG12018221HX, SEG12018222HX, SEG12018223HX, SEG12018224HX, SEG12018225HX, SEG12018226HX, SEG12018227HX, SEG12018228HX, SEG12018229HX, SEG12018230HX, SEG12018231HX, SEG12018232HX, SEG12018233HX, SEG12018234HX, SEG12018235HX, SEG12018236HX, SEG12018237HX, SEG12018238HX, SEG12018239HX, SEG12018240HX, SEG12018241HX, SEG12018242HX, SEG12018243HX, SEG12018244HX, SEG12018245HX, SEG12018246HX, SEG12018247HX, SEG12018248HX, SEG12018249HX, SEG12018250HX, SEG12018251HX, SEG12018252HX, SEG12018253HX, SEG12018254HX, SEG12018255HX, SEG12018256HX, SEG12018257HX, SEG12018258HX, SEG12018259HX, SEG12018260HX, SEG12018261HX, SEG12018262HX, SEG12018263HX, SEG12018264HX, SEG12018265HX, SEG12018266HX, SEG12018267HX, SEG12018268HX, SEG12018269HX, SEG12018270HX, SEG12018271HX, SEG12018272HX, SEG12018273HX, SEG12018274HX, SEG12018275HX, SEG12018276HX, SEG12018277HX, SEG12018278HX, SEG12018279HX, SEG12018280HX, SEG12018281HX, SEG12018282HX, SEG12018283HX, SEG12018284HX, SEG12018285HX, SEG12018286HX, SEG12018287HX, SEG12028489HX, SEG12028490HX, SEG12028491HX, SEG12028492HX, SEG12028493HX, SEG12028494HX, SEG12028495HX, SEG12028496HX, SEG12028497HX, SEG12028498HX, SEG12028499HX, SEG12028500HX, SEG12028501HX, SEG12028502HX, SEG12028503HX, SEG12028504HX, SEG12028505HX, SEG12028506HX, SEG12028507HX, SEG12028508HX, SEG12028509HX, SEG12028510HX, SEG12028511HX, SEG12028512HX, SEG12028513HX, SEG12028514HX, SEG12028515HX, SEG12028516HX, SEG12028517HX, SEG12028518HX, SEG12028519HX, SEG12028520HX, SEG12028521HX, SEG12028522HX, SEG12028523HX, SEG12028524HX, SEG12028525HX, SEG12028526HX, SEG12028527HX, SEG12028528HX, SEG12028529HX, SEG12028530HX, SEG12028531HX, SEG12028532HX, SEG12028533HX, SEG12028534HX, SEG12028535HX, SEG12028536HX, SEG12028537HX, SEG12028538HX, SEG12028589HX, SEG12028590HX, SEG12028591HX, SEG12028592HX, SEG12028593HX, SEG12028594HX, SEG12028595HX, SEG12028596HX, SEG12028597HX, SEG12028598HX, SEG12028599HX, SEG12028600HX, SEG12028601HX, SEG12028602HX, SEG12028603HX, SEG12028604HX, SEG12028605HX, SEG12028606HX, SEG12028607HX, SEG12028608HX, SEG12028609HX, SEG12028610HX, SEG12028611HX, SEG12028612HX, SEG12028613HX, SEG12028614HX, SEG12028615HX, SEG12028616HX, SEG12028617HX, SEG12028618HX, SEG12028619HX, SEG12028620HX, SEG12028621HX, SEG12028622HX, SEG12028623HX, SEG12028624HX, SEG12028625HX, SEG12028626HX, SEG12028627HX, SEG12028628HX, SEG12028629HX, SEG12028630HX, SEG12028631HX, SEG12028632HX, SEG12028633HX, SEG12028634HX, SEG12028635HX, SEG12028636HX, SEG12028637HX, SEG12028638HX, SEG12049064HX, SEG12049065HX, SEG12049066HX, SEG12049067HX, SEG12049068HX, SEG12049069HX, SEG12049070HX, SEG12049071HX, SEG12049072HX, SEG12049073HX, SEG12049074HX, SEG12049075HX, SEG12049076HX, SEG12049077HX, SEG12049078HX, SEG12049079HX, SEG12049080HX, SEG12049081HX, SEG12049082HX, SEG12049083HX, SEG12049084HX, SEG12049085HX, SEG12049086HX, SEG12049087HX, SEG12049088HX, SEG12049089HX, SEG12049090HX, SEG12049091HX, SEG12049092HX, SEG12049093HX, SEG12049094HX, SEG12049095HX, SEG12049096HX, SEG12049097HX, SEG12049098HX, SEG12049099HX, SEG12049100HX, SEG12049101HX, SEG12049102HX, SEG12049103HX, SEG12049104HX, SEG12049105HX, SEG12049106HX, SEG12049107HX, SEG12049108HX, SEG12049109HX, SEG12049110HX, SEG12049111HX, SEG12049112HX, SEG12049113HX, SEG12049114HX, SEG12049115HX, SEG12049116HX, SEG12049117HX, SEG12049118HX, SEG12049119HX, SEG12049120HX, SEG12049121HX, SEG12049122HX, SEG12049123HX, SEG12049124HX, SEG12049125HX, SEG12049126HX, SEG12049127HX, SEG12049128HX, SEG12049129HX, SEG12049130HX, SEG12049131HX, SEG12049132HX, SEG12049133HX, SEG12049134HX, SEG12049135HX, SEG12049136HX, SEG12049137HX, SEG12049138HX, SEG12049139HX, SEG12049140HX, SEG12049141HX, SEG12049142HX, SEG12049143HX, SEG12049144HX, SEG12049145HX, SEG12049146HX, SEG12049147HX, SEG12049148HX, SEG12049149HX, SEG12049150HX, SEG12049151HX, SEG12049152HX, SEG12049153HX, SEG12049154HX, SEG12049155HX, SEG12049156HX, SEG12049157HX, SEG12049158HX, SEG12049159HX, SEG12049160HX, SEG12049161HX, SEG12049162HX, SEG12049163HX, SEG12049164HX, SEG12049165HX, SEG12049166HX, SEG12049167HX, SEG12049168HX, SEG12049169HX, SEG12049170HX, SEG12049171HX, SEG12049172HX, SEG12049173HX, SEG12049174HX, SEG12049175HX, SEG12049176HX, SEG12049177HX, SEG12049178HX, SEG12049179HX, SEG12049180HX, SEG12049181HX, SEG12049182HX, SEG12049183HX, SEG12049184HX, SEG12049185HX, SEG12049186HX, SEG12049187HX, SEG12049188HX, SEG12049189HX, SEG12049190HX, SEG12049191HX, SEG12049192HX, SEG12049193HX, SEG12049194HX, SEG12049195HX, SEG12049196HX, SEG12049197HX, SEG12049198HX, SEG12049199HX, SEG12049200HX, SEG12049201HX, SEG12049202HX, SEG12049203HX, SEG12049204HX, SEG12049205HX, SEG12049206HX, SEG12049207HX, SEG12049208HX, SEG12049209HX, SEG12049210HX, SEG12049211HX, SEG12049212HX, SEG12049213HX, SEG12069654HX, SEG12069655HX, SEG12069656HX, SEG12069657HX, SEG12069658HX, SEG12069659HX, SEG12069660HX, SEG12069661HX, SEG12069662HX, SEG12069663HX, SEG12069664HX, SEG12069665HX, SEG12069666HX, SEG12069667HX, SEG12069668HX, SEG12069669HX, SEG12069670HX, SEG12069671HX, SEG12069672HX, SEG12069673HX, SEG12069674HX, SEG12069675HX, SEG12069676HX, SEG12069677HX, SEG12069678HX, SEG12069679HX, SEG12069680HX, SEG12069681HX, SEG12069682HX, SEG12069683HX, SEG12069685HX, SEG12069686HX, SEG12069687HX, SEG12069688HX, SEG12069689HX, SEG12069690HX, SEG12069691HX, SEG12069692HX, SEG12069693HX, SEG12069694HX, SEG12069695HX, SEG12069696HX, SEG12069697HX, SEG12069698HX, SEG12069699HX, SEG12069700HX, SEG12069701HX, SEG12069702HX, SEG12069703HX, SEG12069755HX, SEG12069756HX, SEG12069757HX, SEG12069758HX, SEG12069759HX, SEG12069760HX, SEG12069761HX, SEG12069762HX, SEG12069763HX, SEG12069764HX, SEG12069765HX, SEG12069766HX, SEG12069767HX, SEG12069768HX, SEG12069769HX, SEG12069770HX, SEG12069771HX, SEG12069772HX, SEG12069773HX, SEG12069774HX, SEG12069775HX, SEG12069776HX, SEG12069777HX, SEG12069778HX, SEG12069779HX, SEG12069780HX, SEG12069781HX, SEG12069782HX, SEG12069783HX, SEG12069784HX, SEG12069785HX, SEG12069786HX, SEG12069787HX, SEG12069788HX, SEG12069789HX, SEG12069790HX, SEG12069791HX, SEG12069792HX, SEG12069793HX, SEG12069794HX, SEG12069795HX, SEG12069796HX, SEG12069797HX, SEG12069798HX, SEG12069799HX, SEG12069800HX, SEG12069801HX, SEG12069802HX, SEG12069803HX, SEG12069804HX, SEG12070000HX, SEG12070001HX, SEG12070002HX, SEG12070003HX, SEG12070004HX, SEG12070005HX, SEG12070006HX, SEG12070007HX, SEG12070008HX, SEG12070009HX, SEG12070010HX, SEG12070011HX, SEG12070012HX, SEG12070013HX, SEG12070014HX, SEG12070015HX, SEG12070016HX, SEG12070017HX, SEG12070018HX, SEG12070019HX, SEG12070020HX, SEG12070021HX, SEG12070022HX, SEG12070023HX, SEG12070024HX, SEG12070025HX, SEG12070026HX, SEG12070027HX, SEG12070028HX, SEG12070029HX, SEG12070030HX, SEG12070031HX, SEG12070032HX, SEG12070033HX, SEG12070034HX, SEG12070035HX, SEG12070036HX, SEG12070037HX, SEG12070038HX, SEG12070039HX, SEG12070040HX, SEG12070041HX, SEG12070042HX, SEG12070043HX, SEG12070044HX, SEG12070045HX, SEG12070046HX, SEG12079947HX, SEG12079948HX, SEG12079949HX, SEG12079950HX, SEG12079951HX, SEG12079952HX, SEG12079953HX, SEG12079954HX, SEG12079955HX, SEG12079956HX, SEG12079957HX, SEG12079958HX, SEG12079959HX, SEG12079960HX, SEG12079961HX, SEG12079962HX, SEG12079963HX, SEG12079964HX, SEG12079965HX, SEG12079966HX, SEG12079967HX, SEG12079968HX, SEG12079969HX, SEG12079970HX, SEG12079971HX, SEG12079972HX, SEG12079973HX, SEG12079974HX, SEG12079975HX, SEG12079976HX, SEG12079977HX, SEG12079978HX, SEG12079979HX, SEG12079980HX, SEG12079981HX, SEG12079982HX, SEG12079983HX, SEG12079984HX, SEG12079985HX, SEG12079986HX, SEG12079987HX, SEG12079988HX, SEG12079989HX, SEG12079990HX, SEG12079991HX, SEG12079992HX, SEG12079993HX, SEG12079994HX, SEG12079995HX, SEG12079996HX, SEG12079997HX, SEG12079998HX, SEG12079999HX, SEG12080122HX, SEG12080123HX, SEG12080124HX, SEG12080125HX, SEG12080126HX, SEG12080127HX, SEG12080128HX, SEG12080129HX, SEG12080130HX, SEG12080131HX, SEG12080132HX, SEG12080133HX, SEG12080134HX, SEG12080135HX, SEG12080136HX, SEG12080137HX, SEG12080138HX, SEG12080139HX, SEG12080140HX, SEG12080141HX, SEG12080142HX, SEG12080143HX, SEG12080144HX, SEG12080145HX, SEG12080146HX, SEG12080147HX, SEG12080148HX, SEG12080149HX, SEG12080150HX, SEG12080151HX, SEG12080152HX, SEG12080153HX, SEG12080154HX, SEG12080155HX, SEG12080156HX, SEG12080157HX, SEG12080158HX, SEG12080159HX, SEG12080160HX, SEG12080161HX, SEG12080162HX, SEG12080163HX, SEG12080164HX, SEG12080165HX, SEG12080166HX, SEG12080167HX, SEG12080168HX, SEG12080169HX, SEG12080170HX, SEG12080171HX, SEG12080341HX, SEG12080342HX, SEG12080343HX, SEG12080344HX, SEG12080345HX, SEG12080346HX, SEG12080347HX, SEG12080348HX, SEG12080349HX, SEG12080350HX, SEG12080351HX, SEG12080352HX, SEG12080353HX, SEG12080354HX, SEG12080355HX, SEG12080356HX, SEG12080357HX, SEG12080358HX, SEG12080359HX, SEG12080360HX, SEG12090361HX, SEG12090362HX, SEG12090363HX, SEG12090364HX, SEG12090365HX, SEG12090366HX, SEG12090367HX, SEG12090368HX, SEG12090369HX, SEG12090370HX, SEG12090371HX, SEG12090372HX, SEG12090373HX, SEG12090374HX, SEG12090375HX, SEG12090376HX, SEG12090377HX, SEG12090378HX, SEG12090379HX, SEG12090380HX, SEG12090391HX, SEG12090392HX, SEG12090393HX, SEG12090394HX, SEG12090395HX, SEG12090396HX, SEG12090397HX, SEG12090398HX, SEG12090399HX, SEG12090400HX, SEG12090401HX, SEG12090402HX, SEG12090403HX, SEG12090404HX, SEG12090405HX, SEG12090406HX, SEG12090407HX, SEG12090408HX, SEG12090409HX, SEG12090410HX, SEG12090411HX, SEG12090412HX, SEG12090413HX, SEG12090414HX, SEG12090415HX, SEG12090416HX, SEG12090417HX, SEG12090418HX, SEG12090419HX, SEG12090420HX, SEG12090421HX, SEG12090422HX, SEG12090423HX, SEG12090424HX, SEG12090425HX, SEG12090426HX, SEG12090427HX, SEG12090428HX, SEG12090429HX, SEG12090430HX, SEG12090431HX, SEG12090432HX, SEG12090433HX, SEG12090434HX, SEG12090435HX, SEG12090436HX, SEG12090437HX, SEG12090438HX, SEG12090439HX, SEG12090440HX, SEG12090496HX, SEG12090497HX, SEG12090498HX, SEG12090499HX, SEG12090500HX, SEG12090501HX, SEG12090502HX, SEG12090503HX, SEG12090504HX, SEG12090505HX, SEG12090506HX, SEG12090507HX, SEG12090508HX, SEG12090509HX, SEG12090510HX, SEG12090511HX, SEG12090512HX, SEG12090513HX, SEG12090514HX, SEG12090515HX, SEG12090516HX, SEG12090517HX, SEG12090518HX, SEG12090519HX, SEG12090520HX, SEG12090521HX, SEG12090522HX, SEG12090523HX, SEG12090524HX, SEG12090525HX, SEG12090526HX, SEG12090527HX, SEG12090528HX, SEG12090529HX, SEG12090530HX, SEG12090531HX, SEG12090532HX, SEG12090533HX, SEG12090534HX, SEG12090535HX, SEG12090537HX, SEG12090539HX, SEG12090540HX, SEG12090541HX, SEG12090542HX, SEG12090543HX, SEG12090544HX, SEG12090545HX, SEG12090546HX, SEG12090547HX, SEG12090548HX, SEG12090549HX, SEG12090550HX, SEG12090551HX, SEG12090552HX, SEG12090553HX, SEG12090554HX, SEG12090555HX, SEG12090556HX, SEG12090557HX, SEG12090558HX, SEG12090559HX, SEG12090560HX, SEG12090561HX, SEG12090562HX, SEG12090563HX, SEG12090564HX, SEG12090565HX, SEG12090566HX, SEG12090567HX, SEG12090568HX, SEG12090569HX, SEG12090570HX, SEG12090571HX, SEG12090572HX, SEG12090573HX, SEG12090574HX, SEG12090575HX, SEG12090576HX, SEG12090577HX, SEG12090578HX, SEG12090579HX, SEG12090580HX, SEG12090581HX, SEG12090582HX, SEG12090583HX, SEG12090584HX, SEG12090585HX, SEG12090586HX, SEG12090587HX, SEG12090588HX, SEG12090589HX, SEG12090590HX, SEG12090591HX, SEG12090592HX, SEG12090593HX, SEG12090594HX, SEG12090595HX, SEG12100876HX, SEG12100877HX, SEG12100878HX, SEG12100879HX, SEG12100880HX, SEG12100881HX, SEG12100882HX, SEG12100883HX, SEG12100884HX, SEG12100885HX, SEG12100886HX, SEG12100887HX, SEG12100888HX, SEG12100889HX, SEG12100890HX, SEG12100891HX, SEG12100892HX, SEG12100893HX, SEG12100894HX, SEG12100895HX, SEG12100896HX, SEG12100897HX, SEG12100898HX, SEG12100899HX, SEG12100900HX, SEG12100901HX, SEG12100902HX, SEG12100903HX, SEG12100904HX, SEG12100905HX, SEG12100906HX, SEG12100907HX, SEG12100908HX, SEG12100909HX, SEG12100910HX, SEG12100911HX, SEG12100912HX, SEG12100913HX, SEG12100914HX, SEG12100915HX, SEG12100916HX, SEG12100917HX, SEG12100918HX, SEG12100919HX, SEG12100920HX, SEG12100921HX, SEG12100922HX, SEG12100923HX, SEG12100924HX, SEG12100925HX, SEG12100926HX, SEG12100927HX, SEG12100928HX, SEG12100929HX, SEG12100930HX, SEG12100931HX, SEG12100932HX, SEG12100933HX, SEG12100934HX, SEG12100935HX, SEG12100936HX, SEG12100937HX, SEG12100938HX, SEG12100939HX, SEG12100940HX, SEG12100941HX, SEG12100942HX, SEG12100943HX, SEG12100944HX, SEG12100945HX, SEG12111196HX, SEG12111197HX, SEG12111198HX, SEG12111199HX, SEG12111200HX, SEG12111201HX, SEG12111202HX, SEG12111203HX, SEG12111204HX, SEG12111205HX, SEG12111206HX, SEG12111207HX, SEG12111208HX, SEG12111209HX, SEG12111210HX, SEG12111211HX, SEG12111212HX, SEG12111213HX, SEG12111214HX, SEG12111215HX, SEG12111216HX, SEG12111217HX, SEG12111218HX, SEG12111219HX, SEG12111220HX, SEG12111221HX, SEG12111222HX, SEG12111223HX, SEG12111224HX, SEG12111225HX, SEG12111226HX, SEG12111227HX, SEG12111228HX, SEG12111229HX, SEG12111230HX, SEG12111231HX, SEG12111232HX, SEG12111233HX, SEG12111234HX, SEG12111235HX, SEG12111236HX, SEG12111237HX, SEG12111238HX, SEG12111239HX, SEG12111240HX, SEG12111241HX, SEG12111242HX, SEG12111243HX, SEG12111244HX, SEG12111245HX, SEG12111246HX, SEG12111247HX, SEG12111248HX, SEG12111249HX, SEG12111250HX, SEG12111251HX, SEG12111252HX, SEG12111253HX, SEG12111254HX, SEG12111255HX, SEG12111256HX, SEG12111257HX, SEG12111258HX, SEG12111259HX, SEG12111260HX, SEG12111261HX, SEG12111262HX, SEG12111263HX, SEG12111264HX, SEG12111265HX, SEG12111266HX, SEG12111267HX, SEG12111268HX, SEG12111269HX, SEG12111270HX, SEG12111271HX, SEG12111272HX, SEG12111273HX, SEG12111274HX, SEG12111275HX, SEG12111276HX, SEG12111277HX, SEG12111278HX, SEG12111279HX, SEG12111280HX, SEG12111281HX, SEG12111282HX, SEG12111283HX, SEG12111284HX, SEG12111285HX, SEG12111286HX, SEG12111287HX, SEG12111288HX, SEG12111289HX, SEG12111295HX, SEG12141914HX, SEG12141915HX, SEG12141916HX, SEG12141917HX, SEG12141918HX, SEG12141919HX, SEG12141920HX, SEG12141921HX, SEG12141922HX, SEG12141923HX, SEG12141924HX, SEG12141925HX, SEG12141926HX, SEG12141927HX, SEG12141928HX, SEG12141929HX, SEG12141930HX, SEG12141931HX, SEG12141932HX, SEG12141933HX, SEG12141934HX, SEG12141935HX, SEG12141936HX, SEG12141937HX, SEG12141938HX, SEG12141939HX, SEG12141940HX, SEG12141941HX, SEG12141942HX, SEG12141943HX, SEG12141944HX, SEG12141945HX, SEG12141946HX, SEG12141947HX, SEG12141948HX, SEG12141949HX, SEG12141950HX, SEG12141951HX, SEG12141952HX, SEG12141953HX, SEG12141954HX, SEG12141955HX, SEG12141956HX, SEG12141957HX, SEG12141958HX, SEG12141959HX, SEG12141960HX, SEG12141961HX, SEG12141962HX, SEG12141963HX, SEG12152190HX, SEG12152191HX, SEG12152192HX, SEG12152193HX, SEG12152194HX, SEG12152195HX, SEG12152196HX, SEG12152197HX, SEG12152198HX, SEG12152199HX, SEG12152200HX, SEG12152201HX, SEG12152202HX, SEG12152203HX, SEG12152204HX, SEG12152205HX, SEG12152206HX, SEG12152207HX, SEG12152208HX, SEG12152209HX, SEG12152210HX, SEG12152211HX, SEG12152212HX, SEG12152213HX, SEG12152214HX, SEG12152215HX, SEG12152216HX, SEG12152217HX, SEG12152218HX, SEG12152219HX, SEG12152220HX, SEG12152221HX, SEG12152222HX, SEG12152223HX, SEG12152224HX, SEG12152225HX, SEG12152226HX, SEG12152227HX, SEG12152228HX, SEG12152229HX, SEG12152230HX, SEG12152231HX, SEG12152232HX, SEG12152233HX, SEG12152234HX, SEG12152235HX, SEG12152236HX, SEG12152237HX, SEG12152238HX, SEG12152239HX, SEG12152240HX, SEG12152241HX, SEG12152242HX, SEG12152243HX, SEG12152244HX, SEG12152245HX, SEG12152246HX, SEG12152247HX, SEG12152248HX, SEG12152249HX, SEG12152250HX, SEG12152251HX, SEG12152252HX, SEG12152253HX, SEG12152254HX, SEG12152255HX, SEG12152256HX, SEG12152257HX, SEG12152258HX, SEG12152259HX, SEG12152260HX, SEG12152261HX, SEG12152262HX, SEG12152263HX, SEG12152264HX, SEG12152265HX, SEG12152266HX, SEG12152267HX, SEG12152268HX, SEG12152269HX, SEG12152270HX, SEG12152271HX, SEG12152272HX, SEG12152273HX, SEG12152274HX, SEG12152275HX, SEG12152276HX, SEG12152277HX, SEG12152278HX, SEG12152279HX, SEG12152280HX, SEG12152281HX, SEG12152282HX, SEG12152283HX, SEG12152284HX, SEG12152285HX, SEG12152286HX, SEG12152287HX, SEG12152288HX, SEG12152289HX, SEG12172403HX, SEG12172404HX, SEG12172405HX, SEG12172406HX, SEG12172407HX, SEG12172408HX, SEG12172409HX, SEG12172410HX, SEG12172411HX, SEG12172412HX, SEG12172413HX, SEG12172414HX, SEG12172415HX, SEG12172416HX, SEG12172417HX, SEG12172418HX, SEG12172419HX, SEG12172420HX, SEG12172421HX, SEG12172422HX, SEG12172423HX, SEG12172424HX, SEG12172425HX, SEG12172426HX, SEG12172427HX, SEG12172428HX, SEG12172429HX, SEG12172430HX, SEG12172431HX, SEG12172432HX, SEG12172433HX, SEG12172434HX, SEG12172435HX, SEG12172436HX, SEG12172437HX, SEG12172438HX, SEG12172439HX, SEG12172440HX, SEG12172441HX, SEG12172442HX, SEG12172443HX, SEG12172444HX, SEG12172445HX, SEG12172446HX, SEG12172447HX, SEG12172448HX, SEG12172449HX, SEG12172450HX, SEG12172451HX, SEG12172452HX, SEG12172541HX, SEG12172542HX, SEG12172543HX, SEG12172544HX, SEG12172545HX, SEG12172546HX, SEG12172547HX, SEG12172548HX, SEG12172549HX, SEG12172550HX, SEG12172551HX, SEG12172552HX, SEG12172553HX, SEG12172554HX, SEG12172555HX, SEG12172556HX, SEG12172557HX, SEG12172558HX, SEG12172559HX, SEG12172560HX, SEG12172561HX, SEG12172562HX, SEG12172563HX, SEG12172564HX, SEG12172565HX, SEG12172566HX, SEG12172567HX, SEG12172568HX, SEG12172569HX, SEG12172570HX, SEG12182732HX, SEG12182733HX, SEG12182734HX, SEG12182735HX, SEG12182736HX, SEG12182737HX, SEG12182738HX, SEG12182739HX, SEG12182740HX, SEG12182742HX, SEG12182744HX, SEG12182745HX, SEG12182746HX, SEG12182747HX, SEG12182748HX, SEG12182749HX, SEG12182750HX, SEG12182751HX, SEG12182752HX, SEG12182753HX, SEG12182754HX, SEG12182755HX, SEG12182756HX, SEG12182757HX, SEG12182758HX, SEG12182759HX, SEG12182760HX, SEG12182762HX, SEG12182763HX, SEG12182764HX, SEG12182765HX, SEG12182766HX, SEG12182767HX, SEG12182768HX, SEG12182769HX, SEG12182770HX, SEG12182771HX, SEG12182772HX, SEG12182773HX, SEG12182774HX, SEG12182775HX, SEG12182776HX, SEG12182777HX, SEG12182778HX, SEG12182779HX, SEG12182780HX, SEG12182781HX, SEG12182782HX, SEG12182783HX, SEG12182784HX, SEG12182785HX, SEG12182786HX, SEG12182787HX, SEG12182788HX, SEG12182789HX, SEG12182790HX, SEG12182791HX, SEG12182792HX, SEG12182793HX, SEG12182794HX, SEG12182795HX, SEG12182796HX, SEG12182797HX, SEG12182798HX, SEG12182799HX, SEG12182800HX, SEG12182801HX, SEG12182802HX, SEG12182803HX, SEG12182804HX, SEG12182805HX, SEG12182806HX, SEG12182807HX, SEG12182808HX, SEG12182809HX, SEG12182810HX, SEG12182811HX, SEG12182812HX, SEG12182813HX, SEG12182814HX, SEG12182815HX, SEG12182816HX, SEG12182817HX, SEG12182818HX, SEG12182819HX, SEG12182820HX, SEG12182821HX, SEG12182822HX, SEG12182823HX, SEG12182824HX, SEG12182825HX, SEG12182826HX, SEG12182827HX, SEG12182828HX, SEG12182829HX, SEG12182830HX, SEG12182831HX, SEG12192902HX, SEG12192903HX, SEG12192904HX, SEG12192905HX, SEG12192906HX, SEG12192907HX, SEG12192908HX, SEG12192909HX, SEG12192910HX, SEG12192911HX, SEG12192912HX, SEG12192913HX, SEG12192914HX, SEG12192915HX, SEG12192916HX, SEG12192917HX, SEG12192918HX, SEG12192919HX, SEG12192920HX, SEG12192921HX, SEG12192922HX, SEG12192923HX, SEG12192924HX, SEG12192925HX, SEG12192926HX, SEG12192927HX, SEG12192928HX, SEG12192929HX, SEG12192930HX, SEG12192931HX, SEG12192932HX, SEG12192933HX, SEG12192934HX, SEG12192935HX, SEG12192936HX, SEG12192937HX, SEG12192938HX, SEG12192939HX, SEG12192940HX, SEG12192941HX, SEG12192942HX, SEG12192943HX, SEG12192944HX, SEG12192945HX, SEG12192946HX, SEG12192947HX, SEG12192948HX, SEG12192949HX, SEG12192950HX, SEG12192951HX, SEG12192952HX, SEG12192953HX, SEG12192954HX, SEG12192955HX, SEG12192956HX, SEG12192957HX, SEG12192958HX, SEG12192959HX, SEG12192960HX, SEG12192961HX, SEG12192962HX, SEG12192963HX, SEG12192964HX, SEG12192965HX, SEG12192966HX, SEG12192967HX, SEG12192968HX, SEG12192969HX, SEG12192970HX, SEG12192971HX, SEG12192972HX, SEG12192973HX, SEG12192974HX, SEG12192975HX, SEG12192976HX, SEG12192977HX, SEG12192978HX, SEG12192979HX, SEG12192980HX, SEG12192981HX, SEG12192982HX, SEG12192983HX, SEG12192984HX, SEG12192985HX, SEG12192986HX, SEG12192987HX, SEG12192988HX, SEG12192989HX, SEG12192990HX, SEG12192991HX, SEG12192992HX, SEG12192993HX, SEG12192994HX, SEG12192995HX, SEG12192996HX, SEG12192997HX, SEG12192998HX, SEG12192999HX, SEG12193000HX, SEG12193001HX, SEG12203221HX, SEG12203222HX, SEG12203223HX, SEG12203224HX, SEG12203225HX, SEG12203226HX, SEG12203227HX, SEG12203228HX, SEG12203229HX, SEG12203230HX, SEG12203231HX, SEG12203232HX, SEG12203233HX, SEG12203234HX, SEG12203235HX, SEG12203236HX, SEG12203237HX, SEG12203238HX, SEG12203239HX, SEG12203240HX, SEG12203241HX, SEG12203242HX, SEG12203243HX, SEG12203244HX, SEG12203245HX, SEG12203246HX, SEG12203247HX, SEG12203248HX, SEG12203249HX, SEG12203250HX, SEG12203251HX, SEG12203252HX, SEG12203253HX, SEG12203254HX, SEG12203255HX, SEG12203256HX, SEG12203257HX, SEG12203258HX, SEG12203259HX, SEG12203260HX, SEG12203261HX, SEG12203262HX, SEG12203263HX, SEG12203264HX, SEG12203265HX, SEG12203266HX, SEG12203267HX, SEG12203268HX, SEG12203269HX, SEG12203270HX, SEG12203271HX, SEG12203272HX, SEG12203273HX, SEG12203274HX, SEG12203275HX, SEG12203276HX, SEG12203277HX, SEG12203278HX, SEG12203279HX, SEG12203280HX, SEG12203281HX, SEG12203282HX, SEG12203283HX, SEG12203284HX, SEG12203285HX, SEG12203286HX, SEG12203287HX, SEG12203288HX, SEG12203289HX, SEG12203290HX, SEG12203291HX, SEG12203292HX, SEG12203293HX, SEG12203294HX, SEG12203295HX, SEG12203296HX, SEG12203297HX, SEG12203298HX, SEG12203299HX, SEG12203300HX, SEG12203301HX, SEG12203302HX, SEG12203303HX, SEG12203304HX, SEG12203305HX, SEG12203306HX, SEG12203307HX, SEG12203308HX, SEG12203309HX, SEG12203310HX, SEG12203311HX, SEG12203312HX, SEG12203313HX, SEG12203314HX, SEG12203315HX, SEG12203316HX, SEG12203317HX, SEG12203318HX, SEG12203319HX, SEG12203320HX, SEG12213680HX, SEG12213681HX, SEG12213682HX, SEG12213683HX, SEG12213684HX, SEG12213685HX, SEG12213686HX, SEG12213687HX, SEG12213688HX, SEG12213689HX, SEG12213690HX, SEG12213691HX, SEG12213692HX, SEG12213693HX, SEG12213694HX, SEG12213695HX, SEG12213696HX, SEG12213697HX, SEG12213698HX, SEG12213699HX, SEG12213700HX, SEG12213701HX, SEG12213702HX, SEG12213703HX, SEG12213704HX, SEG12213705HX, SEG12213706HX, SEG12213707HX, SEG12213708HX, SEG12213709HX, SEG12213710HX, SEG12213711HX, SEG12213712HX, SEG12213713HX, SEG12213714HX, SEG12213715HX, SEG12213716HX, SEG12213717HX, SEG12213718HX, SEG12213719HX, SEG12213720HX, SEG12213721HX, SEG12213722HX, SEG12213723HX, SEG12213724HX, SEG12213725HX, SEG12213726HX, SEG12213727HX, SEG12213728HX, SEG12213729HX, SEG12213730HX, SEG12213731HX, SEG12213732HX, SEG12213733HX, SEG12213734HX, SEG12213735HX, SEG12213736HX, SEG12213737HX, SEG12213738HX, SEG12213739HX, SEG12213740HX, SEG12213741HX, SEG12213742HX, SEG12213743HX, SEG12213744HX, SEG12213745HX, SEG12213746HX, SEG12213747HX, SEG12213748HX, SEG12213749HX, SEG12213750HX, SEG12213751HX, SEG12213752HX, SEG12213753HX, SEG12213754HX, SEG12213755HX, SEG12213756HX, SEG12213757HX, SEG12213758HX, SEG12213759HX, SEG12213760HX, SEG12213761HX, SEG12213762HX, SEG12213763HX, SEG12213764HX, SEG12213765HX, SEG12213766HX, SEG12213767HX, SEG12213768HX, SEG12213769HX, SEG12213770HX, SEG12213771HX, SEG12213772HX, SEG12213773HX, SEG12213774HX, SEG12213775HX, SEG12213776HX, SEG12213777HX, SEG12213778HX, SEG12213779HX, SEG12223802HX, SEG12223803HX, SEG12223804HX, SEG12223805HX, SEG12223806HX, SEG12223807HX, SEG12223808HX, SEG12223809HX, SEG12223810HX, SEG12223811HX, SEG12223812HX, SEG12223813HX, SEG12223814HX, SEG12223815HX, SEG12223816HX, SEG12223817HX, SEG12223818HX, SEG12223819HX, SEG12223820HX, SEG12223821HX, SEG12223822HX, SEG12223823HX, SEG12223824HX, SEG12223825HX, SEG12223826HX, SEG12223827HX, SEG12223828HX, SEG12223829HX, SEG12223830HX, SEG12223831HX, SEG12223832HX, SEG12223833HX, SEG12223834HX, SEG12223835HX, SEG12223836HX, SEG12223837HX, SEG12223838HX, SEG12223839HX, SEG12223840HX, SEG12223841HX, SEG12223842HX, SEG12223843HX, SEG12223844HX, SEG12223845HX, SEG12223846HX, SEG12223847HX, SEG12223848HX, SEG12223849HX, SEG12223850HX, SEG12223851HX, SEG12234212HX, SEG12234213HX, SEG12234214HX, SEG12234215HX, SEG12234216HX, SEG12234217HX, SEG12234218HX, SEG12234219HX, SEG12234220HX, SEG12234221HX, SEG12234222HX, SEG12234223HX, SEG12234224HX, SEG12234225HX, SEG12234226HX, SEG12234227HX, SEG12234228HX, SEG12234229HX, SEG12234230HX, SEG12234231HX, SEG12244893HX, SEG12244894HX, SEG12244895HX, SEG12244896HX, SEG12244897HX, SEG12244898HX, SEG12244899HX, SEG12244900HX, SEG12244901HX, SEG12244902HX, SEG12244903HX, SEG12244904HX, SEG12244905HX, SEG12244906HX, SEG12244907HX, SEG12244908HX, SEG12244909HX, SEG12244910HX, SEG12244911HX, SEG12244912HX, SEG12244913HX, SEG12244914HX, SEG12244915HX, SEG12244916HX, SEG12244917HX, SEG12244918HX, SEG12244919HX, SEG12244920HX, SEG12244921HX, SEG12244922HX, SEG12244923HX, SEG12244924HX, SEG12244925HX, SEG12244926HX, SEG12244927HX, SEG12244928HX, SEG12244929HX, SEG12244930HX, SEG12244931HX, SEG12244932HX, SEG12244933HX, SEG12244934HX, SEG12244935HX, SEG12244936HX, SEG12244937HX, SEG12244938HX, SEG12244939HX, SEG12244940HX, SEG12244941HX, SEG12244942HX, SEG12245152HX, SEG12245153HX, SEG12245154HX, SEG12245155HX, SEG12245156HX, SEG12245157HX, SEG12245158HX, SEG12245159HX, SEG12245160HX, SEG12245161HX, SEG12245162HX, SEG12245163HX, SEG12245164HX, SEG12245165HX, SEG12245166HX, SEG12245167HX, SEG12245168HX, SEG12245169HX, SEG12245170HX, SEG12245171HX, SEG12245172HX, SEG12245173HX, SEG12245174HX, SEG12245175HX, SEG12245176HX, SEG12245177HX, SEG12245178HX, SEG12245179HX, SEG12245180HX, SEG12245181HX, SEG12245182HX, SEG12245183HX, SEG12245184HX, SEG12245185HX, SEG12245186HX, SEG12245187HX, SEG12245188HX, SEG12245189HX, SEG12245190HX, SEG12245191HX, SEG12245192HX, SEG12245193HX, SEG12245194HX, SEG12245195HX, SEG12245196HX, SEG12245197HX, SEG12245198HX, SEG12245199HX, SEG12245200HX, SEG12245201HX, SEG12255268HX, SEG12255269HX, SEG12255270HX, SEG12255271HX, SEG12255272HX, SEG12255273HX, SEG12255274HX, SEG12255276HX, SEG12255277HX, SEG12255278HX, SEG12255279HX, SEG12255280HX, SEG12255281HX, SEG12255282HX, SEG12255283HX, SEG12255284HX, SEG12255285HX, SEG12255286HX, SEG12255287HX, SEG12255288HX, SEG12255289HX, SEG12255290HX, SEG12255291HX, SEG12255292HX, SEG12255293HX, SEG12255294HX, SEG12255295HX, SEG12255296HX, SEG12255297HX, SEG12255298HX, SEG12255299HX, SEG12255301HX, SEG12255302HX, SEG12255303HX, SEG12255304HX, SEG12255305HX, SEG12255306HX, SEG12255307HX, SEG12255308HX, SEG12255309HX, SEG12255310HX, SEG12255311HX, SEG12255312HX, SEG12255313HX, SEG12255314HX, SEG12255315HX, SEG12255316HX, SEG12255317HX, SEG12255318HX, SEG12265379HX, SEG12265380HX, SEG12265381HX, SEG12265382HX, SEG12265383HX, SEG12265384HX, SEG12265385HX, SEG12265386HX, SEG12265387HX, SEG12265388HX, SEG12265389HX, SEG12265390HX, SEG12265391HX, SEG12265392HX, SEG12265393HX, SEG12265394HX, SEG12265395HX, SEG12265396HX, SEG12265397HX, SEG12265398HX, SEG12265399HX, SEG12265400HX, SEG12265401HX, SEG12265402HX, SEG12265403HX, SEG12265404HX, SEG12265405HX, SEG12265406HX, SEG12265407HX, SEG12265408HX, SEG12265409HX, SEG12265410HX, SEG12265411HX, SEG12265412HX, SEG12265413HX, SEG12265414HX, SEG12265415HX, SEG12265416HX, SEG12265417HX, SEG12265418HX, SEG12265589HX, SEG12296104HX, SEG12296105HX, SEG12296106HX, SEG12296107HX, SEG12296108HX, SEG12296109HX, SEG12296110HX, SEG12296111HX, SEG12296112HX, SEG12296113HX, SEG12296114HX, SEG12296115HX, SEG12296116HX, SEG12296117HX, SEG12296118HX, SEG12296119HX, SEG12296120HX, SEG12296121HX, SEG12296122HX, SEG12296123HX, SEG12296124HX, SEG12296125HX, SEG12296126HX, SEG12296127HX, SEG12296128HX, SEG12296129HX, SEG12296130HX, SEG12296131HX, SEG12296132HX, SEG12296133HX, SEG12296134HX, SEG12296135HX, SEG12296136HX, SEG12296137HX, SEG12296138HX, SEG12296139HX, SEG12296140HX, SEG12296141HX, SEG12296142HX, SEG12296143HX, SEG12296144HX, SEG12296145HX, SEG12296146HX, SEG12296147HX, SEG12296148HX, SEG12296149HX, SEG12296150HX, SEG12296151HX, SEG12296152HX, SEG12296153HX, SEG12296154HX, SEG12296155HX, SEG12296156HX, SEG12296157HX, SEG12296158HX, SEG12296159HX, SEG12296160HX, SEG12296161HX, SEG12296162HX, SEG12296163HX, SEG12296164HX, SEG12296165HX, SEG12296166HX, SEG12296167HX, SEG12296168HX, SEG12296169HX, SEG12296170HX, SEG12296171HX, SEG12296172HX, SEG12296173HX, SEG12296174HX, SEG12296175HX, SEG12296176HX, SEG12296177HX, SEG12296178HX, SEG12316669HX, SEG12316670HX, SEG12316671HX, SEG12316672HX, SEG12316673HX, SEG12316674HX, SEG12316675HX, SEG12316676HX, SEG12316677HX, SEG12316678HX, SEG12316680HX, SEG12316681HX, SEG12316682HX, SEG12316683HX, SEG12316684HX, SEG12316686HX, SEG12316689HX, SEG12316690HX, SEG12316691HX, SEG12316692HX, SEG12316693HX, SEG12316694HX, SEG12316695HX, SEG12336994HX, SEG12336995HX, SEG12336996HX, SEG12336997HX, SEG12336998HX, SEG12336999HX, SEG12337000HX, SEG12337001HX, SEG12337002HX, SEG12337003HX, SEG12337004HX, SEG12337005HX, SEG12337006HX, SEG12337007HX, SEG12337008HX, SEG12337009HX, SEG12337010HX, SEG12337011HX, SEG12337012HX, SEG12337013HX, SEG12337014HX, SEG12337015HX, SEG12337016HX, SEG12337017HX, SEG12337018HX, SEG12337019HX, SEG12337020HX, SEG12337021HX, SEG12337022HX, SEG12337023HX, SEG12347026HX, SEG12347027HX, SEG12347028HX, SEG12347029HX, SEG12347030HX, SEG12347031HX, SEG12347032HX, SEG12347033HX, SEG12347034HX, SEG12347035HX, SEG12347036HX, SEG12347037HX, SEG12347038HX, SEG12347039HX, SEG12347040HX, SEG12347041HX, SEG12347042HX, SEG12347043HX, SEG12347044HX, SEG12347045HX, SEG12347046HX, SEG12347047HX, SEG12347048HX, SEG12347049HX, SEG12347050HX, SEG12347051HX, SEG12347052HX, SEG12347053HX, SEG12347054HX, SEG12347055HX, SEG12347056HX, SEG12347057HX, SEG12347058HX, SEG12347059HX, SEG12347060HX, SEG12347061HX, SEG12347062HX, SEG12347063HX, SEG12347064HX, SEG12347065HX, SEG12347066HX, SEG12347067HX, SEG12347068HX, SEG12347069HX, SEG12347070HX, SEG12347071HX, SEG12347072HX, SEG12347073HX, SEG12347074HX, SEG12347075HX, SEG12347076HX, SEG12347077HX, SEG12347078HX, SEG12347079HX, SEG12347080HX, SEG12347081HX, SEG12347082HX, SEG12347083HX, SEG12347084HX, SEG12347085HX, SEG12347086HX, SEG12347087HX, SEG12347088HX, SEG12347089HX, SEG12347090HX, SEG12347091HX, SEG12347092HX, SEG12347093HX, SEG12347094HX, SEG12347095HX, SEG12347096HX, SEG12347097HX, SEG12347098HX, SEG12347099HX, SEG12347100HX, SEG12367734HX, SEG12367735HX, SEG12367736HX, SEG12367737HX, SEG12367738HX, SEG12367739HX, SEG12367740HX, SEG12367741HX, SEG12367742HX, SEG12367743HX, SEG12367744HX, SEG12367745HX, SEG12367746HX, SEG12367747HX, SEG12367748HX, SEG12367749HX, SEG12367750HX, SEG12367751HX, SEG12367752HX, SEG12367753HX, SEG12367754HX, SEG12367755HX, SEG12367756HX, SEG12367757HX, SEG12367758HX, SEG12367759HX, SEG12367760HX, SEG12367761HX, SEG12367762HX, SEG12367763HX, SEG12367764HX, SEG12367765HX, SEG12367766HX, SEG12367767HX, SEG12367768HX, SEG12367769HX, SEG12367770HX, SEG12367771HX, SEG12367772HX, SEG12367773HX, SEG12367774HX, SEG12367775HX, SEG12367776HX, SEG12367777HX, SEG12367778HX, SEG12367779HX, SEG12367780HX, SEG12367781HX, SEG12367782HX, SEG12367783HX, SEG12367834HX, SEG12367835HX, SEG12367836HX, SEG12367837HX, SEG12367838HX, SEG12367839HX, SEG12367840HX, SEG12367841HX, SEG12367842HX, SEG12367843HX, SEG12367844HX, SEG12367845HX, SEG12367846HX, SEG12367847HX, SEG12367848HX, SEG12367849HX, SEG12367850HX, SEG12367851HX, SEG12367852HX, SEG12367853HX, SEG12367854HX, SEG12367855HX, SEG12367856HX, SEG12367857HX, SEG12367858HX, SEG12367859HX, SEG12367860HX, SEG12367861HX, SEG12367862HX, SEG12367863HX, SEG12367864HX, SEG12367865HX, SEG12367866HX, SEG12367867HX, SEG12367868HX, SEG12367869HX, SEG12367870HX, SEG12367871HX, SEG12367872HX, SEG12367873HX, SEG12367874HX, SEG12367875HX, SEG12367876HX, SEG12367877HX, SEG12367878HX, SEG12367879HX, SEG12367880HX, SEG12367881HX, SEG12367882HX, SEG12367883HX, SEG12368084HX, SEG12368085HX, SEG12368086HX, SEG12368087HX, SEG12368088HX, SEG12368089HX, SEG12368090HX, SEG12368091HX, SEG12368092HX, SEG12368093HX, SEG12368094HX, SEG12368095HX, SEG12368096HX, SEG12368097HX, SEG12368098HX, SEG12368099HX, SEG12368100HX, SEG12368101HX, SEG12368102HX, SEG12368103HX, SEG12368104HX, SEG12368105HX, SEG12368106HX, SEG12368107HX, SEG12368108HX, SEG12368109HX, SEG12368110HX, SEG12368111HX, SEG12368112HX, SEG12368113HX, SEG12368114HX, SEG12368115HX, SEG12368116HX, SEG12368117HX, SEG12368118HX, SEG12368119HX, SEG12368120HX, SEG12368121HX, SEG12368122HX, SEG12368123HX, SEG12368124HX, SEG12368125HX, SEG12368126HX, SEG12368127HX, SEG12368128HX, SEG12368129HX, SEG12368130HX, SEG12368131HX, SEG12368132HX, SEG12368133HX, SEG12368134HX, SEG12368135HX, SEG12368136HX, SEG12368137HX, SEG12368138HX, SEG12368139HX, SEG12368140HX, SEG12368141HX, SEG12368142HX, SEG12368143HX, SEG12368144HX, SEG12368145HX, SEG12368146HX, SEG12368147HX, SEG12368148HX, SEG12368149HX, SEG12368150HX, SEG12368151HX, SEG12368152HX, SEG12368153HX, SEG12368154HX, SEG12368155HX, SEG12368156HX, SEG12368157HX, SEG12368158HX, SEG12368159HX, SEG12368160HX, SEG12368161HX, SEG12368162HX, SEG12368163HX, SEG12368164HX, SEG12368165HX, SEG12368166HX, SEG12368167HX, SEG12368168HX, SEG12368169HX, SEG12368170HX, SEG12368171HX, SEG12368172HX, SEG12368173HX, SEG12368174HX, SEG12368175HX, SEG12368176HX, SEG12368177HX, SEG12368178HX, SEG12368179HX, SEG12368180HX, SEG12368181HX, SEG12368182HX, SEG12368183HX, SEG12368184HX, SEG12368185HX, SEG12368186HX, SEG12368187HX, SEG12368188HX, SEG12368189HX, SEG12368190HX, SEG12368191HX, SEG12368192HX, SEG12368193HX, SEG12368194HX, SEG12368195HX, SEG12368196HX, SEG12368197HX, SEG12368198HX, SEG12368199HX, SEG12368200HX, SEG12368201HX, SEG12368202HX, SEG12368203HX, SEG12368204HX, SEG12368205HX, SEG12368206HX, SEG12368207HX, SEG12368208HX, SEG12368209HX, SEG12368210HX, SEG12368211HX, SEG12368212HX, SEG12368213HX, SEG12368214HX, SEG12368215HX, SEG12368216HX, SEG12368217HX, SEG12368218HX, SEG12368219HX, SEG12368220HX, SEG12368221HX, SEG12368222HX, SEG12368223HX, SEG12368224HX, SEG12368225HX, SEG12368226HX, SEG12368227HX, SEG12368228HX, SEG12368229HX, SEG12368230HX, SEG12368231HX, SEG12368232HX, SEG12368233HX, SEG12368234HX, SEG12368235HX, SEG12368236HX, SEG12368237HX, SEG12368238HX, SEG12368239HX, SEG12368240HX, SEG12368241HX, SEG12368242HX, SEG12368243HX, SEG12368244HX, SEG12368245HX, SEG12368246HX, SEG12368247HX, SEG12368248HX, SEG12368249HX, SEG12368250HX, SEG12368251HX, SEG12368252HX, SEG12368253HX, SEG12368254HX, SEG12368255HX, SEG12368256HX, SEG12368257HX, SEG12368258HX, SEG12368259HX, SEG12368260HX, SEG12368261HX, SEG12368262HX, SEG12368263HX, SEG12368264HX, SEG12368265HX, SEG12368266HX, SEG12368267HX, SEG12368268HX, SEG12368269HX, SEG12388781HX, SEG12388782HX, SEG12388783HX, SEG12388784HX, SEG12388785HX, SEG12388786HX, SEG12388787HX, SEG12388788HX, SEG12388789HX, SEG12388790HX, SEG12388791HX, SEG12388792HX, SEG12388793HX, SEG12388794HX, SEG12388795HX, SEG12388796HX, SEG12388797HX, SEG12388798HX, SEG12388799HX, SEG12388800HX, SEG12388801HX, SEG12388802HX, SEG12388803HX, SEG12388804HX, SEG12388805HX, SEG12388806HX, SEG12388807HX, SEG12388808HX, SEG12388809HX, SEG12388810HX, SEG12388811HX, SEG12388812HX, SEG12388813HX, SEG12388814HX, SEG12388815HX, SEG12388816HX, SEG12388817HX, SEG12388818HX, SEG12388819HX, SEG12388820HX, SEG12388821HX, SEG12388822HX, SEG12388823HX, SEG12388824HX, SEG12388825HX, SEG12388826HX, SEG12388827HX, SEG12388828HX, SEG12388829HX, SEG12388830HX, SEG12389092HX, SEG12389093HX, SEG12389094HX, SEG12389095HX, SEG12389096HX, SEG12389097HX, SEG12389098HX, SEG12389099HX, SEG12389100HX, SEG12389101HX, SEG12389102HX, SEG12389103HX, SEG12389104HX, SEG12389105HX, SEG12389106HX, SEG12389107HX, SEG12389108HX, SEG12389109HX, SEG12389110HX, SEG12389111HX, SEG12389112HX, SEG12389113HX, SEG12389114HX, SEG12389115HX, SEG12389116HX, SEG12389117HX, SEG12389118HX, SEG12389119HX, SEG12389120HX, SEG12389121HX, SEG12389122HX, SEG12389123HX, SEG12389124HX, SEG12389125HX, SEG12389126HX, SEG12389127HX, SEG12389128HX, SEG12389129HX, SEG12389130HX, SEG12389131HX, SEG12389132HX, SEG12389133HX, SEG12389134HX, SEG12389135HX, SEG12389136HX, SEG12389137HX, SEG12389138HX, SEG12389139HX, SEG12389140HX, SEG12389141HX, SEG12389142HX, SEG12389143HX, SEG12389144HX, SEG12389145HX, SEG12389146HX, SEG12389147HX, SEG12389148HX, SEG12389149HX, SEG12389150HX, SEG12389151HX, SEG12389152HX, SEG12389153HX, SEG12389154HX, SEG12389155HX, SEG12389156HX, SEG12389157HX, SEG12389158HX, SEG12389159HX, SEG12389160HX, SEG12389161HX, SEG12389162HX, SEG12389163HX, SEG12389164HX, SEG12389165HX, SEG12389166HX, SEG12389167HX, SEG12389168HX, SEG12389169HX, SEG12389170HX, SEG12389171HX, SEG12389172HX, SEG12389173HX, SEG12389174HX, SEG12389175HX, SEG12389176HX, SEG12389177HX, SEG12389178HX, SEG12389179HX, SEG12389180HX, SEG12389181HX, SEG12389182HX, SEG12389183HX, SEG12389184HX, SEG12389185HX, SEG12389186HX, SEG12389187HX, SEG12389188HX, SEG12389189HX, SEG12389190HX, SEG12389191HX, SEG12389192HX, SEG12389193HX, SEG12389194HX, SEG12389195HX, SEG12389196HX, SEG12389197HX, SEG12389198HX, SEG12389199HX, SEG12389200HX, SEG12389201HX, SEG12389202HX, SEG12389203HX, SEG12389204HX, SEG12389205HX, SEG12389206HX, SEG12389207HX, SEG12389208HX, SEG12389209HX, SEG12389210HX, SEG12389211HX, SEG12389212HX, SEG12389213HX, SEG12389214HX, SEG12389215HX, SEG12389216HX, SEG12389217HX, SEG12389218HX, SEG12389219HX, SEG12389220HX, SEG12389221HX, SEG12389222HX, SEG12389223HX, SEG12389224HX, SEG12389225HX, SEG12389226HX, SEG12389227HX, SEG12389228HX, SEG12389229HX, SEG12389230HX, SEG12389231HX, SEG12389232HX, SEG12389233HX, SEG12389234HX, SEG12389235HX, SEG12389236HX, SEG12389237HX, SEG12389238HX, SEG12389239HX, SEG12389240HX, SEG12389241HX, SEG12409746HX, SEG12409747HX, SEG12409748HX, SEG12409749HX, SEG12409750HX, SEG12409751HX, SEG12409752HX, SEG12409753HX, SEG12409754HX, SEG12409755HX, SEG12409756HX, SEG12409757HX, SEG12409758HX, SEG12409759HX, SEG12409760HX, SEG12409761HX, SEG12409762HX, SEG12409763HX, SEG12409764HX, SEG12409765HX, SEG12409766HX, SEG12409767HX, SEG12409768HX, SEG12409769HX, SEG12409770HX, SEG12409771HX, SEG12409772HX, SEG12409773HX, SEG12409774HX, SEG12409775HX, SEG12409812HX, SEG12409813HX, SEG12409814HX, SEG12409815HX, SEG12409816HX, SEG12409817HX, SEG12409818HX, SEG12409819HX, SEG12409820HX, SEG12409821HX, SEG12409822HX, SEG12409823HX, SEG12409824HX, SEG12409825HX, SEG12409826HX, SEG12409827HX, SEG12409828HX, SEG12409829HX, SEG12409830HX, SEG12409831HX, SEG12409832HX, SEG12409833HX, SEG12409834HX, SEG12409835HX, SEG12409836HX, SEG12409837HX, SEG12409838HX, SEG12409839HX, SEG12409840HX, SEG12409841HX, SEG12409842HX, SEG12409843HX, SEG12409844HX, SEG12409845HX, SEG12409846HX, SEG12409847HX, SEG12409848HX, SEG12409849HX, SEG12409850HX, SEG12409851HX, SEG12409852HX, SEG12409853HX, SEG12409854HX, SEG12409855HX, SEG12409856HX, SEG12409857HX, SEG12409858HX, SEG12409859HX, SEG12409860HX, SEG12409861HX, SEG12410126HX, SEG12410127HX, SEG12410128HX, SEG12410129HX, SEG12410130HX, SEG12410131HX, SEG12410132HX, SEG12410133HX, SEG12410134HX, SEG12410135HX, SEG12410136HX, SEG12410137HX, SEG12410138HX, SEG12410139HX, SEG12410140HX, SEG12410141HX, SEG12410142HX, SEG12410143HX, SEG12410144HX, SEG12410145HX, SEG12410146HX, SEG12410147HX, SEG12410148HX, SEG12410149HX, SEG12410150HX, SEG12410151HX, SEG12410152HX, SEG12410153HX, SEG12410154HX, SEG12410155HX, SEG12410156HX, SEG12410157HX, SEG12410158HX, SEG12410159HX, SEG12410160HX, SEG12410161HX, SEG12410162HX, SEG12410163HX, SEG12410164HX, SEG12410165HX, SEG12410166HX, SEG12410167HX, SEG12410168HX, SEG12410169HX, SEG12410170HX, SEG12410171HX, SEG12410172HX, SEG12410173HX, SEG12410174HX, SEG12410175HX, SEG12410306HX, SEG12410307HX, SEG12410308HX, SEG12410309HX, SEG12410310HX, SEG12410311HX, SEG12410312HX, SEG12410313HX, SEG12410314HX, SEG12410315HX, SEG12410316HX, SEG12410317HX, SEG12410318HX, SEG12410319HX, SEG12410320HX, SEG12410321HX, SEG12410322HX, SEG12410323HX, SEG12410324HX, SEG12410325HX, SEG12410326HX, SEG12410327HX, SEG12410328HX, SEG12410329HX, SEG12410330HX, SEG12410331HX, SEG12410332HX, SEG12410333HX, SEG12410334HX, SEG12410335HX, SEG12410336HX, SEG12410337HX, SEG12410338HX, SEG12410339HX, SEG12410340HX, SEG12410341HX, SEG12410342HX, SEG12410343HX, SEG12410344HX, SEG12410345HX, SEG12410346HX, SEG12410347HX, SEG12410348HX, SEG12410349HX, SEG12410350HX, SEG12410351HX, SEG12410352HX, SEG12410353HX, SEG12410354HX, SEG12410355HX, SEG12420597HX, SEG12420598HX, SEG12420599HX, SEG12420600HX, SEG12420601HX, SEG12420602HX, SEG12420603HX, SEG12420604HX, SEG12420605HX, SEG12420606HX, SEG12420607HX, SEG12420608HX, SEG12420609HX, SEG12420610HX, SEG12420611HX, SEG12420612HX, SEG12420613HX, SEG12420614HX, SEG12420615HX, SEG12420616HX, SEG12420617HX, SEG12420618HX, SEG12420619HX, SEG12420620HX, SEG12420621HX, SEG12420622HX, SEG12420623HX, SEG12420624HX, SEG12420625HX, SEG12420626HX, SEG12420627HX, SEG12420628HX, SEG12420629HX, SEG12420630HX, SEG12420631HX, SEG12420632HX, SEG12420633HX, SEG12420634HX, SEG12420635HX, SEG12420636HX, SEG12420637HX, SEG12420638HX, SEG12420639HX, SEG12420640HX, SEG12420641HX, SEG12420642HX, SEG12420643HX, SEG12420644HX, SEG12420645HX, SEG12420646HX, SEG12420647HX, SEG12420648HX, SEG12420649HX, SEG12420650HX, SEG12420651HX, SEG12420652HX, SEG12420653HX, SEG12420654HX, SEG12420655HX, SEG12420656HX, SEG12420657HX, SEG12420658HX, SEG12420659HX, SEG12420660HX, SEG12420661HX, SEG12420662HX, SEG12420663HX, SEG12420664HX, SEG12420665HX, SEG12420666HX, SEG12420667HX, SEG12420668HX, SEG12420669HX, SEG12420670HX, SEG12420671HX, SEG12420672HX, SEG12420673HX, SEG12420674HX, SEG12420675HX, SEG12420676HX, SEG12420677HX, SEG12420678HX, SEG12420679HX, SEG12420680HX, SEG12420681HX, SEG12420682HX, SEG12420683HX, SEG12420684HX, SEG12420685HX, SEG12420686HX, SEG12420687HX, SEG12420688HX, SEG12420689HX, SEG12420690HX, SEG12420691HX, SEG12420692HX, SEG12420693HX, SEG12420694HX, SEG12420695HX, SEG12420696HX, SEG12431191HX, SEG12431192HX, SEG12431193HX, SEG12431194HX, SEG12431195HX, SEG12431196HX, SEG12431197HX, SEG12431198HX, SEG12431199HX, SEG12431200HX, SEG12431201HX, SEG12431202HX, SEG12431203HX, SEG12431204HX, SEG12431205HX, SEG12431206HX, SEG12431207HX, SEG12431208HX, SEG12431209HX, SEG12431210HX, SEG12431211HX, SEG12431212HX, SEG12431213HX, SEG12431214HX, SEG12431215HX, SEG12431216HX, SEG12431217HX, SEG12431218HX, SEG12431219HX, SEG12431220HX, SEG12431221HX, SEG12431222HX, SEG12431223HX, SEG12431224HX, SEG12431225HX, SEG12431226HX, SEG12431227HX, SEG12431228HX, SEG12431229HX, SEG12431230HX, SEG12431231HX, SEG12431232HX, SEG12431233HX, SEG12431234HX, SEG12431235HX, SEG12431236HX, SEG12431237HX, SEG12431238HX, SEG12431239HX, SEG12431240HX, SEG12431241HX, SEG12431242HX, SEG12431243HX, SEG12431244HX, SEG12431245HX, SEG12431246HX, SEG12431247HX, SEG12431248HX, SEG12431249HX, SEG12431250HX, SEG12431251HX, SEG12431252HX, SEG12431253HX, SEG12431254HX, SEG12431255HX, SEG12431256HX, SEG12431257HX, SEG12431258HX, SEG12431259HX, SEG12431260HX, SEG12431261HX, SEG12431262HX, SEG12431263HX, SEG12431264HX, SEG12431265HX, SEG12431266HX, SEG12431267HX, SEG12431268HX, SEG12431269HX, SEG12431270HX, SEG12431271HX, SEG12431272HX, SEG12431273HX, SEG12431274HX, SEG12431275HX, SEG12431276HX, SEG12431277HX, SEG12431278HX, SEG12431279HX, SEG12431280HX, SEG12431281HX, SEG12431282HX, SEG12431283HX, SEG12431284HX, SEG12431285HX, SEG12431286HX, SEG12431287HX, SEG12431288HX, SEG12431289HX, SEG12431290HX, SEG12441709HX, SEG12441710HX, SEG12441711HX, SEG12441712HX, SEG12441713HX, SEG12441714HX, SEG12441715HX, SEG12441716HX, SEG12441717HX, SEG12441718HX, SEG12441719HX, SEG12441720HX, SEG12441721HX, SEG12441722HX, SEG12441723HX, SEG12441724HX, SEG12441725HX, SEG12441726HX, SEG12441727HX, SEG12441728HX, SEG12441729HX, SEG12441730HX, SEG12441731HX, SEG12441732HX, SEG12441733HX, SEG12441734HX, SEG12441735HX, SEG12441736HX, SEG12441737HX, SEG12441738HX, SEG12441739HX, SEG12441740HX, SEG12441741HX, SEG12441742HX, SEG12441743HX, SEG12441744HX, SEG12441745HX, SEG12441746HX, SEG12441747HX, SEG12441748HX, SEG12441749HX, SEG12441750HX, SEG12441751HX, SEG12441752HX, SEG12441753HX, SEG12441754HX, SEG12441755HX, SEG12441756HX, SEG12441757HX, SEG12441758HX, SEG12441759HX, SEG12441760HX, SEG12441761HX, SEG12441762HX, SEG12441763HX, SEG12441764HX, SEG12441765HX, SEG12441766HX, SEG12441767HX, SEG12441768HX, SEG12441769HX, SEG12441770HX, SEG12441771HX, SEG12441772HX, SEG12441773HX, SEG12441774HX, SEG12441775HX, SEG12441776HX, SEG12441777HX, SEG12441778HX, SEG12441779HX, SEG12441780HX, SEG12441781HX, SEG12441782HX, SEG12441783HX, SEG12441784HX, SEG12441785HX, SEG12441786HX, SEG12441787HX, SEG12441788HX, SEG12441789HX, SEG12441790HX, SEG12441791HX, SEG12441792HX, SEG12441793HX, SEG12441794HX, SEG12441795HX, SEG12441796HX, SEG12441797HX, SEG12441798HX, SEG12441799HX, SEG12441800HX, SEG12441801HX, SEG12441802HX, SEG12441803HX, SEG12441804HX, SEG12441805HX, SEG12441806HX, SEG12441807HX, SEG12441808HX, SEG12452053HX, SEG12452054HX, SEG12452055HX, SEG12452056HX, SEG12452057HX, SEG12452058HX, SEG12452059HX, SEG12452060HX, SEG12452061HX, SEG12452062HX, SEG12452063HX, SEG12452064HX, SEG12452065HX, SEG12452066HX, SEG12452067HX, SEG12452068HX, SEG12452069HX, SEG12452070HX, SEG12452071HX, SEG12452072HX, SEG12452073HX, SEG12452074HX, SEG12452075HX, SEG12452076HX, SEG12452077HX, SEG12452078HX, SEG12452079HX, SEG12452080HX, SEG12452081HX, SEG12452082HX, SEG12452083HX, SEG12452084HX, SEG12452085HX, SEG12452086HX, SEG12452087HX, SEG12452088HX, SEG12452089HX, SEG12452090HX, SEG12452091HX, SEG12452092HX, SEG12452093HX, SEG12452094HX, SEG12452095HX, SEG12452096HX, SEG12452097HX, SEG12452098HX, SEG12452099HX, SEG12452100HX, SEG12452101HX, SEG12452102HX, SEG12452103HX, SEG12452104HX, SEG12452105HX, SEG12452106HX, SEG12452107HX, SEG12452108HX, SEG12452109HX, SEG12452110HX, SEG12452111HX, SEG12452112HX, SEG12452113HX, SEG12452114HX, SEG12452115HX, SEG12452116HX, SEG12452117HX, SEG12452118HX, SEG12452119HX, SEG12452120HX, SEG12452121HX, SEG12452122HX, SEG12452123HX, SEG12452124HX, SEG12452125HX, SEG12452126HX, SEG12452127HX, SEG12452128HX, SEG12452129HX, SEG12452130HX, SEG12452131HX, SEG12452132HX, SEG12452133HX, SEG12452134HX, SEG12452135HX, SEG12452136HX, SEG12452137HX, SEG12452138HX, SEG12452139HX, SEG12452140HX, SEG12452141HX, SEG12452142HX, SEG12452143HX, SEG12452144HX, SEG12452145HX, SEG12452146HX, SEG12452147HX, SEG12452148HX, SEG12452149HX, SEG12452150HX, SEG12452151HX, SEG12452152HX, SEG12462259HX, SEG12462260HX, SEG12462261HX, SEG12462262HX, SEG12462263HX, SEG12462264HX, SEG12462265HX, SEG12462266HX, SEG12462267HX, SEG12462268HX, SEG12462269HX, SEG12462270HX, SEG12462271HX, SEG12462272HX, SEG12462273HX, SEG12462274HX, SEG12462275HX, SEG12462276HX, SEG12462277HX, SEG12462278HX, SEG12462279HX, SEG12462280HX, SEG12462281HX, SEG12462282HX, SEG12462283HX, SEG12462284HX, SEG12462285HX, SEG12462286HX, SEG12462287HX, SEG12462288HX, SEG12462289HX, SEG12462290HX, SEG12462291HX, SEG12462292HX, SEG12462293HX, SEG12462294HX, SEG12462295HX, SEG12462296HX, SEG12462297HX, SEG12462298HX, SEG12462299HX, SEG12462300HX, SEG12462301HX, SEG12462302HX, SEG12462303HX, SEG12462304HX, SEG12462305HX, SEG12462306HX, SEG12462307HX, SEG12462308HX, SEG12462309HX, SEG12462310HX, SEG12462311HX, SEG12462312HX, SEG12462313HX, SEG12462314HX, SEG12462315HX, SEG12462316HX, SEG12462317HX, SEG12462318HX, SEG12462319HX, SEG12462320HX, SEG12462321HX, SEG12462322HX, SEG12462323HX, SEG12462324HX, SEG12462325HX, SEG12462326HX, SEG12462327HX, SEG12462328HX, SEG12462329HX, SEG12462330HX, SEG12462331HX, SEG12462332HX, SEG12462333HX, SEG12462334HX, SEG12462335HX, SEG12462336HX, SEG12462337HX, SEG12462338HX, SEG12462339HX, SEG12462340HX, SEG12462341HX, SEG12462342HX, SEG12462343HX, SEG12462344HX, SEG12462345HX, SEG12462346HX, SEG12462347HX, SEG12462348HX, SEG12462349HX, SEG12462350HX, SEG12462351HX, SEG12462352HX, SEG12462353HX, SEG12462354HX, SEG12462355HX, SEG12462356HX, SEG12462357HX, SEG12462358HX, SEG12472531HX, SEG12472532HX, SEG12472533HX, SEG12472534HX, SEG12472535HX, SEG12472536HX, SEG12472537HX, SEG12472538HX, SEG12472539HX, SEG12472540HX, SEG12472541HX, SEG12472542HX, SEG12472543HX, SEG12472544HX, SEG12472545HX, SEG12472546HX, SEG12472547HX, SEG12472548HX, SEG12472549HX, SEG12472550HX, SEG12472551HX, SEG12472552HX, SEG12472553HX, SEG12472554HX, SEG12472555HX, SEG12472556HX, SEG12472557HX, SEG12472558HX, SEG12472559HX, SEG12472560HX, SEG12472561HX, SEG12472562HX, SEG12472563HX, SEG12472564HX, SEG12472565HX, SEG12472566HX, SEG12472567HX, SEG12472568HX, SEG12472569HX, SEG12472570HX, SEG12472571HX, SEG12472572HX, SEG12472573HX, SEG12472574HX, SEG12472575HX, SEG12472576HX, SEG12472577HX, SEG12472578HX, SEG12472579HX, SEG12472580HX, SEG12472607HX, SEG12472608HX, SEG12472609HX, SEG12472610HX, SEG12472611HX, SEG12472612HX, SEG12472613HX, SEG12472614HX, SEG12472615HX, SEG12472616HX, SEG12472617HX, SEG12472618HX, SEG12472619HX, SEG12472620HX, SEG12472621HX, SEG12472622HX, SEG12472623HX, SEG12472624HX, SEG12472625HX, SEG12472626HX, SEG12472627HX, SEG12472628HX, SEG12472629HX, SEG12472630HX, SEG12472631HX, SEG12472632HX, SEG12472633HX, SEG12472634HX, SEG12472635HX, SEG12472636HX, SEG12472637HX, SEG12472638HX, SEG12472639HX, SEG12472640HX, SEG12472641HX, SEG12472642HX, SEG12472643HX, SEG12472644HX, SEG12472645HX, SEG12472646HX, SEG12472647HX, SEG12472648HX, SEG12472649HX, SEG12472650HX, SEG12472651HX, SEG12472652HX, SEG12472653HX, SEG12472654HX, SEG12472655HX, SEG12472656HX, SEG12472657HX, SEG12472658HX, SEG12472659HX, SEG12472660HX, SEG12472661HX, SEG12472662HX, SEG12472663HX, SEG12472664HX, SEG12472665HX, SEG12472666HX, SEG12472667HX, SEG12472668HX, SEG12472669HX, SEG12472670HX, SEG12472671HX, SEG12472672HX, SEG12472673HX, SEG12472674HX, SEG12472675HX, SEG12472676HX, SEG12472677HX, SEG12472678HX, SEG12472679HX, SEG12472680HX, SEG12472681HX, SEG12472682HX, SEG12472683HX, SEG12472684HX, SEG12472685HX, SEG12472686HX, SEG12472687HX, SEG12472688HX, SEG12472689HX, SEG12472690HX, SEG12472691HX, SEG12472692HX, SEG12472693HX, SEG12472694HX, SEG12472695HX, SEG12472696HX, SEG12472697HX, SEG12472698HX, SEG12472699HX, SEG12472700HX, SEG12472701HX, SEG12472702HX, SEG12472703HX, SEG12472704HX, SEG12472705HX, SEG12472706HX, SEG12482825HX, SEG12482826HX, SEG12482827HX, SEG12482828HX, SEG12482829HX, SEG12482830HX, SEG12482831HX, SEG12482832HX, SEG12482833HX, SEG12482834HX, SEG12482835HX, SEG12482836HX, SEG12482837HX, SEG12482838HX, SEG12482839HX, SEG12482840HX, SEG12482841HX, SEG12482842HX, SEG12482843HX, SEG12482844HX, SEG12482845HX, SEG12482846HX, SEG12482847HX, SEG12482848HX, SEG12482849HX, SEG12482850HX, SEG12482851HX, SEG12482852HX, SEG12482853HX, SEG12482854HX, SEG12482855HX, SEG12482856HX, SEG12482857HX, SEG12482858HX, SEG12482859HX, SEG12482860HX, SEG12482861HX, SEG12482862HX, SEG12482863HX, SEG12482864HX, SEG12482865HX, SEG12482866HX, SEG12482867HX, SEG12482868HX, SEG12482869HX, SEG12482870HX, SEG12482871HX, SEG12482872HX, SEG12482873HX, SEG12482874HX, SEG12493319HX, SEG12493320HX, SEG12493321HX, SEG12493322HX, SEG12493323HX, SEG12493324HX, SEG12493325HX, SEG12493326HX, SEG12493327HX, SEG12493328HX, SEG12493329HX, SEG12493330HX, SEG12493331HX, SEG12493332HX, SEG12493333HX, SEG12493334HX, SEG12493335HX, SEG12493336HX, SEG12493337HX, SEG12493338HX, SEG12493339HX, SEG12493340HX, SEG12493341HX, SEG12493342HX, SEG12493343HX, SEG12493344HX, SEG12493345HX, SEG12493346HX, SEG12493347HX, SEG12493348HX, SEG12493349HX, SEG12493350HX, SEG12493351HX, SEG12493352HX, SEG12493353HX, SEG12493354HX, SEG12493355HX, SEG12493356HX, SEG12493357HX, SEG12493358HX, SEG12493359HX, SEG12493360HX, SEG12493361HX, SEG12493362HX, SEG12493363HX, SEG12493364HX, SEG12493365HX, SEG12493366HX, SEG12493367HX, SEG12493368HX, SEG12493369HX, SEG12493370HX, SEG12493371HX, SEG12493372HX, SEG12493373HX, SEG12493374HX, SEG12493375HX, SEG12493376HX, SEG12493377HX, SEG12493378HX, SEG12493379HX, SEG12493380HX, SEG12493381HX, SEG12493382HX, SEG12493383HX, SEG12493384HX, SEG12493385HX, SEG12493386HX, SEG12493387HX, SEG12493388HX, SEG12493389HX, SEG12493390HX, SEG12493391HX, SEG12493392HX, SEG12493393HX, SEG12493394HX, SEG12493395HX, SEG12493396HX, SEG12493397HX, SEG12493398HX, SEG12493399HX, SEG12493400HX, SEG12493401HX, SEG12493402HX, SEG12493403HX, SEG12493404HX, SEG12493405HX, SEG12493406HX, SEG12493407HX, SEG12493408HX, SEG12493409HX, SEG12493410HX, SEG12493411HX, SEG12493412HX, SEG12493413HX, SEG12493414HX, SEG12493415HX, SEG12493416HX, SEG12493417HX, SEG12493418HX, SEG12493619HX, SEG12493620HX, SEG12493621HX, SEG12493622HX, SEG12493623HX, SEG12493624HX, SEG12493625HX, SEG12493626HX, SEG12493627HX, SEG12493628HX, SEG12493629HX, SEG12493630HX, SEG12493631HX, SEG12493632HX, SEG12493633HX, SEG12493634HX, SEG12493635HX, SEG12493636HX, SEG12493637HX, SEG12493638HX, SEG12493639HX, SEG12493640HX, SEG12493641HX, SEG12493642HX, SEG12493643HX, SEG12493644HX, SEG12493645HX, SEG12493646HX, SEG12493647HX, SEG12493648HX, SEG12493649HX, SEG12493650HX, SEG12493651HX, SEG12493652HX, SEG12493653HX, SEG12493654HX, SEG12493655HX, SEG12493656HX, SEG12493657HX, SEG12493658HX, SEG12493659HX, SEG12493660HX, SEG12493661HX, SEG12493662HX, SEG12493663HX, SEG12493664HX, SEG12493665HX, SEG12493666HX, SEG12493667HX, SEG12493668HX, SEG12493669HX, SEG12493670HX, SEG12493671HX, SEG12493672HX, SEG12493673HX, SEG12493674HX, SEG12493675HX, SEG12493676HX, SEG12493677HX, SEG12493678HX, SEG12493679HX, SEG12493680HX, SEG12493681HX, SEG12493682HX, SEG12493683HX, SEG12493684HX, SEG12493685HX, SEG12493686HX, SEG12493687HX, SEG12493688HX, SEG12493689HX, SEG12493690HX, SEG12493691HX, SEG12493692HX, SEG12493693HX, SEG12493694HX, SEG12493695HX, SEG12493696HX, SEG12493697HX, SEG12493698HX, SEG12493699HX, SEG12493700HX, SEG12493701HX, SEG12493702HX, SEG12493703HX, SEG12493704HX, SEG12493705HX, SEG12493706HX, SEG12493707HX, SEG12493708HX, SEG12493709HX, SEG12493710HX, SEG12493711HX, SEG12493712HX, SEG12493713HX, SEG12493714HX, SEG12493715HX, SEG12493716HX, SEG12493717HX, SEG12493718HX, SEG12505076HX, SEG12505077HX, SEG12505078HX, SEG12505079HX, SEG12505080HX, SEG12505081HX, SEG12505082HX, SEG12505083HX, SEG12505084HX, SEG12505085HX, SEG12505086HX, SEG12505087HX, SEG12505088HX, SEG12505089HX, SEG12505090HX, SEG12505091HX, SEG12505092HX, SEG12505093HX, SEG12505094HX, SEG12505095HX, SEG12505096HX, SEG12505097HX, SEG12505098HX, SEG12505099HX, SEG12505100HX, SEG12505101HX, SEG12505102HX, SEG12505103HX, SEG12505104HX, SEG12505105HX, SEG12505106HX, SEG12505107HX, SEG12505108HX, SEG12505109HX, SEG12505110HX, SEG12505111HX, SEG12505112HX, SEG12505113HX, SEG12505114HX, SEG12505115HX, SEG12505116HX, SEG12505117HX, SEG12505118HX, SEG12505119HX, SEG12505120HX, SEG12505121HX, SEG12505122HX, SEG12505123HX, SEG12505124HX, SEG12505125HX, SEG12505126HX, SEG12505127HX, SEG12505128HX, SEG12505129HX, SEG12505130HX, SEG12505131HX, SEG12505132HX, SEG12505133HX, SEG12505134HX, SEG12505135HX, SEG12505136HX, SEG12505137HX, SEG12505138HX, SEG12505139HX, SEG12505140HX, SEG12505141HX, SEG12505142HX, SEG12505143HX, SEG12505144HX, SEG12505145HX, SEG12505146HX, SEG12505147HX, SEG12505148HX, SEG12505149HX, SEG12505150HX, SEG12505151HX, SEG12505152HX, SEG12505153HX, SEG12505154HX, SEG12505155HX, SEG12505156HX, SEG12505157HX, SEG12505158HX, SEG12505159HX, SEG12505160HX, SEG12505161HX, SEG12505162HX, SEG12505163HX, SEG12505164HX, SEG12505165HX, SEG12505166HX, SEG12505167HX, SEG12505168HX, SEG12505169HX, SEG12505170HX, SEG12505171HX, SEG12505172HX, SEG12505173HX, SEG12505174HX, SEG12505175HX, SEG12515938HX, SEG12515939HX, SEG12515940HX, SEG12515941HX, SEG12515942HX, SEG12515943HX, SEG12515944HX, SEG13016171HX, SEG13016172HX, SEG13016173HX, SEG13016174HX, SEG13016175HX, SEG13016176HX, SEG13016177HX, SEG13016178HX, SEG13016179HX, SEG13016180HX, SEG13016181HX, SEG13016185HX, SEG13016186HX, SEG13016187HX, SEG13016188HX, SEG13016189HX, SEG13016190HX, SEG13016191HX, SEG13016192HX, SEG13016195HX, SEG13016196HX, SEG13016197HX, SEG13016198HX, SEG13016199HX, SEG13016200HX, SEG13016201HX, SEG13016202HX, SEG13016203HX, SEG13016206HX, SEG13016207HX, SEG13016208HX, SEG13016209HX, SEG13016210HX, SEG13016211HX, SEG13016212HX, SEG13016213HX, SEG13016214HX, SEG13016217HX, SEG13016218HX, SEG13016219HX, SEG13016220HX, SEG13016221HX, SEG13016222HX, SEG13016223HX, SEG13016224HX, SEG13016225HX, SEG13016228HX, SEG13016229HX, SEG13016230HX, SEG13016231HX, SEG13016232HX, SEG13016233HX, SEG13016234HX, SEG13016235HX, SEG13016236HX, SEG13016237HX, SEG13016238HX, SEG13016239HX, SEG13016240HX, SEG13016241HX, SEG13026409HX, SEG13026410HX, SEG13026411HX, SEG13026412HX, SEG13026413HX, SEG13026414HX, SEG13026415HX, SEG13026416HX, SEG13026417HX, SEG13026418HX, SEG13026419HX, SEG13026420HX, SEG13026421HX, SEG13026422HX, SEG13026423HX, SEG13026424HX, SEG13026425HX, SEG13026426HX, SEG13026427HX, SEG13026428HX, SEG13026429HX, SEG13026430HX, SEG13026431HX, SEG13026432HX, SEG13026433HX, SEG13026434HX, SEG13026435HX, SEG13026436HX, SEG13026437HX, SEG13026438HX, SEG13026439HX, SEG13026440HX, SEG13026441HX, SEG13026442HX, SEG13026443HX, SEG13026444HX, SEG13026445HX, SEG13026446HX, SEG13026447HX, SEG13026448HX, SEG13026449HX, SEG13026450HX, SEG13026451HX, SEG13026452HX, SEG13026453HX, SEG13026454HX, SEG13026455HX, SEG13026456HX, SEG13026457HX, SEG13026458HX, SEG13026459HX, SEG13026460HX, SEG13026461HX, SEG13026462HX, SEG13026463HX, SEG13026464HX, SEG13026465HX, SEG13026466HX, SEG13026467HX, SEG13026468HX, SEG13026469HX, SEG13026470HX, SEG13026471HX, SEG13026472HX, SEG13026473HX, SEG13026474HX, SEG13026475HX, SEG13026476HX, SEG13026477HX, SEG13026478HX, SEG13026479HX, SEG13026480HX, SEG13026481HX, SEG13026482HX, SEG13026483HX, SEG13026484HX, SEG13026485HX, SEG13026486HX, SEG13026487HX, SEG13026488HX, SEG13026489HX, SEG13026490HX, SEG13026491HX, SEG13026492HX, SEG13026493HX, SEG13026494HX, SEG13026495HX, SEG13026496HX, SEG13026497HX, SEG13026498HX, SEG13026499HX, SEG13026500HX, SEG13026501HX, SEG13026502HX, SEG13026503HX, SEG13026504HX, SEG13026505HX, SEG13026506HX, SEG13026507HX, SEG13026508HX, SEG13026509HX, SEG13026510HX, SEG13026511HX, SEG13026512HX, SEG13026513HX, SEG13026514HX, SEG13026515HX, SEG13026516HX, SEG13026517HX, SEG13026518HX, SEG13026519HX, SEG13026520HX, SEG13026521HX, SEG13026522HX, SEG13026523HX, SEG13026524HX, SEG13026525HX, SEG13026526HX, SEG13026527HX, SEG13026528HX, SEG13037086HX, SEG13037087HX, SEG13037088HX, SEG13037089HX, SEG13037090HX, SEG13037091HX, SEG13037092HX, SEG13037093HX, SEG13037094HX, SEG13037095HX, SEG13037096HX, SEG13037097HX, SEG13037098HX, SEG13037099HX, SEG13037100HX, SEG13037101HX, SEG13037102HX, SEG13037103HX, SEG13037104HX, SEG13037105HX, SEG13037106HX, SEG13037107HX, SEG13037108HX, SEG13037109HX, SEG13037110HX, SEG13037111HX, SEG13037112HX, SEG13037113HX, SEG13037114HX, SEG13037115HX, SEG13037116HX, SEG13037117HX, SEG13037118HX, SEG13037119HX, SEG13037120HX, SEG13037121HX, SEG13037122HX, SEG13037123HX, SEG13037124HX, SEG13037125HX, SEG13037126HX, SEG13037127HX, SEG13037128HX, SEG13037129HX, SEG13037130HX, SEG13047529HX, SEG13047530HX, SEG13047531HX, SEG13047532HX, SEG13047533HX, SEG13047534HX, SEG13047535HX, SEG13047536HX, SEG13047537HX, SEG13047538HX, SEG13047539HX, SEG13047540HX, SEG13047541HX, SEG13047542HX, SEG13047543HX, SEG13047544HX, SEG13047545HX, SEG13047546HX, SEG13047547HX, SEG13047548HX, SEG13057834HX, SEG13057835HX, SEG13057836HX, SEG13057837HX, SEG13057838HX, SEG13057839HX, SEG13057840HX, SEG13057841HX, SEG13057842HX, SEG13057843HX, SEG13057844HX, SEG13057845HX, SEG13057846HX, SEG13057847HX, SEG13057848HX, SEG13057849HX, SEG13057850HX, SEG13057851HX, SEG13057852HX, SEG13057853HX, SEG13057854HX, SEG13057855HX, SEG13057856HX, SEG13057857HX, SEG13057858HX, SEG13057859HX, SEG13057860HX, SEG13057861HX, SEG13057862HX, SEG13057863HX, SEG13057864HX, SEG13057865HX, SEG13057866HX, SEG13057867HX, SEG13057868HX, SEG13057869HX, SEG13057870HX, SEG13057871HX, SEG13057872HX, SEG13057873HX, SEG13057874HX, SEG13057875HX, SEG13057876HX, SEG13057877HX, SEG13057878HX, SEG13057879HX, SEG13057880HX, SEG13057881HX, SEG13057882HX, SEG13057883HX, SEG13057889HX, SEG13057890HX, SEG13057891HX, SEG13057892HX, SEG13057893HX, SEG13057894HX, SEG13057895HX, SEG13057896HX, SEG13057897HX, SEG13057898HX, SEG13057899HX, SEG13057900HX, SEG13057901HX, SEG13057902HX, SEG13057903HX, SEG13057904HX, SEG13057905HX, SEG13057906HX, SEG13057907HX, SEG13057908HX, SEG13057909HX, SEG13057910HX, SEG13057911HX, SEG13057912HX, SEG13057913HX, SEG13057914HX, SEG13057915HX, SEG13057916HX, SEG13057917HX, SEG13057918HX, SEG13057919HX, SEG13057920HX, SEG13057921HX, SEG13057922HX, SEG13057923HX, SEG13057924HX, SEG13057925HX, SEG13057926HX, SEG13057927HX, SEG13057928HX, SEG13057929HX, SEG13057930HX, SEG13057931HX, SEG13057932HX, SEG13057933HX, SEG13057934HX, SEG13057935HX, SEG13057936HX, SEG13057937HX, SEG13057938HX, SEG13068379HX, SEG13068380HX, SEG13068381HX, SEG13068382HX, SEG13068383HX, SEG13068384HX, SEG13068385HX, SEG13068386HX, SEG13068387HX, SEG13068388HX, SEG13068389HX, SEG13068390HX, SEG13068391HX, SEG13068392HX, SEG13068393HX, SEG13068394HX, SEG13068395HX, SEG13068396HX, SEG13068397HX, SEG13068398HX, SEG13068399HX, SEG13068400HX, SEG13068401HX, SEG13068402HX, SEG13068403HX, SEG13068404HX, SEG13068405HX, SEG13068406HX, SEG13068407HX, SEG13068408HX, SEG13068409HX, SEG13068410HX, SEG13068411HX, SEG13068412HX, SEG13068413HX, SEG13068414HX, SEG13068415HX, SEG13068416HX, SEG13068417HX, SEG13068418HX, SEG13068419HX, SEG13068420HX, SEG13068421HX, SEG13068422HX, SEG13068423HX, SEG13068424HX, SEG13068425HX, SEG13068426HX, SEG13068427HX, SEG13068428HX, SEG13068429HX, SEG13068430HX, SEG13068431HX, SEG13068432HX, SEG13068433HX, SEG13068434HX, SEG13068522HX, SEG13068523HX, SEG13068524HX, SEG13068525HX, SEG13068526HX, SEG13068527HX, SEG13068528HX, SEG13068529HX, SEG13068530HX, SEG13068531HX, SEG13068532HX, SEG13068533HX, SEG13068534HX, SEG13068535HX, SEG13068536HX, SEG13068537HX, SEG13068538HX, SEG13068539HX, SEG13068540HX, SEG13068541HX, SEG13068542HX, SEG13068543HX, SEG13068544HX, SEG13068545HX, SEG13068546HX, SEG13068547HX, SEG13068548HX, SEG13068549HX, SEG13068550HX, SEG13068551HX, SEG13068552HX, SEG13068553HX, SEG13068554HX, SEG13068555HX, SEG13068556HX, SEG13068557HX, SEG13068558HX, SEG13068559HX, SEG13068560HX, SEG13068561HX, SEG13068562HX, SEG13068563HX, SEG13068564HX, SEG13068565HX, SEG13068566HX, SEG13068567HX, SEG13068568HX, SEG13068569HX, SEG13068570HX, SEG13068571HX, SEG13078682HX, SEG13078683HX, SEG13078684HX, SEG13078685HX, SEG13078686HX, SEG13078687HX, SEG13078688HX, SEG13078689HX, SEG13078690HX, SEG13078691HX, SEG13078692HX, SEG13078693HX, SEG13078694HX, SEG13078695HX, SEG13078696HX, SEG13078697HX, SEG13078698HX, SEG13078699HX, SEG13078700HX, SEG13078701HX, SEG13078702HX, SEG13078703HX, SEG13078704HX, SEG13078705HX, SEG13078706HX, SEG13078707HX, SEG13078708HX, SEG13078709HX, SEG13078710HX, SEG13078711HX, SEG13078712HX, SEG13078713HX, SEG13078714HX, SEG13078715HX, SEG13078716HX, SEG13078717HX, SEG13078718HX, SEG13078719HX, SEG13078720HX, SEG13078721HX, SEG13078722HX, SEG13078723HX, SEG13078724HX, SEG13078725HX, SEG13078726HX, SEG13078727HX, SEG13078728HX, SEG13078729HX, SEG13078730HX, SEG13078731HX, SEG13078732HX, SEG13078733HX, SEG13078734HX, SEG13078735HX, SEG13078736HX, SEG13078737HX, SEG13078738HX, SEG13078739HX, SEG13078740HX, SEG13078741HX, SEG13078742HX, SEG13078743HX, SEG13078744HX, SEG13078745HX, SEG13078746HX, SEG13078747HX, SEG13078748HX, SEG13078749HX, SEG13078750HX, SEG13078751HX, SEG13078752HX, SEG13078753HX, SEG13078754HX, SEG13078755HX, SEG13078756HX, SEG13078757HX, SEG13078758HX, SEG13078759HX, SEG13078760HX, SEG13078761HX, SEG13078762HX, SEG13078763HX, SEG13078764HX, SEG13078765HX, SEG13078766HX, SEG13078767HX, SEG13078768HX, SEG13078769HX, SEG13078770HX, SEG13078771HX, SEG13078772HX, SEG13078773HX, SEG13078774HX, SEG13078775HX, SEG13078776HX, SEG13078777HX, SEG13078778HX, SEG13078779HX, SEG13078780HX, SEG13078781HX, SEG13089252HX, SEG13089253HX, SEG13089254HX, SEG13089255HX, SEG13089256HX, SEG13089257HX, SEG13089258HX, SEG13089259HX, SEG13089260HX, SEG13089261HX, SEG13089262HX, SEG13089263HX, SEG13089264HX, SEG13089265HX, SEG13089266HX, SEG13089267HX, SEG13089268HX, SEG13089269HX, SEG13089270HX, SEG13089271HX, SEG13099413HX, SEG13099414HX, SEG13099415HX, SEG13099416HX, SEG13099417HX, SEG13099418HX, SEG13099419HX, SEG13099420HX, SEG13099421HX, SEG13099422HX, SEG13099423HX, SEG13099424HX, SEG13099425HX, SEG13099426HX, SEG13099427HX, SEG13099428HX, SEG13099429HX, SEG13099430HX, SEG13099431HX, SEG13099432HX, SEG13099433HX, SEG13099434HX, SEG13099435HX, SEG13099436HX, SEG13099437HX, SEG13099438HX, SEG13099439HX, SEG13099440HX, SEG13099441HX, SEG13099442HX, SEG13099443HX, SEG13099444HX, SEG13099445HX, SEG13099446HX, SEG13099447HX, SEG13099448HX, SEG13099449HX, SEG13099450HX, SEG13099451HX, SEG13099452HX, SEG13099470HX, SEG13099471HX, SEG13099472HX, SEG13099473HX, SEG13099474HX, SEG13099475HX, SEG13099476HX, SEG13099477HX, SEG13099478HX, SEG13099479HX, SEG13099480HX, SEG13099481HX, SEG13099482HX, SEG13099483HX, SEG13099484HX, SEG13099485HX, SEG13099486HX, SEG13099487HX, SEG13099488HX, SEG13099489HX, SEG13099490HX, SEG13099491HX, SEG13099492HX, SEG13099493HX, SEG13099494HX, SEG13099495HX, SEG13099496HX, SEG13099497HX, SEG13099498HX, SEG13099499HX, SEG13099500HX, SEG13099501HX, SEG13099502HX, SEG13099503HX, SEG13099504HX, SEG13099505HX, SEG13099506HX, SEG13099507HX, SEG13099508HX, SEG13099509HX, SEG13099510HX, SEG13099511HX, SEG13099512HX, SEG13099513HX, SEG13099514HX, SEG13099515HX, SEG13099516HX, SEG13099517HX, SEG13099518HX, SEG13099519HX, SEG13099690HX, SEG13099691HX, SEG13099692HX, SEG13099693HX, SEG13099694HX, SEG13099695HX, SEG13099696HX, SEG13099697HX, SEG13099698HX, SEG13099699HX, SEG13099700HX, SEG13099701HX, SEG13099702HX, SEG13099703HX, SEG13099704HX, SEG13099705HX, SEG13099706HX, SEG13099707HX, SEG13099708HX, SEG13099709HX, SEG13099710HX, SEG13099711HX, SEG13099712HX, SEG13099713HX, SEG13099714HX, SEG13099715HX, SEG13099716HX, SEG13099717HX, SEG13099718HX, SEG13099719HX, SEG13099720HX, SEG13099721HX, SEG13099722HX, SEG13099723HX, SEG13099724HX, SEG13099725HX, SEG13099726HX, SEG13099727HX, SEG13099728HX, SEG13099729HX, SEG13099730HX, SEG13099731HX, SEG13099732HX, SEG13099733HX, SEG13099734HX, SEG13099735HX, SEG13099736HX, SEG13099737HX, SEG13099738HX, SEG13099739HX, SEG13100062HX, SEG13100063HX, SEG13100064HX, SEG13100065HX, SEG13100066HX, SEG13100067HX, SEG13100068HX, SEG13100069HX, SEG13100070HX, SEG13100071HX, SEG13100072HX, SEG13100073HX, SEG13100074HX, SEG13100075HX, SEG13100076HX, SEG13100077HX, SEG13100078HX, SEG13100079HX, SEG13100080HX, SEG13100081HX, SEG13100082HX, SEG13100083HX, SEG13100084HX, SEG13100085HX, SEG13100086HX, SEG13100087HX, SEG13100088HX, SEG13100089HX, SEG13100090HX, SEG13100091HX, SEG13100092HX, SEG13100093HX, SEG13100094HX, SEG13100095HX, SEG13100096HX, SEG13100097HX, SEG13100098HX, SEG13100099HX, SEG13100100HX, SEG13100101HX, SEG13100102HX, SEG13100103HX, SEG13100104HX, SEG13100105HX, SEG13100106HX, SEG13100107HX, SEG13100108HX, SEG13100109HX, SEG13100110HX, SEG13100111HX, SEG13109937HX, SEG13109938HX, SEG13109939HX, SEG13109940HX, SEG13109941HX, SEG13109942HX, SEG13109943HX, SEG13109944HX, SEG13109945HX, SEG13109946HX, SEG13109947HX, SEG13109948HX, SEG13109949HX, SEG13109950HX, SEG13109951HX, SEG13109952HX, SEG13109953HX, SEG13109954HX, SEG13109955HX, SEG13109956HX, SEG13109957HX, SEG13109958HX, SEG13109959HX, SEG13109960HX, SEG13109961HX, SEG13109962HX, SEG13109963HX, SEG13109964HX, SEG13109965HX, SEG13109966HX, SEG13109967HX, SEG13109968HX, SEG13109969HX, SEG13109970HX, SEG13109971HX, SEG13109972HX, SEG13109973HX, SEG13109974HX, SEG13109975HX, SEG13109976HX, SEG13109977HX, SEG13109978HX, SEG13109979HX, SEG13109980HX, SEG13109981HX, SEG13109982HX, SEG13109983HX, SEG13109984HX, SEG13109985HX, SEG13109986HX, SEG13110321HX, SEG13110322HX, SEG13110323HX, SEG13110324HX, SEG13110325HX, SEG13110326HX, SEG13110327HX, SEG13110328HX, SEG13110329HX, SEG13110330HX, SEG13110331HX, SEG13110332HX, SEG13110333HX, SEG13110334HX, SEG13110335HX, SEG13110336HX, SEG13110337HX, SEG13110338HX, SEG13110339HX, SEG13110340HX, SEG13110341HX, SEG13110342HX, SEG13110343HX, SEG13110344HX, SEG13110345HX, SEG13110346HX, SEG13110347HX, SEG13110348HX, SEG13110349HX, SEG13110350HX, SEG13110351HX, SEG13110352HX, SEG13110353HX, SEG13110354HX, SEG13110355HX, SEG13110356HX, SEG13110357HX, SEG13110358HX, SEG13110359HX, SEG13110360HX, SEG13110361HX, SEG13110362HX, SEG13110363HX, SEG13110364HX, SEG13110365HX, SEG13110366HX, SEG13110367HX, SEG13110368HX, SEG13110369HX, SEG13110370HX, SEG13120593HX, SEG13120594HX, SEG13120595HX, SEG13120596HX, SEG13120597HX, SEG13120598HX, SEG13120599HX, SEG13120600HX, SEG13120601HX, SEG13120602HX, SEG13120603HX, SEG13120604HX, SEG13120605HX, SEG13120606HX, SEG13120607HX, SEG13120608HX, SEG13120609HX, SEG13120610HX, SEG13120611HX, SEG13120612HX, SEG13120613HX, SEG13120614HX, SEG13120615HX, SEG13120616HX, SEG13120617HX, SEG13120618HX, SEG13120619HX, SEG13120620HX, SEG13120621HX, SEG13120622HX, SEG13120623HX, SEG13120624HX, SEG13120625HX, SEG13120626HX, SEG13120627HX, SEG13120628HX, SEG13120629HX, SEG13120630HX, SEG13120631HX, SEG13120632HX, SEG13120633HX, SEG13120634HX, SEG13120635HX, SEG13120636HX, SEG13120637HX, SEG13120638HX, SEG13120639HX, SEG13120640HX, SEG13120641HX, SEG13120642HX, SEG13120881HX, SEG13120882HX, SEG13120883HX, SEG13120884HX, SEG13120885HX, SEG13120886HX, SEG13120887HX, SEG13120888HX, SEG13120889HX, SEG13120890HX, SEG13120891HX, SEG13120892HX, SEG13120893HX, SEG13120894HX, SEG13120895HX, SEG13120896HX, SEG13120897HX, SEG13120898HX, SEG13120899HX, SEG13120900HX, SEG13120901HX, SEG13120902HX, SEG13120903HX, SEG13120904HX, SEG13120905HX, SEG13120906HX, SEG13120907HX, SEG13120908HX, SEG13120909HX, SEG13120910HX, SEG13120911HX, SEG13120912HX, SEG13120913HX, SEG13120914HX, SEG13120915HX, SEG13120916HX, SEG13120917HX, SEG13120918HX, SEG13120919HX, SEG13120920HX, SEG13120921HX, SEG13120922HX, SEG13120923HX, SEG13120924HX, SEG13120925HX, SEG13120926HX, SEG13120927HX, SEG13120928HX, SEG13120929HX, SEG13120930HX, SEG13131026HX, SEG13131027HX, SEG13131028HX, SEG13131029HX, SEG13131030HX, SEG13131031HX, SEG13131032HX, SEG13131033HX, SEG13131034HX, SEG13131035HX, SEG13131036HX, SEG13131037HX, SEG13131038HX, SEG13131039HX, SEG13131040HX, SEG13131041HX, SEG13131042HX, SEG13131043HX, SEG13131044HX, SEG13131045HX, SEG13131046HX, SEG13131047HX, SEG13131048HX, SEG13131049HX, SEG13131050HX, SEG13131051HX, SEG13131052HX, SEG13131053HX, SEG13131054HX, SEG13131055HX, SEG13141399HX, SEG13141400HX, SEG13141401HX, SEG13141402HX, SEG13141403HX, SEG13141404HX, SEG13141405HX, SEG13141406HX, SEG13141407HX, SEG13141408HX, SEG13141409HX, SEG13141410HX, SEG13141411HX, SEG13141412HX, SEG13141413HX, SEG13141414HX, SEG13141415HX, SEG13141416HX, SEG13141417HX, SEG13141418HX, SEG13141419HX, SEG13141420HX, SEG13141421HX, SEG13141422HX, SEG13141423HX, SEG13141424HX, SEG13141425HX, SEG13141426HX, SEG13141427HX, SEG13141428HX, SEG13141429HX, SEG13141430HX, SEG13141431HX, SEG13141432HX, SEG13141433HX, SEG13141434HX, SEG13141435HX, SEG13141436HX, SEG13141437HX, SEG13141438HX, SEG13141439HX, SEG13141440HX, SEG13141441HX, SEG13141442HX, SEG13141443HX, SEG13141444HX, SEG13141445HX, SEG13141446HX, SEG13141447HX, SEG13141448HX, SEG13151637HX, SEG13151638HX, SEG13151639HX, SEG13151640HX, SEG13151641HX, SEG13151642HX, SEG13151643HX, SEG13151644HX, SEG13151645HX, SEG13151646HX, SEG13151647HX, SEG13151648HX, SEG13151649HX, SEG13151650HX, SEG13151651HX, SEG13151652HX, SEG13151653HX, SEG13151654HX, SEG13151655HX, SEG13151656HX, SEG13151657HX, SEG13151658HX, SEG13151659HX, SEG13151660HX, SEG13151661HX, SEG13151662HX, SEG13151663HX, SEG13151664HX, SEG13151665HX, SEG13151666HX, SEG13151667HX, SEG13151668HX, SEG13151669HX, SEG13151670HX, SEG13151671HX, SEG13151672HX, SEG13151673HX, SEG13151674HX, SEG13151675HX, SEG13151676HX, SEG13151677HX, SEG13151678HX, SEG13151679HX, SEG13151680HX, SEG13151681HX, SEG13151682HX, SEG13151683HX, SEG13151684HX, SEG13151685HX, SEG13151686HX, SEG13151788HX, SEG13151789HX, SEG13151790HX, SEG13151791HX, SEG13151792HX, SEG13151793HX, SEG13151794HX, SEG13151795HX, SEG13151796HX, SEG13151797HX, SEG13151798HX, SEG13151799HX, SEG13151800HX, SEG13151801HX, SEG13151802HX, SEG13151803HX, SEG13151804HX, SEG13151805HX, SEG13151806HX, SEG13151807HX, SEG13151808HX, SEG13151809HX, SEG13151810HX, SEG13151811HX, SEG13151812HX, SEG13151813HX, SEG13151814HX, SEG13151815HX, SEG13151816HX, SEG13151817HX, SEG13151818HX, SEG13151819HX, SEG13151820HX, SEG13151821HX, SEG13151822HX, SEG13151823HX, SEG13151824HX, SEG13151825HX, SEG13151826HX, SEG13151827HX, SEG13151828HX, SEG13151829HX, SEG13151830HX, SEG13151831HX, SEG13151832HX, SEG13151833HX, SEG13151834HX, SEG13151835HX, SEG13151836HX, SEG13151837HX, SEG13162218HX, SEG13162219HX, SEG13162220HX, SEG13162221HX, SEG13162222HX, SEG13162223HX, SEG13162224HX, SEG13162225HX, SEG13162226HX, SEG13162227HX, SEG13162228HX, SEG13162229HX, SEG13162230HX, SEG13162231HX, SEG13162232HX, SEG13162233HX, SEG13162234HX, SEG13162235HX, SEG13162236HX, SEG13162237HX, SEG13162238HX, SEG13162239HX, SEG13162240HX, SEG13162241HX, SEG13162242HX, SEG13162243HX, SEG13162244HX, SEG13162245HX, SEG13162246HX, SEG13162247HX, SEG13162248HX, SEG13162249HX, SEG13162250HX, SEG13162251HX, SEG13162252HX, SEG13162253HX, SEG13162254HX, SEG13162255HX, SEG13162256HX, SEG13162257HX, SEG13162258HX, SEG13162259HX, SEG13162260HX, SEG13162261HX, SEG13162262HX, SEG13162263HX, SEG13162264HX, SEG13162265HX, SEG13162266HX, SEG13162267HX, SEG13172501HX, SEG13172502HX, SEG13172503HX, SEG13172504HX, SEG13172505HX, SEG13172506HX, SEG13172507HX, SEG13172508HX, SEG13172509HX, SEG13172510HX, SEG13172511HX, SEG13172512HX, SEG13172513HX, SEG13172514HX, SEG13172515HX, SEG13172516HX, SEG13172517HX, SEG13172518HX, SEG13172519HX, SEG13172520HX, SEG13172521HX, SEG13172522HX, SEG13172523HX, SEG13172524HX, SEG13172525HX, SEG13172526HX, SEG13172527HX, SEG13172528HX, SEG13172529HX, SEG13172530HX, SEG13172531HX, SEG13172532HX, SEG13172533HX, SEG13172534HX, SEG13172535HX, SEG13172536HX, SEG13172537HX, SEG13172538HX, SEG13172539HX, SEG13172540HX, SEG13172541HX, SEG13172542HX, SEG13172543HX, SEG13172544HX, SEG13172545HX, SEG13172546HX, SEG13172548HX, SEG13172550HX, SEG13182907HX, SEG13182908HX, SEG13182909HX, SEG13182910HX, SEG13182911HX, SEG13182912HX, SEG13182913HX, SEG13182914HX, SEG13182915HX, SEG13182916HX, SEG13182917HX, SEG13182918HX, SEG13182919HX, SEG13182920HX, SEG13182921HX, SEG13182922HX, SEG13182923HX, SEG13182924HX, SEG13182925HX, SEG13182926HX, SEG13182927HX, SEG13182928HX, SEG13182929HX, SEG13182930HX, SEG13182931HX, SEG13182932HX, SEG13182933HX, SEG13182934HX, SEG13182935HX, SEG13182936HX, SEG13182937HX, SEG13182938HX, SEG13182939HX, SEG13182940HX, SEG13182941HX, SEG13182942HX, SEG13182943HX, SEG13182944HX, SEG13182945HX, SEG13182946HX, SEG13193110HX, SEG13193111HX, SEG13193112HX, SEG13193113HX, SEG13193114HX, SEG13193115HX, SEG13193116HX, SEG13193117HX, SEG13193118HX, SEG13193119HX, SEG13193120HX, SEG13193121HX, SEG13193122HX, SEG13193123HX, SEG13193124HX, SEG13193125HX, SEG13193126HX, SEG13193127HX, SEG13193128HX, SEG13193129HX, SEG13193130HX, SEG13193131HX, SEG13193132HX, SEG13193133HX, SEG13193134HX, SEG13193135HX, SEG13193136HX, SEG13193137HX, SEG13193138HX, SEG13193139HX, SEG13193140HX, SEG13193141HX, SEG13193142HX, SEG13193143HX, SEG13193144HX, SEG13193145HX, SEG13193146HX, SEG13193147HX, SEG13193148HX, SEG13193149HX, SEG13193150HX, SEG13193151HX, SEG13193152HX, SEG13193153HX, SEG13193154HX, SEG13193155HX, SEG13193156HX, SEG13193157HX, SEG13193158HX, SEG13193159HX, SEG13193165HX, SEG13193166HX, SEG13193167HX, SEG13193168HX, SEG13193169HX, SEG13193170HX, SEG13193171HX, SEG13193172HX, SEG13193173HX, SEG13193174HX, SEG13193175HX, SEG13193176HX, SEG13193177HX, SEG13193178HX, SEG13193179HX, SEG13193180HX, SEG13193181HX, SEG13193182HX, SEG13193183HX, SEG13193184HX, SEG13193185HX, SEG13193186HX, SEG13193187HX, SEG13193188HX, SEG13193189HX, SEG13193190HX, SEG13193191HX, SEG13193192HX, SEG13193193HX, SEG13193194HX, SEG13193195HX, SEG13193196HX, SEG13193197HX, SEG13193198HX, SEG13193199HX, SEG13193200HX, SEG13193201HX, SEG13193202HX, SEG13193203HX, SEG13193204HX, SEG13193205HX, SEG13193206HX, SEG13193207HX, SEG13193208HX, SEG13193209HX, SEG13213518HX, SEG13213519HX, SEG13213520HX, SEG13213521HX, SEG13213522HX, SEG13213523HX, SEG13213524HX, SEG13213525HX, SEG13213526HX, SEG13213527HX, SEG13213528HX, SEG13213529HX, SEG13213530HX, SEG13213531HX, SEG13213532HX, SEG13213533HX, SEG13213534HX, SEG13213535HX, SEG13213536HX, SEG13213537HX, SEG13213538HX, SEG13213539HX, SEG13213540HX, SEG13213541HX, SEG13213542HX, SEG13213543HX, SEG13213544HX, SEG13213545HX, SEG13213546HX, SEG13213547HX, SEG13213548HX, SEG13213549HX, SEG13213550HX, SEG13213551HX, SEG13213552HX, SEG13213553HX, SEG13213554HX, SEG13213555HX, SEG13213556HX, SEG13213557HX, SEG13213558HX, SEG13213559HX, SEG13213560HX, SEG13213561HX, SEG13213562HX, SEG13213563HX, SEG13213564HX, SEG13213565HX, SEG13213566HX, SEG13213567HX, SEG13213830HX, SEG13213831HX, SEG13213832HX, SEG13213833HX, SEG13213834HX, SEG13213835HX, SEG13213836HX, SEG13213837HX, SEG13213838HX, SEG13213839HX, SEG13213840HX, SEG13213841HX, SEG13223869HX, SEG13223870HX, SEG13223871HX, SEG13223872HX, SEG13223873HX, SEG13223874HX, SEG13223875HX, SEG13223876HX, SEG13223877HX, SEG13223878HX, SEG13223879HX, SEG13223880HX, SEG13223881HX, SEG13223882HX, SEG13223883HX, SEG13223884HX, SEG13223885HX, SEG13223886HX, SEG13223887HX, SEG13223888HX, SEG13223889HX, SEG13223890HX, SEG13223891HX, SEG13223892HX, SEG13223893HX, SEG13223894HX, SEG13223895HX, SEG13223896HX, SEG13223897HX, SEG13223898HX, SEG13223899HX, SEG13223900HX, SEG13223901HX, SEG13223902HX, SEG13223903HX, SEG13223904HX, SEG13223905HX, SEG13223906HX, SEG13223907HX, SEG13223908HX, SEG13223909HX, SEG13223910HX, SEG13223911HX, SEG13223912HX, SEG13223913HX, SEG13223914HX, SEG13223915HX, SEG13223916HX, SEG13223917HX, SEG13223918HX, SEG13234041HX, SEG13234042HX, SEG13234043HX, SEG13234044HX, SEG13234045HX, SEG13234046HX, SEG13234047HX, SEG13234048HX, SEG13234049HX, SEG13234050HX, SEG13234051HX, SEG13234052HX, SEG13234053HX, SEG13234054HX, SEG13234055HX, SEG13234056HX, SEG13234057HX, SEG13234058HX, SEG13234059HX, SEG13234060HX, SEG13234061HX, SEG13234062HX, SEG13234063HX, SEG13234064HX, SEG13234065HX, SEG13234066HX, SEG13234067HX, SEG13234068HX, SEG13234069HX, SEG13234070HX, SEG13234071HX, SEG13234072HX, SEG13234073HX, SEG13234074HX, SEG13234075HX, SEG13234076HX, SEG13234077HX, SEG13234078HX, SEG13234079HX, SEG13234080HX, SEG13234081HX, SEG13234082HX, SEG13234083HX, SEG13234084HX, SEG13234085HX, SEG13234086HX, SEG13234087HX, SEG13234088HX, SEG13234089HX, SEG13234090HX, SEG13234284HX, SEG13234285HX, SEG13234286HX, SEG13234287HX, SEG13234288HX, SEG13234289HX, SEG13234290HX, SEG13234291HX, SEG13234292HX, SEG13234293HX, SEG13234294HX, SEG13234295HX, SEG13234296HX, SEG13234297HX, SEG13234298HX, SEG13234299HX, SEG13234300HX, SEG13234301HX, SEG13234302HX, SEG13234303HX, SEG13234304HX, SEG13234305HX, SEG13234306HX, SEG13234307HX, SEG13234308HX, SEG13234309HX, SEG13234310HX, SEG13234311HX, SEG13234312HX, SEG13234313HX, SEG13234314HX, SEG13234315HX, SEG13234316HX, SEG13234317HX, SEG13234318HX, SEG13234319HX, SEG13234320HX, SEG13234321HX, SEG13234322HX, SEG13234323HX, SEG13234324HX, SEG13234325HX, SEG13234326HX, SEG13244854HX, SEG13244855HX, SEG13244856HX, SEG13244857HX, SEG13244858HX, SEG13244859HX, SEG13244860HX, SEG13244861HX, SEG13244862HX, SEG13244863HX, SEG13244864HX, SEG13244865HX, SEG13244866HX, SEG13244867HX, SEG13244868HX, SEG13244869HX, SEG13244870HX, SEG13244871HX, SEG13244872HX, SEG13244873HX, SEG13265455HX, SEG13265456HX, SEG13265457HX, SEG13265458HX, SEG13265459HX, SEG13265460HX, SEG13265461HX, SEG13265462HX, SEG13265463HX, SEG13265464HX, SEG13265465HX, SEG13265466HX, SEG13265467HX, SEG13265468HX, SEG13265469HX, SEG13265470HX, SEG13265471HX, SEG13265472HX, SEG13265473HX, SEG13265474HX, SEG13265475HX, SEG13265476HX, SEG13265477HX, SEG13265478HX, SEG13265479HX, SEG13265480HX, SEG13265481HX, SEG13265482HX, SEG13265483HX, SEG13265484HX, SEG13265485HX, SEG13265486HX, SEG13265487HX, SEG13265488HX, SEG13265489HX, SEG13265490HX, SEG13265491HX, SEG13265492HX, SEG13265493HX, SEG13265494HX, SEG13265495HX, SEG13265496HX, SEG13265497HX, SEG13265498HX, SEG13265499HX, SEG13265500HX, SEG13265501HX, SEG13265502HX, SEG13265503HX, SEG13265595HX, SEG13265596HX, SEG13265597HX, SEG13265598HX, SEG13265599HX, SEG13265600HX, SEG13265601HX, SEG13265602HX, SEG13265603HX, SEG13265604HX, SEG13265605HX, SEG13265606HX, SEG13265607HX, SEG13265608HX, SEG13265609HX, SEG13265610HX, SEG13265611HX, SEG13265612HX, SEG13265613HX, SEG13265614HX, SEG13265615HX, SEG13265616HX, SEG13265617HX, SEG13265618HX, SEG13265619HX, SEG13265620HX, SEG13265621HX, SEG13265622HX, SEG13265623HX, SEG13265624HX, SEG13265625HX, SEG13265626HX, SEG13265627HX, SEG13265628HX, SEG13265629HX, SEG13265630HX, SEG13265631HX, SEG13265632HX, SEG13265633HX, SEG13265634HX, SEG13265635HX, SEG13265636HX, SEG13265637HX, SEG13265638HX, SEG13265639HX, SEG13265640HX, SEG13265641HX, SEG13265642HX, SEG13265643HX, SEG13265644HX, SEG13275809HX, SEG13275810HX, SEG13275811HX, SEG13275812HX, SEG13275813HX, SEG13275814HX, SEG13275815HX, SEG13275816HX, SEG13275817HX, SEG13275818HX, SEG13275819HX, SEG13275820HX, SEG13275821HX, SEG13275822HX, SEG13275823HX, SEG13275824HX, SEG13275825HX, SEG13275826HX, SEG13275827HX, SEG13275828HX, SEG13275829HX, SEG13275830HX, SEG13275831HX, SEG13275832HX, SEG13275833HX, SEG13275834HX, SEG13275835HX, SEG13275836HX, SEG13275837HX, SEG13275838HX, SEG13275839HX, SEG13275840HX, SEG13275841HX, SEG13275842HX, SEG13275843HX, SEG13275844HX, SEG13275845HX, SEG13275846HX, SEG13275847HX, SEG13275848HX, SEG13275849HX, SEG13275850HX, SEG13275851HX, SEG13275852HX, SEG13275853HX, SEG13275854HX, SEG13275855HX, SEG13275856HX, SEG13275857HX, SEG13275858HX, SEG13275859HX, SEG13275860HX, SEG13275861HX, SEG13275862HX, SEG13275863HX, SEG13275864HX, SEG13275865HX, SEG13275866HX, SEG13275867HX, SEG13275868HX, SEG13275869HX, SEG13275870HX, SEG13275871HX, SEG13275872HX, SEG13275873HX, SEG13275874HX, SEG13275875HX, SEG13275876HX, SEG13275877HX, SEG13275878HX, SEG13275879HX, SEG13275880HX, SEG13275881HX, SEG13275882HX, SEG13275883HX, SEG13275884HX, SEG13275885HX, SEG13275886HX, SEG13275887HX, SEG13275888HX, SEG13275889HX, SEG13275890HX, SEG13275891HX, SEG13275892HX, SEG13275893HX, SEG13275894HX, SEG13275895HX, SEG13275896HX, SEG13275897HX, SEG13275898HX, SEG13275899HX, SEG13275900HX, SEG13275901HX, SEG13275902HX, SEG13275903HX, SEG13275904HX, SEG13275905HX, SEG13275906HX, SEG13275907HX, SEG13275908HX, SEG13286114HX, SEG13286115HX, SEG13286116HX, SEG13286117HX, SEG13286118HX, SEG13286119HX, SEG13286120HX, SEG13286121HX, SEG13286122HX, SEG13286123HX, SEG13286124HX, SEG13286125HX, SEG13286126HX, SEG13286127HX, SEG13286128HX, SEG13286129HX, SEG13286130HX, SEG13286131HX, SEG13286132HX, SEG13286133HX, SEG13286134HX, SEG13286135HX, SEG13286136HX, SEG13286137HX, SEG13286138HX, SEG13286139HX, SEG13286140HX, SEG13286141HX, SEG13286142HX, SEG13286143HX, SEG13286144HX, SEG13286146HX, SEG13286147HX, SEG13286148HX, SEG13286149HX, SEG13286150HX, SEG13286152HX, SEG13286153HX, SEG13286154HX, SEG13286155HX, SEG13286156HX, SEG13286157HX, SEG13286158HX, SEG13286159HX, SEG13286160HX, SEG13286161HX, SEG13286162HX, SEG13286163HX, SEG13286342HX, SEG13286343HX, SEG13286344HX, SEG13286345HX, SEG13286346HX, SEG13286347HX, SEG13286348HX, SEG13286349HX, SEG13286350HX, SEG13286351HX, SEG13286352HX, SEG13286353HX, SEG13286354HX, SEG13286355HX, SEG13286356HX, SEG13286357HX, SEG13286358HX, SEG13286359HX, SEG13286360HX, SEG13286361HX, SEG13286362HX, SEG13286363HX, SEG13286364HX, SEG13286365HX, SEG13286366HX, SEG13286367HX, SEG13286368HX, SEG13286369HX, SEG13286370HX, SEG13286371HX, SEG13286372HX, SEG13286373HX, SEG13286374HX, SEG13286375HX, SEG13286376HX, SEG13286377HX, SEG13286378HX, SEG13286379HX, SEG13286380HX, SEG13286381HX, SEG13286382HX, SEG13286383HX, SEG13286384HX, SEG13286385HX, SEG13286386HX, SEG13286387HX, SEG13286388HX, SEG13286389HX, SEG13286390HX, SEG13286391HX, SEG13286505HX, SEG13286506HX, SEG13286507HX, SEG13286508HX, SEG13286509HX, SEG13286510HX, SEG13286511HX, SEG13286512HX, SEG13286513HX, SEG13286514HX, SEG13286515HX, SEG13286516HX, SEG13286517HX, SEG13286518HX, SEG13286519HX, SEG13286520HX, SEG13286521HX, SEG13286522HX, SEG13286523HX, SEG13286524HX, SEG13286525HX, SEG13286526HX, SEG13286527HX, SEG13286528HX, SEG13286529HX, SEG13286530HX, SEG13286531HX, SEG13286532HX, SEG13286533HX, SEG13286534HX, SEG13286535HX, SEG13286536HX, SEG13286537HX, SEG13286538HX, SEG13286539HX, SEG13286540HX, SEG13286541HX, SEG13286542HX, SEG13286543HX, SEG13286544HX, SEG13286545HX, SEG13286546HX, SEG13286547HX, SEG13286548HX, SEG13286549HX, SEG13286550HX, SEG13286551HX, SEG13286552HX, SEG13286553HX, SEG13286554HX, SEG13286555HX, SEG13286556HX, SEG13286557HX, SEG13286558HX, SEG13286559HX, SEG13286560HX, SEG13286561HX, SEG13286562HX, SEG13286563HX, SEG13286564HX, SEG13286565HX, SEG13286566HX, SEG13286567HX, SEG13286568HX, SEG13286569HX, SEG13286570HX, SEG13286571HX, SEG13286572HX, SEG13286573HX, SEG13286574HX, SEG13286575HX, SEG13286576HX, SEG13286577HX, SEG13286578HX, SEG13286579HX, SEG13306990HX, SEG13306991HX, SEG13306992HX, SEG13306993HX, SEG13306994HX, SEG13306995HX, SEG13306996HX, SEG13306997HX, SEG13306998HX, SEG13306999HX, SEG13307000HX, SEG13307001HX, SEG13307002HX, SEG13307003HX, SEG13307004HX, SEG13307005HX, SEG13307006HX, SEG13307007HX, SEG13307008HX, SEG13307009HX, SEG13307010HX, SEG13307011HX, SEG13307012HX, SEG13307013HX, SEG13307014HX, SEG13307015HX, SEG13307016HX, SEG13307017HX, SEG13307018HX, SEG13307019HX, SEG13307020HX, SEG13307021HX, SEG13307022HX, SEG13307023HX, SEG13307024HX, SEG13307025HX, SEG13307026HX, SEG13307027HX, SEG13307028HX, SEG13307029HX, SEG13307030HX, SEG13307031HX, SEG13307032HX, SEG13307033HX, SEG13307034HX, SEG13307035HX, SEG13307036HX, SEG13307037HX, SEG13307038HX, SEG13307039HX, SEG13317240HX, SEG13317254HX, SEG13317255HX, SEG13317256HX, SEG13317257HX, SEG13317258HX, SEG13317259HX, SEG13317260HX, SEG13317261HX, SEG13317262HX, SEG13317263HX, SEG13317264HX, SEG13317265HX, SEG13317266HX, SEG13317267HX, SEG13317268HX, SEG13317269HX, SEG13317270HX, SEG13317271HX, SEG13317272HX, SEG13317273HX, SEG13317274HX, SEG13317275HX, SEG13317276HX, SEG13317277HX, SEG13317278HX, SEG13317279HX, SEG13317280HX, SEG13317281HX, SEG13317282HX, SEG13317283HX, SEG13317284HX, SEG13317285HX, SEG13317286HX, SEG13317287HX, SEG13317288HX, SEG13317289HX, SEG13317290HX, SEG13317291HX, SEG13317292HX, SEG13317293HX, SEG13317294HX, SEG13317295HX, SEG13317296HX, SEG13317297HX, SEG13317298HX, SEG13317299HX, SEG13317300HX, SEG13317301HX, SEG13317302HX, SEG13317303HX, SEG13317304HX, SEG13317305HX, SEG13317306HX, SEG13317307HX, SEG13317308HX, SEG13317309HX, SEG13317310HX, SEG13317311HX, SEG13317312HX, SEG13317313HX, SEG13317314HX, SEG13317315HX, SEG13317316HX, SEG13317317HX, SEG13317318HX, SEG13317319HX, SEG13317320HX, SEG13317321HX, SEG13317322HX, SEG13317323HX, SEG13317324HX, SEG13317325HX, SEG13317326HX, SEG13317327HX, SEG13317328HX, SEG13317329HX, SEG13317330HX, SEG13317331HX, SEG13317332HX, SEG13317333HX, SEG13317334HX, SEG13317335HX, SEG13317336HX, SEG13317337HX, SEG13317338HX, SEG13317339HX, SEG13317340HX, SEG13317341HX, SEG13317342HX, SEG13317343HX, SEG13317344HX, SEG13317345HX, SEG13317346HX, SEG13317347HX, SEG13317348HX, SEG13317349HX, SEG13317350HX, SEG13317351HX, SEG13317352HX, SEG13317353HX, SEG13327841HX, SEG13327842HX, SEG13327843HX, SEG13327844HX, SEG13327845HX, SEG13327846HX, SEG13327847HX, SEG13327848HX, SEG13327849HX, SEG13327850HX, SEG13327851HX, SEG13327852HX, SEG13327853HX, SEG13327854HX, SEG13327855HX, SEG13327856HX, SEG13327857HX, SEG13327858HX, SEG13327859HX, SEG13327860HX, SEG13327861HX, SEG13327862HX, SEG13327863HX, SEG13327864HX, SEG13327865HX, SEG13327866HX, SEG13327867HX, SEG13327868HX, SEG13327869HX, SEG13327870HX, SEG13327871HX, SEG13327872HX, SEG13327873HX, SEG13327874HX, SEG13327875HX, SEG13327876HX, SEG13327877HX, SEG13327878HX, SEG13327879HX, SEG13327880HX, SEG13327881HX, SEG13327882HX, SEG13327883HX, SEG13327884HX, SEG13327885HX, SEG13327886HX, SEG13327887HX, SEG13327888HX, SEG13327889HX, SEG13327890HX, SEG13327891HX, SEG13327892HX, SEG13327893HX, SEG13327894HX, SEG13327895HX, SEG13327896HX, SEG13327897HX, SEG13327898HX, SEG13327899HX, SEG13327900HX, SEG13327901HX, SEG13327902HX, SEG13327903HX, SEG13327904HX, SEG13327905HX, SEG13327906HX, SEG13327907HX, SEG13327908HX, SEG13327909HX, SEG13327910HX, SEG13327911HX, SEG13327912HX, SEG13327913HX, SEG13327914HX, SEG13327915HX, SEG13327916HX, SEG13327917HX, SEG13327918HX, SEG13327919HX, SEG13327920HX, SEG13327921HX, SEG13327922HX, SEG13327923HX, SEG13327924HX, SEG13327925HX, SEG13327926HX, SEG13327927HX, SEG13327928HX, SEG13327929HX, SEG13327930HX, SEG13327931HX, SEG13327932HX, SEG13327933HX, SEG13327934HX, SEG13327935HX, SEG13327936HX, SEG13327937HX, SEG13327938HX, SEG13327939HX, SEG13327940HX, SEG13338377HX, SEG13338378HX, SEG13338379HX, SEG13338380HX, SEG13338381HX, SEG13338382HX, SEG13338383HX, SEG13338384HX, SEG13338385HX, SEG13338386HX, SEG13338387HX, SEG13338388HX, SEG13338389HX, SEG13338390HX, SEG13338391HX, SEG13338392HX, SEG13338393HX, SEG13338394HX, SEG13338395HX, SEG13338396HX, SEG13338397HX, SEG13338398HX, SEG13338399HX, SEG13338400HX, SEG13338401HX, SEG13338402HX, SEG13338403HX, SEG13338404HX, SEG13338405HX, SEG13338406HX, SEG13338407HX, SEG13338408HX, SEG13338409HX, SEG13338410HX, SEG13338411HX, SEG13338412HX, SEG13338413HX, SEG13338414HX, SEG13338415HX, SEG13338416HX, SEG13338417HX, SEG13338418HX, SEG13338419HX, SEG13338420HX, SEG13338421HX, SEG13338422HX, SEG13338423HX, SEG13338424HX, SEG13338425HX, SEG13338426HX, SEG13348636HX, SEG13348637HX, SEG13348638HX, SEG13348639HX, SEG13348640HX, SEG13348641HX, SEG13348642HX, SEG13348643HX, SEG13348644HX, SEG13348645HX, SEG13348646HX, SEG13348647HX, SEG13348648HX, SEG13348649HX, SEG13348650HX, SEG13348651HX, SEG13348652HX, SEG13348653HX, SEG13348654HX, SEG13348655HX, SEG13348656HX, SEG13348657HX, SEG13348658HX, SEG13348659HX, SEG13348660HX, SEG13348661HX, SEG13348662HX, SEG13348663HX, SEG13348664HX, SEG13348665HX, SEG13348666HX, SEG13348667HX, SEG13348668HX, SEG13348669HX, SEG13348670HX, SEG13348671HX, SEG13348672HX, SEG13348673HX, SEG13348674HX, SEG13348675HX, SEG13348676HX, SEG13348677HX, SEG13348678HX, SEG13348679HX, SEG13348680HX, SEG13348681HX, SEG13348682HX, SEG13348683HX, SEG13348684HX, SEG13348685HX, SEG13348686HX, SEG13348687HX, SEG13348688HX, SEG13348689HX, SEG13348690HX, SEG13348691HX, SEG13348692HX, SEG13348693HX, SEG13348694HX, SEG13348695HX, SEG13348696HX, SEG13348697HX, SEG13348698HX, SEG13348699HX, SEG13348700HX, SEG13348701HX, SEG13348702HX, SEG13348703HX, SEG13348704HX, SEG13348705HX, SEG13348706HX, SEG13348707HX, SEG13348708HX, SEG13348709HX, SEG13348710HX, SEG13348726HX, SEG13348727HX, SEG13348728HX, SEG13348729HX, SEG13348730HX, SEG13348731HX, SEG13348732HX, SEG13348733HX, SEG13348734HX, SEG13348735HX, SEG13369004HX, SEG13369005HX, SEG13369006HX, SEG13369007HX, SEG13369008HX, SEG13369009HX, SEG13369010HX, SEG13369011HX, SEG13369012HX, SEG13369013HX, SEG13369014HX, SEG13369015HX, SEG13369016HX, SEG13369017HX, SEG13369018HX, SEG13369019HX, SEG13369020HX, SEG13369021HX, SEG13369022HX, SEG13369023HX, SEG13369024HX, SEG13369025HX, SEG13369026HX, SEG13369027HX, SEG13369028HX, SEG13369029HX, SEG13369030HX, SEG13369031HX, SEG13369032HX, SEG13369033HX, SEG13369034HX, SEG13369035HX, SEG13369036HX, SEG13369037HX, SEG13369038HX, SEG13369039HX, SEG13369040HX, SEG13369041HX, SEG13369042HX, SEG13369043HX, SEG13369044HX, SEG13369045HX, SEG13369046HX, SEG13369047HX, SEG13369048HX, SEG13369049HX, SEG13369050HX, SEG13369051HX, SEG13379541HX, SEG13379542HX, SEG13379543HX, SEG13379544HX, SEG13379545HX, SEG13379546HX, SEG13379547HX, SEG13379568HX, SEG13379571HX, SEG13379578HX, SEG13379579HX, SEG13379580HX, SEG13379581HX, SEG13379582HX, SEG13379583HX, SEG13379584HX, SEG13379585HX, SEG13379586HX, SEG13379587HX. |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
Technical Support 262-513-4122
|
Manufacturer Reason for Recall |
There is a potential for communication loss associated with the F5-01 Frame when connected to CARESCAPE Patient Data Module (PDM) in the CARESCAPE Monitor B850. The PDM patient parameter acquisition module can sometimes spontaneously lose the communication with the host patient monitor when attached to the F5-01 Frame due to an issue with the PDM Ethernet communication hardware. The communicatio
|
FDA Determined Cause 2 |
Component change control |
Action |
GE Healthcare sent an "Urgent Medical Device Correction" letter dated December 20, 2013 to all affected customers. The letter was addressed to Healthcare Administrator / Risk Manager, Chief of Nursing and Director of Biomedical Engineering. The letter addressed the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information (Call Technical Support at 1-800-558-7044 or your local Service Representative). |
Quantity in Commerce |
12, 945 units (6,239 US, 6,706 OUS) |
Distribution |
Worldwide Distribution - US Nationwide including Puerto Rico and the countries of: Australia, Austria, Azerbaijan, Bahrain, Belgium, BRAZIL, Canada, CHILE, COSTA RICA, Cyprus, Czech Republic, Denmark, ECUADOR, Egypt, Estonia, Finland, France, Germany, Greece, GUAM, India, INDONESIA, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Korea, Republic of, Kuwait, Lithuania, Malaysia, Malta, Mexico, NEPAL, Netherlands, New Zealand, Norway, Pakistan, PANAMA, PARAGUAY, PERU, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, United Republic of, Turkey, United Arab Emirates, United Kingdom, URUGUAY. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MHX and Original Applicant = GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES
|
|
|
|